Current through Register Vol. 47, No. 5, March 10, 2024
(a) The
requirements of this section constitute the regulations for total coliforms and
E. coli. This regulation establishes a maximum contaminant
level and treatment technique requirements.
11.16(1)
Applicability and
Definitions
(a) All public water
systems must comply with the requirements specified in this rule, unless
otherwise specified.
(b) "CLEAN
COMPLIANCE HISTORY" means a record of no MCL violations under 11.45(1), no
sampling violations under 11.16, and no treatment technique triggers or
treatment technique violations under 11.16 for a minimum of 12
months.
(c) Failure to comply with
the applicable requirements of 11.16 is a violation of the Colorado Primary
Drinking Water Regulations.
(d) The
supplier must have the system operated by qualified personnel who meet the
requirements of Regulation 100, the Water and Wastewater Facility
Operators Certification Requirements.
11.16(2)
Analytical Methods and
Laboratory Certification
(a) Suppliers
must analyze all compliance samples, required by 11.16 and 11.46(2), using a
Department-certified laboratory using a certified method.
11.16(3)
Sample Siting Plan
Requirements
(a) As part of the
monitoring plan specified in 11.5, the supplier must develop a written sample
siting plan, pursuant to 11.16(4) and 11.16(5), that identifies all of the
following:
(i) Sampling sites and a sample
collection schedule that are representative of water throughout the
distribution system. The supplier must collect total coliform samples according
to the written sample siting plan. Monitoring locations may include a
customer's premises, dedicated sampling station, or other designated compliance
sampling site. The sample siting plan must include routine and repeat sample
sites and any other sampling sites necessary to meet the requirements of
11.11.
(ii) Suppliers must identify
repeat sample sites in the sample siting plan. Unless the requirements of
11.16(3)(a)(ii)(A) or 11.16(4)(a)(v)(A) are met, the supplier must collect at
least one repeat total coliform sample at the site where the original total
coliform-positive sample was collected, at least one repeat total coliform
sample at a site within five service connections upstream from the site where
the original total-coliform positive sample was collected, and at least one
repeat total coliform sample at a site within five service connections
downstream from the site where the original total-coliform positive sample was
collected.
(A) Alternatively, suppliers may
propose repeat monitoring locations to the Department that the supplier
believes to be more representative of a pathway for contamination of the
distribution system. A supplier may elect to:
(I) Identify alternative fixed repeat
sampling sites that the supplier believes to be representative of a pathway for
contamination of the distribution system; or
(II) Develop criteria for selecting repeat
sampling sites on a situational basis that the supplier believes to best verify
and determine the extent of potential contamination and a potential pathway for
contamination of the distribution system in a standard operating procedure
(SOP) that is included in the sampling plan. The Department may modify the SOP
or require alternative repeat sampling sites.
(B) If the supplier collected the original
total coliform-positive sample from the end of the distribution system or one
site away from the end of the distribution system, the Department may allow an
alternative sampling site for collecting repeat samples at the upstream or
downstream sites.
(iii)
The Department may review, revise, and approve the written sample siting plan,
as specified in 11.16(3). The supplier must demonstrate that the sample siting
plan remains representative of the water quality in the distribution
system.
11.16(4)
Sampling Requirements
(a) To determine compliance with the MCL for
E. coli or to determine if a treatment technique is triggered,
the supplier must collect total coliform samples as specified in the supplier's
sample siting plan in 11.16(3) and as specified in 11.16(4) and 11.16(5).
(i) The supplier must collect total coliform
samples at regular time intervals throughout the month, except:
(A) For groundwater systems that supply less
than or equal to ([LESS THAN EQUAL TO]) 4,900 people, the supplier may collect
all required samples on a single day if the samples are collected from
different sites.
(ii)
The supplier may collect more samples than the minimum number of routine total
coliform samples required as specified in Table 11.16-I as a tool to
investigate potential problems in the distribution system.
(A) The supplier must use these sample
results to determine if a coliform treatment technique in 11.16(8)(a)(i) or
11.16(8)(a)(ii) has been triggered if the supplier collects these samples in
accordance with the sample siting plan and are representative of water
throughout the distribution system.
