6 CCR 1007-1 - RULES AND REGULATIONS PERTAINING TO THE ADMINISTRATION AND ENFORCEMENT OF THE PESTICIDE APPLICATORS' ACT
Part 06 - RADIATION CONTROL - X-RAY IMAGING IN THE HEALING ARTS
Appendix 6 CCR 1007-1-6-I - TABLE OF HALF VALUE LAYERS FOR A SPECIFIED kVp AND SYSTEM

Current through Register Vol. 48, No. 6, March 25, 2025

X-Ray Tube Voltage (kilovolt peak)

Design

Operating

Range

Measured Operating Potential

Minimum HVL (mm in Aluminum)

Specified Dental Systems \1\

Other X-Ray Systems \2\

Other X-Ray Systems \3\

Below 51

30

1.5

0.3

0.3

40

1.5

0.4

0.4

50

1.5

0.5

0.5

51 to 70

51

1.5

1.2

1.3

60

1.5

1.3

1.5

70

1.5

1.5

1.8

Above 70

71

2.1

2.1

2.5

80

2.3

2.3

2.9

90

2.5

2.5

3.2

100

2.7

2.7

3.6

110

3.0

3.0

3.9

120

3.2

3.2

4.3

130

3.5

3.5

4.7

140

3.8

3.8

5.0

150

4.1

4.1

5.4

\1\ Dental x-ray systems designed for use with intraoral image receptors and manufactured after December 1, 1980.

\2\ Dental x-ray systems designed for use with intraoral image receptors and manufactured before or on December 1, 1980, and all other x-ray systems subject to this section and manufactured before June 10, 2006.

\3\ All x-ray systems, except dental x-ray systems designed for use with intraoral image receptors, subject to this section and manufactured on or after June 10, 2006.

Disclaimer: These regulations may not be the most recent version. Colorado may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.