Microbiology Devices; Reclassification of Human Immunodeficiency Virus Viral Load Monitoring Tests, 66982-66988 [2021-25372]
Download as PDF
66982
Proposed Rules
Federal Register
Vol. 86, No. 224
Wednesday, November 24, 2021
This section of the FEDERAL REGISTER
contains notices to the public of the proposed
issuance of rules and regulations. The
purpose of these notices is to give interested
persons an opportunity to participate in the
rule making prior to the adoption of the final
rules.
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA–2020–N–2297]
Microbiology Devices; Reclassification
of Human Immunodeficiency Virus
Viral Load Monitoring Tests
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed amendment; proposed
order.
The Food and Drug
Administration (FDA or the Agency) is
proposing to reclassify human
immunodeficiency virus (HIV) viral
load monitoring tests, a
postamendments class III device with
the product code MZF, into class II
(special controls), subject to premarket
notification. FDA is also proposing a
new device classification regulation
along with special controls that the
Agency believes are necessary to
provide a reasonable assurance of safety
and effectiveness for this device type.
FDA is proposing this reclassification
on its own initiative. If finalized, this
order will reclassify this device type
from class III (premarket approval) to
class II (special controls) and reduce the
regulatory burdens associated with
these devices because manufacturers
will no longer be required to submit a
premarket approval application (PMA)
for this device type but can instead
submit a less burdensome premarket
notification (510(k)) and receive
clearance before marketing their device.
DATES: Submit either electronic or
written comments on the proposed
order by January 24, 2022. See section
XI of this document for the proposed
effective date of any final order based on
this proposed order.
ADDRESSES: You may submit comments
as follows. Please note that late,
untimely filed comments will not be
considered. Electronic comments must
khammond on DSKJM1Z7X2PROD with PROPOSALS
SUMMARY:
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
be submitted on or before January 24,
2022. The https://www.regulations.gov
electronic filing system will accept
comments until 11:59 p.m. Eastern Time
at the end of January 24, 2022.
Comments received by mail/hand
delivery/courier (for written/paper
submissions) will be considered timely
if they are postmarked or the delivery
service acceptance receipt is on or
before that date.
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal:
https://www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand Delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2020–N–2297 for ‘‘Microbiology
Devices; Reclassification of Human
PO 00000
Frm 00001
Fmt 4702
Sfmt 4702
Immunodeficiency Virus Viral Load
Monitoring Tests.’’ Received comments,
those filed in a timely manner (see
ADDRESSES), will be placed in the docket
and, except for those submitted as
‘‘Confidential Submissions,’’ publicly
viewable at https://www.regulations.gov
or at the Dockets Management Staff
between 9 a.m. and 4 p.m., Monday
through Friday, 240–402–7500.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://
www.govinfo.gov/content/pkg/FR-201509-18/pdf/2015-23389.pdf.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852, 240–402–7500.
FOR FURTHER INFORMATION CONTACT:
Myrna Hanna, Center for Biologics
Evaluation and Review, Food and Drug
Administration, 10903 New Hampshire
E:\FR\FM\24NOP1.SGM
24NOP1
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
Ave., Bldg. 66, Rm. 5513, Silver Spring,
MD 20993–0002, 301–796–5739.
SUPPLEMENTARY INFORMATION:
khammond on DSKJM1Z7X2PROD with PROPOSALS
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (FD&C Act), as amended by the
Medical Device Amendments of 1976
(the 1976 amendments) (Pub. L. 94–
295), the Safe Medical Devices Act of
1990 (Pub. L. 101–629), the Food and
Drug Administration Modernization Act
of 1997 (Pub. L. 105–115), the Medical
Device User Fee and Modernization Act
of 2002 (Pub. L. 107–250), the Medical
Devices Technical Corrections Act (Pub.
L. 108–214), the Food and Drug
Administration Amendments Act of
2007 (Pub. L. 110–85), and the Food and
Drug Administration Safety and
Innovation Act (Pub. L. 112–144),
among other amendments, establishes a
comprehensive system for the regulation
of medical devices intended for human
use. Section 513 of the FD&C Act (21
U.S.C. 360c) established three categories
(classes) of devices, reflecting the
regulatory controls needed to provide
reasonable assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (general controls and special
controls), and class III (general controls
and premarket approval).
Section 513(a)(1) of the FD&C Act
defines the three classes of devices.
Class I devices are those devices for
which the general controls of the FD&C
Act (controls authorized by or under
sections 501, 502, 510, 516, 518, 519, or
520 (21 U.S.C. 351, 352, 360, 360f, 360h,
360i, or 360j) or any combination of
such sections) are sufficient to provide
reasonable assurance of safety and
effectiveness; or those devices for which
insufficient information exists to
determine that general controls are
sufficient to provide reasonable
assurance of safety and effectiveness or
to establish special controls to provide
such assurance, but because the devices
are not purported or represented to be
for a use in supporting or sustaining
human life or for a use which is of
substantial importance in preventing
impairment of human health, and do
not present a potential unreasonable
risk of illness or injury, are to be
regulated by general controls (section
513(a)(1)(A) of the FD&C Act). Class II
devices are those devices for which
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness
and for which there is sufficient
information to establish special controls
to provide such assurance, including the
promulgation of performance standards,
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
postmarket surveillance, patient
registries, development and
dissemination of guidelines,
recommendations, and other
appropriate actions the Agency deems
necessary to provide such assurance
(section 513(a)(1)(B) of the FD&C Act).
Class III devices are those devices for
which insufficient information exists to
determine that general controls and
special controls would provide a
reasonable assurance of safety and
effectiveness, and are purported or
represented to be for a use in supporting
or sustaining human life or for a use
which is of substantial importance in
preventing impairment of human
health, or present a potential
unreasonable risk of illness or injury
(section 513(a)(1)(C) of the FD&C Act).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices) are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, (1) FDA reclassifies the device
into class I or class II, or (2) FDA issues
an order finding the device to be
substantially equivalent, in accordance
with section 513(i) of the FD&C Act, to
a predicate device that does not require
premarket approval. FDA determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act and part 807 (21 CFR part
807), subpart E, of the regulations.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or II
under section 513(f)(3) of the FD&C Act.
Section 513(f)(3) of the FD&C Act
provides that FDA, acting by
administrative order, can reclassify the
device into class I or class II on its own
initiative, or in response to a petition
from the manufacturer or importer of
the device. To change the classification
of the device, the proposed new class
must have sufficient regulatory controls
to provide a reasonable assurance of the
safety and effectiveness of the device for
its intended use.
FDA relies upon ‘‘valid scientific
evidence,’’ as defined in section
513(a)(3) and 21 CFR 860.7(c)(2), in the
classification process to determine the
level of regulation for devices. To be
considered in the reclassification
process, the ‘‘valid scientific evidence’’
upon which the Agency relies must be
publicly available (see section 520(c) of
the FD&C Act). Publicly available
PO 00000
Frm 00002
Fmt 4702
Sfmt 4702
66983
information excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA (see
section 520(c) of the FD&C Act).
In accordance with section 513(f)(3) of
the FD&C Act, the Agency is proposing
to reclassify HIV viral load monitoring
tests, a postamendments class III device,
into class II (special controls), subject to
premarket notification because the
Agency believes the standard in
513(a)(1)(B) of the FD&C Act is met
because there is sufficient information
to establish special controls, which, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.1
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the 510(k) premarket
notification requirements under section
510(k) of the FD&C Act if the Agency
determines that premarket notification
is not necessary to reasonably assure the
safety and effectiveness of the device.
FDA has determined that premarket
notification is necessary to reasonably
assure the safety and effectiveness of
HIV viral load monitoring tests.
Therefore, the Agency does not intend
to exempt this proposed class II device
from premarket notification (510(k))
submission under section 510(m) of the
FD&C Act.
II. Regulatory History of the Devices
This proposed order addresses HIV
viral load monitoring tests. These are
prescription tests that measure HIV
RNA as an aid in monitoring patient
status and are assigned product code
MZF.2 These postamendments devices
are currently regulated as class III
devices under section 513(f)(1) of the
FD&C Act. Based on our review
experience and consistent with the
FD&C Act and FDA’s regulations in 21
CFR 860.134, FDA believes that these
devices should be reclassified from class
III into class II with special controls
because special controls, in addition to
general controls, are necessary and
sufficient to provide reasonable
assurance of the safety and effectiveness
of these devices and there is sufficient
1 In December 2019, FDA began adding the term
‘‘Proposed amendment’’ to the ‘‘ACTION’’ caption
for these documents, typically styled ‘‘Proposed
order’’, to indicate that they ‘‘propose to amend’’
the Code of Federal Regulations. This editorial
change was made in accordance with the Office of
Federal Register’s (OFR) interpretations of the
Federal Register Act (44 U.S.C. chapter 15), its
implementing regulations (1 CFR 5.9 and parts 21
and 22), and the Document Drafting Handbook.
