Current through Register Vol. 30, No. 38, September 20, 2024
A. Definitions:
1. "CT" means computed tomography.
2. "CT conditions of operation" means all
selectable parameters governing the operation of a CT including nominal
tomographic section thickness, and technique factors.
3. "CTDI" means computed tomography dose
index, the integral of the dose profile along a line perpendicular to the
tomographic plane divided by the product of the nominal tomographic thickness
and the number of tomogram produced in a single scan.
4. "CTDI vol" means a value of a
volume-weighted tomography dose index. The unit of the CTDI vol is Gray or
subunits of the Gray. The value of the CTDI vol for patient scan is used to
trigger a notification when the value exceeds or will exceed a threshold
value.
5. "CTN" means CT number,
the number used to represent the x-ray attenuation associated with each
elemental area of the CT image.
6.
"Dose profile" means the dose as a function of position along a line.
7. "DLP" means the dose-length product. The
DLP is the mathematical product of the CTDI vol and the length of the scan. The
unit DLP is the Gray-cm of subunits of the Gray-cm. The DLP is used to trigger
a notification when the value exceeds or will exceed a threshold
value.
8. "Elemental area" means
the smallest area within a tomogram for which the x-ray attenuation properties
of a body are depicted.
9.
"Multiple tomogram system" means a CT system that obtains x-ray transmissions
data simultaneously during a single scan to produce more than one
tomogram.
10. "Nominal tomographic
section thickness" means the full width at half-maximum of the sensitivity
profile taken at the center of the cross section volume over which x-ray
transmission data are collected.
11. "Reference plane" means a plane that is
displaced from and parallel to the tomographic plane.
12. "Scan" means the complete process of
collecting x-ray transmission data for the production of a tomogram. Data can
be collected simultaneously during a single scan for the production of one or
more tomograms.
B.
Facility: A registrant shall ensure that a CT facility has:
1. An operable two-way communication system
between the patient and the operator in each CT room.
2. A viewing system that will allow the
operator to continuously view the patient from the control panel during each
examination. If the viewing system malfunctions the CT shall not be used until
the viewing system is repaired.
C. Equipment. A registrant shall ensure that:
1. There is a means to terminate x-ray
exposure automatically in the event of equipment failure by:
a. De-energizing the x-ray source,
or
b. Shuttering the x-ray
beam.
2. The equipment
shall provide the operator the ability to terminate the x-ray exposure at any
time during the examination, provided the scan or series of scans is greater
than one-half second duration.
a. If an
operator terminates an x-ray exposure, the operator shall reset the CT
conditions of operation before the initiation of another scan.
b. A visible signal shall indicate when an
x-ray exposure has been terminated because of equipment failure.
3. A means is provided to permit
visual determination of the tomographic plane for a single tomogram system, or
the location of a reference plane offset from a single tomograph or multiple
tomogram system.
a. If a light source is used
to satisfy this requirement, it shall provide illumination of the tomographic
plane or reference plane under ambient light conditions.
b. The difference between the actual plane
location and the indicated location of a tomographic plane or reference plane
shall not exceed 5 millimeters.
c.
The deviation of indicated scan increment versus actual increment shall not
exceed plus or minus 1 millimeter with any mass from 0 to 100 kilograms resting
on the patient support device.
4. The control panel and gantry provides a
visual indication, if x-rays are produced.
5. Emergency buttons and switches are marked
by function.
6. Parameters of CT
operation used during a patient examination are visible to the operator upon
initiation of the scan. If an operational parameter is not adjustable by the
operator, this subsection may be met by indicating on the control panel the
parameter is not adjustable by the operator.
7. Radiation exposure does not exceed 100 mR
in one hour at one meter in any direction from the tube port of an operating
CT.
8. The angular position or
positions where the maximum surface CTDI occurs is identified to allow for
reproducible positioning of a CT dosimetry phantom, except in those cases where
the x-ray tubes are designed to move, in which case, the maximum dose and
associated tube position shall be evaluated according to manufacturer
recommendations.
D.
Operating Procedures. A registrant shall ensure that:
1. Operating procedures are available at the
control panel, or by electronic means, regarding the operation of a CT and
evaluation of a CT's operation.
2.
