Medical Devices; Technical Amendments, 18439-18444 [2020-06354]
Download as PDF
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
interim measures with respect to this
court order. See https://
www.bis.doc.gov/index.php/
component/docman/?task=doc_
download&gid=2535. For additional
information about the court ordered
injunction pertaining to revisions to the
U.S. https://www.pmddtc.state.gov/
ddtc_public?id=ddtc_public_portal_
news_and_events&timeframe=week.
Dated: March 16, 2020.
Matthew S. Borman,
Deputy Assistant Secretary for Export
Administration.
[FR Doc. 2020–05934 Filed 4–1–20; 8:45 am]
BILLING CODE 3510–33–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Parts 5, 801, 803, 807, 814, 820,
821, 822, 830, 860, 884, 900, and 1002
[Docket No. FDA–2020–N–0011]
Medical Devices; Technical
Amendments
AGENCY:
Food and Drug Administration,
HHS.
Final rule; technical
amendments.
ACTION:
The Food and Drug
Administration (FDA, the Agency, or
we) is amending its medical device
regulations. These revisions are
necessary to reflect changes to the
Agency’s Center for Devices and
Radiological Health’s organizational
structure, including the reorganization
of its offices. The revisions replace
references to the obsolete offices and
positions with the current information,
update the physical addresses for such
offices, and correct inaccurate citations.
In addition, as part of this effort we
made other editorial non-substantive
changes to correct other addresses,
references, and citations, as appropriate.
The rule does not impose any new
regulatory requirements on affected
parties. This action is editorial in nature
and is intended to improve the accuracy
of the Agency’s regulations.
DATES: This rule is effective April 1,
2020.
SUMMARY:
jbell on DSKJLSW7X2PROD with RULES
FOR FURTHER INFORMATION CONTACT:
Madhusoodana Nambiar, Office of
Policy, Center for Devices and
Radiological Health, 10903 New
Hampshire Ave., Bldg. 66, Rm. 5519,
Silver Spring, MD 20993–0002, 301–
796–5837.
SUPPLEMENTARY INFORMATION:
VerDate Sep<11>2014
18:44 Apr 01, 2020
Jkt 250001
I. Background
The FDA Center for Devices and
Radiological Health (CDRH) has
reorganized (84 FR 22854, May 20,
2019) to create an agile infrastructure
that can adapt to future organizational,
regulatory, and scientific needs. The
goal of this change is to implement more
efficient, consistent work processes
across CDRH that better support and
advance CDRH’s public health mission
and vision. The reorganization will
integrate CDRH’s premarket and
postmarket program functions along
product lines, allowing experts to
leverage their knowledge to optimize
decision making across the product life
cycle. Implementation took a phased
approach starting on March 18, 2019,
and was completed on September 30,
2019.
Historically, CDRH has been
organized according to the stage of the
product’s life cycle, e.g., premarket
review, postmarket surveillance, and
compliance, rather than by the type of
product regulated. The reorganization
integrates these functions by product
type within the Office of Product
Evaluation and Quality (OPEQ). OPEQ
was formed by combining the Office of
Compliance, the Office of Device
Evaluation, the Office of Surveillance
and Biometrics, and the Office of In
Vitro Diagnostics and Radiological
Health into one super office focused on
a Total Product Lifecycle approach to
medical device oversight. Within OPEQ,
there are offices divided by product
type, referred to as Offices of Health
Technology (OHT), as well as crosscutting offices focusing on specific
policy and programmatic needs
including the Office of Regulatory
Programs and the Office of Clinical
Evidence and Analysis. In addition, the
reorganization established the Office of
Policy, which includes two teams, the
Guidance, Legislation and Special
Projects Team and the Regulatory
Documents and Special Projects Team,
with no changes in the functions for
CDRH Policy. The reorganization also
established the Office of Strategic
Partnerships and Technology
Innovation (OST), which combined the
Science and Strategic Partnerships,
Digital Health, Health Informatics and
Innovation teams. There are no changes
in functions within the different OST
teams. CDRH reorganization also
realigned Management Services within
the Center to ensure administrative
functions in CDRH are optimally
aligned, structured, and deliver
excellent service. The reorganization
streamlined the Center’s communication
functions, by combining the internal
PO 00000
Frm 00027
Fmt 4700
Sfmt 4700
18439
and external communication functions,
including CDRH Executive Secretary
and Speaker Liaison, into the renamed
Division of Communication in the
Office of Communication and
Education, and created an Internal
Communication Branch. The structure
of the Office of Science and Engineering
Laboratories remains unchanged.
As part of this effort, we are also
making other editorial non-substantive
changes to correct other addresses,
references, and citations, as appropriate.
II. Description of the Technical
Amendments
The regulations specified in this rule
have been revised to replace all
references to ‘‘Office of Device
Evaluation’’, ‘‘Office of Compliance’’,
‘‘Office of Surveillance and Biometrics’’
with ‘‘Office of Product Evaluation and
Quality,’’ and where, appropriate, we
have used the term ‘‘Office,’’
‘‘Division,’’ ‘‘Team’’ or ‘‘Office of Health
Technology’’ to reflect the responsible
unit within CDRH. We have also made
conforming edits, as appropriate. In
addition, because of the reorganization,
the physical location for many of the
offices changed, and thus, we have
made non-substantive amendments to
ensure that the room numbers and
addresses reflect the current
information, and other changes as
necessary to update outdated addresses,
references, and citations in the
regulations pertaining to medical
devices. The rule does not impose any
new regulatory requirements on affected
parties. The amendments are editorial in
nature and should not be construed as
modifying any substantive standards or
requirements.
III. Notice and Public Comment
Publication of this document
constitutes final action of these changes
under the Administrative Procedure Act
(5 U.S.C. 553). Section 553 of the
Administrative Procedure Act (APA)
exempts ‘‘rules of agency organization,
procedure, or practice’’ from proposed
rulemaking (i.e., notice and comment
rulemaking). 5 U.S.C. 553(b)(3)(A).
Rules are also exempt when an agency
finds ‘‘good cause’’ that notice and
comment rulemaking procedures would
be ‘‘impracticable, unnecessary, or
contrary to the public interest.’’ 5 U.S.C.
553(b)(3)(B).
FDA has determined that this
rulemaking meets the notice and
comment exemption requirements in 5
U.S.C. 553(b)(3)(A) and (B). FDA’s
revisions make technical or nonsubstantive changes that pertain solely
to the CDRH reorganization and office
move and do not alter any substantive
E:\FR\FM\02APR1.SGM
02APR1
18440
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
standard. FDA does not believe public
comment is necessary for these minor
revisions.
The APA allows an effective date less
than 30 days after publication as
‘‘provided by the agency for good cause
found and published with the rule’’ (5
U.S.C. 553(d)(3)). A delayed effective
date is unnecessary in this case because
the amendments do not impose any new
regulatory requirements on affected
parties. As a result, affected parties do
not need time to prepare before the rule
takes effect. Therefore, FDA finds good
cause for the amendments to become
effective on the date of publication of
this action.
List of Subjects
1. The authority citation for part 5
continues to read as follows:
■
Authority: 5 U.S.C. 552; 21 U.S.C. 301–
397.
§ 5.1100
21 CFR Part 801
Labeling, Medical devices, Reporting
and recordkeeping requirements.
21 CFR Parts 803 and 821
Imports, Medical devices, Reporting
and recordkeeping requirements.
21 CFR Part 807
Confidential business information,
Imports, Medical devices, Reporting and
recordkeeping requirements.
21 CFR Part 814
Administrative practice and
procedure, Confidential business
information, Medical devices, Medical
research, Reporting and recordkeeping
requirements.
21 CFR Parts 820 and 822
Medical devices, Reporting and
recordkeeping requirements.
21 CFR Part 830
Administrative practice and
procedure, Labeling, Medical devices,
Reporting and recordkeeping
requirements.
21 CFR Part 860
Administrative practice and
procedure, Medical devices.
21 CFR Part 884
Medical Devices.
jbell on DSKJLSW7X2PROD with RULES
PART 5—ORGANIZATION
2. In § 5.1100, revise the entries for
‘‘Center for Devices and Radiological
Health’’ through ‘‘Office of In Vitro
Diagnostics and Radiological Health’’ to
read as follows:
Authority delegations (Government
agencies), Imports, Organization and
functions (Government agencies).
21 CFR Part 900
Electronic products, Health facilities,
Medical devices, Radiation protection,
Reporting and recordkeeping
requirements, X-rays.
16:30 Apr 01, 2020
Electronic products, Radiation
protection, Reporting and recordkeeping
requirements.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under the
authority delegated to the Commissioner
of Food and Drugs, 21 CFR parts 5, 801,
803, 807, 814, 820, 821, 822, 830, 860,
884, 900, and 1002 are amended as
follows:
■
21 CFR Part 5
VerDate Sep<11>2014
21 CFR Part 1002
Jkt 250001
Headquarters.
*
*
*
*
*
Center for Devices and Radiological
Health.12
Office of the Center Director.
Quality Management Staff.
Office of Communication and
Education.
Digital Communication Media Staff.
Program Management Operations
Staff.
Division of Communication.
External Communications Branch.
Web and Graphics Branch.
Internal Communication Branch.
Division of Employee Training and
Development.
Employee Development Branch.
Technology and Learning
Management Branch.
Division of Industry and Consumer
Education.
Premarket Programs Branch.
Postmarket and Consumer Branch.
Division of Information Disclosure.
Freedom of Information Branch A.
Freedom of Information Branch B.
Office of Management.
Planning and Program Analysis Staff.
Division of Acquisition Services.
Advanced Acquisitions.
Simplified Acquisitions.
Acquisition Planning Assistance.
Division of Workforce Management.
Recruitment.
Human Capital Management.
Special Programs.
Division of Management Services.
Travel and Conference Management.
Committee Management and
Planning.
Space and Facilities Management.
12 Mailing address: 10903 New Hampshire Ave.,
Bldg. 66, Silver Spring, MD 20993.
PO 00000
Frm 00028
Fmt 4700
Sfmt 4700
Division of Financial Management.
Budget Formulation.
Budget Execution.
Financial Accountability.
Office of Policy.
Guidance, Legislation and Special
Projects.
Regulatory Documents and Special
Projects.
Office of Product Evaluation and
Quality.
Quality and Analytics Staff.
Clinical and Scientific Policy Staff.
Strategic Initiatives Staff.
Regulation, Policy and Guidance
Staff.
Compliance and Quality Staff.
Operations Staff.
Office of Regulatory Programs.
Division of Regulatory Programs 1
(Division of Submission Support).
Division of Regulatory Programs 2
(Division of Establishment
Support).
Division of Regulatory Programs 3
(Division of Market Intelligence).
Office of Clinical Evidence and
Analysis.
Division of Clinical Evidence and
Analysis 1 (Division of Clinical
Science and Quality).
Division of Clinical Evidence and
Analysis 2 (Division of
Biostatistics).
Office of Health Technology 1 (OHT1:
Office of Ophthalmic, Anesthesia,
Respiratory, ENT and Dental
Devices).
Division of Health Technology 1A
(Division of Ophthalmic Devices).
Division of Health Technology 1B
(Division of Dental Devices).
Division of Health Technology 1C
(Division of ENT, Sleep Disordered
Breathing, Respiratory and
Anesthesia Devices).
Office of Health Technology 2 (OHT2:
Office of Cardiovascular Devices).
Division of Health Technology 2A
(Division of Cardiac
Electrophysiology, Diagnostics, and
Monitoring Devices).
Division of Health Technology 2B
(Division of Circulatory Support,
Structural and Vascular Devices).
Division of Health Technology 2C
(Division of Coronary and
Peripheral Interventional Devices).
Office of Health Technology 3 (OHT3:
Office of Gastrorenal, ObGyn,
General Hospital and Urology
Devices).
Division of Health Technology 3A
(Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices).
Division of Health Technology 3B
(Division of Reproductive,
Gynecology and Urology Devices).
Division of Health Technology 3C
E:\FR\FM\02APR1.SGM
02APR1
jbell on DSKJLSW7X2PROD with RULES
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
(Division of Drug Delivery and
General Hospital Devices, and
Human Factors).
Office of Health Technology 4 (OHT4:
Office of Surgical and Infection
Control Devices).
Division of Health Technology 4A
(Division of General Surgery
Devices).
Division of Health Technology 4B
(Division of Infection Control and
Plastic Surgery Devices).
Office of Health Technology 5 (OHT5:
Office of Neurological and Physical
Medicine Devices).
Division of Health Technology 5A
(Division of Neurological,
Neurointerventional and
Neurodiagnostic Devices).
Division of Health Technology 5B
(Division of Neuromodulation and
Physical Medicine Devices).
Office of Health Technology 6 (OHT6:
Office of Orthopedic Devices).
Division of Health Technology 6A
(Division of Joint Arthroplasty
Devices).
Division of Health Technology 6B
(Division of Spinal Devices).
Division of Health Technology 6C
(Division of Stereotaxic, Trauma
and Restorative, Devices).
Office of Health Technology 7 (OHT7:
Office of In Vitro Diagnostics and
Radiological Health).
Division of Program Operations and
Management.
Division of Chemistry and Toxicology
Devices.
Chemistry Branch.
Diabetes Branch.
Toxicology Branch.
Cardio-Renal Diagnostics Branch.
Division of Molecular Genetics and
Pathology.
Molecular Pathology and Cytology
Branch.
Molecular Genetics Branch.
Division of Immunology and
Hematology Devices.
Hematology Branch.
Immunology and Flow-Cytometry
Branch.
Division of Microbiology Devices.
Viral Respiratory and HPV Branch.
General Viral and Hepatitis Branch.
General Bacterial and Antimicrobial
Susceptibility Branch.
Bacterial Respiratory and Medical
Countermeasures Branch.
Division of Radiological Health.
Magnetic Resonance and Electronic
Products Branch.
Diagnostic X-Ray Systems Branch.
Nuclear Medicine and Radiation
Therapy Branch.
Mammography, Ultrasound and
Imaging Software Branch.
Division of Mammography Quality
VerDate Sep<11>2014
16:30 Apr 01, 2020
Jkt 250001
7. In § 801.57, revise paragraph (c)(2)
to read as follows:
Standards.
Office of Science and Engineering
Laboratories.
Immediate Office of the Director
Division of Applied Mechanics.
Division of Biomedical Physics.
Division of Biology, Chemistry and
Materials Science.
Division of Imaging, Diagnostics, and
Software Reliability.
Division of Administrative and
Laboratory Support.
Office of Strategic Partnerships and
Technology Innovation.
Innovation.
Division of All Hazards Response
Science and Strategic Partnerships.
Medical Device Development Tools.
Health of Women.
Pediatrics and Special Populations.
All Hazards Readiness Response and
Cybersecurity.
Patient Science and Engagement.
Partnerships to Advance Innovation
and Regulatory Science.
Science and Special Projects
Incubator.
Standards and Conformity
Assessment Program.
Division of Digital Health.
Operational Excellence.
Technical and Policy Leadership
Strategic Partnerships and
Initiatives 1.
Technical and Policy Leadership
Strategic Partnerships and
Initiatives 2.
Strategic Initiatives and Special
Projects.
Division of Technology and Data
Services.
Business and Transformation
Services.
Data Services.
Technology Services.
*
*
*
*
*
■
PART 801—LABELING
*
5. The authority citation for part 801
continues to read as follows:
■
Authority: 21 U.S.C. 321, 331, 351, 352,
360d, 360i, 360j, 371, 374.
6. In § 801.55, revise paragraph (b)(2)
to read as follows:
■
§ 801.55 Request for an exception from or
alternative to a unique device identifier
requirement.
*
*
*
*
*
(b) * * *
(2) In all other cases, by email to:
GUDIDSupport@fda.hhs.gov, or by
correspondence to: UDI Regulatory
Policy Support, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 32, Rm. 3293, Silver Spring,
MD 20993–0002.
*
*
*
*
*
PO 00000
Frm 00029
Fmt 4700
Sfmt 4700
18441
§ 801.57 Discontinuation of legacy FDA
identification numbers assigned to devices.
*
*
*
*
*
(c) * * *
(2) No later than September 24, 2014,
the labeler submits, and obtains FDA
approval of, a request for continued use
of the assigned labeler code. A request
for continued use of an assigned labeler
code must be submitted by email to:
GUDIDSupport@fda.hhs.gov, or by
correspondence to: UDI Regulatory
Policy Support, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 32, Rm. 3293, Silver Spring,
MD 20993–0002.
*
*
*
*
*
PART 803—MEDICAL DEVICE
REPORTING
8. The authority citation for part 803
continues to read as follows:
■
Authority: 21 U.S.C. 352, 360, 360i, 360j,
371, 374.
9. In § 803.11, revise paragraph (d) to
read as follows:
■
§ 803.11 What form should I use to submit
reports of individual adverse events and
where do I obtain these forms?
*
*
*
*
*
(d) Form FDA 3500A is available on
the internet at https://
www.accessdata.fda.gov/scripts/
medwatch/index.cfm.
■ 10. In § 803.19(b), revise the second
sentence to read as follows:
§ 803.19 Are there exemptions, variances,
or alternative forms of adverse event
reporting requirements?
*
*
*
*
(b) * * * You must submit the
request to us in writing at the following
address: MDR Exemption Requests,
Medical Device Report (MDR) Team,
Division of Regulatory Programs 3,
Office of Regulatory Programs, Office of
Product Evaluation and Quality, 10903
New Hampshire Ave., Bldg. 66,
Rm.1523, Silver Spring, MD 20993–
0002. * * *
*
*
*
*
*
■ 11. In § 803.21, revise paragraph (a) to
read as follows:
§ 803.21 Where can I find the reporting
codes for adverse events that I use with
medical device reports?
(a) The MedWatch Medical Device
Reporting Code Instruction Manual
contains adverse event codes for use
with Form FDA 3500A. You may obtain
the coding manual from FDA’s website
E:\FR\FM\02APR1.SGM
02APR1
18442
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
at: https://www.fda.gov/medicaldevices/mandatory-reportingrequirements-manufacturers-importersand-device-user-facilities/mdr-adverseevent-codes.
*
*
*
*
*
■ 12. In § 803.33, revise paragraph (a) to
read as follows:
§ 803.33 If I am a user facility, what must
I include when I submit an annual report?
(a) You must submit to us an annual
report on Form FDA 3419. You must
submit an annual report by January 1, of
each year. You may obtain this form on
the internet at: https://www.fda.gov/
media/72292/download.
*
*
*
*
*
PART 807—ESTABLISHMENT
REGISTRATION AND DEVICE LISTING
FOR MANUFACTURERS AND INITIAL
IMPORTERS OF DEVICES
13. The authority citation for part 807
continues to read as follows:
■
Authority: 21 U.S.C. 321, 331, 351, 352,
360, 360c, 360e, 360i, 360j, 360bbb-8b, 371,
374, 379k–1, 381, 393; 42 U.S.C. 264, 271.
14. In § 807.21, revise paragraph (b)
introductory text to read as follows:
■
§ 807.21 How to register establishments
and list devices.
*
*
*
*
*
(b) If the information under
§ 807.21(a) cannot be submitted
electronically, a waiver may be
requested. Waivers will be granted only
if use of electronic means is not
reasonable for the person requesting the
waiver. To request a waiver, applicants
must send a letter to the Imports and
Registration and Listing Team, Division
of Regulatory Programs 2, Office of
Regulatory Programs, Office of Product
Evaluation and Quality, Center for
Devices and Radiological Health, Food
and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm.1432,
Silver Spring, MD 20993–0002, that
includes the following information:
*
*
*
*
*
■ 15. In § 807.34, revise paragraph (a) to
read as follows:
jbell on DSKJLSW7X2PROD with RULES
§ 807.34 Summary of requirements for
owners or operators granted a waiver from
submitting required information
electronically.
(a) For initial registration and listing,
owners or operators who have been
granted a waiver from electronic filing
using the procedures set forth in
§ 807.21(b) must send a letter containing
all of the registration and listing
information described in §§ 807.22,
807.25 (and § 807.26 when such
information is requested by FDA), at the
VerDate Sep<11>2014
16:30 Apr 01, 2020
Jkt 250001
times described in § 807.22, to: The
Imports and Registration and Listing
Team, Division of Regulatory Programs
2, Office of Regulatory Programs, Office
of Product Evaluation and Quality,
Center for Devices and Radiological
Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 66,
Rm. 1432, Silver Spring, MD 20993–
0002.
*
*
*
*
*
■ 16. In § 807.37, revise paragraph (a) to
read as follows:
§ 807.37 Public availability of
establishment registration and device
listing information.
(a) Establishment registration and
device listing information is available
for public inspection in accordance with
section 510(f) of the Federal Food, Drug,
and Cosmetic Act and will be posted on
the FDA website, with the exception of
the information identified in paragraph
(b) of this section. Requests for
information by persons who do not have
access to the internet should be directed
to the Imports and Registration and
Listing Team, Division of Regulatory
Programs 2, Office of Regulatory
Programs, Office of Product Evaluation
and Quality, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm.1432, Silver Spring,
MD 20993–0002. In addition, there will
be available for inspection at each of the
Food and Drug Administration district
offices the same information for firms
within the geographical area of such
district offices. Upon request,
verification of a registration number or
location of a registered establishment
will be provided.
*
*
*
*
*
PART 814—PREMARKET APPROVAL
OF MEDICAL DEVICES
17. The authority citation for part 814
continues to read as follows:
■
Authority: 21 U.S.C. 351, 352, 353, 360,
360c–360j, 360bbb–8b, 371, 372, 373, 374,
375, 379, 379e, 379k–1, 381.
18. In § 814.42, revise paragraph (d)(2)
to read as follows:
■
§ 814.42
Filing a PMA.
*
*
*
*
*
(d) * * *
(2) Request in writing within 10
working days of the date of receipt of
the notice refusing to file the PMA, an
informal conference with the Director of
the associated Office of Health
Technology to review FDA’s decision
not to file the PMA. FDA will hold the
informal conference within 10 working
days of its receipt of the request and
PO 00000
Frm 00030
Fmt 4700
Sfmt 4700
will render its decision on filing within
5 working days after the informal
conference. If, after the informal
conference, FDA accepts the PMA for
filing, the date of filing will be the date
of the decision to accept the PMA for
filing. If FDA does not reverse its
decision not to file the PMA, the
applicant may request reconsideration
of the decision from the Director of the
Office of Product Evaluation and
Quality, the Director of the Center for
Biologics Evaluation and Research, or
the Director of the Center for Drug
Evaluation and Research, as applicable.
The Director’s decision will constitute
final administrative action for the
purpose of judicial review.
*
*
*
*
*
■ 19. In § 814.100, revise paragraph
(e)(2) to read as follows:
§ 814.100
Purpose and scope.
*
*
*
*
*
(e) * * *
(2) Submitting an HDE to the Office of
Product Evaluation and Quality (OPEQ),
Center for Devices and Radiological
Health (CDRH), the Center for Biologics
Evaluation and Research (CBER), or the
Center for Drug Evaluation and Research
(CDER), as applicable.
*
*
*
*
*
PART 820—QUALITY SYSTEM
REGULATION
20. The authority citation for part 820
continues to read as follows:
■
Authority: 21 U.S.C. 351, 352, 360, 360c,
360d, 360e, 360h, 360i, 360j, 360l, 371, 374,
381, 383; 42 U.S.C. 216, 262, 263a, 264.
21. In § 820.1, revise paragraph (e)(1)
to read as follows:
■
§ 820.1
Scope.
*
*
*
*
*
(e) * * * (1) Any person who wishes
to petition for an exemption or variance
from any device quality system
requirement is subject to the
requirements of section 520(f)(2) of the
Federal Food, Drug, and Cosmetic Act.
Petitions for an exemption or variance
shall be submitted according to the
procedures set forth in § 10.30 of this
chapter, the FDA’s administrative
procedures. For guidance on how to
proceed for a request for a variance,
contact Division of Regulatory Programs
2, Office of Regulatory Programs, Office
of Product Evaluation and Quality,
Center for Devices and Radiological
Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 66,
Rm. 1438, Silver Spring, MD 20993–
0002.
*
*
*
*
*
E:\FR\FM\02APR1.SGM
02APR1
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
§ 830.110 Application for accreditation as
an issuing agency.
PART 821—MEDICAL DEVICE
TRACKING REQUIREMENTS
22. The authority citation for part 821
continues to read as follows:
■
Authority: 21 U.S.C. 331, 351, 352, 360,
360e, 360h, 360i, 371, 374.
23. In § 821.2, revise paragraph (b)
introductory text to read as follows:
■
§ 821.2
Exemptions and variances.
*
*
*
*
*
(b) A request for an exemption or
variance shall be submitted in the form
of a petition under § 10.30 of this
chapter and shall comply with the
requirements set out therein, except that
a response shall be issued in 90 days.
The Director or Deputy Directors,
CDRH, or the Director of the Office of
Regulatory Program, CDRH, shall issue
responses to requests under this section.
The petition shall also contain the
following:
*
*
*
*
*
PART 822—POSTMARKET
SURVEILLANCE
24. The authority citation for part 822
continues to read as follows:
■
Authority: 21 U.S.C. 331, 352, 360i, 360l,
371, 374.
25. In § 822.7, revise paragraph (a)(1)
to read as follows:
■
§ 822.7 What should I do if I do not agree
that postmarket surveillance is
appropriate?
(a) * * *
(1) Requesting a meeting with the
individual who issued the order for
postmarket surveillance;
*
*
*
*
*
§ 830.320 Submission of unique device
identification information.
*
*
*
*
*
(c) * * * (1) A labeler may request a
waiver from electronic submission of
UDI data by submitting a letter
addressed to the appropriate Center
Director explaining why electronic
submission is not technologically
feasible; send the request by email to:
udi@fda.hhs.gov, or by correspondence
to: UDI Regulatory Policy Support,
Center for Devices and Radiological
Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 32,
Rm. 3293, Silver Spring, MD 20993–
0002.
*
*
*
*
*
(3) A labeler that has a waiver from
electronic submission of UDI data must
send a letter containing all of the
information required by § 830.310, as
well as any ancillary information
permitted to be submitted under
§ 830.340 that the labeler wishes to
submit, within the time permitted by
§ 830.330, addressed to: UDI Regulatory
Policy Support, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 32, Rm. 3293, Silver Spring,
MD 20993–0002.
Radiological Health, Office of Policy
Staff, 10903 New Hampshire Ave., Bldg.
66, Rm. 5445, Silver Spring, MD 20993–
0002; for devices regulated by the
Center for Biologics Evaluation and
Research, addressed to the Food and
Drug Administration, Center for
Biologics Evaluation and Research,
Document Control Center, 10903 New
Hampshire Ave., Bldg. 71, Rm. G112,
Silver Spring, MD 20993–0002; for
devices regulated by the Center for Drug
Evaluation and Research, addressed to
the Food and Drug Administration,
Center for Drug Evaluation and
Research, Central Document Control
Room, 5901–B Ammendale Rd.,
Beltsville, MD 20705–1266, as
applicable.
*
*
*
*
*
PART 884—OBSTETRICAL AND
GYNECOLOGICAL DEVICES
32. The authority citation for part 884
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
33. In § 884.5360, remove and reserve
paragraph (c).
■
PART 900—MAMMOGRAPHY
34. The authority citation for part 900
continues to read as follows:
■
Authority: 21 U.S.C. 360i, 360nn, 374(e);
42 U.S.C. 263b.
35. In § 900.3, revise paragraph (b)(1)
to read as follows:
■
§ 900.3 Application for approval as an
accreditation body.
PART 830—UNIQUE DEVICE
IDENTIFICATION
■
*
*
*
*
(b) * * *(1) An applicant seeking
initial FDA approval as an accreditation
body shall inform the Division of
Mammography Quality Standards,
Center for Devices and Radiological
Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 66,
Rm. 3621, Silver Spring, MD 20993,
Attn: Program Management Branch, of
its desire to be approved as an
accreditation body and of its requested
scope of authority.
*
*
*
*
*
■ 36. In § 900.15, revise paragraph
(d)(3)(i) to read as follows:
27. The authority citation for part 830
continues to read as follows:
§ 860.123 Reclassification petition:
Content and form.
§ 900.15 Appeals of adverse accreditation
or reaccreditation decisions that preclude
certification or recertification.
*
*
26. In § 822.22, revise paragraph (a)(1)
to read as follows:
■
§ 822.22 What recourse do I have if I do
not agree with your decision?
PART 860—MEDICAL DEVICE
CLASSIFICATION PROCEDURES
(a) * * *
(1) Requesting a meeting with the
individual who issued the order for
postmarket surveillance;
*
*
*
*
*
■
jbell on DSKJLSW7X2PROD with RULES
(a) * * *(1) An applicant seeking
initial FDA accreditation as an issuing
agency shall notify FDA of its desire to
be accredited by sending a notification
by email to: GUDIDSupport@
fda.hhs.gov, or by correspondence to:
UDI Regulatory Policy Support, Center
for Devices and Radiological Health,
Food and Drug Administration, 10903
New Hampshire Ave., Bldg. 32, Rm.
3293, Silver Spring, MD 20993–0002.
*
*
*
*
*
■ 29. In § 830.320, revise paragraphs
(c)(1) and (3) to read as follows:
18443
Authority: 21 U.S.C. 321, 331, 352, 353,
360, 360d, 360i, 360j, 371.
*
*
*
*
*
28. In § 830.110, revise paragraph
(a)(1) to read as follows:
■
VerDate Sep<11>2014
16:30 Apr 01, 2020
Jkt 250001
30. The authority citation for part 860
continues to read as follows:
■
Authority: 21 U.S.C. 360c, 360d, 360e,
360i, 360j, 371, 374.
31. In § 860.123, revise paragraph
(b)(1) to read as follows:
*
*
*
*
(b) * * *
(1) For devices regulated by the
Center for Devices and Radiological
Health, addressed to the Food and Drug
Administration, Center for Devices and
PO 00000
Frm 00031
Fmt 4700
Sfmt 4700
*
*
*
*
*
(d) * * *
(3) * * *
(i) A facility must request
reconsideration by DMQS within 60
days of the accreditation body’s adverse
E:\FR\FM\02APR1.SGM
02APR1
18444
Federal Register / Vol. 85, No. 64 / Thursday, April 2, 2020 / Rules and Regulations
appeals decision, at the following
address: Food and Drug Administration,
Center for Devices and Radiological
Health, Division of Mammography
Quality Standards, Attn: Facility
Accreditation Review Committee, 10903
New Hampshire Ave., Bldg. 66, Rm.
3621, Silver Spring, MD 20993–0002.
*
*
*
*
*
■ 37. In § 900.18, revise paragraph (c)
introductory text to read as follows:
Devices and Radiological Health,
Division of Mammography Quality
Standards, 10903 New Hampshire Ave.,
Bldg. 66, Rm. 3621, Silver Spring, MD
20993–0002.
*
*
*
*
*
Dated: March 23, 2020.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2020–06354 Filed 4–1–20; 8:45 am]
BILLING CODE 4164–01–P
§ 900.18 Alternative requirements for
§ 900.12 quality standards.
*
*
*
*
*
(c) Applications for approval of an
alternative standard. An application for
approval of an alternative standard or
for an amendment or extension of the
alternative standard shall be submitted
in an original and two copies to the
Food and Drug Administration, Center
for Devices and Radiological Health,
Director, Division of Mammography
Quality Standards, 10903 New
Hampshire Ave., Bldg. 66, Rm. 3621,
Silver Spring, MD 20993–0002. The
application for approval of an
alternative standard shall include the
following information:
*
*
*
*
*
■ 38. In § 900.21, revise paragraph (b)(1)
to read as follows:
§ 900.21 Application for approval as a
certification agency.
*
*
*
*
*
(b) * * * (1) An applicant seeking
FDA approval as a certification agency
shall inform the Food and Drug
Administration, Center for Devices and
Radiological Health, Director, Division
of Mammography Quality Standards,
Attn: Program Management Branch,
10903 New Hampshire Ave., Bldg. 66,
Rm. 3621, Silver Spring, MD 20993–
0002, in writing, of its desire to be
approved as a certification agency.
*
*
*
*
*
PART 1002—RECORDS AND
REPORTS
39. The authority citation for part
1002 continues to read as follows:
■
40. In § 1002.50, revise paragraph
(c)(3) to read as follows:
■
jbell on DSKJLSW7X2PROD with RULES
Special exemptions.
*
*
*
*
*
(c) * * *
(3) Such conditions as are deemed
necessary to protect the public health
and safety. Copies of exemptions shall
be available upon request from the Food
and Drug Administration, Center for
VerDate Sep<11>2014
16:30 Apr 01, 2020
Jkt 250001
Food and Drug Administration
21 CFR Parts 862 and 866
[Docket No. FDA–2020–N–0011]
Medical Devices; Technical
Amendment
AGENCY:
Food and Drug Administration,
HHS.
Final rule; technical
amendment.
ACTION:
The Food and Drug
Administration (FDA or Agency) is
amending certain medical device
regulations to accurately reflect the
devices exempted from premarket
notification (510(k)) as indicated in the
lists published on April 13, 2017, and
July 11, 2017. FDA published a final
amendment, final order in the Federal
Register of December 30, 2019 (‘‘Final
Order’’) codifying the two Federal
Register notices. The present revisions
are necessary to correct editorial errors
to ensure that the codified is consistent
with the exemptions in the Federal
Register notices. The rule does not
impose any new regulatory
requirements on affected parties. This
action is editorial in nature and is
intended to improve the accuracy of the
Agency’s regulations.
DATES: This rule is effective April 1,
2020.
SUMMARY:
FOR FURTHER INFORMATION CONTACT:
Authority: 21 U.S.C. 352, 360, 360i, 360j,
360hh–360ss, 371, 374.
§ 1002.50
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Ryan Lubert, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3574, Silver Spring,
MD 20993–0002, 240–402–6357.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with sections 510(l)(2)
and 510(m)(1)(A) of the Federal Food,
Drug, and Cosmetic Act (FD&C Act) (21
U.S.C. 360(l)(2) and 360(m)(1)(A)), FDA
issued two separate notices of final
determination exempting a list of class
PO 00000
Frm 00032
Fmt 4700
Sfmt 4700
I and II devices from section 510(k) of
the FD&C Act, respectively, subject to
certain limitations published in the
Federal Register April 13, 2017 (82 FR
17841) and July 11, 2017 (82 FR 31976).
The devices included in these lists were
exempt upon publication of the final
determination notices in the Federal
Register notices (see sections
510(l)(2)(A) and 510(m)(3) of the FD&C
Act). On December 30, 2019 (84 FR
71794), FDA issued an amendment,
final order, which amended the codified
for the classification regulations
implicated in the Federal Register
notices to reflect the exemptions and
limitations on exemptions in those
notices. This Final Order incorrectly
amended the codified for three device
types such that the exemption in the
current codified is inconsistent with the
scope of the device exemptions
described in the Federal Register
notices. Specifically, for the three
implicated device types, FDA indicated
in the Federal Register notices that a
device with a particular intended use
was exempt from the premarket
notification requirements in section
510(k) of the FD&C Act; however, the
codified currently indicates that the
entire device type is exempt from
section 510(k) of the FD&C Act, which
is not the case.
As such, FDA is amending the
codified for §§ 862.1345, 862.1775, and
866.2900 (21 CFR 862.1345, 862.1775,
and 866.2900) to be consistent with the
exemptions as stated in the Federal
Register notices. These amendments are
not substantive changes because the
Federal Register notices exempted the
affected devices from the section 510(k)
of the FD&C Act, but are intended to
correct the codified and to clarify which
devices under those classification
regulations are exempt from the
premarket notification requirements in
section 510(k) of the FD&C Act and
which device types remain subject to
such requirements.
II. Description of the Technical
Amendments
The regulations specified in this rule
have been revised to correct and clarify
the codified language of the regulations
specified in this technical amendment,
specifically §§ 862.1345, 862.1775, and
866.2900, to be consistent with the
exemptions as stated in the Federal
Register notices. FDA is making no
substantive changes to the following
regulations:
1. FDA is revising § 862.1345(b) by
replacing ‘‘The device is exempt from
the premarket notification procedures in
subpart E of part 807 of this chapter
subject to the limitations in § 862.9’’
E:\FR\FM\02APR1.SGM
02APR1
Agencies
[Federal Register Volume 85, Number 64 (Thursday, April 2, 2020)]
[Rules and Regulations]
[Pages 18439-18444]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2020-06354]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Parts 5, 801, 803, 807, 814, 820, 821, 822, 830, 860, 884,
900, and 1002
[Docket No. FDA-2020-N-0011]
Medical Devices; Technical Amendments
AGENCY: Food and Drug Administration, HHS.
ACTION: Final rule; technical amendments.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA, the Agency, or we) is
amending its medical device regulations. These revisions are necessary
to reflect changes to the Agency's Center for Devices and Radiological
Health's organizational structure, including the reorganization of its
offices. The revisions replace references to the obsolete offices and
positions with the current information, update the physical addresses
for such offices, and correct inaccurate citations. In addition, as
part of this effort we made other editorial non-substantive changes to
correct other addresses, references, and citations, as appropriate. The
rule does not impose any new regulatory requirements on affected
parties. This action is editorial in nature and is intended to improve
the accuracy of the Agency's regulations.
DATES: This rule is effective April 1, 2020.
FOR FURTHER INFORMATION CONTACT: Madhusoodana Nambiar, Office of
Policy, Center for Devices and Radiological Health, 10903 New Hampshire
Ave., Bldg. 66, Rm. 5519, Silver Spring, MD 20993-0002, 301-796-5837.
SUPPLEMENTARY INFORMATION:
I. Background
The FDA Center for Devices and Radiological Health (CDRH) has
reorganized (84 FR 22854, May 20, 2019) to create an agile
infrastructure that can adapt to future organizational, regulatory, and
scientific needs. The goal of this change is to implement more
efficient, consistent work processes across CDRH that better support
and advance CDRH's public health mission and vision. The reorganization
will integrate CDRH's premarket and postmarket program functions along
product lines, allowing experts to leverage their knowledge to optimize
decision making across the product life cycle. Implementation took a
phased approach starting on March 18, 2019, and was completed on
September 30, 2019.
Historically, CDRH has been organized according to the stage of the
product's life cycle, e.g., premarket review, postmarket surveillance,
and compliance, rather than by the type of product regulated. The
reorganization integrates these functions by product type within the
Office of Product Evaluation and Quality (OPEQ). OPEQ was formed by
combining the Office of Compliance, the Office of Device Evaluation,
the Office of Surveillance and Biometrics, and the Office of In Vitro
Diagnostics and Radiological Health into one super office focused on a
Total Product Lifecycle approach to medical device oversight. Within
OPEQ, there are offices divided by product type, referred to as Offices
of Health Technology (OHT), as well as cross-cutting offices focusing
on specific policy and programmatic needs including the Office of
Regulatory Programs and the Office of Clinical Evidence and Analysis.
In addition, the reorganization established the Office of Policy, which
includes two teams, the Guidance, Legislation and Special Projects Team
and the Regulatory Documents and Special Projects Team, with no changes
in the functions for CDRH Policy. The reorganization also established
the Office of Strategic Partnerships and Technology Innovation (OST),
which combined the Science and Strategic Partnerships, Digital Health,
Health Informatics and Innovation teams. There are no changes in
functions within the different OST teams. CDRH reorganization also
realigned Management Services within the Center to ensure
administrative functions in CDRH are optimally aligned, structured, and
deliver excellent service. The reorganization streamlined the Center's
communication functions, by combining the internal and external
communication functions, including CDRH Executive Secretary and Speaker
Liaison, into the renamed Division of Communication in the Office of
Communication and Education, and created an Internal Communication
Branch. The structure of the Office of Science and Engineering
Laboratories remains unchanged.
As part of this effort, we are also making other editorial non-
substantive changes to correct other addresses, references, and
citations, as appropriate.
II. Description of the Technical Amendments
The regulations specified in this rule have been revised to replace
all references to ``Office of Device Evaluation'', ``Office of
Compliance'', ``Office of Surveillance and Biometrics'' with ``Office
of Product Evaluation and Quality,'' and where, appropriate, we have
used the term ``Office,'' ``Division,'' ``Team'' or ``Office of Health
Technology'' to reflect the responsible unit within CDRH. We have also
made conforming edits, as appropriate. In addition, because of the
reorganization, the physical location for many of the offices changed,
and thus, we have made non-substantive amendments to ensure that the
room numbers and addresses reflect the current information, and other
changes as necessary to update outdated addresses, references, and
citations in the regulations pertaining to medical devices. The rule
does not impose any new regulatory requirements on affected parties.
The amendments are editorial in nature and should not be construed as
modifying any substantive standards or requirements.
III. Notice and Public Comment
Publication of this document constitutes final action of these
changes under the Administrative Procedure Act (5 U.S.C. 553). Section
553 of the Administrative Procedure Act (APA) exempts ``rules of agency
organization, procedure, or practice'' from proposed rulemaking (i.e.,
notice and comment rulemaking). 5 U.S.C. 553(b)(3)(A). Rules are also
exempt when an agency finds ``good cause'' that notice and comment
rulemaking procedures would be ``impracticable, unnecessary, or
contrary to the public interest.'' 5 U.S.C. 553(b)(3)(B).
FDA has determined that this rulemaking meets the notice and
comment exemption requirements in 5 U.S.C. 553(b)(3)(A) and (B). FDA's
revisions make technical or non-substantive changes that pertain solely
to the CDRH reorganization and office move and do not alter any
substantive
[[Page 18440]]
standard. FDA does not believe public comment is necessary for these
minor revisions.
The APA allows an effective date less than 30 days after
publication as ``provided by the agency for good cause found and
published with the rule'' (5 U.S.C. 553(d)(3)). A delayed effective
date is unnecessary in this case because the amendments do not impose
any new regulatory requirements on affected parties. As a result,
affected parties do not need time to prepare before the rule takes
effect. Therefore, FDA finds good cause for the amendments to become
effective on the date of publication of this action.
List of Subjects
21 CFR Part 5
Authority delegations (Government agencies), Imports, Organization
and functions (Government agencies).
21 CFR Part 801
Labeling, Medical devices, Reporting and recordkeeping
requirements.
21 CFR Parts 803 and 821
Imports, Medical devices, Reporting and recordkeeping requirements.
21 CFR Part 807
Confidential business information, Imports, Medical devices,
Reporting and recordkeeping requirements.
21 CFR Part 814
Administrative practice and procedure, Confidential business
information, Medical devices, Medical research, Reporting and
recordkeeping requirements.
21 CFR Parts 820 and 822
Medical devices, Reporting and recordkeeping requirements.
21 CFR Part 830
Administrative practice and procedure, Labeling, Medical devices,
Reporting and recordkeeping requirements.
21 CFR Part 860
Administrative practice and procedure, Medical devices.
21 CFR Part 884
Medical Devices.
21 CFR Part 900
Electronic products, Health facilities, Medical devices, Radiation
protection, Reporting and recordkeeping requirements, X-rays.
21 CFR Part 1002
Electronic products, Radiation protection, Reporting and
recordkeeping requirements.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
the authority delegated to the Commissioner of Food and Drugs, 21 CFR
parts 5, 801, 803, 807, 814, 820, 821, 822, 830, 860, 884, 900, and
1002 are amended as follows:
PART 5--ORGANIZATION
0
1. The authority citation for part 5 continues to read as follows:
Authority: 5 U.S.C. 552; 21 U.S.C. 301-397.
0
2. In Sec. 5.1100, revise the entries for ``Center for Devices and
Radiological Health'' through ``Office of In Vitro Diagnostics and
Radiological Health'' to read as follows:
Sec. 5.1100 Headquarters.
* * * * *
Center for Devices and Radiological Health.\12\
---------------------------------------------------------------------------
\12\ Mailing address: 10903 New Hampshire Ave., Bldg. 66, Silver
Spring, MD 20993.
---------------------------------------------------------------------------
Office of the Center Director.
Quality Management Staff.
Office of Communication and Education.
Digital Communication Media Staff.
Program Management Operations Staff.
Division of Communication.
External Communications Branch.
Web and Graphics Branch.
Internal Communication Branch.
Division of Employee Training and Development.
Employee Development Branch.
Technology and Learning Management Branch.
Division of Industry and Consumer Education.
Premarket Programs Branch.
Postmarket and Consumer Branch.
Division of Information Disclosure.
Freedom of Information Branch A.
Freedom of Information Branch B.
Office of Management.
Planning and Program Analysis Staff.
Division of Acquisition Services.
Advanced Acquisitions.
Simplified Acquisitions.
Acquisition Planning Assistance.
Division of Workforce Management.
Recruitment.
Human Capital Management.
Special Programs.
Division of Management Services.
Travel and Conference Management.
Committee Management and Planning.
Space and Facilities Management.
Division of Financial Management.
Budget Formulation.
Budget Execution.
Financial Accountability.
Office of Policy.
Guidance, Legislation and Special Projects.
Regulatory Documents and Special Projects.
Office of Product Evaluation and Quality.
Quality and Analytics Staff.
Clinical and Scientific Policy Staff.
Strategic Initiatives Staff.
Regulation, Policy and Guidance Staff.
Compliance and Quality Staff.
Operations Staff.
Office of Regulatory Programs.
Division of Regulatory Programs 1 (Division of Submission Support).
Division of Regulatory Programs 2 (Division of Establishment
Support).
Division of Regulatory Programs 3 (Division of Market
Intelligence).
Office of Clinical Evidence and Analysis.
Division of Clinical Evidence and Analysis 1 (Division of Clinical
Science and Quality).
Division of Clinical Evidence and Analysis 2 (Division of
Biostatistics).
Office of Health Technology 1 (OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT and Dental Devices).
Division of Health Technology 1A (Division of Ophthalmic Devices).
Division of Health Technology 1B (Division of Dental Devices).
Division of Health Technology 1C (Division of ENT, Sleep Disordered
Breathing, Respiratory and Anesthesia Devices).
Office of Health Technology 2 (OHT2: Office of Cardiovascular Devices).
Division of Health Technology 2A (Division of Cardiac
Electrophysiology, Diagnostics, and Monitoring Devices).
Division of Health Technology 2B (Division of Circulatory Support,
Structural and Vascular Devices).
Division of Health Technology 2C (Division of Coronary and
Peripheral Interventional Devices).
Office of Health Technology 3 (OHT3: Office of Gastrorenal, ObGyn,
General Hospital and Urology Devices).
Division of Health Technology 3A (Division of Renal,
Gastrointestinal, Obesity, and Transplant Devices).
Division of Health Technology 3B (Division of Reproductive,
Gynecology and Urology Devices).
Division of Health Technology 3C
[[Page 18441]]
(Division of Drug Delivery and General Hospital Devices, and Human
Factors).
Office of Health Technology 4 (OHT4: Office of Surgical and Infection
Control Devices).
Division of Health Technology 4A (Division of General Surgery
Devices).
Division of Health Technology 4B (Division of Infection Control and
Plastic Surgery Devices).
Office of Health Technology 5 (OHT5: Office of Neurological and
Physical Medicine Devices).
Division of Health Technology 5A (Division of Neurological,
Neurointerventional and Neurodiagnostic Devices).
Division of Health Technology 5B (Division of Neuromodulation and
Physical Medicine Devices).
Office of Health Technology 6 (OHT6: Office of Orthopedic Devices).
Division of Health Technology 6A (Division of Joint Arthroplasty
Devices).
Division of Health Technology 6B (Division of Spinal Devices).
Division of Health Technology 6C (Division of Stereotaxic, Trauma
and Restorative, Devices).
Office of Health Technology 7 (OHT7: Office of In Vitro Diagnostics and
Radiological Health).
Division of Program Operations and Management.
Division of Chemistry and Toxicology Devices.
Chemistry Branch.
Diabetes Branch.
Toxicology Branch.
Cardio-Renal Diagnostics Branch.
Division of Molecular Genetics and Pathology.
Molecular Pathology and Cytology Branch.
Molecular Genetics Branch.
Division of Immunology and Hematology Devices.
Hematology Branch.
Immunology and Flow-Cytometry Branch.
Division of Microbiology Devices.
Viral Respiratory and HPV Branch.
General Viral and Hepatitis Branch.
General Bacterial and Antimicrobial Susceptibility Branch.
Bacterial Respiratory and Medical Countermeasures Branch.
Division of Radiological Health.
Magnetic Resonance and Electronic Products Branch.
Diagnostic X-Ray Systems Branch.
Nuclear Medicine and Radiation Therapy Branch.
Mammography, Ultrasound and Imaging Software Branch.
Division of Mammography Quality Standards.
Office of Science and Engineering Laboratories.
Immediate Office of the Director
Division of Applied Mechanics.
Division of Biomedical Physics.
Division of Biology, Chemistry and Materials Science.
Division of Imaging, Diagnostics, and Software Reliability.
Division of Administrative and Laboratory Support.
Office of Strategic Partnerships and Technology Innovation.
Innovation.
Division of All Hazards Response Science and Strategic
Partnerships.
Medical Device Development Tools.
Health of Women.
Pediatrics and Special Populations.
All Hazards Readiness Response and Cybersecurity.
Patient Science and Engagement.
Partnerships to Advance Innovation and Regulatory Science.
Science and Special Projects Incubator.
Standards and Conformity Assessment Program.
Division of Digital Health.
Operational Excellence.
Technical and Policy Leadership Strategic Partnerships and
Initiatives 1.
Technical and Policy Leadership Strategic Partnerships and
Initiatives 2.
Strategic Initiatives and Special Projects.
Division of Technology and Data Services.
Business and Transformation Services.
Data Services.
Technology Services.
* * * * *
PART 801--LABELING
0
5. The authority citation for part 801 continues to read as follows:
Authority: 21 U.S.C. 321, 331, 351, 352, 360d, 360i, 360j, 371,
374.
0
6. In Sec. 801.55, revise paragraph (b)(2) to read as follows:
Sec. 801.55 Request for an exception from or alternative to a unique
device identifier requirement.
* * * * *
(b) * * *
(2) In all other cases, by email to: [email protected], or
by correspondence to: UDI Regulatory Policy Support, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 32, Rm. 3293, Silver Spring, MD 20993-0002.
* * * * *
0
7. In Sec. 801.57, revise paragraph (c)(2) to read as follows:
Sec. 801.57 Discontinuation of legacy FDA identification numbers
assigned to devices.
* * * * *
(c) * * *
(2) No later than September 24, 2014, the labeler submits, and
obtains FDA approval of, a request for continued use of the assigned
labeler code. A request for continued use of an assigned labeler code
must be submitted by email to: [email protected], or by
correspondence to: UDI Regulatory Policy Support, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 32, Rm. 3293, Silver Spring, MD 20993-0002.
* * * * *
PART 803--MEDICAL DEVICE REPORTING
0
8. The authority citation for part 803 continues to read as follows:
Authority: 21 U.S.C. 352, 360, 360i, 360j, 371, 374.
0
9. In Sec. 803.11, revise paragraph (d) to read as follows:
Sec. 803.11 What form should I use to submit reports of individual
adverse events and where do I obtain these forms?
* * * * *
(d) Form FDA 3500A is available on the internet at https://www.accessdata.fda.gov/scripts/medwatch/index.cfm.
0
10. In Sec. 803.19(b), revise the second sentence to read as follows:
Sec. 803.19 Are there exemptions, variances, or alternative forms of
adverse event reporting requirements?
* * * * *
(b) * * * You must submit the request to us in writing at the
following address: MDR Exemption Requests, Medical Device Report (MDR)
Team, Division of Regulatory Programs 3, Office of Regulatory Programs,
Office of Product Evaluation and Quality, 10903 New Hampshire Ave.,
Bldg. 66, Rm.1523, Silver Spring, MD 20993-0002. * * *
* * * * *
0
11. In Sec. 803.21, revise paragraph (a) to read as follows:
Sec. 803.21 Where can I find the reporting codes for adverse events
that I use with medical device reports?
(a) The MedWatch Medical Device Reporting Code Instruction Manual
contains adverse event codes for use with Form FDA 3500A. You may
obtain the coding manual from FDA's website
[[Page 18442]]
at: https://www.fda.gov/medical-devices/mandatory-reporting-requirements-manufacturers-importers-and-device-user-facilities/mdr-adverse-event-codes.
* * * * *
0
12. In Sec. 803.33, revise paragraph (a) to read as follows:
Sec. 803.33 If I am a user facility, what must I include when I
submit an annual report?
(a) You must submit to us an annual report on Form FDA 3419. You
must submit an annual report by January 1, of each year. You may obtain
this form on the internet at: https://www.fda.gov/media/72292/download.
* * * * *
PART 807--ESTABLISHMENT REGISTRATION AND DEVICE LISTING FOR
MANUFACTURERS AND INITIAL IMPORTERS OF DEVICES
0
13. The authority citation for part 807 continues to read as follows:
Authority: 21 U.S.C. 321, 331, 351, 352, 360, 360c, 360e, 360i,
360j, 360bbb-8b, 371, 374, 379k-1, 381, 393; 42 U.S.C. 264, 271.
0
14. In Sec. 807.21, revise paragraph (b) introductory text to read as
follows:
Sec. 807.21 How to register establishments and list devices.
* * * * *
(b) If the information under Sec. 807.21(a) cannot be submitted
electronically, a waiver may be requested. Waivers will be granted only
if use of electronic means is not reasonable for the person requesting
the waiver. To request a waiver, applicants must send a letter to the
Imports and Registration and Listing Team, Division of Regulatory
Programs 2, Office of Regulatory Programs, Office of Product Evaluation
and Quality, Center for Devices and Radiological Health, Food and Drug
Administration, 10903 New Hampshire Ave., Bldg. 66, Rm.1432, Silver
Spring, MD 20993-0002, that includes the following information:
* * * * *
0
15. In Sec. 807.34, revise paragraph (a) to read as follows:
Sec. 807.34 Summary of requirements for owners or operators granted
a waiver from submitting required information electronically.
(a) For initial registration and listing, owners or operators who
have been granted a waiver from electronic filing using the procedures
set forth in Sec. 807.21(b) must send a letter containing all of the
registration and listing information described in Sec. Sec. 807.22,
807.25 (and Sec. 807.26 when such information is requested by FDA), at
the times described in Sec. 807.22, to: The Imports and Registration
and Listing Team, Division of Regulatory Programs 2, Office of
Regulatory Programs, Office of Product Evaluation and Quality, Center
for Devices and Radiological Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 66, Rm. 1432, Silver Spring, MD 20993-
0002.
* * * * *
0
16. In Sec. 807.37, revise paragraph (a) to read as follows:
Sec. 807.37 Public availability of establishment registration and
device listing information.
(a) Establishment registration and device listing information is
available for public inspection in accordance with section 510(f) of
the Federal Food, Drug, and Cosmetic Act and will be posted on the FDA
website, with the exception of the information identified in paragraph
(b) of this section. Requests for information by persons who do not
have access to the internet should be directed to the Imports and
Registration and Listing Team, Division of Regulatory Programs 2,
Office of Regulatory Programs, Office of Product Evaluation and
Quality, Center for Devices and Radiological Health, Food and Drug
Administration, 10903 New Hampshire Ave., Bldg. 66, Rm.1432, Silver
Spring, MD 20993-0002. In addition, there will be available for
inspection at each of the Food and Drug Administration district offices
the same information for firms within the geographical area of such
district offices. Upon request, verification of a registration number
or location of a registered establishment will be provided.
* * * * *
PART 814--PREMARKET APPROVAL OF MEDICAL DEVICES
0
17. The authority citation for part 814 continues to read as follows:
Authority: 21 U.S.C. 351, 352, 353, 360, 360c-360j, 360bbb-8b,
371, 372, 373, 374, 375, 379, 379e, 379k-1, 381.
0
18. In Sec. 814.42, revise paragraph (d)(2) to read as follows:
Sec. 814.42 Filing a PMA.
* * * * *
(d) * * *
(2) Request in writing within 10 working days of the date of
receipt of the notice refusing to file the PMA, an informal conference
with the Director of the associated Office of Health Technology to
review FDA's decision not to file the PMA. FDA will hold the informal
conference within 10 working days of its receipt of the request and
will render its decision on filing within 5 working days after the
informal conference. If, after the informal conference, FDA accepts the
PMA for filing, the date of filing will be the date of the decision to
accept the PMA for filing. If FDA does not reverse its decision not to
file the PMA, the applicant may request reconsideration of the decision
from the Director of the Office of Product Evaluation and Quality, the
Director of the Center for Biologics Evaluation and Research, or the
Director of the Center for Drug Evaluation and Research, as applicable.
The Director's decision will constitute final administrative action for
the purpose of judicial review.
* * * * *
0
19. In Sec. 814.100, revise paragraph (e)(2) to read as follows:
Sec. 814.100 Purpose and scope.
* * * * *
(e) * * *
(2) Submitting an HDE to the Office of Product Evaluation and
Quality (OPEQ), Center for Devices and Radiological Health (CDRH), the
Center for Biologics Evaluation and Research (CBER), or the Center for
Drug Evaluation and Research (CDER), as applicable.
* * * * *
PART 820--QUALITY SYSTEM REGULATION
0
20. The authority citation for part 820 continues to read as follows:
Authority: 21 U.S.C. 351, 352, 360, 360c, 360d, 360e, 360h,
360i, 360j, 360l, 371, 374, 381, 383; 42 U.S.C. 216, 262, 263a, 264.
0
21. In Sec. 820.1, revise paragraph (e)(1) to read as follows:
Sec. 820.1 Scope.