2 On February 21, 2020, FDA published a separate
proposed order to reclassify certain HIV serological
diagnostic and supplemental tests and HIV nucleic
acid diagnostic and supplemental tests, which are
also currently assigned product code MZF, from
class III into class II (85 FR 10110).
E:\FR\FM\24NOP1.SGM
24NOP1
66984
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
khammond on DSKJM1Z7X2PROD with PROPOSALS
information to establish special controls
to provide such assurance.
FDA approved the first in vitro
nucleic acid amplification-based HIV
viral load test on June 3, 1996, for the
quantitation of HIV–1 RNA in human
plasma. Currently, there are six HIV
viral load monitoring tests on the
market, all of which met the
performance criteria specified in the
proposed special controls, considered
necessary for the intended use of the
test, when they were approved by FDA
(Ref. 1).
A review of the FDA’s medical device
reporting database, MAUDE
(Manufacturer and User Facility Device
Experience), indicates a low number of
reported events for HIV viral load
monitoring tests. Millions of devices
intended for use as HIV viral load
monitoring tests have been sold since
1996 with fewer than 200 reported
adverse events as of October 2020. Of
these events, fewer than 10 are reported
to involve injuries due to incorrect
results; the remainder are malfunctions,
such as user errors or incorrect results,
that had no reported effect on the
individual being monitored. There has
been one class II recall and no class I
(highest risk) recalls specific to HIV
viral load monitoring tests, which,
coupled with the low number of
reported events, indicates a good safety
record for this device class.
III. Device Description
This proposed order applies to HIV
viral load monitoring tests that are
prescription in vitro diagnostic devices
for monitoring of HIV viral load in body
fluids. As such, these prescription
devices must satisfy prescription
labeling requirements for in vitro
diagnostic products (see 21 CFR
809.10(a)(4) and (b)(5)(ii)). HIV viral
load monitoring tests are intended for
use in the clinical management of
individuals living with HIV and are for
professional use only. These devices are
not intended for use as an aid in
diagnosis or for screening donors of
blood or blood products or human cells,
tissues, or cellular and tissue-based
products (HCT/Ps).
HIV viral load monitoring tests are
quantitative in vitro diagnostic tests that
measure the amount of HIV RNA in
human body fluids such as plasma and
whole blood. The HIV RNA is isolated,
amplified, and detected by labeled
probes that produce a quantitative
output that determines the amount of
HIV in the sample. The test results then
are used as part of patient management
decisions in conjunction with other
relevant clinical and laboratory
findings.
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
Approval of HIV viral load monitoring
tests has been based on studies and
established clinical decision points that
demonstrate that changes in viral load
correlate with clinically meaningful
outcomes, meaning that HIV RNA
measurements could reliably assess the
success or failure of antiretroviral
therapy (ART) (Ref. 1).
IV. Proposed Reclassification
FDA is proposing to reclassify HIV
viral load monitoring tests. At a public
meeting held on July 19, 2018, the
Blood Products Advisory Committee,
convened as a medical device panel
(‘‘the Panel’’), unanimously agreed that
special controls, in addition to general
controls, are sufficient to mitigate the
risk to health from incorrect results from
HIV nucleic acid and serological
diagnostic and supplemental tests. The
Panel believed that class II with the
special controls would provide
reasonable assurance of the safety and
effectiveness of those devices. In
February 2020, FDA issued a proposed
order that, if finalized, would reclassify
those devices from class III into class II
(85 FR 10110). As part of the Panel’s
discussion, the Panel also recommended
that FDA consider reclassification of
quantitative HIV tests indicated for use
for monitoring HIV viral load from class
III to class II (Ref. 2).
In accordance with section 513(f)(3) of
the FD&C Act and 21 CFR part 860,
subpart C, FDA is proposing to
reclassify postamendments HIV viral
load monitoring tests from class III into
class II. FDA believes that there are
sufficient data and information available
through FDA’s accumulated experience
with these devices and from published
literature to demonstrate that the
proposed special controls, along with
general controls, would effectively
mitigate the risks to health identified in
section V of this document and provide
reasonable assurance of safety and
effectiveness of these devices. Absent
the special controls identified in this
proposed order, general controls
applicable to the device are insufficient
to provide reasonable assurance of the
safety and effectiveness.
FDA is proposing to create a separate
classification regulation for HIV viral
load monitoring tests. Under this
proposed order, if finalized, HIV viral
load monitoring tests will be reclassified
from class III to class II and identified
as prescription devices. In this proposed
order the Agency has proposed the
special controls under section
513(a)(1)(B) of the FD&C Act that,
together with general controls, would
provide reasonable assurance of the
PO 00000
Frm 00003
Fmt 4702
Sfmt 4702
safety and effectiveness of HIV viral
load monitoring tests.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C if FDA determines that premarket
notification is not necessary to provide
reasonable assurance of the safety and
effectiveness of the device. For these
HIV viral load monitoring tests, FDA
has determined that premarket
notification is necessary to provide a
reasonable assurance of safety and
effectiveness. Therefore, FDA does not
intend to exempt this proposed class II
device from the 510(k) requirements. If
this proposed order is finalized, persons
who intend to market this type of device
must submit to FDA a 510(k) and
receive clearance prior to marketing the
device.
This proposed order, if finalized, will
decrease regulatory burden on industry
because manufacturers will no longer
have to submit a PMA for this device
type, but can instead submit a 510(k) to
the Agency for review prior to
marketing their device. A 510(k) is a
less-burdensome pathway to market a
device, which typically results in a
shorter premarket review timeline
compared with a PMA, helping to
provide more timely access to this
device type to patients. FDA expects
that the reclassification of these devices
would enable more manufacturers to
develop HIV viral load monitoring tests
such that patients would benefit from
increased access to safe and effective
tests.
V. Risks to Health
Viral load is one of the important
markers for monitoring the effectiveness
of ART and disease progression. The
Department of Health and Human
Services (HHS) in 2014 issued
guidelines on the treatment of HIV in
adults and adolescents in United States.
The guidelines are updated periodically
based on new data. Regarding viral load
monitoring, the HHS guidelines define
optimal viral load suppression as
suppressing viral load levels
persistently to <20 to 75 copies per
milliliter (/mL) of HIV RNA, depending
on the assay used (Ref. 3). Virologic
failure, at which point changes in
treatment are considered, is defined as
the inability to achieve or maintain
suppression of viral replication to an
HIV RNA level <200 copies/mL (Ref. 3).
The HHS guidelines recommend that
viral load testing should be performed
for all patients with HIV at entry into
care, initiation of therapy, and on a
regular basis thereafter (Ref. 3).
Therefore, improving access to HIV viral
E:\FR\FM\24NOP1.SGM
24NOP1
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
khammond on DSKJM1Z7X2PROD with PROPOSALS
load monitoring tests is a public health
priority. After considering FDA’s
accumulated experience with these
devices from review submissions and
the published literature, FDA has
identified the following probable risks
to health associated with HIV viral load
monitoring tests:
(1) An inaccurately low viral load test
result may influence patient
management decisions, such as
continuation of ineffective treatment,
which can lead to serious injury
including death. Inaccurately low viral
load test results also may contribute to
public health risk by leading to
inadvertent transmission of the virus by
an individual living with HIV. Factors
that may cause decreased test sensitivity
and/or increased risk of inaccurately
low viral load test results include, but
are not limited to, viral strain
variability, acquisition of de novo
mutations in genomic regions of HIV
targeted by the device, and the presence
of interfering substances in the sample.
Inaccurately low results also can be
caused by improper sample collection
or sample handling, loss of sensitivity of
the device, failure of detection reagents,
and failure of instruments.
(2) An inaccurately high viral load
test result may contribute to an
unnecessary change in therapy,
potentially ending effective treatment
and leading to less effective
management of disease, as well as
significant emotional stress. Factors that
may cause an increased rate of
inaccurately high viral load test results
include, but are not limited to, crossreactivity with other substances in the
sample, carryover, viral strain
variability, or acquisition of de novo
mutations in genes other than HIV.
Inaccurately high results also can be
caused by improper sample collection
and sample contamination.
(3) Incorrect interpretation of test
results by healthcare professionals may
result in patient management decisions,
such as continuation of ineffective
therapy or an unnecessary change in
therapy, that could lead to serious
injury, including death, and less
effective management of the disease.
Incorrect interpretation of results may
be caused by inadequate labeling,
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
including insufficient limitations,
warnings, and explanations of test
procedure.
