Arizona Administrative Code
Title 9 - HEALTH SERVICES
Chapter 7 - DEPARTMENT OF HEALTH SERVICES - RADIATION CONTROL
Article 11 - INDUSTRIAL USES OF X-RAYS, NOT INCLUDING ANALYTICAL X-RAY SYSTEMS
Section R9-7-1114 - Inspection and Maintenance of Radiation Machines, Survey Instruments, and Associated Equipment
Current through Register Vol. 30, No. 38, September 20, 2024
A. A registrant shall perform visual and operability checks on survey instruments and radiation machines before use on each day the equipment is to be used to ensure that the equipment is in good working condition and required labeling is present. Survey instrument operability checks shall be performed using check sources or other authorized means. If equipment problems are found, the registrant shall remove the equipment from service until it is repaired.
B. A registrant shall have written inspection and maintenance procedures for radiation machines and survey instruments that require inspection and maintenance, at intervals that do not exceed three months or before first use of the equipment and to ensure the proper functioning of components important to safety. Replacement components shall meet design specifications. If equipment problems are discovered, the registrant shall remove the equipment from service until the equipment is repaired.
C. A registrant shall maintain records of equipment problems found in daily checks and quarterly inspections and retain each record for three years after it is made. The record shall include the date of the check or inspection, name of the inspector, equipment involved, any problems found, and any repair or needed maintenance performed.