Medical Devices; Neurological Devices; Classification of the Cerebrospinal Fluid Shunt System, 73973-73975 [2021-28157]
Download as PDF
Federal Register / Vol. 86, No. 247 / Wednesday, December 29, 2021 / Rules and Regulations
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. In order for
a device to fall within this classification,
and thus avoid automatic classification
in class III, it would have to comply
with the special controls named in this
final order. The necessary special
controls appear in the regulation
codified by this order. This device is
subject to premarket notification
requirements under section 510(k).
At the time of classification,
diagnostic neurosurgical microscope
filters are for prescription use only.
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
khammond on DSKJM1Z7X2PROD with RULES
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
VerDate Sep<11>2014
16:59 Dec 28, 2021
Jkt 256001
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 882 is
amended as follows:
PART 882—NEUROLOGICAL DEVICES
1. The authority citation for part 882
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 882.4950 to subpart E to read
as follows:
■
§ 882.4950 Diagnostic neurosurgical
microscope filter.
(a) Identification. A diagnostic
neurosurgical microscope filter is a
device intended for use during
neurosurgery to visualize fluorescence
and enhance visualization of tissue
associated with a specific disease or
condition.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use, and verify and
validate filter specifications and
functional characteristics, including the
following:
(i) Spectrum and intensity of the
illumination source;
(ii) Spectrum of the excitation and
emission filter modules when integrated
in the surgical operating microscope;
(iii) Excitation power and power
density;
(iv) Optical path loss from
illumination source to objective lens or
microscope camera;
(v) Homogeneity of the excitation
light at the focal plane;
(vi) Fluorescence detection
sensitivity;
(vii) Verification of calibration or
preoperative procedures; and
(viii) If camera-based, spectral
sensitivity of the camera.
(2) Labeling must include:
(i) Identification of the filter
characteristics in conjunction with a
compatible surgical operating
microscope, to include the following:
(A) Illumination spectrum and power
density; and
(B) Excitation and emission filter
spectra.
(ii) Instructions for calibration or
preoperative checks to ensure device
functionality prior to each use;
(iii) Instructions for use with
compatible surgical operating
PO 00000
Frm 00027
Fmt 4700
Sfmt 4700
73973
microscopes, external light sources, and
cameras;
(iv) A warning that the device should
only be used with fluorophores
approved for use within the specified
spectral ranges; and
(v) A warning that the device is not
a standalone diagnostic.
Dated: December 17, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–28160 Filed 12–28–21; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 882
[Docket No. FDA–2021–N–0855]
Medical Devices; Neurological
Devices; Classification of the
Cerebrospinal Fluid Shunt System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final amendment; final order.
The Food and Drug
Administration (FDA or we) is
classifying the cerebrospinal fluid shunt
system into class II (special controls).
The special controls that apply to the
device type are identified in this order
and will be part of the codified language
for the cerebrospinal fluid shunt
system’s classification. We are taking
this action because we have determined
that classifying the device into class II
(special controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices.
DATES: This order is effective December
29, 2021. The classification was
applicable on August 22, 2014.
FOR FURTHER INFORMATION CONTACT:
Xiaolin Zheng, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2674, Silver Spring,
MD 20993–0002, 301–796–2823,
Xiaolin.Zheng@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
Upon request, FDA has classified the
cerebrospinal fluid shunt system as
class II (special controls), which we
have determined will provide a
reasonable assurance of safety and
effectiveness. In addition, we believe
this action will enhance patients’ access
to beneficial innovation.
E:\FR\FM\29DER1.SGM
29DER1
73974
Federal Register / Vol. 86, No. 247 / Wednesday, December 29, 2021 / Rules and Regulations
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C. 360(k)
and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
placed within class III, the De Novo
classification is considered to be the
initial classification of the device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see
section 513(f)(2)(B)(i) of the FD&C Act).
As a result, other device sponsors do not
have to submit a De Novo request or
premarket approval application to
market a substantially equivalent device
(see section 513(i) of the FD&C Act,
defining ‘‘substantial equivalence’’).
