Medical Devices; Orthopedic Devices; Classification of the Intraoperative Orthopedic Strain Sensor, 68403-68405 [2021-26183]
Download as PDF
Federal Register / Vol. 86, No. 229 / Thursday, December 2, 2021 / Rules and Regulations
68403
TABLE 1—CONDITIONING TOOL FOR EATING DISORDERS RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
Ineffective treatment leading to worsening condition of the patient, progression of disease, and/or delay of alternative treatments.
Adverse tissue reaction ............................................................................
Electrical shock or burns ..........................................................................
Nonclinical performance testing; Software validation, verification and
hazard analysis; and Labeling.
Biocompatibility evaluation.
Electrical safety testing, Electromagnetic compatibility (EMC) testing,
and Labeling.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. In order for
a device to fall within this classification,
and thus avoid automatic classification
in class III, it would have to comply
with the special controls named in this
final order. The necessary special
controls appear in the regulation
codified by this order. This device is
subject to premarket notification
requirements under section 510(k).
At the time of classification,
conditioning tools for eating disorders
are for prescription use only.
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act and 21 CFR
801.5, as long as the conditions of 21
CFR 801.109 are met (referring to 21
U.S.C. 352(f)(1)).
lotter on DSK11XQN23PROD with RULES1
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in part 814, subparts A
through E, regarding premarket
approval, have been approved under
OMB control number 0910–0231; the
collections of information in part 807,
subpart E, regarding premarket
notification submissions, have been
VerDate Sep<11>2014
16:00 Dec 01, 2021
Jkt 256001
approved under OMB control number
0910–0120; and the collections of
information in part 801, regarding
labeling, have been approved under
OMB control number 0910–0485.
Dated: November 26, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–26176 Filed 12–1–21; 8:45 am]
BILLING CODE 4164–01–P
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 882 is
amended as follows:
PART 882—NEUROLOGICAL DEVICES
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 888
[Docket No. FDA–2021–N–0648]
1. The authority citation for part 882
continues to read as follows:
Medical Devices; Orthopedic Devices;
Classification of the Intraoperative
Orthopedic Strain Sensor
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
AGENCY:
■
2. Add § 882.5060 to subpart F to read
as follows:
■
§ 882.5060
disorders.
Conditioning tool for eating
(a) Identification. A conditioning tool
for eating disorders is a prescription
device that non-invasively measures the
mass of food eaten during a meal and
provides feedback in the form of eating
rate, patient satiety, and eating pattern
information to the patient.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Nonclinical performance testing
must demonstrate:
(i) Device measurement accuracy and
repeatability; and
(ii) Device feedback accuracy.
(2) Software verification, validation,
and hazard analysis must be performed.
(3) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(4) Performance testing must
demonstrate the electromagnetic
compatibility (EMC) and electrical
safety of the device.
(5) Labeling and patient labeling must
be provided which includes the
following:
(i) Information identifying and
explaining how to use the device and its
components; and
(ii) Information on how the device
operates and the typical course of
treatment.
PO 00000
Frm 00015
Fmt 4700
Sfmt 4700
Food and Drug Administration,
Department of Health and Human
Services (HHS).
ACTION: Final amendment; final order.
The Food and Drug
Administration (FDA or we) is
classifying the intraoperative orthopedic
strain sensor into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the intraoperative
orthopedic strain sensor’s classification.
We are taking this action because we
have determined that classifying the
device into class II (special controls)
will provide a reasonable assurance of
safety and effectiveness of the device.
We believe this action will also enhance
patients’ access to beneficial innovative
devices.
DATES: This order is effective December
2, 2021. The classification was
applicable on March 28, 2019.
