Radiology Devices; Reclassification of Medical Image Analyzers, 3545-3548 [2020-00494]
Download as PDF
khammond on DSKJM1Z7X2PROD with RULES
Federal Register / Vol. 85, No. 14 / Wednesday, January 22, 2020 / Rules and Regulations
a reading queue. The device operates in
parallel with the standard of care, which
remains the default option for all cases.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Design verification and validation
must include:
(i) A detailed description of the
notification and triage algorithms and
all underlying image analysis algorithms
including, but not limited to, a detailed
description of the algorithm inputs and
outputs, each major component or
block, how the algorithm affects or
relates to clinical practice or patient
care, and any algorithm limitations.
(ii) A detailed description of prespecified performance testing protocols
and dataset(s) used to assess whether
the device will provide effective triage
(e.g., improved time to review of
prioritized images for pre-specified
clinicians).
(iii) Results from performance testing
that demonstrate that the device will
provide effective triage. The
performance assessment must be based
on an appropriate measure to estimate
the clinical effectiveness. The test
dataset must contain sufficient numbers
of cases from important cohorts (e.g.,
subsets defined by clinically relevant
confounders, effect modifiers,
associated diseases, and subsets defined
by image acquisition characteristics)
such that the performance estimates and
confidence intervals for these individual
subsets can be characterized with the
device for the intended use population
and imaging equipment.
(iv) Stand-alone performance testing
protocols and results of the device.
(v) Appropriate software
documentation (e.g., device hazard
analysis; software requirements
specification document; software design
specification document; traceability
analysis; description of verification and
validation activities including system
level test protocol, pass/fail criteria, and
results).
(2) Labeling must include the
following:
(i) A detailed description of the
patient population for which the device
is indicated for use;
(ii) A detailed description of the
intended user and user training that
addresses appropriate use protocols for
the device;
(iii) Discussion of warnings,
precautions, and limitations must
include situations in which the device
may fail or may not operate at its
expected performance level (e.g., poor
image quality for certain
subpopulations), as applicable;
VerDate Sep<11>2014
15:46 Jan 21, 2020
Jkt 250001
(iv) A detailed description of
compatible imaging hardware, imaging
protocols, and requirements for input
images;
(v) Device operating instructions; and
(vi) A detailed summary of the
performance testing, including: test
methods, dataset characteristics, triage
effectiveness (e.g., improved time to
review of prioritized images for prespecified clinicians), diagnostic
accuracy of algorithms informing triage
decision, and results with associated
statistical uncertainty (e.g., confidence
intervals), including a summary of
subanalyses on case distributions
stratified by relevant confounders, such
as lesion and organ characteristics,
disease stages, and imaging equipment.
Dated: January 9, 2020.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2020–00496 Filed 1–21–20; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 892
[Docket No. FDA–2018–N–1553]
Radiology Devices; Reclassification of
Medical Image Analyzers
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final amendment; final order.
The Food and Drug
Administration (FDA or the Agency) is
issuing a final order to reclassify
medical image analyzers applied to
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection, postamendments class
III devices (regulated under product
code MYN), into class II (special
controls), subject to premarket
notification. These devices are intended
to direct the clinician’s attention to
portions of an image that may reveal
abnormalities during interpretation of
patient radiology images by the
clinician. FDA is also identifying the
special controls that the Agency
believes are necessary to provide a
reasonable assurance of safety and
effectiveness of the device type.
DATES: This order is effective February
21, 2020.
FOR FURTHER INFORMATION CONTACT:
Robert Ochs, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
SUMMARY:
PO 00000
Frm 00007
Fmt 4700
Sfmt 4700
3545
Ave., Bldg. 66, Rm. 3680, Silver Spring,
MD 20993–0002, 301–796–6661,
Robert.Ochs@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (FD&C Act), as amended, establishes
a comprehensive system for the
regulation of medical devices intended
for human use. Section 513 of the FD&C
Act (21 U.S.C. 360c) established three
classes of devices, reflecting the
regulatory controls needed to provide
reasonable assurance of their safety and
effectiveness. The three classes of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices), are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II, or FDA issues an order finding
the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part
807 (21 CFR part 807).
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or II
under section 513(f)(3). Section 513(f)(3)
of the FD&C Act provides that FDA
acting by order can reclassify the device
into class I or II on its own initiative, or
in response to a petition from the
manufacturer or importer of the device.
To change the classification of the
device, the proposed new class must
have sufficient regulatory controls to
provide a reasonable assurance of the
safety and effectiveness of the device for
its intended use.
On June 4, 2018 (83 FR 25598), FDA
published in the Federal Register a
proposed order to reclassify the device
type from class III to class II, subject to
premarket notification. The comment
period on the proposed order closed on
August 3, 2018.
II. Comments on the Proposed Order
In response to the June 4, 2018,
proposed order (83 FR 25598), FDA
received two comments including from
E:\FR\FM\22JAR1.SGM
22JAR1
khammond on DSKJM1Z7X2PROD with RULES
3546
Federal Register / Vol. 85, No. 14 / Wednesday, January 22, 2020 / Rules and Regulations
a healthcare professional in the medical
device industry and a professional
society by the close of the comment
period, each containing one or more
comments on one or more issues. We
describe and respond to the comments
in this section of the document. The
order of response to the commenters is
purely for organizational purposes and
does not signify the comment’s value or
importance nor the order in which
comments were received.
(Comment 1) One commenter
supported the proposed reclassification
and requested that the Agency provide
examples to further clarify what devices
may fall within the scope of the
proposed reclassification action. The
commenter offered three use cases as
examples: (1) A concurrent-read
software device that shows computer
assisted or aided detection (CADe)
marks (on the regions of interest)
intended to draw the clinicians’
attention during their standard review
workflow, (2) a software device that
provides quantitative measures of
disease risk, and (3) a software device
that suggests prioritization of the cases
in a review list/worklist for a clinician’s
reading session.
(Response 1) FDA agrees that
providing examples as part of the
preamble would provide further clarity
in the scope of the reclassification order.
FDA stated in Section II of the proposed
order (83 FR 25598) that, if finalized,
the reclassification would cover medical
image analyzers including CADe
devices for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection that
are assigned product code MYN.
Therefore, if example number 1 from the
commenter above provided is a CADe
device intended for use in the clinical
applications noted above, it likely falls
within the scope of this reclassification
order. FDA believes that examples
number 2 and number 3 identified in
Comment 1 would not be appropriately
classified as medical image analyzers or
CADe devices. Therefore, these two
examples are device types not covered
by this reclassification. Specifically,
FDA believes that example number 2 is
likely a computer-aided diagnosis
device, which FDA has classified
separately under 21 CFR 892.2060
Radiological computer-assisted
diagnostic (CADx) software for lesions
suspicious for cancer, and example
number 3 is likely a computer-aided
triage device, which FDA has classified
separately under 21 CFR 892.2080,
Radiological Computer Aided Triage
and Notification Software.
VerDate Sep<11>2014
15:46 Jan 21, 2020
Jkt 250001
FDA believes that the identification
language in § 892.2070(a) (21 CFR
892.2070) of the classification regulation
as described in the proposed order
adequately identifies the types of
devices that would be affected by this
reclassification action and declines to
further modify the identification
language for medical image analyzers.
(Comment 2) The commenter
generally supported the proposed
reclassification and proposed special
controls but recommended the
following additional special controls to
ensure the safety and effectiveness of
the devices:
• Collect postmarket data: The
commenter recommends that FDA
consider adding special controls that
require the collection of postmarket data
when necessary to ensure that the
device performance is acceptable in
clinical use and to allow for a
comparison to premarket data submitted
for 510(k) clearance.
• Employ periodic retraining: The
commenter recommends that FDA
consider adding special controls for
periodic retraining of clinicians to
ensure that they understand the use of
the device according to device labeling
and the appropriate reading protocol to
follow.
• Implement quality assurance
requirements: The commenter notes that
there are no quality assurance
requirements in the proposed special
controls to assure the medical image
analyzer does not fail or to ensure that
it operates at its expected performance
level. Accordingly, the commenter asks
FDA to consider adding special controls
for a well-defined device-specific
quality assurance process.