Instead, sponsors can use the lessburdensome 510(k) process, when
necessary, to market their device.
II. De Novo Classification
For this device, FDA issued an order
on November 27, 2012, finding the
Medtronic DUETTM External Drainage
and Monitoring System (EDMS) not
substantially equivalent to a predicate
device and not subject to a premarket
approval application (PMA). Thus, the
device remained in class III in
accordance with section 513(f)(1) of the
FD&C Act when we issued the order.
On December 21, 2012, FDA received
Medtronic Neurosurgery’s request for De
Novo classification of the Medtronic
DUETTM EDMS. FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1) of the FD&C
Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see section 513
(a)(1)(B) of the FD&C Act). After review
of the information submitted in the
request, we determined that the device
can be classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on August 22, 2014, FDA
issued an order to the requester
classifying the device into class II. In
this final order, FDA is codifying the
classification of the device by adding 21
CFR 882.5560.1 We have named the
generic type of device cerebrospinal
fluid shunt system, and it is identified
as a prescription device used to monitor
and divert fluid from the brain or other
part of the central nervous system to an
internal delivery site or an external
receptacle for the purpose of preventing
spinal cord ischemia or injury during
procedures that require reduction in
central nervous system pressure. A
cerebrospinal fluid shunt system may
include catheters, valved catheters,
valves, connectors, and pressure
monitors intended to facilitate use of the
shunt or evaluation of a patient with a
shunt.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
khammond on DSKJM1Z7X2PROD with RULES
TABLE 1—CEREBROSPINAL FLUID SHUNT SYSTEM RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
Pyrogenicity/adverse tissue reaction ........................................................
Biocompatibility testing, Pyrogenicity testing, Labeling, Shelf-life testing,
and Sterility testing.
Labeling, Sterility testing, and Package integrity testing.
Labeling, and Non-clinical performance testing.
Labeling, and Non-clinical performance testing.
Infection (including meningitis) .................................................................
Cerebrospinal fluid (CSF) leakage ...........................................................
Over- and under-drainage
1 FDA notes that the ACTION caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
VerDate Sep<11>2014
16:59 Dec 28, 2021
Jkt 256001
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
PO 00000
Frm 00028
Fmt 4700
Sfmt 4700
E:\FR\FM\29DER1.SGM
29DER1
Federal Register / Vol. 86, No. 247 / Wednesday, December 29, 2021 / Rules and Regulations
73975
TABLE 1—CEREBROSPINAL FLUID SHUNT SYSTEM RISKS AND MITIGATION MEASURES—Continued
Identified risks
Mitigation measures
• Spinal headache with and without CSF leakage.
• Intracranial hemorrhage.
• Hematoma (e.g., spinal, subdural).
• Paraplegia.
• Foreign body obstruction.
Procedural/use errors ...............................................................................
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k) of the FD&C Act.
At the time of classification,
cerebrospinal fluid shunt systems are
for prescription use only. Prescription
devices are exempt from the
requirement for adequate directions for
use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
khammond on DSKJM1Z7X2PROD with RULES
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 820,
regarding quality system regulation,
have been approved under OMB control
number 0910–0073; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
VerDate Sep<11>2014
16:59 Dec 28, 2021
Jkt 256001
Labeling.
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 882 is
amended as follows:
PART 882—NEUROLOGICAL DEVICES
1. The authority citation for part 882
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 882.5560 to subpart F to read
as follows:
■
§ 882.5560
system.
Cerebrospinal fluid shunt
(a) Identification. A cerebrospinal
fluid shunt system is a prescription
device used to monitor and divert fluid
from the brain or other part of the
central nervous system to an internal
delivery site or an external receptacle
for the purpose of preventing spinal
cord ischemia or injury during
procedures that require reduction in
central nervous system pressure. A
cerebrospinal fluid shunt system may
include catheters, valved catheters,
valves, connectors, and pressure
monitors intended to facilitate use of the
shunt or evaluation of a patient with a
shunt.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The device description must
include a detailed summary of the
device technical parameters, including
design configuration, dimensions,
engineering drawings, and a list of all
components with identification of their
materials of construction.
