Section R9-7-1136 - Permanent Radiographic Installations
Current through Register Vol. 30, No. 38, September 20, 2024
A. If a registrant maintains a permanent radiographic installation that does not fall within the definition of "enclosed radiography" in R9-7-102, the registrant shall ensure that each entrance used for personnel access to the high radiation area has either:
B. A registrant shall test the alarm system for proper operation with a radiation source each day before the installation is used for radiographic operations. The test shall include a check of both the visible and audible signals. The registrant shall test each device referenced in subsection (A)(1) monthly. If an entrance control device or alarm signal is operating improperly, the registrant shall immediately label the device or signal as "defective" and repair the device or signal within seven calendar days. The registrant may continue to use the facility during this seven-day period, if the registrant implements continuous surveillance requirements of R9-7-1116 and uses an alarm rate meter.
C. A registrant shall maintain each record of alarm system and entrance control device tests for three years after the record is made.