Current through Register 2024 Notice Reg. No. 38, September 20, 2024
(a) Scope.
(1.0) This test method covers the
determination of the total pressure, exerted in vacuum, by air-containing,
volatile, petroleum products. The test method is suitable for testing samples
with boiling points above 0°C (32°F) that exert a vapor pressure
between 7 and 130 kPa (1.0 and 19 psi) at 37.8°C (100°F) at a
vapor-to-liquid ratio of 4:
1. The test method
is suitable for testing gasoline samples which contain oxygenates. No account
is made of dissolved water in the sample. (Samples can also be tested at other
vapor-to-liquid ratios, temperatures and pressures, but the Precision and Bias
as described in section (k) do not necessarily apply.)
(2.0) This test method covers the use of
automated vapor pressure instruments that perform measurements on liquid
specimen sizes in the range from 1 to 10 ml.
(3.0) Standard values are specified in SI
units (International System of Units). The values given in parentheses are
provided for information purposes only.
(4.0) This test method may involve hazardous
materials, operations, and equipment. This test method does not purport to
address all of the safety problems associated with its use. It is the
responsibility of the user of this test method to establish appropriate safety
and health practices and determine the applicability of regulatory limitations
prior to use. For specific hazard statements, see section
(g)(5.0).
(b) Summary of
Test Method.
(1.0) A known volume of chilled,
air-saturated sample is introduced into a thermostatically controlled test
chamber, the internal volume of which is five times that of the total test
specimen introduced into the chamber. A vacuum is applied to the chamber in
accordance with the manufacturer's instructions. After introduction into the
test chamber the test specimen is allowed to reach thermal equilibrium at the
test temperature, 37.8°C (100°F). The resulting rise in pressure in the
chamber is measured using a pressure transducer sensor and indicator.
(2.0) Only the sum of the partial pressure of
the sample and the partialpressure of the dissolved air (commonly known as the
total pressure) are used in this test method. Note that some instruments may
call this pressure measurement by another term. Also note that some instruments
are capable of measuring the absolute pressure of the specimen as
well.
(3.0) The measured total
vapor pressure is converted to a Reid vapor pressure equivalent (RVPE) by use
of a calibration equation (section (i)(1.0)). This calculation converts the
measured total pressure to the Reid vapor pressure (RVP) expected from the
American Society of Testing and Materials (ASTM) Test Method D
323-58.
(c) Apparatus.
(1.0) Vapor Pressure Apparatus - An
appropriate instrument, designed for the intended use should be selected. The
minimum performance level for the automated vapor pressure test instrument is
that the instrument shall perform as well as, or better than, the precision
criteria set forth in the ASTM D323-58, which is incorporated herein by
reference. The ASTM D323-58 states a repeatability value of 0.2 psi and a
reproducibility value of 0.3 psi. The instrument shall provide accurate results
which are comparable to the RVP measured by the ASTM 323-58. Typically, the
type of apparatus suitable for use in this test method employs a small volume
test chamber incorporating a transducer for pressure measurements and
associated equipment for thermostatically controlling the chamber temperature
and for evacuating the test chamber.
(1.1) The test chamber shall be designed to
contain between 5 and 50 ml of liquid and vapor and be capable of maintaining a
vapor-to-liquid ratio between 3.95 to 1.00 and 4.05 to 1.00.
(1.2) The pressure transducer shall have a
minimum operational range from 0 to 177 kPa (0 to 25.6 psi) with a minimum
resolution of 0.1 kPa (0.01 psi) and a minimum accuracy of ± 0.3 kPa
(± 0.05 psi). The pressure measurement system shall include associated
electronics and readout devices to display the resulting pressure
reading.
(1.3) The thermostatically
controlled heater shall be used to maintain the test chamber at 37.8
±0.1°C (100 ±0.2°F) for the duration of the
test.
(1.4) A platinum resistance
thermometer shall be used for measuring the temperature of the test chamber.
The minimum temperature range of the measuring device shall be from ambient to
60°C (140°F) with a resolution of 0.1°C (0.2°F) and accuracy of
0.1°C (0.2°F).
(1.5) The
vapor pressure apparatus shall have provisions for introduction of the test
specimen into the test chamber and for the cleaning or purging of the chamber
following the test.
(2.0) A vacuum
pump (if required by the manufacturer's instructions) shall be capable of
reducing the pressure in the test chamber to less than 0.01 kPa (0.001 psi)
absolute.