(iii) If any of the sample results collected
under 11.16(4) are total coliform-positive, the supplier must comply with the
repeat monitoring requirements in 11.16(5) and E. coli
analytical requirements in 11.16(4)(e).
(iv) The supplier is not required to submit
special purpose samples, as defined in 11.2(5)(78) unless the sample result is
E. coli-positive and is representative of water in the
distribution system. The supplier must submit E. coli-positive
special purpose sample results to the Department as specified in
11.35(2)(a).
(v) If an E.
coli MCL violation occurs under 11.16(11) or if a coliform treatment
technique is triggered under 11.16(8), the supplier must still collect at least
the minimum number of required samples.
(vi) For groundwater systems, the supplier
must collect triggered source water monitoring samples as specified in 11.11(4)
in addition to repeat samples required in 11.16(5).
(A) For a groundwater system with a single
well supplying less than or equal to ([LESS THAN EQUAL TO]) 1,000 people, if
the supplier is required to collect a triggered source water monitoring sample,
the supplier, with written Department approval, may collect one of the repeat
total coliform samples at the sample site required for triggered source water
monitoring under 11.11(4), if the supplier demonstrates to the Department's
satisfaction that the sample siting plan remains representative of water
quality in the distribution system.
(I) If
approved by the Department, the supplier may use the repeat total coliform
sample to meet both the triggered source water monitoring requirements
specified in 11.11(4) and the total coliform repeat sampling requirements
specified in this section, 11.16(5).
(II) If the repeat sample collected from the
groundwater source is E. coli-positive, the supplier must
comply with the requirements in 11.11(4).
(b)
Routine Sampling
Requirements for Total Coliform
(i) For all public water systems, the
supplier must collect the number of routine total coliform samples specified in
Table 11.16-I each month except:
(A) For
non-community groundwater systems that supply less than or equal to ([LESS THAN
EQUAL TO]) 1,000 people, the supplier must collect one total coliform sample
during each quarter that water is supplied to the public, unless the supplier
is required to increase the routine sampling frequency as specified in
11.16(4)(c).
(I) In any month where the system
supplies greater than (>) 1,000 people, the supplier must collect the number
of routine total coliform samples specified in Table 11.16-I each month.
(a) The supplier must have written
Department-approval to alternate between quarterly and monthly sampling
frequencies based on when the population supplied is less than or equal to
([LESS THAN EQUAL TO]) 1,000 people or when the population supplied is greater
than (>) 1,000 people.
(ii) For public water systems that haul
water, the water hauler must collect at least one total coliform sample from
the outlet port of each tank or container each month that the tank or container
is used to supply water to the public.
(iii) For hand-pumped wells, the supplier
must collect at least one total coliform sample from each hand-pumped well each
month that it supplies water to the public.
(iv) For the following public water systems,
the supplier is not eligible for a quarterly sampling frequency as specified in
11.16(4)(b)(i)(A):
(A) Seasonal
systems.
(B) Public water systems
that do not provide chemical disinfection.
(C) Public water systems that haul
water.
(D) Groundwater systems with
hand-pumped wells.
(v)
The Department must perform a special monitoring evaluation during each
sanitary survey to review the status of the system, including the distribution
system, and determine whether the supplier is on an appropriate monitoring
schedule.
(A) Based on the Department's
special monitoring evaluation, the Department may modify the Supplier's
monitoring schedule, consistent with 11.16(4) and 11.16(5)..