PO 00000
Frm 00029
Fmt 4700
Sfmt 9990
(2) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(3) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
(i) Simulated use testing must be
conducted to characterize fluid flow and
resistance to leakage; and
(ii) Mechanical integrity testing of all
connections must be conducted.
(4) Performance data must support the
shelf life of the device by demonstrating
continued sterility, package integrity,
and device functionality over the
specified shelf life.
(5) Performance data must
demonstrate the sterility and
pyrogenicity of patient-contacting
components of the device.
(6) The labeling must include:
(i) Contraindications with respect to
patients who should not receive a
lumbar drain;
(ii) A warning that the device should
have 24-hour-a-day availability of
trained personnel to supervise
monitoring and drainage;
(iii) Instructions on proper device
setup, positioning, and monitoring;
(iv) Warnings and precautions to
inform the user of serious hazards and
special care associated with the use of
the device;
(v) A statement that the device is not
to be reused, reprocessed, or resterilized
when open but unused; and
(vi) Cleaning instructions for the
injection sites.
Dated: December 17, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–28157 Filed 12–28–21; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\29DER1.SGM
29DER1
Agencies
[Federal Register Volume 86, Number 247 (Wednesday, December 29, 2021)]
[Rules and Regulations]
[Pages 73973-73975]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2021-28157]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 882
[Docket No. FDA-2021-N-0855]
Medical Devices; Neurological Devices; Classification of the
Cerebrospinal Fluid Shunt System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the cerebrospinal fluid shunt system into class II (special controls).
The special controls that apply to the device type are identified in
this order and will be part of the codified language for the
cerebrospinal fluid shunt system's classification. We are taking this
action because we have determined that classifying the device into
class II (special controls) will provide a reasonable assurance of
safety and effectiveness of the device. We believe this action will
also enhance patients' access to beneficial innovative devices.
DATES: This order is effective December 29, 2021. The classification
was applicable on August 22, 2014.
FOR FURTHER INFORMATION CONTACT: Xiaolin Zheng, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2674, Silver Spring, MD 20993-0002, 301-796-2823,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the cerebrospinal fluid shunt
system as class II (special controls), which we have determined will
provide a reasonable assurance of safety and effectiveness. In
addition, we believe this action will enhance patients' access to
beneficial innovation.
[[Page 73974]]
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate by means of the procedures
for premarket notification under section 510(k) of the FD&C Act (21
U.S.C. 360(k) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation. When FDA classifies a device into
class I or II via the De Novo process, the device can serve as a
predicate for future devices of that type, including for 510(k)s (see
section 513(f)(2)(B)(i) of the FD&C Act). As a result, other device
sponsors do not have to submit a De Novo request or premarket approval
application to market a substantially equivalent device (see section
513(i) of the FD&C Act, defining ``substantial equivalence''). Instead,
sponsors can use the less-burdensome 510(k) process, when necessary, to
market their device.
II. De Novo Classification
For this device, FDA issued an order on November 27, 2012, finding
the Medtronic DUETTM External Drainage and Monitoring System
(EDMS) not substantially equivalent to a predicate device and not
subject to a premarket approval application (PMA). Thus, the device
remained in class III in accordance with section 513(f)(1) of the FD&C
Act when we issued the order.
On December 21, 2012, FDA received Medtronic Neurosurgery's request
for De Novo classification of the Medtronic DUETTM EDMS. FDA
reviewed the request in order to classify the device under the criteria
for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see section 513 (a)(1)(B) of the FD&C Act). After
review of the information submitted in the request, we determined that
the device can be classified into class II with the establishment of
special controls. FDA has determined that these special controls, in
addition to the general controls, will provide reasonable assurance of
the safety and effectiveness of the device.