(3.0) A syringe
(optional, depending on sample introduction mechanism employed with each
instrument) shall be gas-tight. The syringe shall be 1 to 20-ml capacity with a
± 1% or better precision. The capacity of the syringe should not exceed
two times the volume of the test specimen being dispensed.
(4.0) Ice Water Bath or Refrigerator (Air
Bath): for chilling the samplesand syringe to temperatures betweens and 1°C
(32 to 34°F).
(5.0) Mercury
Barometer (if required by the manufacturer's instructions): in the 0 to 120 kPa
(0 to 17.4 psi) range.
(6.0) McLeod
Vacuum Gage (if required by the manufacturer's instructions): to cover at least
the range from 0 to 0.67 kPa (0 to 5mm Hg).
(d) Sampling.
(1.0) Obtain a sample in accordance with
title 13, California Code of Regulations, section
2296.
(2.0) The extreme sensitivity of vapor
pressure measurements to losses through evaporation and the resulting changes
in composition is such as to require the utmost precaution and most meticulous
care in the handling of samples.
(3.0) Protect samples from excessive high
temperatures prior to testing. This can be accomplished by storage in an
appropriate ice water bath or refrigerator.
(4.0) Do not test samples stored in leaky
containers. Discard and obtain another sample if leaks are
detected.
(e) Preparation
of Apparatus.
(1.0) Prepare the instrument
for operation in accordance with the manufacturer's instructions.
(2.0) Clean and prepare the test chamber as
required to avoid contamination of the test specimen.
(3.0) For instruments that require that the
test chamber be evacuated prior to the introduction of the test specimen: Prior
to specimen introduction, visually determine from the instrument display that
the test chamber pressure is stable and does not exceed 0.1 kPa (0.01 psi).
When the pressure is not stable or exceeds this value, check that the chamber
is clean of volatile materials remaining in the chamber from a previous
specimen or check the calibration of the transducer.
(4.0) If a syringe is used for introduction
of the specimen, chill it to between 0 and 4.5°C (32 and 40°F) in an
ice water bath or a refrigerator before drawing in the specimen. Avoid water
contamination of the syringe reservoir by suitably sealing the outlet of the
syringe during the cooling process.
(5.0) For instruments using a pre-heated test
chamber: Prior to introduction of the test specimen check that the temperature
of the test chamber is within the required range from 37.8 ± 0.1°C
(100 ± 0.2°F).
(f) Calibration.
(1.0) Pressure Transducer:
(1.1) Check the calibration of the pressure
transducer on a monthly basisor when needed as indicated from the quality
control checks (section (g)). The calibration of the pressure transducer is
checked using two reference points, zero pressure ( <0.1kPa) and the ambient
baromtric pressure.
(1.2) Connect a
McLeod gage to the vacuum source in line with the test chamber. Apply a vacuum
to the test chamber. When the McLeod gage registers a pressure less than 0.1
kPa (0.8mm Hg, or 0.01 psi), adjust the pressure transducer control to zero or
to the actual reading on the McLeod gage as dictated by the instrument design
and manufacturer's instructions.
(1.3) Open the test chamber to the atmosphere
and observe the pressure transducer reading. If the pressure reading is not
equal to the ambient barometric pressure, then adjust the pressure transducer
span control until the appropriate reading is observed. Ensure that the
instrument is set to display the total pressure and not a calculated or
corrected value.
(1.4) Repeat steps
(f)(1.2) and (f)(1.3) until the zero and barometric pressures read correctly
without further adjustments.
(2.0)
Thermometer - Check the calibration of the platinum resistance thermometer used
to monitor the temperature of the test chamber at least every six months
against a National Institute on Standards and Technology (NIST) traceable
thermometer.
(g) Quality
Control Checks.
(1.0) Check the performance of
the instrument each day it is in use by running a quality control sample
consisting of a pure solvent of known vapor pressure similar to the vapor
pressure of the samples to be tested. Treat the pure solvent quality control
check sample in the same manner as a sample (section (h)). Record the total
vapor pressure (do not calculate a Reid vapor pressure equivalent) in a log for
the purpose of tracking the instrument's performance. If the total vapor
pressure differs from the previous entry (for the same pure solvent) in the log
by more than ± 1.0 kPa (0.15 psi), then check the instrument calibration
(section (f)). If the trend of the log shows variations of more than ±
1.0 kPa (0.15 psi) (for the same pure solvent), also check the instrument
calibration.
