Medical Devices; Ophthalmic Devices; Classification of the Retinal Diagnostic Software Device, 3203-3205 [2022-01147]
Download as PDF
3203
Federal Register / Vol. 87, No. 14 / Friday, January 21, 2022 / Rules and Regulations
APPENDIX A TO PART 11—FEE
SCHEDULE FOR FY 2022—Continued
State
County
West Virginia .....
jspears on DSK121TN23PROD with RULES1
Wisconsin ..........
VerDate Sep<11>2014
Pierce ................
San Juan ...........
Skagit ................
Skamania ..........
Snohomish ........
Spokane ............
Stevens .............
Thurston ............
Wahkiakum ........
Walla Walla .......
Whatcom ...........
Whitman ............
Yakima ..............
Barbour ..............
Berkeley ............
Boone ................
Braxton ..............
Brooke ...............
Cabell ................
Calhoun .............
Clay ...................
Doddridge ..........
Fayette ..............
Gilmer ................
Grant .................
Greenbrier .........
Hampshire .........
Hancock ............
Hardy .................
Harrison .............
Jackson .............
Jefferson ............
Kanawha ...........
Lewis .................
Lincoln ...............
Logan ................
Marion ...............
Marshall .............
Mason ................
McDowell ...........
Mercer ...............
Mineral ...............
Mingo .................
Monongalia ........
Monroe ..............
Morgan ..............
Nicholas .............
Ohio ...................
Pendleton ..........
Pleasants ...........
Pocahontas .......
Preston ..............
Putnam ..............
Raleigh ..............
Randolph ...........
Ritchie ...............
Roane ................
Summers ...........
Taylor ................
Tucker ...............
Tyler ..................
Upshur ...............
Wayne ...............
Webster .............
Wetzel ...............
Wirt ....................
Wood .................
Wyoming ...........
Adams ...............
Ashland .............
Barron ................
Bayfield ..............
Brown ................
Buffalo ...............
Burnett ...............
Calumet .............
Chippewa ..........
Clark ..................
Columbia ...........
Crawford ............
16:23 Jan 20, 2022
APPENDIX A TO PART 11—FEE
SCHEDULE FOR FY 2022—Continued
Fee/acre/yr
380.82
167.58
179.51
214.10
342.58
66.06
27.80
210.52
85.15
44.85
297.60
30.94
48.82
63.52
145.52
63.63
55.88
76.86
96.96
49.60
46.85
57.92
79.25
35.83
71.33
70.90
81.73
124.44
87.41
68.12
60.15
159.78
105.58
58.76
50.14
67.31
80.64
70.38
66.11
168.56
68.42
75.85
30.36
123.24
72.42
142.39
71.14
98.59
61.21
62.79
51.01
74.73
77.97
100.90
65.97
49.11
52.51
61.81
83.66
77.89
52.04
71.96
54.66
62.55
52.42
49.19
90.68
91.06
120.55
59.99
91.82
58.84
228.13
105.81
73.13
211.46
95.58
108.98
156.51
85.40
Jkt 256001
State
County
Wyoming ............
PO 00000
Frm 00029
Dane ..................
Dodge ................
Door ...................
Douglas .............
Dunn ..................
Eau Claire .........
Florence ............
Fond du Lac ......
Forest ................
Grant .................
Green ................
Green Lake .......
Iowa ...................
Iron ....................
Jackson .............
Jefferson ............
Juneau ..............
Kenosha ............
Kewaunee .........
La Crosse ..........
Lafayette ............
Langlade ............
Lincoln ...............
Manitowoc .........
Marathon ...........
Marinette ...........
Marquette ..........
Menominee ........
Milwaukee .........
Monroe ..............
Oconto ...............
Oneida ...............
Outagamie .........
Ozaukee ............
Pepin .................
Pierce ................
Polk ...................
Portage ..............
Price ..................
Racine ...............
Richland ............
Rock ..................
Rusk ..................
Sauk ..................
Sawyer ..............
Shawano ...........
Sheboygan ........
St. Croix ............
Taylor ................
Trempealeau .....
Vernon ...............
Vilas ...................
Walworth ...........
Washburn ..........
Washington .......
Waukesha .........
Waupaca ...........
Waushara ..........
Winnebago ........
Wood .................
Albany ...............
Big Horn ............
Campbell ...........
Carbon ...............
Converse ...........
Crook .................
Fremont .............
Goshen ..............
Hot Springs .......
Johnson .............
Laramie .............
Lincoln ...............
Natrona ..............
Niobrara .............
Park ...................
Platte .................
Sheridan ............
Sublette .............
Sweetwater ........
Teton .................
Fmt 4700
Sfmt 4700
APPENDIX A TO PART 11—FEE
SCHEDULE FOR FY 2022—Continued
State
Fee/acre/yr
221.26
156.94
127.75
52.66
96.70
122.65
67.86
195.37
65.15
126.67
145.74
153.61
130.62
91.33
102.05
165.27
99.47
203.43
150.92
133.93
160.51
87.87
87.04
183.26
127.59
104.12
112.15
46.62
239.76
106.53
111.88
109.17
193.54
176.01
104.04
124.07
94.99
110.11
66.04
206.30
90.13
176.95
66.73
112.98
69.63
125.20
177.08
125.90
78.82
106.29
104.30
158.79
186.19
84.00
189.41
147.89
121.28
113.62
187.21
88.92
10.74
23.35
8.32
8.08
7.77
14.38
18.72
12.67
9.12
8.64
12.46
26.86
6.67
9.21
21.95
12.90
17.98
24.26
4.35
59.49
County
Fee/acre/yr
Uinta ..................
Washakie ...........
Weston ..............
15.75
17.18
9.83
[FR Doc. 2022–01105 Filed 1–20–22; 8:45 am]
BILLING CODE 6717–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA–2021–N–0993]
Medical Devices; Ophthalmic Devices;
Classification of the Retinal Diagnostic
Software Device
Food and Drug Administration,
Department of Health and Human
Services (HHS).
ACTION: Final amendment; final order.
AGENCY:
The Food and Drug
Administration (FDA, Agency, or we) is
classifying the retinal diagnostic
software device into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the retinal
diagnostic software device’s
classification. We are taking this action
because we have determined that
classifying the device into class II
(special controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices.
DATES: This order is effective January
21, 2022. The classification was
applicable on April 11, 2018.
FOR FURTHER INFORMATION CONTACT:
Elvin Ng, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1304, Silver Spring,
MD 20993–0002, 240–402–4662,
Elvin.Ng@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
Upon request, FDA has classified the
retinal diagnostic software device as
class II (special controls), which we
have determined will provide a
reasonable assurance of safety and
effectiveness. In addition, we believe
this action will enhance patients’ access
to beneficial innovation, by placing the
device into a lower device class than the
automatic class III assignment.
E:\FR\FM\21JAR1.SGM
21JAR1
3204
Federal Register / Vol. 87, No. 14 / Friday, January 21, 2022 / Rules and Regulations
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (see 21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
device by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act. Although the device was
automatically placed within class III,
the De Novo classification is considered
to be the initial classification of the
device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see
section 513(f)(2)(B)(i)) of the FD&C Act).
As a result, other device sponsors do not
have to submit a De Novo request or
premarket approval application to
market a substantially equivalent device
(see section 513(i) of the FD&C Act,
defining ‘‘substantial equivalence’’).
Instead, sponsors can use the lessburdensome 510(k) process, when
necessary, to market their device.
II. De Novo Classification
On January 12, 2018, FDA received
IDx, LLC’s request for De Novo
classification of the IDx-DR. FDA
reviewed the request in order to classify
the device under the criteria for
classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on April 11, 2018, FDA
issued an order to the requester
classifying the device into class II. In
this final order, FDA is codifying the
classification of the device by adding 21
CFR 886.1100.1 We have named the
generic type of device retinal diagnostic
software device, and it is identified as
a prescription software device that
incorporates an adaptive algorithm to
evaluate ophthalmic images for
diagnostic screening to identify retinal
diseases or conditions.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
jspears on DSK121TN23PROD with RULES1
TABLE 1—RETINAL DIAGNOSTIC SOFTWARE DEVICE RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
False positive results leading to additional unnecessary medical procedures.
• Diagnostic software failure ............................................
• Software failure ..............................................................
False negative results leading to delay of further evaluation
or treatment.
• Diagnostic software failure ............................................
• Software failure ..............................................................
Operator failure to provide images that meet input quality
specifications.
Clinical performance testing; Software verification, validation, and hazard analysis; and Protocol for technical specification changes.
1 FDA notes that the ACTION caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
VerDate Sep<11>2014
16:23 Jan 20, 2022
Jkt 256001
Clinical performance testing; Software verification, validation, and hazard analysis; Protocol for technical specification changes; and Labeling.
Labeling, Training, and Human factors validation testing.
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
PO 00000
Frm 00030
Fmt 4700
Sfmt 4700
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
E:\FR\FM\21JAR1.SGM
21JAR1
Federal Register / Vol. 87, No. 14 / Friday, January 21, 2022 / Rules and Regulations
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k) of the FD&C Act.
At the time of classification, retinal
diagnostic software devices are for
prescription use only. Prescription
devices are exempt from the
requirement for adequate directions for
use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
jspears on DSK121TN23PROD with RULES1
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
VerDate Sep<11>2014
16:23 Jan 20, 2022
Jkt 256001
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods
and services.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 886 is
amended as follows:
PART 886—OPHTHALMIC DEVICES
1. The authority citation for part 886
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 886.1100 to subpart B to read
as follows:
■
§ 886.1100
device.
