Current through Register Vol. 42, No. 11, August 30, 2024
(1)
Purpose and Scope. This Rule
420-3-26-.11 provides special
requirements for analytical x-ray equipment; provided, however, that nothing in
this Rule shall apply to x-ray equipment used to detect, measure, gauge or
control the density, level, interface location, thickness of materials, or
equipment used for industrial radiography as defined in Rule
420-3-26-.04, or sources of
radiation used in the healing arts. The requirements of this Rule are in
addition to, and not in substitution for applicable requirements in other
Rules. Note that Rules
420-3-26-.01,
420-3-26-.03,
420-3-26-.05, and
420-3-26-.10 also apply to
analytical x-ray equipment users.
(2)
Definitions.
(a) "Analytical x-ray equipment" means any
device which utilizes x-rays for the purpose of examining the microstructure of
materials. This includes all types of x-ray diffraction, fluorescence, and
spectographic analysis equipment.
(b) "Analytical x-ray system" means a group
of local and remote components utilizing x-rays to determine the elemental
composition or to examine the microstructure of materials. Local components
include those that are struck by x-rays such as radiation source housings, port
and shutter assemblies, collimators, sample holders, cameras, gonimeters,
detectors and shielding. Remote components include power supplies,
transformers, amplifiers, readout devices, and control panels.
(c) "Fail-safe characteristics" means a
design feature which causes beam port shutters to close, or otherwise prevents
emergence of the primary beam, upon a failure of a safety or warning
device.
(d) "Normal operating
procedures" means operating procedures for conditions suitable for analytical
purposes with shielding and barriers in place. These do not include maintenance
but do include routine alignment procedures. Routine and emergency radiation
safety considerations are part of these procedures.
(e) "Open-beam configuration" means an
analytical x-ray system in which an individual could accidentally place some
part of his body in the primary beam path during normal operation.
(f) "Positive visual warning light" means a
warning light which has redundant lights so that a single failure will not
prevent the warning light from functioning.
(g) "Primary Beam" means ionizing radiation
which passes through the aperture of the source housing by a direct path from
the x-ray tube or a radioactive source located in the radiation source
housing.
(h) "Unattended operation"
means any operation in which the analytical x-ray system is generating x-rays
and an operator trained in accordance with
420-3-26-.11(6)
of this Rule
420-3-26-.11 is not physically
present in the area sufficiently near the local components to prevent any
operation which could cause an individual to exceed the limits given in
420-3-26-.03(6)
of these rules.
(3)
Equipment Requirements.
(a) A device such as a guard or interlock
which prevents the entry of any portion of an individual's body into the
primary x-ray beam path or which causes the beam to be shut off upon entry into
its path shall be provided on all open-beam configurations. Prior to operation
a registrant may apply to the Agency for an exemption from the requirements of
a safety device. Such application shall include a description of the
alternative methods that will be employed to minimize the possibility of an
accidental exposure, including procedures to assure that operators and others
in the area will be informed of the absence of safety devices.
(b) Warning Devices.
1. A positive visible warning light labeled
with the words "X-RAY ON," or words having a similar intent, shall be located;
(i) Near any switch that energizes an x-ray
tube and shall be illuminated only when the tube is energized; and
(ii) At a conspicuous location that may be
visible at all local components; and
2. Open-beam configurations shall be provided
with a readily discernible indication of;
(i)
x-ray tube status (ON-OFF) located near the radiation source housing, if the
primary beam is controlled in this manner; and/or
(ii) shutter status (OPEN-CLOSED) located
near each port on the radiation source housing, if the primary beam is
controlled in this manner.
Warning devices shall be labeled so that their purpose is
easily identified. On equipment transferred after January 1, 1977, warning
devices shall have fail-safe characteristics.
(c) Unused ports on radiation
source housings shall be secured in the closed position in the manner which
will prevent casual opening.
(d)
All analytical x-ray equipment shall be labeled with a readily discernible sign
or signs bearing the radiation symbol and the words:
1. "CAUTION-HIGH INTENSITY X-RAY BEAM," or
words having a similar intent on the x-ray source housing; and
2. "CAUTION RADIATION-THIS EQUIPMENT PRODUCES
X-RAYS WHEN ENERGIZED," or words having a similar intent, near any switch that
energizes an x-ray tube and at a conspicuous location if the radiation source
is an x-ray tube.
(e) On
open-beam configurations transferred after January 1, 1977, each port on the
radiation source housing shall be equipped with a shutter that cannot be opened
unless a collimator or a shielding coupling has been connected to the
port.
(f) Each x-ray tube housing
shall be so constructed that with all shutters closed the leakage radiation
measured at a distance of 5 cm from its surface is not capable of producing a
dose in excess of 2.5 mrem in one hour at any specified tube rating.
(g) Each x-ray generator, including high
voltage rectifiers, transformers, and amplifiers, shall be supplied with a
protective cabinet which limits leakage radiation measured at a distance of 5
cm from its surface such that it is not capable of producing a dose in excess
of 0.25 mrem in one hour.
(h) Each
entrance to a room containing analytical x-ray equipment in unattended
operations shall have a warning light with the words "X-RAY ON" or words having
a similar intent. In addition, for an open beam configuration unattended
operation, there shall be a device to shut off analytical x-ray equipment upon
the entrance of any person not trained in accordance with
420-3-26-.11(6)
of this Rule.
(4)
Area Requirements.
(a) The local components of an analytical
x-ray system shall be located and arranged and shall include sufficient
shielding or access control such that no radiation levels exist in any area
surrounding the local component group which could result in a dose to an
individual present therein in excess of the dose limits given in
420-3-26-.03(6)
of these rules. These levels shall be met at any specified rube
rating.
(b) Surveys. Radiation
surveys with appropriate radiation detection devices as required by
420-3-26-.03(9),
of all operable analytical x-ray systems sufficient to show compliance with
paragraph
420-3-26-.11(4)(a)
shall be performed quarterly and;
1. Upon installation of the
equipment;
2. Following any change
in the initial arrangement, number, or type of local components in the
system;
3. Following any
maintenance requiring the disassembly or removal of a local component in the
system;
4. During the performance
of maintenance and alignment procedures require the presence of a primary x-ray
beam when any local component in the system is disassembled or
removed;
5. Any time a visual
inspection of the local components in the system reveals an abnormal condition;
and
6. Whenever personnel
monitoring devices show a significant increase over the previous monitoring
period or the readings are approaching the limits specified in
420-3-26-.03(2).
7. Notwithstanding the radiation survey
requirements in
420-3-26-.11(4)(b),
quarterly radiation surveys are not required for devices with a maximum energy
of 70kVp or less.
(c)
Each area or room containing analytical x-ray equipment that is not under
constant surveillance shall be conspicuously posted with a sign or signs
bearing the radiation symbol and the words "CAUTION--X-RAY EQUIPMENT," or words
having a similar intent.
(5)
Operating
Requirements.
(a) Normal
operating procedures shall be written and available to all analytical x-ray
equipment workers. No person shall be permitted to operate analytical x-ray
equipment in any manner other than that specified in the procedures unless such
person has obtained written approval of the individual designated to the Agency
as the Radiation Safety officer.
(b) No person shall bypass a safety device
unless such person has obtained the approval of the designated Radiation Safety
Officer. When a safety device has been bypassed, a readily discernible sign
bearing the words "SAFETY DEVICE NOT WORKING," or words having a similar
intent, shall be placed on the radiation source housing.
(6) Personnel Requirements.
(a) Instruction. No person shall be permitted
to operate or maintain analytical x-ray equipment unless such person has
received instructions in and demonstrated competence as to:
1. Identification of radiation hazards
associated with the use of the equipment.
2. Significance of the various radiation
warning and safety devices incorporated into the equipment, or the reasons they
have not been installed on certain pieces of equipment and the extra
precautions required in such cases.
3. Proper operating procedures for the
equipment.
4. Biological effects of
radiation, including symptoms of an acute localized exposure; and
5. Proper procedures for reporting an actual
or suspected exposure.
(b) Personnel Monitoring. Finger or wrist
dosimetric devices shall be provided to and shall be used by:
1. Analytical x-ray equipment workers using
systems having an open-beam configuration and not equipped with a safety
device, and
2. Personnel
maintaining analytical x-ray equipment, if the maintenance procedures require
the presence of a primary x-ray beam when any local component in the analytical
x-ray system is disassembled or removed. In reporting dose values, due
consideration should be given to the energy of the x-ray beam and the size of
the x-ray beam.
Filed September 1, 1982. Readopted with
amendments filed October 24, 1983. Amended Filed January 19,
1990.
Author: Karl David Walter
Statutory Authority:
Code of Ala.
1975, §§
22-14-4,
22-14-7,
22-14-8.