Medical Devices; Gastroenterology-Urology Devices; Classification of the Pressure Ulcer Management Tool, 70733-70735 [2021-26924]
Download as PDF
Federal Register / Vol. 86, No. 236 / Monday, December 13, 2021 / Rules and Regulations
This device is subject to premarket
notification requirements under section
510(k) of the FD&C Act.
At the time of classification,
temporary coil embolization assist
devices are for prescription use only.
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
khammond on DSKJM1Z7X2PROD with RULES
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 882 is
amended as follows:
VerDate Sep<11>2014
15:56 Dec 10, 2021
Jkt 256001
PART 882—NEUROLOGICAL DEVICES
1. The authority citation for part 882
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 882.5955 to subpart F to read
as follows:
■
§ 882.5955 Temporary coil embolization
assist device.
(a) Identification. A temporary coil
embolization assist device is a
prescription device intended for
temporary use in the neurovasculature
to mechanically assist in the
embolization of intracranial aneurysms
with embolic coils. The device is
delivered into the neurovasculature
with an endovascular approach. This
device is not intended to be
permanently implanted and is removed
from the body when the procedure is
completed.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Clinical performance testing of the
device must demonstrate the device
performs as intended for temporary use
as an endovascular device to assist in
the coil embolization of intracranial
aneurysms and must evaluate all
adverse events, including tissue or
vessel damage that could lead to
dissection, perforation, hemorrhage, or
vasospasm, thrombo-embolic events,
and coil entanglement.
(2) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(3) Non-clinical performance testing
must demonstrate the device performs
as intended under anticipated
conditions of use, including:
(i) Mechanical testing to demonstrate
the device can withstand anticipated
tensile, torsional, compressive, and tip
deflection forces;
(ii) Mechanical testing to evaluate the
radial forces exerted by the device;
(iii) Simulated use testing to
demonstrate the device can be delivered
to the target location in the
neurovasculature and is compatible
with embolic coils;
(iv) Dimensional verification testing;
(v) Radiopacity testing; and
(vi) Performance testing to evaluate
the coating integrity and particulates
under simulated use conditions.
(4) Animal testing under anticipated
use conditions must evaluate all adverse
events, including damage to vessels or
tissues.
(5) Performance data must support the
sterility and pyrogenicity of the device.
(6) Performance data must support the
shelf life of the device by demonstrating
PO 00000
Frm 00045
Fmt 4700
Sfmt 4700
70733
continued sterility, package integrity,
and device functionality over the
labeled shelf life.
(7) The labeling must include:
(i) Instructions for use;
(ii) A detailed summary of the device
technical parameters, including
compatible delivery catheter
dimensions and device sizing
information;
(iii) A summary of the clinical testing
results, including a detailed summary of
the device- and procedure-related
complications and adverse events; and
(iv) A shelf life.
Dated: December 6, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–26926 Filed 12–10–21; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 876
[Docket No. FDA–2021–N–0898]
Medical Devices; GastroenterologyUrology Devices; Classification of the
Pressure Ulcer Management Tool
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final amendment; final order.
The Food and Drug
Administration (FDA or we) is
classifying the pressure ulcer
management tool into class I. We are
taking this action because we have
determined that classifying the device
into class I will provide a reasonable
assurance of safety and effectiveness of
the device. We believe this action will
also enhance patients’ access to
beneficial innovative devices.
DATES: This order is effective December
13, 2021. The classification was
applicable on December 20, 2018.
FOR FURTHER INFORMATION CONTACT:
Gema Gonzalez, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2530, Silver Spring,
MD 20993–0002, 301–796–6519,
Gema.Gonzalez@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
Upon request, FDA has classified the
pressure ulcer management tool as class
I, which we have determined will
provide a reasonable assurance of safety
and effectiveness. In addition, we
believe this action will enhance
E:\FR\FM\13DER1.SGM
13DER1
khammond on DSKJM1Z7X2PROD with RULES
70734
Federal Register / Vol. 86, No. 236 / Monday, December 13, 2021 / Rules and Regulations
patients’ access to beneficial innovation
by placing the device into a lower
device class than the automatic class III
assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (see 21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
device by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
VerDate Sep<11>2014
15:56 Dec 10, 2021
Jkt 256001
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
placed within class III, the De Novo
classification is considered to be the
initial classification of the device.