FOR FURTHER INFORMATION CONTACT:
Colin O’Neill, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4458, Silver Spring,
MD 20993–0002, 301–796–6428,
Colin.ONeill@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
Upon request, FDA has classified the
intraoperative orthopedic strain sensor
as class II (special controls), which we
have determined will provide a
E:\FR\FM\02DER1.SGM
02DER1
68404
Federal Register / Vol. 86, No. 229 / Thursday, December 2, 2021 / Rules and Regulations
reasonable assurance of safety and
effectiveness. In addition, we believe
this action will enhance patients’ access
to beneficial innovation, by placing the
device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
device by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
placed within class III, the De Novo
classification is considered to be the
initial classification of the device.
When FDA classifies a device into
class I or II via the De Novo process, the
device can serve as a predicate for
future devices of that type, including for
510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)).
As a result, other device sponsors do not
have to submit a De Novo request or
premarket approval application to
market a substantially equivalent device
(see 21 U.S.C. 360c(i), defining
‘‘substantial equivalence’’). Instead,
sponsors can use the less-burdensome
510(k) process, when necessary, to
market their device.
II. De Novo Classification
On July 19, 2018, Intellirod Spine,
Inc. submitted a request for De Novo
classification of the LOADPROTM
Intraoperative Rod Strain Sensor. FDA
reviewed the request in order to classify
the device under the criteria for
classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on March 28, 2019, FDA
issued an order to the requester
classifying the device into class II. In
this final order, FDA is codifying the
classification of the device by adding 21
CFR 888.3090.1 We have named the
generic type of device intraoperative
orthopedic strain sensor, and it is
identified as an adjunct tool intended to
measure strain on an orthopedic
implant in the intraoperative setting
only. The device is not intended to
provide diagnostic information or
influence clinical decision making.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
lotter on DSK11XQN23PROD with RULES1
TABLE 1—INTRAOPERATIVE ORTHOPEDIC STRAIN SENSOR RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
Prolonged operative time due to device error or
use error.
Electrical shock or device failure due to interference from other devices.
Infection ..............................................................
Adverse tissue reaction ......................................
Usability testing; Non-clinical performance testing; Software verification, validation, and hazard
analysis; and Labeling.
Electromagnetic compatibility testing, and Electrical safety testing.
Sterilization validation, Reprocessing validation, Shelf life testing, and Labeling.
Biocompatibility evaluation.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
1 FDA notes that the ‘‘ACTION’’ caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
VerDate Sep<11>2014
16:00 Dec 01, 2021
Jkt 256001
PO 00000
Frm 00016
Fmt 4700
Sfmt 4700
E:\FR\FM\02DER1.SGM
02DER1
Federal Register / Vol. 86, No. 229 / Thursday, December 2, 2021 / Rules and Regulations
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3521). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820, regarding quality system
regulation, have been approved under
OMB control number 0910–0073; and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 888
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 888 is
amended as follows:
Intraoperative orthopedic strain
(a) Identification. A strain sensor
device is an adjunct tool intended to
measure strain on an orthopedic
implant in the intraoperative setting
only. The device is not intended to
provide diagnostic information or
influence clinical decision making.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance testing must be conducted:
(i) Mechanical testing to evaluate the
effect of the device on the mechanical
performance of the implant and to
characterize the mechanical limits of the
components used with the implant; and
(ii) Accuracy and repeatability testing
of strain measurements.
(2) Usability testing must evaluate the
effect of the device on the performance
of the surgical procedure.
(3) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(4) Performance testing must support
the sterility and shelf life of the patientcontacting components of the device.
(5) Software verification, validation,
and hazard analysis must be performed.
(6) Performance data must validate
the reprocessing instructions for
reusable components of the device.
(7) Performance data must be
provided to demonstrate the
electromagnetic compatibility (EMC)
and electrical safety of the device.
(8) Labeling must include the
following:
(i) A shelf life;
(ii) Instructions for use;
(iii) Reprocessing instructions for any
reusable components; and
(iv) A statement that the device is not
intended to provide diagnostic
information or influence clinical
decision making.
