Physical Medicine Devices; Reclassification of Non-Invasive Bone Growth Stimulators, 49986-49994 [2020-17543]
Download as PDF
49986
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
received, and any final disposition in
person in the Dockets Office (see the
ADDRESSES section for the address and
phone number) between 9:00 a.m. and
5:00 p.m., Monday through Friday,
except federal holidays. An informal
docket may also be examined during
normal business hours at the Northwest
Mountain Regional Office of the Federal
Aviation Administration, Air Traffic
Organization, Western Service Center,
Operations Support Group, 2200 S
216th Street, Des Moines, WA 98198.
Availability and Summary of
Documents for Incorporation by
Reference
This document proposes to amend
FAA Order 7400.11D, Airspace
Designations and Reporting Points,
dated August 8, 2019, and effective
September 15, 2019. FAA Order
7400.11D is publicly available as listed
in the ADDRESSES section of this
document. FAA Order 7400.11D lists
Class A, B, C, D, and E airspace areas,
air traffic service routes, and reporting
points.
The Proposal
The FAA is proposing an amendment
to Title 14 Code of Federal Regulations
(14 CFR) Part 71 by establishing Class E
airspace, extending upward from 700
feet above the surface, at Benton Field
Airport. This area is designed to contain
IFR departures to 1,200 feet above the
surface and IFR arrivals descending
below 1,500 feet above the surface. This
airspace area would be described as
follows: That airspace extending
upward from 700 feet above the surface
within a 3.3-mile radius of the airport,
and within 4 miles east and 2.3 miles
west of the 002° bearing from the
airport, extending from 2.5 miles north
of the airport to 12.4 miles north of the
airport, and within 2.7 miles each side
of the 176° bearing from the airport,
extending from the 3.3-mile radius to
10.8 miles south of Benton Field
Airport.
Class E5 airspace designations are
published in paragraph 6005 of FAA
Order 7400.11D, dated August 8, 2019,
and effective September 15, 2019, which
is incorporated by reference in 14 CFR
71.1. The Class E airspace designations
listed in this document will be
published subsequently in the Order.
FAA Order 7400.11, Airspace
Designations and Reporting Points, is
published yearly and effective on
September 15.
Regulatory Notices and Analyses
The FAA has determined that this
regulation only involves an established
body of technical regulations for which
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
frequent and routine amendments are
necessary to keep them operationally
current, is non-controversial, and
unlikely to result in adverse or negative
comments. It, therefore: (1) Is not a
‘‘significant regulatory action’’ under
Executive Order 12866; (2) is not a
‘‘significant rule’’ under DOT
Regulatory Policies and Procedures (44
FR 11034; February 26, 1979); and (3)
does not warrant preparation of a
regulatory evaluation as the anticipated
impact is so minimal. Since this is a
routine matter that will only affect air
traffic procedures and air navigation, it
is certified that this rule, when
promulgated, would not have a
significant economic impact on a
substantial number of small entities
under the criteria of the Regulatory
Flexibility Act.
Environmental Review
List of Subjects in 14 CFR Part 71
Airspace, Incorporation by reference,
Navigation (air).
The Proposed Amendment
Accordingly, pursuant to the
authority delegated to me, the Federal
Aviation Administration proposes to
amend 14 CFR part 71 as follows:
PART 71—DESIGNATION OF CLASS A,
B, C, D, AND E AIRSPACE AREAS; AIR
TRAFFIC SERVICE ROUTES; AND
REPORTING POINTS
1. The authority citation for 14 CFR
part 71 continues to read as follows:
■
Authority: 49 U.S.C. 106(f), 106(g), 40103,
40113, 40120; E.O. 10854, 24 FR 9565, 3 CFR,
1959–1963 Comp., p. 389.
[Amended]
2. The incorporation by reference in
14 CFR 71.1 of FAA Order 7400.11D,
Airspace Designations and Reporting
Points, dated August 8, 2019, and
effective September 15, 2019, is
amended as follows:
■
Paragraph 6005 Class E Airspace Areas
Extending Upward From 700 Feet or More
Above the Surface of the Earth.
*
*
*
*
*
AWP CA E5 Redding, CA
Benton Field Airport, CA
(Lat. 40°34′25″ N, long. 122°24′26″ W)
That airspace extending upward from 700
feet above the surface within a 3.3-mile
radius of the airport, and within 4 miles east
PO 00000
Frm 00009
Fmt 4702
Sfmt 4702
Issued in Seattle, Washington, on August
10, 2020.
B.G. Chew,
Acting Group Manager, Operations Support
Group, Western Service Center.
[FR Doc. 2020–17875 Filed 8–14–20; 8:45 am]
BILLING CODE 4910–13–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 890
[Docket No. FDA–2020–N–1053]
This proposal will be subject to an
environmental analysis in accordance
with FAA Order 1050.1F,
‘‘Environmental Impacts: Policies and
Procedures’’ prior to any FAA final
regulatory action.
§ 71.1
and 2.3 miles west of the 002° bearing from
the airport, extending from 2.5 miles north of
the airport to 12.4 miles north of the airport,
and within 2.7 miles each side of the 176°
bearing from the airport, extending from the
3.3-mile radius to 10.8 miles south of Benton
Field Airport.
Physical Medicine Devices;
Reclassification of Non-Invasive Bone
Growth Stimulators
Food and Drug Administration,
Health and Human Services (HHS).
ACTION: Proposed amendment; proposed
order; request for comments.
AGENCY:
The Food and Drug
Administration (FDA) is proposing to
reclassify non-invasive bone growth
stimulators, postamendments class III
devices (product codes LOF and LPQ),
into class II (special controls), subject to
premarket notification. FDA is also
proposing a new device classification
with the name ‘‘non-invasive bone
growth stimulators’’ along with the
proposed special controls that the
Agency believes are necessary to
provide a reasonable assurance of safety
and effectiveness of these devices. FDA
is proposing this reclassification on its
own initiative. If finalized, this order
will reclassify these devices from class
III (premarket approval) to class II
(special controls) and reduce the
regulatory burdens associated with
these devices, as these devices will no
longer be required to submit a
premarket approval application (PMA),
but are subject to premarket notification
(510(k)) requirements and general and
special controls.
DATES: Submit either electronic or
written comments on the proposed
order by October 16, 2020. Please see
section XII of this document for the
proposed effective date when the new
requirements apply and for the
proposed effective date of a final order
based on this proposed order.
ADDRESSES: You may submit comments
as follows. Please note that late,
SUMMARY:
E:\FR\FM\17AUP1.SGM
17AUP1
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
untimely filed comments will not be
considered. Electronic comments must
be submitted on or before October 16,
2020. The https://www.regulations.gov
electronic filing system will accept
comments until 11:59 p.m. Eastern Time
at the end of October 16, 2020.
Comments received by mail/hand
delivery/courier (for written/paper
submissions) will be considered timely
if they are postmarked or the delivery
service acceptance receipt is on or
before that date.
Electronic Submissions
Submit electronic comments in the
following way:
• Federal Rulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand Delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2020–N–1053 for ‘‘Physical Medicine
Devices; Reclassification of NonInvasive Bone Growth Stimulators.’’
Received comments, those filed in a
timely manner (see ADDRESSES), will be
placed in the docket and, except for
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
those submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Dockets Management Staff between 9
a.m. and 4 p.m., Monday through
Friday.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://
www.govinfo.gov/content/pkg/FR-201509-18/pdf/2015-23389.pdf.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Jesse Muir, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4508, Silver Spring,
MD 20993, 240–402–6679, Jesse.Muir@
fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (FD&C Act), as amended, establishes
a comprehensive system for the
regulation of medical devices intended
for human use. Section 513 of the FD&C
PO 00000
Frm 00010
Fmt 4702
Sfmt 4702
49987
Act (21 U.S.C. 360c) established three
categories (classes) of devices, reflecting
the regulatory controls needed to
provide reasonable assurance of their
safety and effectiveness. The three
categories of devices are class I (general
controls), class II (general controls and
special controls), and class III (general
controls and premarket approval).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices) are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until: (1) FDA reclassifies the device
into class I or II or (2) FDA issues an
order finding the device to be
substantially equivalent, in accordance
with section 513(i) of the FD&C Act, to
a predicate device that does not require
premarket approval. FDA determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part
807 (21 CFR part 807), subpart E, of the
regulations.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or class
II under section 513(f)(3) of the FD&C
Act. Section 513(f)(3) of the FD&C Act
provides that FDA acting by
administrative order, can reclassify the
device into class I or class II on its own
initiative, or in response to a petition
from the manufacturer or importer of
the device. To change the classification
of the device, the proposed new class
must have sufficient regulatory controls
to provide reasonable assurance of the
safety and effectiveness of the device for
its intended use.
Reevaluation of the data previously
presented before the Agency is an
appropriate basis for subsequent action,
where the reevaluation is made in light
of newly available regulatory authority
(see Bell v. Goddard, 366 F.2d 177, 181
(7th Cir. 1966); Ethicon, Inc. v. FDA, 762
F. Supp. 382, 388–391 (D.D.C. 1991)) or
in light of changes in ‘‘medical science’’
(Upjohn v. Finch, 422 F.2d 944, 951 (6th
Cir. 1970)). Whether data before the
Agency are old or new, the ‘‘new
information’’ to support reclassification
under 513(f)(3) must be ‘‘valid scientific
evidence’’, as defined in section
513(a)(3) of the FD&C Act and
§ 860.7(c)(2) (21 CFR 860.7(c)(2)). (See,
e.g., General Medical Co. v. FDA, 770
F.2d 214 (D.C. Cir. 1985); Contact Lens
Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.
E:\FR\FM\17AUP1.SGM
17AUP1
49988
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
1985), cert. denied, 474 U.S. 1062
(1986).)
FDA relies upon ‘‘valid scientific
evidence,’’ as defined in section
513(a)(3) of the FD&C Act and
§ 860.7(c)(2), in the classification
process to determine the level of
regulation for devices. To be considered
in the reclassification process, the
‘‘valid scientific evidence’’ upon which
the Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA (see
section 520(c) of the FD&C Act (21
U.S.C. 360j(c)). Section 520(h)(4) of the
FD&C Act provides that FDA may use,
for reclassification of a device, certain
information in a PMA 6 years after the
application has been approved (Ref. 1).
This includes information from clinical
and preclinical tests or studies that
demonstrate the safety or effectiveness
of the device, but does not include
descriptions of methods of manufacture
or product composition and other trade
secrets.
In accordance with section 513(f)(3) of
the FD&C Act, FDA is issuing this
proposed order to reclassify noninvasive bone growth stimulator
devices, postamendments class III
devices, into class II (special controls),
subject to premarket notification
because FDA believes the standard in
section 513(a)(1)(B) of the FD&C Act is
met as there is sufficient information to
establish special controls, which in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.1
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the premarket
notification requirements under section
510(k) of the FD&C Act, if the Agency
determines that premarket notification
is not necessary to reasonably assure the
safety and effectiveness of the device.
FDA has determined that premarket
notification is necessary to reasonably
assure the safety and effectiveness of
non-invasive bone growth stimulator
devices. Therefore, the Agency does not
intend to exempt these proposed class II
devices from premarket notification
(510(k)) submission as provided under
section 510(m) of the FD&C Act.
1 In December 2019, FDA began adding the term
‘‘Proposed amendment’’ to the ‘‘ACTION’’ caption
for these documents, typically styled ‘‘Proposed
order’’, to indicate that they ‘‘propose to amend’’
the Code of Federal Regulations. This editorial
change was made in accordance with the Office of
Federal Register’s (OFR) interpretations of the
Federal Register Act (44 U.S.C. chapter 15), its
implementing regulations (1 CFR 5.9 and parts 21
and 22), and the Document Drafting Handbook.
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
II. Regulatory History of Non-Invasive
Bone Growth Stimulator Devices
In accordance with section 513(f)(1) of
the FD&C Act, non-invasive bone
growth stimulator devices were
automatically classified into class III
because they were not introduced or
delivered for introduction into interstate
commerce for commercial distribution
before May 28, 1976, and have not been
found substantially equivalent to a
device placed in commercial
distribution after May 28, 1976, which
was subsequently classified or
reclassified into class II or class I.
Therefore, the device is subject to PMA
requirements under section 515 of the
FD&C Act (21 U.S.C. 360e).
Accordingly, on November 6, 1979,
FDA approved a PMA for the Bio
Osteogen System 204 (P790002) (Ref. 2).
Since that time, five additional original
PMAs have been approved for noninvasive bone growth stimulators
(P850007, P850022, P900009, P910066,
and P030034). On February 9, 2005,
FDA received a reclassification petition
dated February 7, 2005, submitted by RS
Medical Corporation, requesting that
FDA reclassify certain non-invasive
bone growth stimulators from class III to
class II (Ref. 3). As stated in the Notice
of Panel Recommendation discussed
further below, ‘‘the petition was
submitted under section 513(e) of the
act but FDA . . . review[ed] the petition
under section 513(f)(3) of the act
because that section contain[ed] the
appropriate procedures for
reclassification of postamendments
devices’’ (72 FR 1951 at 1952, January
17, 2007). FDA requested additional
information and the petitioner amended
the petition on November 30, 2005
(‘‘amended petition’’). In accordance
with the FD&C Act and regulations,
FDA referred the petition, as amended,
to the FDA Advisory Committee,
specifically the Orthopaedic and
Rehabilitation Devices Panel (‘‘the 2006
Panel’’) for its recommendations on the
requested reclassification.