Retinal diagnostic software
(a) Identification. A retinal diagnostic
software device is a prescription
software device that incorporates an
adaptive algorithm to evaluate
ophthalmic images for diagnostic
screening to identify retinal diseases or
conditions.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Software verification and
validation documentation, based on a
comprehensive hazard analysis, must
fulfill the following:
(i) Software documentation must
provide a full characterization of
technical parameters of the software,
including algorithm(s).
(ii) Software documentation must
describe the expected impact of
applicable image acquisition hardware
characteristics on performance and
associated minimum specifications.
(iii) Software documentation must
include a cybersecurity vulnerability
and management process to assure
software functionality.
(iv) Software documentation must
include mitigation measures to manage
failure of any subsystem components
with respect to incorrect patient reports
and operator failures.
(2) Clinical performance data
supporting the indications for use must
be provided, including the following:
(i) Clinical performance testing must
evaluate sensitivity, specificity, positive
predictive value, and negative
predictive value for each endpoint
reported for the indicated disease or
condition across the range of available
device outcomes.
(ii) Clinical performance testing must
evaluate performance under anticipated
conditions of use.
(iii) Statistical methods must include
the following:
(A) Where multiple samples from the
same patient are used, statistical
PO 00000
Frm 00031
Fmt 4700
Sfmt 9990
3205
analysis must not assume statistical
independence without adequate
justification.
(B) Statistical analysis must provide
confidence intervals for each
performance metric.
(iv) Clinical data must evaluate the
variability in output performance due to
both the user and the image acquisition
device used.
(3) A training program with
instructions on how to acquire and
process quality images must be
provided.
(4) Human factors validation testing
that evaluates the effect of the training
program on user performance must be
provided.
(5) A protocol must be developed that
describes the level of change in device
technical specifications that could
significantly affect the safety or
effectiveness of the device.
(6) Labeling must include:
(i) Instructions for use, including a
description of how to obtain quality
images and how device performance is
affected by user interaction and user
training;
(ii) The type of imaging data used,
what the device outputs to the user, and
whether the output is qualitative or
quantitative;
(iii) Warnings regarding image
acquisition factors that affect image
quality;
(iv) Warnings regarding interpretation
of the provided outcomes, including:
(A) A warning that the device is not
to be used to screen for the presence of
diseases or conditions beyond its
indicated uses;
(B) A warning that the device
provides a screening diagnosis only and
that it is critical that the patient be
advised to receive followup care; and
(C) A warning that the device does not
treat the screened disease;
(v) A summary of the clinical
performance of the device for each
output, with confidence intervals; and
(vi) A summary of the clinical
performance testing conducted with the
device, including a description of the
patient population and clinical
environment under which it was
evaluated.
Dated: January 14, 2022.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2022–01147 Filed 1–20–22; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\21JAR1.SGM
21JAR1
Agencies
[Federal Register Volume 87, Number 14 (Friday, January 21, 2022)]
[Rules and Regulations]
[Pages 3203-3205]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2022-01147]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA-2021-N-0993]
Medical Devices; Ophthalmic Devices; Classification of the
Retinal Diagnostic Software Device
AGENCY: Food and Drug Administration, Department of Health and Human
Services (HHS).
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, Agency, or we) is
classifying the retinal diagnostic software device into class II
(special controls). The special controls that apply to the device type
are identified in this order and will be part of the codified language
for the retinal diagnostic software device's classification. We are
taking this action because we have determined that classifying the
device into class II (special controls) will provide a reasonable
assurance of safety and effectiveness of the device. We believe this
action will also enhance patients' access to beneficial innovative
devices.
DATES: This order is effective January 21, 2022. The classification was
applicable on April 11, 2018.
FOR FURTHER INFORMATION CONTACT: Elvin Ng, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1304, Silver Spring, MD 20993-0002, 240-402-4662,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the retinal diagnostic software
device as class II (special controls), which we have determined will
provide a reasonable assurance of safety and effectiveness. In
addition, we believe this action will enhance patients' access to
beneficial innovation, by placing the device into a lower device class
than the automatic class III assignment.
[[Page 3204]]
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device
that does not require premarket approval. We determine whether a new
device is substantially equivalent to a predicate device by means of
the procedures for premarket notification under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation. When FDA classifies a device into
class I or II via the De Novo process, the device can serve as a
predicate for future devices of that type, including for 510(k)s (see
section 513(f)(2)(B)(i)) of the FD&C Act). As a result, other device
sponsors do not have to submit a De Novo request or premarket approval
application to market a substantially equivalent device (see section
513(i) of the FD&C Act, defining ``substantial equivalence''). Instead,
sponsors can use the less-burdensome 510(k) process, when necessary, to
market their device.