The operating procedures contain the following information:
a. A copy of the latest evaluation of the
CT's operation, to include output for each CT procedure, performed by a
qualified expert;
b. Instructions
on the use of the CT performance phantom by the qualified expert, a schedule of
quality control tests with the results of the most recent quality control test,
and the allowable variations for the indicated parameters;
c. The distance in millimeters between the
tomographic plane and the reference plane if a reference plane is used;
and
d. A current technique chart
that contains the information required in R9-7-607(D)(4)(a) for both adult and
pediatric patients, as applicable, is available at the CT operating console,
and a procedure for determining whether a CT has been performed according to
instructions of a physician.
e. A
written or electronic log that contains the information required in
R9-7-607(D)(5) as well as an entry in the record of any displayed values for
the exam from either a CTDI vol or DLP measurement for each patient exam
completed on equipment manufactured on or after January 1, 2011.
3. If the evaluation of the CT's
operation or quality control test identifies a parameter exceeding the
tolerance established by a qualified expert, the use of a CT for patient
examination is limited to those uses established in written instructions from
the qualified expert.
E.
Quality control tests. A registrant shall have a written quality control test
procedure, developed by a qualified expert, and ensure that the quality control
test procedure:
1. Incorporates the use of a
CT performance phantom that is compatible with an approved accreditation
program approved by the Medicare Improvements for Patients and Providers Act
(MIPPA) or supplied by or approved for use by the manufacturer of the
unit.
2. Is followed in the
evaluation of the CT's operation, that the interval between tests does not
exceed those set forth in the application for accreditation or quarterly if not
accredited by an organization approved by (MIPPA), and that system conditions
are specified by the registrant's qualified expert.
3. Includes obtaining quality control test
images with the CT performance phantom using the same processing mode and CT
conditions of operation that are used to perform the evaluation of the CT's
operation.
4. Requires that images
obtained under subsection (E)(3) be retained until a new evaluation of the CT's
operation is performed.
5. Requires
that any Alerts and Notification settings using CTDI vol or DLP are reviewed
against preloaded techniques in the system and any missing fields are reviewed
with the staff radiologist and noted in the annual report.
6. Requires the quality control test
procedure and records of quality control tests performed be maintained for
three years for Department inspection.
F. Evaluation of a CT's operation. A
registrant shall ensure that:
1. The
evaluation of a CT's operation is performed by, or under the direct supervision
of, a qualified expert who is physically present at the facility during the
evaluation of the CT's operation.
2. The evaluation of a CT's operation:
a. Is performed before initial patient use
and annually (within two months of the annual due date) and after any change or
replacement of components that could, in the opinion of the qualified expert,
cause a change in radiation output; and
b. Shall measure the CTDI in a dosimetry
phantom along the two axes specified in subsection (F)(4)(b).
c. A complete evaluation of a CT unit,
performed before the annual due date shall clearly list if the new survey
changes the annual due date for the unit. It shall be clearly noted on all
documentation for the next three years that the survey has established a new
annual due date based upon the date of the new survey.
3. The evaluation of a CT's x-ray system is
performed with a calibrated dosimetry system that:
a. Has been calibrated using a method that is
traceable to the National Institute of Standards and Technology (NIST),
and
b. Has been calibrated within
the preceding two years.
4. CT dosimetry phantoms used in determining
radiation output are compatible with an approved accreditation program approved
by (MIPPA) or supplied by or approved for use by the manufacturer of the unit;
and
a. Are constructed in a way that the
parameters used to image the most commonly imaged parts of the human body are
evaluated; and
b. At a minimum,
provide means for placement of a dosimeter along the axis of rotation and along
a line parallel to the axis of rotation 1.0 centimeter from the outer surface
and within the phantom.
5. Any effects on the measured dose due to
the removal of phantom material to accommodate the dosimeter are accounted for
in the reported data or included in the statement of maximum deviation for the
measured values.
G. CT
units designated for simulator use, veterinary use, dental use, podiatry use,
and non-diagnostic use on humans are exempt from the annual requirements in
subsections (E) and (F) provided an initial evaluation is conducted by a
qualified expert and the output does not exceed the manufacturers specified
limits. The initial evaluation shall be maintained for Department
review.