Medical Devices; Immunology and Microbiology Devices; Classification of the Whole Exome Sequencing Constituent Device, 73565-73568 [2024-20550]
Download as PDF
73565
Rules and Regulations
Federal Register
Vol. 89, No. 176
Wednesday, September 11, 2024
This section of the FEDERAL REGISTER
contains regulatory documents having general
applicability and legal effect, most of which
are keyed to and codified in the Code of
Federal Regulations, which is published under
50 titles pursuant to 44 U.S.C. 1510.
The Code of Federal Regulations is sold by
the Superintendent of Documents.
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA–2024–N–4018]
Medical Devices; Immunology and
Microbiology Devices; Classification of
the Whole Exome Sequencing
Constituent Device
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final amendment; final order.
The Food and Drug
Administration (FDA, Agency, or we) is
classifying the whole exome sequencing
constituent device into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the whole exome
sequencing constituent device’s
classification. We are taking this action
because we have determined that
classifying the device into class II
(special controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices.
DATES: This order is effective September
11, 2024. The classification was
applicable on December 23, 2020.
FOR FURTHER INFORMATION CONTACT:
Zivana Tezak, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3424, Silver Spring,
MD 20993–0002, 301–796–6206,
Zivana.Tezak@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
lotter on DSK11XQN23PROD with RULES1
SUMMARY:
I. Background
Upon request, FDA has classified the
whole exome sequencing constituent
device as class II (special controls),
which we have determined will provide
a reasonable assurance of safety and
effectiveness.
VerDate Sep<11>2014
15:55 Sep 10, 2024
Jkt 262001
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (see 21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
device by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D
(21 CFR part 860, subpart D)). Section
207 of the Food and Drug
Administration Modernization Act of
1997 (Pub. L. 105–115) established the
first procedure for De Novo
classification. Section 607 of the Food
and Drug Administration Safety and
Innovation Act (Pub. L. 112–144)
modified the De Novo application
process by adding a second procedure.
A device sponsor may utilize either
procedure for De Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
PO 00000
Frm 00001
Fmt 4700
Sfmt 4700
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
placed within class III, the De Novo
classification is considered to be the
initial classification of the device.
When FDA classifies a device into
class I or II via the De Novo process, the
device can serve as a predicate for
future devices of that type, including for
510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device
sponsors do not have to submit a De
Novo request or premarket approval
application to market a substantially
equivalent device (see section 513(i) of
the FD&C Act, defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the 510(k) process, when necessary, to
market their device.
II. De Novo Classification
On August 2, 2019, FDA received
Helix OpCo, LLC’s request for De Novo
classification of the Helix Learning
Platform. FDA reviewed the request in
order to classify the device under the
criteria for classification set forth in
section 513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on December 23, 2020,
FDA issued an order to the requester
classifying the device into class II. In
this final order, FDA is codifying the
classification of the device by adding 21
E:\FR\FM\11SER1.SGM
11SER1
73566
Federal Register / Vol. 89, No. 176 / Wednesday, September 11, 2024 / Rules and Regulations
CFR 866.6000.1 We have named the
generic type of device whole exome
sequencing constituent device, and it is
identified as a device for germline
whole exome sequencing of genomic
deoxyribonucleic acid (DNA) isolated
from human specimens. The DNA
sequence generated by this device is
intended as input for clinical germline
DNA assays that have FDA marketing
authorization and are intended for use
with this device.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
TABLE 1—WHOLE EXOME SEQUENCING CONSTITUENT DEVICE RISKS AND MITIGATION MEASURES
Identified risks to health
Mitigation measures
Inaccurate test results and failure to
provide results.
Incorrect application or interpretation of results.
User error and improper use of the
device.
Certain design verification and validation, including certain analytical studies and clinical studies; and Certain labeling information, including certain performance information and device limitations.
Certain design verification and validation, including certain clinical studies; and Certain labeling information,
including certain performance information and device limitations.