VI. Summary of the Reasons for
Reclassification
FDA believes that HIV viral load
monitoring tests should be reclassified
from class III into class II (special
controls) because special controls, in
addition to general controls, can be
established to mitigate the risks to
health identified in section V and
provide reasonable assurance of the
safety and effectiveness of this device
type. The proposed special controls are
identified by FDA in section VII of this
proposed order. FDA’s reasons for
reclassification are as follows:
(1) There is substantial scientific and
medical information available regarding
the nature and complexity of, and risks
associated with, HIV viral load
monitoring tests (Refs. 3 to 11). The
safety and effectiveness of this device
type has become well-established by the
performance of the approved HIV viral
load tests (Ref. 1).
(2) Risks associated with the failure of
the device to perform as indicated (e.g.,
inaccurately high or low test results)
and risks associated with incorrect
interpretation of results can be mitigated
through a combination of special
controls, including performance criteria,
certain labeling requirements, and
submission of certain manufacturing
information. Performance criteria would
consist primarily of analytical and
method comparison study design
specifications and acceptance criteria
that are based on public information
regarding the performance and
validation of previously approved
devices. FDA expects that a device
would demonstrate acceptable
performance, e.g., analytical sensitivity,
at clinically relevant medical decision
points at the time of clearance. This
would help ensure that devices meet or
exceed the performance of other cleared
or approved HIV viral load tests at
existing clinically relevant medical
decision points and, in the future,
demonstrate similar performance if
there are changes in those medical
decision points that reflect additional
evidence and/or medical advances.
PO 00000
Frm 00004
Fmt 4702
Sfmt 4702
66985
Examples of labeling mitigations
include appropriate limitations,
including that results should be
interpreted in conjunction with the
individual’s clinical presentation,
history, and other laboratory results.
These are necessary to ensure that the
devices are used correctly, and the
results are interpreted appropriately,
given the diversity of settings in which
these devices are intended to be used.
Manufacturing information required to
be submitted would include summaries
of strategies to quantitate new HIV
types, subtypes, genotypes, and
mutations to ensure the tests continue to
monitor clinically relevant forms of
HIV. It also would include a detailed
device description, including
information on number and design of
primers and probes, which should be
designed according to current best
practices and professional
recommendations. It would also include
appropriate and acceptable procedures
to determine the severity of events to
ensure appropriate adverse event
reporting, protocols for assessing
stability, and evaluation of test
performance at the extremes of
specifications to ensure the tests have
been validated to function correctly
under diverse conditions.
Taking into account the established
health benefits of the use of the device
and the nature of the probable risks of
the device (Refs. 1, 3 to 11), FDA, on its
own initiative, is proposing to reclassify
these postamendments devices from
class III into class II. FDA believes that,
when used as indicated, HIV viral load
monitoring tests can provide significant
benefits to clinicians and patients.
VII. Proposed Special Controls
FDA believes that these devices can
be classified into class II with the
establishment of special controls. FDA
believes that these special controls, in
addition to general controls, will
provide a reasonable assurance of the
safety and efficacy of these devices.
Table 1 demonstrates how these
proposed special controls will mitigate
each of the identified risks to health in
section V.
E:\FR\FM\24NOP1.SGM
24NOP1
66986
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR HIV VIRAL LOAD MONITORING TESTS
Identified risks to health
Mitigation measures
An inaccurately low test result may influence patient
management decisions, including continuation of ineffective antiviral therapy which can lead to serious injury
including death. An inaccurately low test result may
contribute to public health risk by leading to inadvertent
transmission of the virus by a person living with HIV.
Certain labeling limitations, warnings, and explanations of the procedures and interpretation results.
Analytical sensitivity and method comparison study performance criteria.
Acceptable strategies for monitoring emergence of and ability of the test to detect
new or altered circulating forms of HIV.
Certain other device verification and validation information, including acceptable
processes for risk analysis, testing performance at extremes of specifications, and
determining severity of adverse events and malfunctions.
An inaccurately high test result may contribute to unnecLabeling instructions for appropriate confirmation of elevated results.
essary change in therapy, potentially disrupting effecAnalytical performance criteria.
tive treatment and leading to less effective manageAcceptable validation of susceptibility to interference and cross-reactivity.
ment of disease, as well as significant emotional stress. Acceptable processes for risk analysis, testing performance at extremes of specifications, and determining severity of adverse events and malfunctions.
Incorrect interpretation of test results may result in paCertain labeling limitations, warnings, and explanations of the procedures and intertient management decisions, such as continuation of
pretation results.
ineffective therapy or an unnecessary change in therapy, that could lead to serious injury, including death,
and less effective management of the disease.
If this proposed order is finalized,
HIV viral load monitoring tests will be
reclassified into class II (special
controls). As discussed below, the
reclassification will be codified in 21
CFR 866.3958. Firms submitting a
510(k) for an HIV viral load monitoring
test will be required to comply with the
particular mitigation measures set forth
in the special controls. Adherence to the
special controls, in addition to the
general controls, is necessary to provide
a reasonable assurance of the safety and
effectiveness of the devices.
VIII. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
khammond on DSKJM1Z7X2PROD with PROPOSALS
IX. Paperwork Reduction Act of 1995
FDA tentatively concludes that this
proposed order contains no new
collection of information. Therefore,
clearance by the Office of Management
and Budget (OMB) under the Paperwork
Reduction Act of 1995 (PRA) (44 U.S.C.
3501–3521) is not required. This
proposed order refers to previously
approved FDA collections of
information. These collections of
information are subject to review by the
OMB under the PRA. The collections of
information in part 807, subpart E,
regarding premarket notification
submissions have been approved under
OMB control number 0910–0120; the
collections of information in 21 CFR
part 820 have been approved under
OMB control number 0910–0073; the
collections of information in 21 CFR
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
part 803 have been approved under
OMB control number 0910–0437; and
the collections of information in 21 CFR
parts 801 and 809 have been approved
under OMB control number 0910–0485.
X. Codification of Orders
Under section 513(f)(3) of the FD&C
Act, FDA may issue final orders to
reclassify devices. FDA will continue to
codify classifications and
reclassifications in the Code of Federal
Regulations (CFR). Changes resulting
from final orders will appear in the CFR
as newly codified orders. Therefore,
under section 513(f)(3), in the proposed
order, we are proposing to codify HIV
viral load monitoring tests in the new 21
CFR 866.3958, under which HIV viral
load monitoring tests would be
reclassified from class III to class II.
XI. Proposed Effective Date
FDA proposes that any final order
based on this proposed order become
effective 30 days after its date of
publication in the Federal Register.
XII. References
The following references are on
display at the Dockets Management Staff
(see ADDRESSES) and are available for
viewing by interested persons between
9 a.m. and 4 p.m., Monday through
Friday; they are also available
electronically at https://
www.regulations.gov. FDA has verified
the website addresses, as of the date this
document publishes in the Federal
Register, but websites are subject to
change over time.
1. List of approved HIV viral load monitoring
tests with supporting information can be
found at https://www.fda.gov/vaccinesblood-biologics/approved-bloodproducts/premarket-approvals-and-
PO 00000
Frm 00005
Fmt 4702
Sfmt 4702
humanitarian-device-exemptionssupporting-documents.
2. ‘‘Reclassification of HIV Point of Care and
Laboratory-based serological and NAT
diagnostic devices from Class III (PMA)
to Class II 510(k); Issue Summary
Prepared for the July 19, 2018, Meeting
of the Blood Products Advisory
Committee (BPAC)).’’ Available at:
https://www.fda.gov/advisorycommittees/blood-products-advisorycommittee/2018-meeting-materialsblood-products-advisory-committee.
3. ‘‘Guidelines for the Use of Antiretroviral
Agents in Adults and Adolescents Living
with HIV.’’ Department of Health and
Human Services. Accessed November 24,
2020. Available at: https://
clinicalinfo.hiv.gov/en/guidelines/adultand-adolescent-arv/antiretroviraltherapy-prevent-sexual-transmission-hiv.
4. ‘‘Human Immunodeficiency Virus-1
Infection: Developing Antiretroviral
Drugs for Treatment; Guidance for
Industry.’’ Available at: https://
www.fda.gov/media/86284/download.
5. Aberg, J.A., J.E. Gallant, K.H. Ghanem, et
al., ‘‘Primary Care Guidelines for the
Management of Persons Infected with
HIV: 2013 Update by the HIV Medicine
Association of the Infectious Diseases
Society of America,’’ Clinical Infectious
Disease, 58:e1–34, 2014.
6. Saag, M.S., M. Holodniy, D.R. Kuritzkes,
et al., ‘‘HIV Viral Load Markers in
Clinical Practice,’’ Nature Medicine,
2:625–629, 1996.
7. Das, M., P.L. Chu, G-M. Santos, et al.,
‘‘Decreases in Community Viral Load are
Accompanied by Reductions in New HIV
Infections in San Francisco,’’ PLoS ONE,
5:e11068, 2010.
