Section 6887 - Extractors and Centrifuges

Universal Citation: 8 CA Code of Regs 6887

Current through Register 2024 Notice Reg. No. 38, September 20, 2024

(a) Each centrifugal extractor or centrifuge shall be equipped with a metal cover or access door of at least No. 20 gage, or its equivalent which shall entirely cover the opening of the outer shell.

(b) Each centrifugal extractor or centrifuge shall be equipped with an interlock that will disconnect the source of power to the drive motor and prevent starting rotation when the access door to the machine is open. The interlock shall also prevent opening of the access door or cover while the centrifugal units are rotating. The interlock shall allow the operator to rotate the unit by manual control or inch the unit for loading or unloading while the access door or cover is open.

(c) Each centrifugal extractor or centrifuge shall be equipped with a mechanically or electrically operated brake to stop the units when the driving power is shut off.

1. New section filed 9-6-94; operative 10-6-94 (Register 94, No. 36).

Note: Authority cited: Section 142.3, Labor Code. Reference: Section 142.3, Labor Code.

Disclaimer: These regulations may not be the most recent version. California may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.