When FDA classifies a device into
class I or II via the De Novo process, the
device can serve as a predicate for
future devices of that type, including for
510(k)s (see section 513(f)(2)(B)(i) of the
FD&C Act). As a result, other device
sponsors do not have to submit a De
Novo request or premarket approval
application to market a substantially
equivalent device (see section 513(i) of
the FD&C Act, defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the less-burdensome 510(k) process,
when necessary, to market their device.
II. De Novo Classification
On April 3, 2017, FDA received Bruin
Biometrics, LLC’s request for De Novo
classification of the SEM Scanner
(Model 200). FDA reviewed the request
in order to classify the device under the
criteria for classification set forth in
section 513(a)(1) of the FD&C Act.
We classify devices into class I if
general controls are sufficient to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use (see 21 U.S.C.
360c(a)(1)(A)). After review of the
information submitted in the request,
we determined that the device can be
classified into class I. FDA has
determined that general controls will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on December 20, 2018,
FDA issued an order to the requester
classifying the device into class I. In this
final order, FDA is codifying the
classification of the device by adding 21
CFR 876.2100.1 We have named the
generic type of device pressure ulcer
management tool, and it is identified as
a prescription device intended for
patients at risk of developing pressure
ulcers. The device provides output that
supports a user’s decision to increase
intervention. The device is an adjunct
tool for pressure ulcer management that
1 FDA notes that the ‘‘ACTION’’ caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
PO 00000
Frm 00046
Fmt 4700
Sfmt 4700
is not intended for detection or
diagnostic purposes.
FDA has identified the following risks
to health associated specifically with
this type of device: Adverse tissue
reaction, transmission of infection
between patients, electromagnetic
interference with patient monitoring
equipment, and electrical shock. As
previously stated, FDA believes general
controls provide reasonable assurance of
safety and effectiveness for this device
type.
At the time of classification, pressure
ulcer management tools are for
prescription use only. Prescription
devices are exempt from the
requirement for adequate directions for
use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
Section 510(l)(1) of the FD&C Act
provides that a device within a type that
has been classified into class I under
section 513 of the FD&C Act is exempt
from premarket notification under
section 510(k), unless the device is of
substantial importance in preventing
impairment of human health or presents
a potentially unreasonable risk of illness
or injury (21 U.S.C. 360(l)(1)). Devices
within this type are exempt from the
premarket notification requirements
under section 510(k), subject to the
limitations of exemptions in 21 CFR
876.9.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order refers to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; and the collections of
E:\FR\FM\13DER1.SGM
13DER1
Federal Register / Vol. 86, No. 236 / Monday, December 13, 2021 / Rules and Regulations
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120.
List of Subjects in 21 CFR Part 876
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 876 is
amended as follows:
PART 876—GASTROENTEROLOGYUROLOGY DEVICES
1. The authority citation for part 876
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 876.2100 to subpart C to read
as follows:
■
§ 876.2100
tool.
Pressure ulcer management
(a) Identification. A pressure ulcer
management tool is a prescription
device intended for patients at risk of
developing pressure ulcers. The device
provides output that supports a user’s
decision to increase intervention. The
device is an adjunct tool for pressure
ulcer management that is not intended
for detection or diagnostic purposes.
(b) Classification. Class I (general
controls). The device is exempt from the
premarket notification procedures in
subpart E of part 807 of this chapter,
subject to the limitations in § 876.9.
Dated: December 8, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–26924 Filed 12–10–21; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF STATE
22 CFR Part 42
[Public Notice: 11460]
RIN 1400–AF20
Waiver of Personal Appearance and InPerson Oath Requirement for Certain
Immigrant Visa Applicants Due to
COVID–19
Department of State.
Final rule and temporary final
AGENCY:
khammond on DSKJM1Z7X2PROD with RULES
ACTION:
rule.
This temporary final rule
(TFR) provides flexibility for consular
officers to waive the personal
appearance of certain repeat immigrant
visa applicants who were approved for
an immigrant visa in the same
SUMMARY:
VerDate Sep<11>2014
15:56 Dec 10, 2021
Jkt 256001
classification and on the same basis as
the current application on or after
August 4, 2019. It also gives consular
officers discretion to allow this subset of
immigrant visa applicants to affirm the
accuracy of the contents of their
application without appearing in person
before a consular officer. This TFR is
effective immediately and expires after
24 months. The final rule portion of this
document reinstates parts of the
regulations with certain updates after
the expiration of the TFR.
DATES: Amendments in instructions 2
and 3 in this temporary final rule are
effective from December 13, 2021,
through December 13, 2023. The
amendment in instruction 4 is effective
December 13, 2023.
