Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 6 - USE OF X-RAYS IN THE HEALING ARTS
Section R9-7-610.01 - Hand-held Intraoral Dental Radiographic Unit Requirements For Use

Universal Citation: AZ Admin Code R 9-7-610.01

Current through Register Vol. 30, No. 38, September 20, 2024

A. Registrants are subject to the following requirements for Intraoral dental radiographic units designed to be operated as a hand-held unit:

1. For all uses:
a. Operators of hand-held intraoral dental radiographic units shall be specifically trained to operate such equipment.

b. A hand-held intraoral dental radiographic unit shall be held without any motion during a patient examination. A tube stand may be utilized to immobilize a hand-held intraoral dental radiographic unit during patient examination.

c. The operator shall ensure there are no bystanders within a radius of at least six feet from the patient being examined with a hand-held intraoral radio-graphic unit.

2. Additional requirements for operatories in permanent facilities:
a. Hand-held intraoral dental radiographic units shall be used for patient examinations in dental operatories that meet the structural shielding requirements specified by the Department or by a qualified health or medical physicist.

b. Hand-held intraoral dental radiographic units shall not be used for patient examinations in hallways and waiting rooms.

B. Hand-held units may only be used in a manner as specified on the registration issued by the Department.

Disclaimer: These regulations may not be the most recent version. Arizona may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.