TABLE 11.16-I NUMBER OF ROUTINE TOTAL COLIFORM
SAMPLES REQUIRED PER MONITORING PERIOD
|
Population supplied
|
Minimum number of samples
required
|
Population supplied
|
Minimum number of samples
required
|
25 to
1,0001
|
1
|
59,001 to 70,000
|
70
|
1,001 to 2,500
|
2
|
70,001 to 83,000
|
80
|
2,501 to 3,300
|
3
|
83,001 to 96,000
|
90
|
3,301 to 4,100
|
4
|
96,001 to 130,000
|
100
|
4,101 to 4,900
|
5
|
130,001 to 220,000
|
120
|
4,901 to 5,800
|
6
|
220,001 to 320,000
|
150
|
5,801 to 6,700
|
7
|
320,001 to 450,000
|
180
|
6,701 to 7,600
|
8
|
450,001 to 600,000
|
210
|
7,601 to 8,500
|
9
|
600,001 to 780,000
|
240
|
8,501 to 12,900
|
10
|
780,001 to 970,000
|
270
|
12,901 to 17,200
|
15
|
970,001 to 1,230,000
|
300
|
17,201 to 21,500
|
20
|
1,230,001 to 1,520,000
|
330
|
21,501 to 25,000
|
25
|
1,520,001 to 1,850,000
|
360
|
25,001 to 33,000
|
30
|
1,850,001 to 2,270,000
|
390
|
33,001 to 41,000
|
40
|
2,270,001 to 3,020,000
|
420
|
41,001 to 50,000
|
50
|
3,020,001 to 3,960,000
|
450
|
50,001 to 59,000
|
60
|
3,960,001 or more
|
480
|
1 Includes
systems that have greater than or equal to (>=) 15 service connections, but
supply less than (<) 25 people.
(c)
For Non-community Groundwater
Systems Supplying Less Than or Equal to ([LESS THAN EQUAL TO]) 1,000 People -
Increased Routine Sampling Requirements for Total Coliform
(i) If the supplier is sampling quarterly,
the supplier must increase the routine sampling frequency to monthly if any of
the following events occur:
(A) The supplier
triggers a Level 2 assessment or two Level 1 assessments under 11.16(8) that
occur within a rolling 12-month period.
(B) A total coliform treatment technique
violation occurs.
(C) Two sampling
violations under 11.16occur in a rolling 12-month period.
(D) A Level 1 assessment is triggered and a
sampling violation of 11.16 occurs within a rolling12-month period.
(E) The supplier receives an E.
coli MCL violation.
(ii) The supplier must begin the monthly
sampling frequency in the month following the month that the event occurred
under 11.16(4)(c)(i).
(iii) If the
supplier is sampling monthly, the Department may allow the supplier to return
to a routine quarterly sampling frequency if all of the following criteria are
met:
(A) Within the last 12 months, the
Department or a Department-approved party has completed a sanitary survey or a
Level 2 assessment.
(B) The system
is free of sanitary defects and all significant deficiencies have been
corrected, has a protected source water, and meets approved construction
standards.
(C) The system's water
source(s) is protected from the direct influence of surface water or any other
source of contamination.
(D) The
system has a clean compliance history for at least 12 consecutive
months.
(d)
For Non-community Groundwater Systems Supplying Less Than or Equal
to ([LESS THAN EQUAL TO]) 1,000 People - Additional Routine Sampling
Requirements in the Month Following a Total Coliform-positive Sample
Result
(i) If the supplier is
collecting total coliform samples on a quarterly frequency and one or more of
the samples collected is total coliform-positive (with or without a Level 1
treatment technique trigger), the supplier must collect at least three
additional routine samples during the following month.
(A) The supplier may either collect the
samples at regular time intervals throughout the month or collect all required
additional routine samples on a single day if the samples are collected from
different sites.
(ii) If
any of the additional routine sample results are total coliform-positive, the
supplier must collect repeat samples as specified in 11.16(5).
(iii) The supplier must use the results of
additional routine samples to determine whether an E. coli MCL
violation of 11.16(10) has occurred or if a treatment technique requirement is
triggered under 11.16(9).
(iv) If
all three additional routine samples are total coliform-negative, the supplier
may return to collecting one total coliform sample on a quarterly sampling
frequency. The supplier must begin collecting the quarterly sampling frequency
in the calendar quarter following the month that the three additional routine
samples were required.
(e) If any routine or repeat sample result is
total coliform-positive, the supplier must have a laboratory analyze the total
coliform-positive culture medium to determine if
E. coli are
present.
(i) If any routine sample under
11.16(4), repeat sample under 11.16(5), or special purpose sample result under
11.3(78) is
E. coli-positive, the supplier must notify the
Department no later than the end of the day that the supplier is notified of
the sample result.
(A) If the supplier is
notified of the sample result after the Department is closed, the supplier must
contact the Department's after-hours phone line.
(f) If any routine sample
collected under 11.16(4) is total coliform-positive, the supplier must comply
with the repeat monitoring requirements in 11.16(5).