FOR FURTHER INFORMATION CONTACT:
Andrea Lage, Acting Senior Regulatory
Coordinator, Visa Services Directorate,
Bureau of Consular Affairs, Department
of State; telephone (202) 485–7586,
VisaRegs@state.gov.
SUPPLEMENTARY INFORMATION:
I. What changes to 22 CFR 42.62 and
42.67 does this TFR make?
The Department is temporarily
authorizing consular officers, for 24
months, to waive, on a discretionary
basis, the requirements in 22 CFR 42.62
and 42.67 that an immigrant visa
applicant appear in person before and
be interviewed by a consular officer for
certain repeat immigrant visa
applicants. This TFR applies to
immigrant visa applicants who were
issued a U.S. immigrant visa on or after
August 4, 2019, who meet the following
additional criteria: Individuals who
would be eligible for a discretionary
waiver of personal appearance and
interview pursuant to this TFR must be
seeking an immigrant visa in the same
classification (or another classification
as the result of automatic conversion
due to the death or naturalization of the
petitioner of the previously issued
immigrant visa) and pursuant to the
same approved petition as their
previously approved application, and
they must continue to qualify for the
immigrant visa sought.
Under this TFR, the personal
appearance and interview of certain
applicants for an immigrant visa may be
waived in the discretion of the consular
officer, provided that the applicant is
willing to affirm under penalty of
perjury to the information provided on
the Online Immigrant Visa and Alien
Registration Application, Form DS–260
(or Form DS–230, Application for
Immigrant Visa and Alien Registration if
the consular officer authorizes the use of
that form). The consular officer may
PO 00000
Frm 00047
Fmt 4700
Sfmt 4700
70735
communicate with the applicant by
telephone or email, may request that the
applicant provide additional
information that the consular officer
deems necessary, and may request the
applicant to appear in person. If the
applicant identifies the need to change
responses to Form DS–260, the consular
officer or other authorized consular staff
can reopen the DS–260 for the applicant
to make changes to that form and re-sign
it under penalty of perjury.
This TFR will automatically expire 24
months after it takes effect. As the TFR
is designed to help address the problem
of applicants who are unable to travel
due to the COVID–19 pandemic and
who must meet specific time-limited
criteria, this TFR will no longer be
necessary as the pandemic becomes less
acute and ordinary travel resumes. The
Department believes that 24 months is
sufficient to process the cases described.
Pursuant to section 222(a) of the
Immigration and Nationality Act (INA),
8 U.S.C. 1202(a), every immigrant visa
applicant must make an application in
the form, manner, and place prescribed
by regulation. Except as may otherwise
be prescribed by regulations, every
immigrant visa application must ‘‘be
signed by the applicant in the presence
of the consular officer and verified by
the oath of the applicant administered
by the consular officer.’’ INA 222(e), 8
U.S.C. 1202(e). Regulations further
require immigrant visa applicants to be
interviewed by a consular officer. 22
CFR 42.62(b). This TFR provides an
exception to these personal appearance
and interview requirements pursuant to
INA 222(a) and (e), 8 U.S.C. 1202(a) and
(e).
II. Why is the Department promulgating
this TFR?
A. The COVID–19 Pandemic
On January 31, 2020, the Secretary of
Health and Human Services declared a
public health emergency under section
319 of the Public Health Service Act (42
U.S.C. 247d) in response to COVID–19.1
On March 13, 2020, then-President
Trump declared a National Emergency
concerning the COVID–19 outbreak to
control the spread of the virus that
causes COVID–19 in the United States.2
That proclamation declared that the
emergency began in the United States
on March 1, 2020. In addition to the
National Emergency, a variety of
Presidential Proclamations have
1 HHS, Determination of Public Health
Emergency, 85 FR 7316 (Feb. 7, 2020).
2 Proclamation 9994 of March 13, 2020, Declaring
a National Emergency Concerning the Coronavirus
Disease (COVID–19) Outbreak, 85 FR 15337 (Mar.
18, 2020).
E:\FR\FM\13DER1.SGM
13DER1
Agencies
[Federal Register Volume 86, Number 236 (Monday, December 13, 2021)]
[Rules and Regulations]
[Pages 70733-70735]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2021-26924]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 876
[Docket No. FDA-2021-N-0898]
Medical Devices; Gastroenterology-Urology Devices; Classification
of the Pressure Ulcer Management Tool
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the pressure ulcer management tool into class I. We are taking this
action because we have determined that classifying the device into
class I will provide a reasonable assurance of safety and effectiveness
of the device. We believe this action will also enhance patients'
access to beneficial innovative devices.
