Microbiology Devices; Reclassification of Antigen, Antibody, and Nucleic Acid-Based Hepatitis B Virus Assay Devices, 78265-78278 [2024-21932]
Download as PDF
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
Commonwealth of Independent States (CIS).
The FAA is issuing this AD to address the
reassessment of these repairs. The unsafe
condition, if not addressed, could lead to
damage or failure of the affected parts and
the NLG, and possible damage to the airplane
and injury to occupants.
(f) Compliance
Comply with this AD within the
compliance times specified, unless already
done.
(g) Requirements
Except as specified in paragraph (h) of this
AD: Comply with all required actions and
compliance times specified in, and in
accordance with, EASA AD 2024–0022,
dated January 23, 2024 (EASA AD 2024–
0022).
(h) Exceptions to EASA AD 2024–0022
(1) Where EASA AD 2024–0022 refers to its
effective date, this AD requires using the
effective date of this AD.
(2) This AD does not adopt the ‘‘Remarks’’
section of EASA AD 2024–0022.
ddrumheller on DSK120RN23PROD with PROPOSALS1
(i) Additional AD Provisions
The following provisions also apply to this
AD:
(1) Alternative Methods of Compliance
(AMOCs): The Manager, International
Validation Branch, FAA, has the authority to
approve AMOCs for this AD, if requested
using the procedures found in 14 CFR 39.19.
In accordance with 14 CFR 39.19, send your
request to your principal inspector or
responsible Flight Standards Office, as
appropriate. If sending information directly
to the manager of the International Validation
Branch, mail it to the address identified in
paragraph (j) of this AD. Information may be
emailed to: 9-AVS-AIR-730-AMOC@faa.gov.
Before using any approved AMOC, notify
your appropriate principal inspector, or
lacking a principal inspector, the manager of
the responsible Flight Standards Office.
(2) Contacting the Manufacturer: For any
requirement in this AD to obtain instructions
from a manufacturer, the instructions must
be accomplished using a method approved
by the Manager, International Validation
Branch, FAA; or EASA; or Airbus SAS’s
EASA Design Organization Approval (DOA).
If approved by the DOA, the approval must
include the DOA-authorized signature.
(3) Required for Compliance (RC): Except
as required by paragraph (i)(2) of this AD, if
any material contains procedures or tests that
are identified as RC, those procedures and
tests must be done to comply with this AD;
any procedures or tests that are not identified
as RC are recommended. Those procedures
and tests that are not identified as RC may
be deviated from using accepted methods in
accordance with the operator’s maintenance
or inspection program without obtaining
approval of an AMOC, provided the
procedures and tests identified as RC can be
done and the airplane can be put back in an
airworthy condition. Any substitutions or
changes to procedures or tests identified as
RC require approval of an AMOC.
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
(j) Additional Information
For more information about this AD,
contact Timothy Dowling, Aviation Safety
Engineer, FAA, 1600 Stewart Avenue, Suite
410, Westbury, NY 11590; phone 206–231–
3667; email Timothy.P.Dowling@faa.gov.
(k) Material Incorporated by Reference
(1) The Director of the Federal Register
approved the incorporation by reference
(IBR) of the material listed in this paragraph
under 5 U.S.C. 552(a) and 1 CFR part 51.
(2) You must use this material as
applicable to do the actions required by this
AD, unless this AD specifies otherwise.
(i) European Union Aviation Safety Agency
(EASA) AD 2024–0022, dated January 23,
2024.
(ii) [Reserved]
(3) For EASA AD 2024–0022 identified in
this AD, contact EASA, Konrad-AdenauerUfer 3, 50668 Cologne, Germany; telephone
+49 221 8999 000; email ADs@
easa.europa.eu; website easa.europa.eu. You
may find this EASA AD on the EASA website
at ad.easa.europa.eu.
(4) You may view this material at the FAA,
Airworthiness Products Section, Operational
Safety Branch, 2200 South 216th St., Des
Moines, WA. For information on the
availability of this material at the FAA, call
206–231–3195.
(5) You may view this material at the
National Archives and Records
Administration (NARA). For information on
the availability of this material at NARA,
visit www.archives.gov/federal-register/cfr/
ibr-locations, or email fr.inspection@
nara.gov.
Issued on September 19, 2024.
Peter A. White,
Deputy Director, Integrated Certificate
Management Division, Aircraft Certification
Service.
[FR Doc. 2024–21811 Filed 9–24–24; 8:45 am]
BILLING CODE 4910–13–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA–2024–N–3533]
Microbiology Devices; Reclassification
of Antigen, Antibody, and Nucleic
Acid-Based Hepatitis B Virus Assay
Devices
AGENCY:
Food and Drug Administration,
HHS.
Proposed amendment; proposed
order; request for comments.
ACTION:
The Food and Drug
Administration (FDA, the Agency, or
we) is proposing to reclassify qualitative
hepatitis B virus (HBV) antigen assays,
qualitative HBV antibody assays and
quantitative assays that detect anti-HBs
SUMMARY:
PO 00000
Frm 00006
Fmt 4702
Sfmt 4702
78265
(antibodies to HBV surface antigen
(HBsAg)), and quantitative HBV nucleic
acid-based assays, all of which are
postamendments class III devices, into
class II (general controls and special
controls), subject to premarket
notification. FDA is also proposing three
new device classification regulations
along with the special controls that the
Agency believes are necessary to
provide a reasonable assurance of safety
and effectiveness for each device.
DATES: Either electronic or written
comments on the proposed order must
be submitted by November 25, 2024.
Please see section X of this document
for the proposed effective date when the
new requirements apply and for the
proposed effective date of a final order
based on this proposed order.
ADDRESSES: You may submit comments
as follows. Please note that late,
untimely filed comments will not be
considered. The https://
www.regulations.gov electronic filing
system will accept comments until
11:59 p.m. Eastern Time at the end of
November 25, 2024. Comments received
by mail/hand delivery/courier (for
written/paper submissions) will be
considered timely if they are received
on or before that date.
Electronic Submissions
Submit electronic comments in the
following way:
• Federal Rulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
78266
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
• Mail/Hand Delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2024–N–3533 for ‘‘Microbiology
Devices; Reclassification of Antigen,
Antibody, and Nucleic Acid-Based
Hepatitis B Virus Assay Devices.’’
Received comments, those filed in a
timely manner (see ADDRESSES), will be
placed in the docket and, except for
those submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Dockets Management Staff between 9
a.m. and 4 p.m., Monday through Friday
Eastern Time, 240–402–7500.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://
www.govinfo.gov/content/pkg/FR-201509-18/pdf/2015-23389.pdf.
Docket: For access to the docket to
read background documents, the plain
language summary of the proposed
order of not more than 100 words
consistent with the ‘‘Providing
Accountability Through Transparency
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
Act,’’ or the electronic and written/
paper comments received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852, 240–402–7500.
FOR FURTHER INFORMATION CONTACT:
Maria Ines Garcia, Center for Devices
and Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3104, Silver Spring,
MD 20993, 301–796–7017,
Maria.Garcia@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (FD&C Act), as amended, establishes
a comprehensive system for the
regulation of medical devices intended
for human use. Section 513 of the FD&C
Act (21 U.S.C. 360c) established three
categories (classes) of devices, reflecting
the regulatory controls needed to
provide reasonable assurance of their
safety and effectiveness. The three
categories of devices are class I (general
controls), class II (general controls and
special controls), and class III (general
controls and premarket approval).
Section 513(a)(1) of the FD&C Act
defines the three classes of devices.
Class I devices are those devices for
which the general controls of the FD&C
Act (controls authorized by or under
sections 501, 502, 510, 516, 518, 519, or
520 (21 U.S.C. 351, 352, 360, 360f, 360h,
360i, or 360j) or any combination of
such sections) are sufficient to provide
reasonable assurance of safety and
effectiveness; or those devices for which
insufficient information exists to
determine that general controls are
sufficient to provide reasonable
assurance of safety and effectiveness or
to establish special controls to provide
such assurance, but because the devices
are not purported or represented to be
for a use in supporting or sustaining
human life or for a use which is of
substantial importance in preventing
impairment of human health, and do
not present a potential unreasonable
risk of illness or injury, are to be
regulated by general controls (section
513(a)(1)(A) of the FD&C Act). Class II
devices are those devices for which
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
and for which there is sufficient
information to establish special controls
to provide such assurance, including the
issue of performance standards,
postmarket surveillance, patient
PO 00000
Frm 00007
Fmt 4702
Sfmt 4702
registries, development and
dissemination of guidelines,
recommendations, and other
appropriate actions the Agency deems
necessary to provide such assurance
(section 513(a)(1)(B) of the FD&C Act).
Class III devices are those devices for
which insufficient information exists to
determine that general controls and
special controls would provide a
reasonable assurance of safety and
effectiveness, and are purported or
represented to be for a use in supporting
or sustaining human life or for a use
which is of substantial importance in
preventing impairment of human
health, or present a potential
unreasonable risk of illness or injury
(section 513(a)(1)(C) of the FD&C Act).
Devices that were not in commercial
distribution before May 28, 1976
(generally referred to as
‘‘postamendments devices’’) are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval, unless, and
until: (1) FDA reclassifies the device
into class I or II, or (2) FDA issues an
order finding the device to be
substantially equivalent, in accordance
with section 513(i) of the FD&C Act, to
a predicate device that does not require
premarket approval. The Agency
determines whether new devices are
substantially equivalent to predicate
devices by means of the premarket
notification procedures in section 510(k)
of the FD&C Act and part 807, subpart
E (21 CFR part 807, subpart E) of FDA’s
regulations.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or class
II under section 513(f)(3) of the FD&C
Act. Section 513(f)(3) of the FD&C Act
provides that FDA, acting by
administrative order, can reclassify the
device into class I or class II on its own
initiative, or in response to a petition
from the manufacturer or importer of
the device. To change the classification
of the device, the proposed new class
must have sufficient regulatory controls
to provide reasonable assurance of the
safety and effectiveness of the device for
its intended use.
FDA relies upon ‘‘valid scientific
evidence’’, as defined in section
513(a)(3) of the FD&C Act and 21 CFR
860.7(c)(2), in the classification process
to determine the level of regulation for
devices. To be considered in the
reclassification process, the ‘‘valid
scientific evidence’’ upon which the
Agency relies must be publicly available
(see section 520(c) of the FD&C Act).
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
ddrumheller on DSK120RN23PROD with PROPOSALS1
Publicly available information excludes
trade secret and/or confidential
commercial information, e.g., the
contents of a pending premarket
approval application (PMA) (see section
520(c) of the FD&C Act).
In accordance with section 513(f)(3) of
the FD&C Act, FDA is issuing this
proposed order to reclassify qualitative
HBV antigen assays intended for
qualitative detection of HBV antigens as
an aid in the diagnosis of acute or
chronic HBV infection in specific
populations, HBV antibody assays
(including qualitative and quantitative
anti-HBs) intended for use in the
detection of antibodies to HBV, and
quantitative HBV nucleic acid-based
assays intended for use in the detection
of HBV nucleic acid in specimens from
individuals with antibody evidence of
HBV infection, all of which are
postamendments class III devices, into
class II (general controls and special
controls) subject to premarket
notification, under three new device
classification regulations with the
names ‘‘Qualitative Hepatitis B Virus
Antigen Assays,’’ ‘‘Hepatitis B Virus
Antibody Assays,’’ and ‘‘Hepatitis B
Virus Nucleic Acid-Based Assays.’’ FDA
believes the standard in section
513(a)(1)(B) of the FD&C Act is met as
there is sufficient information to
establish special controls, which, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of these
devices.1
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
FDA has determined that premarket
notification is necessary to provide a
reasonable assurance of the safety and
effectiveness of HBV antigen assays,
HBV antibody assays, and HBV nucleic
acid-based assays for their intended
uses, therefore, the Agency does not
intend to exempt these proposed class II
devices from the requirement for
premarket notification (510(k))
submission as provided under section
510(m) of the FD&C Act. If this
1 FDA notes that the ‘‘ACTION’’ caption for this
proposed order is styled as ‘‘Proposed amendment;
proposed order,’’ rather than ‘‘Proposed order.’’
Beginning in December 2019, this editorial change
was made to indicate that the document ‘‘amends’’
the Code of Federal Regulations. The change was
made in accordance with the Office of the Federal
Register’s (OFR) interpretations of the Federal
Register Act (44 U.S.C. chapter 15), its
implementing regulations (1 CFR 5.9 and parts 21
and 22), and the Document Drafting Handbook.
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
proposed order is finalized, persons
who intend to market this type of device
must submit to FDA a premarket
notification under section 510(k) of the
FD&C Act prior to marketing the device.
II. Regulatory History of the Devices
Under section 513(f)(1) of the FD&C
Act, qualitative HBV antigen assays,
HBV antibody assays (including
qualitative and quantitative anti-HBs),
and quantitative HBV nucleic acidbased assays are automatically classified
into class III because they were not
introduced or delivered for introduction
into interstate commerce for commercial
distribution before May 28, 1976, and
have not been found substantially
equivalent to a device placed in
commercial distribution after May 28,
1976, which was subsequently classified
or reclassified into class II or class I.
Therefore, they are subject to PMA
requirements under section 515 of the
FD&C Act (21 U.S.C. 360e). Qualitative
HBV antigen assays and HBV antibody
assays (including qualitative and
quantitative anti-HBs) are prescription
devices and assigned product code
LOM. Quantitative HBV nucleic acidbased assays are prescription devices
and assigned product code MKT.
A. Qualitative HBV Antigen Assays
The first proposed device
reclassification action applies to
qualitative HBV antigen assay devices
that are prescription in vitro diagnostic
devices intended for qualitative
detection of HBV antigens as an aid in
the diagnosis of acute or chronic HBV
infection in specific populations. On
February 8, 2001, FDA approved its first
HBV antigen assay (DiaSorin’s ETI–EBK
PLUS) for use in the qualitative
detection of hepatitis Be antigen
(HBeAg) in human serum or plasma
(ethylenediaminetetraacetic acid
(EDTA), citrate, or heparin) as indicative
of a laboratory diagnosis of HBV
infection through its PMA process
under section 515 of the FD&C Act. On
June 1, 2001, FDA approved its first
HBV surface antigen (HBsAg) assay
(Roche Elecsys HBsAg Immunoassay,
Elecsys HBsAg Confirmatory, and
Precicontrol HBsAg) for the qualitative
detection of HBsAg in human serum or
plasma (heparin, EDTA, sodium citrate)
in adult pregnant and non-pregnant
individuals. In a May 22, 2002, Federal
Register notice (67 FR 36009), FDA
announced the approval order and the
availability of the Summary of Safety
and Effectiveness Data (SSED) for these
devices. Since the first approval order
for an HBV antigen assay issued on
February 8, 2001, FDA has approved 16
additional original PMAs for qualitative
PO 00000
Frm 00008
Fmt 4702
Sfmt 4702
78267
HBV antigen assays that are prescription
devices intended for the detection of
HBV antigens. These assays are
intended as an aid in the diagnosis of
acute or chronic HBV infection in
conjunction with clinical findings and
other diagnostic procedures (e.g., HBV
serology and antigen testing, liver
function, etc.). These assays are not
intended for use in screening of blood,
plasma, cells, or tissue donors.
A review of the medical device
reporting (MDR) databases indicates that
there were 625 reported events for
qualitative HBV antigen assays as of
June 2024. Of these reported events, a
significant majority of these were
determined to be of no known impact or
consequence to the patient. Events
reported included false reactive results,
false non-reactive results, incorrect or
inadequate assay results, incorrect/
inadequate/imprecise readings,
improper or incorrect procedure or
method, device operates differently than
expected, and adverse event without
identified device or use problems.
Where incorrect results were obtained,
it was not clear what the correct result
should have been. As of June 2024,
there have been no class III recalls, six
class II recalls, and no class I recalls 2
involving qualitative HBV antigen
assays. The class II recalls occurred
since 2006 due to defective caps, device
design, no marketing application, signal
for reactive results, and biased results
for biotin concentrations that were
lower than indicated. No patient harm
was identified. These facts, coupled
with the low number of reported events
that caused patient harm, indicate a
good safety record for this device class.
These recall events reflect the risks to
health identified in section V below,
and FDA believes the special controls
proposed herein, in addition to general
controls, can effectively mitigate the
risks identified in these recalls.
B. HBV Antibody Assays (Including
Qualitative and Quantitative Anti-HBs)
The second type of devices this
proposed reclassification order applies
to are qualitative HBV antibody assays
and quantitative anti-HBs assays that are
prescription in vitro diagnostic devices
intended for use in the detection of
antibodies to HBV. These devices are
intended to aid in the diagnosis of HBV
infection in persons with signs and
symptoms of hepatitis and in persons at
risk for HBV infection. On September
29, 2000, FDA approved its first
qualitative HBV antibody assay (OrthoClinical Diagnostics, Inc.’s Vitros
2 Class I, II, and III recalls are defined in 21 CFR
7.3(m).
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
78268
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
Immunodiagnostic Products: Anti-HBS
Reagent Pack/Anti-HBS Calibrators) for
the qualitative in vitro determination of
total antibody to hepatitis B surface
antigen (anti-HBs) in human serum as
an aid in determining susceptibility to
HBV infection for individuals prior to or
following HBV vaccination, or where
vaccination status is unknown, and for
use with other HBV serological markers
for the laboratory diagnosis of HBV
disease associated with HBV infection,
through its PMA process under section
515 of the FD&C Act. In a March 12,
2001, Federal Register notice (66 FR
14390), FDA announced the approval
order and the availability of the SSED
for this device. On July 22, 2002, FDA
approved its first quantitative Anti-HBs
(Siemens Healthcare Diagnostics
Products Ltd.’s Immulite 2000 XPI AntiHBs) for the quantitative measurement
of total antibodies to the hepatitis B
surface antigen (anti-HBs) in human
serum and plasma (heparinized or
EDTA) as an aid in the determination of
susceptibility to HBV infection for
individuals prior to or following HBV
vaccination, or where vaccination status
is unknown, or for use with other HBV
serological markers for the laboratory
diagnosis of HBV disease associated
with HBV infection, through its PMA
process under section 515 of the FD&C
Act.
Since the first approval order of a
qualitative HBV antibody assay on
September 29, 2000, FDA has approved
31 additional original PMAs for
qualitative HBV antibody assays for the
detection of antibodies to HBV. FDA has
also approved six assays for quantitative
anti-HBs detection. Qualitative HBV
antibody assays and quantitative antiHBs assays are intended to aid in the
diagnosis of HBV infection in persons
with signs and symptoms of hepatitis
and in persons at risk for HBV infection
in conjunction with clinical findings
and other diagnostic procedures (e.g.,
HBV serology and antigen testing, liver
function, etc.). These assays are not
intended for use in screening of blood,
plasma, cells, or tissue donors.