11.16(5)
Repeat Sampling Requirements
for Total Coliform
(a) For each routine
sample result collected under 11.16(4) that is total coliform-positive, the
supplier must collect a sample set of at least three repeat total coliform
samples no later than 24 hours after being notified of the positive sample
result.
(i) If the supplier has a logistical
problem beyond their control that prevents the supplier from collecting the
repeat samples within the 24-hour limit, the Department may extend the 24hour
limit on a case-by-case basis.
(A) If the
Department grants the extension, the Department shall specify how much time the
supplier has to collect the repeat samples.
(ii) The Department shall not waive the
requirement to collect repeat samples.
(b) The supplier must collect repeat samples
in accordance with the written sample siting plan required under
11.16(3).
(c) The supplier must
collect all repeat samples on the same day.
(i) If the system has only one service
connection, the Department may allow the supplier to collect a larger volume
repeat sample(s) in one or more sample containers of any size, as long as the
total volume collected is at least 300 ml.
(d) If a treatment technique is triggered
based only on routine sample results, the supplier is required to collect only
one repeat sample set for each routine total coliform-positive routine sample,
collected under 11.16(4), and is not required to comply with the requirements
specified in 11.16(5)(e).
(e) If
one or more of the repeat sample results is total coliform-positive, the
supplier must:
(i) Collect an additional
repeat sample set as specified in 11.16(5)(a-d) for each site that had a total
coliform-positive sample result.
(A) The
additional repeat sample set(s) must be collected no later than 24 hours after
being notified of the total coliform-positive sample result(s), unless the
Department extends the 24-hour limit as specified in 11.16(5)(a)(i).
(ii) Continue to collect
additional repeat sample sets as specified in 11.16(5)(e)(i) until either:
(A) Total coliforms are not detected in one
complete repeat sample set; or
(B)
A treatment technique is triggered as specified in 11.16(8) based on total
coliform-positive repeat sample results and the supplier has notified the
Department.
(iii) If a
trigger under 11.16(8) is exceeded as a result of a total coliform-positive
routine sample, the supplier is required to conduct only one round of repeat
monitoring for each total coliform-positive routine sample.
(f) If the supplier collects a
routine sample, which after analysis is found to be total coliform-positive,
but before receiving that sample result the supplier collects another routine
sample within five service connections of the original sample, the supplier may
use the subsequent routine sample as a repeat sample instead of as a routine
sample.
(g) Results of all routine
and repeat samples collected under 11.16(4) and or 11.16(5) not invalidated by
the Department under 11.16(7) must be used to determine if a coliform treatment
technique is triggered under 11.16(8).
11.16(6)
Additional Requirements for
Seasonal Systems
(a) The supplier must
complete Department-approved start-up procedures and certify that the startup
procedures were completed before supplying water to the public each season.
(i) No later than the 10th of the month
following the month that the system began supplying water to the public, the
supplier must submit the certification that start-up procedures were
completed.
(b) The
supplier must either submit start-up procedures for Department approval or use
the pre-approved procedures in the Department's Revised Total Coliform
Rule Start-up Procedures for Seasonal Systems Handbook.
(c) As part of the start-up procedures, the
supplier must collect a total coliform sample in the distribution system before
supplying water to the public.
(d)
All seasonal systems are required to collect monthly total coliform samples,
pursuant to 11.16(4), during the operating season according to Table
11.16-I.
11.16(7)
Invalidation of Total Coliform Samples
(a) A total coliform-positive sample result
invalidated under 11.16(7) does not count towards meeting the minimum reporting
requirements of 11.16.
(b) The
Department may invalidate a total coliform-positive sample result only if one
or more of the following conditions are met:
(i) The laboratory establishes that improper
sample analysis caused the total coliform-
positive sample result.
(ii) Based on repeat sample results collected
under 11.16(5), the Department determines that the total coliform-positive
sample resulted from a domestic or other non-distribution system plumbing
problem that is limited to the specific service connection from which the total
coliform sample was collected. The Department shall not invalidate a total
coliform-positive sample result on the basis of repeat sample results unless
all repeat sample(s) collected at the same site as the original total
coliform-positive sample are also total coliform-positive, and all repeat
samples collected at a site other than the original site are total
coliform-negative. The Department shall not invalidate a total
coliform-positive sample result solely on the basis that all repeat sample
results are total coliform-negative, or if the system has only one service
connection.