Certain design verification and validation, including certain analytical studies and clinical studies; and Certain labeling information, including certain performance information and device limitations.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
lotter on DSK11XQN23PROD with RULES1
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in part 860,
subpart D, regarding De Novo
classification have been approved under
OMB control number 0910–0844; the
collections of information in 21 CFR
part 814, subparts A through E,
regarding premarket approval, have
been approved under OMB control
number 0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
1 FDA notes that the ‘‘ACTION’’ caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
VerDate Sep<11>2014
15:55 Sep 10, 2024
Jkt 262001
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
parts 801 and 809, regarding labeling,
have been approved under OMB control
number 0910–0485.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical
devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 866 is
amended as follows:
PART 866—IMMUNOLOGY AND
MICROBIOLOGY DEVICES
1. The authority citation for part 866
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 866.6000 to subpart G to read
as follows:
■
§ 866.6000 Whole exome sequencing
constituent device.
(a) Identification. A whole exome
sequencing constituent device is for
germline whole exome sequencing of
genomic deoxyribonucleic acid (DNA)
isolated from human specimens. The
DNA sequence generated by this device
is intended as input for clinical
germline DNA assays that have FDA
marketing authorization and are
intended for use with this device.
(b) Classification. Class II (special
controls). The special controls for this
device are:
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
PO 00000
Frm 00002
Fmt 4700
Sfmt 4700
(1) The intended use on the device’s
label and labeling required under
§ 809.10 of this chapter must include:
(i) The indicated variant types for
which acceptable, as determined by
FDA, validation data has been provided.
Distinct variant types are considered as
single nucleotide variant, insertion,
deletion, tandem repeats, copy number
variants, or gene rearrangements, and
validated for specific sizes and lengths,
as applicable.
(ii) The indicated specimen type(s) for
which acceptable, as determined by
FDA, validation data has been provided.
(2) The labeling required under
§ 809.10(b) of this chapter must include:
(i) The identification of, or the
specifications for, the collection device
or devices to be used for sample
collection, as applicable.
(ii) A description of the reportable
range, which is the region of the genome
for which the assay is intended to
provide results, as well as a description
of the targeted regions of the genome
that have enhanced coverage. This must
include a description of any genomic
regions that are excluded from the
reportable region due to unacceptable
risk of erroneous results, or for other
reasons. A description of the clinically
relevant genes excluded from the
reportable range must also be included,
if applicable.
(iii) A description of the design
features and control elements, including
the quality metrics and thresholds
which are used for reporting the
analytical range (the genomic DNA in
the reportable range that passed the
quality metrics in the run required for
reporting to the user) that are
incorporated into the testing procedure,
that mitigate the risk of incorrect
clinical results. The following metrics
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
E:\FR\FM\11SER1.SGM
11SER1
lotter on DSK11XQN23PROD with RULES1
Federal Register / Vol. 89, No. 176 / Wednesday, September 11, 2024 / Rules and Regulations
are considered applicable in the
generation of high confidence data and
the established thresholds for these
metrics for reporting must be described
and be determined to be acceptable by
FDA: cluster density and percent of
cluster pass quality filter, percent of
bases meeting the minimum base
quality score, average coverage of reads,
percent of reads mapped on target,
percent of reportable region with
coverage meeting the minimum
requirement, percent of unassigned read
indices, percent of reads for non-human
DNA, allele fraction, and strand bias.
Any alternate metrics used must be
described and an acceptable, as
determined by FDA, rationale for
applicability must be provided.
(iv) A representative sample of the
device output report(s) provided to
users, which must include any relevant
limitations of the device, as determined
applicable by FDA.
(3) Design verification and validation
must include:
(i) A detailed description of the
impact of any software, including
software applications and hardwarebased devices that incorporate software,
on the device’s function.
(ii) Acceptable data, as determined by
FDA, demonstrating how the key quality
metrics and quality metric thresholds in
the list in paragraph (b)(2)(iii) of this
section for reporting were established
and optimized for accuracy using
appropriate DNA standards with
established reference genomic sequence.
Data must include, as applicable, base
quality score, allele fraction for
heterozygosity and coverage, and other
applicable metrics.