Dated: November 29, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
PART 888—ORTHOPEDIC DEVICES
lotter on DSK11XQN23PROD with RULES1
§ 888.3090
sensor.
1. The authority citation for part 888
continues to read as follows:
■
[FR Doc. 2021–26183 Filed 12–1–21; 8:45 am]
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
BILLING CODE 4164–01–P
2. Add § 888.3090 to subpart D to read
as follows:
■
VerDate Sep<11>2014
16:00 Dec 01, 2021
Jkt 256001
PO 00000
Frm 00017
Fmt 4700
Sfmt 4700
68405
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Part 100
[Docket No. USCG–2021–0879]
Special Local Regulations; Charleston
Parade of Boats, Charleston, SC
Coast Guard, DHS.
Notification of enforcement of
regulation.
AGENCY:
ACTION:
The Coast Guard will enforce
the special local regulation for the
Charleston Parade of Boats on December
11, 2021. This action is necessary to
ensure safety of life on navigable waters
of the United States during the
Charleston Parade of Boats. During the
enforcement period, no person or vessel
may enter, transit through, anchor in, or
remain within the designated area
unless authorized by the Captain of the
Port Charleston (COTP) or a designated
representative.
DATES: The regulations in 33 CFR
100.704, Table 1 to § 100.704, Item No.
10, will be enforced from 4:00 p.m. until
8:30 p.m. on December 11, 2021.
FOR FURTHER INFORMATION CONTACT: If
you have questions about this notice of
enforcement, call or email LCDR Chad
Ray, Sector Charleston Office of
Waterways Management, Coast Guard;
telephone (843) 740–3184, email
Chad.L.Ray@uscg.mil.
SUPPLEMENTARY INFORMATION: The Coast
Guard will enforce the special local
regulation in 33 CFR 100.704, Item No.
10, for the Charleston Parade of Boats
from 4:00 p.m. through 8:30 p.m. on
December 11, 2021. This action is being
taken to provide for the safety of life on
navigable waterways during this event.
Our regulation for marine events within
the Seventh Coast Guard District
§ 100.704, Item No. 10, specifies the
location of the regulated area for the
Charleston Parade of Boats, which
encompasses a portion of the waterways
during the parade transit from
Charleston Harbor Anchorage A through
Shutes Folly, Bennis Reach, Horse
Reach, Hog Island Reach, Town Creek
Lower Reach, Ashley River, and
finishing at City Marina. During the
enforcement period, if you are the
operator of a vessel in the regulated area
you must comply with directions from
the Patrol Commander or any Official
Patrol displaying a Coast Guard ensign.
In addition to this notice of
enforcement in the Federal Register, the
Coast Guard plans to provide
notification of this enforcement period
SUMMARY:
E:\FR\FM\02DER1.SGM
02DER1
Agencies
[Federal Register Volume 86, Number 229 (Thursday, December 2, 2021)]
[Rules and Regulations]
[Pages 68403-68405]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2021-26183]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 888
[Docket No. FDA-2021-N-0648]
Medical Devices; Orthopedic Devices; Classification of the
Intraoperative Orthopedic Strain Sensor
AGENCY: Food and Drug Administration, Department of Health and Human
Services (HHS).
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the intraoperative orthopedic strain sensor into class II (special
controls). The special controls that apply to the device type are
identified in this order and will be part of the codified language for
the intraoperative orthopedic strain sensor's classification. We are
taking this action because we have determined that classifying the
device into class II (special controls) will provide a reasonable
assurance of safety and effectiveness of the device. We believe this
action will also enhance patients' access to beneficial innovative
devices.
DATES: This order is effective December 2, 2021. The classification was
applicable on March 28, 2019.