* * * * *
(e) * * * (1) Any person who wishes to petition for an exemption or
variance from any device quality system requirement is subject to the
requirements of section 520(f)(2) of the Federal Food, Drug, and
Cosmetic Act. Petitions for an exemption or variance shall be submitted
according to the procedures set forth in Sec. 10.30 of this chapter,
the FDA's administrative procedures. For guidance on how to proceed for
a request for a variance, contact Division of Regulatory Programs 2,
Office of Regulatory Programs, Office of Product Evaluation and
Quality, Center for Devices and Radiological Health, Food and Drug
Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 1438, Silver
Spring, MD 20993-0002.
* * * * *
[[Page 18443]]
PART 821--MEDICAL DEVICE TRACKING REQUIREMENTS
0
22. The authority citation for part 821 continues to read as follows:
Authority: 21 U.S.C. 331, 351, 352, 360, 360e, 360h, 360i, 371,
374.
0
23. In Sec. 821.2, revise paragraph (b) introductory text to read as
follows:
Sec. 821.2 Exemptions and variances.
* * * * *
(b) A request for an exemption or variance shall be submitted in
the form of a petition under Sec. 10.30 of this chapter and shall
comply with the requirements set out therein, except that a response
shall be issued in 90 days. The Director or Deputy Directors, CDRH, or
the Director of the Office of Regulatory Program, CDRH, shall issue
responses to requests under this section. The petition shall also
contain the following:
* * * * *
PART 822--POSTMARKET SURVEILLANCE
0
24. The authority citation for part 822 continues to read as follows:
Authority: 21 U.S.C. 331, 352, 360i, 360l, 371, 374.
0
25. In Sec. 822.7, revise paragraph (a)(1) to read as follows:
Sec. 822.7 What should I do if I do not agree that postmarket
surveillance is appropriate?
(a) * * *
(1) Requesting a meeting with the individual who issued the order
for postmarket surveillance;
* * * * *
0
26. In Sec. 822.22, revise paragraph (a)(1) to read as follows:
Sec. 822.22 What recourse do I have if I do not agree with your
decision?
(a) * * *
(1) Requesting a meeting with the individual who issued the order
for postmarket surveillance;
* * * * *
PART 830--UNIQUE DEVICE IDENTIFICATION
0
27. The authority citation for part 830 continues to read as follows:
Authority: 21 U.S.C. 321, 331, 352, 353, 360, 360d, 360i, 360j,
371.
* * * * *
0
28. In Sec. 830.110, revise paragraph (a)(1) to read as follows:
Sec. 830.110 Application for accreditation as an issuing agency.
(a) * * *(1) An applicant seeking initial FDA accreditation as an
issuing agency shall notify FDA of its desire to be accredited by
sending a notification by email to: [email protected], or by
correspondence to: UDI Regulatory Policy Support, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 32, Rm. 3293, Silver Spring, MD 20993-0002.
* * * * *
0
29. In Sec. 830.320, revise paragraphs (c)(1) and (3) to read as
follows:
Sec. 830.320 Submission of unique device identification information.
* * * * *
(c) * * * (1) A labeler may request a waiver from electronic
submission of UDI data by submitting a letter addressed to the
appropriate Center Director explaining why electronic submission is not
technologically feasible; send the request by email to:
[email protected], or by correspondence to: UDI Regulatory Policy
Support, Center for Devices and Radiological Health, Food and Drug
Administration, 10903 New Hampshire Ave., Bldg. 32, Rm. 3293, Silver
Spring, MD 20993-0002.
* * * * *
(3) A labeler that has a waiver from electronic submission of UDI
data must send a letter containing all of the information required by
Sec. 830.310, as well as any ancillary information permitted to be
submitted under Sec. 830.340 that the labeler wishes to submit, within
the time permitted by Sec. 830.330, addressed to: UDI Regulatory
Policy Support, Center for Devices and Radiological Health, Food and
Drug Administration, 10903 New Hampshire Ave., Bldg. 32, Rm. 3293,
Silver Spring, MD 20993-0002.
PART 860--MEDICAL DEVICE CLASSIFICATION PROCEDURES
0
30. The authority citation for part 860 continues to read as follows:
Authority: 21 U.S.C. 360c, 360d, 360e, 360i, 360j, 371, 374.
0
31. In Sec. 860.123, revise paragraph (b)(1) to read as follows:
Sec. 860.123 Reclassification petition: Content and form.
* * * * *
(b) * * *
(1) For devices regulated by the Center for Devices and
Radiological Health, addressed to the Food and Drug Administration,
Center for Devices and Radiological Health, Office of Policy Staff,
10903 New Hampshire Ave., Bldg. 66, Rm. 5445, Silver Spring, MD 20993-
0002; for devices regulated by the Center for Biologics Evaluation and
Research, addressed to the Food and Drug Administration, Center for
Biologics Evaluation and Research, Document Control Center, 10903 New
Hampshire Ave., Bldg. 71, Rm. G112, Silver Spring, MD 20993-0002; for
devices regulated by the Center for Drug Evaluation and Research,
addressed to the Food and Drug Administration, Center for Drug
Evaluation and Research, Central Document Control Room, 5901-B
Ammendale Rd., Beltsville, MD 20705-1266, as applicable.
* * * * *
PART 884--OBSTETRICAL AND GYNECOLOGICAL DEVICES
0
32. The authority citation for part 884 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
33. In Sec. 884.5360, remove and reserve paragraph (c).
PART 900--MAMMOGRAPHY
0
34. The authority citation for part 900 continues to read as follows:
Authority: 21 U.S.C. 360i, 360nn, 374(e); 42 U.S.C. 263b.
0
35. In Sec. 900.3, revise paragraph (b)(1) to read as follows:
Sec. 900.3 Application for approval as an accreditation body.
* * * * *
(b) * * *(1) An applicant seeking initial FDA approval as an
accreditation body shall inform the Division of Mammography Quality
Standards, Center for Devices and Radiological Health, Food and Drug
Administration, 10903 New Hampshire Ave., Bldg. 66, Rm. 3621, Silver
Spring, MD 20993, Attn: Program Management Branch, of its desire to be
approved as an accreditation body and of its requested scope of
authority.
* * * * *
0
36. In Sec. 900.15, revise paragraph (d)(3)(i) to read as follows:
Sec. 900.15 Appeals of adverse accreditation or reaccreditation
decisions that preclude certification or recertification.
* * * * *
(d) * * *
(3) * * *
(i) A facility must request reconsideration by DMQS within 60 days
of the accreditation body's adverse
[[Page 18444]]
appeals decision, at the following address: Food and Drug
Administration, Center for Devices and Radiological Health, Division of
Mammography Quality Standards, Attn: Facility Accreditation Review
Committee, 10903 New Hampshire Ave., Bldg. 66, Rm. 3621, Silver Spring,
MD 20993-0002.
* * * * *
0
37. In Sec. 900.18, revise paragraph (c) introductory text to read as
follows:
Sec. 900.18 Alternative requirements for Sec. 900.12 quality
standards.
* * * * *
(c) Applications for approval of an alternative standard. An
application for approval of an alternative standard or for an amendment
or extension of the alternative standard shall be submitted in an
original and two copies to the Food and Drug Administration, Center for
Devices and Radiological Health, Director, Division of Mammography
Quality Standards, 10903 New Hampshire Ave., Bldg. 66, Rm. 3621, Silver
Spring, MD 20993-0002. The application for approval of an alternative
standard shall include the following information:
* * * * *
0
38. In Sec. 900.21, revise paragraph (b)(1) to read as follows:
Sec. 900.21 Application for approval as a certification agency.
* * * * *
(b) * * * (1) An applicant seeking FDA approval as a certification
agency shall inform the Food and Drug Administration, Center for
Devices and Radiological Health, Director, Division of Mammography
Quality Standards, Attn: Program Management Branch, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3621, Silver Spring, MD 20993-0002, in writing, of
its desire to be approved as a certification agency.
* * * * *
PART 1002--RECORDS AND REPORTS
0
39. The authority citation for part 1002 continues to read as follows:
Authority: 21 U.S.C. 352, 360, 360i, 360j, 360hh-360ss, 371,
374.
0
40. In Sec. 1002.50, revise paragraph (c)(3) to read as follows:
Sec. 1002.50 Special exemptions.
* * * * *
(c) * * *
(3) Such conditions as are deemed necessary to protect the public
health and safety. Copies of exemptions shall be available upon request
from the Food and Drug Administration, Center for Devices and
Radiological Health, Division of Mammography Quality Standards, 10903
New Hampshire Ave., Bldg. 66, Rm. 3621, Silver Spring, MD 20993-0002.
* * * * *
Dated: March 23, 2020.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2020-06354 Filed 4-1-20; 8:45 am]