Section 5154.2 - Ventilation Requirements for Biological Safety Cabinets
Current through Register 2024 Notice Reg. No. 38, September 20, 2024
(a) Scope and Application. When biological safety cabinets, as defined below, are used to prevent harmful exposure from biohazard agents or biohazardous materials or hazardous substances they shall conform to the provisions of this Section and Section 5143.
EXCEPTION: Inspection doors or clean-out doors in exhaust ducts required by Section 5143(a)(3) do not apply to exhaust systems used in conjunction with biological safety cabinets.
(b) Definitions.
Class I. The Class I biological safety cabinet is an open-fronted, negative pressure, ventilated cabinet. Exhaust air from the cabinet is filtered by a high efficiency particulate air (HEPA) filter and discharged without internal recirculation. This cabinet may be used in three operational modes; with a full width open front, with an installed front closure panel not equipped with gloves, and with an installed front closure panel equipped with arm-length protective gloves.
Class II. The Class II vertical laminar flow biological safety cabinet is an open fronted, ventilated cabinet. Exhaust air is filtered with a high efficiency particulate air filter (HEPA). This cabinet provides HEPA-filtered downward air flow within the workspace. Class II biological safety cabinets are further classified as type A, type B1, type B2, and type B3. Class II type A cabinets may have positive pressure contaminated internal ducts and may exhaust HEPA filtered air into the laboratory. Class II type B1 cabinets have all biologically contaminated internal ducts or plenums under negative pressure or surrounded by negative pressure ducts or plenums, exhaust HEPA filtered air through external ducts to space outside the laboratory, and have HEPA filtered downflow air composed largely of unrecirculated inflow air. Class II type B2 cabinets (also know as "total exhaust" cabinets) have all biologically contaminated internal ducts or plenums under negative pressure or surrounded by negative pressure ducts or plenums, exhaust HEPA filtered air through external ducts to space outside the laboratory, and have HEPA filtered downflow air drawn from the laboratory or outside air. Class II type B3 cabinets (also known as "convertible" cabinets) have all biologically contaminated internal ducts or plenums under negative pressure or surrounded by negative pressure ducts or plenums, exhaust HEPA filtered air through external ducts to space outside the laboratory, and have HEPA filtered downflow air that is a portion of the mixed downflow and inflow air from a common exhaust plenum.
NOTE: Design, construction, and performance standards are available from the NSF International (the National Sanitation Foundation), Ann Arbor, Michigan. That standard is "National Sanitation Foundation Standard 49 Class II (Laminar Flow) Biohazard Cabinetry".
Class III. The Class III biological safety cabinet is a totally enclosed, negative pressure, ventilated cabinet of gas-tight construction. Operations within the Class III cabinet are conducted through attached protective gloves. Supply air is drawn into the cabinet through high efficiency particulate air filters. Exhaust air is filtered by two high efficiency particulate air filters placed in series or by high efficiency particulate air filtration and incineration, and discharged to the outdoor environment without re-circulation.
(c) Use. Where biological safety cabinets are used to prevent exposure to biohazard materials they shall be used according to the biosafety level assigned by the National Institute of Health, Centers for Disease Control, in HHS publication No. 93-8395, "Biosafety in Microbiological and Biomedical Laboratories", 1993, to the particular microorganism, HHS publication No. 93-8395, "Biosafety in Microbiological and Biomedical Laboratories", 1993, is available from the Superintendent of Documents, U.S. Government Printing Office, Washington, D.C. 20402 and is hereby incorporated by reference. When a particular microorganism is not assigned a level or listed, the criteria given in that publication shall be used to determine the appropriate biosafety level.
Class II biological safety cabinets may be used to prevent harmful exposure to cytotoxic agents during their compounding or preparation for parenteral use. Biological safety cabinets may be used to control harmful exposure to aerosols and particulate matter, provided the presence of the substance in the biological safety cabinet does not present a risk of fire or explosion. When biological safety cabinets are used to control exposure to these hazards they shall meet the requirements of this Section.
NOTE: The U.S. Department of Labor recommends the use of externally vented biological safety cabinets for the preparation of cytotoxic drugs in its work practice guidelines for cytotoxic drugs, OSHA Instruction PUB 8-1.1, January 29, 1986.
(d) Operation. In addition to the operation requirements of Section 5143(b) the mechanical ventilation system used to provide air flow and or negative pressures shall be operated until readily accessible interior surfaces are decontaminated. The manufacturers' recommendations for start-up and shut-down should also be followed. In the absence of such procedures, Appendix B contains a suggested start up and shut down procedure for Class II biological safety cabinets, the manufacturers' recommendations should also be followed.
(e) Ventilation rates and negative pressure. All Class I and II biological safety cabinet face velocities shall be sufficient to maintain an inward flow of air at all openings into the cabinet under operating conditions.
(f) Airflow measurements and HEPA filter leak testing. Biological safety cabinets shall be tested after installation, alterations, or maintenance, and at least annually. Records of tests performed shall be retained for at least five years.
EXCEPTION: Cabinets in which the exhaust filter is not accessible can be measured directly at the work access opening using a calibrated total capture air flow hood to measure the air volume entering the cabinet, and dividing this measurement by the area of the work access opening to determine the average face velocity.
EXCEPTION: Average intake velocity can also be measured directly at the work access opening using a calibrated total capture air flow hood to measure the air volume entering the cabinet, and dividing this measurement by the area of the work access opening to determine the average face velocity.
(g) Special requirements.
(h) Appendices. The information contained in the appendices is not intended, by itself, to create any additional obligations not otherwise imposed or detract from any existing obligation.
(i) Implementation. The requirements of subsection (g)(2) and (g)(3) shall be complied with no later than one year from the effective date of this Section. All other requirements of this Section shall be complied upon the effective date of this Section.
1. New section filed 10-11-94; operative 11-10-94 (Register 94, No. 41).
Note: Authority cited: Section 142.3, Labor code. Reference: Section 142.3, Labor Code.