8. Stadhouders, R., S.D. Pas, J. Anber, et al.,
‘‘The Effect of Primer-Template
Mismatches on the Detection and
Quantification of Nucleic Acids Using
the 5’ Nuclease Assay,’’ Journal of
Molecular Diagnostics, 12:109–117,
2010.
9. Swenson, L.C., B. Cobb, A.M. Geretti, et
al., ‘‘Comparative Performances of HIV–
E:\FR\FM\24NOP1.SGM
24NOP1
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
1 RNA Load Assays at Low Viral Load
Levels: Results of an International
Collaboration,’’ Journal of Clinical
Microbiology, 52(2):517–523, 2014.
10. Caniglia, E.C., C. Sabin, J.M. Robins, et
al., ‘‘When to Monitor CD4 Cell Count
and HIV RNA to Reduce Mortality and
AIDS-Defining Illness in Virologically
Suppressed HIV-Positive Persons on
Antiretroviral Therapy in High-Income
Countries: A Prospective Observational
Study,’’ Journal of Acquired Immune
Deficiency Syndromes, 72:214–221,
2016.
11. Shoko, C. and D. Chikobvu, ‘‘A
Superiority of Viral Load Over CD4 Cell
Count When Predicting Mortality in HIV
Patients on Therapy.’’ BioMed Central
Infectious Diseases, 19:169, 2019.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical
devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act it is proposed
that 21 CFR part 866 be amended as
follows:
PART 866—IMMUNOLOGY AND
MICROBIOLOGY DEVICES
1. The authority citation for part 866
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 866.3958 to subpart D to read
as follows:
■
khammond on DSKJM1Z7X2PROD with PROPOSALS
§ 866.3958 Human immunodeficiency virus
(HIV) viral load monitoring test.
(a) Identification. A human
immunodeficiency virus (HIV) viral
load monitoring test is an in vitro
diagnostic prescription device for the
quantitation of the amount of HIV RNA
in human body fluids. The test is
intended for use in the clinical
management of individuals living with
HIV and is for professional use only.
The test results are intended to be
interpreted in conjunction with other
relevant clinical and laboratory
findings. The test is not intended to be
used as an aid in diagnosis or for
screening donors of blood or blood
products or human cells, tissues, or
cellular and tissue-based products
(HCT/Ps).
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The labeling must include:
(i) An intended use that states that the
device is not intended for use as an aid
in diagnosis or for use in screening
donors of blood or blood products, or
HCT/Ps.
(ii) A detailed explanation of the
principles of operation and procedures
used for assay performance.
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
(iii) A detailed explanation of the
interpretation of results and
recommended actions to take based on
current clinical guidelines.
(iv) Limitations, which must be
updated to reflect current clinical
practice and patient management. The
limitations must include, but are not
limited to, statements that indicate:
(A) The matrices and sample types
with which the device has been cleared
and that use of this test with specimen
types other than those specifically
cleared for this device may cause
inaccurate test results.
(B) Mutations in highly conserved
regions may affect binding of primers
and/or probes resulting in the underquantitation of virus or failure to detect
the presence of virus.
(C) All test results should be
interpreted in conjunction with the
individual’s clinical presentation,
history, and other laboratory results.
(2) Device verification and validation
must include:
(i) Detailed device description,
including the device components,
ancillary reagents required but not
provided, and an explanation of the
device methodology. Additional
information appropriate to the
technology must be included, such as
detailed information on the design of
primers and probes.
(ii) For devices with assay calibrators,
the design and nature of all primary,
secondary, and subsequent quantitation
standards used for calibration as well as
their traceability to a reference material.
In addition, analytical testing must be
performed following the release of a
new lot of the standard material that
was used for device clearance, or when
there is a transition to a new calibration
standard.
(iii) Detailed documentation of
analytical performance studies
conducted as appropriate to the
technology, specimen types tested, and
intended use of the device, including
but not limited to, limit of blank, limit
of detection, limit of quantitation, cutoff
determination, precision, linearity,
endogenous and exogenous
interferences, cross-reactivity, carryover, quality control, matrix
equivalency, sample and reagent
stability. Samples selected for use in
analytical studies or used to prepare
samples for use in analytical studies
must be from subjects with clinically
relevant genotypes circulating in the
United States.
(iv) Multisite reproducibility study
that includes the testing of three
independent production lots.
(v) Analytical sensitivity of the device
must demonstrate acceptable
PO 00000
Frm 00006
Fmt 4702
Sfmt 4702
66987
performance at current clinically
relevant medical decision points.
Samples tested to demonstrate
analytical sensitivity must include
appropriate numbers and types of
samples, including real clinical samples
near the lower limit of quantitation and
any clinically relevant medical decision
points. Analytical specificity of the
device must demonstrate acceptable
performance. Samples tested to
demonstrate analytical specificity must
include appropriate numbers and types
of samples from patients with different
underlying illnesses and infection and
from patients with potential interfering
substances.
(vi) Detailed documentation of
performance from a multisite clinical
study or a multisite analytical method
comparison study.
(A) For devices evaluated in a
multisite clinical study, the study must
use specimens from individuals living
with HIV being monitored for changes
in viral load, and the test results must
be compared to the clinical status of the
patients.
(B) For tests evaluated in a multisite
analytical method comparison study,
the performance of the test must be
compared to an FDA-cleared or
approved comparator. The multisite
method comparison study must include
appropriate numbers and types of
samples with analyte concentrations
across the measuring range of the assay,
representing clinically relevant
genotypes. The multisite method
comparison study design, including
number of samples tested, must be
sufficient to meet the following criteria:
(1) Agreement between the two tests
across the measuring range of the assays
must have an r2 of greater than or equal
to 0.95.
(2) The bias between the test and
comparator assay, as determined by
difference plots, must be less than or
equal to 0.5 log copies/mL.
(vii) If a multisite clinical study is
performed under paragraph (b)(2)(vi) of
this section, detailed documentation of
a single-site analytical method
comparison study between the device
and an FDA-cleared or approved
comparator. The analytical method
comparison study must use appropriate
numbers and types of samples with
analyte concentrations across the
measuring range of the assay,
representing clinically relevant
genotypes. The results must meet the
criteria in paragraphs (b)(2)(vi)(B)(1) and
(2) of this section.
(viii) Strategies for detection of new
strains, types, subtypes, genotypes, and
genetic mutations as they emerge.
E:\FR\FM\24NOP1.SGM
24NOP1
66988
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 / Proposed Rules
(ix) Risk analysis and management
strategies, such as Failure Modes Effects
Analysis and/or Hazard Analysis and
Critical Control Points summaries and
their impact on test performance.
(x) Final release criteria to be used for
manufactured device lots with an
appropriate justification that lots
released at the extremes of the
specifications will meet the claimed
analytical and clinical performance
characteristics as well as the stability
claims.
(xi) All stability protocols, including
acceptance criteria.
(xii) Appropriate and acceptable
procedure(s) for addressing complaints
and other device information that
determines when to submit a medical
device report.
(xiii) Premarket notification
submissions must include the
information contained in paragraphs
(b)(2)(i) through (xii) of this section.
Dated: November 16, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021–25372 Filed 11–23–21; 8:45 am]
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Part 117
[Docket No. USCG–2021–0778]
RIN 1625–AA09
Drawbridge Operation Regulation;
Willamette River, Portland, OR
Coast Guard, Department of
Homeland Security (DHS).
ACTION: Notice of proposed rulemaking.
AGENCY:
The Coast Guard proposes to
modify the operating schedule that
governs the Morrison Bridge across the
Willamette River, mile 12.8, at Portland,
OR. Multnomah County, Oregon, the
bridge owner, is requesting to change
the current regulation to allow painting
and preservation of the Morrison Bridge
including the double bascule span. The
modified rule would change from a full
span opening to a single leaf, half
opening, and operation. We invite your
comments on this proposed rulemaking.
DATES: Comments and related material
must reach the Coast Guard on or before
December 27, 2021. The Coast Guard
anticipates that this proposed rule will
be effective from 7 p.m. on April 1,
2022, through 7 p.m. on May 31, 2023.
ADDRESSES: You may submit comments
identified by docket number USCG–
khammond on DSKJM1Z7X2PROD with PROPOSALS
VerDate Sep<11>2014
16:21 Nov 23, 2021
Jkt 256001
If
you have questions on this proposed
rule, call or email Steven Fischer,
Thirteenth District Bridge
Administrator, U.S. Coast Guard;
telephone 206–220–7282, email d13smb-d13-bridges@uscg.mil.
FOR FURTHER INFORMATION CONTACT:
SUPPLEMENTARY INFORMATION:
I. Table of Abbreviations
CFR Code of Federal Regulations
DHS Department of Homeland Security
FR Federal Register
OMB Office of Management and Budget
NPRM Notice of proposed rulemaking
(Advance, Supplemental)
§ Section
U.S.C. United States Code
County Multnomah County
II. Background, Purpose and Legal
Basis
BILLING CODE 4164–01–P
SUMMARY:
2021–0778 using Federal Decision
Making Portal at https://
www.regulations.gov. See the ‘‘Public
Participation and Request for
Comments’’ portion of the
SUPPLEMENTARY INFORMATION section
below for instructions on submitting
comments.