A review of the MDR databases
indicates that there were 1,107 reported
events for HBV antibody assays between
years 2001 and June 2024. Of these
reported events, a significant majority of
these were of no known impact to the
patient, and only four resulted in impact
to patients such as misdiagnosis or viral
infection. Events reported included
adverse events without identified device
or use problem, disconnection/low
assay results, false non-reactive results,
false reactive results, false high assay
results (for example, the first assay
result had a low signal to cutoff (s/co)
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
value and repeat testing produced a
higher s/co value), incorrect assay
results, inadequate assay results, and
low assay results (for example, the first
assay result was in the equivocal zone,
repeat testing produced a non-reactive
result, and testing with an alternate
device produced a reactive result). In
numerous cases, it was not possible to
determine what the correct result
should have been (further testing was
not performed, insufficient sample
volume, different assays were used). As
of June 2024, FDA is aware of 4 class III
recalls, 12 class II recalls, and no class
I recalls for these devices. The class II
recalls occurred in 2007, 2008, 2009,
2011, 2012, 2013, 2014, 2018, and 2019,
and were related to issues such as false
reactive results, false high assay results,
defective caps, and errors in labeling,
packaging, or software. No patient harm
has been identified. These facts,
coupled with the low number of
reported events that impacted the
patient, indicate a good safety record for
this device class. These recall events
reflect the risks to health identified in
section V below, and FDA believes the
special controls proposed herein, in
addition to general controls, can
effectively mitigate the risks identified
in these recalls.
C. Quantitative HBV Nucleic AcidBased Assays
Finally, the third type of device this
proposed reclassification order applies
to are quantitative HBV nucleic acidbased assay devices for use as a
prescription in vitro diagnostic device
intended for use in the detection of HBV
nucleic acid in specimens from
individuals with antibody evidence of
HBV infection. On September 4, 2008,
FDA approved its first quantitative HBV
nucleic assay (Roche Molecular
Systems, Inc.’s COBAS TaqMan HBV
Test For Use With The High Pure
System), an in vitro nucleic acid
amplification assay for the quantitation
of HBV deoxyribonucleic acid (DNA) in
human serum or plasma (EDTA)
intended for use as an aid in the
management of patients with chronic
HBV infection undergoing antiviral
therapy, through its PMA process under
section 515 of the FD&C Act.
Since the first approval order, FDA
has approved four additional original
PMAs for quantitative HBV nucleic
acid-based assays for the quantitative
detection of HBV DNA. The detection of
HBV DNA is used for management of
patients undergoing antiviral therapy for
assessing response to treatment and not
as a diagnostic for HBV infection.
The following section provides
examples of the different technologies
PO 00000
Frm 00009
Fmt 4702
Sfmt 4702
used. The different technologies begin
with specimen lysis and HBV DNA
through hybridization with magnetic
particles. The differences in the
technologies occur with the method of
amplification:
• In one technology, the target HBV
DNA sequence is amplified. The
presence of HBV amplification products
is detected by measuring the
fluorescence of the HBV probe that
binds to the target. Similarly, the
presence of the internal control
amplification product is detected. In the
absence of HBV or internal control
target sequences, probe fluorescence is
quenched. In the presence of HBV or
internal control target, the HBV or
internal control probes bind to their
target.
• In another technology, target
amplification occurs via transcriptionbased nucleic acid amplification by
fluorescent labeled probes (torches).
More torches hybridize when more
amplicon is present creating a higher
fluorescent signal. The time taken for
the fluorescent signal to reach a
threshold proportional to the starting
HBV DNA concentration is measured in
relation to internal controls.
A review of the MDR databases
indicates that as of June 2024 there were
13 reported events for nucleic acidbased HBV DNA assays since the first
reported event in 2009. MDRs were for
the following reasons: (1) incorrect,
inadequate, or imprecise result or
readings; (2) high readings; and (3) nonreproducible results. Of these, two had
no known impact or consequence to the
patient and two occurred when the
patient had no signs, symptoms, or
conditions. As of June 2024, FDA is
aware of one class III recall, five class
II recalls, and no class I recalls for these
devices. The class II recalls occurred
between 2005 and 2022 and were
related to issues such as misquantitation
of high results for negative samples
(carryover from a high positive sample
tested adjacent to a negative sample may
produce an incorrect positive result),
liquid level detection of reagent
cassette, under filled and over filled
enzyme reagent vials in assay kits,
software, and low level of recombinant
HBV DNA found in one lot of reagent.
These facts, coupled with the low
number of reported events that
impacted the patient, indicate a good
safety record for this device class. These
recall events reflect the risks to health
identified in section V below, and FDA
believes the special controls proposed
herein, in addition to general controls,
can effectively mitigate the risks
identified in these recalls.
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
III. Device Description
The HBV assays that are the subject of
this proposed order are
postamendments prescription in vitro
diagnostic devices classified into class
III under section 513(f)(1) of the FD&C
Act.
A. Qualitative HBV Antigen Assays
A qualitative HBV antigen assay is a
prescription in vitro diagnostic device
intended for use in the qualitative
detection of HBV antigens and for use
as an aid in the diagnosis of HBV
infection in specific populations. HBV
antigen assays aid in the diagnosis of
acute or chronic HBV infection. HBV
antigen assays typically detect the
presence of Hepatitis B surface antigen
(HBsAg) or Hepatitis B e antigen
(HBeAg). HBV antigens (HBsAg and
HBeAg), when present in samples, bind
to anti-HBs or anti-HBe antibodies to
form a complex that is bound to a solid
phase (e.g., microparticles, microtiter
plate or other technology). Detection of
the complexes can be performed using
different methods which measure the
presence/absence of anti-HBs or antiHBe antibodies in the sample.
Diagnosis of HBV infection should not
be established based on a single assay
result but should be determined in
conjunction with clinical findings and
other diagnostic procedures (e.g., HBV
serology and antigen testing, liver
function, etc.). These assays are not
intended for use in screening of blood,
plasma, cells, or tissue donors.
ddrumheller on DSK120RN23PROD with PROPOSALS1
B. HBV Antibody Assays (Including
Qualitative and Quantitative Anti-HBs)
A qualitative HBV antibody assay is a
prescription in vitro diagnostic device
intended for use in the qualitative
detection of antibodies to HBV and for
use as an aid in the diagnosis of HBV
infection in specific populations. HBV
antibody assays aid in the diagnosis of
HBV infection in persons with signs and
symptoms of hepatitis and in persons at
risk for HBV infection. Antibody assays
typically detect the presence of
antibodies to HBsAg (anti-HBs),
Hepatitis B core antigen (anti-HBc), or
HBeAg (anti-HBe). Diagnosis of HBV
infection should not be established
based on a single assay result, but
should be determined in conjunction
with clinical findings and other
diagnostic procedures (e.g., HBV
serology and antigen testing, liver
function, etc.). These assays are not
intended for use in screening of blood,
plasma, cells, or tissue donors.
A quantitative assay that detects antiHBs (antibodies to HBV surface antigen
(HBsAg)) is a prescription in vitro
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
diagnostic device that is intended for
quantitative use to aid in the diagnosis
of HBV infection in persons with signs
and symptoms of hepatitis and in
persons at risk for HBV infection.
Detection of anti-HBs indicates a
present or past infection with HBV and
can be used in conjunction with clinical
findings such as other HBV serological
markers (detection of other HBV
antigens and antibodies to HBV) for
diagnosis of HBV infection. Anti-HBs
assay results may be used as an aid in
the determination of susceptibility to
HBV infection in individuals prior to
vaccination or when vaccination status
is unknown.
In some device designs, HBV
antibodies, when present in the sample,
bind to HBV antigens to form a complex
that is bound to a solid phase (e.g.,
microparticles, microtiter plate, or other
technology). Detection of complexes can
be performed using different methods
that measure the presence/absence of
HBV antibodies in the sample.
C. Quantitative HBV Nucleic AcidBased Assays
A quantitative HBV nucleic acidbased assay is a prescription in vitro
diagnostic device intended for use in
the detection of HBV nucleic acid in
specimens from individuals with
antibody evidence of HBV infection. In
these devices, the detection of HBV
nucleic acid is used for management of
patients undergoing antiviral therapy for
assessing response to treatment and
NOT as a diagnostic for HBV infection.
FDA is proposing to reclassify
qualitative HBV antigen, HBV antibody
assays (including qualitative and
quantitative anti-HBs), and quantitative
HBV nucleic acid-based assays from
class III (general controls and premarket
approval) to class II (general controls
and special controls) and to establish
new names for the device types that will
be within the classification regulations.
FDA proposes to revise 21 CFR part 866
to create three new device classification
regulations with the names ‘‘Qualitative
Hepatitis B Virus Antigen Assays,’’
‘‘Hepatitis B Virus Antibody Assays,’’
and ‘‘Hepatitis B Virus Nucleic AcidBased Assays.’’ FDA believes that these
names and proposed identification
language most accurately describe these
devices.
• A Qualitative Hepatitis B Virus
(HBV) Antigen Assay is tentatively
identified as an in vitro diagnostic
device intended for prescription use for
qualitative use with human serum,
plasma, or other matrices that aids in
the diagnosis of chronic or acute HBV
infection. HBV surface antigen (HBsAg)
is also used for screening of HBV
PO 00000
Frm 00010
Fmt 4702
Sfmt 4702
78269
infection in pregnant women to identify
neonates who are at risk of acquiring
hepatitis B during perinatal period. The
assay is not intended for screening of
blood, plasma, cells, or tissue donors.
• A Hepatitis B Virus (HBV) Antibody
Assay is tentatively identified as an in
vitro diagnostic device intended for
prescription use in the detection of
antibodies to HBV in human serum and
plasma, or other matrices, and or as an
aid in the diagnosis of HBV infection in
persons with signs and symptoms of
hepatitis and in persons at risk for
hepatitis B infection. In addition, antiHBc IgM (IgM antibodies to core
antigen) assay is indicative of recent
HBV infection. Anti-HBs (antibodies to
surface antigen) assay results may be
used as an aid in the determination of
susceptibility to HBV infection in
individuals prior to or following HBV
vaccination or when vaccination status
is unknown. The assay is not intended
for screening of blood, plasma, cells, or
tissue donors. The assay is intended as
an aid in diagnosis in conjunction with
clinical findings and other diagnostic
procedures.
• A Hepatitis B Virus (HBV) Nucleic
Acid-Based Assay is tentatively
identified as an in vitro diagnostic
device intended for prescription use in
the detection of HBV nucleic acid in
specimens from individuals with
antibody evidence of HBV infection. In
these devices, the detection of HBV
nucleic acid is used as an aid in the
management of HBV-infected
individuals. The assay is intended for
use with human serum or plasma (and
other matrices as applicable) from
individuals with HBV. The assay is not
intended for use as a donor screening
assay for the presence of HBV nucleic
acids in blood, blood products, plasma,
cells, or tissue donors.
Based upon our review experience
and consistent with the FD&C Act and
FDA’s regulations in 21 CFR 860.134,
FDA believes that these devices should
be reclassified from class III into class
II with special controls because there is
sufficient information to establish
special controls that, along with general
controls, can provide reasonable
assurance of the devices’ safety and
effectiveness.
IV. Proposed Reclassification and
Summary of Reasons for
Reclassification
FDA is proposing to reclassify the
HBV assays that are the subject of this
proposed order. On September 7, 2023,
the Microbiology Devices Panel (Panel)
of the Medical Devices Advisory
Committee convened to discuss and
make recommendations regarding the
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
78270
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
reclassification of HBV assays from class
III (general controls and premarket
approval) to class II (general controls
and special controls) (https://
www.fda.gov/media/173609/download).
Panel members unanimously agreed that
special controls, in addition to general
controls, are necessary and sufficient to
mitigate the risks to health of patients
presented by these devices and to
provide reasonable assurance of the
safety and effectiveness of these devices
(Refs. 1 and 2). The Panel agreed with
FDA-identified risks and identified
additional risk(s) and benefit(s) to
include in the overall risk assessment.
The Panel also discussed potential
mitigation measure(s)/control(s) FDA
should consider for each of the
identified risks and recommended that,
as part of any reclassification, the
expected performance for these devices
should remain the same. Notably, the
performance of approved HBV antigen
assays has generally been at least 97
percent sensitivity and 99 percent
specificity. For approved anti-HBs, antiHbe, and anti-HBc total assays the
sensitivity has generally been at least 95
percent, for approved anti-HBc IgM
assays the sensitivity has been at least
86 percent, and for all HBV approved
antibody assays the specificity has
generally been above 97 percent.
FDA believes that at this time,
sufficient data and information exist
such that the risks identified in section
V below can be mitigated by
establishing special controls, and that
these special controls, together with
general controls, are necessary to
provide a reasonable assurance of the
safety and effectiveness of these HBV
assays and therefore proposes these
devices to be reclassified from class III
(general controls and premarket
approval) to class II (general controls
and special controls). In accordance
with section 513(f)(3) of the FD&C Act
and 21 CFR part 860, subpart C, FDA is
proposing to reclassify qualitative HBV
antigen assays, HBV antibody assays
(including qualitative and quantitative
anti-HBs), and quantitative HBV nucleic
acid-based assays from class III into
class II, subject to premarket notification
(510(k)) requirements. FDA believes that
there is sufficient information available
to FDA through FDA’s accumulated
experience with these devices from
reviewing the PMAs for these HBV
assays, and the Panel considerations
and recommendations regarding the
proposed special controls that FDA
believes would effectively mitigate the
risks to health identified in section V.
Absent the special controls identified in
this proposed order, general controls
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
applicable to the devices are insufficient
to provide reasonable assurance of the
safety and effectiveness of the devices.
FDA expects that the reclassification of
these devices would enable more
manufacturers to develop these assays
such that patients would benefit from
increased access to safe and effective
tests.
FDA is proposing to create three
separate classification regulations for
HBV assays that will be reclassified
from class III to class II. HBV assays are
prescription in vitro diagnostic devices,
and under this proposed order, if
finalized, these devices will be
identified as prescription in vitro
diagnostic devices. As such, the devices
must satisfy prescription labeling
requirements for in vitro diagnostic
products (see 21 CFR 809.10(a)(4) and
(b)(5)(ii)). In this proposed order, if
finalized, FDA has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, will provide a reasonable
assurance of the safety and effectiveness
of these assays.
FDA is also proposing to create a new
product code for HBV antibody assays
(including qualitative and quantitative
anti-HBs) that will be assigned upon any
finalization of this proposed order.
Qualitative HBV antigen assays will
continue to be assigned the product
code LOM upon any finalization of this
proposed order.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For these HBV assays, FDA has
determined that premarket notification
is necessary to provide a reasonable
assurance of the safety and effectiveness
of these devices. Therefore, the Agency
does not intend to exempt these
proposed class II devices from 510(k)
requirements. If this proposed order is
finalized, persons who intend to market
a new HBV assay will no longer need to
have a PMA for these devices but can
instead submit to FDA a 510(k) and
receive clearance prior to marketing the
device. A 510(k) typically results in a
shorter premarket review timeline
compared to a PMA, which ultimately
provides more timely access of these
types of devices to patients.
V. Public Health Benefits and Risks to
Health
FDA is providing a substantive
summary of the valid scientific evidence
concerning the public health benefits of
PO 00000
Frm 00011
Fmt 4702
Sfmt 4702
the use of HBV assays (see also https://
www.fda.gov/media/171770/download),
and the nature (and if known, the
incidence) of the risks of the devices
(see further discussion of the special
controls being proposed to mitigate
these risks in section VII of this
proposed order).
HBV infection represents a significant
global public health burden. According
to the World Health Organization
(WHO), in 2019 there were
approximately 296 million people
chronically infected people worldwide,
with 1.5 million new HBV infections
each year.3 It is estimated by the Centers
for Disease Control and Prevention
(CDC) that chronic HBV infection in the
United States affects at least between
580,000 to 1.17 million people with
HBV infection in the United States; twothirds of whom may be unaware of their
infection.4 HBV infection can be
asymptomatic, and accordingly, many
HBV-infected individuals are unaware
of their HBV infection. Approximately
95 percent of adult patients with acute
infection, defined as the first 6 months
after infection, recover completely, and
5 percent of adults develop chronic
HBV.5 Infants born to women who are
HbsAg-positive are at high risk of HBV
infection. In absence of treatment,
infants infected with HBV have a 90
percent risk of progression to chronic
HBV and up to 25 percent of infants
who acquire chronic HBV infection will
die prematurely from HBV-related
hepatocellular carcinoma or cirrhosis.6
Patients who are tested and become
aware that they are HBV infected may
modify risk behaviors to prevent
transmission to others and can be
referred for treatment. Patients with
chronic HBV infection have a risk of
developing liver damage, liver cancer,
or liver failure. They can also spread
their infection to others. HBV can be
reactivated in patients receiving
immunosuppressive therapies, resulting
in serious risk of liver failure or liverassociated death (Ref. 3). HBV is a
vaccine-preventable liver infection.
With the initiation of the WHO Viral
Hepatitis Elimination Plan 7 and the
Department of Health and Human
Services (HHS) Viral Hepatitis National
3 https://www.who.int/news-room/fact-sheets/
detail/hepatitis-b. Accessed on July 12, 2024.
4 Centers for Disease Control and Prevention—
Clinical Overview of Hepatitis B (Available at
https://www.cdc.gov/hepatitis-b/hcp/clinicaloverview/). Accessed on July 12, 2024.
5 Ibid.
6 Ibid.
7 https://www.who.int/health-topics/hepatitis/
elimination-of-hepatitis-by-2030#tab=tab_1.
Accessed on July 12, 2024.
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
Strategic Plan for the United States,8 it
is important for individuals to know
their HBV infected status, to link HBV
infected individuals to care, and to
eliminate virus transmission. Therefore,
diagnosis of patients with HBV infection
through devices such as HBV antibody
and antigen assays is essential to ensure
that patients are linked to the
appropriate care. Current CDC HBV
Screening and Testing
Recommendations include testing of the
following groups: all adults 18 and older
at least once in their lifetime using a
triple panel test, pregnant women
during pregnancy, people who are at
ongoing risk for exposure, and anyone
who requests HBV testing.9
FDA considered our accumulated
experience with the regulation of these
HBV assays, input from the Panel
meeting, and postmarket information
regarding these HBV assays, i.e.,
information from FDA’s publicly
available MDR, Manufacturer and User
Facility Device Experience (MAUDE),
and Medical Device Recall databases.
These HBV assays provide a benefit to
the public health by informing
individuals of their HBV infected status,
linking HBV infected individuals to
appropriate care, and aiding in
eliminating virus transmission. Once an
individual is tested and diagnosed as
HBV infected, HBV nucleic acid testing
is performed to inform treatment
decisions. While HBV infection is
treatable, it is not curable, which means
that most people who start HBV
antiviral treatment must continue it for
life. The goal of current treatment is to
suppress the virus and reduce the
likelihood of long-term complications
and transmission (Refs. 3 and 4). Thus,
identifying individuals who are HBV
infected, linking them to care, and
managing their HBV infection to
alleviate development of liver damage,
liver cancer, liver failure, and potential
HBV transmission would not only
greatly impact public health but also go
a long way towards helping the United
States achieve HBV elimination.
Probable risks to health associated
with the use of HBV assays include risks
related to the risk of false results (false
positives, false negatives, inaccurate low
assay results, inaccurate high assay
results, false reactive results, or false
non-reactive results), failure to correctly
interpret assay results, and failure to
correctly operate the device. For HBV
antigen and antibody assays, false
8 https://www.hhs.gov/sites/default/files/ViralHepatitis-National-Strategic-Plan-2021-2025.pdf.
Accessed on July 12, 2024.
9 https://www.cdc.gov/hepatitis/hbv/index.htm.
Accessed on July 12, 2024.
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
positive results are generally referred to
as false reactive results and false
negative results are generally referred to
as false non-reactive results. False
results can lead to uninfected
individuals receiving unnecessary
further testing and treatment or infected
individuals remaining undiagnosed and
untreated. Undiagnosed and untreated
individuals are likely to experience
increases in morbidity and mortality
and can spread the infection to others.
FDA has identified the following
additional specific risks to health
associated with each of the HBV assays
listed below.
A. Qualitative HBV Antigen Assays
Factors that may cause decreased
assay sensitivity and/or an increased
rate of false non-reactive results include,
but are not limited to, the presence of
interfering substances in the sample,
acute infection at a stage that is too early
for a device to detect the infection, and
antigen concentrations that are too low
to be detected by the device. Factors
that may lead to false reactive results
include device contamination from
reactive samples, cross-reactivity with
other antigens, or misinterpretation of
invalid results as reactive.
• A false reactive assay result for
HbeAg. Incorrectly interpreting the
assay results as a reactive assay result or
failing to correctly operate the assay
causing a false reactive assay result may
lead to continued treatment for hepatitis
B with antiviral medication when it
otherwise would not be indicated.
Antiviral medication has risks including
toxicity and more rarely allergic
reactions. Over time, viral resistance in
patients who are co-infected but
undiagnosed with other viruses that are
treated with the same antiviral
medication, such as HIV, can lead to
viral resistance.
• A false reactive assay result for
HbsAg. Incorrectly interpreting the
assay results as a reactive assay result or
failing to correctly operate the assay
causing a false reactive assay result may
contribute to unnecessary additional
testing, potentially delaying diagnosis of
alternative causes of liver disease when
present and may impact the
psychological well-being of the patient.
Factors that may increase the rate of
false reactive assay reporting include
cross-reactivity with antigens from other
microorganisms or other disease
conditions.
• A false non-reactive result for
HbeAg. Incorrectly interpreting the
assay results as a non-reactive assay
result or failing to correctly operate the
assay causing a false non-reactive assay
result may lead to missing the
PO 00000
Frm 00012
Fmt 4702
Sfmt 4702
78271
opportunity for treatment of an HBV
infected individual with antiviral
medication or premature
discontinuation of antiviral treatment
when continuation of treatment is
otherwise indicated should a clinician
be falsely led to determine a patient has
seroconverted HbeAg to anti-Hbe.
Premature discontinuation of antiviral
medication could result in adverse
effects on patient health, such as
cirrhosis, liver cancer, and liver damage,
all of which are known to contribute to
patient morbidity and mortality, or may
contribute to public health risk by
leading to virus transmission.
• A false non-reactive assay result for
HbsAg. Incorrectly interpreting the
assay results as a non-reactive assay
result or failing to correctly operate the
assay causing a false non-reactive assay
result may delay or prevent a patient
with HBV infection from being
identified and linked to care. Missed
identification of patients with chronic
HBV infection could lead to adverse
effects on patient health such as
cirrhosis, liver cancer, and liver damage,
all of which are known to contribute to
patient morbidity and mortality. A false
non-reactive HbsAg assay incorrectly
interpreted as non-reactive also may
contribute to public health risk by
leading to virus transmission.
B. HBV Antibody Assays (Including
Qualitative and Quantitative Anti-HBs)
Factors that may cause decreased
assay sensitivity and/or an increased
rate of false non-reactive results include,
but are not limited to, the presence of
interfering substances in the sample,
acute infection at a stage that is too early
for a device to detect the infection, and
antibody concentrations that are too low
to be detected by the device. They also
can be caused by misinterpretation of
invalid results as non-reactive. Factors
that may lead to false reactive results
include device contamination from
reactive samples, cross-reactivity with
other antibodies, or misinterpretation of
invalid results as reactive.
• A false reactive assay result for
anti-HBs and anti-HBc. Incorrectly
interpreting the assay results as a
reactive assay result or failing to
correctly operate the assay causing a
false reactive assay result may lead to
improper patient management. A false
reactive antibody assay result could
result in the unnecessary continuation
of antiviral treatment. Antiviral
medication has risks including toxicity
and more rarely allergic reactions. Over
time, viral resistance in patients who are
co-infected but undiagnosed with other
viruses that are treated with the same
antiviral medication, such as HIV, can
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
78272
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
lead to viral resistance. Consequently,
repeatedly false reactive results have the
potential to lead to inappropriate
patient management decisions.
• A false reactive assay result for
anti-HBs. Incorrectly interpreting the
assay results as a reactive assay result or
failing to correctly operate the assay
causing a false reactive assay result
when the device is used as an aid in the
determination of susceptibility to HBV
infection in individuals prior to or
following HBV vaccination or where
vaccination status is unknown may
cause a patient to be considered
previously exposed and therefore
immune to HBV or that the patient was
successfully vaccinated. A false reactive
result may cause the patient to not
receive a vaccine, vaccine booster,
hyperimmune globulin, and would be at
higher risk of infection if exposed to
HBV.
• A false reactive assay result for
anti-Hbe. Incorrectly interpreting the
assay results as a reactive assay result,
or failing to correctly operate the assay
causing a false reactive assay result may
lead to missing the opportunity for
treatment of HBV infection with
antiviral medications in a subset of
individuals for whom treatment would
otherwise be indicated, or premature
discontinuation of antiviral treatment
when continuation of treatment is
otherwise indicated should a clinician
be falsely led to determine a patient has
seroconverted HbeAg to anti-Hbe.
Premature discontinuation of antiviral
medication could result in adverse
effects on patient health such as
cirrhosis, liver cancer, and liver damage,
all of which are known to contribute to
patient morbidity and mortality, or may
contribute to public health risk by
leading to inadvertent transmission of
virus by an infected individual.
• A false non-reactive assay result for
anti-HBc. When the device is used as an
aid in the diagnosis of HBV infection in
patients with symptoms of hepatitis or
who may be at risk for HBV infection,
incorrectly interpreting the assay results
as non-reactive assay result, or failing to
correctly operate the assay causing a
false non-reactive assay result may lead
to non-diagnosis or a delay in diagnosis
of HBV infection with an associated
delay in therapy and potentially
increased risk of HBV-related morbidity
or mortality. Patients with active
infection may unknowingly continue to
infect others. False non-reactive results
can also lead to unnecessary diagnostic
evaluation if alternative etiologies of
hepatitis are pursued. False non-reactive
assay results may occur if the level of
antibody in a specimen is below the
limit of detection of the assay.
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
• A false non-reactive assay result for
anti-HBs. When the device is used as an
aid in the determination of
susceptibility to HBV infection in
individuals prior to or following HBV
vaccination or where vaccination status
is unknown, incorrectly interpreting the
assay results as a non-reactive assay
result or failing to correctly operate the
assay causing a false non-reactive assay
result may lead to unnecessary repeated
vaccination for HBV.
• A false non-reactive assay result for
anti-Hbe. Incorrectly interpreting the
assay results as non-reactive assay result
or failing to correctly operate the assay
causing a false non-reactive assay result
may lead to improper patient
management, including continued
treatment for HBV with antiviral
medication. Antiviral medication has
risks including toxicity and more rarely
allergic reactions. Over time, viral
resistance in patients who are coinfected but undiagnosed with other
viruses using the same antiviral
medication, such as HIV, can lead to
viral resistance.
C. Quantitative HBV Nucleic AcidBased Assays
Decreased assay sensitivity and/or an
increased rate of false negative assay
reporting may occur with patient
samples that contain different genotypes
or rare de novo mutations in HBV
genomic regions targeted by the device.
In these situations, HBV viral load can
transiently decrease and/or become
undetectable in samples before the virus
enters chronic replication.
• A false positive or falsely elevated
quantitative HBV nucleic acid assay
result. Incorrectly interpreting the assay
results as a positive assay result or
failing to correctly operate the assay
causing a false positive assay result may
negatively influence patient
management decisions. Such decisions
may include the administration or
continuation of unnecessary antiviral
treatment in patients with chronic HBV
infection with its known toxicities and
more rarely allergic reactions. Certain
patients with falsely elevated HBV
nucleic acid assay results may not
undergo liver biopsy to investigate other
causes of liver disease when the biopsy
would otherwise be indicated for certain
patients.
• A false negative or falsely decreased
quantitative HBV nucleic acid assay
result. Incorrectly interpreting the assay
results as a negative assay result, or
failing to correctly operate the assay
causing a false negative assay result may
negatively influence patient
management decisions for patients with
chronic HBV infection, including the
PO 00000
Frm 00013
Fmt 4702
Sfmt 4702
withholding of treatment, failure to
treat, or premature discontinuation of
treating HBV infection when antiviral
treatment is otherwise indicated or the
choice of an inappropriate treatment.
This could lead to adverse effects on
patient health such as progressive liver
disease, cirrhosis and/or hepatocellular
carcinoma, and other cancers. Patients
with active HBV replication also risk
spreading the virus to others. Certain
patients with falsely low HBV nucleic
acid assay results may undergo liver
biopsy to investigate other causes of
liver disease.
VI. Summary of Data Upon Which the
Reclassification Is Based
The safety and effectiveness of these
device types has become well
established since the initial approval of
the first qualitative HBV antibody assay
in 2000, the first HBV antigen assay in
2001, and the first quantitative HBV
nucleic acid-based assay in 2008. FDA
has considered and analyzed the
following information: (1) accumulated
experience regulating these HBV assays,
(2) input from the Panel meeting, and
(3) postmarket information regarding
HBV assays, i.e., information from
FDA’s publicly available MDR, MAUDE,
and Medical Device Recall databases.
The available evidence demonstrates
that there are public health benefits
derived from the use of HBV assays
indicated for use to aid in diagnosis of
HBV infection and/or for use to aid in
the management of HBV infected
patients, or as an aid in the
determination of susceptibility to HBV
infection (anti-HBs). In addition, the
nature of the associated risks to health
are known, and special controls can be
established to sufficiently mitigate these
risks.
Based on our review of the
information described above, FDA has
determined that special controls, in
addition to general controls, are
necessary to provide a reasonable
assurance of safety and effectiveness for
HBV assays, and that sufficient
information exists to establish such
special controls. Therefore, FDA, on its
own initiative, is proposing to reclassify
these postamendments devices from
class III (general controls and premarket
approval) into class II (general controls
and special controls), subject to
premarket notification (510(k))
requirements.
VII. Proposed Special Controls
FDA believes that these devices can
be classified into class II with the
establishment of special controls. FDA
believes that the following proposed
special controls would mitigate each of
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
the risks to health described in section
V and that these special controls, in
addition to general controls, would
provide a reasonable assurance of safety
and effectiveness for HBV assays. Tables
1 through 3 below demonstrate how
FDA believes each risk to health
described in section V would be
mitigated by the proposed special
controls for each device type.
A. Qualitative HBV Antigen Assays
The risk of inaccurate interpretation
of assay results can be mitigated by
special controls requiring certain
labeling, including providing clearly
stated warnings and limitations and
information on principles of operation
and procedures in performing the assay.
Risks associated with false results
(e.g., false non-reactive and false
reactive assay results) and with the
failure to correctly operate the device
can be mitigated through a combination
of special controls, including certain
labeling requirements, certain design
verification and validation information,
and performance studies. Examples of
verification and validation information
to be included in the design of the
device include documentation of
performance specifications including
analytical and clinical performance
criteria. In addition, design verification
78273
and validation activities must include
documentation of a complete device
description, critical reagents, risk
analysis strategies, lot release criteria,
stability studies, and protocols.
Required statements in labeling can aid
in mitigating the failure of the device to
perform as indicated, for example
including a statement that use of the
assay with specimen types other than
those specifically identified for use with
this device may cause inaccurate assay
results. Special controls requiring
additional labeling to provide a brief
summary of the instructions for use can
also mitigate these risks.
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR QUALITATIVE HBV ANTIGEN ASSAYS
Identified risks to health
Mitigation measures
False reactive/non-reactive assay result ......................................
Failure to correctly interpret the assay results .............................
Failure to correctly operate the device .........................................
B. HBV Antibody Assays (Including
Qualitative and Quantitative Anti-HBs)
The risk of falsely reactive, nonreactive, elevated, or lowered assay
results can be mitigated by special
controls requiring certain labeling,
including providing clearly stated
warnings and limitations and
information on principles of operation
and procedures in performing the assay.
Risks associated with the failure of
the device to perform as indicated (e.g.,
Certain labeling information, including limitations, explanation of procedures,
and results interpretation information.
Certain design verification and validation information, including certain device
description information, risk analysis strategies, lot release criteria, stability
studies and protocols, and performance criteria including analytical studies
and clinical studies.
Certain labeling information, including warnings, limitations, results interpretation information, and explanation of procedures.
Certain design verification and validation information, including certain device
description information, critical reagent information, risk analysis strategies,
lot release criteria, and stability studies and protocols.
Certain labeling information, including warnings, limitations, results interpretation information, and explanation of procedures.
Certain design verification and validation information, including certain device
description, critical reagent information, risk analysis strategies, lot release
criteria, and stability studies and protocols.
false non-reactive and false reactive
assay results) can be mitigated through
a combination of special controls,
including certain labeling requirements,
certain design verification and
validation information, and
performance studies. Examples of
verification and validation information
to be included in the design of the
device include documentation of
performance specifications including
analytical and clinical performance
criteria. In addition, design verification
and validation activities must include
documentation of a complete device
description, critical reagents, risk
analysis strategies, lot release criteria,
stability studies, and protocols.
Required statements in labeling can aid
in mitigating the failure of the device to
perform as indicated; for example,
including a statement that use of the
assay with specimen types other than
those specifically identified for use with
this device may cause inaccurate assay
results.
TABLE 2—RISKS TO HEALTH AND MITIGATION MEASURES FOR HBV ANTIBODY ASSAYS (INCLUDING QUALITATIVE AND
QUANTITATIVE ANTI-HBS)
ddrumheller on DSK120RN23PROD with PROPOSALS1
Identified risks to health
Mitigation measures
False reactive/false non-reactive assay result. In addition, for
quantitative assays: Falsely elevated/falsely lowered assay
result.
Failure to correctly interpret the assay results .............................
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
PO 00000
Frm 00014
Certain labeling information, including limitations, explanation of procedures,
and results interpretation information.
Certain design verification and validation information including certain device
description information, risk analysis strategies, lot release criteria, stability
studies and protocols, and performance criteria including analytical studies
and clinical studies.
Certain labeling information, including warnings, limitations, results interpretation information, and explanation of procedures.
Certain design verification and validation information including certain device
description, critical reagent information, risk analysis strategies, lot release
criteria, and stability studies and protocols.
Fmt 4702
Sfmt 4702
E:\FR\FM\25SEP1.SGM
25SEP1
78274
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
TABLE 2—RISKS TO HEALTH AND MITIGATION MEASURES FOR HBV ANTIBODY ASSAYS (INCLUDING QUALITATIVE AND
QUANTITATIVE ANTI-HBS)—Continued
Identified risks to health
Mitigation measures
Failure to correctly operate the devices .......................................
C. Quantitative HBV Nucleic AcidBased Assays
The risk of falsely positive, negative,
elevated, or lowered assay results can be
mitigated by special controls requiring
certain labeling, including providing
clearly stated warnings and limitations,
device description information, and
detailed instructions in the device
labeling regarding the interpretation of
assay results and principles of operation
and procedures in performing the assay.
Risks associated with the failure of
the device to perform as indicated (e.g.,
Certain labeling information, warnings, limitations, results interpretation information, and explanation of procedures.
Certain design verification and validation information including certain device
description, critical reagent information, risk analysis strategies, lot release
criteria, and stability studies and protocols.
inaccurately low or high results, false
negative results, and false positive assay
results) can be mitigated through a
combination of special controls related
to certain labeling requirements, design
verification and validation activities,
and performance studies. Examples of
verification and validation information
to be included in the design of the
device include documentation of a
complete device description, calibrators,
critical reagents, traceability, and lot
release criteria. In addition, design
verification and validation must include
documentation of performance
specifications, including analytical and
clinical performance criteria. Required
statements in labeling can aid in
mitigating the occurrence of inaccurate
results. The risks of false positive/false
negative/falsely elevated/falsely
lowered results due to decreased assay
sensitivity can be mitigated by special
controls related to certain labeling,
design verification and validation
activities, risk analysis strategies, and
performance studies.
TABLE 3—RISKS TO HEALTH AND MITIGATION MEASURES FOR QUANTITATIVE HBV NUCLEIC ACID-BASED ASSAYS
Identified risks to health
Mitigation measures
False positive/false negative/falsely elevated/falsely lowered result.
Failure to correctly interpret the assay results .............................
ddrumheller on DSK120RN23PROD with PROPOSALS1
Failure to correctly operate the device .........................................
If this proposed order is finalized,
qualitative HBV antigen assays, HBV
antibody assays (including qualitative
and quantitative anti-HBs), and
quantitative HBV nucleic acid-based
assays will be reclassified into class II
(general controls and special controls)
and would be subject to premarket
notification requirements under section
510(k) of the FD&C Act. Firms
submitting a 510(k) of the FD&C Act for
such devices will be required to comply
with the particular mitigation measures
set forth in the special controls. FDA
believes that adherence to the special
controls, in addition to the general
controls, is necessary to provide a
reasonable assurance of safety and
effectiveness of HBV assays.
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
Certain labeling information, including limitations, explanation of procedures,
and results interpretation information.
Certain design verification and validation information, including certain device
description information, risk analysis strategies, lot release criteria, stability
studies and protocols, and performance criteria including analytical studies
and clinical studies.
Certain labeling information, including warnings, limitations, results interpretation information, and explanation of procedures.
Certain design verification and validation information, including certain device
description, critical reagent information, risk analysis strategies, lot release
criteria, and stability studies and protocols.
Certain labeling warnings, limitations, results interpretation information, and
explanation of procedures.
Certain design verification and validation information including certain device
description, critical reagent information, risk analysis strategies, lot release
criteria, and stability studies and protocols.
VIII. Analysis of Environmental Impact
We have determined under 21 CFR
25.34(b) that this action is of a type that
does not individually or cumulatively
have a significant effect on the human
environment. Therefore, neither an
environmental assessment nor an
environmental impact statement is
required.
21 CFR part 820 have been approved
under OMB control number 0910–0073;
the collections of information in part
807, subpart E, have been approved
under OMB control number 0910–0120;
and the collections of information in 21
CFR parts 801 and 809 have been
approved under OMB control number
0910–0485.
IX. Paperwork Reduction Act of 1995
While this proposed order contains no
new collections of information, it does
refer to previously approved FDA
collections of information. The
previously approved FDA collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (PRA) (44 U.S.C. 3501–
3521). The collections of information in
X. Proposed Effective Date
PO 00000
Frm 00015
Fmt 4702
Sfmt 4702
FDA proposes that any final order
based on this proposed order become
effective 30 days after the date of its
publication in the Federal Register.
XI. Codification of Orders
Under section 513(f)(3) of the FD&C
Act, FDA may issue final orders to
reclassify devices. FDA will continue to
codify classifications and
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
reclassifications in the Code of Federal
Regulations (CFR). Changes resulting
from final orders will appear in the CFR
as newly codified orders. Therefore,
under section 513(f)(3) of the FD&C Act,
in the proposed order, we are proposing
to codify qualitative hepatitis B virus
antigen assays in the new § 866.3178,
hepatitis B virus antibody assays
(including qualitative and quantitative
anti-HBs) in the new § 866.3179, and
quantitative hepatitis B virus nucleic
acid-based assays in the new § 866.3180,
under which these HBV assays would
be reclassified from class III into class
II.
XII. References
The following references marked with
an asterisk (*) are on display at the
Dockets Management Staff (see
ADDRESSES) and are available for
viewing by interested persons between
9 a.m. and 4 p.m., Monday through
Friday; they also are available
electronically at https://
www.regulations.gov. References
without asterisks are not on public
display at https://www.regulations.gov
because they have copyright restriction.
Some may be available at the website
address, if listed. References without
asterisks are available for viewing only
at the Dockets Management Staff.
Although FDA verified the website
addresses in this document, please note
that websites are subject to change over
time.
ddrumheller on DSK120RN23PROD with PROPOSALS1
*1. Summary Minutes Prepared for the
September 7, 2023, Meeting of the
Microbiology Devices Panel (available at
https://www.fda.gov/media/173610/
download).
*2. Meeting Transcript Prepared for the
September 7, 2023, Meeting of the
Microbiology Devices Panel (available at
https://www.fda.gov/media/173609/
download).
3. Terrault, N.A., A.S.F. Lok, B.J. McMahon,
et al., ‘‘Update on Prevention, Diagnosis,
and Treatment of Chronic Hepatitis B:
AASLD 2018 Hepatitis B Guidance.’’
Hepatology, 67(4): 1560–1599, 2018.
4. CDC, ‘‘Clinical Testing and Diagnosis for
Hepatitis B,’’ https://www.cdc.gov/
hepatitis-b/hcp/diagnosis-testing/
index.html. Accessed July 11, 2024.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical
devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 866 be amended as follows:
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
PART 866—IMMUNOLOGY AND
MICROBIOLOGY DEVICES
1. The authority citation for part 866
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 866.3178 to subpart D to read
as follows:
■
§ 866.3178 Qualitative hepatitis B virus
antigen assays.
(a) Identification. A qualitative
hepatitis B virus (HBV) antigen assay is
identified as an in vitro diagnostic
device intended for prescription use for
qualitative use with human serum,
plasma, or other matrices that aids in
the diagnosis of chronic or acute HBV
infection. HBV surface antigen (HbsAg)
is also used for screening of HBV
infection in pregnant women to identify
neonates who are at risk of acquiring
hepatitis B during perinatal period. The
assay is not intended for screening of
blood, plasma, cells, or tissue donors.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The labeling required under
§ 809.10(b) of this chapter must include:
(i) A prominent statement that the
assay is not intended for the screening
of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the
principles of operation and procedures
for performing the assay.
(iii) A detailed explanation of the
interpretation of results.
(iv) Limitations, which must be
updated to reflect current clinical
practice and disease presentation and
management. The limitations must
include statements that indicate:
(A) The specimen types for which the
device has been cleared, and that use of
this assay with specimen types other
than those specifically cleared for this
device may result in inaccurate assay
results.
(B) When appropriate, performance
characteristics of the assay have not
been established in populations of
immunocompromised or
immunosuppressed patients or other
populations where assay performance
may be affected.
(C) Diagnosis of hepatitis B infection
should not be established on the basis
of a single assay result but should be
determined by a licensed healthcare
professional in conjunction with the
clinical presentation, history, and other
diagnostic procedures.
(D) Detection of HBV antigens
indicates a current infection with
hepatitis B virus but does not
differentiate between acute or chronic
PO 00000
Frm 00016
Fmt 4702
Sfmt 4702
78275
infection. False reactive HbsAg result
may occur for up to 2 weeks after
vaccination with HbsAg containing
vaccine.
(E) Current methods for the detection
of hepatitis B antigens may not detect
all potentially infected individuals. A
non-reactive assay result does not
exclude the possibility of exposure to or
infection with hepatitis B virus. A nonreactive assay result in individuals with
prior exposure to hepatitis B may be due
to but not limited to antigen levels
below the detection limit of this assay
or lack of antigen reactivity to the
antibodies in this assay. HBV mutants
lacking the ability to produce antigens
have been reported. These may occur as
‘‘escape’’ mutants in the presence of
anti-HBV antibodies and such patients
may be infectious.
(F) Results obtained with this assay
may not be used interchangeably with
results obtained with a different
manufacturer’s assay.
(2) Design verification and validation
must include the following:
(i) A detailed device description,
including all parts that make up the
device, ancillary reagents required but
not provided, an explanation of the
device methodology, design of the
capture antibody(ies), external controls,
and computational path from collected
raw data to reported result (e.g., how
collected raw signals are converted into
a reported signal and result), as
applicable to the detection method and
device design.
(ii) For devices with assay calibrators,
the design and composition of all
primary, secondary, and subsequent
quantitation standards used for
calibration as well as their traceability
to a standardized reference material that
FDA has determined is appropriate (e.g.,
a recognized consensus standard). In
addition, analytical testing must be
performed following the release of a
new lot of the standard material that
was used for device clearance or
approval, or when there is a transition
to a new calibration standard.
(iii) Documentation and
characterization (e.g., supplier,
determination of identity, purity, and
stability) of all critical reagents
(including description of the capture
antibody(ies)), and protocols for
maintaining product integrity
throughout its labeled shelf life.
(iv) Risk analysis and management
strategies, such as Failure Modes Effects
Analysis and/or Hazard Analysis and
Critical Control Points summaries and
their impact on assay performance.
(v) Final release criteria to be used for
manufactured assay lots with
appropriate evidence that lots released
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
78276
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
at the extremes of the specifications will
meet the identified analytical and
clinical performance characteristics as
well as stability.
(vi) Stability studies for reagents must
include documentation of an assessment
of real-time stability for multiple reagent
lots using the indicated specimen types
and must use acceptance criteria that
ensure that analytical and clinical
performance characteristics are met
when stability is assigned based on the
extremes of the acceptance range.
(vii) All stability protocols, including
acceptance criteria.
(viii) Final release assay results for
each lot used in clinical studies.
(ix) Reproducibility study data that
includes the testing of three
independent production lots.
(x) Detailed documentation of
analytical performance studies
conducted, as appropriate to the
technology, specimen types tested, and
intended use of the device, including,
the limit of blank (LoB), limit of
detection (LoD), cutoff, precision
(reproducibility) including lot-to-lot
and/or instrument-to-instrument
precision, interference, cross reactivity,
carryover, hook effect, seroconversion
panel testing, matrix equivalency,
prominent mutants/variants detection
(e.g., for HbsAg), specimen stability,
reagent stability, and cross-genotype
antigen detection sensitivity, when
appropriate.
(xi) Analytical sensitivity of the assay
is the same or better than that of other
cleared or approved assays.
(xii) For devices with associated
software or instrumentation,
documentation must include a detailed
description of device software,
including software applications and
hardware-based devices that incorporate
software. The detailed description must
include documentation of verification,
validation, and hazard analysis and risk
assessment activities, including an
assessment of the impact of threats and
vulnerabilities on device functionality
and end users/patients as part of
cybersecurity review.
(xiii) Detailed documentation and
results from a clinical study.
Performance must be analyzed relative
to an FDA cleared or approved HBV
antigen assay or a comparator that FDA
has determined is appropriate. This
study must be conducted using
appropriate patient samples, with an
appropriate number of HBV reactive and
non-reactive samples in applicable risk
and disease categories, and any
applicable confirmatory testing.
Additional relevant patient groups must
be validated as appropriate. The
samples must include prospective
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
(sequential) samples for each identified
specimen type and, as appropriate,
additional characterized clinical
samples. Samples must be sourced from
geographically diverse areas. This study
must be conducted in the appropriate
settings by the intended users to
demonstrate clinical performance.
■ 3. Add § 866.3179 to subpart D to read
as follows:
§ 866.3179 Hepatitis B virus antibody
assays (including qualitative and
quantitative anti-HBs).
(a) Identification. A hepatitis B virus
(HBV) antibody assay is identified as an
in vitro diagnostic device intended for
prescription use in the detection of
antibodies to HBV in human serum,
plasma, or other matrices, and as a
device that aids in the diagnosis of HBV
infection in persons with signs and
symptoms of hepatitis and in persons at
risk for hepatitis B infection. In
addition, results from an anti-HBc IgM
(IgM antibodies to core antigen) assay
indicating the presence of anti-HBc IgM
are indicative of recent HBV infection.
Anti-HBs (antibodies to surface antigen)
assay results may be used as an aid in
the determination of susceptibility to
HBV infection in individuals prior to or
following HBV vaccination or when
vaccination status is unknown. The
assay is not intended for screening of
blood, plasma, cells, or tissue donors.
The assay is intended as an aid in
diagnosis in conjunction with clinical
findings and other diagnostic
procedures.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The labeling required under
§ 809.10(b) of this chapter must include:
(i) A prominent statement that the
assay is not intended for the screening
of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the
principles of operation and procedures
for performing the assay.
(iii) A detailed explanation of the
interpretation of results.
(iv) Limitations, which must be
updated to reflect current clinical
practice and disease presentation and
management. The limitations must
include statements that indicate:
(A) When appropriate, performance
characteristics of the assay have not
been established in populations of
immunocompromised or
immunosuppressed patients or other
special populations where assay
performance may be affected.
(B) Detection of HBV antibodies to a
single viral antigen indicates a present
or past infection with hepatitis B virus,
PO 00000
Frm 00017
Fmt 4702
Sfmt 4702
but does not differentiate between acute,
chronic, or resolved infection.
(C) The specimen types for which the
device has been cleared, and that use of
the assay with specimen types other
than those specifically cleared for this
device may result in inaccurate assay
results.
(D) Diagnosis of hepatitis B infection
should not be established on the basis
of a single assay result but should be
determined by a licensed healthcare
professional in conjunction with the
clinical presentation, history, and other
diagnostic procedures.
(E) A non-reactive assay result may
occur early during acute infection, prior
to development of a host antibody
response to infection, or when analyte
levels are below the limit of detection of
the assay.
(F) Results obtained with this assay
may not be used interchangeably with
results obtained with a different
manufacturer’s assay.
(v) For devices intended for the
quantitative detection of HBV
antibodies (anti-HBs), in addition to the
special controls listed in paragraphs
(b)(1) and (2) of this section, labeling
required under § 809.10(b) of this
chapter must include:
(A) The assay calibrators’ traceability
to a standardized reference material that
FDA has determined is appropriate (e.g.,
a recognized consensus standard) and
the limit of blank (LoB), limit of
detection (LoD), limit of quantitation
(LoQ), linearity, and precision to define
the analytical measuring interval.
(B) Performance results of the
analytical sensitivity study testing a
standardized reference material that
FDA has determined is appropriate (e.g.,
a recognized consensus standard).
(2) Design verification and validation
must include the following:
(i) Detailed device description,
including all parts that make up the
device, ancillary reagents required but
not provided, an explanation of the
device methodology, and design of the
antigen(s) and capture antibody(ies)
sequences, rationale for the selected
epitope(s), degree of amino acid
sequence conservation of the target, and
the design and composition of all
primary, secondary and subsequent
standards used for calibration.
(ii) Documentation and
characterization (e.g., supplier,
determination of identity, and stability)
of all critical reagents (including
description of the antigen(s) and capture
antibody(ies)), and protocols for
maintaining product integrity
throughout its labeled shelf life.
(iii) Risk analysis and management
strategies, such as Failure Modes Effects
E:\FR\FM\25SEP1.SGM
25SEP1
ddrumheller on DSK120RN23PROD with PROPOSALS1
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
Analysis and/or Hazard Analysis and
Critical Control Points summaries and
their impact on assay performance.
(iv) Final release criteria to be used
for manufactured assay lots with
appropriate evidence that lots released
at the extremes of the specifications will
meet the identified analytical and
clinical performance characteristics as
well as stability.
(v) Stability studies for reagents must
include documentation of an assessment
of real-time stability for multiple reagent
lots using the indicated specimen types
and must use acceptance criteria that
ensure that analytical and clinical
performance characteristics are met
when stability is assigned based on the
extremes of the acceptance range.
(vi) All stability protocols, including
acceptance criteria.
(vii) When applicable, analytical
sensitivity of the assay is the same or
better than that of other cleared or
approved assays.
(viii) Analytical performance studies
and results for determining the limit of
blank (LoB), limit of detection (LoD),
cutoff, precision (reproducibility),
including lot-to-lot and/or instrumentto-instrument precision, interference,
cross reactivity, carryover, hook effect,
seroconversion panel testing, matrix
equivalency, specimen stability, reagent
stability, and cross-genotype antibody
detection sensitivity, when appropriate.
(ix) For devices intended for the
detection of antibodies for which a
standardized reference material (that
FDA has determined is appropriate) is
available, the analytical sensitivity
study and results testing the
standardized reference material.
Detailed documentation of that study
and its results must be provided,
including the study protocol, study
report, testing results, and all statistical
analyses.
(x) For devices with associated
software or instrumentation,
documentation must include a detailed
description of device software,
including software applications and
hardware-based devices that incorporate
software. The detailed description must
include documentation of verification,
validation, and hazard analysis and risk
assessment activities, including an
assessment of the impact of threats and
vulnerabilities on device functionality
and end users/patients as part of
cybersecurity review.
(xi) Detailed documentation of
clinical performance testing from a
clinical study with an appropriate
number of HBV reactive and nonreactive samples in applicable risk
categories and conducted in the
appropriate settings by the intended
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
users. Performance must be analyzed
relative to an FDA cleared or approved
HBV antibody assay or a comparator
that FDA has determined is appropriate.
Additional relevant patient groups must
be validated as appropriate. The
samples must include prospective
(sequential) samples for each identified
specimen type and, as appropriate,
additional characterized clinical
samples. Samples must be sourced from
geographically diverse areas.
(3) For any HBV antibody assay
intended for quantitative detection of
anti-HBV antibodies, the following
special controls, in addition to those
special controls listed in paragraphs
(b)(1) and (2) of this section, also apply:
(i) Detailed documentation of the
metrological calibration traceability
hierarchy to a standardized reference
material that FDA has determined is
appropriate.
(ii) Detailed documentation of the
following analytical performance
studies conducted, as appropriate to the
technology, specimen types tested, and
intended use of the device, including
upper and lower limits of quantitation
(UloQ and LloQ, respectively), linearity
using clinical samples, and an accuracy
study using the recognized international
standard material.
4. Add § 866.3180 to subpart D to read
as follows:
§ 866.3180 Hepatitis B virus nucleic acidbased assays.
(a) Identification. A nucleic acidbased hepatitis B virus (HBV) assay is
identified as an in vitro diagnostic
device intended for prescription use in
the detection of HBV nucleic acid in
specimens from individuals with
antibody evidence of HBV infection. In
these devices, the detection of HBV
nucleic acid is used as an aid in the
management of HBV-infected
individuals. The assay is intended for
use with human serum or plasma (and
other matrices as applicable) from
individuals with HBV. The assay is not
intended for use as a donor screening
assay for the presence of HBV nucleic
acids in blood, blood products, plasma,
cells, or tissue donors, or as a diagnostic
assay to confirm the presence of HBV
infection.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Labeling required under
§ 809.10(b) of this chapter must include:
(i) A prominent statement that the
assay is not intended for use as a
screening assay for the presence of HBV
DNA in blood or blood products,
plasma, cells, or tissue donors, or as a
PO 00000
Frm 00018
Fmt 4702
Sfmt 4702
78277
diagnostic assay to confirm the presence
of HBV infection.
(ii) A detailed explanation of the
principles of operation and procedures
for performing the assay.
(iii) A detailed explanation of the
interpretation of results.
(iv) Limitations, which must be
updated to reflect current clinical
practice and disease presentation and/or
management. These limitations must
include statements that indicate:
(A) Management of patients
undergoing hepatitis B virus treatment
should not be established on the basis
of a single assay result but should be
determined by a licensed healthcare
professional in conjunction with the
clinical presentation, history, and other
diagnostic procedures, e.g., HBV
serologic testing, liver function assays,
liver elastography, etc.
(B) The specimen types for which the
device has been cleared, and that use of
this assay with specimen types other
than those specifically cleared for this
device may result in inaccurate assay
results.
(C) The results obtained with this
assay may not be used interchangeably
with results obtained with a different
manufacturer’s assay.
(2) Design verification and validation
must include the following:
(i) Detailed device description,
including the device components,
ancillary reagents required but not
provided, and an explanation of the
device methodology. Additional
information appropriate to the
technology must be included such as
design of primers and probes, rationale
for the selected gene targets,
specifications for amplicon size, and
degree of nucleic acid sequence
conservation.
(ii) For devices with assay calibrators,
the design and composition of all
primary, secondary, and subsequent
quantitation standards used for
calibration as well as their traceability
to a standardized reference material that
FDA has determined is appropriate (e.g.,
a recognized consensus standard). In
addition, analytical testing must be
performed following the release of a
new lot of the standard material that
was used for device clearance or
approval, or when there is a transition
to a new calibration standard.
(iii) Documentation and
characterization (e.g., determination of
the identity, supplier, purity, and
stability) of all critical reagents
(including nucleic acid sequences for
primers and probes) and protocols for
maintaining product integrity.
(iv) Risk analysis and management
strategies demonstrating how risk
E:\FR\FM\25SEP1.SGM
25SEP1
78278
Federal Register / Vol. 89, No. 186 / Wednesday, September 25, 2024 / Proposed Rules
ddrumheller on DSK120RN23PROD with PROPOSALS1
control measures are implemented to
address device system hazards, such as
Failure Modes Effects Analysis and/or
Hazard Analysis and Critical Control
Points summaries and their impact on
assay performance.
(v) Final release criteria to be used for
manufactured assay lots with
appropriate evidence that lots released
at the extremes of the specification will
meet the identified analytical and
clinical performance characteristics as
well as stability.
(vi) Stability studies for reagents must
include documentation of an assessment
of real-time stability for multiple reagent
lots using the indicated specimen types
and must use acceptance criteria that
ensure that analytical and clinical
performance characteristics are met
when stability is assigned based on the
extremes of the acceptance range.
(vii) All stability protocols, including
acceptance criteria.
(viii) Detailed documentation of
analytical performance studies
conducted as appropriate to the
technology, specimen types tested, and
intended use of the device, including
limit of detection (LoD), linearity,
precision, endogenous and exogenous
interferences, cross-reactivity, carryover,
matrix equivalency, sample and
reagents stability, and as applicable,
upper and lower limits of quantitation
(ULoQ and LLoQ, respectively).
Samples selected for use must be from
subjects with clinically relevant
circulating genotypes in the United
States. Cross-reactivity studies must
include samples from HBV nucleic acid
negative subjects with other viral or
non-viral causes of liver disease,
including autoimmune hepatitis,
alcoholic liver disease, chronic hepatitis
C virus (HCV), primary biliary cirrhosis,
and nonalcoholic steatohepatitis, when
applicable. The effect of each identified
nucleic-acid isolation and purification
procedure on detection must be
evaluated.
(ix) For devices with associated
software or instrumentation,
documentation must include a detailed
description of device software,
including software applications and
hardware-based devices that incorporate
software. The detailed description must
include documentation of verification,
VerDate Sep<11>2014
16:25 Sep 24, 2024
Jkt 262001
validation, and hazard analysis and risk
assessment activities, including an
assessment of the impact of threats and
vulnerabilities on device functionality
and end users/patients as part of
cybersecurity review.
(x) Detailed documentation of
performance from a clinical study with
a design and number of clinical samples
(appropriately statistically powered)
that is appropriate for the intended use
of the device as well as conducted in the
appropriate settings by the intended
users. The samples must include
prospective (sequential) samples for
each claimed specimen type and, as
appropriate, additional characterized
clinical samples. Samples must be
sourced from geographically diverse
areas.
Dated: September 20, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024–21932 Filed 9–24–24; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF STATE
22 CFR Parts 120 and 121
[Public Notice: 12552; DOS–2024–0023]
RIN 1400–AF29
International Traffic in Arms
Regulations: Revisions to Definition
and Controls Related to Defense
Services; Extension of Comment
Period
Department of State.
Proposed rule; extension of
comment period.
AGENCY:
ACTION:
The Department of State is
extending the comment period for a
proposed rule published July 29, 2024.
The original comment period required
submission of comments on or before
September 27, 2024. In response to
requests from the public, the
Department extends the comment
period through October 15, 2024.
DATES: The comment period for the
proposed rule published July 29, 2024,
at 89 FR 60980, is extended. Comments
should be received on or before October
15, 2024.
SUMMARY:
PO 00000
Frm 00019
Fmt 4702
Sfmt 9990
Interested parties may
submit comments by one of the
following methods:
• Email: DDTCPublicComments@
state.gov with the subject line:
‘‘Regulatory Change: Defense Service
Definition’’.
• Internet: At www.regulations.gov,
search for this notice, by docket number
DOS–2024–0023. Additional
instructions regarding submission of
comments can be found in the
document published at 89 FR 60980,
July 29, 2024.
ADDRESSES:
FOR FURTHER INFORMATION CONTACT:
Sarah Heidema, Director, Office of
Defense Trade Controls Policy,
Department of State, telephone (202)
663–1282; email
DDTCCustomerService@state.gov.
ATTN: Revisions to Definition and
Controls Related to Defense Services.
On July
29, 2024, the Department of State
published a proposed rule proposing
revisions to the definition of defense
service at 22 CFR 120.32 of the
International Traffic in Arms
Regulations (22 CFR parts 120 through
130) and to the United States Munitions
List at 22 CFR 121.1 (89 FR 60980). On
the same day, the Department of
Commerce published a complementary
proposed rule proposing changes to
existing restrictions under the Export
Administration Regulations (15 CFR
parts 730 through 744) on military and
intelligence end uses and end users and
related controls on certain activities of
U.S. persons, as well as the proposed
addition of a military-support end-user
control (89 FR 60985). In response to
requests from the public received by the
Department of Commerce, and due to
their plan to extend the comment period
for their complementary proposed rule
for 15 more days, as published via
separate notice, the Department of State
is similarly extending the comment
period for its proposed rule for 15 days.
SUPPLEMENTARY INFORMATION:
Stanley L. Brown,
Acting Assistant Secretary, Bureau of
Political-Military Affairs, Department of
State.
[FR Doc. 2024–22041 Filed 9–23–24; 4:15 pm]
BILLING CODE 4710–25–P
E:\FR\FM\25SEP1.SGM
25SEP1
Agencies
[Federal Register Volume 89, Number 186 (Wednesday, September 25, 2024)]
[Proposed Rules]
[Pages 78265-78278]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2024-21932]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA-2024-N-3533]
Microbiology Devices; Reclassification of Antigen, Antibody, and
Nucleic Acid-Based Hepatitis B Virus Assay Devices
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed amendment; proposed order; request for comments.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, the Agency, or we) is
proposing to reclassify qualitative hepatitis B virus (HBV) antigen
assays, qualitative HBV antibody assays and quantitative assays that
detect anti-HBs (antibodies to HBV surface antigen (HBsAg)), and
quantitative HBV nucleic acid-based assays, all of which are
postamendments class III devices, into class II (general controls and
special controls), subject to premarket notification. FDA is also
proposing three new device classification regulations along with the
special controls that the Agency believes are necessary to provide a
reasonable assurance of safety and effectiveness for each device.
DATES: Either electronic or written comments on the proposed order must
be submitted by November 25, 2024. Please see section X of this
document for the proposed effective date when the new requirements
apply and for the proposed effective date of a final order based on
this proposed order.
ADDRESSES: You may submit comments as follows. Please note that late,
untimely filed comments will not be considered. The https://www.regulations.gov electronic filing system will accept comments until
11:59 p.m. Eastern Time at the end of November 25, 2024. Comments
received by mail/hand delivery/courier (for written/paper submissions)
will be considered timely if they are received on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal Rulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
[[Page 78266]]
Mail/Hand Delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2024-N-3533 for ``Microbiology Devices; Reclassification of
Antigen, Antibody, and Nucleic Acid-Based Hepatitis B Virus Assay
Devices.'' Received comments, those filed in a timely manner (see
ADDRESSES), will be placed in the docket and, except for those
submitted as ``Confidential Submissions,'' publicly viewable at https://www.regulations.gov or at the Dockets Management Staff between 9 a.m.
and 4 p.m., Monday through Friday Eastern Time, 240-402-7500.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.govinfo.gov/content/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents, the
plain language summary of the proposed order of not more than 100 words
consistent with the ``Providing Accountability Through Transparency
Act,'' or the electronic and written/paper comments received, go to
https://www.regulations.gov and insert the docket number, found in
brackets in the heading of this document, into the ``Search'' box and
follow the prompts and/or go to the Dockets Management Staff, 5630
Fishers Lane, Rm. 1061, Rockville, MD 20852, 240-402-7500.
FOR FURTHER INFORMATION CONTACT: Maria Ines Garcia, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 3104, Silver Spring, MD 20993, 301-796-
7017, [email protected].
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended,
establishes a comprehensive system for the regulation of medical
devices intended for human use. Section 513 of the FD&C Act (21 U.S.C.
360c) established three categories (classes) of devices, reflecting the
regulatory controls needed to provide reasonable assurance of their
safety and effectiveness. The three categories of devices are class I
(general controls), class II (general controls and special controls),
and class III (general controls and premarket approval).
Section 513(a)(1) of the FD&C Act defines the three classes of
devices. Class I devices are those devices for which the general
controls of the FD&C Act (controls authorized by or under sections 501,
502, 510, 516, 518, 519, or 520 (21 U.S.C. 351, 352, 360, 360f, 360h,
360i, or 360j) or any combination of such sections) are sufficient to
provide reasonable assurance of safety and effectiveness; or those
devices for which insufficient information exists to determine that
general controls are sufficient to provide reasonable assurance of
safety and effectiveness or to establish special controls to provide
such assurance, but because the devices are not purported or
represented to be for a use in supporting or sustaining human life or
for a use which is of substantial importance in preventing impairment
of human health, and do not present a potential unreasonable risk of
illness or injury, are to be regulated by general controls (section
513(a)(1)(A) of the FD&C Act). Class II devices are those devices for
which general controls by themselves are insufficient to provide
reasonable assurance of safety and effectiveness, and for which there
is sufficient information to establish special controls to provide such
assurance, including the issue of performance standards, postmarket
surveillance, patient registries, development and dissemination of
guidelines, recommendations, and other appropriate actions the Agency
deems necessary to provide such assurance (section 513(a)(1)(B) of the
FD&C Act). Class III devices are those devices for which insufficient
information exists to determine that general controls and special
controls would provide a reasonable assurance of safety and
effectiveness, and are purported or represented to be for a use in
supporting or sustaining human life or for a use which is of
substantial importance in preventing impairment of human health, or
present a potential unreasonable risk of illness or injury (section
513(a)(1)(C) of the FD&C Act).
Devices that were not in commercial distribution before May 28,
1976 (generally referred to as ``postamendments devices'') are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval, unless, and until: (1) FDA
reclassifies the device into class I or II, or (2) FDA issues an order
finding the device to be substantially equivalent, in accordance with
section 513(i) of the FD&C Act, to a predicate device that does not
require premarket approval. The Agency determines whether new devices
are substantially equivalent to predicate devices by means of the
premarket notification procedures in section 510(k) of the FD&C Act and
part 807, subpart E (21 CFR part 807, subpart E) of FDA's regulations.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or class II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA, acting by administrative
order, can reclassify the device into class I or class II on its own
initiative, or in response to a petition from the manufacturer or
importer of the device. To change the classification of the device, the
proposed new class must have sufficient regulatory controls to provide
reasonable assurance of the safety and effectiveness of the device for
its intended use.
FDA relies upon ``valid scientific evidence'', as defined in
section 513(a)(3) of the FD&C Act and 21 CFR 860.7(c)(2), in the
classification process to determine the level of regulation for
devices. To be considered in the reclassification process, the ``valid
scientific evidence'' upon which the Agency relies must be publicly
available (see section 520(c) of the FD&C Act).
[[Page 78267]]
Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
premarket approval application (PMA) (see section 520(c) of the FD&C
Act).
In accordance with section 513(f)(3) of the FD&C Act, FDA is
issuing this proposed order to reclassify qualitative HBV antigen
assays intended for qualitative detection of HBV antigens as an aid in
the diagnosis of acute or chronic HBV infection in specific
populations, HBV antibody assays (including qualitative and
quantitative anti-HBs) intended for use in the detection of antibodies
to HBV, and quantitative HBV nucleic acid-based assays intended for use
in the detection of HBV nucleic acid in specimens from individuals with
antibody evidence of HBV infection, all of which are postamendments
class III devices, into class II (general controls and special
controls) subject to premarket notification, under three new device
classification regulations with the names ``Qualitative Hepatitis B
Virus Antigen Assays,'' ``Hepatitis B Virus Antibody Assays,'' and
``Hepatitis B Virus Nucleic Acid-Based Assays.'' FDA believes the
standard in section 513(a)(1)(B) of the FD&C Act is met as there is
sufficient information to establish special controls, which, in
addition to general controls, will provide reasonable assurance of the
safety and effectiveness of these devices.\1\
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this proposed
order is styled as ``Proposed amendment; proposed order,'' rather
than ``Proposed order.'' Beginning in December 2019, this editorial
change was made to indicate that the document ``amends'' the Code of
Federal Regulations. The change was made in accordance with the
Office of the Federal Register's (OFR) interpretations of the
Federal Register Act (44 U.S.C. chapter 15), its implementing
regulations (1 CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
---------------------------------------------------------------------------
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. FDA has determined that premarket
notification is necessary to provide a reasonable assurance of the
safety and effectiveness of HBV antigen assays, HBV antibody assays,
and HBV nucleic acid-based assays for their intended uses, therefore,
the Agency does not intend to exempt these proposed class II devices
from the requirement for premarket notification (510(k)) submission as
provided under section 510(m) of the FD&C Act. If this proposed order
is finalized, persons who intend to market this type of device must
submit to FDA a premarket notification under section 510(k) of the FD&C
Act prior to marketing the device.
II. Regulatory History of the Devices
Under section 513(f)(1) of the FD&C Act, qualitative HBV antigen
assays, HBV antibody assays (including qualitative and quantitative
anti-HBs), and quantitative HBV nucleic acid-based assays are
automatically classified into class III because they were not
introduced or delivered for introduction into interstate commerce for
commercial distribution before May 28, 1976, and have not been found
substantially equivalent to a device placed in commercial distribution
after May 28, 1976, which was subsequently classified or reclassified
into class II or class I. Therefore, they are subject to PMA
requirements under section 515 of the FD&C Act (21 U.S.C. 360e).
Qualitative HBV antigen assays and HBV antibody assays (including
qualitative and quantitative anti-HBs) are prescription devices and
assigned product code LOM. Quantitative HBV nucleic acid-based assays
are prescription devices and assigned product code MKT.
A. Qualitative HBV Antigen Assays
The first proposed device reclassification action applies to
qualitative HBV antigen assay devices that are prescription in vitro
diagnostic devices intended for qualitative detection of HBV antigens
as an aid in the diagnosis of acute or chronic HBV infection in
specific populations. On February 8, 2001, FDA approved its first HBV
antigen assay (DiaSorin's ETI-EBK PLUS) for use in the qualitative
detection of hepatitis Be antigen (HBeAg) in human serum or plasma
(ethylenediaminetetraacetic acid (EDTA), citrate, or heparin) as
indicative of a laboratory diagnosis of HBV infection through its PMA
process under section 515 of the FD&C Act. On June 1, 2001, FDA
approved its first HBV surface antigen (HBsAg) assay (Roche Elecsys
HBsAg Immunoassay, Elecsys HBsAg Confirmatory, and Precicontrol HBsAg)
for the qualitative detection of HBsAg in human serum or plasma
(heparin, EDTA, sodium citrate) in adult pregnant and non-pregnant
individuals. In a May 22, 2002, Federal Register notice (67 FR 36009),
FDA announced the approval order and the availability of the Summary of
Safety and Effectiveness Data (SSED) for these devices. Since the first
approval order for an HBV antigen assay issued on February 8, 2001, FDA
has approved 16 additional original PMAs for qualitative HBV antigen
assays that are prescription devices intended for the detection of HBV
antigens. These assays are intended as an aid in the diagnosis of acute
or chronic HBV infection in conjunction with clinical findings and
other diagnostic procedures (e.g., HBV serology and antigen testing,
liver function, etc.). These assays are not intended for use in
screening of blood, plasma, cells, or tissue donors.
A review of the medical device reporting (MDR) databases indicates
that there were 625 reported events for qualitative HBV antigen assays
as of June 2024. Of these reported events, a significant majority of
these were determined to be of no known impact or consequence to the
patient. Events reported included false reactive results, false non-
reactive results, incorrect or inadequate assay results, incorrect/
inadequate/imprecise readings, improper or incorrect procedure or
method, device operates differently than expected, and adverse event
without identified device or use problems. Where incorrect results were
obtained, it was not clear what the correct result should have been. As
of June 2024, there have been no class III recalls, six class II
recalls, and no class I recalls \2\ involving qualitative HBV antigen
assays. The class II recalls occurred since 2006 due to defective caps,
device design, no marketing application, signal for reactive results,
and biased results for biotin concentrations that were lower than
indicated. No patient harm was identified. These facts, coupled with
the low number of reported events that caused patient harm, indicate a
good safety record for this device class. These recall events reflect
the risks to health identified in section V below, and FDA believes the
special controls proposed herein, in addition to general controls, can
effectively mitigate the risks identified in these recalls.
---------------------------------------------------------------------------
\2\ Class I, II, and III recalls are defined in 21 CFR 7.3(m).
---------------------------------------------------------------------------
B. HBV Antibody Assays (Including Qualitative and Quantitative Anti-
HBs)
The second type of devices this proposed reclassification order
applies to are qualitative HBV antibody assays and quantitative anti-
HBs assays that are prescription in vitro diagnostic devices intended
for use in the detection of antibodies to HBV. These devices are
intended to aid in the diagnosis of HBV infection in persons with signs
and symptoms of hepatitis and in persons at risk for HBV infection. On
September 29, 2000, FDA approved its first qualitative HBV antibody
assay (Ortho-Clinical Diagnostics, Inc.'s Vitros
[[Page 78268]]
Immunodiagnostic Products: Anti-HBS Reagent Pack/Anti-HBS Calibrators)
for the qualitative in vitro determination of total antibody to
hepatitis B surface antigen (anti-HBs) in human serum as an aid in
determining susceptibility to HBV infection for individuals prior to or
following HBV vaccination, or where vaccination status is unknown, and
for use with other HBV serological markers for the laboratory diagnosis
of HBV disease associated with HBV infection, through its PMA process
under section 515 of the FD&C Act. In a March 12, 2001, Federal
Register notice (66 FR 14390), FDA announced the approval order and the
availability of the SSED for this device. On July 22, 2002, FDA
approved its first quantitative Anti-HBs (Siemens Healthcare
Diagnostics Products Ltd.'s Immulite 2000 XPI Anti-HBs) for the
quantitative measurement of total antibodies to the hepatitis B surface
antigen (anti-HBs) in human serum and plasma (heparinized or EDTA) as
an aid in the determination of susceptibility to HBV infection for
individuals prior to or following HBV vaccination, or where vaccination
status is unknown, or for use with other HBV serological markers for
the laboratory diagnosis of HBV disease associated with HBV infection,
through its PMA process under section 515 of the FD&C Act.
Since the first approval order of a qualitative HBV antibody assay
on September 29, 2000, FDA has approved 31 additional original PMAs for
qualitative HBV antibody assays for the detection of antibodies to HBV.
FDA has also approved six assays for quantitative anti-HBs detection.
Qualitative HBV antibody assays and quantitative anti-HBs assays are
intended to aid in the diagnosis of HBV infection in persons with signs
and symptoms of hepatitis and in persons at risk for HBV infection in
conjunction with clinical findings and other diagnostic procedures
(e.g., HBV serology and antigen testing, liver function, etc.). These
assays are not intended for use in screening of blood, plasma, cells,
or tissue donors.
A review of the MDR databases indicates that there were 1,107
reported events for HBV antibody assays between years 2001 and June
2024. Of these reported events, a significant majority of these were of
no known impact to the patient, and only four resulted in impact to
patients such as misdiagnosis or viral infection. Events reported
included adverse events without identified device or use problem,
disconnection/low assay results, false non-reactive results, false
reactive results, false high assay results (for example, the first
assay result had a low signal to cutoff (s/co) value and repeat testing
produced a higher s/co value), incorrect assay results, inadequate
assay results, and low assay results (for example, the first assay
result was in the equivocal zone, repeat testing produced a non-
reactive result, and testing with an alternate device produced a
reactive result). In numerous cases, it was not possible to determine
what the correct result should have been (further testing was not
performed, insufficient sample volume, different assays were used). As
of June 2024, FDA is aware of 4 class III recalls, 12 class II recalls,
and no class I recalls for these devices. The class II recalls occurred
in 2007, 2008, 2009, 2011, 2012, 2013, 2014, 2018, and 2019, and were
related to issues such as false reactive results, false high assay
results, defective caps, and errors in labeling, packaging, or
software. No patient harm has been identified. These facts, coupled
with the low number of reported events that impacted the patient,
indicate a good safety record for this device class. These recall
events reflect the risks to health identified in section V below, and
FDA believes the special controls proposed herein, in addition to
general controls, can effectively mitigate the risks identified in
these recalls.
C. Quantitative HBV Nucleic Acid-Based Assays
Finally, the third type of device this proposed reclassification
order applies to are quantitative HBV nucleic acid-based assay devices
for use as a prescription in vitro diagnostic device intended for use
in the detection of HBV nucleic acid in specimens from individuals with
antibody evidence of HBV infection. On September 4, 2008, FDA approved
its first quantitative HBV nucleic assay (Roche Molecular Systems,
Inc.'s COBAS TaqMan HBV Test For Use With The High Pure System), an in
vitro nucleic acid amplification assay for the quantitation of HBV
deoxyribonucleic acid (DNA) in human serum or plasma (EDTA) intended
for use as an aid in the management of patients with chronic HBV
infection undergoing antiviral therapy, through its PMA process under
section 515 of the FD&C Act.
Since the first approval order, FDA has approved four additional
original PMAs for quantitative HBV nucleic acid-based assays for the
quantitative detection of HBV DNA. The detection of HBV DNA is used for
management of patients undergoing antiviral therapy for assessing
response to treatment and not as a diagnostic for HBV infection.
The following section provides examples of the different
technologies used. The different technologies begin with specimen lysis
and HBV DNA through hybridization with magnetic particles. The
differences in the technologies occur with the method of amplification:
In one technology, the target HBV DNA sequence is
amplified. The presence of HBV amplification products is detected by
measuring the fluorescence of the HBV probe that binds to the target.
Similarly, the presence of the internal control amplification product
is detected. In the absence of HBV or internal control target
sequences, probe fluorescence is quenched. In the presence of HBV or
internal control target, the HBV or internal control probes bind to
their target.
In another technology, target amplification occurs via
transcription-based nucleic acid amplification by fluorescent labeled
probes (torches). More torches hybridize when more amplicon is present
creating a higher fluorescent signal. The time taken for the
fluorescent signal to reach a threshold proportional to the starting
HBV DNA concentration is measured in relation to internal controls.
A review of the MDR databases indicates that as of June 2024 there
were 13 reported events for nucleic acid-based HBV DNA assays since the
first reported event in 2009. MDRs were for the following reasons: (1)
incorrect, inadequate, or imprecise result or readings; (2) high
readings; and (3) non-reproducible results. Of these, two had no known
impact or consequence to the patient and two occurred when the patient
had no signs, symptoms, or conditions. As of June 2024, FDA is aware of
one class III recall, five class II recalls, and no class I recalls for
these devices. The class II recalls occurred between 2005 and 2022 and
were related to issues such as misquantitation of high results for
negative samples (carryover from a high positive sample tested adjacent
to a negative sample may produce an incorrect positive result), liquid
level detection of reagent cassette, under filled and over filled
enzyme reagent vials in assay kits, software, and low level of
recombinant HBV DNA found in one lot of reagent. These facts, coupled
with the low number of reported events that impacted the patient,
indicate a good safety record for this device class. These recall
events reflect the risks to health identified in section V below, and
FDA believes the special controls proposed herein, in addition to
general controls, can effectively mitigate the risks identified in
these recalls.
[[Page 78269]]
III. Device Description
The HBV assays that are the subject of this proposed order are
postamendments prescription in vitro diagnostic devices classified into
class III under section 513(f)(1) of the FD&C Act.
A. Qualitative HBV Antigen Assays
A qualitative HBV antigen assay is a prescription in vitro
diagnostic device intended for use in the qualitative detection of HBV
antigens and for use as an aid in the diagnosis of HBV infection in
specific populations. HBV antigen assays aid in the diagnosis of acute
or chronic HBV infection. HBV antigen assays typically detect the
presence of Hepatitis B surface antigen (HBsAg) or Hepatitis B e
antigen (HBeAg). HBV antigens (HBsAg and HBeAg), when present in
samples, bind to anti-HBs or anti-HBe antibodies to form a complex that
is bound to a solid phase (e.g., microparticles, microtiter plate or
other technology). Detection of the complexes can be performed using
different methods which measure the presence/absence of anti-HBs or
anti-HBe antibodies in the sample.
Diagnosis of HBV infection should not be established based on a
single assay result but should be determined in conjunction with
clinical findings and other diagnostic procedures (e.g., HBV serology
and antigen testing, liver function, etc.). These assays are not
intended for use in screening of blood, plasma, cells, or tissue
donors.
B. HBV Antibody Assays (Including Qualitative and Quantitative Anti-
HBs)
A qualitative HBV antibody assay is a prescription in vitro
diagnostic device intended for use in the qualitative detection of
antibodies to HBV and for use as an aid in the diagnosis of HBV
infection in specific populations. HBV antibody assays aid in the
diagnosis of HBV infection in persons with signs and symptoms of
hepatitis and in persons at risk for HBV infection. Antibody assays
typically detect the presence of antibodies to HBsAg (anti-HBs),
Hepatitis B core antigen (anti-HBc), or HBeAg (anti-HBe). Diagnosis of
HBV infection should not be established based on a single assay result,
but should be determined in conjunction with clinical findings and
other diagnostic procedures (e.g., HBV serology and antigen testing,
liver function, etc.). These assays are not intended for use in
screening of blood, plasma, cells, or tissue donors.
A quantitative assay that detects anti-HBs (antibodies to HBV
surface antigen (HBsAg)) is a prescription in vitro diagnostic device
that is intended for quantitative use to aid in the diagnosis of HBV
infection in persons with signs and symptoms of hepatitis and in
persons at risk for HBV infection. Detection of anti-HBs indicates a
present or past infection with HBV and can be used in conjunction with
clinical findings such as other HBV serological markers (detection of
other HBV antigens and antibodies to HBV) for diagnosis of HBV
infection. Anti-HBs assay results may be used as an aid in the
determination of susceptibility to HBV infection in individuals prior
to vaccination or when vaccination status is unknown.
In some device designs, HBV antibodies, when present in the sample,
bind to HBV antigens to form a complex that is bound to a solid phase
(e.g., microparticles, microtiter plate, or other technology).
Detection of complexes can be performed using different methods that
measure the presence/absence of HBV antibodies in the sample.
C. Quantitative HBV Nucleic Acid-Based Assays
A quantitative HBV nucleic acid-based assay is a prescription in
vitro diagnostic device intended for use in the detection of HBV
nucleic acid in specimens from individuals with antibody evidence of
HBV infection. In these devices, the detection of HBV nucleic acid is
used for management of patients undergoing antiviral therapy for
assessing response to treatment and NOT as a diagnostic for HBV
infection.
FDA is proposing to reclassify qualitative HBV antigen, HBV
antibody assays (including qualitative and quantitative anti-HBs), and
quantitative HBV nucleic acid-based assays from class III (general
controls and premarket approval) to class II (general controls and
special controls) and to establish new names for the device types that
will be within the classification regulations. FDA proposes to revise
21 CFR part 866 to create three new device classification regulations
with the names ``Qualitative Hepatitis B Virus Antigen Assays,''
``Hepatitis B Virus Antibody Assays,'' and ``Hepatitis B Virus Nucleic
Acid-Based Assays.'' FDA believes that these names and proposed
identification language most accurately describe these devices.
A Qualitative Hepatitis B Virus (HBV) Antigen Assay is
tentatively identified as an in vitro diagnostic device intended for
prescription use for qualitative use with human serum, plasma, or other
matrices that aids in the diagnosis of chronic or acute HBV infection.
HBV surface antigen (HBsAg) is also used for screening of HBV infection
in pregnant women to identify neonates who are at risk of acquiring
hepatitis B during perinatal period. The assay is not intended for
screening of blood, plasma, cells, or tissue donors.
A Hepatitis B Virus (HBV) Antibody Assay is tentatively
identified as an in vitro diagnostic device intended for prescription
use in the detection of antibodies to HBV in human serum and plasma, or
other matrices, and or as an aid in the diagnosis of HBV infection in
persons with signs and symptoms of hepatitis and in persons at risk for
hepatitis B infection. In addition, anti-HBc IgM (IgM antibodies to
core antigen) assay is indicative of recent HBV infection. Anti-HBs
(antibodies to surface antigen) assay results may be used as an aid in
the determination of susceptibility to HBV infection in individuals
prior to or following HBV vaccination or when vaccination status is
unknown. The assay is not intended for screening of blood, plasma,
cells, or tissue donors. The assay is intended as an aid in diagnosis
in conjunction with clinical findings and other diagnostic procedures.
A Hepatitis B Virus (HBV) Nucleic Acid-Based Assay is
tentatively identified as an in vitro diagnostic device intended for
prescription use in the detection of HBV nucleic acid in specimens from
individuals with antibody evidence of HBV infection. In these devices,
the detection of HBV nucleic acid is used as an aid in the management
of HBV-infected individuals. The assay is intended for use with human
serum or plasma (and other matrices as applicable) from individuals
with HBV. The assay is not intended for use as a donor screening assay
for the presence of HBV nucleic acids in blood, blood products, plasma,
cells, or tissue donors.
Based upon our review experience and consistent with the FD&C Act
and FDA's regulations in 21 CFR 860.134, FDA believes that these
devices should be reclassified from class III into class II with
special controls because there is sufficient information to establish
special controls that, along with general controls, can provide
reasonable assurance of the devices' safety and effectiveness.
IV. Proposed Reclassification and Summary of Reasons for
Reclassification
FDA is proposing to reclassify the HBV assays that are the subject
of this proposed order. On September 7, 2023, the Microbiology Devices
Panel (Panel) of the Medical Devices Advisory Committee convened to
discuss and make recommendations regarding the
[[Page 78270]]
reclassification of HBV assays from class III (general controls and
premarket approval) to class II (general controls and special controls)
(https://www.fda.gov/media/173609/download). Panel members unanimously
agreed that special controls, in addition to general controls, are
necessary and sufficient to mitigate the risks to health of patients
presented by these devices and to provide reasonable assurance of the
safety and effectiveness of these devices (Refs. 1 and 2). The Panel
agreed with FDA-identified risks and identified additional risk(s) and
benefit(s) to include in the overall risk assessment. The Panel also
discussed potential mitigation measure(s)/control(s) FDA should
consider for each of the identified risks and recommended that, as part
of any reclassification, the expected performance for these devices
should remain the same. Notably, the performance of approved HBV
antigen assays has generally been at least 97 percent sensitivity and
99 percent specificity. For approved anti-HBs, anti-Hbe, and anti-HBc
total assays the sensitivity has generally been at least 95 percent,
for approved anti-HBc IgM assays the sensitivity has been at least 86
percent, and for all HBV approved antibody assays the specificity has
generally been above 97 percent.
FDA believes that at this time, sufficient data and information
exist such that the risks identified in section V below can be
mitigated by establishing special controls, and that these special
controls, together with general controls, are necessary to provide a
reasonable assurance of the safety and effectiveness of these HBV
assays and therefore proposes these devices to be reclassified from
class III (general controls and premarket approval) to class II
(general controls and special controls). In accordance with section
513(f)(3) of the FD&C Act and 21 CFR part 860, subpart C, FDA is
proposing to reclassify qualitative HBV antigen assays, HBV antibody
assays (including qualitative and quantitative anti-HBs), and
quantitative HBV nucleic acid-based assays from class III into class
II, subject to premarket notification (510(k)) requirements. FDA
believes that there is sufficient information available to FDA through
FDA's accumulated experience with these devices from reviewing the PMAs
for these HBV assays, and the Panel considerations and recommendations
regarding the proposed special controls that FDA believes would
effectively mitigate the risks to health identified in section V.
Absent the special controls identified in this proposed order, general
controls applicable to the devices are insufficient to provide
reasonable assurance of the safety and effectiveness of the devices.
FDA expects that the reclassification of these devices would enable
more manufacturers to develop these assays such that patients would
benefit from increased access to safe and effective tests.
FDA is proposing to create three separate classification
regulations for HBV assays that will be reclassified from class III to
class II. HBV assays are prescription in vitro diagnostic devices, and
under this proposed order, if finalized, these devices will be
identified as prescription in vitro diagnostic devices. As such, the
devices must satisfy prescription labeling requirements for in vitro
diagnostic products (see 21 CFR 809.10(a)(4) and (b)(5)(ii)). In this
proposed order, if finalized, FDA has identified the special controls
under section 513(a)(1)(B) of the FD&C Act that, together with general
controls, will provide a reasonable assurance of the safety and
effectiveness of these assays.
FDA is also proposing to create a new product code for HBV antibody
assays (including qualitative and quantitative anti-HBs) that will be
assigned upon any finalization of this proposed order. Qualitative HBV
antigen assays will continue to be assigned the product code LOM upon
any finalization of this proposed order.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For these HBV assays, FDA has determined
that premarket notification is necessary to provide a reasonable
assurance of the safety and effectiveness of these devices. Therefore,
the Agency does not intend to exempt these proposed class II devices
from 510(k) requirements. If this proposed order is finalized, persons
who intend to market a new HBV assay will no longer need to have a PMA
for these devices but can instead submit to FDA a 510(k) and receive
clearance prior to marketing the device. A 510(k) typically results in
a shorter premarket review timeline compared to a PMA, which ultimately
provides more timely access of these types of devices to patients.
V. Public Health Benefits and Risks to Health
FDA is providing a substantive summary of the valid scientific
evidence concerning the public health benefits of the use of HBV assays
(see also https://www.fda.gov/media/171770/download), and the nature
(and if known, the incidence) of the risks of the devices (see further
discussion of the special controls being proposed to mitigate these
risks in section VII of this proposed order).
HBV infection represents a significant global public health burden.
According to the World Health Organization (WHO), in 2019 there were
approximately 296 million people chronically infected people worldwide,
with 1.5 million new HBV infections each year.\3\ It is estimated by
the Centers for Disease Control and Prevention (CDC) that chronic HBV
infection in the United States affects at least between 580,000 to 1.17
million people with HBV infection in the United States; two-thirds of
whom may be unaware of their infection.\4\ HBV infection can be
asymptomatic, and accordingly, many HBV-infected individuals are
unaware of their HBV infection. Approximately 95 percent of adult
patients with acute infection, defined as the first 6 months after
infection, recover completely, and 5 percent of adults develop chronic
HBV.\5\ Infants born to women who are HbsAg-positive are at high risk
of HBV infection. In absence of treatment, infants infected with HBV
have a 90 percent risk of progression to chronic HBV and up to 25
percent of infants who acquire chronic HBV infection will die
prematurely from HBV-related hepatocellular carcinoma or cirrhosis.\6\
Patients who are tested and become aware that they are HBV infected may
modify risk behaviors to prevent transmission to others and can be
referred for treatment. Patients with chronic HBV infection have a risk
of developing liver damage, liver cancer, or liver failure. They can
also spread their infection to others. HBV can be reactivated in
patients receiving immunosuppressive therapies, resulting in serious
risk of liver failure or liver-associated death (Ref. 3). HBV is a
vaccine-preventable liver infection.
---------------------------------------------------------------------------
\3\ https://www.who.int/news-room/fact-sheets/detail/hepatitis-b. Accessed on July 12, 2024.
\4\ Centers for Disease Control and Prevention--Clinical
Overview of Hepatitis B (Available at https://www.cdc.gov/hepatitis-b/hcp/clinical-overview/). Accessed on July 12, 2024.
\5\ Ibid.
\6\ Ibid.
---------------------------------------------------------------------------
With the initiation of the WHO Viral Hepatitis Elimination Plan \7\
and the Department of Health and Human Services (HHS) Viral Hepatitis
National
[[Page 78271]]
Strategic Plan for the United States,\8\ it is important for
individuals to know their HBV infected status, to link HBV infected
individuals to care, and to eliminate virus transmission. Therefore,
diagnosis of patients with HBV infection through devices such as HBV
antibody and antigen assays is essential to ensure that patients are
linked to the appropriate care. Current CDC HBV Screening and Testing
Recommendations include testing of the following groups: all adults 18
and older at least once in their lifetime using a triple panel test,
pregnant women during pregnancy, people who are at ongoing risk for
exposure, and anyone who requests HBV testing.\9\
---------------------------------------------------------------------------
\7\ https://www.who.int/health-topics/hepatitis/elimination-of-hepatitis-by-2030#tab=tab_1. Accessed on July 12, 2024.
\8\ https://www.hhs.gov/sites/default/files/Viral-Hepatitis-National-Strategic-Plan-2021-2025.pdf. Accessed on July 12, 2024.
\9\ https://www.cdc.gov/hepatitis/hbv/index.htm. Accessed on
July 12, 2024.
---------------------------------------------------------------------------
FDA considered our accumulated experience with the regulation of
these HBV assays, input from the Panel meeting, and postmarket
information regarding these HBV assays, i.e., information from FDA's
publicly available MDR, Manufacturer and User Facility Device
Experience (MAUDE), and Medical Device Recall databases.
