Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 7 - MEDICAL USES OF RADIOACTIVE MATERIAL
Section R9-7-724 - Surveys after Brachytherapy Source Implant and Removal; Accountability
Current through Register Vol. 30, No. 38, September 20, 2024
A. A licensee shall make a survey to locate and account for all sources that have not been implanted immediately after implanting sources in a patient or a human research subject.
B. A licensee shall make a survey of the patient or the human research subject with a radiation detection survey instrument immediately after removing the last temporary implant source to confirm that all sources have been removed.
C. A licensee shall maintain accountability at all times for all sources in storage or use.
D. A licensee shall return brachytherapy sources to a secure storage area as soon as possible after removing sources from a patient or a human research subject.
E. A licensee shall record the procedures performed in subsections (A) through (D) and retain the records for three years following completion of the record.
F. A licensee must use only brachytherapy sources: