Medical Devices; Cardiovascular Devices; Classification of the Adjunctive Open Loop Fluid Therapy Recommender, 72317-72319 [2024-19736]
Download as PDF
Federal Register / Vol. 89, No. 172 / Thursday, September 5, 2024 / Rules and Regulations
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
parts 801 and 809, regarding labeling,
have been approved under OMB control
number 0910–0485.
List of Subjects in 21 CFR Part 864
Blood, Medical devices, Packaging
and containers.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 864 is
amended as follows:
PART 864—HEMATOLOGY AND
PATHOLOGY DEVICES
1. The authority citation for part 864
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 864.7295 to subpart H to read
as follows:
■
khammond on DSKJM1Z7X2PROD with RULES
§ 864.7295 Heparin and direct oral factor
Xa inhibitor drug test system.
(a) Identification. A heparin and
direct oral factor Xa inhibitor drug test
system is intended for the detection of
heparin and direct oral factor Xa
inhibitors in human specimens
collected from patients taking heparin
or direct oral factor Xa inhibitors. This
device is intended to aid in the
management of therapy in conjunction
with other clinical and laboratory
findings.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Design verification and validation
must include the following:
(i) Detailed documentation of
analytical device performance studies
and results demonstrating acceptable
analytical performance with a sufficient
number of specimens tested in order to
obtain unbiased estimates of analytical
performance. This documentation shall
include the following as appropriate to
the technology, specimen types tested,
and intended use of the device:
(A) Studies and results for that
demonstrate device precision including
repeatability and reproducibility, using
quality controls and clinical samples,
when appropriate. Precision studies
must assess specimens for each
indicated drug at concentrations
throughout the measuring range of the
device including near clinically relevant
levels, as appropriate. The study must
evaluate different sources of variability
including, as appropriate, between-run,
between-operator, between-lot, between-
VerDate Sep<11>2014
21:04 Sep 04, 2024
Jkt 262001
instrument, between-day, and betweensite;
(B) Studies and results that
demonstrate that the device is free from
clinically significant interference, from
endogenous and exogenous interferents
associated with the target population(s),
and interferents that are specific for, or
related to, the technology or
methodology of the device;
(C) Data to demonstrate appropriate
specimen stability for the intended
sample matrices under the intended
conditions for specimen collection,
handling, and storage described in the
device labeling;
(D) Studies and results that
demonstrate the linear range, limit of
blank (LoB), limit of detection (LoD),
and limit of quantitation (LoQ), as
applicable to the technology of the
device; and
(E) For any devices intended for use
for near patient testing, studies and
results that demonstrate the robustness
of the device in the hands of the
intended user, including the entire
testing procedure, pre-analytical
specimen processing steps, and results
interpretation.
(ii) Detailed documentation of clinical
performance testing in which the
performance is analyzed relative to a
comparator that FDA has determined is
appropriate. Specimens must be
representative of the intended use
population(s) and must cover the full
range of the device output and any
clinically relevant decision points as
appropriate.
(2) The labeling required under
§ 809.10(b) of this chapter must include:
(i) Identification of any known
interferents, including all endogenous,
exogenous, technology-specific, and
patient population-specific interferents,
specific to the test outputs. The
information must include the
concentration(s) or level(s) of the
interferent at which clinically
significant interference was found to
occur, and the concentration range or
levels at which interference was not
found to occur;
(ii) A prominent statement that the
device is not intended for use in
monitoring patients taking heparin or
direct oral factor Xa inhibitors; and
(iii) Limiting statements indicating, as
applicable:
(A) That the device should only be
used in conjunction with information
available from clinical evaluations and
other diagnostic procedures; and
(B) That the device is not specific to
the direct oral factor Xa inhibitor that
has been evaluated and may detect the
presence of other direct factor Xa
inhibitors that have not been evaluated.
PO 00000
Frm 00019
Fmt 4700
Sfmt 4700
72317
Dated: August 29, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024–19824 Filed 9–4–24; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA–2024–N–3947]
Medical Devices; Cardiovascular
Devices; Classification of the
Adjunctive Open Loop Fluid Therapy
Recommender
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final amendment; final order.
The Food and Drug
Administration (FDA, Agency, or we) is
classifying the adjunctive open loop
fluid therapy recommender into class II
(special controls). The special controls
that apply to the device type are
identified in this order and will be part
of the codified language for the
adjunctive open loop fluid therapy
recommender’s classification. We are
taking this action because we have
determined that classifying the device
into class II (special controls) will
provide a reasonable assurance of safety
and effectiveness of the device. We
believe this action will also enhance
patients’ access to beneficial innovative
devices.
DATES: This order is effective September
5, 2024. The classification was
applicable on November 13, 2020.
FOR FURTHER INFORMATION CONTACT:
Biniyam Taddese, Center for Devices
and Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2544, Silver Spring,
MD 20993–0002, 240–402–6570,
Biniyam.Taddese@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
Upon request, FDA has classified the
adjunctive open loop fluid therapy
recommender as class II (special
controls), which we have determined
will provide a reasonable assurance of
safety and effectiveness.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
E:\FR\FM\05SER1.SGM
05SER1
72318
Federal Register / Vol. 89, No. 172 / Thursday, September 5, 2024 / Rules and Regulations
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (see 21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
device by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D
(21 CFR part 860, subpart D)). Section
207 of the Food and Drug
Administration Modernization Act of
1997 (Pub. L. 105–115) established the
first procedure for De Novo
classification. Section 607 of the Food
and Drug Administration Safety and
Innovation Act (Pub. L. 112–144)
modified the De Novo application
process by adding a second procedure.
A device sponsor may utilize either
procedure for De Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
placed within class III, the De Novo
classification is considered to be the
initial classification of the device.
When FDA classifies a device into
class I or II via the De Novo process, the
device can serve as a predicate for
future devices of that type, including for
510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device
sponsors do not have to submit a De
Novo request or premarket approval
application to market a substantially
equivalent device (see section 513(i) of
the FD&C Act, defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the 510(k) process, when necessary, to
market their device.
II. De Novo Classification
On June 4, 2019, FDA received
Edwards Lifesciences’ request for De
Novo classification of the Acumen
Assisted Fluid Management (AFM)
Software Feature. FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1) of the FD&C
Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on November 13, 2020,
FDA issued an order to the requester
classifying the device into class II. In
this final order, FDA is codifying the
classification of the device by adding 21
CFR 870.5600.1 We have named the
generic type of device adjunctive open
loop fluid therapy recommender, and it
is identified as a prescription device
that uses software algorithms to analyze
cardiovascular vital signs and predict a
patient’s estimated response to fluid
therapy. The device is intended for
adjunctive use with other physical vital
sign parameters and patient information
and is not intended to independently
direct therapy.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
TABLE 1—ADJUNCTIVE OPEN LOOP FLUID THERAPY RECOMMENDER RISKS AND MITIGATION MEASURES
Identified risks to health
Mitigation measures
Delay in monitoring or treatment ..............................................................
Software verification, validation, and hazard analysis; Usability assessment; and Labeling.
Software verification, validation, and hazard analysis; Non-clinical performance testing; Usability assessment; Clinical performance testing;
and Labeling.
Software verification, validation, and hazard analysis; Non-clinical performance testing; Usability assessment; Clinical performance testing;
and Labeling.
khammond on DSKJM1Z7X2PROD with RULES
Inappropriate or missed treatment due to over-reliance on software recommendation, which is affected by algorithm or software error, or inaccurate input from sensors or users.
Fluid overload due to over-reliance on software recommendations,
which are affected by algorithm or software error, or inaccurate input
from sensors or users.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
1 FDA notes that the ACTION caption for this final
order is styled as ‘‘Final amendment; final order,’’
rather than ‘‘Final order.’’ Beginning in December
2019, this editorial change was made to indicate
that the document ‘‘amends’’ the Code of Federal
Regulations. The change was made in accordance
with the Office of Federal Register’s (OFR)
interpretations of the Federal Register Act (44
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
VerDate Sep<11>2014
21:04 Sep 04, 2024
Jkt 262001
PO 00000
Frm 00020
Fmt 4700
Sfmt 4700
E:\FR\FM\05SER1.SGM
05SER1
Federal Register / Vol. 89, No. 172 / Thursday, September 5, 2024 / Rules and Regulations
premarket notification requirements
under section 510(k) of the FD&C Act.
At the time of classification, the
adjunctive open loop fluid therapy
recommender is for prescription use
only. Prescription devices are exempt
from the requirement for adequate
directions for use for the layperson
under section 502(f)(1) of the FD&C Act
(21 U.S.C. 352(f)(1)) and 21 CFR 801.5,
as long as the conditions of 21 CFR
801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in part 860,
subpart D, regarding De Novo
classification have been approved under
OMB control number 0910–0844; the
collections of information in 21 CFR
part 814, subparts A through E,
regarding premarket approval, have
been approved under OMB control
number 0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
khammond on DSKJM1Z7X2PROD with RULES
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 870 is
amended as follows:
PART 870—CARDIOVASCULAR
DEVICES
1. The authority citation for part 870
continues to read as follows:
■
VerDate Sep<11>2014
21:04 Sep 04, 2024
Jkt 262001
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 870.5600 to subpart F to read
as follows:
■
§ 870.5600 Adjunctive open loop fluid
therapy recommender
(a) Identification. The adjunctive open
loop fluid therapy recommender is a
prescription device that uses software
algorithms to analyze cardiovascular
vital signs and predict a patient’s
estimated response to fluid therapy. The
device is intended for adjunctive use
with other physical vital sign
parameters and patient information and
is not intended to independently direct
therapy.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Clinical performance testing under
anticipated conditions of use must
fulfill the following:
(i) A summary of the clinical
performance testing must include the
relevant patient demographics, and any
statistical techniques used for analyzing
the data;
(ii) Subjects must be representative of
the intended use population for the
device. Any selection criteria or sample
limitations must be fully described and
justified;
(iii) Testing must demonstrate the
recommendation consistency using the
expected range of data sources and data
quality encountered in the intended
patients, users, and environments; and
(iv) Testing must evaluate the
relationship between algorithm
recommendations, therapeutic actions,
and predicted physiological event or
status.
(2) A software description and the
results of verification and validation
testing based on a comprehensive
hazard analysis and risk assessment
must be provided, including:
(i) A full characterization of the
software technical parameters,
including algorithms;
(ii) A description of the expected
recommendation, accounting for
differences in patient condition and
environment;
(iii) A description of all mitigations
for user error or failure of any subsystem
components (including signal detection,
signal analysis, data display, and
storage) that affect the device’s
recommendations;
(iv) A characterization of algorithm
sensitivity to variations in user inputs;
(v) A characterization of sensor
accuracy and performance;
(vi) A description of sensor data
quality control measures; and
(vii) Safeguards to reduce the
possibility of fluid overload.
PO 00000
Frm 00021
Fmt 4700
Sfmt 9990
72319
(3) A scientific justification for the
validity of the algorithm(s) must be
provided. This justification must
include non-clinical verification and
validation of the algorithm calculations
and clinical validation using an
independent data set.
(4) A human factors and usability
engineering assessment must be
provided.
(5) Labeling must include:
(i) A description of what the device
measures, how the device decides to
issue recommendations, and the
expected range of frequency of
recommendations, while accounting for
differences in patient condition and
environment;
(ii) Detailed information regarding
limitations of the device’s algorithm,
and key assumptions made when the
device issues a recommendation;
(iii) Warnings identifying sensor
acquisition factors that may impact
measurement results;
(iv) Warnings identifying user errors
that affect the device’s
recommendations;
(v) Detailed information regarding the
expected impact of user input errors on
the device recommendations;
(vi) Guidance for interpretation of the
device’s recommendations, including a
description that the recommendation is
adjunctive to other physical vital sign
parameters and patient information;
(vii) Description of the impact of the
compatible sensor(s) on the device’s
performance;
(viii) The expected performance of the
device for all intended patients, users,
and environments;
(ix) Relevant characteristics of the
patients studied in the clinical
validation (such as age, gender, race or
ethnicity, and patient condition) and a
summary of validation results; and
(x) Description of the software
safeguards that are in place to prevent
fluid overload, and description of any
limitation of the software safeguards.
Dated: August 28, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024–19736 Filed 9–4–24; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\05SER1.SGM
05SER1
Agencies
[Federal Register Volume 89, Number 172 (Thursday, September 5, 2024)]
[Rules and Regulations]
[Pages 72317-72319]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2024-19736]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA-2024-N-3947]
Medical Devices; Cardiovascular Devices; Classification of the
Adjunctive Open Loop Fluid Therapy Recommender
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, Agency, or we) is
classifying the adjunctive open loop fluid therapy recommender into
class II (special controls). The special controls that apply to the
device type are identified in this order and will be part of the
codified language for the adjunctive open loop fluid therapy
recommender's classification. We are taking this action because we have
determined that classifying the device into class II (special controls)
will provide a reasonable assurance of safety and effectiveness of the
device. We believe this action will also enhance patients' access to
beneficial innovative devices.
DATES: This order is effective September 5, 2024. The classification
was applicable on November 13, 2020.
FOR FURTHER INFORMATION CONTACT: Biniyam Taddese, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 2544, Silver Spring, MD 20993-0002, 240-
402-6570, [email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the adjunctive open loop fluid
therapy recommender as class II (special controls), which we have
determined will provide a reasonable assurance of safety and
effectiveness.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
[[Page 72318]]
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device
that does not require premarket approval. We determine whether a new
device is substantially equivalent to a predicate device by means of
the procedures for premarket notification under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)).
Section 207 of the Food and Drug Administration Modernization Act of
1997 (Pub. L. 105-115) established the first procedure for De Novo
classification. Section 607 of the Food and Drug Administration Safety
and Innovation Act (Pub. L. 112-144) modified the De Novo application
process by adding a second procedure. A device sponsor may utilize
either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
When FDA classifies a device into class I or II via the De Novo
process, the device can serve as a predicate for future devices of that
type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C
Act). As a result, other device sponsors do not have to submit a De
Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
510(k) process, when necessary, to market their device.
II. De Novo Classification
On June 4, 2019, FDA received Edwards Lifesciences' request for De
Novo classification of the Acumen Assisted Fluid Management (AFM)
Software Feature. FDA reviewed the request in order to classify the
device under the criteria for classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on November 13, 2020, FDA issued an order to the
requester classifying the device into class II. In this final order,
FDA is codifying the classification of the device by adding 21 CFR
870.5600.\1\ We have named the generic type of device adjunctive open
loop fluid therapy recommender, and it is identified as a prescription
device that uses software algorithms to analyze cardiovascular vital
signs and predict a patient's estimated response to fluid therapy. The
device is intended for adjunctive use with other physical vital sign
parameters and patient information and is not intended to independently
direct therapy.
---------------------------------------------------------------------------
\1\ FDA notes that the ACTION caption for this final order is
styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Adjunctive Open Loop Fluid Therapy Recommender Risks and
Mitigation Measures
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Delay in monitoring or treatment....... Software verification,
validation, and hazard
analysis; Usability
assessment; and Labeling.
Inappropriate or missed treatment due Software verification,
to over-reliance on software validation, and hazard
recommendation, which is affected by analysis; Non-clinical
algorithm or software error, or performance testing; Usability
inaccurate input from sensors or users. assessment; Clinical
performance testing; and
Labeling.
Fluid overload due to over-reliance on Software verification,
software recommendations, which are validation, and hazard
affected by algorithm or software analysis; Non-clinical
error, or inaccurate input from performance testing; Usability
sensors or users. assessment; Clinical
performance testing; and
Labeling.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to
[[Page 72319]]
premarket notification requirements under section 510(k) of the FD&C
Act.
At the time of classification, the adjunctive open loop fluid
therapy recommender is for prescription use only. Prescription devices
are exempt from the requirement for adequate directions for use for the
layperson under section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1))
and 21 CFR 801.5, as long as the conditions of 21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in part 860, subpart D, regarding De Novo classification
have been approved under OMB control number 0910-0844; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions, have been approved under OMB
control number 0910-0120; the collections of information in 21 CFR part
820, regarding quality system regulation, have been approved under OMB
control number 0910-0073; and the collections of information in 21 CFR
part 801, regarding labeling, have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
870 is amended as follows:
PART 870--CARDIOVASCULAR DEVICES
0
1. The authority citation for part 870 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 870.5600 to subpart F to read as follows:
Sec. 870.5600 Adjunctive open loop fluid therapy recommender
(a) Identification. The adjunctive open loop fluid therapy
recommender is a prescription device that uses software algorithms to
analyze cardiovascular vital signs and predict a patient's estimated
response to fluid therapy. The device is intended for adjunctive use
with other physical vital sign parameters and patient information and
is not intended to independently direct therapy.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Clinical performance testing under anticipated conditions of
use must fulfill the following:
(i) A summary of the clinical performance testing must include the
relevant patient demographics, and any statistical techniques used for
analyzing the data;
(ii) Subjects must be representative of the intended use population
for the device. Any selection criteria or sample limitations must be
fully described and justified;
(iii) Testing must demonstrate the recommendation consistency using
the expected range of data sources and data quality encountered in the
intended patients, users, and environments; and
(iv) Testing must evaluate the relationship between algorithm
recommendations, therapeutic actions, and predicted physiological event
or status.
(2) A software description and the results of verification and
validation testing based on a comprehensive hazard analysis and risk
assessment must be provided, including:
(i) A full characterization of the software technical parameters,
including algorithms;
(ii) A description of the expected recommendation, accounting for
differences in patient condition and environment;
(iii) A description of all mitigations for user error or failure of
any subsystem components (including signal detection, signal analysis,
data display, and storage) that affect the device's recommendations;
(iv) A characterization of algorithm sensitivity to variations in
user inputs;
(v) A characterization of sensor accuracy and performance;
(vi) A description of sensor data quality control measures; and
(vii) Safeguards to reduce the possibility of fluid overload.
(3) A scientific justification for the validity of the algorithm(s)
must be provided. This justification must include non-clinical
verification and validation of the algorithm calculations and clinical
validation using an independent data set.
(4) A human factors and usability engineering assessment must be
provided.
(5) Labeling must include:
(i) A description of what the device measures, how the device
decides to issue recommendations, and the expected range of frequency
of recommendations, while accounting for differences in patient
condition and environment;
(ii) Detailed information regarding limitations of the device's
algorithm, and key assumptions made when the device issues a
recommendation;
(iii) Warnings identifying sensor acquisition factors that may
impact measurement results;
(iv) Warnings identifying user errors that affect the device's
recommendations;
(v) Detailed information regarding the expected impact of user
input errors on the device recommendations;
(vi) Guidance for interpretation of the device's recommendations,
including a description that the recommendation is adjunctive to other
physical vital sign parameters and patient information;
(vii) Description of the impact of the compatible sensor(s) on the
device's performance;
(viii) The expected performance of the device for all intended
patients, users, and environments;
(ix) Relevant characteristics of the patients studied in the
clinical validation (such as age, gender, race or ethnicity, and
patient condition) and a summary of validation results; and
(x) Description of the software safeguards that are in place to
prevent fluid overload, and description of any limitation of the
software safeguards.
Dated: August 28, 2024.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2024-19736 Filed 9-4-24; 8:45 am]