These HBV assays provide a benefit to the public health by
informing individuals of their HBV infected status, linking HBV
infected individuals to appropriate care, and aiding in eliminating
virus transmission. Once an individual is tested and diagnosed as HBV
infected, HBV nucleic acid testing is performed to inform treatment
decisions. While HBV infection is treatable, it is not curable, which
means that most people who start HBV antiviral treatment must continue
it for life. The goal of current treatment is to suppress the virus and
reduce the likelihood of long-term complications and transmission
(Refs. 3 and 4). Thus, identifying individuals who are HBV infected,
linking them to care, and managing their HBV infection to alleviate
development of liver damage, liver cancer, liver failure, and potential
HBV transmission would not only greatly impact public health but also
go a long way towards helping the United States achieve HBV
elimination.
Probable risks to health associated with the use of HBV assays
include risks related to the risk of false results (false positives,
false negatives, inaccurate low assay results, inaccurate high assay
results, false reactive results, or false non-reactive results),
failure to correctly interpret assay results, and failure to correctly
operate the device. For HBV antigen and antibody assays, false positive
results are generally referred to as false reactive results and false
negative results are generally referred to as false non-reactive
results. False results can lead to uninfected individuals receiving
unnecessary further testing and treatment or infected individuals
remaining undiagnosed and untreated. Undiagnosed and untreated
individuals are likely to experience increases in morbidity and
mortality and can spread the infection to others. FDA has identified
the following additional specific risks to health associated with each
of the HBV assays listed below.
A. Qualitative HBV Antigen Assays
Factors that may cause decreased assay sensitivity and/or an
increased rate of false non-reactive results include, but are not
limited to, the presence of interfering substances in the sample, acute
infection at a stage that is too early for a device to detect the
infection, and antigen concentrations that are too low to be detected
by the device. Factors that may lead to false reactive results include
device contamination from reactive samples, cross-reactivity with other
antigens, or misinterpretation of invalid results as reactive.
A false reactive assay result for HbeAg. Incorrectly
interpreting the assay results as a reactive assay result or failing to
correctly operate the assay causing a false reactive assay result may
lead to continued treatment for hepatitis B with antiviral medication
when it otherwise would not be indicated. Antiviral medication has
risks including toxicity and more rarely allergic reactions. Over time,
viral resistance in patients who are co-infected but undiagnosed with
other viruses that are treated with the same antiviral medication, such
as HIV, can lead to viral resistance.
A false reactive assay result for HbsAg. Incorrectly
interpreting the assay results as a reactive assay result or failing to
correctly operate the assay causing a false reactive assay result may
contribute to unnecessary additional testing, potentially delaying
diagnosis of alternative causes of liver disease when present and may
impact the psychological well-being of the patient. Factors that may
increase the rate of false reactive assay reporting include cross-
reactivity with antigens from other microorganisms or other disease
conditions.
A false non-reactive result for HbeAg. Incorrectly
interpreting the assay results as a non-reactive assay result or
failing to correctly operate the assay causing a false non-reactive
assay result may lead to missing the opportunity for treatment of an
HBV infected individual with antiviral medication or premature
discontinuation of antiviral treatment when continuation of treatment
is otherwise indicated should a clinician be falsely led to determine a
patient has seroconverted HbeAg to anti-Hbe. Premature discontinuation
of antiviral medication could result in adverse effects on patient
health, such as cirrhosis, liver cancer, and liver damage, all of which
are known to contribute to patient morbidity and mortality, or may
contribute to public health risk by leading to virus transmission.
A false non-reactive assay result for HbsAg. Incorrectly
interpreting the assay results as a non-reactive assay result or
failing to correctly operate the assay causing a false non-reactive
assay result may delay or prevent a patient with HBV infection from
being identified and linked to care. Missed identification of patients
with chronic HBV infection could lead to adverse effects on patient
health such as cirrhosis, liver cancer, and liver damage, all of which
are known to contribute to patient morbidity and mortality. A false
non-reactive HbsAg assay incorrectly interpreted as non-reactive also
may contribute to public health risk by leading to virus transmission.
B. HBV Antibody Assays (Including Qualitative and Quantitative Anti-
HBs)
Factors that may cause decreased assay sensitivity and/or an
increased rate of false non-reactive results include, but are not
limited to, the presence of interfering substances in the sample, acute
infection at a stage that is too early for a device to detect the
infection, and antibody concentrations that are too low to be detected
by the device. They also can be caused by misinterpretation of invalid
results as non-reactive. Factors that may lead to false reactive
results include device contamination from reactive samples, cross-
reactivity with other antibodies, or misinterpretation of invalid
results as reactive.
A false reactive assay result for anti-HBs and anti-HBc.
Incorrectly interpreting the assay results as a reactive assay result
or failing to correctly operate the assay causing a false reactive
assay result may lead to improper patient management. A false reactive
antibody assay result could result in the unnecessary continuation of
antiviral treatment. Antiviral medication has risks including toxicity
and more rarely allergic reactions. Over time, viral resistance in
patients who are co-infected but undiagnosed with other viruses that
are treated with the same antiviral medication, such as HIV, can
[[Page 78272]]
lead to viral resistance. Consequently, repeatedly false reactive
results have the potential to lead to inappropriate patient management
decisions.
A false reactive assay result for anti-HBs. Incorrectly
interpreting the assay results as a reactive assay result or failing to
correctly operate the assay causing a false reactive assay result when
the device is used as an aid in the determination of susceptibility to
HBV infection in individuals prior to or following HBV vaccination or
where vaccination status is unknown may cause a patient to be
considered previously exposed and therefore immune to HBV or that the
patient was successfully vaccinated. A false reactive result may cause
the patient to not receive a vaccine, vaccine booster, hyperimmune
globulin, and would be at higher risk of infection if exposed to HBV.
A false reactive assay result for anti-Hbe. Incorrectly
interpreting the assay results as a reactive assay result, or failing
to correctly operate the assay causing a false reactive assay result
may lead to missing the opportunity for treatment of HBV infection with
antiviral medications in a subset of individuals for whom treatment
would otherwise be indicated, or premature discontinuation of antiviral
treatment when continuation of treatment is otherwise indicated should
a clinician be falsely led to determine a patient has seroconverted
HbeAg to anti-Hbe. Premature discontinuation of antiviral medication
could result in adverse effects on patient health such as cirrhosis,
liver cancer, and liver damage, all of which are known to contribute to
patient morbidity and mortality, or may contribute to public health
risk by leading to inadvertent transmission of virus by an infected
individual.
A false non-reactive assay result for anti-HBc. When the
device is used as an aid in the diagnosis of HBV infection in patients
with symptoms of hepatitis or who may be at risk for HBV infection,
incorrectly interpreting the assay results as non-reactive assay
result, or failing to correctly operate the assay causing a false non-
reactive assay result may lead to non-diagnosis or a delay in diagnosis
of HBV infection with an associated delay in therapy and potentially
increased risk of HBV-related morbidity or mortality. Patients with
active infection may unknowingly continue to infect others. False non-
reactive results can also lead to unnecessary diagnostic evaluation if
alternative etiologies of hepatitis are pursued. False non-reactive
assay results may occur if the level of antibody in a specimen is below
the limit of detection of the assay.
A false non-reactive assay result for anti-HBs. When the
device is used as an aid in the determination of susceptibility to HBV
infection in individuals prior to or following HBV vaccination or where
vaccination status is unknown, incorrectly interpreting the assay
results as a non-reactive assay result or failing to correctly operate
the assay causing a false non-reactive assay result may lead to
unnecessary repeated vaccination for HBV.
A false non-reactive assay result for anti-Hbe.
Incorrectly interpreting the assay results as non-reactive assay result
or failing to correctly operate the assay causing a false non-reactive
assay result may lead to improper patient management, including
continued treatment for HBV with antiviral medication. Antiviral
medication has risks including toxicity and more rarely allergic
reactions. Over time, viral resistance in patients who are co-infected
but undiagnosed with other viruses using the same antiviral medication,
such as HIV, can lead to viral resistance.
C. Quantitative HBV Nucleic Acid-Based Assays
Decreased assay sensitivity and/or an increased rate of false
negative assay reporting may occur with patient samples that contain
different genotypes or rare de novo mutations in HBV genomic regions
targeted by the device. In these situations, HBV viral load can
transiently decrease and/or become undetectable in samples before the
virus enters chronic replication.
A false positive or falsely elevated quantitative HBV
nucleic acid assay result. Incorrectly interpreting the assay results
as a positive assay result or failing to correctly operate the assay
causing a false positive assay result may negatively influence patient
management decisions. Such decisions may include the administration or
continuation of unnecessary antiviral treatment in patients with
chronic HBV infection with its known toxicities and more rarely
allergic reactions. Certain patients with falsely elevated HBV nucleic
acid assay results may not undergo liver biopsy to investigate other
causes of liver disease when the biopsy would otherwise be indicated
for certain patients.
A false negative or falsely decreased quantitative HBV
nucleic acid assay result. Incorrectly interpreting the assay results
as a negative assay result, or failing to correctly operate the assay
causing a false negative assay result may negatively influence patient
management decisions for patients with chronic HBV infection, including
the withholding of treatment, failure to treat, or premature
discontinuation of treating HBV infection when antiviral treatment is
otherwise indicated or the choice of an inappropriate treatment. This
could lead to adverse effects on patient health such as progressive
liver disease, cirrhosis and/or hepatocellular carcinoma, and other
cancers. Patients with active HBV replication also risk spreading the
virus to others. Certain patients with falsely low HBV nucleic acid
assay results may undergo liver biopsy to investigate other causes of
liver disease.
VI. Summary of Data Upon Which the Reclassification Is Based
The safety and effectiveness of these device types has become well
established since the initial approval of the first qualitative HBV
antibody assay in 2000, the first HBV antigen assay in 2001, and the
first quantitative HBV nucleic acid-based assay in 2008. FDA has
considered and analyzed the following information: (1) accumulated
experience regulating these HBV assays, (2) input from the Panel
meeting, and (3) postmarket information regarding HBV assays, i.e.,
information from FDA's publicly available MDR, MAUDE, and Medical
Device Recall databases. The available evidence demonstrates that there
are public health benefits derived from the use of HBV assays indicated
for use to aid in diagnosis of HBV infection and/or for use to aid in
the management of HBV infected patients, or as an aid in the
determination of susceptibility to HBV infection (anti-HBs). In
addition, the nature of the associated risks to health are known, and
special controls can be established to sufficiently mitigate these
risks.
Based on our review of the information described above, FDA has
determined that special controls, in addition to general controls, are
necessary to provide a reasonable assurance of safety and effectiveness
for HBV assays, and that sufficient information exists to establish
such special controls. Therefore, FDA, on its own initiative, is
proposing to reclassify these postamendments devices from class III
(general controls and premarket approval) into class II (general
controls and special controls), subject to premarket notification
(510(k)) requirements.
VII. Proposed Special Controls
FDA believes that these devices can be classified into class II
with the establishment of special controls. FDA believes that the
following proposed special controls would mitigate each of
[[Page 78273]]
the risks to health described in section V and that these special
controls, in addition to general controls, would provide a reasonable
assurance of safety and effectiveness for HBV assays. Tables 1 through
3 below demonstrate how FDA believes each risk to health described in
section V would be mitigated by the proposed special controls for each
device type.
A. Qualitative HBV Antigen Assays
The risk of inaccurate interpretation of assay results can be
mitigated by special controls requiring certain labeling, including
providing clearly stated warnings and limitations and information on
principles of operation and procedures in performing the assay.
Risks associated with false results (e.g., false non-reactive and
false reactive assay results) and with the failure to correctly operate
the device can be mitigated through a combination of special controls,
including certain labeling requirements, certain design verification
and validation information, and performance studies. Examples of
verification and validation information to be included in the design of
the device include documentation of performance specifications
including analytical and clinical performance criteria. In addition,
design verification and validation activities must include
documentation of a complete device description, critical reagents, risk
analysis strategies, lot release criteria, stability studies, and
protocols. Required statements in labeling can aid in mitigating the
failure of the device to perform as indicated, for example including a
statement that use of the assay with specimen types other than those
specifically identified for use with this device may cause inaccurate
assay results. Special controls requiring additional labeling to
provide a brief summary of the instructions for use can also mitigate
these risks.
Table 1--Risks to Health and Mitigation Measures for Qualitative HBV
Antigen Assays
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
False reactive/non-reactive assay Certain labeling information,
result. including limitations,
explanation of procedures, and
results interpretation
information.
Certain design verification and
validation information,
including certain device
description information, risk
analysis strategies, lot release
criteria, stability studies and
protocols, and performance
criteria including analytical
studies and clinical studies.
Failure to correctly interpret the Certain labeling information,
assay results. including warnings, limitations,
results interpretation
information, and explanation of
procedures.
Certain design verification and
validation information,
including certain device
description information,
critical reagent information,
risk analysis strategies, lot
release criteria, and stability
studies and protocols.
Failure to correctly operate the Certain labeling information,
device. including warnings, limitations,
results interpretation
information, and explanation of
procedures.
Certain design verification and
validation information,
including certain device
description, critical reagent
information, risk analysis
strategies, lot release
criteria, and stability studies
and protocols.
------------------------------------------------------------------------
B. HBV Antibody Assays (Including Qualitative and Quantitative Anti-
HBs)
The risk of falsely reactive, non-reactive, elevated, or lowered
assay results can be mitigated by special controls requiring certain
labeling, including providing clearly stated warnings and limitations
and information on principles of operation and procedures in performing
the assay.
Risks associated with the failure of the device to perform as
indicated (e.g., false non-reactive and false reactive assay results)
can be mitigated through a combination of special controls, including
certain labeling requirements, certain design verification and
validation information, and performance studies. Examples of
verification and validation information to be included in the design of
the device include documentation of performance specifications
including analytical and clinical performance criteria. In addition,
design verification and validation activities must include
documentation of a complete device description, critical reagents, risk
analysis strategies, lot release criteria, stability studies, and
protocols. Required statements in labeling can aid in mitigating the
failure of the device to perform as indicated; for example, including a
statement that use of the assay with specimen types other than those
specifically identified for use with this device may cause inaccurate
assay results.
Table 2--Risks to Health and Mitigation Measures for HBV Antibody Assays
(Including Qualitative and Quantitative Anti-HBs)
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
False reactive/false non-reactive Certain labeling information,
assay result. In addition, for including limitations,
quantitative assays: Falsely explanation of procedures, and
elevated/falsely lowered assay results interpretation
result. information.
Certain design verification and
validation information including
certain device description
information, risk analysis
strategies, lot release
criteria, stability studies and
protocols, and performance
criteria including analytical
studies and clinical studies.
Failure to correctly interpret the Certain labeling information,
assay results. including warnings, limitations,
results interpretation
information, and explanation of
procedures.
Certain design verification and
validation information including
certain device description,
critical reagent information,
risk analysis strategies, lot
release criteria, and stability
studies and protocols.
[[Page 78274]]
Failure to correctly operate the Certain labeling information,
devices. warnings, limitations, results
interpretation information, and
explanation of procedures.
Certain design verification and
validation information including
certain device description,
critical reagent information,
risk analysis strategies, lot
release criteria, and stability
studies and protocols.
------------------------------------------------------------------------
C. Quantitative HBV Nucleic Acid-Based Assays
The risk of falsely positive, negative, elevated, or lowered assay
results can be mitigated by special controls requiring certain
labeling, including providing clearly stated warnings and limitations,
device description information, and detailed instructions in the device
labeling regarding the interpretation of assay results and principles
of operation and procedures in performing the assay.
Risks associated with the failure of the device to perform as
indicated (e.g., inaccurately low or high results, false negative
results, and false positive assay results) can be mitigated through a
combination of special controls related to certain labeling
requirements, design verification and validation activities, and
performance studies. Examples of verification and validation
information to be included in the design of the device include
documentation of a complete device description, calibrators, critical
reagents, traceability, and lot release criteria. In addition, design
verification and validation must include documentation of performance
specifications, including analytical and clinical performance criteria.
Required statements in labeling can aid in mitigating the occurrence of
inaccurate results. The risks of false positive/false negative/falsely
elevated/falsely lowered results due to decreased assay sensitivity can
be mitigated by special controls related to certain labeling, design
verification and validation activities, risk analysis strategies, and
performance studies.
Table 3--Risks to Health and Mitigation Measures for Quantitative HBV
Nucleic Acid-Based Assays
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
False positive/false negative/falsely Certain labeling information,
elevated/falsely lowered result. including limitations,
explanation of procedures, and
results interpretation
information.
Certain design verification and
validation information,
including certain device
description information, risk
analysis strategies, lot release
criteria, stability studies and
protocols, and performance
criteria including analytical
studies and clinical studies.
Failure to correctly interpret the Certain labeling information,
assay results. including warnings, limitations,
results interpretation
information, and explanation of
procedures.
Certain design verification and
validation information,
including certain device
description, critical reagent
information, risk analysis
strategies, lot release
criteria, and stability studies
and protocols.
Failure to correctly operate the Certain labeling warnings,
device. limitations, results
interpretation information, and
explanation of procedures.
Certain design verification and
validation information including
certain device description,
critical reagent information,
risk analysis strategies, lot
release criteria, and stability
studies and protocols.
------------------------------------------------------------------------
If this proposed order is finalized, qualitative HBV antigen
assays, HBV antibody assays (including qualitative and quantitative
anti-HBs), and quantitative HBV nucleic acid-based assays will be
reclassified into class II (general controls and special controls) and
would be subject to premarket notification requirements under section
510(k) of the FD&C Act. Firms submitting a 510(k) of the FD&C Act for
such devices will be required to comply with the particular mitigation
measures set forth in the special controls. FDA believes that adherence
to the special controls, in addition to the general controls, is
necessary to provide a reasonable assurance of safety and effectiveness
of HBV assays.
VIII. Analysis of Environmental Impact
We have determined under 21 CFR 25.34(b) that this action is of a
type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IX. Paperwork Reduction Act of 1995
While this proposed order contains no new collections of
information, it does refer to previously approved FDA collections of
information. The previously approved FDA collections of information are
subject to review by the Office of Management and Budget (OMB) under
the Paperwork Reduction Act of 1995 (PRA) (44 U.S.C. 3501-3521). The
collections of information in 21 CFR part 820 have been approved under
OMB control number 0910-0073; the collections of information in part
807, subpart E, have been approved under OMB control number 0910-0120;
and the collections of information in 21 CFR parts 801 and 809 have
been approved under OMB control number 0910-0485.
X. Proposed Effective Date
FDA proposes that any final order based on this proposed order
become effective 30 days after the date of its publication in the
Federal Register.