On June 2, 2006, the 2006 Panel
deliberated on the information in RS
Medical’s petition; the presentations
made by RS Medical, FDA, and
members of the public; and their own
experience with certain non-invasive
bone growth stimulators (Ref. 4).
The 2006 Panel identified the
following risks to health associated with
non-invasive bone growth stimulators:
Electric shock; burn; skin irritation and/
or allergic reaction; inconsistent or
ineffective treatment; adverse
interaction with electrical implants;
adverse interactions with internal/
external fixation devices; and biological
PO 00000
Frm 00011
Fmt 4702
Sfmt 4702
risks. The 2006 Panel did not
specifically address risks associated
with ultrasound-based devices, as these
were outside the scope of RS Medical’s
petition; however, as discussed below,
based upon FDA’s review of information
since the Panel meeting, the risks
identified with ultrasound-based
devices, along with their reported
benefits, are comparable to those of noninvasive bone growth stimulators
incorporating other modalities.
The majority of the 2006 Panel
recommended that non-invasive bone
growth stimulators should be retained
in class III because there was
insufficient information in the petition
by RS Medical to establish that special
controls in conjunction with general
controls would provide a reasonable
assurance of the safety and effectiveness
of the device. Specifically, the Panel
recommended that the proposed special
controls by RS Medical were sufficient
to control for the risk of electric shock,
burn, skin irritation, and/or allergic
reaction; adverse interaction with
electrical implants; adverse interactions
with internal/external fixation devices;
and biological risks. However, the Panel
believed that there was insufficient
evidence presented by RS Medical to
control for the risk of inconsistent or
ineffective treatment because there is a
lack of knowledge about how waveform
characteristics (e.g., pulse duration,
amplitude, power, frequency), including
potential modifications to the device,
affect the clinical response to treatment.
The Panel requested additional clinical
data and/or special controls, which was
not adequately devised by the
petitioner, to control for the risk of
inconsistent or ineffective treatment.
FDA concurred with the 2006 Panel’s
recommendation, and similarly believed
that RS Medical’s petition was
inadequate in that FDA had concerns
about the petitioner’s proposed special
controls to control the risk of
inconsistent or ineffective treatment. In
the Federal Register of January 17, 2007
(72 FR 1951), FDA published a Notice
of Panel Recommendations (‘‘the 2007
Notice’’), as referenced above.
In a letter dated April 2, 2007, RS
Medical requested that its petition be
withdrawn (Ref. 5). On July 10, 2007,
FDA granted RS Medical’s request for
withdrawal of the petition and did not
take any further action on the petition
(Ref. 6). FDA has not received any
subsequent petition requesting
reclassification of these devices.
Subsequently, as part of the Center for
Devices and Radiological Health’s 2014–
2015 strategic priority, ‘‘Strike the Right
Balance Between Premarket and
Postmarket Data Collection,’’ a
E:\FR\FM\17AUP1.SGM
17AUP1
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
retrospective review of all PMA product
codes with active PMAs approved prior
to 2010 was conducted to determine
whether, based on our current
understanding of the technology, certain
devices could be reclassified (downclassified). On April 29, 2015, FDA
published a document in the Federal
Register identifying certain product
codes as potential candidates for
reclassification (80 FR 23798), including
non-invasive bone growth stimulators
under product codes LOF and LPQ,
from class III to class II (Ref. 7). One
comment was received in response to
this proposal for reclassification of LOF
and LPQ; this comment did not support
FDA’s intention to reclassify these
devices, citing the concerns discussed
during the 2006 Panel. This comment
was considered in development of this
proposed order. Note that invasive bone
growth stimulators, designated under
product code LOE, are outside the scope
of this proposed reclassification. As
noted in the 2006 Panel, invasive bone
growth stimulator devices have added
risks compared to non-invasive bone
growth stimulators, and therefore would
require a separate classification
discussion. Furthermore, invasive bone
growth stimulators were also considered
as a part of the aforementioned PMA
retrospective review and FDA
determined that these devices should
remain as class III (Ref. 8). Therefore,
FDA will continue to regulate invasive
bone growth stimulators as a class III
device, subject to PMA requirements.
While RS Medical’s petition
inadequately addressed all of the risks
associated with non-invasive bone
growth stimulators for reclassification,
FDA is, on its own initiative, proposing
to reclassify these devices from class III
to class II, and believes that sufficient
information exists to establish special
controls, as identified in this proposed
order, that, together with general
controls, can provide a reasonable
assurance of safety and effectiveness for
this device type. Additionally, RS
Medical in its petition excluded use of
these devices as an adjunct to cervical
fusion surgery in patients at high risk
for nonfusion, as well as for use in
congenital pseudarthrosis. Based upon
the review of the evidence and FDA’s
ability to establish special controls, FDA
believes these indications that have
been approved for currently marketed
non-invasive bone growth stimulator
devices should be included in this
proposed reclassification.
III. Device Description
Non-invasive bone growth simulators,
currently designated under product
codes LOF and LPQ, are typically
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
composed of a waveform generator and
transducer (e.g., coils, electrodes, and/or
ultrasound transducers). Patientcontacting surfaces include the
transducers, lead wires, and the device
outer casing.
Non-invasive bone growth stimulators
utilize an electrical component to
produce an output electrical, magnetic,
or ultrasonic waveform that is delivered
to a treatment site via a non-invasively
applied transducer (e.g., electromagnetic
coil or ultrasound transducer) or
electrodes (e.g., capacitor plates). The
device also incorporates an internal
means to monitor the output waveform
and delivery of treatment, and to
provide visual and/or audible alarms to
alert the user of improper device
function. The induced electrical and/or
magnetic fields are generated using one
of the following modalities:
• Capacitive coupling (CC), in which
a pair of electrodes are placed on the
skin such that a current can be driven
across the target site;
• pulsed electromagnetic fields
(PEMF), in which a modulated
electromagnetic field is generated near
the treatment site though an external
coil; or
• combined magnetic fields (CMF), in
which a coil generates a combination of
a static and pulsed magnetic field near
the treatment site.
The ultrasonic waveform is generated
using:
• Low intensity pulsed ultrasound
(LIPUS), in which pulsed ultrasonic
signals are generated using ultrasonic
transducers.
The non-invasive nature of the device
obviates the need for sterile
components; however, patientcontacting surfaces should be capable of
being cleaned as needed and
biocompatibility must be ensured.
Non-invasive bone growth stimulators
are generally intended to promote
osteogenesis as adjunct to primary
treatments for fracture fixation or spinal
fusion. The indications for use for this
device type depend on the specific
device characteristics, but have
included:
• Treatment of an established nonunion secondary to trauma of the
appendicular system,
• treatment of congenital
pseudarthrosis,
• treatment of failed fusions of the
appendicular system,
• early treatment of certain fresh
fractures, and
• as an adjunct to lumbar or cervical
spinal fusion.
In addition, non-invasive bone growth
stimulators are currently prescription
PO 00000
Frm 00012
Fmt 4702
Sfmt 4702
49989
use only devices under § 801.109 (21
CFR 801.109).
IV. Proposed Reclassification
In accordance with section 513(f)(3) of
the FD&C Act and 21 CFR part 860,
subpart C, FDA is proposing to
reclassify non-invasive bone growth
stimulator devices under product codes
LOF and LPQ from class III into class II.
This includes devices that generate
electrical or magnetic fields using CC,
PEMF, and CMF, and ultrasonic signals.
FDA believes that there is sufficient
information available by way of FDA’s
accumulated experience with these
devices from review of premarket
submissions, peer-reviewed literature,
medical device reports (MDRs), and
recalls to understand the risks
associated with these devices to
establish special controls that effectively
mitigate the risks to health identified in
section V. In this proposed order, the
Agency has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, would provide a reasonable
assurance of the safety and effectiveness
for non-invasive bone growth
stimulators to be in class II. Absent the
special controls identified in this
proposed order, general controls
applicable to this device type are
insufficient to provide reasonable
assurance of safety and effectiveness of
the device.
FDA is proposing to create a
classification regulation for noninvasive bone growth stimulators,
which would include devices
designated under product codes LOF
and LPQ. Under this proposed order, if
finalized, a non-invasive bone growth
stimulator will be identified as a
prescription device. As such, the
prescription device must satisfy
prescription labeling requirements (see
§ 801.109, Prescription devices).
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of § 801.109 are met. In
this proposed order, if finalized, the
Agency has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, will provide a reasonable
assurance of the safety and effectiveness
for non-invasive bone growth stimulator
devices.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act if FDA determines that
premarket notification is not necessary
E:\FR\FM\17AUP1.SGM
17AUP1
49990
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
to provide reasonable assurance of the
safety and effectiveness of the device.
For non-invasive bone growth
stimulators, FDA has determined that
premarket notification is necessary to
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, FDA does not propose to
exempt these proposed class II devices
from 510(k) requirements. If this order
is finalized, persons who intend to
market this type of device would need
to submit to FDA a 510(k) and receive
clearance prior to marketing the device.
V. Risks to Health
Based on available information for
non-invasive bone growth stimulators,
including the 2005 reclassification
petition request from RS Medical Corp,
input from the 2006 Panel, data in PMA
applications P030034, P850022/S009,
and P910066/S011 available to FDA
under section 520(h)(4) of the FD&C
Act, published literature, and
postmarket experience associated with
use of these devices, FDA identifies the
following risks to health associated with
non-invasive bone growth stimulators:
a. Failure or delay of osteogenesis—A
patient could receive ineffective
treatment, contributing to failure or
delay of osteogenesis that may lead to
clinical symptoms (e.g., pain) and the
need for surgical interventions.
Ineffective treatment could be a result of
various circumstances (e.g., inadequate
therapeutic signal output or device
malfunction or misuse).
b. Burn—A patient or health care
professional could be burned from the
use and operation of the device. This
could be a result of various
circumstances including device
malfunction (e.g., electrical fault) or
misuse of the device (e.g., use while
sleeping).
c. Electrical shock—A patient or
health care professional could be
shocked from the use and operation of
the device. This could be a result of
various circumstances including device
malfunction (e.g., electrical fault) or
misuse of the device (e.g., use of
alternating current source during
treatment).
d. Electromagnetic interference
(EMI)—A patient with electrically
powered implants (such as cardiac
pacemakers, cardiac defibrillators, and
neurostimulators) could experience an
adverse interaction with the implanted
electrical device via EMI or
radiofrequency interference.
e. Adverse tissue reaction—A patient
could experience skin irritation and/or
allergic reaction associated with the use
and operation of the device via the use
of non-biocompatible device materials.
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
f. Adverse interaction with internal/
external fixation devices—The signal
output could be impacted by certain
metallic internal or external fixation
devices leading to inadequate treatment
signals, device malfunction, or tissue
damage.
g. Adverse biologic effects—A patient
may experience adverse biologic effects
resulting from prolonged exposure to
the treatment signal via biologic
interaction with the treatment signal at
a cellular level. Excessive energy
transmission could cause tissue damage
or aberrant tissue behavior if signal
output parameters exceed established
safety thresholds.
The risks to health identified within
this proposed order are consistent with
those identified in the 2005
reclassification petition, as amended.
The 2006 Panel agreed with these
identified risks; however, in some cases
the risk or accompanying description
was reworded for clarity in this
proposed order (e.g., ‘‘inconsistent
treatment or ineffective treatment’’ is
described in terms of risk to health,
which may entail ‘‘failure or delay in
osteogenesis’’). Also, the risk of adverse
biologic effects previously specified
risks of carcinogenicity, genotoxicity,
mutagenicity, and teratological effects.
The petitioner notes in the amended
petition that ‘‘. . . the evidence points
to lack of genotoxic, carcinogenic, and
teratologic potential of the subject
waveforms,’’ which is corroborated by
the lack of such reports identified in the
literature. Although FDA similarly has
found a lack of such reports, it considers
this risk more generally as potential
deleterious effects at the tissue or
cellular level due to signal output
parameters that exceed established
safety thresholds.
VI. Summary of Reasons for
Reclassification
FDA believes that non-invasive bone
growth stimulator devices, which are
intended to promote osteogenesis as an
adjunct to primary treatments for
fracture fixation or spinal fusion, should
be reclassified from class III to class II
and that there is sufficient information
to establish special controls for the risks
identified in section V which, in
addition to general controls, can provide
reasonable assurance of safety and
effectiveness.
Specifically, FDA proposes to require
clinical performance data as a special
control to address the risk of failure or
delay of osteogenesis. FDA review of the
literature suggests a high variability of
treatment efficacy, depending on
therapeutic signal and anatomic
location. This would also address the
PO 00000
Frm 00013
Fmt 4702
Sfmt 4702
main concern cited by the 2006 Panel
and FDA with RS Medical’s proposal,
which led to the recommendation to
retain non-invasive bone growth
stimulators in class III, and various
comments received in response to the
2007 Notice.
FDA’s proposal would require that
clinical performance of any noninvasive bone growth stimulator device
be evaluated in support of the intended
use. Rather than prescribe specific study
requirements, FDA’s proposal would
allow for flexibility in study design and
the level of clinical evidence needed by
taking into consideration certain
parameters, e.g., the intended use,
treatment population, and technological
characteristics of the device, including
any similarities between the device and
legally marketed predicate device, as
appropriate.
VII. Summary of Data Upon Which the
Reclassification Is Based
The available evidence demonstrates
that there are probable health benefits
derived from the use of these devices,
and that the nature and incidence of
risks are well known so that special
controls can be established to
adequately mitigate the risks to health.
FDA is proposing a single device class
for non-invasive bone growth
stimulators, considering that FDA did
not identify any unique risks associated
with the different modalities included
in this proposed order. FDA has
considered and analyzed the following:
Data in PMA applications P030034,
P850022/S009, and P910066/S011
available to FDA under section 520(h)(4)
of the FD&C Act; information presented
at the 2006 Panel concerning RS
Medical’s petition to down-classify
certain non-invasive bone growth
stimulators (Ref. 3) and the 2007 Notice;
peer-reviewed articles that discussed
the use of, as well as the probable
benefits and risks of these devices;
reported adverse events identified
through a search of FDA’s Medical
Device Reporting (MDR) system; and a
review of any recalls associated with
these devices through a search of FDA’s
Medical Device Recall database.
In accordance with the ‘‘6-year rule’’
described in section 520(h)(4) of the
FD&C Act, FDA considered data
contained in three original PMAs or
supplements, P030034, P850022/S009,
and P910066/S011, approved for noninvasive bone growth stimulators (Refs.
9 to 11). These PMAs/supplements
include three different device
modalities: A PEMF device (P030034), a
CC device (P850022/S009), and a CMF
device (P910066/S011). In review of the
reported clinical data in the summary of
E:\FR\FM\17AUP1.SGM
17AUP1
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
safety and effectiveness data documents
(SSEDs), the studies conducted in
support of these devices include a total
study size of 831 enrolled subjects. The
adverse event profile for the devices in
each study were similar to the control
group, with a similar distribution of
event types. With regards to benefit, the
clinical data reported in the SSEDs
demonstrate an improved rate of bone
fusion compared to placebo controls,
with an 83.6 percent vs. 68.6 percent
fusion rate at 6 months in P030034
(cervical spine), an 85 percent vs. 75
percent clinical success shown in
P850022/S009 (lumbar spine), and a 67
percent vs. 43 percent fusion rate at 9
months in P910066/S011 (lumbar
spine).
Further, FDA performed a literature
review to evaluate data related to noninvasive bone growth stimulator
devices, including studies up to the date
of the 2006 Panel, as well as any new
clinical information published since the
2006 Panel.
Literature published at the time of the
2006 Panel includes a 1953 seminal
paper on the use of electrical signals to
stimulate bone formation by Yasuda,
that reported bone formation in rabbits
exposed to direct current (DC)
stimulation (Ref. 12). In the following
decades, other researchers expanded on
this finding in animal and clinical
models. In a canine study, a DC
stimulation was shown to cause
complete ossification of the femoral
medullary canal (Ref. 13). The first
clinical case report demonstrated that
electrical stimulation could treat a nonunion fracture (Ref. 14). An early
publication regarding the effects of DC
stimulation on spinal fusion was
published by Dwyer (Ref. 15). Another
early clinical study published by
Becker, et al. showed successful fracture
fusion with a success rate of 77 percent
(Ref. 16).
In the 1990s and early 2000s, several
literature articles were identified
assessing the effects of non-invasive
bone growth stimulators on various
anatomic locations. These studies
generally included various therapeutic
modalities (magnitude, frequency,
duration, etc.) and demonstrated
varying results regarding the efficacy of
these treatments. In two studies of
PEMF devices, Basset and SchinkAscani (Ref. 17) found a 72 percent
fusion rate in patients with congenital
pseudarthrosis of the tibia, and in a
study of non-unions of the scaphoid,
Adams, et al. (Ref. 18) reported a fusion
rate of 69 percent, as a followup to an
earlier study that found a fusion rate of
80 percent. When looking at the rate of
compliance of PEMF devices as a factor
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
of effectiveness, Garland, et al. (Ref. 19)
found that fusion rates ranged from 35.7
percent to 80 percent, depending on
how often the devices were used. In
studies of CC devices, fusion rates in
long bones varied from 60 percent (Ref.
20), 68.8 percent (Ref. 21), and 72.7
percent (Ref. 22), to no difference
between treatment and a placebo-treated
group in a study by Fourie and
Bowerbank (Ref. 23). While there was a
large range of observed efficacies, there
was no reporting of treatment-related
adverse events. These reported
variabilities in efficacy and low adverse
event rates were consistent with the
findings by the 2006 Panel.
FDA performed a systematic review of
published literature to identify any new
clinical findings since the 2006 Panel.
FDA identified 14 papers that included
a combination of retrospective and
prospective studies. For studies that
assessed medical or insurance claims
databases, radiographs were not always
available to determine actual fusion.
Instead, results were presented in terms
of healing rate based on patient records
or reported outcomes. When
radiographs were available and
analyzed to assess union, results were
reported as fusion rate.
Phillips, et al. (Ref. 24) looked at
registry data of 2,370 subjects who were
treated with OL1000 (DJO), a CMF
device, at various fracture sites and
reported an average healing rate of 75.1
percent (ranging from 57.2 percent in
the humerus to 89.7 percent in the
finger phalanx). DeVries, et al. (Ref. 25)
also evaluated the OL1000 device in a
retrospective analysis of 144 subjects,
finding a fusion rate of 57.1 percent in
tibiotalocalcaneal fusions of the ankle.
With respect to LIPUS, Zura, et al.
(Refs. 26 and 27) published two papers
evaluating subjects in the Exogen
(Bioventus) Post Market Registry. One of
the studies assessed how various patient
risk factors affected healing rate in 4,190
subjects. The study demonstrated an
overall healing rate of 95.7 percent, and
in another single arm study of 767
subjects, showed a healing rate varying
from 81.8 percent to 87.9 percent
depending on fracture site. Nolte, et al.
(Ref. 28) evaluated the Exogen registry
in conjunction with a medical claims
database to examine metatarsal fractures
and reported a healing rate of 97.4
percent overall, while Elvey, et al. (Ref.
29) evaluated 26 cases with use of
Exogen in hand and wrist non-unions,
and found a fusion rate of 54 percent to
58 percent. In two smaller studies of the
Exogen device, Majeed, et al. (Ref. 30)
and Mizra, et al. (Ref. 31) both evaluated
foot and ankle fractures and found 78.7
percent and 67 percent fusion rate in a
PO 00000
Frm 00014
Fmt 4702
Sfmt 4702
49991
47 and 18 patient study, respectively.
Biglari (Ref. 32) also performed an
observational study using the Exogen
device and found a much lower fusion
rate of 32.8 percent in 60 subjects
having existing non-unions of various
long bones.
For PEMF devices, a retrospective
study by Coric, et al. (Ref. 33) on the
effects of the CervicalStim (Orthofix)
device on 593 subjects showed a 73.2
percent fusion rate in the cervical spine
at 6 months. In a single arm prospective
study by Assiotis, et al. (Ref. 34), a 77.3
percent fusion rate in the tibia was
demonstrated with use of the
Physiostim (Orthofix). Murray and
Pethica (Ref. 35) performed a 1,382subject retrospective study of use of the
EBI device (Zimmer Biomet) for nonunions of the scaphoid, tibia, and fibula,
and while an assessment of healing rates
was not performed, the data showed
reduction in time to healing between 35
percent and 40 percent when the device
was used as prescribed.
In addition, two randomized control
studies on PEMF devices were
conducted by Foley, et al. (Ref. 36) and
by Streit, et al. (Ref. 37). Foley evaluated
323 subjects using the Orthofix
CervicalStim device in cervical fusion
and found an 83.6 percent fusion rate in
the treatment group compared to a 68.6
percent fusion rate in the control group,
with no difference in pain scores or
adverse events between groups. Streit, et
al. performed a small, eight subject
clinical study using the EBI device to
treat non-unions of the fifth metatarsal
and found the time to fusion was
reduced on average from 14.7 weeks to
8.9 weeks with the use of the device.
In summary, FDA’s literature review
resulted in findings that are consistent
with available clinical data from PMA
submissions. These studies suggest that
there are probable benefits to the use of
these devices; however, differences in
methodology, including differences in
devices used, treatment waveform and
frequency, patient populations, as well
as anatomic location, could have had
significant effects on reported device
effectiveness, which ranged from 32.8
percent to 97.4 percent. Regarding
safety, the findings from these studies
demonstrate that the devices are
relatively safe as the adverse event
profile associated with these devices
using various modalities was similar to
controls. Overall, the studies involved
10,566 subjects (including control
subjects), with only a single report of a
serious adverse event (Biglari, Ref. 32);
however, a direct link to the use of the
device could not be established for this
event.
E:\FR\FM\17AUP1.SGM
17AUP1
49992
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
Further, a search of FDA’s MDR
database was conducted to identify all
adverse events submitted to FDA up to
October 31, 2019, for devices approved
under product codes LOF and LPQ. The
results of the identified reports are
consistent with the risk profiles
identified in both PMA applications and
literature that were reviewed. FDA’s
search yielded a total of 270 unique
MDRs. The most frequently reported
events were categorized as ‘‘skin
reaction/issue’’ (n = 187) followed by
‘‘pain’’ (n = 59) and ‘‘device functional
issue’’ (n = 21). A review of the adverse
events regarding skin reactions found
that a majority were due to irritation
from the electrode adhesive or
ultrasound gel used. There was no
apparent difference in risk profile across
the various device modalities, though
the risk of skin irritation was primarily
observed in the skin-contacting devices
(due to the electrodes in the CC device
and the gel in the LIPUS device). For
cases where followup was described,
patients recovered when treatment was
discontinued. In addition, 11 reports of
‘‘mass/tumor’’ were identified; however,
the nature of the relationship between
the mass/tumor to the device was
unrelated or unclear. Based upon FDA’s
assessment of other systematic reviews
of these devices, no other reports of
mass/tumors have been identified (Refs.
38 to 42).
Finally, a search of FDA’s Medical
Device Recall database was conducted.
No recalls were found when searching
the database for devices under product
code LOF. Two class 2 recalls were
reported for devices under product code
LPQ; specifically, there was a recall for
the Exogen Express Bone Healing
System and a recall for the Exogen
4000+ Ultrasound Bone Healing System.
Both were posted on August 4, 2009,
and initiated by the manufacturer
because of problems with the
transducer, which may have resulted in
a reduced ultrasound output. These
recalls were terminated on November
18, 2010. These recall events reflect the
risks to health identified in section V,
and FDA believes the special controls
proposed, in addition to general
controls, can effectively mitigate the
risks identified.
VIII. Proposed Special Controls
Table 1 outlines the risks to health
identified in section V and the
corresponding mitigation measures
proposed to reasonably assure safety
and effectiveness, which are discussed
in more detail below.
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR NON-INVASIVE BONE GROWTH STIMULATORS
Identified risk to health
Mitigation measures
Failure or delay of osteogenesis ..............................................................
Burn ..........................................................................................................
Electrical shock .........................................................................................
Electromagnetic interference ....................................................................
Adverse tissue reaction ............................................................................
Adverse interaction with internal/external fixation devices ......................
Adverse biological effects .........................................................................
The risk of failure or delay of
osteogenesis is clinically significant. To
mitigate this risk, FDA proposes that
manufacturers provide clinical
performance data to demonstrate that
the device yields positive outcomes
(e.g., fusion of the non-union) in
accordance with its intended use.
Further, FDA proposes non-clinical
performance testing to demonstrate that
the device performs as intended under
anticipated conditions of use to achieve
the identified successful clinical
performance characteristics. This would
include verification and validation of
critical performance characteristics,
including characterization of the
designed outputs of the device as well
as the outputs that are delivered to the
patient, thermal safety and reliability
testing, reliability testing consistent
with the expected device use-life, and
validation that signal characteristics are
within safe physiologic limits. Also,
FDA proposes appropriate software
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
Clinical performance data.
Non-clinical performance testing.
Software verification, validation, and hazard analysis.
Labeling.
Non-clinical performance testing.
Electrical safety testing.
Labeling.
Electrical safety testing.
Labeling.
Electromagnetic compatibility (EMC) testing.
Labeling.
Biocompatibility evaluation.
Labeling.
Labeling.
Non-clinical performance testing.
Software verification, validation, and hazard analysis.
verification, validation, and hazard
analysis to ensure that any device
software performs as intended. Lastly,
FDA proposes labeling to provide
appropriate instructions (e.g., duration,
frequency of use) to the end user.
To mitigate the risk of skin burns,
FDA proposes non-clinical performance
testing of the device to verify and
validate critical performance
characteristics, demonstrate thermal
safety and reliability, validate that
signal characteristics are within safe
physiologic limits, and demonstrate
reliability of the device consistent with
its expected use-life. FDA also proposes
electrical safety testing to minimize the
risk of thermal burns to the patient, and
specific instructions regarding proper
usage and specific warnings associated
with the risk of burns.
To mitigate electrical shocks, FDA
proposes electrical safety testing to
minimize the risk of shock to the
patient. Furthermore, FDA proposes
PO 00000
Frm 00015
Fmt 4702
Sfmt 4702
labeling provisions, including
instructions on appropriate usage and
maintenance, and specific warnings
regarding electrical shock.
To mitigate electromagnetic
interference, FDA proposes
electromagnetic compatibility testing
and labeling to minimize the risk of
adverse interaction with other electronic
devices such as implanted electronic
devices.
To mitigate the risk of adverse tissue
reactions, FDA proposes a
biocompatibility evaluation to ensure
that the materials used in patientcontacting components of the device are
safe for skin contact and labeling that
includes warnings against use on
compromised skin or when there are
known sensitivities, as well as
instructions on appropriate cleaning of
any reusable components.
To mitigate the risk of adverse
interaction with internal/external
fixation devices, FDA proposes labeling,
E:\FR\FM\17AUP1.SGM
17AUP1
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
specifically inclusion of appropriate
warnings for patients with implanted
internal/external devices.
To mitigate the risk of adverse
biologic effects, FDA proposes nonclinical performance testing to verify
and validate critical performance
characteristics of the device,
demonstrate thermal safety and
reliability, validate safety of the signal
by reference to known biological safety
limits, and demonstrate reliability of the
device over the expected use-life.
Furthermore, FDA proposes software
verification, validation, and hazard
analysis.
If this reclassification is finalized,
non-invasive bone growth stimulators
will be reclassified into class II and
would be subject to premarket
notification (510(k)) requirements under
§ 807.81. As discussed below, the intent
is for the reclassification to be codified
in 21 CFR 890.5870. Firms submitting a
510(k) for non-invasive bone growth
stimulators will be required to comply
with the particular mitigation measures
set forth in the special controls.
Adherence to the special controls, in
addition to the general controls, is
necessary to provide a reasonable
assurance of the safety and effectiveness
of these devices.
IX. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this
proposed order contains no new
collections of information. Therefore,
clearance by the Office of Management
and Budget (OMB) under the Paperwork
Reduction Act of 1995 (PRA) (44 U.S.C.
3501–3521) is not required. This
proposed order refers to previously
approved FDA collections of
information. These collections of
information are subject to review by
OMB under the PRA. The collections of
information in part 807, subpart E have
been approved under OMB control
number 0910–0120; the collections of
information in 21 CFR part 814,
subparts A through E, have been
approved under OMB control number
0910–0231; and the collections of
information under 21 CFR part 801 have
been approved under OMB control
number 0910–0485.
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
XI. Codification of Orders
Under section 513(f)(3) of the FD&C
Act, FDA may issue final orders to
reclassify devices. FDA will continue to
codify classifications and
reclassifications in the Code of Federal
Regulations (CFR). Changes resulting
from final orders will appear in the CFR
as newly codified orders. Therefore,
under section 513(f)(3), in the proposed
order, we are proposing to codify noninvasive bone growth stimulators in the
new 21 CFR 890.5870, under which
non-invasive bone growth stimulators
would be reclassified from class III to
class II.
XII. Proposed Effective Date
FDA proposes that any final order
based on this proposal become effective
30 days after the date of its publication
in the Federal Register.
XIII. References
The following references marked with
an asterisk (*) are on display at the
Dockets Management Staff (see
ADDRESSES) and are available for
viewing by interested persons between
9 a.m. and 4 p.m., Monday through
Friday; they also are available
electronically at https://
www.regulations.gov. References
without asterisks are not on public
display at https://www.regulations.gov
because they have copyright restriction.
Some may be available at the website
address, if listed. References without
asterisks are available for viewing only
at the Dockets Management Staff. FDA
has verified the website addresses, as of
the date this document publishes in the
Federal Register, but websites are
subject to change over time.
1. *Guidance on Section 216 of the Food
and Drug Administration Modernization Act
of 1997, available at https://www.fda.gov/
regulatory-information/search-fda-guidancedocuments/guidance-section-216-food-anddrug-administration-modernization-act-1997guidance-industry-and-fda.
2. *P790002 Approval available at: https://
www.accessdata.fda.gov/scripts/cdrh/cfdocs/
cfpma/pma.cfm?id=P790002.
3. *RS Medical Corporation
Reclassification Petition, available at https://
wayback.archive-it.org/7993/
20170405072021/https://www.fda.gov/
ohrms/dockets/ac/06/briefing/2006-4224b106-TabB-RSMEDICAL-Petition.pdf.
4. *FDA’s Orthopaedic and Rehabilitation
Devices Panel transcript and other meeting
materials for the June 2, 2006, meeting are
available at: https://wayback.archive-it.org/
7993/20170403222246/https://www.fda.gov/
ohrms/dockets/ac/cdrh06.html#orthopaedic.
5. *Letter from RS Medical requesting
withdrawal of petition, available at https://
www.regulations.gov/document?D=FDA2005-P-0052-0007.
PO 00000
Frm 00016
Fmt 4702
Sfmt 4702
49993
6. *FDA letter granting RS Medical’s
withdrawal request, available at https://
www.regulations.gov/document?D=FDA2005-P-0052-0006.
7. *First Cohort of Results of the 2014–
2015 Strategic Priority: Strike the Right
Balance Between Premarket and Postmarket
Data Collection (April 2015), available at
https://www.fda.gov/media/91437/download.
8. *Second and Final Cohort of Results of
the 2014–2015 Strategic Priority: Strike the
Right Balance Between Premarket and
Postmarket Data Collection (August 2016),
available at https://www.fda.gov/media/
99822/download.
9. *P030034 Summary of Safety and
Effectiveness, available at https://
www.accessdata.fda.gov/cdrh_docs/pdf3/
P030034B.pdf.
10. *P850022/S009 Summary of Safety and
Effectiveness, available at https://
www.accessdata.fda.gov/cdrh_docs/pdf/
P850022S009B.pdf.
11. *P910066/S011 Summary of Safety and
Effectiveness, available at https://
www.accessdata.fda.gov/cdrh_docs/pdf/
P910066S011B.pdf.
12. Yasuda, I., ‘‘The Classic: Fundamental
Aspects of Fracture Treatment.’’ J. Kyoto
Med. Soc., 4:395–406, 1953.
13. Bassett, C.A., R.J. Pawluk, and R.O.
Becker, ‘‘Effects of Electric Currents on Bone
In Vivo.’’ Nature, 204: 652–654, 1964.
14. Friedenberg, Z.B., M.C. Harlow, and
C.T. Brighton, ‘‘Healing of Nonunion of the
Medial Malleolus by Means of Direct Current:
A Case Report.’’ The Journal of Trauma,
11(10):883–885, 1971.
15. Dwyer, A.F., ‘‘The Use of Electrical
Current Stimulation in Spinal Fusion.’’ The
Orthopedic Clinics of North America,
6(1):265–273, 1975.
16. Becker, R.O., J.A. Spadaro, and A.A.
Marino, ‘‘Clinical Experiences With Low
Direct Current Stimulation of Bone Growth.’’
Clinical Orthopaedics and Related Research,
124:75–83, 1977.
17. Bassett, C.A. and M. Schink-Ascani,
‘‘Long-Term Pulsed Electromagnetic Field
(PEMF) Results in Congenital
Pseudarthrosis.’’ Calcified Tissue
International, 49(3):216–220, 1991.
18. Adams, B.D., G.K. Frykman, and J.
Taleisnik, ‘‘Treatment of Scaphoid Nonunion
With Casting and Pulsed Electromagnetic
Fields: A Study Continuation.’’ The Journal
of Hand Surgery, 17(5):910–914, 1992.
19. Garland, D.E., B. Moses, and W. Salyer,
‘‘Long-Term Follow-up of Fracture
Nonunions Treated With PEMFs.’’
Contemporary Orthopaedics, 22(3):295–302,
1991.
20. Scott, G. and J.B. King, ‘‘A Prospective,
Double-Blind Trial of Electrical Capacitive
Coupling in the Treatment of Non-Union of
Long Bones.’’ The Journal of Bone and Joint
Surgery. American volume, 76(6):820–826,
1994.
21. Abeed, R.I., M. Naseer, and E.W. Abel,
‘‘Capacitively Coupled Electrical Stimulation
Treatment: Results from Patients With Failed
Long Bone Fracture Unions.’’ Journal of
Orthopaedic Trauma, 12(7):510–513, 1998.
22. Zamora-Navas, P., A. Borras Verdera, R.
Antelo Lorenzo, et al., ‘‘Electrical
Stimulation of Bone Nonunion With the
E:\FR\FM\17AUP1.SGM
17AUP1
49994
Federal Register / Vol. 85, No. 159 / Monday, August 17, 2020 / Proposed Rules
Presence of a Gap.’’ Acta Orthopaedica
Belgica, 61(3):169–176, 1995.
23. Fourie, J.A. and P. Bowerbank,
‘‘Stimulation of Bone Healing in New
Fractures of the Tibial Shaft Using
Interferential Currents.’’ Physiotherapy
Research International, 2(4):255–268, 1997.
24. Philips, M., J. Baumhauer, S. Sprague,
et al., ‘‘Use of Combined Magnetic Field
Treatment for Fracture Nonunion.’’ Journal of
Long Term Effects of Medical Implants,
26(3):277–284, 2016.
25. DeVries, J.G., G.C. Berlet, and C.F.
Hyer, ‘‘Union Rate of Tibiotalocalcaneal
Nails With Internal or External Bone
Stimulation.’’ Foot and Ankle International,
33(11):969–978, 2012.
26. Zura, R., S. Mehta, G.J. Della Rocca, et
al., ‘‘A Cohort Study of 4,190 Patients
Treated With Low-Intensity Pulsed
Ultrasound (LIPUS): Findings in the Elderly
Versus All Patients.’’ BioMed Central
Musculoskeletal Disorders, 16:45, 2015.
27. Zura, R., G.J. Della Rocca, S. Mehta, et
al., ‘‘Treatment of Chronic (>1 Year) Fracture
Nonunion: Heal Rate in a Cohort of 767
Patients Treated With Low-Intensity Pulsed
Ultrasound (LIPUS).’’ Injury, 46(10):2036–
2041, 2015.
28. Nolte, P., R. Anderson, E. Strauss, et al.,
‘‘Heal Rate of Metatarsal Fractures: A
Propensity-Matching Study of Patients
Treated With Low-Intensity Pulsed
Ultrasound (LIPUS) vs. Surgical and Other
Treatments.’’ Injury, 47(11):2584–2590, 2016.
29. Elvey, M.H., R. Miller, K.S. Khor, et al.,
‘‘The Use of Low-Intensity Pulsed Ultrasound
in Hand and Wrist Nonunions.’’ Journal of
Plastic Surgery and Hand Surgery, 1–6, 2019
(published online ahead of print, November
26, 2019).
30. Majeed, H., T. Karim, J. Davenport, et
al., ‘‘Clinical and Patient-Reported Outcomes
Following Low Intensity Pulsed Ultrasound
(LIPUS, Exogen) for Established PostTraumatic and Post-Surgical Nonunion in the
Foot and Ankle.’’ Foot and Ankle Surgery,
S1268–7731(19)30072–4, 2019 (published
online ahead of print, May 19, 2019).
31. Mizra, Y.H., K.H. Teoh, D. Golding, et
al., ‘‘Is There a Role for Low Intensity Pulsed
Ultrasound (LIPUS) in Delayed or Nonunion
Following Arthrodesis in Foot and Ankle
Surgery?’’ Foot and Ankle Surgery,
25(6):842–848, 2019.
32. Biglari, B., T.M. Yildirim, T. Swing, et
al., ‘‘Failed Treatment of Long Bone
Nonunions With Low Intensity Pulsed
Ultrasound.’’ Archives of Orthopaedic and
Trauma Surgery, 136(8):1121–1134, 2016.
33. Coric, D., D.E. Bullard, V.V. Patel, et al.,
‘‘Pulsed Electromagnetic Field Stimulation
May Improve Fusion Rates in Cervical
Arthrodesis in High-Risk Populations.’’ Bone
and Joint Research, 7(2):124–130, 2018.
34. Assiotis, A., N.P. Sachinis, and B.E.
Chalidis, ‘‘Pulsed Electromagnetic Fields for
the Treatment of Tibial Delayed Unions and
Nonunions. A Prospective Clinical Study and
Review of the Literature.’’ Journal of
Orthopaedic Surgery and Research, 7:24,
2012.
35. Murray, H.B. and B.A. Pethica, ‘‘A
Follow-up Study of the In-Practice Results of
Pulsed Electromagnetic Field Therapy in the
Management of Nonunion Fractures.’’
VerDate Sep<11>2014
16:51 Aug 14, 2020
Jkt 250001
Orthopaedic Research and Reviews, 8:67–72,
2016.
36. Foley, K.T., T.E. Mroz, P.M. Arnold, et
al., ‘‘Randomized, Prospective, and
Controlled Clinical Trial of Pulsed
Electromagnetic Field Stimulation for
Cervical Fusion.’’ Spine Journal, 8(3):436–
442, 2008.
37. Streit, A., B.C. Watson, J.D. Granata, et
al., ‘‘Effect on Clinical Outcome and Growth
Factor Synthesis With Adjunctive Use of
Pulsed Electromagnetic Fields for Fifth
Metatarsal Nonunion Fracture: A DoubleBlind Randomized Study.’’ Foot and Ankle
International, 37(9):919–923, 2016.
38. Aleem, I.S., I. Aleem, N. Evaniew, et
al., ‘‘Efficacy of Electrical Stimulators for
Bone Healing: A Meta-Analysis of
Randomized Sham-Controlled Trials.’’
Science Reports, 6:31724, 2016.
39. Behrens, S.B., M.E. Deren, and K O.
Monchik, ‘‘A Review of Bone Growth
Stimulation for Fracture Treatment.’’ Current
Orthopedic Practice, 24(1):84–91, 2013.
40. Griffin, X.L., M.L. Costa, N. Parsons, et
al., ‘‘Electromagnetic Field Stimulation for
Treating Delayed Union or Non-Union of
Long Bone Fractures in Adults.’’ The
Cochrane Database of Systematic Reviews,
4:CD008471, 2011.
41. Griffin, X.L., N. Smith, N. Parsons, et
al., ‘‘Ultrasound and Shockwave Therapy for
Acute Fractures in Adults.’’ The Cochrane
Database of Systematic Reviews,
2:CD008579, 2012.
42. Hannemann, P.F., E.H.H. Mommers,
J.P.M. Schots, et al., ‘‘The Effects of LowIntensity Pulsed Ultrasound and Pulsed
Electromagnetic Fields Bone Growth
Stimulation in Acute Fractures: A Systematic
Review and Meta-Analysis of Randomized
Controlled Trials.’’ Archives of Orthopaedic
and Trauma Surgery, 134(8):1093–1106,
2014.
List of Subjects in 21 CFR Part 890
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 890 be amended as follows:
PART 890—PHYSICAL MEDICINE
DEVICES
1. The authority citation for part 890
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 890.5870 to subpart F to read
as follows:
■
§ 890.5870 Non-invasive bone growth
stimulator.
(a) Identification. A non-invasive
bone growth stimulator provides
stimulation through electrical, magnetic,
or ultrasonic fields. The device is for
prescription use and is intended to be
used externally to promote osteogenesis
as an adjunct to primary treatments for
fracture fixation or spinal fusion.
PO 00000
Frm 00017
Fmt 4702
Sfmt 9990
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Clinical performance data must
support the intended use of the device.
(2) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following must
be provided:
(i) Verification and validation of
critical performance characteristics of
the device, including characterization of
the designed outputs of the device as
well as the outputs that are delivered to
the patient.
(ii) Thermal safety and thermal
reliability testing.
(iii) Validation that signal
characteristics are within safe
physiologic limits.
(iv) Reliability testing consistent with
the expected use-life of the device.
(3) Patient-contacting components of
the device must be demonstrated to be
biocompatible.
(4) Performance data must
demonstrate the electrical safety and
electromagnetic compatibility of the
device.
(5) Appropriate software verification,
validation, and hazard analysis must be
performed.
(6) Labeling for the device must
include the following:
(i) Warning against use on
compromised skin or when there are
known sensitivities;
(ii) Appropriate warnings for patients
with implanted medical devices;
(iii) A detailed summary of the
clinical testing, which includes the
clinical outcomes associated with the
use of the device, and a summary of
adverse events and complications that
occurred with the device;
(iv) A clear description of the device;
(v) Instructions on appropriate usage,
duration, and frequency of use;
(vi) Instructions for maintenance and
safe disposal;
(vii) Instructions for appropriate
cleaning of any reusable components;
(viii) Specific warnings regarding user
burns, electrical shock, and skin
irritation; and
(ix) The risks and benefits associated
with use of the device.
Dated: August 4, 2020.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2020–17543 Filed 8–14–20; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\17AUP1.SGM
17AUP1
Agencies
[Federal Register Volume 85, Number 159 (Monday, August 17, 2020)]
[Proposed Rules]
[Pages 49986-49994]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2020-17543]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 890
[Docket No. FDA-2020-N-1053]
Physical Medicine Devices; Reclassification of Non-Invasive Bone
Growth Stimulators
AGENCY: Food and Drug Administration, Health and Human Services (HHS).
ACTION: Proposed amendment; proposed order; request for comments.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is proposing to
reclassify non-invasive bone growth stimulators, postamendments class
III devices (product codes LOF and LPQ), into class II (special
controls), subject to premarket notification. FDA is also proposing a
new device classification with the name ``non-invasive bone growth
stimulators'' along with the proposed special controls that the Agency
believes are necessary to provide a reasonable assurance of safety and
effectiveness of these devices. FDA is proposing this reclassification
on its own initiative. If finalized, this order will reclassify these
devices from class III (premarket approval) to class II (special
controls) and reduce the regulatory burdens associated with these
devices, as these devices will no longer be required to submit a
premarket approval application (PMA), but are subject to premarket
notification (510(k)) requirements and general and special controls.
DATES: Submit either electronic or written comments on the proposed
order by October 16, 2020. Please see section XII of this document for
the proposed effective date when the new requirements apply and for the
proposed effective date of a final order based on this proposed order.
ADDRESSES: You may submit comments as follows. Please note that late,
[[Page 49987]]
untimely filed comments will not be considered. Electronic comments
must be submitted on or before October 16, 2020. The https://www.regulations.gov electronic filing system will accept comments until
11:59 p.m. Eastern Time at the end of October 16, 2020. Comments
received by mail/hand delivery/courier (for written/paper submissions)
will be considered timely if they are postmarked or the delivery
service acceptance receipt is on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal Rulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand Delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2020-N-1053 for ``Physical Medicine Devices; Reclassification of
Non-Invasive Bone Growth Stimulators.'' Received comments, those filed
in a timely manner (see ADDRESSES), will be placed in the docket and,
except for those submitted as ``Confidential Submissions,'' publicly
viewable at https://www.regulations.gov or at the Dockets Management
Staff between 9 a.m. and 4 p.m., Monday through Friday.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.govinfo.gov/content/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Dockets Management Staff, 5630 Fishers Lane,
Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Jesse Muir, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4508, Silver Spring, MD 20993, 240-402-6679,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended,
establishes a comprehensive system for the regulation of medical
devices intended for human use. Section 513 of the FD&C Act (21 U.S.C.
360c) established three categories (classes) of devices, reflecting the
regulatory controls needed to provide reasonable assurance of their
safety and effectiveness. The three categories of devices are class I
(general controls), class II (general controls and special controls),
and class III (general controls and premarket approval).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices) are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until: (1) FDA
reclassifies the device into class I or II or (2) FDA issues an order
finding the device to be substantially equivalent, in accordance with
section 513(i) of the FD&C Act, to a predicate device that does not
require premarket approval. FDA determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807), subpart E, of the regulations.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or class II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA acting by administrative
order, can reclassify the device into class I or class II on its own
initiative, or in response to a petition from the manufacturer or
importer of the device. To change the classification of the device, the
proposed new class must have sufficient regulatory controls to provide
reasonable assurance of the safety and effectiveness of the device for
its intended use.
Reevaluation of the data previously presented before the Agency is
an appropriate basis for subsequent action, where the reevaluation is
made in light of newly available regulatory authority (see Bell v.
Goddard, 366 F.2d 177, 181 (7th Cir. 1966); Ethicon, Inc. v. FDA, 762
F. Supp. 382, 388-391 (D.D.C. 1991)) or in light of changes in
``medical science'' (Upjohn v. Finch, 422 F.2d 944, 951 (6th Cir.
1970)). Whether data before the Agency are old or new, the ``new
information'' to support reclassification under 513(f)(3) must be
``valid scientific evidence'', as defined in section 513(a)(3) of the
FD&C Act and Sec. 860.7(c)(2) (21 CFR 860.7(c)(2)). (See, e.g.,
General Medical Co. v. FDA, 770 F.2d 214 (D.C. Cir. 1985); Contact Lens
Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.
[[Page 49988]]
1985), cert. denied, 474 U.S. 1062 (1986).)
FDA relies upon ``valid scientific evidence,'' as defined in
section 513(a)(3) of the FD&C Act and Sec. 860.7(c)(2), in the
classification process to determine the level of regulation for
devices. To be considered in the reclassification process, the ``valid
scientific evidence'' upon which the Agency relies must be publicly
available. Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
PMA (see section 520(c) of the FD&C Act (21 U.S.C. 360j(c)). Section
520(h)(4) of the FD&C Act provides that FDA may use, for
reclassification of a device, certain information in a PMA 6 years
after the application has been approved (Ref. 1). This includes
information from clinical and preclinical tests or studies that
demonstrate the safety or effectiveness of the device, but does not
include descriptions of methods of manufacture or product composition
and other trade secrets.
In accordance with section 513(f)(3) of the FD&C Act, FDA is
issuing this proposed order to reclassify non-invasive bone growth
stimulator devices, postamendments class III devices, into class II
(special controls), subject to premarket notification because FDA
believes the standard in section 513(a)(1)(B) of the FD&C Act is met as
there is sufficient information to establish special controls, which in
addition to general controls, will provide reasonable assurance of the
safety and effectiveness of the device.\1\
---------------------------------------------------------------------------
\1\ In December 2019, FDA began adding the term ``Proposed
amendment'' to the ``ACTION'' caption for these documents, typically
styled ``Proposed order'', to indicate that they ``propose to
amend'' the Code of Federal Regulations. This editorial change was
made in accordance with the Office of Federal Register's (OFR)
interpretations of the Federal Register Act (44 U.S.C. chapter 15),
its implementing regulations (1 CFR 5.9 and parts 21 and 22), and
the Document Drafting Handbook.
---------------------------------------------------------------------------
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the premarket notification requirements under section
510(k) of the FD&C Act, if the Agency determines that premarket
notification is not necessary to reasonably assure the safety and
effectiveness of the device. FDA has determined that premarket
notification is necessary to reasonably assure the safety and
effectiveness of non-invasive bone growth stimulator devices.
Therefore, the Agency does not intend to exempt these proposed class II
devices from premarket notification (510(k)) submission as provided
under section 510(m) of the FD&C Act.
II. Regulatory History of Non-Invasive Bone Growth Stimulator Devices
In accordance with section 513(f)(1) of the FD&C Act, non-invasive
bone growth stimulator devices were automatically classified into class
III because they were not introduced or delivered for introduction into
interstate commerce for commercial distribution before May 28, 1976,
and have not been found substantially equivalent to a device placed in
commercial distribution after May 28, 1976, which was subsequently
classified or reclassified into class II or class I. Therefore, the
device is subject to PMA requirements under section 515 of the FD&C Act
(21 U.S.C. 360e).
Accordingly, on November 6, 1979, FDA approved a PMA for the Bio
Osteogen System 204 (P790002) (Ref. 2). Since that time, five
additional original PMAs have been approved for non-invasive bone
growth stimulators (P850007, P850022, P900009, P910066, and P030034).
On February 9, 2005, FDA received a reclassification petition dated
February 7, 2005, submitted by RS Medical Corporation, requesting that
FDA reclassify certain non-invasive bone growth stimulators from class
III to class II (Ref. 3). As stated in the Notice of Panel
Recommendation discussed further below, ``the petition was submitted
under section 513(e) of the act but FDA . . . review[ed] the petition
under section 513(f)(3) of the act because that section contain[ed] the
appropriate procedures for reclassification of postamendments devices''
(72 FR 1951 at 1952, January 17, 2007). FDA requested additional
information and the petitioner amended the petition on November 30,
2005 (``amended petition''). In accordance with the FD&C Act and
regulations, FDA referred the petition, as amended, to the FDA Advisory
Committee, specifically the Orthopaedic and Rehabilitation Devices
Panel (``the 2006 Panel'') for its recommendations on the requested
reclassification.
On June 2, 2006, the 2006 Panel deliberated on the information in
RS Medical's petition; the presentations made by RS Medical, FDA, and
members of the public; and their own experience with certain non-
invasive bone growth stimulators (Ref. 4).
The 2006 Panel identified the following risks to health associated
with non-invasive bone growth stimulators: Electric shock; burn; skin
irritation and/or allergic reaction; inconsistent or ineffective
treatment; adverse interaction with electrical implants; adverse
interactions with internal/external fixation devices; and biological
risks. The 2006 Panel did not specifically address risks associated
with ultrasound-based devices, as these were outside the scope of RS
Medical's petition; however, as discussed below, based upon FDA's
review of information since the Panel meeting, the risks identified
with ultrasound-based devices, along with their reported benefits, are
comparable to those of non-invasive bone growth stimulators
incorporating other modalities.
The majority of the 2006 Panel recommended that non-invasive bone
growth stimulators should be retained in class III because there was
insufficient information in the petition by RS Medical to establish
that special controls in conjunction with general controls would
provide a reasonable assurance of the safety and effectiveness of the
device. Specifically, the Panel recommended that the proposed special
controls by RS Medical were sufficient to control for the risk of
electric shock, burn, skin irritation, and/or allergic reaction;
adverse interaction with electrical implants; adverse interactions with
internal/external fixation devices; and biological risks. However, the
Panel believed that there was insufficient evidence presented by RS
Medical to control for the risk of inconsistent or ineffective
treatment because there is a lack of knowledge about how waveform
characteristics (e.g., pulse duration, amplitude, power, frequency),
including potential modifications to the device, affect the clinical
response to treatment. The Panel requested additional clinical data
and/or special controls, which was not adequately devised by the
petitioner, to control for the risk of inconsistent or ineffective
treatment.
FDA concurred with the 2006 Panel's recommendation, and similarly
believed that RS Medical's petition was inadequate in that FDA had
concerns about the petitioner's proposed special controls to control
the risk of inconsistent or ineffective treatment. In the Federal
Register of January 17, 2007 (72 FR 1951), FDA published a Notice of
Panel Recommendations (``the 2007 Notice''), as referenced above.
In a letter dated April 2, 2007, RS Medical requested that its
petition be withdrawn (Ref. 5). On July 10, 2007, FDA granted RS
Medical's request for withdrawal of the petition and did not take any
further action on the petition (Ref. 6). FDA has not received any
subsequent petition requesting reclassification of these devices.
Subsequently, as part of the Center for Devices and Radiological
Health's 2014-2015 strategic priority, ``Strike the Right Balance
Between Premarket and Postmarket Data Collection,'' a
[[Page 49989]]
retrospective review of all PMA product codes with active PMAs approved
prior to 2010 was conducted to determine whether, based on our current
understanding of the technology, certain devices could be reclassified
(down-classified). On April 29, 2015, FDA published a document in the
Federal Register identifying certain product codes as potential
candidates for reclassification (80 FR 23798), including non-invasive
bone growth stimulators under product codes LOF and LPQ, from class III
to class II (Ref. 7). One comment was received in response to this
proposal for reclassification of LOF and LPQ; this comment did not
support FDA's intention to reclassify these devices, citing the
concerns discussed during the 2006 Panel. This comment was considered
in development of this proposed order. Note that invasive bone growth
stimulators, designated under product code LOE, are outside the scope
of this proposed reclassification. As noted in the 2006 Panel, invasive
bone growth stimulator devices have added risks compared to non-
invasive bone growth stimulators, and therefore would require a
separate classification discussion. Furthermore, invasive bone growth
stimulators were also considered as a part of the aforementioned PMA
retrospective review and FDA determined that these devices should
remain as class III (Ref. 8). Therefore, FDA will continue to regulate
invasive bone growth stimulators as a class III device, subject to PMA
requirements.
While RS Medical's petition inadequately addressed all of the risks
associated with non-invasive bone growth stimulators for
reclassification, FDA is, on its own initiative, proposing to
reclassify these devices from class III to class II, and believes that
sufficient information exists to establish special controls, as
identified in this proposed order, that, together with general
controls, can provide a reasonable assurance of safety and
effectiveness for this device type. Additionally, RS Medical in its
petition excluded use of these devices as an adjunct to cervical fusion
surgery in patients at high risk for nonfusion, as well as for use in
congenital pseudarthrosis. Based upon the review of the evidence and
FDA's ability to establish special controls, FDA believes these
indications that have been approved for currently marketed non-invasive
bone growth stimulator devices should be included in this proposed
reclassification.
III. Device Description
Non-invasive bone growth simulators, currently designated under
product codes LOF and LPQ, are typically composed of a waveform
generator and transducer (e.g., coils, electrodes, and/or ultrasound
transducers). Patient-contacting surfaces include the transducers, lead
wires, and the device outer casing.
Non-invasive bone growth stimulators utilize an electrical
component to produce an output electrical, magnetic, or ultrasonic
waveform that is delivered to a treatment site via a non-invasively
applied transducer (e.g., electromagnetic coil or ultrasound
transducer) or electrodes (e.g., capacitor plates). The device also
incorporates an internal means to monitor the output waveform and
delivery of treatment, and to provide visual and/or audible alarms to
alert the user of improper device function. The induced electrical and/
or magnetic fields are generated using one of the following modalities:
Capacitive coupling (CC), in which a pair of electrodes
are placed on the skin such that a current can be driven across the
target site;
pulsed electromagnetic fields (PEMF), in which a modulated
electromagnetic field is generated near the treatment site though an
external coil; or
combined magnetic fields (CMF), in which a coil generates
a combination of a static and pulsed magnetic field near the treatment
site.
The ultrasonic waveform is generated using:
Low intensity pulsed ultrasound (LIPUS), in which pulsed
ultrasonic signals are generated using ultrasonic transducers.
The non-invasive nature of the device obviates the need for sterile
components; however, patient-contacting surfaces should be capable of
being cleaned as needed and biocompatibility must be ensured.
Non-invasive bone growth stimulators are generally intended to
promote osteogenesis as adjunct to primary treatments for fracture
fixation or spinal fusion. The indications for use for this device type
depend on the specific device characteristics, but have included:
Treatment of an established non-union secondary to trauma
of the appendicular system,
treatment of congenital pseudarthrosis,
treatment of failed fusions of the appendicular system,
early treatment of certain fresh fractures, and
as an adjunct to lumbar or cervical spinal fusion.
In addition, non-invasive bone growth stimulators are currently
prescription use only devices under Sec. 801.109 (21 CFR 801.109).
IV. Proposed Reclassification
In accordance with section 513(f)(3) of the FD&C Act and 21 CFR
part 860, subpart C, FDA is proposing to reclassify non-invasive bone
growth stimulator devices under product codes LOF and LPQ from class
III into class II. This includes devices that generate electrical or
magnetic fields using CC, PEMF, and CMF, and ultrasonic signals.
FDA believes that there is sufficient information available by way
of FDA's accumulated experience with these devices from review of
premarket submissions, peer-reviewed literature, medical device reports
(MDRs), and recalls to understand the risks associated with these
devices to establish special controls that effectively mitigate the
risks to health identified in section V. In this proposed order, the
Agency has identified the special controls under section 513(a)(1)(B)
of the FD&C Act that, together with general controls, would provide a
reasonable assurance of the safety and effectiveness for non-invasive
bone growth stimulators to be in class II. Absent the special controls
identified in this proposed order, general controls applicable to this
device type are insufficient to provide reasonable assurance of safety
and effectiveness of the device.
FDA is proposing to create a classification regulation for non-
invasive bone growth stimulators, which would include devices
designated under product codes LOF and LPQ. Under this proposed order,
if finalized, a non-invasive bone growth stimulator will be identified
as a prescription device. As such, the prescription device must satisfy
prescription labeling requirements (see Sec. 801.109, Prescription
devices). Prescription devices are exempt from the requirement for
adequate directions for use for the layperson under section 502(f)(1)
of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the
conditions of Sec. 801.109 are met. In this proposed order, if
finalized, the Agency has identified the special controls under section
513(a)(1)(B) of the FD&C Act that, together with general controls, will
provide a reasonable assurance of the safety and effectiveness for non-
invasive bone growth stimulator devices.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act if FDA determines that premarket notification is
not necessary
[[Page 49990]]
to provide reasonable assurance of the safety and effectiveness of the
device. For non-invasive bone growth stimulators, FDA has determined
that premarket notification is necessary to provide reasonable
assurance of the safety and effectiveness of the device. Therefore, FDA
does not propose to exempt these proposed class II devices from 510(k)
requirements. If this order is finalized, persons who intend to market
this type of device would need to submit to FDA a 510(k) and receive
clearance prior to marketing the device.
V. Risks to Health
Based on available information for non-invasive bone growth
stimulators, including the 2005 reclassification petition request from
RS Medical Corp, input from the 2006 Panel, data in PMA applications
P030034, P850022/S009, and P910066/S011 available to FDA under section
520(h)(4) of the FD&C Act, published literature, and postmarket
experience associated with use of these devices, FDA identifies the
following risks to health associated with non-invasive bone growth
stimulators:
a. Failure or delay of osteogenesis--A patient could receive
ineffective treatment, contributing to failure or delay of osteogenesis
that may lead to clinical symptoms (e.g., pain) and the need for
surgical interventions. Ineffective treatment could be a result of
various circumstances (e.g., inadequate therapeutic signal output or
device malfunction or misuse).
b. Burn--A patient or health care professional could be burned from
the use and operation of the device. This could be a result of various
circumstances including device malfunction (e.g., electrical fault) or
misuse of the device (e.g., use while sleeping).
c. Electrical shock--A patient or health care professional could be
shocked from the use and operation of the device. This could be a
result of various circumstances including device malfunction (e.g.,
electrical fault) or misuse of the device (e.g., use of alternating
current source during treatment).
d. Electromagnetic interference (EMI)--A patient with electrically
powered implants (such as cardiac pacemakers, cardiac defibrillators,
and neurostimulators) could experience an adverse interaction with the
implanted electrical device via EMI or radiofrequency interference.
e. Adverse tissue reaction--A patient could experience skin
irritation and/or allergic reaction associated with the use and
operation of the device via the use of non-biocompatible device
materials.
f. Adverse interaction with internal/external fixation devices--The
signal output could be impacted by certain metallic internal or
external fixation devices leading to inadequate treatment signals,
device malfunction, or tissue damage.
g. Adverse biologic effects--A patient may experience adverse
biologic effects resulting from prolonged exposure to the treatment
signal via biologic interaction with the treatment signal at a cellular
level. Excessive energy transmission could cause tissue damage or
aberrant tissue behavior if signal output parameters exceed established
safety thresholds.
The risks to health identified within this proposed order are
consistent with those identified in the 2005 reclassification petition,
as amended. The 2006 Panel agreed with these identified risks; however,
in some cases the risk or accompanying description was reworded for
clarity in this proposed order (e.g., ``inconsistent treatment or
ineffective treatment'' is described in terms of risk to health, which
may entail ``failure or delay in osteogenesis''). Also, the risk of
adverse biologic effects previously specified risks of carcinogenicity,
genotoxicity, mutagenicity, and teratological effects. The petitioner
notes in the amended petition that ``. . . the evidence points to lack
of genotoxic, carcinogenic, and teratologic potential of the subject
waveforms,'' which is corroborated by the lack of such reports
identified in the literature. Although FDA similarly has found a lack
of such reports, it considers this risk more generally as potential
deleterious effects at the tissue or cellular level due to signal
output parameters that exceed established safety thresholds.
VI. Summary of Reasons for Reclassification
FDA believes that non-invasive bone growth stimulator devices,
which are intended to promote osteogenesis as an adjunct to primary
treatments for fracture fixation or spinal fusion, should be
reclassified from class III to class II and that there is sufficient
information to establish special controls for the risks identified in
section V which, in addition to general controls, can provide
reasonable assurance of safety and effectiveness.
Specifically, FDA proposes to require clinical performance data as
a special control to address the risk of failure or delay of
osteogenesis. FDA review of the literature suggests a high variability
of treatment efficacy, depending on therapeutic signal and anatomic
location. This would also address the main concern cited by the 2006
Panel and FDA with RS Medical's proposal, which led to the
recommendation to retain non-invasive bone growth stimulators in class
III, and various comments received in response to the 2007 Notice.
FDA's proposal would require that clinical performance of any non-
invasive bone growth stimulator device be evaluated in support of the
intended use. Rather than prescribe specific study requirements, FDA's
proposal would allow for flexibility in study design and the level of
clinical evidence needed by taking into consideration certain
parameters, e.g., the intended use, treatment population, and
technological characteristics of the device, including any similarities
between the device and legally marketed predicate device, as
appropriate.
VII. Summary of Data Upon Which the Reclassification Is Based
The available evidence demonstrates that there are probable health
benefits derived from the use of these devices, and that the nature and
incidence of risks are well known so that special controls can be
established to adequately mitigate the risks to health. FDA is
proposing a single device class for non-invasive bone growth
stimulators, considering that FDA did not identify any unique risks
associated with the different modalities included in this proposed
order. FDA has considered and analyzed the following: Data in PMA
applications P030034, P850022/S009, and P910066/S011 available to FDA
under section 520(h)(4) of the FD&C Act; information presented at the
2006 Panel concerning RS Medical's petition to down-classify certain
non-invasive bone growth stimulators (Ref. 3) and the 2007 Notice;
peer-reviewed articles that discussed the use of, as well as the
probable benefits and risks of these devices; reported adverse events
identified through a search of FDA's Medical Device Reporting (MDR)
system; and a review of any recalls associated with these devices
through a search of FDA's Medical Device Recall database.
In accordance with the ``6-year rule'' described in section
520(h)(4) of the FD&C Act, FDA considered data contained in three
original PMAs or supplements, P030034, P850022/S009, and P910066/S011,
approved for non-invasive bone growth stimulators (Refs. 9 to 11).
These PMAs/supplements include three different device modalities: A
PEMF device (P030034), a CC device (P850022/S009), and a CMF device
(P910066/S011). In review of the reported clinical data in the summary
of
[[Page 49991]]
safety and effectiveness data documents (SSEDs), the studies conducted
in support of these devices include a total study size of 831 enrolled
subjects. The adverse event profile for the devices in each study were
similar to the control group, with a similar distribution of event
types. With regards to benefit, the clinical data reported in the SSEDs
demonstrate an improved rate of bone fusion compared to placebo
controls, with an 83.6 percent vs. 68.6 percent fusion rate at 6 months
in P030034 (cervical spine), an 85 percent vs. 75 percent clinical
success shown in P850022/S009 (lumbar spine), and a 67 percent vs. 43
percent fusion rate at 9 months in P910066/S011 (lumbar spine).
Further, FDA performed a literature review to evaluate data related
to non-invasive bone growth stimulator devices, including studies up to
the date of the 2006 Panel, as well as any new clinical information
published since the 2006 Panel.
Literature published at the time of the 2006 Panel includes a 1953
seminal paper on the use of electrical signals to stimulate bone
formation by Yasuda, that reported bone formation in rabbits exposed to
direct current (DC) stimulation (Ref. 12). In the following decades,
other researchers expanded on this finding in animal and clinical
models. In a canine study, a DC stimulation was shown to cause complete
ossification of the femoral medullary canal (Ref. 13). The first
clinical case report demonstrated that electrical stimulation could
treat a non-union fracture (Ref. 14). An early publication regarding
the effects of DC stimulation on spinal fusion was published by Dwyer
(Ref. 15). Another early clinical study published by Becker, et al.
showed successful fracture fusion with a success rate of 77 percent
(Ref. 16).
In the 1990s and early 2000s, several literature articles were
identified assessing the effects of non-invasive bone growth
stimulators on various anatomic locations. These studies generally
included various therapeutic modalities (magnitude, frequency,
duration, etc.) and demonstrated varying results regarding the efficacy
of these treatments. In two studies of PEMF devices, Basset and Schink-
Ascani (Ref. 17) found a 72 percent fusion rate in patients with
congenital pseudarthrosis of the tibia, and in a study of non-unions of
the scaphoid, Adams, et al. (Ref. 18) reported a fusion rate of 69
percent, as a followup to an earlier study that found a fusion rate of
80 percent. When looking at the rate of compliance of PEMF devices as a
factor of effectiveness, Garland, et al. (Ref. 19) found that fusion
rates ranged from 35.7 percent to 80 percent, depending on how often
the devices were used. In studies of CC devices, fusion rates in long
bones varied from 60 percent (Ref. 20), 68.8 percent (Ref. 21), and
72.7 percent (Ref. 22), to no difference between treatment and a
placebo-treated group in a study by Fourie and Bowerbank (Ref. 23).
While there was a large range of observed efficacies, there was no
reporting of treatment-related adverse events. These reported
variabilities in efficacy and low adverse event rates were consistent
with the findings by the 2006 Panel.
FDA performed a systematic review of published literature to
identify any new clinical findings since the 2006 Panel. FDA identified
14 papers that included a combination of retrospective and prospective
studies. For studies that assessed medical or insurance claims
databases, radiographs were not always available to determine actual
fusion. Instead, results were presented in terms of healing rate based
on patient records or reported outcomes. When radiographs were
available and analyzed to assess union, results were reported as fusion
rate.
Phillips, et al. (Ref. 24) looked at registry data of 2,370
subjects who were treated with OL1000 (DJO), a CMF device, at various
fracture sites and reported an average healing rate of 75.1 percent
(ranging from 57.2 percent in the humerus to 89.7 percent in the finger
phalanx). DeVries, et al. (Ref. 25) also evaluated the OL1000 device in
a retrospective analysis of 144 subjects, finding a fusion rate of 57.1
percent in tibiotalocalcaneal fusions of the ankle.
With respect to LIPUS, Zura, et al. (Refs. 26 and 27) published two
papers evaluating subjects in the Exogen (Bioventus) Post Market
Registry. One of the studies assessed how various patient risk factors
affected healing rate in 4,190 subjects. The study demonstrated an
overall healing rate of 95.7 percent, and in another single arm study
of 767 subjects, showed a healing rate varying from 81.8 percent to
87.9 percent depending on fracture site. Nolte, et al. (Ref. 28)
evaluated the Exogen registry in conjunction with a medical claims
database to examine metatarsal fractures and reported a healing rate of
97.4 percent overall, while Elvey, et al. (Ref. 29) evaluated 26 cases
with use of Exogen in hand and wrist non-unions, and found a fusion
rate of 54 percent to 58 percent. In two smaller studies of the Exogen
device, Majeed, et al. (Ref. 30) and Mizra, et al. (Ref. 31) both
evaluated foot and ankle fractures and found 78.7 percent and 67
percent fusion rate in a 47 and 18 patient study, respectively. Biglari
(Ref. 32) also performed an observational study using the Exogen device
and found a much lower fusion rate of 32.8 percent in 60 subjects
having existing non-unions of various long bones.
For PEMF devices, a retrospective study by Coric, et al. (Ref. 33)
on the effects of the CervicalStim (Orthofix) device on 593 subjects
showed a 73.2 percent fusion rate in the cervical spine at 6 months. In
a single arm prospective study by Assiotis, et al. (Ref. 34), a 77.3
percent fusion rate in the tibia was demonstrated with use of the
Physiostim (Orthofix). Murray and Pethica (Ref. 35) performed a 1,382-
subject retrospective study of use of the EBI device (Zimmer Biomet)
for non-unions of the scaphoid, tibia, and fibula, and while an
assessment of healing rates was not performed, the data showed
reduction in time to healing between 35 percent and 40 percent when the
device was used as prescribed.
In addition, two randomized control studies on PEMF devices were
conducted by Foley, et al. (Ref. 36) and by Streit, et al. (Ref. 37).
Foley evaluated 323 subjects using the Orthofix CervicalStim device in
cervical fusion and found an 83.6 percent fusion rate in the treatment
group compared to a 68.6 percent fusion rate in the control group, with
no difference in pain scores or adverse events between groups. Streit,
et al. performed a small, eight subject clinical study using the EBI
device to treat non-unions of the fifth metatarsal and found the time
to fusion was reduced on average from 14.7 weeks to 8.9 weeks with the
use of the device.
In summary, FDA's literature review resulted in findings that are
consistent with available clinical data from PMA submissions. These
studies suggest that there are probable benefits to the use of these
devices; however, differences in methodology, including differences in
devices used, treatment waveform and frequency, patient populations, as
well as anatomic location, could have had significant effects on
reported device effectiveness, which ranged from 32.8 percent to 97.4
percent. Regarding safety, the findings from these studies demonstrate
that the devices are relatively safe as the adverse event profile
associated with these devices using various modalities was similar to
controls. Overall, the studies involved 10,566 subjects (including
control subjects), with only a single report of a serious adverse event
(Biglari, Ref. 32); however, a direct link to the use of the device
could not be established for this event.
[[Page 49992]]
Further, a search of FDA's MDR database was conducted to identify
all adverse events submitted to FDA up to October 31, 2019, for devices
approved under product codes LOF and LPQ. The results of the identified
reports are consistent with the risk profiles identified in both PMA
applications and literature that were reviewed. FDA's search yielded a
total of 270 unique MDRs. The most frequently reported events were
categorized as ``skin reaction/issue'' (n = 187) followed by ``pain''
(n = 59) and ``device functional issue'' (n = 21). A review of the
adverse events regarding skin reactions found that a majority were due
to irritation from the electrode adhesive or ultrasound gel used. There
was no apparent difference in risk profile across the various device
modalities, though the risk of skin irritation was primarily observed
in the skin-contacting devices (due to the electrodes in the CC device
and the gel in the LIPUS device). For cases where followup was
described, patients recovered when treatment was discontinued. In
addition, 11 reports of ``mass/tumor'' were identified; however, the
nature of the relationship between the mass/tumor to the device was
unrelated or unclear. Based upon FDA's assessment of other systematic
reviews of these devices, no other reports of mass/tumors have been
identified (Refs. 38 to 42).
Finally, a search of FDA's Medical Device Recall database was
conducted. No recalls were found when searching the database for
devices under product code LOF. Two class 2 recalls were reported for
devices under product code LPQ; specifically, there was a recall for
the Exogen Express Bone Healing System and a recall for the Exogen
4000+ Ultrasound Bone Healing System. Both were posted on August 4,
2009, and initiated by the manufacturer because of problems with the
transducer, which may have resulted in a reduced ultrasound output.
These recalls were terminated on November 18, 2010. These recall events
reflect the risks to health identified in section V, and FDA believes
the special controls proposed, in addition to general controls, can
effectively mitigate the risks identified.
VIII. Proposed Special Controls
Table 1 outlines the risks to health identified in section V and
the corresponding mitigation measures proposed to reasonably assure
safety and effectiveness, which are discussed in more detail below.
Table 1--Risks to Health and Mitigation Measures for Non-Invasive Bone
Growth Stimulators
------------------------------------------------------------------------
Identified risk to health Mitigation measures
------------------------------------------------------------------------
Failure or delay of osteogenesis....... Clinical performance data.
Non-clinical performance
testing.
Software verification,
validation, and hazard
analysis.
Labeling.
Burn................................... Non-clinical performance
testing.
Electrical safety testing.
Labeling.
Electrical shock....................... Electrical safety testing.
Labeling.
Electromagnetic interference........... Electromagnetic compatibility
(EMC) testing.
Labeling.
Adverse tissue reaction................ Biocompatibility evaluation.
Labeling.
Adverse interaction with internal/ Labeling.
external fixation devices.
Adverse biological effects............. Non-clinical performance
testing.
Software verification,
validation, and hazard
analysis.
------------------------------------------------------------------------
The risk of failure or delay of osteogenesis is clinically
significant. To mitigate this risk, FDA proposes that manufacturers
provide clinical performance data to demonstrate that the device yields
positive outcomes (e.g., fusion of the non-union) in accordance with
its intended use. Further, FDA proposes non-clinical performance
testing to demonstrate that the device performs as intended under
anticipated conditions of use to achieve the identified successful
clinical performance characteristics. This would include verification
and validation of critical performance characteristics, including
characterization of the designed outputs of the device as well as the
outputs that are delivered to the patient, thermal safety and
reliability testing, reliability testing consistent with the expected
device use-life, and validation that signal characteristics are within
safe physiologic limits. Also, FDA proposes appropriate software
verification, validation, and hazard analysis to ensure that any device
software performs as intended. Lastly, FDA proposes labeling to provide
appropriate instructions (e.g., duration, frequency of use) to the end
user.
To mitigate the risk of skin burns, FDA proposes non-clinical
performance testing of the device to verify and validate critical
performance characteristics, demonstrate thermal safety and
reliability, validate that signal characteristics are within safe
physiologic limits, and demonstrate reliability of the device
consistent with its expected use-life. FDA also proposes electrical
safety testing to minimize the risk of thermal burns to the patient,
and specific instructions regarding proper usage and specific warnings
associated with the risk of burns.
To mitigate electrical shocks, FDA proposes electrical safety
testing to minimize the risk of shock to the patient. Furthermore, FDA
proposes labeling provisions, including instructions on appropriate
usage and maintenance, and specific warnings regarding electrical
shock.
To mitigate electromagnetic interference, FDA proposes
electromagnetic compatibility testing and labeling to minimize the risk
of adverse interaction with other electronic devices such as implanted
electronic devices.
To mitigate the risk of adverse tissue reactions, FDA proposes a
biocompatibility evaluation to ensure that the materials used in
patient-contacting components of the device are safe for skin contact
and labeling that includes warnings against use on compromised skin or
when there are known sensitivities, as well as instructions on
appropriate cleaning of any reusable components.
To mitigate the risk of adverse interaction with internal/external
fixation devices, FDA proposes labeling,
[[Page 49993]]
specifically inclusion of appropriate warnings for patients with
implanted internal/external devices.
To mitigate the risk of adverse biologic effects, FDA proposes non-
clinical performance testing to verify and validate critical
performance characteristics of the device, demonstrate thermal safety
and reliability, validate safety of the signal by reference to known
biological safety limits, and demonstrate reliability of the device
over the expected use-life. Furthermore, FDA proposes software
verification, validation, and hazard analysis.
If this reclassification is finalized, non-invasive bone growth
stimulators will be reclassified into class II and would be subject to
premarket notification (510(k)) requirements under Sec. 807.81. As
discussed below, the intent is for the reclassification to be codified
in 21 CFR 890.5870. Firms submitting a 510(k) for non-invasive bone
growth stimulators will be required to comply with the particular
mitigation measures set forth in the special controls. Adherence to the
special controls, in addition to the general controls, is necessary to
provide a reasonable assurance of the safety and effectiveness of these
devices.
IX. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this proposed order contains no new
collections of information. Therefore, clearance by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(PRA) (44 U.S.C. 3501-3521) is not required. This proposed order refers
to previously approved FDA collections of information. These
collections of information are subject to review by OMB under the PRA.
The collections of information in part 807, subpart E have been
approved under OMB control number 0910-0120; the collections of
information in 21 CFR part 814, subparts A through E, have been
approved under OMB control number 0910-0231; and the collections of
information under 21 CFR part 801 have been approved under OMB control
number 0910-0485.
XI. Codification of Orders
Under section 513(f)(3) of the FD&C Act, FDA may issue final orders
to reclassify devices. FDA will continue to codify classifications and
reclassifications in the Code of Federal Regulations (CFR). Changes
resulting from final orders will appear in the CFR as newly codified
orders. Therefore, under section 513(f)(3), in the proposed order, we
are proposing to codify non-invasive bone growth stimulators in the new
21 CFR 890.5870, under which non-invasive bone growth stimulators would
be reclassified from class III to class II.
XII. Proposed Effective Date
FDA proposes that any final order based on this proposal become
effective 30 days after the date of its publication in the Federal
Register.
XIII. References
The following references marked with an asterisk (*) are on display
at the Dockets Management Staff (see ADDRESSES) and are available for
viewing by interested persons between 9 a.m. and 4 p.m., Monday through
Friday; they also are available electronically at https://www.regulations.gov. References without asterisks are not on public
display at https://www.regulations.gov because they have copyright
restriction. Some may be available at the website address, if listed.
References without asterisks are available for viewing only at the
Dockets Management Staff. FDA has verified the website addresses, as of
the date this document publishes in the Federal Register, but websites
are subject to change over time.
1. *Guidance on Section 216 of the Food and Drug Administration
Modernization Act of 1997, available at https://www.fda.gov/regulatory-information/search-fda-guidance-documents/guidance-section-216-food-and-drug-administration-modernization-act-1997-guidance-industry-and-fda.
2. *P790002 Approval available at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?id=P790002.
3. *RS Medical Corporation Reclassification Petition, available
at https://wayback.archive-it.org/7993/20170405072021/https://www.fda.gov/ohrms/dockets/ac/06/briefing/2006-4224b1-06-TabB-RSMEDICAL-Petition.pdf.
4. *FDA's Orthopaedic and Rehabilitation Devices Panel
transcript and other meeting materials for the June 2, 2006, meeting
are available at: https://wayback.archive-it.org/7993/20170403222246/https://www.fda.gov/ohrms/dockets/ac/cdrh06.html#orthopaedic.
5. *Letter from RS Medical requesting withdrawal of petition,
available at https://www.regulations.gov/document?D=FDA-2005-P-0052-0007.
6. *FDA letter granting RS Medical's withdrawal request,
available at https://www.regulations.gov/document?D=FDA-2005-P-0052-0006.
7. *First Cohort of Results of the 2014-2015 Strategic Priority:
Strike the Right Balance Between Premarket and Postmarket Data
Collection (April 2015), available at https://www.fda.gov/media/91437/download.
8. *Second and Final Cohort of Results of the 2014-2015
Strategic Priority: Strike the Right Balance Between Premarket and
Postmarket Data Collection (August 2016), available at https://www.fda.gov/media/99822/download.
9. *P030034 Summary of Safety and Effectiveness, available at
https://www.accessdata.fda.gov/cdrh_docs/pdf3/P030034B.pdf.
10. *P850022/S009 Summary of Safety and Effectiveness, available
at https://www.accessdata.fda.gov/cdrh_docs/pdf/P850022S009B.pdf.
11. *P910066/S011 Summary of Safety and Effectiveness, available
at https://www.accessdata.fda.gov/cdrh_docs/pdf/P910066S011B.pdf.
12. Yasuda, I., ``The Classic: Fundamental Aspects of Fracture
Treatment.'' J. Kyoto Med. Soc., 4:395-406, 1953.
13. Bassett, C.A., R.J. Pawluk, and R.O. Becker, ``Effects of
Electric Currents on Bone In Vivo.'' Nature, 204: 652-654, 1964.
14. Friedenberg, Z.B., M.C. Harlow, and C.T. Brighton, ``Healing
of Nonunion of the Medial Malleolus by Means of Direct Current: A
Case Report.'' The Journal of Trauma, 11(10):883-885, 1971.
15. Dwyer, A.F., ``The Use of Electrical Current Stimulation in
Spinal Fusion.'' The Orthopedic Clinics of North America, 6(1):265-
273, 1975.
16. Becker, R.O., J.A. Spadaro, and A.A. Marino, ``Clinical
Experiences With Low Direct Current Stimulation of Bone Growth.''
Clinical Orthopaedics and Related Research, 124:75-83, 1977.
17. Bassett, C.A. and M. Schink-Ascani, ``Long-Term Pulsed
Electromagnetic Field (PEMF) Results in Congenital Pseudarthrosis.''
Calcified Tissue International, 49(3):216-220, 1991.
18. Adams, B.D., G.K. Frykman, and J. Taleisnik, ``Treatment of
Scaphoid Nonunion With Casting and Pulsed Electromagnetic Fields: A
Study Continuation.'' The Journal of Hand Surgery, 17(5):910-914,
1992.
19. Garland, D.E., B. Moses, and W. Salyer, ``Long-Term Follow-
up of Fracture Nonunions Treated With PEMFs.'' Contemporary
Orthopaedics, 22(3):295-302, 1991.
20. Scott, G. and J.B. King, ``A Prospective, Double-Blind Trial
of Electrical Capacitive Coupling in the Treatment of Non-Union of
Long Bones.'' The Journal of Bone and Joint Surgery. American
volume, 76(6):820-826, 1994.
21. Abeed, R.I., M. Naseer, and E.W. Abel, ``Capacitively
Coupled Electrical Stimulation Treatment: Results from Patients With
Failed Long Bone Fracture Unions.'' Journal of Orthopaedic Trauma,
12(7):510-513, 1998.
22. Zamora-Navas, P., A. Borras Verdera, R. Antelo Lorenzo, et
al., ``Electrical Stimulation of Bone Nonunion With the
[[Page 49994]]
Presence of a Gap.'' Acta Orthopaedica Belgica, 61(3):169-176, 1995.
23. Fourie, J.A. and P. Bowerbank, ``Stimulation of Bone Healing
in New Fractures of the Tibial Shaft Using Interferential
Currents.'' Physiotherapy Research International, 2(4):255-268,
1997.
24. Philips, M., J. Baumhauer, S. Sprague, et al., ``Use of
Combined Magnetic Field Treatment for Fracture Nonunion.'' Journal
of Long Term Effects of Medical Implants, 26(3):277-284, 2016.
25. DeVries, J.G., G.C. Berlet, and C.F. Hyer, ``Union Rate of
Tibiotalocalcaneal Nails With Internal or External Bone
Stimulation.'' Foot and Ankle International, 33(11):969-978, 2012.
26. Zura, R., S. Mehta, G.J. Della Rocca, et al., ``A Cohort
Study of 4,190 Patients Treated With Low-Intensity Pulsed Ultrasound
(LIPUS): Findings in the Elderly Versus All Patients.'' BioMed
Central Musculoskeletal Disorders, 16:45, 2015.
27. Zura, R., G.J. Della Rocca, S. Mehta, et al., ``Treatment of
Chronic (>1 Year) Fracture Nonunion: Heal Rate in a Cohort of 767
Patients Treated With Low-Intensity Pulsed Ultrasound (LIPUS).''
Injury, 46(10):2036-2041, 2015.
28. Nolte, P., R. Anderson, E. Strauss, et al., ``Heal Rate of
Metatarsal Fractures: A Propensity-Matching Study of Patients
Treated With Low-Intensity Pulsed Ultrasound (LIPUS) vs. Surgical
and Other Treatments.'' Injury, 47(11):2584-2590, 2016.
29. Elvey, M.H., R. Miller, K.S. Khor, et al., ``The Use of Low-
Intensity Pulsed Ultrasound in Hand and Wrist Nonunions.'' Journal
of Plastic Surgery and Hand Surgery, 1-6, 2019 (published online
ahead of print, November 26, 2019).
30. Majeed, H., T. Karim, J. Davenport, et al., ``Clinical and
Patient-Reported Outcomes Following Low Intensity Pulsed Ultrasound
(LIPUS, Exogen) for Established Post-Traumatic and Post-Surgical
Nonunion in the Foot and Ankle.'' Foot and Ankle Surgery, S1268-
7731(19)30072-4, 2019 (published online ahead of print, May 19,
2019).
31. Mizra, Y.H., K.H. Teoh, D. Golding, et al., ``Is There a
Role for Low Intensity Pulsed Ultrasound (LIPUS) in Delayed or
Nonunion Following Arthrodesis in Foot and Ankle Surgery?'' Foot and
Ankle Surgery, 25(6):842-848, 2019.
32. Biglari, B., T.M. Yildirim, T. Swing, et al., ``Failed
Treatment of Long Bone Nonunions With Low Intensity Pulsed
Ultrasound.'' Archives of Orthopaedic and Trauma Surgery,
136(8):1121-1134, 2016.
33. Coric, D., D.E. Bullard, V.V. Patel, et al., ``Pulsed
Electromagnetic Field Stimulation May Improve Fusion Rates in
Cervical Arthrodesis in High-Risk Populations.'' Bone and Joint
Research, 7(2):124-130, 2018.
34. Assiotis, A., N.P. Sachinis, and B.E. Chalidis, ``Pulsed
Electromagnetic Fields for the Treatment of Tibial Delayed Unions
and Nonunions. A Prospective Clinical Study and Review of the
Literature.'' Journal of Orthopaedic Surgery and Research, 7:24,
2012.
35. Murray, H.B. and B.A. Pethica, ``A Follow-up Study of the
In-Practice Results of Pulsed Electromagnetic Field Therapy in the
Management of Nonunion Fractures.'' Orthopaedic Research and
Reviews, 8:67-72, 2016.
36. Foley, K.T., T.E. Mroz, P.M. Arnold, et al., ``Randomized,
Prospective, and Controlled Clinical Trial of Pulsed Electromagnetic
Field Stimulation for Cervical Fusion.'' Spine Journal, 8(3):436-
442, 2008.
37. Streit, A., B.C. Watson, J.D. Granata, et al., ``Effect on
Clinical Outcome and Growth Factor Synthesis With Adjunctive Use of
Pulsed Electromagnetic Fields for Fifth Metatarsal Nonunion
Fracture: A Double-Blind Randomized Study.'' Foot and Ankle
International, 37(9):919-923, 2016.
38. Aleem, I.S., I. Aleem, N. Evaniew, et al., ``Efficacy of
Electrical Stimulators for Bone Healing: A Meta-Analysis of
Randomized Sham-Controlled Trials.'' Science Reports, 6:31724, 2016.
39. Behrens, S.B., M.E. Deren, and K O. Monchik, ``A Review of
Bone Growth Stimulation for Fracture Treatment.'' Current Orthopedic
Practice, 24(1):84-91, 2013.
40. Griffin, X.L., M.L. Costa, N. Parsons, et al.,
``Electromagnetic Field Stimulation for Treating Delayed Union or
Non-Union of Long Bone Fractures in Adults.'' The Cochrane Database
of Systematic Reviews, 4:CD008471, 2011.
41. Griffin, X.L., N. Smith, N. Parsons, et al., ``Ultrasound
and Shockwave Therapy for Acute Fractures in Adults.'' The Cochrane
Database of Systematic Reviews, 2:CD008579, 2012.
42. Hannemann, P.F., E.H.H. Mommers, J.P.M. Schots, et al.,
``The Effects of Low-Intensity Pulsed Ultrasound and Pulsed
Electromagnetic Fields Bone Growth Stimulation in Acute Fractures: A
Systematic Review and Meta-Analysis of Randomized Controlled
Trials.'' Archives of Orthopaedic and Trauma Surgery, 134(8):1093-
1106, 2014.
List of Subjects in 21 CFR Part 890
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 890 be amended as follows:
PART 890--PHYSICAL MEDICINE DEVICES
0
1. The authority citation for part 890 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 890.5870 to subpart F to read as follows:
Sec. 890.5870 Non-invasive bone growth stimulator.
(a) Identification. A non-invasive bone growth stimulator provides
stimulation through electrical, magnetic, or ultrasonic fields. The
device is for prescription use and is intended to be used externally to
promote osteogenesis as an adjunct to primary treatments for fracture
fixation or spinal fusion.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Clinical performance data must support the intended use of the
device.
(2) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use. The
following must be provided:
(i) Verification and validation of critical performance
characteristics of the device, including characterization of the
designed outputs of the device as well as the outputs that are
delivered to the patient.
(ii) Thermal safety and thermal reliability testing.
(iii) Validation that signal characteristics are within safe
physiologic limits.
(iv) Reliability testing consistent with the expected use-life of
the device.
(3) Patient-contacting components of the device must be
demonstrated to be biocompatible.
(4) Performance data must demonstrate the electrical safety and
electromagnetic compatibility of the device.
(5) Appropriate software verification, validation, and hazard
analysis must be performed.
(6) Labeling for the device must include the following:
(i) Warning against use on compromised skin or when there are known
sensitivities;
(ii) Appropriate warnings for patients with implanted medical
devices;
(iii) A detailed summary of the clinical testing, which includes
the clinical outcomes associated with the use of the device, and a
summary of adverse events and complications that occurred with the
device;
(iv) A clear description of the device;
(v) Instructions on appropriate usage, duration, and frequency of
use;
(vi) Instructions for maintenance and safe disposal;
(vii) Instructions for appropriate cleaning of any reusable
components;
(viii) Specific warnings regarding user burns, electrical shock,
and skin irritation; and
(ix) The risks and benefits associated with use of the device.
Dated: August 4, 2020.
Lauren K. Roth,
Associate Commissioner for Policy.
[FR Doc. 2020-17543 Filed 8-14-20; 8:45 am]
BILLING CODE 4164-01-P