(iii) The Department
has substantial grounds to believe that a total coliform-positive sample result
was due to a circumstance or condition that does not reflect water quality in
the distribution system. If the Department makes this determination, the
supplier must still collect the required number of repeat samples under
11.16(5) and use them to determine if a treatment technique is triggered as
specified in 11.16(8). The Department must document the decision and supporting
rationale for invalidating a total coliform-positive sample result in writing,
have it approved and signed by a supervisor of the Department official who
recommended the decision, and make this document available to the EPA and the
public. The written documentation must state the specific cause of the total
coliform-positive sample result and what action the supplier has taken, or will
take, to correct the problem. The Department shall not invalidate a total
coliform-positive sample result solely on the basis that all repeat sample
results are total coliform-negative.
(iv) The laboratory shall invalidate a total
coliform sample result (unless total coliforms are detected) if the sample
produces a turbid culture in the absence of gas production using an analytical
method where gas formation is examined (e.g., the Multiple-Tube Fermentation
Technique), the sample produces a turbid culture in the absence of an acid
reaction in the Presence-Absence (P-A) Coliform Test, or the sample exhibits
confluent growth or produces colonies too numerous to count with an analytical
method using a membrane filter (e.g., Membrane Filter Technique). If the
laboratory invalidates a sample result because of such interferences, the
supplier must collect a replacement total coliform sample from the same site as
the invalidated sample no later than 24 hours after being notified of the
invalidation, and have it analyzed for the presence of total coliforms. The
supplier must continue to re-sample within 24 hours and have the samples
analyzed for the presence of total coliforms until the supplier obtains a valid
result. The Department may extend the 24-hour limit on a case-by-case basis.
(A) "CONFLUENT GROWTH" means, in the context
of bacterial testing, a continuous bacterial growth covering the entire
filtration area of a membrane filter, or a portion thereof, in which bacterial
colonies are not discrete.
(B) "TOO
NUMEROUS TO COUNT" means that the total number of bacterial colonies exceeds
200 on a 47-millimeter (mm) diameter membrane filter used for coliform
detection.
11.16(8)
Total Coliform Treatment
Technique Triggers
(a) The treatment
technique triggers for a Level 1 assessment are as follows:
(i) If the supplier collects greater than or
equal to (>=) 40 samples per month, more than 5.0
percent of the samples collected for the month are total
coliform-positive.
(ii) If
the supplier collects less than (<) 40 samples per month, two or more
samples collected for the monitoring period are total
coliform-positive.
(iii) The
supplier fails to collect all required repeat samples after any single total
coliform-positive sample.
(b) The treatment technique triggers for a
Level 2 assessment are as follows:
(i) An
E. coli MCL violation occurs as specified in
11.16(11)(a).
(ii) A second
treatment technique trigger for a Level 1 assessment, as specified in
11.16(8)(a), occurred within 12 consecutive months, except:
(A) If the Department has determined the
likely reason(s) for the total coliform-positive sample(s) that caused the
first Level 1 assessment to be triggered and the Department has established
that the supplier has corrected the problem(s).
11.16(9)
Treatment Technique
Requirements: Level 1 and Level 2 Assessment
(a) Once all of the monitoring required by
11.16(4) and or 11.16(5) has been completed, the supplier must determine if any
treatment technique has been triggered as specified in 11.16(8). If at any time
a treatment technique trigger has been exceeded, the supplier must complete the
assessments as required by 11.16(9).
(b)
General Requirements for
Assessments(i) To identify the
possible presence of sanitary defects and defects in distribution system
coliform sampling practices, the supplier must ensure that a Level 1 or Level 2
assessment is conducted.
(ii) The
supplier must ensure that the assessor evaluates at least all of the following
elements:
(A) Inadequacies in sample
sites.
(B) Inadequacies in sampling
protocol.
(C) Inadequacies in
sample processing.
(D) Atypical
events that could affect distributed water quality or indicate that distributed
water quality was impaired.
(E)
Changes in distribution system maintenance and operation, including water
storage, that could affect distributed water quality.