Multnomah County, Oregon, owns
and operates the Morrison Bridge across
the Willamette River at mile 12.8. The
County is requesting a temporary
change to the existing operating
regulation. The County is proposing to
open the Morrison Bridge’s span in
single leaf mode, half of the double
bascule span, to marine vessels with a
minimum of two-hour notice, or fourhour notice if a tug assist is needed. The
County needs to maintain half of the
draw closed to allow for preservation
and paint efforts. The proposed
regulation change would allow the
Morrison Bridge to alternate operation
of the east or west leaf span from April
1, 2022, through May 31, 2023. The west
span will be operational at the
beginning of construction and the east
span will be closed to navigation. The
dates to switch operational spans will
be determined later and published in
the Local and Broadcast Notice to
Mariners. This proposal also allows a
containment system under the bridge
that reduces the non-opening half of the
bridge’s vertical clearance by 5 feet from
69 feet center to 64 feet, and from 48
feet on the sides to 43 feet above the
Columbia River Datum 0.0. Marine
traffic on this section of the Willamette
River consists of vessels ranging from
small pleasure craft up to large
commercial vessels and barges. The
subject bridge currently operates in
accordance with 33 CFR
117.897(c)(3)(iv).
PO 00000
Frm 00007
Fmt 4702
Sfmt 4702
III. Discussion of Proposed Rule
The Coast Guard proposes a
temporary change to 33 CFR
117.897(c)(3)(iv) to be in effect from 7
p.m. on 1 April, 2022, through 7 p.m.
on 31 May, 2023. This temporary rule
will suspend the current regulatory cite
regarding the Morrison Bridge, and add
a temporary 33 CFR 117.897(c)(3)(vi)
which will amend the operating
schedule of the Morrison Bridge by
requiring a two-hour notice, or fourhour notice with tug assist, for all draw
openings, and alternate the operation of
the double bascule spans to single span
which will reduce the horizontal
clearances of the bridge. The temporary
rule is necessary to accommodate
preservation and painting of the
Morrison Bridge. This bridge provides a
vertical clearance approximately 69 feet,
at the center, above Columbia River
Datum 0.0 when in the closed-tonavigation position. One half of the
bascule bridge will have a containment
system installed on the non-opening
half of the span, which will reduce the
vertical clearance by 5 feet to 64 feet
center and 43 feet on the sides. A tug
will be available for assists to mariners
as needed when a request is given with
a notice of four hours for an opening.
The horizontal clearance with a full
opening is 185 feet, therefore, in single
leaf operations, a temporary rule change
will reduced the horizontal clearance to
approximately 90 feet. Vessels able to
transit under the Morrison Bridge
without an opening may do so at any
time. Marine vessels are advised to be
aware of fall hazards. This section of the
Willamette River has no alternate
routes. During the Portland Rose
Festival, both leafs of the double bascule
span will be fully operational. If any
mariner submits a full opening request
to the County prior to construction
beginning, a full opening can be
scheduled. All marine emergency
vessels can navigate under the Morrison
Bridge without an opening, and
therefore do not need to contact the
Hawthorne Bridge for an emergency
opening.
This regulatory action determination
is based on the ability of the Morrison
Bridge to open on signal after the
Hawthorne Bridge, at Willamette River
mile 13.1, has received at least a twohour notice, or four-hour notice for tug
assist, by telephone at 503–988–3452 or
VHF radio request. The Coast Guard has
made this finding based on the fact that
the proposed change allows any vessel
needing a drawbridge opening to transit
through the Morrison Bridge after
providing adequate notice and being
provided with tug assistance if required.
E:\FR\FM\24NOP1.SGM
24NOP1
Agencies
[Federal Register Volume 86, Number 224 (Wednesday, November 24, 2021)]
[Proposed Rules]
[Pages 66982-66988]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2021-25372]
========================================================================
Proposed Rules
Federal Register
________________________________________________________________________
This section of the FEDERAL REGISTER contains notices to the public of
the proposed issuance of rules and regulations. The purpose of these
notices is to give interested persons an opportunity to participate in
the rule making prior to the adoption of the final rules.
========================================================================
Federal Register / Vol. 86, No. 224 / Wednesday, November 24, 2021 /
Proposed Rules
[[Page 66982]]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 866
[Docket No. FDA-2020-N-2297]
Microbiology Devices; Reclassification of Human Immunodeficiency
Virus Viral Load Monitoring Tests
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed amendment; proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or the Agency) is
proposing to reclassify human immunodeficiency virus (HIV) viral load
monitoring tests, a postamendments class III device with the product
code MZF, into class II (special controls), subject to premarket
notification. FDA is also proposing a new device classification
regulation along with special controls that the Agency believes are
necessary to provide a reasonable assurance of safety and effectiveness
for this device type. FDA is proposing this reclassification on its own
initiative. If finalized, this order will reclassify this device type
from class III (premarket approval) to class II (special controls) and
reduce the regulatory burdens associated with these devices because
manufacturers will no longer be required to submit a premarket approval
application (PMA) for this device type but can instead submit a less
burdensome premarket notification (510(k)) and receive clearance before
marketing their device.
DATES: Submit either electronic or written comments on the proposed
order by January 24, 2022. See section XI of this document for the
proposed effective date of any final order based on this proposed
order.
ADDRESSES: You may submit comments as follows. Please note that late,
untimely filed comments will not be considered. Electronic comments
must be submitted on or before January 24, 2022. The https://www.regulations.gov electronic filing system will accept comments until
11:59 p.m. Eastern Time at the end of January 24, 2022. Comments
received by mail/hand delivery/courier (for written/paper submissions)
will be considered timely if they are postmarked or the delivery
service acceptance receipt is on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand Delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2020-N-2297 for ``Microbiology Devices; Reclassification of Human
Immunodeficiency Virus Viral Load Monitoring Tests.'' Received
comments, those filed in a timely manner (see ADDRESSES), will be
placed in the docket and, except for those submitted as ``Confidential
Submissions,'' publicly viewable at https://www.regulations.gov or at
the Dockets Management Staff between 9 a.m. and 4 p.m., Monday through
Friday, 240-402-7500.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.govinfo.gov/content/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Dockets Management Staff, 5630 Fishers Lane,
Rm. 1061, Rockville, MD 20852, 240-402-7500.
FOR FURTHER INFORMATION CONTACT: Myrna Hanna, Center for Biologics
Evaluation and Review, Food and Drug Administration, 10903 New
Hampshire
[[Page 66983]]
Ave., Bldg. 66, Rm. 5513, Silver Spring, MD 20993-0002, 301-796-5739.
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended by
the Medical Device Amendments of 1976 (the 1976 amendments) (Pub. L.
94-295), the Safe Medical Devices Act of 1990 (Pub. L. 101-629), the
Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-
115), the Medical Device User Fee and Modernization Act of 2002 (Pub.
L. 107-250), the Medical Devices Technical Corrections Act (Pub. L.
108-214), the Food and Drug Administration Amendments Act of 2007 (Pub.
L. 110-85), and the Food and Drug Administration Safety and Innovation
Act (Pub. L. 112-144), among other amendments, establishes a
comprehensive system for the regulation of medical devices intended for
human use. Section 513 of the FD&C Act (21 U.S.C. 360c) established
three categories (classes) of devices, reflecting the regulatory
controls needed to provide reasonable assurance of their safety and
effectiveness. The three categories of devices are class I (general
controls), class II (general controls and special controls), and class
III (general controls and premarket approval).
Section 513(a)(1) of the FD&C Act defines the three classes of
devices. Class I devices are those devices for which the general
controls of the FD&C Act (controls authorized by or under sections 501,
502, 510, 516, 518, 519, or 520 (21 U.S.C. 351, 352, 360, 360f, 360h,
360i, or 360j) or any combination of such sections) are sufficient to
provide reasonable assurance of safety and effectiveness; or those
devices for which insufficient information exists to determine that
general controls are sufficient to provide reasonable assurance of
safety and effectiveness or to establish special controls to provide
such assurance, but because the devices are not purported or
represented to be for a use in supporting or sustaining human life or
for a use which is of substantial importance in preventing impairment
of human health, and do not present a potential unreasonable risk of
illness or injury, are to be regulated by general controls (section
513(a)(1)(A) of the FD&C Act). Class II devices are those devices for
which general controls by themselves are insufficient to provide
reasonable assurance of safety and effectiveness and for which there is
sufficient information to establish special controls to provide such
assurance, including the promulgation of performance standards,
postmarket surveillance, patient registries, development and
dissemination of guidelines, recommendations, and other appropriate
actions the Agency deems necessary to provide such assurance (section
513(a)(1)(B) of the FD&C Act). Class III devices are those devices for
which insufficient information exists to determine that general
controls and special controls would provide a reasonable assurance of
safety and effectiveness, and are purported or represented to be for a
use in supporting or sustaining human life or for a use which is of
substantial importance in preventing impairment of human health, or
present a potential unreasonable risk of illness or injury (section
513(a)(1)(C) of the FD&C Act).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices) are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until, (1) FDA
reclassifies the device into class I or class II, or (2) FDA issues an
order finding the device to be substantially equivalent, in accordance
with section 513(i) of the FD&C Act, to a predicate device that does
not require premarket approval. FDA determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act and part 807
(21 CFR part 807), subpart E, of the regulations.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA, acting by administrative
order, can reclassify the device into class I or class II on its own
initiative, or in response to a petition from the manufacturer or
importer of the device. To change the classification of the device, the
proposed new class must have sufficient regulatory controls to provide
a reasonable assurance of the safety and effectiveness of the device
for its intended use.
