Code of Colorado Regulations
1000 - Department of Public Health and Environment
1007 - Hazardous Materials and Waste Management Division
6 CCR 1007-1 - RULES AND REGULATIONS PERTAINING TO THE ADMINISTRATION AND ENFORCEMENT OF THE PESTICIDE APPLICATORS' ACT
Part 06 - RADIATION CONTROL - X-RAY IMAGING IN THE HEALING ARTS
Appendix 6 CCR 1007-1-6-E - USE OF HAND-HELD X-RAY EQUIPMENT

Current through Register Vol. 48, No. 6, March 25, 2025

6E.1 The following requirements are applicable, as determined by the Department, to any x-ray radiographic device that is designed to be held in the hand during operation.

6E.1.1 Requirements for any location:

(1) The facility shall adopt and follow procedures provided by the manufacturer regarding the safe operation of the device.

(2) The facility shall maintain documentation that each operator has completed training as specified by the manufacturer.

(3) The operator shall ensure there are no bystanders within a radius of at least 2 meters (more than 6 feet) from the patient being examined with a hand-held radiographic unit.

(4) If a hand-held device was designed with an optional, removable scatter radiation block, it shall be installed and used during patient examination and shall:

(a) Provide not less than 0.25 mm lead equivalent;

(b) Be at least 15.2 cm (6 inches) in diameter;

(c) Be positioned as close as practicable to the distal end of the position indication device.

(5) When operating a hand-held x-ray system, operators shall:

(a) Wear whole body dosimetry in accordance with Part 4, Section 4.6.3; and

(b) Wear 0.25 mm lead-equivalent protective apparel, unless the device is used with a scatter shield meeting the requirements of 6E.1.1(4) or as otherwise exempted in writing by the Department.

(6) In order to prevent repeat imaging due to motion that reduces image quality, motion shall be minimized as much as possible when holding and operating the device. If the operator has difficulty in holding the device stationary, the operator shall use a stand or tripod to immobilize the device.

6E.1.2 Additional requirements for use of hand-held x-ray equipment in permanent facilities:

(1) A hand-held device shall not be used for patient examinations in hallways and waiting rooms.

6E.2 When not under the control of the operator, the registrant shall secure the hand-held device from unauthorized removal or use.

Disclaimer: These regulations may not be the most recent version. Colorado may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.