Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 7 - MEDICAL USES OF RADIOACTIVE MATERIAL
Section R9-7-736 - Full Calibration Measurements on Gamma Stereotactic Radiosurgery Units

Universal Citation: AZ Admin Code R 9-7-736

Current through Register Vol. 30, No. 38, September 20, 2024

A. A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform full calibration measurements on each unit:

1. Before the first medical use of the unit;

2. Before medical use under the following conditions:
a. Whenever spot-check measurements indicate that the output differs by more than 5 percent from the output obtained at the last full calibration corrected mathematically for radioactive decay;

b. Following replacement of the sources or following reinstallation of the gamma stereotactic radiosurgery unit in a new location; and

c. Following any repair of the gamma stereotactic radiosurgery unit that includes removal of the sources or major repair of the components associated with the source assembly; and

3. At intervals not exceeding one year, with the exception that relative helmet factors need only be determined before the first medical use of a helmet and following any damage to a helmet.

B. To satisfy the requirement of subsection (A), full calibration measurements shall include determination of:

1. The output within plusmn; 3 percent;

2. Relative helmet factors;

3. Isocenter coincidence;

4. Timer accuracy and linearity over the range of use;

5. On-off error;

6. Trunnion centricity;

7. Treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit off;

8. Helmet microswitches;

9. Emergency timing circuits; and

10. Stereotactic frames and localizing devices (trunnions).

C. A licensee shall use the dosimetry system described in R9-7-733(A) to measure the output for one set of exposure conditions. The remaining radiation measurements required in subsection (B)(1) may be made using a dosimetry system that indicates relative dose rates.

D. A licensee shall make full calibration measurements required by subsection (A) in accordance with published protocols accepted by nationally recognized bodies.

E. A licensee shall mathematically correct the outputs determined in subsection (B)(1) at intervals not exceeding one month for cobalt-60 and at intervals consistent with 1 percent physical decay for all other radionuclides.

F. Full calibration measurements required by subsection (A) and physical decay corrections required by subsection (E) shall be performed by an authorized medical physicist.

G. A licensee shall retain a record of each calibration for three years from the date of the procedure.

Disclaimer: These regulations may not be the most recent version. Arizona may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.