Therefore, on August 22, 2014, FDA issued an order to the requester
classifying the device into class II. In this final order, FDA is
codifying the classification of the device by adding 21 CFR
882.5560.\1\ We have named the generic type of device cerebrospinal
fluid shunt system, and it is identified as a prescription device used
to monitor and divert fluid from the brain or other part of the central
nervous system to an internal delivery site or an external receptacle
for the purpose of preventing spinal cord ischemia or injury during
procedures that require reduction in central nervous system pressure. A
cerebrospinal fluid shunt system may include catheters, valved
catheters, valves, connectors, and pressure monitors intended to
facilitate use of the shunt or evaluation of a patient with a shunt.
---------------------------------------------------------------------------
\1\ FDA notes that the ACTION caption for this final order is
styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Cerebrospinal Fluid Shunt System Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Pyrogenicity/adverse tissue reaction... Biocompatibility testing,
Pyrogenicity testing,
Labeling, Shelf-life testing,
and Sterility testing.
Infection (including meningitis)....... Labeling, Sterility testing,
and Package integrity testing.
Cerebrospinal fluid (CSF) leakage...... Labeling, and Non-clinical
performance testing.
Over- and under-drainage Labeling, and Non-clinical
performance testing.
[[Page 73975]]
Spinal headache with and
without CSF leakage.
Intracranial hemorrhage...
Hematoma (e.g., spinal,
subdural).
Paraplegia................
Foreign body obstruction..
Procedural/use errors.................. Labeling.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
At the time of classification, cerebrospinal fluid shunt systems
are for prescription use only. Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR
801.5, as long as the conditions of 21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in the guidance document ``De Novo Classification
Process (Evaluation of Automatic Class III Designation)'' have been
approved under OMB control number 0910-0844; the collections of
information in 21 CFR part 820, regarding quality system regulation,
have been approved under OMB control number 0910-0073; the collections
of information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions, have been approved under OMB
control number 0910-0120; and the collections of information in 21 CFR
part 801, regarding labeling, have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act, and
under authority delegated to the Commissioner of Food and Drugs, 21 CFR
part 882 is amended as follows:
PART 882--NEUROLOGICAL DEVICES
0
1. The authority citation for part 882 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 882.5560 to subpart F to read as follows:
Sec. 882.5560 Cerebrospinal fluid shunt system.
(a) Identification. A cerebrospinal fluid shunt system is a
prescription device used to monitor and divert fluid from the brain or
other part of the central nervous system to an internal delivery site
or an external receptacle for the purpose of preventing spinal cord
ischemia or injury during procedures that require reduction in central
nervous system pressure. A cerebrospinal fluid shunt system may include
catheters, valved catheters, valves, connectors, and pressure monitors
intended to facilitate use of the shunt or evaluation of a patient with
a shunt.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The device description must include a detailed summary of the
device technical parameters, including design configuration,
dimensions, engineering drawings, and a list of all components with
identification of their materials of construction.
(2) The patient-contacting components of the device must be
demonstrated to be biocompatible.
(3) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance characteristics must be tested:
(i) Simulated use testing must be conducted to characterize fluid
flow and resistance to leakage; and
(ii) Mechanical integrity testing of all connections must be
conducted.
(4) Performance data must support the shelf life of the device by
demonstrating continued sterility, package integrity, and device
functionality over the specified shelf life.
(5) Performance data must demonstrate the sterility and
pyrogenicity of patient-contacting components of the device.
(6) The labeling must include:
(i) Contraindications with respect to patients who should not
receive a lumbar drain;
(ii) A warning that the device should have 24-hour-a-day
availability of trained personnel to supervise monitoring and drainage;
(iii) Instructions on proper device setup, positioning, and
monitoring;
(iv) Warnings and precautions to inform the user of serious hazards
and special care associated with the use of the device;
(v) A statement that the device is not to be reused, reprocessed,
or resterilized when open but unused; and
(vi) Cleaning instructions for the injection sites.
Dated: December 17, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021-28157 Filed 12-28-21; 8:45 am]
BILLING CODE 4164-01-P