FOR FURTHER INFORMATION CONTACT: Colin O'Neill, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4458, Silver Spring, MD 20993-0002, 301-796-6428,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the intraoperative orthopedic
strain sensor as class II (special controls), which we have determined
will provide a
[[Page 68404]]
reasonable assurance of safety and effectiveness. In addition, we
believe this action will enhance patients' access to beneficial
innovation, by placing the device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate device by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
When FDA classifies a device into class I or II via the De Novo
process, the device can serve as a predicate for future devices of that
type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application to market a substantially equivalent
device (see 21 U.S.C. 360c(i), defining ``substantial equivalence'').
Instead, sponsors can use the less-burdensome 510(k) process, when
necessary, to market their device.
II. De Novo Classification
On July 19, 2018, Intellirod Spine, Inc. submitted a request for De
Novo classification of the LOADPROTM Intraoperative Rod
Strain Sensor. FDA reviewed the request in order to classify the device
under the criteria for classification set forth in section 513(a)(1) of
the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on March 28, 2019, FDA issued an order to the requester
classifying the device into class II. In this final order, FDA is
codifying the classification of the device by adding 21 CFR
888.3090.\1\ We have named the generic type of device intraoperative
orthopedic strain sensor, and it is identified as an adjunct tool
intended to measure strain on an orthopedic implant in the
intraoperative setting only. The device is not intended to provide
diagnostic information or influence clinical decision making.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Intraoperative Orthopedic Strain Sensor Risks and Mitigation
Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Prolonged operative time due Usability testing; Non-clinical
to device error or use error. performance testing; Software
verification, validation, and hazard
analysis; and Labeling.
Electrical shock or device Electromagnetic compatibility testing,
failure due to interference and Electrical safety testing.
from other devices.
Infection.................... Sterilization validation, Reprocessing
validation, Shelf life testing, and
Labeling.
Adverse tissue reaction...... Biocompatibility evaluation.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
[[Page 68405]]
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections
of information in the guidance document ``De Novo Classification
Process (Evaluation of Automatic Class III Designation)'' have been
approved under OMB control number 0910-0844; the collections of
information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collections of information in part 807, subpart E, regarding
premarket notification submissions, have been approved under OMB
control number 0910-0120; the collections of information in 21 CFR part
820, regarding quality system regulation, have been approved under OMB
control number 0910-0073; and the collections of information in 21 CFR
part 801, regarding labeling, have been approved under OMB control
number 0910-0485.
List of Subjects in 21 CFR Part 888
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act, and
under authority delegated to the Commissioner of Food and Drugs, 21 CFR
part 888 is amended as follows:
PART 888--ORTHOPEDIC DEVICES
0
1. The authority citation for part 888 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 888.3090 to subpart D to read as follows:
Sec. 888.3090 Intraoperative orthopedic strain sensor.
(a) Identification. A strain sensor device is an adjunct tool
intended to measure strain on an orthopedic implant in the
intraoperative setting only. The device is not intended to provide
diagnostic information or influence clinical decision making.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance testing must be conducted:
(i) Mechanical testing to evaluate the effect of the device on the
mechanical performance of the implant and to characterize the
mechanical limits of the components used with the implant; and
(ii) Accuracy and repeatability testing of strain measurements.
(2) Usability testing must evaluate the effect of the device on the
performance of the surgical procedure.
(3) The patient-contacting components of the device must be
demonstrated to be biocompatible.
(4) Performance testing must support the sterility and shelf life
of the patient-contacting components of the device.
(5) Software verification, validation, and hazard analysis must be
performed.
(6) Performance data must validate the reprocessing instructions
for reusable components of the device.
(7) Performance data must be provided to demonstrate the
electromagnetic compatibility (EMC) and electrical safety of the
device.
(8) Labeling must include the following:
(i) A shelf life;
(ii) Instructions for use;
(iii) Reprocessing instructions for any reusable components; and
(iv) A statement that the device is not intended to provide
diagnostic information or influence clinical decision making.
Dated: November 29, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021-26183 Filed 12-1-21; 8:45 am]
BILLING CODE 4164-01-P