(F) Source and treatment considerations that
affect distributed water quality.
(G) Existing water quality monitoring
data.
(iii) The supplier
or the Department may request a consultation with the other party at any time
during the assessment or corrective action phase. The consultation may be used
to determine appropriate actions to be taken or to discuss relevant information
that may impact the supplier's ability to comply with the requirements
specified in 11.16(9).
(iv) If
required by the Department, the supplier must ensure that the assessment is
conducted consistent with any Department-specified modifications to assessment
elements based on the size and type of the system and the size, type, and
characteristics of the distribution system.
(v) The supplier must correct sanitary
defects found through either Level 1 or Level 2
assessments, pursuant to 11.16(9)(b)(ii). If the supplier has
not completed corrective action for any sanitary defect before the submission
of the assessment form, the supplier, in consultation with the Department, must
complete the corrective action(s) on a Department-approved schedule. The
supplier must notify the Department when each scheduled corrective action is
completed.
(c)
Level 1 Assessments
(i) If any treatment technique for a Level 1
assessment is triggered under 11.16(8)(a), the supplier must complete a Level 1
assessment, consistent with Department requirements, as soon as
practical.
(ii) No later than 30
days after learning of a treatment technique trigger for a Level 1
assessment, the supplier must submit for review a completed
Level 1 assessment form. In the completed form, the supplier must state whether
sanitary defects were identified and if so, describe all of the
following:
(A) Sanitary defects
identified.
(B) The likely
reason(s) for the treatment technique trigger.
(C) If sanitary defects are identified,
corrective actions completed.
(D)
If sanitary defects are identified, a proposed schedule for any corrective
actions not already completed.
(E)
The assessment form may also indicate that no sanitary defects were
found.
(iii) If the
Department reviews the Level 1 assessment form and determines that the
assessment is not sufficient or the assessment form is not complete (including
any proposed schedule for any corrective actions not already completed), the
Department shall consult with the supplier. If the Department requires
revisions after consultation, the supplier must submit a revised assessment
form to the Department on an agreed-upon date no later than 30 days from the
date of the consultation.
(iv) Upon
completion and submission of the assessment form by the supplier, the
Department shall determine if the supplier identified the
likely reason(s) for the Level 1 trigger. If the supplier identified the likely
reason(s) for the treatment technique trigger, the Department shall determine
if the supplier corrected the problem or included a Department-approved
schedule for correcting the problem.
(d)
Level 2
Assessments
(i) If any treatment
technique for a Level 2 assessment is triggered under 11.16(8)(b), the supplier
must ensure that a Level 2 assessment is conducted as soon as practical. The
supplier must ensure that the Level 2 assessment, consistent with Department
requirements, is completed by the Department or Department-approved
party.
(ii) The supplier must
comply with any expedited actions or additional actions required by the
Department in the case of an E. coli
violation.
(iii) No later
than 30 days after learning of a Level 2 treatment technique trigger
exceedance, the supplier must submit for review a completed Level 2 assessment
form. The supplier must state whether sanitary defects were identified and if
so, describe all of the following:
(A)
Sanitary defects identified.
(B)
The likely reason(s) for the Level 2 treatment technique trigger.
(C) If sanitary defects are identified,
corrective actions completed.
(D)
If sanitary defects are identified, a proposed schedule for any corrective
actions not already completed.
(E)
The assessment form may also indicate that no sanitary defects were
found.
(iv) If the
Department reviews the Level 2 assessment form and determines that the
assessment was not sufficient or the assessment form is not complete (including
any proposed schedule for any corrective actions not already completed), the
Department shall consult with the supplier. If the Department requires
revisions after consultation, the supplier must submit a revised assessment
form to the Department on an agreed-upon schedule no later than 30 days from
the date of the consultation.
(v)
Upon completion and submission of the assessment form by the supplier, the
Department shall determine if the supplier identified the likely reason(s) for
the Level 2 treatment technique trigger. If the supplier identified the likely
reason(s) for the treatment technique trigger, the Department shall determine
if the supplier corrected the problem or included a Department-approved
schedule for correcting the problem.