(iii) Data demonstrating acceptable, as
determined by FDA, analytical device
performance using patient specimens
representing the full spectrum of
expected variant types reported across
the genome and in genomic regions that
are difficult to sequence. The number of
specimens tested must be sufficient to
obtain estimates of device performance
that are representative of the device
performance that can be expected for
the reportable region and clinically
relevant subsets of the reportable region,
as applicable. For each study, data must
include a summary of the key quality
metric data; the number and percentage
of true positives (TP), false positives
(FP), and false negatives (FN); number
and percentage of no-calls; positive
percent agreement (PPA); negative
percent agreement (NPA); positive
predictive value (PPV); technical
positive percent value (TPPV); and nonreference concordance (NRC). These
data must be provided per sample and
stratified by variant type. The variant
VerDate Sep<11>2014
15:55 Sep 10, 2024
Jkt 262001
data must also be further stratified by
size and zygosity (homozygous common
allele, heterozygous, homozygous rare
allele). Data demonstrating the accuracy
assay based on guanine and cytosine
(GC) content, pseudogenes, and
proximity to short tandem repeats must
also be presented. The data must be
presented for the entire exome and also
for clinically relevant subsets of the
reportable region. For each study, the
number of run failures and repeat/
requeued specimens must be
summarized.
(iv) Documentation of acceptance
criteria that are applied to analytical
and clinical validation studies, which
must be justified based on the estimated
risk of erroneous results on clinically
significant genes and variants and must
be clinically acceptable, as determined
by FDA. The acceptance criteria must be
pre-specified prior to clinical and
analytical validation studies, and all
validation testing results must be
documented with respect to those
acceptance criteria.
(v) Analytical validation must be
demonstrated by conducting studies
that provide:
(A) Data demonstrating acceptable, as
determined by FDA, accuracy based on
agreement with an acceptable, as
determined by FDA, comparator
method(s) that has been validated to
have high accuracy and reproducibility.
Accuracy of the test shall be evaluated
with reference standards and clinical
specimens for each indicated specimen
type of a number determined acceptable
by FDA, collected and processed in a
manner consistent with the test’s
instructions for use.
(B) Data demonstrating acceptable, as
determined by FDA, precision from a
precision study using clinical samples
to adequately evaluate intra-run, interrun, and total variability across
operator, instrument, lot, day, and site,
as applicable. The samples must include
the indicated range of DNA input.
Precision, including repeatability and
reproducibility, must be assessed by
agreement between replicates, and also
supported by sequencing quality metrics
for targeted regions across the panel.
Precision must be demonstrated per
specimen and in aggregate. Precision
data must be calculated and presented
with and without no calls/invalid
results.
(C) Data demonstrating acceptable, as
determined by FDA, accuracy in the
presence of clinically relevant levels of
potential interfering substances that are
present in the specimen type and
intended use population, including, for
example, endogenous substances,
PO 00000
Frm 00003
Fmt 4700
Sfmt 4700
73567
exogenous substances, and microbes, as
applicable.
(D) Data demonstrating the absence of
sample cross contamination due to
index swapping (misassignment).
(E) Data demonstrating that the preanalytical steps such as DNA extraction
are robust such that sources of
variability in these steps and procedures
do not diminish the accuracy and
precision of the device.
(F) Data demonstrating that
acceptable, as determined by FDA,
device performance is maintained
across the range of claimed DNA input
concentrations for the assay.
(vi) Design verification and validation
for software within the whole exome
sequencing constituent device must
include the following:
(A) Detailed description of the
software, including specifications and
requirements for the format of data
input and output, such that users can
determine if the device conforms to user
needs and intended uses.
(B) Device design must include a
detailed strategy to ensure cybersecurity
risks that could lead to loss of genetic
data security, are adequately addressed
and mitigated (including device
interface specifications and how safe
reporting of the genetic test is
maintained when software is updated).
Verification and validation must
include security testing to demonstrate
effectiveness of the associated controls.
(C) Device design must ensure that a
record of critical events, including a
record of all genetic test orders using the
whole exome sequencing constituent
device, device malfunctions, and
associated acknowledgments, is stored
and accessible for an adequate period to
allow for auditing of communications
between the whole exome sequencing
constituent device and downstream
clinical genetic tests, and to facilitate
the sharing of pertinent information
with the responsible parties for those
devices.
(vii) A protocol reviewed and
determined acceptable by FDA, that
specifies the verification and validation
activities that will be performed for
anticipated bioinformatic software
modifications to reevaluate performance
claims or performance specifications.
This protocol must include a process for
assessing whether a modification to the
bioinformatics software could
significantly affect the safety or
effectiveness of the device. The protocol
must include assessment metrics,
acceptance criteria, and analytical
methods for the performance testing of
changes, as applicable. The protocol
must also include the process for
communicating to developers of
E:\FR\FM\11SER1.SGM
11SER1
73568
Federal Register / Vol. 89, No. 176 / Wednesday, September 11, 2024 / Rules and Regulations
downstream clinical genetic tests the
impact of the bioinformatics software
change on the whole exome sequencing
constituent system genetic data output
so they may implement appropriate
corresponding actions.
Dated: September 6, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
ENVIRONMENTAL PROTECTION
AGENCY
40 CFR Part 52
[EPA–R09–OAR–2023–0494; FRL–11442–
02–R9]
Air Plan Approval; California; South
Coast Air Quality Management District
Environmental Protection
Agency (EPA).
ACTION: Final rule.
AGENCY:
[FR Doc. 2024–20550 Filed 9–10–24; 8:45 am]
BILLING CODE 4164–01–P
The Environmental Protection
Agency (EPA) is taking final action on
a revision to the South Coast Air Quality
Management District (SCAQMD or ‘‘the
District’’) portion of the California State
Implementation Plan (SIP). This
revision concerns the regulation of
emissions of oxides of nitrogen (NOX)
and particulate matter (PM) associated
with warehouses as indirect sources that
attract or may attract mobile source
emissions. The EPA is approving
SCAQMD Rule 2305, ‘‘Warehouse
Indirect Source Rule—Warehouse
Actions and Investments to Reduce
Emissions (WAIRE) Program,’’ to
regulate these emission sources under
the Clean Air Act (CAA or ‘‘the Act’’) as
a SIP strengthening.
DATES: This rule is effective October 11,
2024.
ADDRESSES: The EPA has established a
docket for this action under Docket ID
No. EPA–R09–OAR–2023–0494. All
documents in the docket are listed on
the https://www.regulations.gov
website. Although listed in the index,
some information is not publicly
available, e.g., Confidential Business
Information (CBI) or other information
SUMMARY:
DEPARTMENT OF THE TREASURY
Internal Revenue Service
26 CFR Part 1
Income Taxes
CFR Correction
This rule is being published by the
Office of the Federal Register to correct
an editorial or technical error that
appeared in the most recent annual
revision of the Code of Federal
Regulations.
In Title 26 of the Code of Federal
Regulations, Part 1 (§§ 1.410 to 1.440),
revised as of April 1, 2024, in section
1.430(h)(2)–1, remove paragraph (ii)
immediately following paragraph (b)(2).
[FR Doc. 2024–20701 Filed 9–10–24; 8:45 am]
BILLING CODE 0099–10–P
whose disclosure is restricted by statute.
Certain other material, such as
copyrighted material, is not placed on
the internet and will be publicly
available only in hard copy form.
Publicly available docket materials are
available through https://
www.regulations.gov, or please contact
the person identified in the FOR FURTHER
INFORMATION CONTACT section for
additional availability information. If
you need assistance in a language other
than English or if you are a person with
a disability who needs a reasonable
accommodation at no cost to you, please
contact the person identified in the FOR
FURTHER INFORMATION CONTACT section.
FOR FURTHER INFORMATION CONTACT: La
Kenya Evans-Hopper, EPA Region IX, 75
Hawthorne St., San Francisco, CA
94105; phone: (415) 972–3245; email:
evanshopper.lakenya@epa.gov.
SUPPLEMENTARY INFORMATION:
Throughout this document, ‘‘we,’’ ‘‘us,’’
and ‘‘our’’ refer to the EPA.
Table of Contents
I. Proposed Action
II. Public Comments and EPA Responses
III. EPA Action
IV. Incorporation by Reference
V. Statutory and Executive Order Reviews
I. Proposed Action
On October 12, 2023 (88 FR 70616)
(‘‘proposed rule’’), the EPA proposed to
approve SCAQMD Rule 2305 as a
revision to the SCAQMD portion of the
California SIP. Table 1 lists the
SCAQMD rule addressed by the
proposed rule with the dates that it was
adopted by the SCAQMD and submitted
by the California Air Resources Board
(CARB).
TABLE 1—SUBMITTED RULE
Local agency
Rule No.
lotter on DSK11XQN23PROD with RULES1
SCAQMD ................................
2305
As described in the proposed rule, the
purpose of SCAQMD Rule 2305 is to
reduce local and regional emissions of
NOX and PM, and to facilitate local and
regional emission reductions associated
with warehouses and the mobile sources
attracted to warehouses in the
SCAQMD, to meet State and Federal air
quality standards for ozone and fine PM
(PM2.5).1 The rule applies within the
jurisdiction of the SCAQMD, which
includes all of Orange County, the nondesert portions of Los Angeles and San
Bernardino counties, and all of
1 88
FR 70616, 70617 (October 12, 2023).
VerDate Sep<11>2014
15:55 Sep 10, 2024
Jkt 262001
Rule title
Adopted
Warehouse Indirect Source Rule—Warehouse Actions and
Investments to Reduce Emissions (WAIRE) Program.
Riverside County (except for the Palo
Verde Valley in far eastern Riverside
County).
Through the adoption of the 2016
South Coast Air Quality Management
Plan (AQMP), the SCAQMD adopted
certain ‘‘facility-based mobile source
measures,’’ including a measure under
which the SCAQMD committed to
assess and identify potential actions to
further reduce emissions from emission
sources associated with warehouse
distribution centers.2 In 2019, the EPA
2 SCAQMD, Final 2016 Air Quality Management
Plan, March 2017, pp. 4–25, 4–28 and 4–29. The
PO 00000
Frm 00004
Fmt 4700
Sfmt 4700
05/07/2021
Submitted
08/13/2021
approved the ozone portions of the 2016
South Coast AQMP, including the
commitment to develop facility-based
mobile source measures, including the
measure focused on warehouse
distribution centers.3 The 2016 AQMP
does not include an emission reduction
estimate for the facility-based mobile
source measure related to warehouses.
In 2021, after assessing potential actions
to further reduce emissions associated
2016 South Coast AQMP designates the warehouse
measure as MOB–03 (‘‘Emission Reductions at
Warehouse Distribution Centers’’).
3 84 FR 52005 (October 1, 2019).
E:\FR\FM\11SER1.SGM
11SER1
Agencies
[Federal Register Volume 89, Number 176 (Wednesday, September 11, 2024)]
[Rules and Regulations]
[Pages 73565-73568]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2024-20550]
========================================================================
Rules and Regulations
Federal Register
________________________________________________________________________
This section of the FEDERAL REGISTER contains regulatory documents
having general applicability and legal effect, most of which are keyed
to and codified in the Code of Federal Regulations, which is published
under 50 titles pursuant to 44 U.S.C. 1510.
The Code of Federal Regulations is sold by the Superintendent of Documents.
========================================================================
Federal Register / Vol. 89, No. 176 / Wednesday, September 11, 2024 /
Rules and Regulations
[[Page 73565]]
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA-2024-N-4018]
Medical Devices; Immunology and Microbiology Devices;
Classification of the Whole Exome Sequencing Constituent Device
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, Agency, or we) is
classifying the whole exome sequencing constituent device into class II
(special controls). The special controls that apply to the device type
are identified in this order and will be part of the codified language
for the whole exome sequencing constituent device's classification. We
are taking this action because we have determined that classifying the
device into class II (special controls) will provide a reasonable
assurance of safety and effectiveness of the device. We believe this
action will also enhance patients' access to beneficial innovative
devices.
DATES: This order is effective September 11, 2024. The classification
was applicable on December 23, 2020.
FOR FURTHER INFORMATION CONTACT: Zivana Tezak, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3424, Silver Spring, MD 20993-0002, 301-796-6206,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the whole exome sequencing
constituent device as class II (special controls), which we have
determined will provide a reasonable assurance of safety and
effectiveness.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device
that does not require premarket approval. We determine whether a new
device is substantially equivalent to a predicate device by means of
the procedures for premarket notification under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo application
process by adding a second procedure. A device sponsor may utilize
either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
When FDA classifies a device into class I or II via the De Novo
process, the device can serve as a predicate for future devices of that
type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C
Act). As a result, other device sponsors do not have to submit a De
Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
510(k) process, when necessary, to market their device.
II. De Novo Classification
On August 2, 2019, FDA received Helix OpCo, LLC's request for De
Novo classification of the Helix Learning Platform. FDA reviewed the
request in order to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on December 23, 2020, FDA issued an order to the
requester classifying the device into class II. In this final order,
FDA is codifying the classification of the device by adding 21
[[Page 73566]]
CFR 866.6000.\1\ We have named the generic type of device whole exome
sequencing constituent device, and it is identified as a device for
germline whole exome sequencing of genomic deoxyribonucleic acid (DNA)
isolated from human specimens. The DNA sequence generated by this
device is intended as input for clinical germline DNA assays that have
FDA marketing authorization and are intended for use with this device.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Whole Exome Sequencing Constituent Device Risks and Mitigation
Measures
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Inaccurate test results and Certain design verification and
failure to provide results. validation, including certain
analytical studies and clinical
studies; and Certain labeling
information, including certain
performance information and device
limitations.