FDA relies upon ``valid scientific evidence,'' as defined in
section 513(a)(3) and 21 CFR 860.7(c)(2), in the classification process
to determine the level of regulation for devices. To be considered in
the reclassification process, the ``valid scientific evidence'' upon
which the Agency relies must be publicly available (see section 520(c)
of the FD&C Act). Publicly available information excludes trade secret
and/or confidential commercial information, e.g., the contents of a
pending PMA (see section 520(c) of the FD&C Act).
In accordance with section 513(f)(3) of the FD&C Act, the Agency is
proposing to reclassify HIV viral load monitoring tests, a
postamendments class III device, into class II (special controls),
subject to premarket notification because the Agency believes the
standard in 513(a)(1)(B) of the FD&C Act is met because there is
sufficient information to establish special controls, which, in
addition to general controls, will provide reasonable assurance of the
safety and effectiveness of the device.\1\
---------------------------------------------------------------------------
\1\ In December 2019, FDA began adding the term ``Proposed
amendment'' to the ``ACTION'' caption for these documents, typically
styled ``Proposed order'', to indicate that they ``propose to
amend'' the Code of Federal Regulations. This editorial change was
made in accordance with the Office of Federal Register's (OFR)
interpretations of the Federal Register Act (44 U.S.C. chapter 15),
its implementing regulations (1 CFR 5.9 and parts 21 and 22), and
the Document Drafting Handbook.
---------------------------------------------------------------------------
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the 510(k) premarket notification requirements under
section 510(k) of the FD&C Act if the Agency determines that premarket
notification is not necessary to reasonably assure the safety and
effectiveness of the device. FDA has determined that premarket
notification is necessary to reasonably assure the safety and
effectiveness of HIV viral load monitoring tests. Therefore, the Agency
does not intend to exempt this proposed class II device from premarket
notification (510(k)) submission under section 510(m) of the FD&C Act.
II. Regulatory History of the Devices
This proposed order addresses HIV viral load monitoring tests.
These are prescription tests that measure HIV RNA as an aid in
monitoring patient status and are assigned product code MZF.\2\ These
postamendments devices are currently regulated as class III devices
under section 513(f)(1) of the FD&C Act. Based on our review experience
and consistent with the FD&C Act and FDA's regulations in 21 CFR
860.134, FDA believes that these devices should be reclassified from
class III into class II with special controls because special controls,
in addition to general controls, are necessary and sufficient to
provide reasonable assurance of the safety and effectiveness of these
devices and there is sufficient
[[Page 66984]]
information to establish special controls to provide such assurance.
---------------------------------------------------------------------------
\2\ On February 21, 2020, FDA published a separate proposed
order to reclassify certain HIV serological diagnostic and
supplemental tests and HIV nucleic acid diagnostic and supplemental
tests, which are also currently assigned product code MZF, from
class III into class II (85 FR 10110).
---------------------------------------------------------------------------
FDA approved the first in vitro nucleic acid amplification-based
HIV viral load test on June 3, 1996, for the quantitation of HIV-1 RNA
in human plasma. Currently, there are six HIV viral load monitoring
tests on the market, all of which met the performance criteria
specified in the proposed special controls, considered necessary for
the intended use of the test, when they were approved by FDA (Ref. 1).
A review of the FDA's medical device reporting database, MAUDE
(Manufacturer and User Facility Device Experience), indicates a low
number of reported events for HIV viral load monitoring tests. Millions
of devices intended for use as HIV viral load monitoring tests have
been sold since 1996 with fewer than 200 reported adverse events as of
October 2020. Of these events, fewer than 10 are reported to involve
injuries due to incorrect results; the remainder are malfunctions, such
as user errors or incorrect results, that had no reported effect on the
individual being monitored. There has been one class II recall and no
class I (highest risk) recalls specific to HIV viral load monitoring
tests, which, coupled with the low number of reported events, indicates
a good safety record for this device class.
III. Device Description
This proposed order applies to HIV viral load monitoring tests that
are prescription in vitro diagnostic devices for monitoring of HIV
viral load in body fluids. As such, these prescription devices must
satisfy prescription labeling requirements for in vitro diagnostic
products (see 21 CFR 809.10(a)(4) and (b)(5)(ii)). HIV viral load
monitoring tests are intended for use in the clinical management of
individuals living with HIV and are for professional use only. These
devices are not intended for use as an aid in diagnosis or for
screening donors of blood or blood products or human cells, tissues, or
cellular and tissue-based products (HCT/Ps).
HIV viral load monitoring tests are quantitative in vitro
diagnostic tests that measure the amount of HIV RNA in human body
fluids such as plasma and whole blood. The HIV RNA is isolated,
amplified, and detected by labeled probes that produce a quantitative
output that determines the amount of HIV in the sample. The test
results then are used as part of patient management decisions in
conjunction with other relevant clinical and laboratory findings.
Approval of HIV viral load monitoring tests has been based on
studies and established clinical decision points that demonstrate that
changes in viral load correlate with clinically meaningful outcomes,
meaning that HIV RNA measurements could reliably assess the success or
failure of antiretroviral therapy (ART) (Ref. 1).
IV. Proposed Reclassification
FDA is proposing to reclassify HIV viral load monitoring tests. At
a public meeting held on July 19, 2018, the Blood Products Advisory
Committee, convened as a medical device panel (``the Panel''),
unanimously agreed that special controls, in addition to general
controls, are sufficient to mitigate the risk to health from incorrect
results from HIV nucleic acid and serological diagnostic and
supplemental tests. The Panel believed that class II with the special
controls would provide reasonable assurance of the safety and
effectiveness of those devices. In February 2020, FDA issued a proposed
order that, if finalized, would reclassify those devices from class III
into class II (85 FR 10110). As part of the Panel's discussion, the
Panel also recommended that FDA consider reclassification of
quantitative HIV tests indicated for use for monitoring HIV viral load
from class III to class II (Ref. 2).
In accordance with section 513(f)(3) of the FD&C Act and 21 CFR
part 860, subpart C, FDA is proposing to reclassify postamendments HIV
viral load monitoring tests from class III into class II. FDA believes
that there are sufficient data and information available through FDA's
accumulated experience with these devices and from published literature
to demonstrate that the proposed special controls, along with general
controls, would effectively mitigate the risks to health identified in
section V of this document and provide reasonable assurance of safety
and effectiveness of these devices. Absent the special controls
identified in this proposed order, general controls applicable to the
device are insufficient to provide reasonable assurance of the safety
and effectiveness.
FDA is proposing to create a separate classification regulation for
HIV viral load monitoring tests. Under this proposed order, if
finalized, HIV viral load monitoring tests will be reclassified from
class III to class II and identified as prescription devices. In this
proposed order the Agency has proposed the special controls under
section 513(a)(1)(B) of the FD&C Act that, together with general
controls, would provide reasonable assurance of the safety and
effectiveness of HIV viral load monitoring tests.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C if FDA determines that premarket notification is not
necessary to provide reasonable assurance of the safety and
effectiveness of the device. For these HIV viral load monitoring tests,
FDA has determined that premarket notification is necessary to provide
a reasonable assurance of safety and effectiveness. Therefore, FDA does
not intend to exempt this proposed class II device from the 510(k)
requirements. If this proposed order is finalized, persons who intend
to market this type of device must submit to FDA a 510(k) and receive
clearance prior to marketing the device.
This proposed order, if finalized, will decrease regulatory burden
on industry because manufacturers will no longer have to submit a PMA
for this device type, but can instead submit a 510(k) to the Agency for
review prior to marketing their device. A 510(k) is a less-burdensome
pathway to market a device, which typically results in a shorter
premarket review timeline compared with a PMA, helping to provide more
timely access to this device type to patients. FDA expects that the
reclassification of these devices would enable more manufacturers to
develop HIV viral load monitoring tests such that patients would
benefit from increased access to safe and effective tests.
