Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 14 - REGISTRATION OF NONIONIZING RADIATION SOURCES AND STANDARDS FOR PROTECTION AGAINST NONIONIZING RADIATION
Section R9-7-1440 - Medical Lasers
Current through Register Vol. 30, No. 38, September 20, 2024
A. A registrant shall ensure that a Class 3 and Class 4 laser product used in the practice of medicine has a means for measuring the level of laser radiation with an error in measurement of no greater than +20%, when calibrated in accordance with the laser product manufacturer's calibration procedure.
B. A registrant shall calibrate a laser used in the practice of medicine according to the manufacturer's specified calibration procedure, at intervals that do not exceed those specified by the manufacturer.
C. In a medical facility where several medical disciplines or a number of different practitioners use Class 3b and Class 4 lasers, a registrant shall form a Laser Safety Committee to govern laser activity, establish use criteria, and approve operating procedures, as follows:
D. A registrant shall use Class 3b and Class 4 Lasers that have a guard mechanism on the switch to control patient exposure and prevent inadvertent exposure.
E. A registrant shall establish a written laser safety training program that provides a thorough understanding of established procedures for each type of laser in use and the medical procedures associated with use of the laser. The registrant shall makeprogram documentation available for Department review and, at minimum, address all of the following in the documentation: