(a) All water used in the reprocessing of dialyzers including solution makeup water shall be first passed through a reverse osmosis (RO) membrane unit.
(b) Water shall not exceed 200 colony forming units (CFU) as tested using a standard plate count method.
(c) Bacterial testing of the water shall be done monthly and the results documented in the Equipment Record.
(d) The water treatment system shall be regularly maintained and verification of maintenance so noted according to the schedule in the Equipment Record.
(e) Dialysis facilities may store water according to written water storage procedures.