Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 7 - MEDICAL USES OF RADIOACTIVE MATERIAL
Section R9-7-742 - Five-year Inspection for Teletherapy and Gamma Stereotactic Radiosurgery Units
Current through Register Vol. 30, No. 38, September 20, 2024
A. A licensee shall have each teletherapy unit and gamma stereo-tactic radiosurgery unit fully inspected and serviced during source replacement or at intervals not to exceed five years, whichever comes first, to assure proper functioning of the source exposure mechanism.
B. This inspection and servicing may only be performed by persons specifically licensed to do so by the Department, the NRC, or an Agreement State.
C. A licensee shall keep a record of each five-year inspection for three years from the date of the inspection, if the inspection determined that service was unnecessary, and three years from the date of the completed service if the inspection determined that service was needed.