11.16(10)
Compliance Determination for
the E. coli MCL
(a)
To determine if an E. coli MCL violation has occurred, the
supplier must include the results of all routine and repeat samples collected
in the monitoring period under 11.16(4) and 11.16(5).
(b) The BATs for achieving compliance with
the E. coli MCL are specified in
40 CFR
141.63(e-f).
11.16(11)
Violations
(a) The following constitute
E.
coli MCL violations, pursuant to 11.45(1) and Table 11.45-I:
(i) A repeat sample is E.
coli-positive following a total coliform-positive routine
sample.
(ii) A repeat sample is
total coliform-positive following an E. coli-positive routine
sample.
(iii) The supplier fails to
collect all required repeat samples following an E.
coli-positive routine sample.
(iv) The supplier fails to analyze a total
coliform-positive repeat sample for E. coli.
(v) If a repeat sample collected at the
monitoring location for triggered source water monitoring is E.
coli-positive. The supplier must also comply with 11.11(4)(d). If a
supplier collects more than one repeat sample at the monitoring location for
triggered source water monitoring, the supplier may reduce the number of
additional source water samples required under 11.11(4) by the number of repeat
samples collected at that location that were not E.
coli-positive.
(vi) If a
supplier collects more than one repeat sample at the triggered source water
monitoring location under 11.11(4), and more than one repeat sample is
E. coli-positive, the supplier has violated the E.
coli MCL and must also comply with 11.11(6).
(vii) If all of the repeat samples collected
at the triggered source water monitoring location are E. coli
negative and a repeat sample collected other than the one for triggered source
water monitoring is E. coli-positive, the supplier has
violated the E. coli MCL under 11.16(11)(a) and the supplier
is not required to comply with 11.11(4)(d).
(b) The following constitute treatment
technique violations:
(i) A treatment
technique was triggered under 11.16(8) and the supplier failed to conduct the
required assessment or corrective action(s) within the timeframe as specified
in 11.16(9).
(ii) For seasonal
systems, the supplier fails to complete Department-approved start-up procedures
before supplying water to the public.
(c) The following constitute monitoring
violations:
(i) Failure to collect every
required routine or additional routine sample in the supplier's compliance
period.
(ii) Failure to analyze for
E. coli following a total coliform-positive routine
sample.
(d) The
following constitute reporting violations:
(i) Failure to submit monitoring results or a
completed assessment form after a supplier conducts the required monitoring or
assessment in a timely manner.
(ii)
Failure to notify the Department following an E. coli-positive
sample as required by 11.16(4)(e)(i) in a timely manner.
(iii) For seasonal systems, failure to submit
certification of completion of Department-approved start-up
procedures.
11.16(12)
Reporting Requirements
(a) In the event of an
E.
coli MCL violation, the supplier must:
(i) Notify the Department no later than the
end of the day that the supplier learns of the violation. If the supplier
learns of the violation after the Department is closed, the supplier must
contact the Department's 24 Hour Environmental Release/Incident report line
(i.e., after-hours phone line).
(ii) Distribute Tier 1 public notice as
specified in 11.33.
(b)
In the event of a coliform treatment technique violation under 11.16(11)(b),
the supplier must:
(i) Notify the Department
no later than the end of the next business day after the supplier learns of the
violation.
(ii) Distribute Tier 2
public notice as specified in 11.33.
(c) In the event of an E.
coli-positive routine or repeat sample event, the supplier must notify
the Department no later than the end of the day when the supplier is notified.
If the supplier learns of the violation after the Department is closed, the
Supplier must contact the Department's 24 Hour Environmental Release/Incident
report line (i.e., after-hours phone line).
(d) A supplier required to conduct an
assessment under 11.16(9) must submit the assessment report within 30 days. The
supplier must notify the Department, in accordance with 11.16(9)(c)(iv) or
11.16(9)(d)(v), when each scheduled corrective action is completed for
corrections not completed at the time of submission of the assessment
form.
(e) In the event of a
coliform monitoring violation, the supplier must:
(i) Notify the Department within 10 days
after the system discovers the violation.
(ii) Conduct Tier 3 public notice as
specified in 11.33.
(f)
No later than the 10th of the month following the
month that the seasonal system began supplying water to the public, the
supplier must submit certification that start-up procedures were
completed.