Section 6 CCR 1007-1-24.10 - Quality Assurance for Image Guided Radiation Therapy
Universal Citation: 1000 CO Code Regs 6 CCR 1007-1-24.10
Current through Register Vol. 48, No. 6, March 25, 2025
24.10.1 Quality assurance for a radiographic/fluoroscopic IGRT system shall be performed by or under the direct supervision of a Registered Medical Physicist, and shall include acceptance testing and periodic verification of system performance in accordance with manufacturer's specifications and the AAPM Task Group 179 report unless the Registered Medical Physicist determines, and documents in writing, that a particular recommendation of these reports is not warranted for the clinical tasks for which the equipment will be used.
Disclaimer: These regulations may not be the most recent version. Colorado may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.