Section R9-7-739 - Periodic Spot-checks for Gamma Stereotactic Radiosurgery Units
Current through Register Vol. 30, No. 38, September 20, 2024
A. A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform spot-checks of each gamma stereotactic radiosurgery facility and on each unit:
B. A licensee shall:
C. To satisfy the requirements of subsection (A)(1), spot-checks shall, at a minimum:
D. To satisfy the requirements of subsections (A)(2) and (A)(3), spot-checks shall assure proper operation of:
E. A licensee shall arrange for the repair of any system identified in subsection (C) that is not operating properly as soon as possible.
F. If the results of the checks required in subsection (D) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
G. A licensee shall retain a record of each check required by subsections (C) and (D) for three years from the date of the procedure, and a copy of the procedures required by subsection (B) until licensee terminates all medical activities involving the radiosurgery unit.