XI. Codification of Orders
Under section 513(f)(3) of the FD&C Act, FDA may issue final orders
to reclassify devices. FDA will continue to codify classifications and
[[Page 78275]]
reclassifications in the Code of Federal Regulations (CFR). Changes
resulting from final orders will appear in the CFR as newly codified
orders. Therefore, under section 513(f)(3) of the FD&C Act, in the
proposed order, we are proposing to codify qualitative hepatitis B
virus antigen assays in the new Sec. 866.3178, hepatitis B virus
antibody assays (including qualitative and quantitative anti-HBs) in
the new Sec. 866.3179, and quantitative hepatitis B virus nucleic
acid-based assays in the new Sec. 866.3180, under which these HBV
assays would be reclassified from class III into class II.
XII. References
The following references marked with an asterisk (*) are on display
at the Dockets Management Staff (see ADDRESSES) and are available for
viewing by interested persons between 9 a.m. and 4 p.m., Monday through
Friday; they also are available electronically at https://www.regulations.gov. References without asterisks are not on public
display at https://www.regulations.gov because they have copyright
restriction. Some may be available at the website address, if listed.
References without asterisks are available for viewing only at the
Dockets Management Staff. Although FDA verified the website addresses
in this document, please note that websites are subject to change over
time.
*1. Summary Minutes Prepared for the September 7, 2023, Meeting of
the Microbiology Devices Panel (available at https://www.fda.gov/media/173610/download).
*2. Meeting Transcript Prepared for the September 7, 2023, Meeting
of the Microbiology Devices Panel (available at https://www.fda.gov/media/173609/download).
3. Terrault, N.A., A.S.F. Lok, B.J. McMahon, et al., ``Update on
Prevention, Diagnosis, and Treatment of Chronic Hepatitis B: AASLD
2018 Hepatitis B Guidance.'' Hepatology, 67(4): 1560-1599, 2018.
4. CDC, ``Clinical Testing and Diagnosis for Hepatitis B,'' https://www.cdc.gov/hepatitis-b/hcp/diagnosis-testing/. Accessed
July 11, 2024.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act, and
under authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 866 be amended as follows:
PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES
0
1. The authority citation for part 866 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 866.3178 to subpart D to read as follows:
Sec. 866.3178 Qualitative hepatitis B virus antigen assays.
(a) Identification. A qualitative hepatitis B virus (HBV) antigen
assay is identified as an in vitro diagnostic device intended for
prescription use for qualitative use with human serum, plasma, or other
matrices that aids in the diagnosis of chronic or acute HBV infection.
HBV surface antigen (HbsAg) is also used for screening of HBV infection
in pregnant women to identify neonates who are at risk of acquiring
hepatitis B during perinatal period. The assay is not intended for
screening of blood, plasma, cells, or tissue donors.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The labeling required under Sec. 809.10(b) of this chapter
must include:
(i) A prominent statement that the assay is not intended for the
screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and
procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical
practice and disease presentation and management. The limitations must
include statements that indicate:
(A) The specimen types for which the device has been cleared, and
that use of this assay with specimen types other than those
specifically cleared for this device may result in inaccurate assay
results.
(B) When appropriate, performance characteristics of the assay have
not been established in populations of immunocompromised or
immunosuppressed patients or other populations where assay performance
may be affected.
(C) Diagnosis of hepatitis B infection should not be established on
the basis of a single assay result but should be determined by a
licensed healthcare professional in conjunction with the clinical
presentation, history, and other diagnostic procedures.
(D) Detection of HBV antigens indicates a current infection with
hepatitis B virus but does not differentiate between acute or chronic
infection. False reactive HbsAg result may occur for up to 2 weeks
after vaccination with HbsAg containing vaccine.
(E) Current methods for the detection of hepatitis B antigens may
not detect all potentially infected individuals. A non-reactive assay
result does not exclude the possibility of exposure to or infection
with hepatitis B virus. A non-reactive assay result in individuals with
prior exposure to hepatitis B may be due to but not limited to antigen
levels below the detection limit of this assay or lack of antigen
reactivity to the antibodies in this assay. HBV mutants lacking the
ability to produce antigens have been reported. These may occur as
``escape'' mutants in the presence of anti-HBV antibodies and such
patients may be infectious.
(F) Results obtained with this assay may not be used
interchangeably with results obtained with a different manufacturer's
assay.
(2) Design verification and validation must include the following:
(i) A detailed device description, including all parts that make up
the device, ancillary reagents required but not provided, an
explanation of the device methodology, design of the capture
antibody(ies), external controls, and computational path from collected
raw data to reported result (e.g., how collected raw signals are
converted into a reported signal and result), as applicable to the
detection method and device design.
(ii) For devices with assay calibrators, the design and composition
of all primary, secondary, and subsequent quantitation standards used
for calibration as well as their traceability to a standardized
reference material that FDA has determined is appropriate (e.g., a
recognized consensus standard). In addition, analytical testing must be
performed following the release of a new lot of the standard material
that was used for device clearance or approval, or when there is a
transition to a new calibration standard.
(iii) Documentation and characterization (e.g., supplier,
determination of identity, purity, and stability) of all critical
reagents (including description of the capture antibody(ies)), and
protocols for maintaining product integrity throughout its labeled
shelf life.
(iv) Risk analysis and management strategies, such as Failure Modes
Effects Analysis and/or Hazard Analysis and Critical Control Points
summaries and their impact on assay performance.
(v) Final release criteria to be used for manufactured assay lots
with appropriate evidence that lots released
[[Page 78276]]
at the extremes of the specifications will meet the identified
analytical and clinical performance characteristics as well as
stability.
(vi) Stability studies for reagents must include documentation of
an assessment of real-time stability for multiple reagent lots using
the indicated specimen types and must use acceptance criteria that
ensure that analytical and clinical performance characteristics are met
when stability is assigned based on the extremes of the acceptance
range.
(vii) All stability protocols, including acceptance criteria.
(viii) Final release assay results for each lot used in clinical
studies.
(ix) Reproducibility study data that includes the testing of three
independent production lots.
(x) Detailed documentation of analytical performance studies
conducted, as appropriate to the technology, specimen types tested, and
intended use of the device, including, the limit of blank (LoB), limit
of detection (LoD), cutoff, precision (reproducibility) including lot-
to-lot and/or instrument-to-instrument precision, interference, cross
reactivity, carryover, hook effect, seroconversion panel testing,
matrix equivalency, prominent mutants/variants detection (e.g., for
HbsAg), specimen stability, reagent stability, and cross-genotype
antigen detection sensitivity, when appropriate.
(xi) Analytical sensitivity of the assay is the same or better than
that of other cleared or approved assays.
(xii) For devices with associated software or instrumentation,
documentation must include a detailed description of device software,
including software applications and hardware-based devices that
incorporate software. The detailed description must include
documentation of verification, validation, and hazard analysis and risk
assessment activities, including an assessment of the impact of threats
and vulnerabilities on device functionality and end users/patients as
part of cybersecurity review.
(xiii) Detailed documentation and results from a clinical study.
Performance must be analyzed relative to an FDA cleared or approved HBV
antigen assay or a comparator that FDA has determined is appropriate.
This study must be conducted using appropriate patient samples, with an
appropriate number of HBV reactive and non-reactive samples in
applicable risk and disease categories, and any applicable confirmatory
testing. Additional relevant patient groups must be validated as
appropriate. The samples must include prospective (sequential) samples
for each identified specimen type and, as appropriate, additional
characterized clinical samples. Samples must be sourced from
geographically diverse areas. This study must be conducted in the
appropriate settings by the intended users to demonstrate clinical
performance.
0
3. Add Sec. 866.3179 to subpart D to read as follows:
Sec. 866.3179 Hepatitis B virus antibody assays (including
qualitative and quantitative anti-HBs).
(a) Identification. A hepatitis B virus (HBV) antibody assay is
identified as an in vitro diagnostic device intended for prescription
use in the detection of antibodies to HBV in human serum, plasma, or
other matrices, and as a device that aids in the diagnosis of HBV
infection in persons with signs and symptoms of hepatitis and in
persons at risk for hepatitis B infection. In addition, results from an
anti-HBc IgM (IgM antibodies to core antigen) assay indicating the
presence of anti-HBc IgM are indicative of recent HBV infection. Anti-
HBs (antibodies to surface antigen) assay results may be used as an aid
in the determination of susceptibility to HBV infection in individuals
prior to or following HBV vaccination or when vaccination status is
unknown. The assay is not intended for screening of blood, plasma,
cells, or tissue donors. The assay is intended as an aid in diagnosis
in conjunction with clinical findings and other diagnostic procedures.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The labeling required under Sec. 809.10(b) of this chapter
must include:
(i) A prominent statement that the assay is not intended for the
screening of blood, plasma, cells, or tissue donors.
(ii) A detailed explanation of the principles of operation and
procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical
practice and disease presentation and management. The limitations must
include statements that indicate:
(A) When appropriate, performance characteristics of the assay have
not been established in populations of immunocompromised or
immunosuppressed patients or other special populations where assay
performance may be affected.
(B) Detection of HBV antibodies to a single viral antigen indicates
a present or past infection with hepatitis B virus, but does not
differentiate between acute, chronic, or resolved infection.
(C) The specimen types for which the device has been cleared, and
that use of the assay with specimen types other than those specifically
cleared for this device may result in inaccurate assay results.
(D) Diagnosis of hepatitis B infection should not be established on
the basis of a single assay result but should be determined by a
licensed healthcare professional in conjunction with the clinical
presentation, history, and other diagnostic procedures.
(E) A non-reactive assay result may occur early during acute
infection, prior to development of a host antibody response to
infection, or when analyte levels are below the limit of detection of
the assay.
(F) Results obtained with this assay may not be used
interchangeably with results obtained with a different manufacturer's
assay.
(v) For devices intended for the quantitative detection of HBV
antibodies (anti-HBs), in addition to the special controls listed in
paragraphs (b)(1) and (2) of this section, labeling required under
Sec. 809.10(b) of this chapter must include:
(A) The assay calibrators' traceability to a standardized reference
material that FDA has determined is appropriate (e.g., a recognized
consensus standard) and the limit of blank (LoB), limit of detection
(LoD), limit of quantitation (LoQ), linearity, and precision to define
the analytical measuring interval.
(B) Performance results of the analytical sensitivity study testing
a standardized reference material that FDA has determined is
appropriate (e.g., a recognized consensus standard).
(2) Design verification and validation must include the following:
(i) Detailed device description, including all parts that make up
the device, ancillary reagents required but not provided, an
explanation of the device methodology, and design of the antigen(s) and
capture antibody(ies) sequences, rationale for the selected epitope(s),
degree of amino acid sequence conservation of the target, and the
design and composition of all primary, secondary and subsequent
standards used for calibration.
(ii) Documentation and characterization (e.g., supplier,
determination of identity, and stability) of all critical reagents
(including description of the antigen(s) and capture antibody(ies)),
and protocols for maintaining product integrity throughout its labeled
shelf life.
(iii) Risk analysis and management strategies, such as Failure
Modes Effects
[[Page 78277]]
Analysis and/or Hazard Analysis and Critical Control Points summaries
and their impact on assay performance.
(iv) Final release criteria to be used for manufactured assay lots
with appropriate evidence that lots released at the extremes of the
specifications will meet the identified analytical and clinical
performance characteristics as well as stability.
(v) Stability studies for reagents must include documentation of an
assessment of real-time stability for multiple reagent lots using the
indicated specimen types and must use acceptance criteria that ensure
that analytical and clinical performance characteristics are met when
stability is assigned based on the extremes of the acceptance range.
(vi) All stability protocols, including acceptance criteria.
(vii) When applicable, analytical sensitivity of the assay is the
same or better than that of other cleared or approved assays.
(viii) Analytical performance studies and results for determining
the limit of blank (LoB), limit of detection (LoD), cutoff, precision
(reproducibility), including lot-to-lot and/or instrument-to-instrument
precision, interference, cross reactivity, carryover, hook effect,
seroconversion panel testing, matrix equivalency, specimen stability,
reagent stability, and cross-genotype antibody detection sensitivity,
when appropriate.
(ix) For devices intended for the detection of antibodies for which
a standardized reference material (that FDA has determined is
appropriate) is available, the analytical sensitivity study and results
testing the standardized reference material. Detailed documentation of
that study and its results must be provided, including the study
protocol, study report, testing results, and all statistical analyses.
(x) For devices with associated software or instrumentation,
documentation must include a detailed description of device software,
including software applications and hardware-based devices that
incorporate software. The detailed description must include
documentation of verification, validation, and hazard analysis and risk
assessment activities, including an assessment of the impact of threats
and vulnerabilities on device functionality and end users/patients as
part of cybersecurity review.
(xi) Detailed documentation of clinical performance testing from a
clinical study with an appropriate number of HBV reactive and non-
reactive samples in applicable risk categories and conducted in the
appropriate settings by the intended users. Performance must be
analyzed relative to an FDA cleared or approved HBV antibody assay or a
comparator that FDA has determined is appropriate. Additional relevant
patient groups must be validated as appropriate. The samples must
include prospective (sequential) samples for each identified specimen
type and, as appropriate, additional characterized clinical samples.
Samples must be sourced from geographically diverse areas.
(3) For any HBV antibody assay intended for quantitative detection
of anti-HBV antibodies, the following special controls, in addition to
those special controls listed in paragraphs (b)(1) and (2) of this
section, also apply:
(i) Detailed documentation of the metrological calibration
traceability hierarchy to a standardized reference material that FDA
has determined is appropriate.
(ii) Detailed documentation of the following analytical performance
studies conducted, as appropriate to the technology, specimen types
tested, and intended use of the device, including upper and lower
limits of quantitation (UloQ and LloQ, respectively), linearity using
clinical samples, and an accuracy study using the recognized
international standard material.
4. Add Sec. 866.3180 to subpart D to read as follows:
Sec. 866.3180 Hepatitis B virus nucleic acid-based assays.
(a) Identification. A nucleic acid-based hepatitis B virus (HBV)
assay is identified as an in vitro diagnostic device intended for
prescription use in the detection of HBV nucleic acid in specimens from
individuals with antibody evidence of HBV infection. In these devices,
the detection of HBV nucleic acid is used as an aid in the management
of HBV-infected individuals. The assay is intended for use with human
serum or plasma (and other matrices as applicable) from individuals
with HBV. The assay is not intended for use as a donor screening assay
for the presence of HBV nucleic acids in blood, blood products, plasma,
cells, or tissue donors, or as a diagnostic assay to confirm the
presence of HBV infection.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Labeling required under Sec. 809.10(b) of this chapter must
include:
(i) A prominent statement that the assay is not intended for use as
a screening assay for the presence of HBV DNA in blood or blood
products, plasma, cells, or tissue donors, or as a diagnostic assay to
confirm the presence of HBV infection.
(ii) A detailed explanation of the principles of operation and
procedures for performing the assay.
(iii) A detailed explanation of the interpretation of results.
(iv) Limitations, which must be updated to reflect current clinical
practice and disease presentation and/or management. These limitations
must include statements that indicate:
(A) Management of patients undergoing hepatitis B virus treatment
should not be established on the basis of a single assay result but
should be determined by a licensed healthcare professional in
conjunction with the clinical presentation, history, and other
diagnostic procedures, e.g., HBV serologic testing, liver function
assays, liver elastography, etc.
(B) The specimen types for which the device has been cleared, and
that use of this assay with specimen types other than those
specifically cleared for this device may result in inaccurate assay
results.
(C) The results obtained with this assay may not be used
interchangeably with results obtained with a different manufacturer's
assay.
(2) Design verification and validation must include the following:
(i) Detailed device description, including the device components,
ancillary reagents required but not provided, and an explanation of the
device methodology. Additional information appropriate to the
technology must be included such as design of primers and probes,
rationale for the selected gene targets, specifications for amplicon
size, and degree of nucleic acid sequence conservation.
(ii) For devices with assay calibrators, the design and composition
of all primary, secondary, and subsequent quantitation standards used
for calibration as well as their traceability to a standardized
reference material that FDA has determined is appropriate (e.g., a
recognized consensus standard). In addition, analytical testing must be
performed following the release of a new lot of the standard material
that was used for device clearance or approval, or when there is a
transition to a new calibration standard.
(iii) Documentation and characterization (e.g., determination of
the identity, supplier, purity, and stability) of all critical reagents
(including nucleic acid sequences for primers and probes) and protocols
for maintaining product integrity.
(iv) Risk analysis and management strategies demonstrating how risk
[[Page 78278]]
control measures are implemented to address device system hazards, such
as Failure Modes Effects Analysis and/or Hazard Analysis and Critical
Control Points summaries and their impact on assay performance.
(v) Final release criteria to be used for manufactured assay lots
with appropriate evidence that lots released at the extremes of the
specification will meet the identified analytical and clinical
performance characteristics as well as stability.
(vi) Stability studies for reagents must include documentation of
an assessment of real-time stability for multiple reagent lots using
the indicated specimen types and must use acceptance criteria that
ensure that analytical and clinical performance characteristics are met
when stability is assigned based on the extremes of the acceptance
range.
(vii) All stability protocols, including acceptance criteria.
(viii) Detailed documentation of analytical performance studies
conducted as appropriate to the technology, specimen types tested, and
intended use of the device, including limit of detection (LoD),
linearity, precision, endogenous and exogenous interferences, cross-
reactivity, carryover, matrix equivalency, sample and reagents
stability, and as applicable, upper and lower limits of quantitation
(ULoQ and LLoQ, respectively). Samples selected for use must be from
subjects with clinically relevant circulating genotypes in the United
States. Cross-reactivity studies must include samples from HBV nucleic
acid negative subjects with other viral or non-viral causes of liver
disease, including autoimmune hepatitis, alcoholic liver disease,
chronic hepatitis C virus (HCV), primary biliary cirrhosis, and
nonalcoholic steatohepatitis, when applicable. The effect of each
identified nucleic-acid isolation and purification procedure on
detection must be evaluated.
(ix) For devices with associated software or instrumentation,
documentation must include a detailed description of device software,
including software applications and hardware-based devices that
incorporate software. The detailed description must include
documentation of verification, validation, and hazard analysis and risk
assessment activities, including an assessment of the impact of threats
and vulnerabilities on device functionality and end users/patients as
part of cybersecurity review.
(x) Detailed documentation of performance from a clinical study
with a design and number of clinical samples (appropriately
statistically powered) that is appropriate for the intended use of the
device as well as conducted in the appropriate settings by the intended
users. The samples must include prospective (sequential) samples for
each claimed specimen type and, as appropriate, additional
characterized clinical samples. Samples must be sourced from
geographically diverse areas.
Dated: September 20, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024-21932 Filed 9-24-24; 8:45 am]
BILLING CODE 4164-01-P