V. Risks to Health
Viral load is one of the important markers for monitoring the
effectiveness of ART and disease progression. The Department of Health
and Human Services (HHS) in 2014 issued guidelines on the treatment of
HIV in adults and adolescents in United States. The guidelines are
updated periodically based on new data. Regarding viral load
monitoring, the HHS guidelines define optimal viral load suppression as
suppressing viral load levels persistently to <20 to 75 copies per
milliliter (/mL) of HIV RNA, depending on the assay used (Ref. 3).
Virologic failure, at which point changes in treatment are considered,
is defined as the inability to achieve or maintain suppression of viral
replication to an HIV RNA level <200 copies/mL (Ref. 3). The HHS
guidelines recommend that viral load testing should be performed for
all patients with HIV at entry into care, initiation of therapy, and on
a regular basis thereafter (Ref. 3). Therefore, improving access to HIV
viral
[[Page 66985]]
load monitoring tests is a public health priority. After considering
FDA's accumulated experience with these devices from review submissions
and the published literature, FDA has identified the following probable
risks to health associated with HIV viral load monitoring tests:
(1) An inaccurately low viral load test result may influence
patient management decisions, such as continuation of ineffective
treatment, which can lead to serious injury including death.
Inaccurately low viral load test results also may contribute to public
health risk by leading to inadvertent transmission of the virus by an
individual living with HIV. Factors that may cause decreased test
sensitivity and/or increased risk of inaccurately low viral load test
results include, but are not limited to, viral strain variability,
acquisition of de novo mutations in genomic regions of HIV targeted by
the device, and the presence of interfering substances in the sample.
Inaccurately low results also can be caused by improper sample
collection or sample handling, loss of sensitivity of the device,
failure of detection reagents, and failure of instruments.
(2) An inaccurately high viral load test result may contribute to
an unnecessary change in therapy, potentially ending effective
treatment and leading to less effective management of disease, as well
as significant emotional stress. Factors that may cause an increased
rate of inaccurately high viral load test results include, but are not
limited to, cross-reactivity with other substances in the sample,
carryover, viral strain variability, or acquisition of de novo
mutations in genes other than HIV. Inaccurately high results also can
be caused by improper sample collection and sample contamination.
(3) Incorrect interpretation of test results by healthcare
professionals may result in patient management decisions, such as
continuation of ineffective therapy or an unnecessary change in
therapy, that could lead to serious injury, including death, and less
effective management of the disease. Incorrect interpretation of
results may be caused by inadequate labeling, including insufficient
limitations, warnings, and explanations of test procedure.
VI. Summary of the Reasons for Reclassification
FDA believes that HIV viral load monitoring tests should be
reclassified from class III into class II (special controls) because
special controls, in addition to general controls, can be established
to mitigate the risks to health identified in section V and provide
reasonable assurance of the safety and effectiveness of this device
type. The proposed special controls are identified by FDA in section
VII of this proposed order. FDA's reasons for reclassification are as
follows:
(1) There is substantial scientific and medical information
available regarding the nature and complexity of, and risks associated
with, HIV viral load monitoring tests (Refs. 3 to 11). The safety and
effectiveness of this device type has become well-established by the
performance of the approved HIV viral load tests (Ref. 1).
(2) Risks associated with the failure of the device to perform as
indicated (e.g., inaccurately high or low test results) and risks
associated with incorrect interpretation of results can be mitigated
through a combination of special controls, including performance
criteria, certain labeling requirements, and submission of certain
manufacturing information. Performance criteria would consist primarily
of analytical and method comparison study design specifications and
acceptance criteria that are based on public information regarding the
performance and validation of previously approved devices. FDA expects
that a device would demonstrate acceptable performance, e.g.,
analytical sensitivity, at clinically relevant medical decision points
at the time of clearance. This would help ensure that devices meet or
exceed the performance of other cleared or approved HIV viral load
tests at existing clinically relevant medical decision points and, in
the future, demonstrate similar performance if there are changes in
those medical decision points that reflect additional evidence and/or
medical advances. Examples of labeling mitigations include appropriate
limitations, including that results should be interpreted in
conjunction with the individual's clinical presentation, history, and
other laboratory results. These are necessary to ensure that the
devices are used correctly, and the results are interpreted
appropriately, given the diversity of settings in which these devices
are intended to be used. Manufacturing information required to be
submitted would include summaries of strategies to quantitate new HIV
types, subtypes, genotypes, and mutations to ensure the tests continue
to monitor clinically relevant forms of HIV. It also would include a
detailed device description, including information on number and design
of primers and probes, which should be designed according to current
best practices and professional recommendations. It would also include
appropriate and acceptable procedures to determine the severity of
events to ensure appropriate adverse event reporting, protocols for
assessing stability, and evaluation of test performance at the extremes
of specifications to ensure the tests have been validated to function
correctly under diverse conditions.
Taking into account the established health benefits of the use of
the device and the nature of the probable risks of the device (Refs. 1,
3 to 11), FDA, on its own initiative, is proposing to reclassify these
postamendments devices from class III into class II. FDA believes that,
when used as indicated, HIV viral load monitoring tests can provide
significant benefits to clinicians and patients.
VII. Proposed Special Controls
FDA believes that these devices can be classified into class II
with the establishment of special controls. FDA believes that these
special controls, in addition to general controls, will provide a
reasonable assurance of the safety and efficacy of these devices. Table
1 demonstrates how these proposed special controls will mitigate each
of the identified risks to health in section V.
[[Page 66986]]
Table 1--Risks to Health and Mitigation Measures for HIV Viral Load
Monitoring Tests
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
An inaccurately low test result Certain labeling limitations,
may influence patient management warnings, and explanations of the
decisions, including continuation procedures and interpretation
of ineffective antiviral therapy results.
which can lead to serious injury Analytical sensitivity and method
including death. An inaccurately comparison study performance
low test result may contribute to criteria.
public health risk by leading to Acceptable strategies for monitoring
inadvertent transmission of the emergence of and ability of the
virus by a person living with HIV. test to detect new or altered
circulating forms of HIV.
Certain other device verification
and validation information,
including acceptable processes for
risk analysis, testing performance
at extremes of specifications, and
determining severity of adverse
events and malfunctions.
An inaccurately high test result Labeling instructions for
may contribute to unnecessary appropriate confirmation of
change in therapy, potentially elevated results.
disrupting effective treatment Analytical performance criteria.
and leading to less effective Acceptable validation of
management of disease, as well as susceptibility to interference and
significant emotional stress. cross-reactivity.
Acceptable processes for risk
analysis, testing performance at
extremes of specifications, and
determining severity of adverse
events and malfunctions.
Incorrect interpretation of test Certain labeling limitations,
results may result in patient warnings, and explanations of the
management decisions, such as procedures and interpretation
continuation of ineffective results.
therapy or an unnecessary change
in therapy, that could lead to
serious injury, including death,
and less effective management of
the disease.
------------------------------------------------------------------------
If this proposed order is finalized, HIV viral load monitoring
tests will be reclassified into class II (special controls). As
discussed below, the reclassification will be codified in 21 CFR
866.3958. Firms submitting a 510(k) for an HIV viral load monitoring
test will be required to comply with the particular mitigation measures
set forth in the special controls. Adherence to the special controls,
in addition to the general controls, is necessary to provide a
reasonable assurance of the safety and effectiveness of the devices.
VIII. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IX. Paperwork Reduction Act of 1995
FDA tentatively concludes that this proposed order contains no new
collection of information. Therefore, clearance by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(PRA) (44 U.S.C. 3501-3521) is not required. This proposed order refers
to previously approved FDA collections of information. These
collections of information are subject to review by the OMB under the
PRA. The collections of information in part 807, subpart E, regarding
premarket notification submissions have been approved under OMB control
number 0910-0120; the collections of information in 21 CFR part 820
have been approved under OMB control number 0910-0073; the collections
of information in 21 CFR part 803 have been approved under OMB control
number 0910-0437; and the collections of information in 21 CFR parts
801 and 809 have been approved under OMB control number 0910-0485.
X. Codification of Orders
Under section 513(f)(3) of the FD&C Act, FDA may issue final orders
to reclassify devices. FDA will continue to codify classifications and
reclassifications in the Code of Federal Regulations (CFR). Changes
resulting from final orders will appear in the CFR as newly codified
orders. Therefore, under section 513(f)(3), in the proposed order, we
are proposing to codify HIV viral load monitoring tests in the new 21
CFR 866.3958, under which HIV viral load monitoring tests would be
reclassified from class III to class II.
XI. Proposed Effective Date
FDA proposes that any final order based on this proposed order
become effective 30 days after its date of publication in the Federal
Register.
XII. References
The following references are on display at the Dockets Management
Staff (see ADDRESSES) and are available for viewing by interested
persons between 9 a.m. and 4 p.m., Monday through Friday; they are also
available electronically at https://www.regulations.gov. FDA has
verified the website addresses, as of the date this document publishes
in the Federal Register, but websites are subject to change over time.
1. List of approved HIV viral load monitoring tests with supporting
information can be found at https://www.fda.gov/vaccines-blood-biologics/approved-blood-products/premarket-approvals-and-humanitarian-device-exemptions-supporting-documents.
2. ``Reclassification of HIV Point of Care and Laboratory-based
serological and NAT diagnostic devices from Class III (PMA) to Class
II 510(k); Issue Summary Prepared for the July 19, 2018, Meeting of
the Blood Products Advisory Committee (BPAC)).'' Available at:
https://www.fda.gov/advisory-committees/blood-products-advisory-committee/2018-meeting-materials-blood-products-advisory-committee.
3. ``Guidelines for the Use of Antiretroviral Agents in Adults and
Adolescents Living with HIV.'' Department of Health and Human
Services. Accessed November 24, 2020. Available at: https://clinicalinfo.hiv.gov/en/guidelines/adult-and-adolescent-arv/antiretroviral-therapy-prevent-sexual-transmission-hiv.
4. ``Human Immunodeficiency Virus-1 Infection: Developing
Antiretroviral Drugs for Treatment; Guidance for Industry.''
Available at: https://www.fda.gov/media/86284/download.
5. Aberg, J.A., J.E. Gallant, K.H. Ghanem, et al., ``Primary Care
Guidelines for the Management of Persons Infected with HIV: 2013
Update by the HIV Medicine Association of the Infectious Diseases
Society of America,'' Clinical Infectious Disease, 58:e1-34, 2014.
6. Saag, M.S., M. Holodniy, D.R. Kuritzkes, et al., ``HIV Viral Load
Markers in Clinical Practice,'' Nature Medicine, 2:625-629, 1996.
7. Das, M., P.L. Chu, G-M. Santos, et al., ``Decreases in Community
Viral Load are Accompanied by Reductions in New HIV Infections in
San Francisco,'' PLoS ONE, 5:e11068, 2010.
8. Stadhouders, R., S.D. Pas, J. Anber, et al., ``The Effect of
Primer-Template Mismatches on the Detection and Quantification of
Nucleic Acids Using the 5' Nuclease Assay,'' Journal of Molecular
Diagnostics, 12:109-117, 2010.
9. Swenson, L.C., B. Cobb, A.M. Geretti, et al., ``Comparative
Performances of HIV-
[[Page 66987]]
1 RNA Load Assays at Low Viral Load Levels: Results of an
International Collaboration,'' Journal of Clinical Microbiology,
52(2):517-523, 2014.
10. Caniglia, E.C., C. Sabin, J.M. Robins, et al., ``When to Monitor
CD4 Cell Count and HIV RNA to Reduce Mortality and AIDS-Defining
Illness in Virologically Suppressed HIV-Positive Persons on
Antiretroviral Therapy in High-Income Countries: A Prospective
Observational Study,'' Journal of Acquired Immune Deficiency
Syndromes, 72:214-221, 2016.
11. Shoko, C. and D. Chikobvu, ``A Superiority of Viral Load Over
CD4 Cell Count When Predicting Mortality in HIV Patients on
Therapy.'' BioMed Central Infectious Diseases, 19:169, 2019.
List of Subjects in 21 CFR Part 866
Biologics, Laboratories, Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act it is
proposed that 21 CFR part 866 be amended as follows:
PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES
0
1. The authority citation for part 866 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 866.3958 to subpart D to read as follows:
Sec. 866.3958 Human immunodeficiency virus (HIV) viral load
monitoring test.
(a) Identification. A human immunodeficiency virus (HIV) viral load
monitoring test is an in vitro diagnostic prescription device for the
quantitation of the amount of HIV RNA in human body fluids. The test is
intended for use in the clinical management of individuals living with
HIV and is for professional use only. The test results are intended to
be interpreted in conjunction with other relevant clinical and
laboratory findings. The test is not intended to be used as an aid in
diagnosis or for screening donors of blood or blood products or human
cells, tissues, or cellular and tissue-based products (HCT/Ps).
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The labeling must include:
(i) An intended use that states that the device is not intended for
use as an aid in diagnosis or for use in screening donors of blood or
blood products, or HCT/Ps.
(ii) A detailed explanation of the principles of operation and
procedures used for assay performance.
(iii) A detailed explanation of the interpretation of results and
recommended actions to take based on current clinical guidelines.
(iv) Limitations, which must be updated to reflect current clinical
practice and patient management. The limitations must include, but are
not limited to, statements that indicate:
(A) The matrices and sample types with which the device has been
cleared and that use of this test with specimen types other than those
specifically cleared for this device may cause inaccurate test results.
(B) Mutations in highly conserved regions may affect binding of
primers and/or probes resulting in the under-quantitation of virus or
failure to detect the presence of virus.
(C) All test results should be interpreted in conjunction with the
individual's clinical presentation, history, and other laboratory
results.
(2) Device verification and validation must include:
(i) Detailed device description, including the device components,
ancillary reagents required but not provided, and an explanation of the
device methodology. Additional information appropriate to the
technology must be included, such as detailed information on the design
of primers and probes.
(ii) For devices with assay calibrators, the design and nature of
all primary, secondary, and subsequent quantitation standards used for
calibration as well as their traceability to a reference material. In
addition, analytical testing must be performed following the release of
a new lot of the standard material that was used for device clearance,
or when there is a transition to a new calibration standard.
(iii) Detailed documentation of analytical performance studies
conducted as appropriate to the technology, specimen types tested, and
intended use of the device, including but not limited to, limit of
blank, limit of detection, limit of quantitation, cutoff determination,
precision, linearity, endogenous and exogenous interferences, cross-
reactivity, carry-over, quality control, matrix equivalency, sample and
reagent stability. Samples selected for use in analytical studies or
used to prepare samples for use in analytical studies must be from
subjects with clinically relevant genotypes circulating in the United
States.
(iv) Multisite reproducibility study that includes the testing of
three independent production lots.
(v) Analytical sensitivity of the device must demonstrate
acceptable performance at current clinically relevant medical decision
points. Samples tested to demonstrate analytical sensitivity must
include appropriate numbers and types of samples, including real
clinical samples near the lower limit of quantitation and any
clinically relevant medical decision points. Analytical specificity of
the device must demonstrate acceptable performance. Samples tested to
demonstrate analytical specificity must include appropriate numbers and
types of samples from patients with different underlying illnesses and
infection and from patients with potential interfering substances.
(vi) Detailed documentation of performance from a multisite
clinical study or a multisite analytical method comparison study.
(A) For devices evaluated in a multisite clinical study, the study
must use specimens from individuals living with HIV being monitored for
changes in viral load, and the test results must be compared to the
clinical status of the patients.
(B) For tests evaluated in a multisite analytical method comparison
study, the performance of the test must be compared to an FDA-cleared
or approved comparator. The multisite method comparison study must
include appropriate numbers and types of samples with analyte
concentrations across the measuring range of the assay, representing
clinically relevant genotypes. The multisite method comparison study
design, including number of samples tested, must be sufficient to meet
the following criteria:
(1) Agreement between the two tests across the measuring range of
the assays must have an r2 of greater than or equal to 0.95.
(2) The bias between the test and comparator assay, as determined
by difference plots, must be less than or equal to 0.5 log copies/mL.
(vii) If a multisite clinical study is performed under paragraph
(b)(2)(vi) of this section, detailed documentation of a single-site
analytical method comparison study between the device and an FDA-
cleared or approved comparator. The analytical method comparison study
must use appropriate numbers and types of samples with analyte
concentrations across the measuring range of the assay, representing
clinically relevant genotypes. The results must meet the criteria in
paragraphs (b)(2)(vi)(B)(1) and (2) of this section.
(viii) Strategies for detection of new strains, types, subtypes,
genotypes, and genetic mutations as they emerge.
[[Page 66988]]
(ix) Risk analysis and management strategies, such as Failure Modes
Effects Analysis and/or Hazard Analysis and Critical Control Points
summaries and their impact on test performance.
(x) Final release criteria to be used for manufactured device lots
with an appropriate justification that lots released at the extremes of
the specifications will meet the claimed analytical and clinical
performance characteristics as well as the stability claims.
(xi) All stability protocols, including acceptance criteria.
(xii) Appropriate and acceptable procedure(s) for addressing
complaints and other device information that determines when to submit
a medical device report.
(xiii) Premarket notification submissions must include the
information contained in paragraphs (b)(2)(i) through (xii) of this
section.
Dated: November 16, 2021.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2021-25372 Filed 11-23-21; 8:45 am]
BILLING CODE 4164-01-P