Code of Colorado Regulations
1000 - Department of Public Health and Environment
1007 - Hazardous Materials and Waste Management Division
6 CCR 1007-1 - RULES AND REGULATIONS PERTAINING TO THE ADMINISTRATION AND ENFORCEMENT OF THE PESTICIDE APPLICATORS' ACT
Part 24 - RADIATION CONTROL - PARTICLE ACCELERATORS AND THERAPEUTIC RADIATION MACHINES IN THE HEALING ARTS
SPECIFIC REQUIREMENTS
Section 6 CCR 1007-1-24.11 - Possession of Survey Instrument(s)

Current through Register Vol. 48, No. 6, March 25, 2025

24.11.1 Each facility location authorized to use a therapeutic radiation machine in accordance with 24.7 or 24.8 shall possess appropriately calibrated portable radiation monitoring equipment, including:

24.11.1.1 At least one portable radiation measurement survey instrument that is:
(1) Capable of measuring dose rates over the range 10 µSv (1 mrem) per hour to 10 mSv (1000 mrem) per hour;

(2) Operable; and

(3) Calibrated in accordance with 24.12.

Disclaimer: These regulations may not be the most recent version. Colorado may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.