Incorrect application or Certain design verification and
interpretation of results. validation, including certain
clinical studies; and Certain
labeling information, including
certain performance information and
device limitations.
User error and improper use of the Certain design verification and
device. validation, including certain
analytical studies and clinical
studies; and Certain labeling
information, including certain
performance information and device
limitations.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions, have been approved under OMB
control number 0910-0120; the collections of information in 21 CFR part
820, regarding quality system regulation, have been approved under OMB
control number 0910-0073; and the collections of information in 21 CFR
parts 801 and 809, regarding labeling, have been approved under OMB
control number 0910-0485.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
866 is amended as follows:
PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES
0
1. The authority citation for part 866 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 866.6000 to subpart G to read as follows:
Sec. 866.6000 Whole exome sequencing constituent device.
(a) Identification. A whole exome sequencing constituent device is
for germline whole exome sequencing of genomic deoxyribonucleic acid
(DNA) isolated from human specimens. The DNA sequence generated by this
device is intended as input for clinical germline DNA assays that have
FDA marketing authorization and are intended for use with this device.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The intended use on the device's label and labeling required
under Sec. 809.10 of this chapter must include:
(i) The indicated variant types for which acceptable, as determined
by FDA, validation data has been provided. Distinct variant types are
considered as single nucleotide variant, insertion, deletion, tandem
repeats, copy number variants, or gene rearrangements, and validated
for specific sizes and lengths, as applicable.
(ii) The indicated specimen type(s) for which acceptable, as
determined by FDA, validation data has been provided.
(2) The labeling required under Sec. 809.10(b) of this chapter
must include:
(i) The identification of, or the specifications for, the
collection device or devices to be used for sample collection, as
applicable.
(ii) A description of the reportable range, which is the region of
the genome for which the assay is intended to provide results, as well
as a description of the targeted regions of the genome that have
enhanced coverage. This must include a description of any genomic
regions that are excluded from the reportable region due to
unacceptable risk of erroneous results, or for other reasons. A
description of the clinically relevant genes excluded from the
reportable range must also be included, if applicable.
(iii) A description of the design features and control elements,
including the quality metrics and thresholds which are used for
reporting the analytical range (the genomic DNA in the reportable range
that passed the quality metrics in the run required for reporting to
the user) that are incorporated into the testing procedure, that
mitigate the risk of incorrect clinical results. The following metrics
[[Page 73567]]
are considered applicable in the generation of high confidence data and
the established thresholds for these metrics for reporting must be
described and be determined to be acceptable by FDA: cluster density
and percent of cluster pass quality filter, percent of bases meeting
the minimum base quality score, average coverage of reads, percent of
reads mapped on target, percent of reportable region with coverage
meeting the minimum requirement, percent of unassigned read indices,
percent of reads for non-human DNA, allele fraction, and strand bias.
Any alternate metrics used must be described and an acceptable, as
determined by FDA, rationale for applicability must be provided.
(iv) A representative sample of the device output report(s)
provided to users, which must include any relevant limitations of the
device, as determined applicable by FDA.
(3) Design verification and validation must include:
(i) A detailed description of the impact of any software, including
software applications and hardware-based devices that incorporate
software, on the device's function.
(ii) Acceptable data, as determined by FDA, demonstrating how the
key quality metrics and quality metric thresholds in the list in
paragraph (b)(2)(iii) of this section for reporting were established
and optimized for accuracy using appropriate DNA standards with
established reference genomic sequence. Data must include, as
applicable, base quality score, allele fraction for heterozygosity and
coverage, and other applicable metrics.
