Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 7 - MEDICAL USES OF RADIOACTIVE MATERIAL
Section R9-7-735 - Full Calibration Measurements on Remote After-loader Units
Current through Register Vol. 30, No. 38, September 20, 2024
A. A licensee authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit:
B. To satisfy the requirement of subsection (A), full calibration measurements shall include, as applicable, determination of:
C. A licensee shall use the dosimetry system described in R9-7-733(A) to measure the output.
D. A licensee shall make full calibration measurements required by subsection (A) in accordance with published protocols accepted by nationally recognized bodies.
E. In addition to the requirements for full calibrations for low dose-rate remote afterloader units in subsection (B), a licensee shall perform an autoradiograph of the sources to verify inventory and source arrangement at intervals not exceeding one quarter.
F. For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made in accordance with subsections (A) through (E).
G. A licensee shall mathematically correct the outputs determined in subsection (B)(1) for physical decay at intervals consistent with 1 percent physical decay.
H. Full calibration measurements required by subsection (A) and physical decay corrections required by subsection (G) shall be performed by an authorized medical physicist.
I. A licensee shall retain a record of each calibration for three years from the date it was completed.