Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 5 - SEALED SOURCE INDUSTRIAL RADIOGRAPHY
Section R9-7-504 - Radiation Survey Instruments
Current through Register Vol. 30, No. 52, December 27, 2024
A. A licensee shall maintain at least two calibrated and operable radiation survey instruments at each location where sources of radiation are present to make radiation surveys required by this Article and Article 4 of this Chapter. Instrumentation required by this Section shall be capable of measuring a range from 0.02 millisieverts (2 millirems) per hour through 0.01 sievert (1 rem) per hour.
B. A licensee shall ensure that each radiation survey instrument required under subsection (A) is calibrated:
C. A licensee shall maintain calibration records for each radiation survey instrument, and maintain each record for three years after it is made.