(iii) Data demonstrating acceptable, as determined by FDA,
analytical device performance using patient specimens representing the
full spectrum of expected variant types reported across the genome and
in genomic regions that are difficult to sequence. The number of
specimens tested must be sufficient to obtain estimates of device
performance that are representative of the device performance that can
be expected for the reportable region and clinically relevant subsets
of the reportable region, as applicable. For each study, data must
include a summary of the key quality metric data; the number and
percentage of true positives (TP), false positives (FP), and false
negatives (FN); number and percentage of no-calls; positive percent
agreement (PPA); negative percent agreement (NPA); positive predictive
value (PPV); technical positive percent value (TPPV); and non-reference
concordance (NRC). These data must be provided per sample and
stratified by variant type. The variant data must also be further
stratified by size and zygosity (homozygous common allele,
heterozygous, homozygous rare allele). Data demonstrating the accuracy
assay based on guanine and cytosine (GC) content, pseudogenes, and
proximity to short tandem repeats must also be presented. The data must
be presented for the entire exome and also for clinically relevant
subsets of the reportable region. For each study, the number of run
failures and repeat/requeued specimens must be summarized.
(iv) Documentation of acceptance criteria that are applied to
analytical and clinical validation studies, which must be justified
based on the estimated risk of erroneous results on clinically
significant genes and variants and must be clinically acceptable, as
determined by FDA. The acceptance criteria must be pre-specified prior
to clinical and analytical validation studies, and all validation
testing results must be documented with respect to those acceptance
criteria.
(v) Analytical validation must be demonstrated by conducting
studies that provide:
(A) Data demonstrating acceptable, as determined by FDA, accuracy
based on agreement with an acceptable, as determined by FDA, comparator
method(s) that has been validated to have high accuracy and
reproducibility. Accuracy of the test shall be evaluated with reference
standards and clinical specimens for each indicated specimen type of a
number determined acceptable by FDA, collected and processed in a
manner consistent with the test's instructions for use.
(B) Data demonstrating acceptable, as determined by FDA, precision
from a precision study using clinical samples to adequately evaluate
intra-run, inter-run, and total variability across operator,
instrument, lot, day, and site, as applicable. The samples must include
the indicated range of DNA input. Precision, including repeatability
and reproducibility, must be assessed by agreement between replicates,
and also supported by sequencing quality metrics for targeted regions
across the panel. Precision must be demonstrated per specimen and in
aggregate. Precision data must be calculated and presented with and
without no calls/invalid results.
(C) Data demonstrating acceptable, as determined by FDA, accuracy
in the presence of clinically relevant levels of potential interfering
substances that are present in the specimen type and intended use
population, including, for example, endogenous substances, exogenous
substances, and microbes, as applicable.
(D) Data demonstrating the absence of sample cross contamination
due to index swapping (misassignment).
(E) Data demonstrating that the pre-analytical steps such as DNA
extraction are robust such that sources of variability in these steps
and procedures do not diminish the accuracy and precision of the
device.
(F) Data demonstrating that acceptable, as determined by FDA,
device performance is maintained across the range of claimed DNA input
concentrations for the assay.
(vi) Design verification and validation for software within the
whole exome sequencing constituent device must include the following:
(A) Detailed description of the software, including specifications
and requirements for the format of data input and output, such that
users can determine if the device conforms to user needs and intended
uses.
(B) Device design must include a detailed strategy to ensure
cybersecurity risks that could lead to loss of genetic data security,
are adequately addressed and mitigated (including device interface
specifications and how safe reporting of the genetic test is maintained
when software is updated). Verification and validation must include
security testing to demonstrate effectiveness of the associated
controls.
(C) Device design must ensure that a record of critical events,
including a record of all genetic test orders using the whole exome
sequencing constituent device, device malfunctions, and associated
acknowledgments, is stored and accessible for an adequate period to
allow for auditing of communications between the whole exome sequencing
constituent device and downstream clinical genetic tests, and to
facilitate the sharing of pertinent information with the responsible
parties for those devices.
(vii) A protocol reviewed and determined acceptable by FDA, that
specifies the verification and validation activities that will be
performed for anticipated bioinformatic software modifications to
reevaluate performance claims or performance specifications. This
protocol must include a process for assessing whether a modification to
the bioinformatics software could significantly affect the safety or
effectiveness of the device. The protocol must include assessment
metrics, acceptance criteria, and analytical methods for the
performance testing of changes, as applicable. The protocol must also
include the process for communicating to developers of
[[Page 73568]]
downstream clinical genetic tests the impact of the bioinformatics
software change on the whole exome sequencing constituent system
genetic data output so they may implement appropriate corresponding
actions.
Dated: September 6, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024-20550 Filed 9-10-24; 8:45 am]