DATES: This order is effective December 13, 2021. The classification
was applicable on December 20, 2018.
FOR FURTHER INFORMATION CONTACT: Gema Gonzalez, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2530, Silver Spring, MD 20993-0002, 301-796-6519,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the pressure ulcer management tool
as class I, which we have determined will provide a reasonable
assurance of safety and effectiveness. In addition, we believe this
action will enhance
[[Page 70734]]
patients' access to beneficial innovation by placing the device into a
lower device class than the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device
that does not require premarket approval. We determine whether a new
device is substantially equivalent to a predicate device by means of
the procedures for premarket notification under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
When FDA classifies a device into class I or II via the De Novo
process, the device can serve as a predicate for future devices of that
type, including for 510(k)s (see section 513(f)(2)(B)(i) of the FD&C
Act). As a result, other device sponsors do not have to submit a De
Novo request or premarket approval application to market a
substantially equivalent device (see section 513(i) of the FD&C Act,
defining ``substantial equivalence''). Instead, sponsors can use the
less-burdensome 510(k) process, when necessary, to market their device.
II. De Novo Classification
On April 3, 2017, FDA received Bruin Biometrics, LLC's request for
De Novo classification of the SEM Scanner (Model 200). FDA reviewed the
request in order to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class I if general controls are sufficient
to provide reasonable assurance of the safety and effectiveness of the
device for its intended use (see 21 U.S.C. 360c(a)(1)(A)). After review
of the information submitted in the request, we determined that the
device can be classified into class I. FDA has determined that general
controls will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on December 20, 2018, FDA issued an order to the
requester classifying the device into class I. In this final order, FDA
is codifying the classification of the device by adding 21 CFR
876.2100.\1\ We have named the generic type of device pressure ulcer
management tool, and it is identified as a prescription device intended
for patients at risk of developing pressure ulcers. The device provides
output that supports a user's decision to increase intervention. The
device is an adjunct tool for pressure ulcer management that is not
intended for detection or diagnostic purposes.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device: Adverse tissue reaction,
transmission of infection between patients, electromagnetic
interference with patient monitoring equipment, and electrical shock.
As previously stated, FDA believes general controls provide reasonable
assurance of safety and effectiveness for this device type.
At the time of classification, pressure ulcer management tools are
for prescription use only. Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR
801.5, as long as the conditions of 21 CFR 801.109 are met.
Section 510(l)(1) of the FD&C Act provides that a device within a
type that has been classified into class I under section 513 of the
FD&C Act is exempt from premarket notification under section 510(k),
unless the device is of substantial importance in preventing impairment
of human health or presents a potentially unreasonable risk of illness
or injury (21 U.S.C. 360(l)(1)). Devices within this type are exempt
from the premarket notification requirements under section 510(k),
subject to the limitations of exemptions in 21 CFR 876.9.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order refers to previously approved collections of
information found in other FDA regulations and guidance. These
collections of information are subject to review by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(44 U.S.C. 3501-3521). The collections of information in the guidance
document ``De Novo Classification Process (Evaluation of Automatic
Class III Designation)'' have been approved under OMB control number
0910-0844; the collections of information in 21 CFR part 814, subparts
A through E, regarding premarket approval, have been approved under OMB
control number 0910-0231; and the collections of
[[Page 70735]]
information in part 807, subpart E, regarding premarket notification
submissions, have been approved under OMB control number 0910-0120.
List of Subjects in 21 CFR Part 876
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
876 is amended as follows:
PART 876--GASTROENTEROLOGY-UROLOGY DEVICES
0
1. The authority citation for part 876 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 876.2100 to subpart C to read as follows:
Sec. 876.2100 Pressure ulcer management tool.
(a) Identification. A pressure ulcer management tool is a
prescription device intended for patients at risk of developing
pressure ulcers. The device provides output that supports a user's
decision to increase intervention. The device is an adjunct tool for
pressure ulcer management that is not intended for detection or
diagnostic purposes.
(b) Classification. Class I (general controls). The device is
exempt from the premarket notification procedures in subpart E of part
807 of this chapter, subject to the limitations in Sec. 876.9.
Dated: December 8, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021-26924 Filed 12-10-21; 8:45 am]
BILLING CODE 4164-01-P