II. De Novo Classification
On January 12, 2018, FDA received IDx, LLC's request for De Novo
classification of the IDx-DR. FDA reviewed the request in order to
classify the device under the criteria for classification set forth in
section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on April 11, 2018, FDA issued an order to the requester
classifying the device into class II. In this final order, FDA is
codifying the classification of the device by adding 21 CFR
886.1100.\1\ We have named the generic type of device retinal
diagnostic software device, and it is identified as a prescription
software device that incorporates an adaptive algorithm to evaluate
ophthalmic images for diagnostic screening to identify retinal diseases
or conditions.
---------------------------------------------------------------------------
\1\ FDA notes that the ACTION caption for this final order is
styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Retinal Diagnostic Software Device Risks and Mitigation
Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
False positive results leading Clinical performance testing; Software
to additional unnecessary verification, validation, and hazard
medical procedures. analysis; and Protocol for technical
Diagnostic software specification changes.
failure.
Software failure.....
False negative results leading Clinical performance testing; Software
to delay of further verification, validation, and hazard
evaluation or treatment. analysis; Protocol for technical
Diagnostic software specification changes; and Labeling.
failure.
Software failure.....
Operator failure to provide Labeling, Training, and Human factors
images that meet input validation testing.
quality specifications.
------------------------------------------------------------------------
[[Page 3205]]
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
At the time of classification, retinal diagnostic software devices
are for prescription use only. Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR
801.5, as long as the conditions of 21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in the guidance document ``De Novo Classification
Process (Evaluation of Automatic Class III Designation)'' have been
approved under OMB control number 0910-0844; the collections of
information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions, have been approved under OMB
control number 0910-0120; the collections of information in 21 CFR part
820, regarding quality system regulation, have been approved under OMB
control number 0910-0073; and the collections of information in 21 CFR
part 801, regarding labeling, have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods and services.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
886 is amended as follows:
PART 886--OPHTHALMIC DEVICES
0
1. The authority citation for part 886 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 886.1100 to subpart B to read as follows:
Sec. 886.1100 Retinal diagnostic software device.
(a) Identification. A retinal diagnostic software device is a
prescription software device that incorporates an adaptive algorithm to
evaluate ophthalmic images for diagnostic screening to identify retinal
diseases or conditions.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Software verification and validation documentation, based on a
comprehensive hazard analysis, must fulfill the following:
(i) Software documentation must provide a full characterization of
technical parameters of the software, including algorithm(s).
(ii) Software documentation must describe the expected impact of
applicable image acquisition hardware characteristics on performance
and associated minimum specifications.
(iii) Software documentation must include a cybersecurity
vulnerability and management process to assure software functionality.
(iv) Software documentation must include mitigation measures to
manage failure of any subsystem components with respect to incorrect
patient reports and operator failures.
(2) Clinical performance data supporting the indications for use
must be provided, including the following:
(i) Clinical performance testing must evaluate sensitivity,
specificity, positive predictive value, and negative predictive value
for each endpoint reported for the indicated disease or condition
across the range of available device outcomes.
(ii) Clinical performance testing must evaluate performance under
anticipated conditions of use.
(iii) Statistical methods must include the following:
(A) Where multiple samples from the same patient are used,
statistical analysis must not assume statistical independence without
adequate justification.
(B) Statistical analysis must provide confidence intervals for each
performance metric.
(iv) Clinical data must evaluate the variability in output
performance due to both the user and the image acquisition device used.
(3) A training program with instructions on how to acquire and
process quality images must be provided.
(4) Human factors validation testing that evaluates the effect of
the training program on user performance must be provided.
(5) A protocol must be developed that describes the level of change
in device technical specifications that could significantly affect the
safety or effectiveness of the device.
(6) Labeling must include:
(i) Instructions for use, including a description of how to obtain
quality images and how device performance is affected by user
interaction and user training;
(ii) The type of imaging data used, what the device outputs to the
user, and whether the output is qualitative or quantitative;
(iii) Warnings regarding image acquisition factors that affect
image quality;
(iv) Warnings regarding interpretation of the provided outcomes,
including:
(A) A warning that the device is not to be used to screen for the
presence of diseases or conditions beyond its indicated uses;
(B) A warning that the device provides a screening diagnosis only
and that it is critical that the patient be advised to receive followup
care; and
(C) A warning that the device does not treat the screened disease;
(v) A summary of the clinical performance of the device for each
output, with confidence intervals; and
(vi) A summary of the clinical performance testing conducted with
the device, including a description of the patient population and
clinical environment under which it was evaluated.
Dated: January 14, 2022.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2022-01147 Filed 1-20-22; 8:45 am]
BILLING CODE 4164-01-P