(2.0) Some of the
possible reference pure materials and their corresponding absolute vapor
pressures include:
| cyclohexane | 22.5 kPa | (3.27
psi) |
| cyclopentane | 68.3 kPa | (9.92
psi) |
| 2,2-dimethylbutane | 67.9
kPa | (9.86 psi) |
| 2,3-dimethylbutane | 51.1
kPa | (1.41 psi) |
| 2-methylpentane | 46.7
kPa | (6.77 psi) |
| toluene | 7.1 kPa | (1.03
psi) |
(The total pressure values cited were obtained from
Phillips Petroleum Co., Bartlesville, OK, or the Table of Physical Constants,
National Gas Producer Association.)
(3.0) Purity of Reagents - Use chemicals of
at least 99% purity for quality control checks. Unless otherwise indicated, it
is intended that all reagents conform to the specifications of the Committee on
Analytical Reagents of the American Chemical Society where such specifications
are available. ("Reagent Chemicals, American Chemical Society Specifications,"
Am. Chemical Soc., Washington, DC. For suggestions on the testing of reagents
not listed by the American Chemical Society, see "Reagent Chemicals and
Standards," by Joseph Rosin, D. Van Nostrand Co, Inc., New York, NY and the
"United States Pharmacopeia.") Lower purities can be used, provided it is first
ascertained that the reagent is of sufficient purity to permit its use without
lessening the accuracy of the determination.
(4.0) The chemicals in this section are
suggested for use in quality control procedures; not for instrument
calibration.
(5.0)
WARNING--Cyclohexane, cyclopentane, 2,2-dimethylbutane, 3,2-dimethylbutane,
2-methylpentane, and toluene are extremely flammable. They are an aspiration
hazard and are harmful if inhaled. They are also a skin irritant on repeated
contact.
(h) Procedure.
(1.0) Sample Temperature - Cool the sample
container and contents in an ice water bath or refrigerator to the 0 to 1°C
(32 to 34°F) range prior to opening the sample container. Allow sufficient
time to reach this temperature.
(2.0) Verification of Sample Container
Filling - After the sample reaches thermal equilibrium at 0 to 1°C, take
the container from the ice water bath or refrigerator, wipe dry with an
absorbent material, unseal and examine the ullage. With a suitable gage,
determine that the liquid content in the container is between 70 to 80% of the
volume of the container capacity.
(2.1) Discard the sample if the liquid
content of the container is less than 70% of the volume of the container
capacity.
(2.2) If the liquid
content of the container is more than 80% of the volume of the container
capacity, pour out enough sample to bring the liquid contents within the 70 to
80% volume range.
(3.0) Air
Saturation of Sample in Sample Container
(3.1) After determining that the liquid
content in the sample container is between 70 to 80% full, reseal the container
and shake vigorously. Return the container to the ice water bath or
refrigerator for a minimum of 2 minutes.
(4.0) Remove the sample from the ice water
bath or refrigerator, dry the exterior of the container with absorbent
material, uncap, insert a transfer tube or syringe (section (e)(4.0)). Draw a
bubble-free aliquot of sample into a gas tight syringe or transfer tube and
deliver this test specimen to the test chamber as rapidly as possible. The
total time between opening the chilled sample container and inserting/securing
the syringe into the sealed test chamber shall not exceed 1 minute.
(5.0) The vapor pressure determination shall
be performed on the first test specimen withdrawn from a sample container.
Successive vapor pressure determinations can be made on the remaining test
material in the same container if the container had been tightly sealed
immediately after the previous vapor pressure determination.
(6.0) Follow the manufacturer's instructions
for the introduction of the test specimen into the test chamber, and for the
operation of the instrument to obtain a total vapor pressure result for the
test specimen.
(7.0) Set the
instrument to read the result in terms of total vapor pressure. If the
instrument is capable of calculating a Reid vapor pressure equivalent value,
ensure that only the parameters described in section (i)(2.0) are
used.
(8.0) Verification of Single
Phase - After drawing a test specimen and introducing it into the instrument
for analysis, check the remaining sample for phase separation. If the sample is
contained in a glass container, this observation can be made prior to sample
transfer. If the sample is contained in a non-transparent container, mix the
sample thoroughly and immediately pour a portion of the remaining sample into a
glass container and observe for evidence of phase separation. If the sample is
not clear and bright or if a second phase is observed, discretion shall be used
to determine if the sample is truly representative.
(9.0) Record the total vapor pressure reading
from the instrument to the nearest 0.1 kPa (0.01 psi). For instruments that do
not automatically record or display a stable pressure value, manually record
the pressure indicator reading every minute to the nearest 0.1 kPa; and, when
three successive readings agree to within 0.1 kPa, record the result to the
nearest 0.1 kPa (0.01 psi).
(i) Calculation.
(1.0) Calibration Equation - Calculate the
Reid vapor pressure equivalent (RVPE) using the following calibration equation.
Ensure that the instrument reading used in this equation corresponds to the
total pressure and has not been corrected by an automatically programmed
correction factor.
Equation 1:
RVPE=aX - b
where:
"RVPE" is the vapor pressure value (in psi) that would be
expected from test method ASTM D323-58;
"a" is the correlative relationship of test data from the
specific automated vapor pressure test instrument and test data from ASTM
D323-58;
"X" is the total vapor pressure value (in psi) as
determined by the specific automated vapor pressure test instrument;
"b" is the offset of the test data between the specific
automated vapor pressure test instrument and the test data from ASTM
D323-58.
The data used for determining the calibration equation for
each instrument shall be obtained during an Air Resources Board vapor pressure
test program. The data shall consist of test results obtained from the analysis
of identical samples by the automated instrument and by ASTM D323-58. Vapor
pressure test programs may be conducted on a periodic basis as needed. The Air
Resources Board conducted such a program and determined that the following
automated vapor pressure test instruments meet the requirements of section (c).
The data from the test program were used to arrive at the calibration equations
for these instruments. The calibration equations are as follows:
| 1. Grabner Instruments, Model: CCA-VP
(laboratory Grabner) | RVPE=(.965) x - .304 |
| 2. Grabner Instruments, Model: CCA-VPS
(portable Grabner) | RVPE=(.972) x - .715 |
| 3. Stanhope-Seta Limited, Model:
Setavap | RVPE=(.961) x - .577 |
(2.0) The calculation described in section
(i)(1.0), above, can be accomplished automatically by the instrument, if so
equipped, and in such cases the user shall not apply any further
corrections.
(j) Report.
(1.0) Report the Reid vapor pressure
equivalent to the nearest 0.1 kPa (0.01 psi).
(k) Precision and Bias.
(1.0) Precision - The precision of this test
method as determined by the statistical examination of interlaboratory test
results is as follows:
(1.1)
Repeatability - The difference between successive test results obtained by the
same operator with the same apparatus under constant operating conditions on
identical test material would, in the long run, in the correct operation of the
test method exceed the following value only in one case in twenty. The
repeatability values for the specific automated vapor pressure test instruments
listed in section (i)(1.0) are:
| 1. Grabner Instruments, Model: CCA-VP
(laboratory Grabner) | 0.084 psi |
| 2. Grabner Instruments, Model: CCA-VPS
(portable Grabner) | 0.084 psi |
| 3. Stanhope-Seta Limited Model:
Setavap | 0.10 psi |
(1.2) Reproducibility -- The difference
between two single and independent test results obtained by different operators
working in different laboratories using the same make and model test instrument
on identical test material would, in the long run, in the correct operation of
the test method exceed the following value only in one case in twenty. The
reproducibility values for the specific automated vapor pressure test
instruments listed in section (i)(1.0) are:
| 1. Grabner Instruments, Model: CCA-VP
(laboratory Grabner) | 0.13 psi |
| 2. Grabner Instruments, Model: CCA-VPS
(portable Grabner) | 0.21 psi |
| 3. Stanhope-Seta Limited Model:
Setavap | 0.32 psi |
(2.0) Bias - A relative bias was observed
between the total pressure obtained using this test method and the Reid vapor
pressure obtained using ASTM Test Method D323-58. This bias is corrected by the
use of the calibration equation in section (i)(1.0) which calculates a Reid
vapor pressure equivalent value from the observed total
pressure.
1. Change
without regulatory effect renumbering former section
2262 to section 2297 filed 9-17-91
pursuant to section
100, title 1, California Code of
Regulations (Register 92, No. 4).
2. Amendment of subsections
(d)(1.0) and (k)(1.1)-(1.2) filed 8-20-2001; operative 8-20-2001 pursuant to
Government Code section
11343.4
(Register 2001, No. 34).
Note: Authority cited: Sections
39600,
39601,
43013,
43018,
43101,
and
43830,
Health and Safety Code; and Western Oil and Gas Ass'n. v. Orange County Air
Pollution Control District, 14 Cal.3d 411, 121 Cal. Rptr. 249 (1975).
Reference: Sections
39000,
39001,
39002,
39003,
39500,
43000,
43013,
43018,
43101
and
43830,
Health and Safety Code; and Western Oil and Gas Ass'n. v. Orange County Air
Pollution Control District, 14 Cal.3d 411, 121 Cal. Rptr. 249
(1975).