(Response 2) FDA disagrees with this
comment. The Agency believes that the
special controls, as identified in the
proposed order, together with general
controls, are sufficient to provide
reasonable assurance of safety and
effectiveness of medical image
analyzers. FDA does not believe that a
requirement for collecting postmarket
data as a special control is necessary to
provide reasonable assurance of the
safety and effectiveness for medical
image analyzers devices. FDA has
nearly 20 years of experience regulating
medical image analyzers as class III
devices considered within the scope of
this reclassification, and postmarket
studies have not been required during
this period. Additionally, there have not
been signals observed during this time
that would suggest postmarket studies
were necessary. As stated in the
proposed order, in the past 10 years,
there have been no medical device
reports related to these CADe devices.
PO 00000
Frm 00008
Fmt 4700
Sfmt 4700
FDA has only classified one recall for
these CADe devices due to distribution
of the CADe device without prior
premarket application approval.
Further, FDA still maintains the
authority to require manufacturers to
conduct postmarket surveillance of class
II devices under section 522 of the
FD&C Act (21 U.S.C. 360l) when certain
conditions are met.
FDA does not believe that an
additional special control regarding user
retraining is necessary to provide
reasonable assurance of the safety and
effectiveness of medical image
analyzers. The Agency believes that it
has identified the risk of misuse by a
physician and that the measures
described in this final order will be
effective in mitigating this probable risk
to health, mainly § 892.2070(b)(2) that
requires the labeling of these devices to
include a detailed description of the
intended user and user training that
addresses appropriate reading protocols
for the device.
While FDA agrees that appropriate
quality assurance (QA) approaches are
helpful in ensuring that a medical image
analyzer consistently operates at its
expected performance level, FDA
disagrees with the need to write specific
QA requirements for the end-users into
the special controls. FDA believes
specifying end-user QA requirements
may be redundant with the special
controls already identified in this final
order (see specifically § 892.2070(b)(1)
and (2)), in combination with general
controls, including especially the
Quality System (QS) Regulations under
part 820 (21 CFR part 820). In addition,
because many different approaches to
address QA exist, coupled with the
general and special controls
requirements that are already applicable
to this device type, FDA does not
believe that defining specific QA
requirements in the special controls is
necessary to reasonably assure safety,
and FDA believes manufacturers should
consider the need for, and most
appropriate methods, to evaluate
potential changes in CADe performance
over time. This information should
address the general controls as well as
special controls (see specifically
§ 892.2070(b)(2)), including providing
user training, identifying compatible
hardware, identifying limitations, and
providing operating instructions.
III. The Final Order
Based on the information discussed in
the preamble to the proposed order (83
FR 25598, June 4, 2018), the comments
received for the proposed order, prior
panel discussions, and FDA’s
experiences over the years in reviewing
E:\FR\FM\22JAR1.SGM
22JAR1
Federal Register / Vol. 85, No. 14 / Wednesday, January 22, 2020 / Rules and Regulations
these device types, FDA concludes that
special controls, in conjunction with
general controls, will provide
reasonable assurance of the safety and
effectiveness of medical image
analyzers. FDA is adopting its findings
under section 513(f)(3) of the FD&C Act,
as published in the preamble to the
proposed order (83 FR 25598). FDA is
issuing this final order to reclassify
medical image analyzers, including
CADe devices, for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection from
class III to class II, and establishing
special controls by revising 21 CFR part
892.1 In this final order, the Agency has
identified the special controls under
section 513(a)(1)(B) of the FD&C Act
that, together with general controls,
provide reasonable assurance of the
safety and effectiveness of these devices.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
FDA has determined that premarket
notification is necessary to provide
reasonable assurance of safety and
effectiveness of medical image
analyzers, and therefore, this device
type is not exempt from premarket
notification requirements. Persons who
intend to market a new CADe device
must obtain clearance of a premarket
notification and demonstrate
compliance with the special controls
included in this final order, prior to
marketing the device.
khammond on DSKJM1Z7X2PROD with RULES
1 FDA notes that the ‘‘ACTION’’ caption for this
final order is styled as ‘‘Final amendment; final
order,’’ rather than ‘‘Final order.’’ Beginning in
December 2019, this editorial change was made to
indicate that the document ‘‘amends’’ the Code of
Federal Regulations. The change was made in
accordance with the Office of Federal Register’s
(OFR) interpretations of the Federal Register Act (44
U.S.C. chapter 15), its implementing regulations (1
CFR 5.9 and parts 21 and 22), and the Document
Drafting Handbook.
VerDate Sep<11>2014
15:46 Jan 21, 2020
Jkt 250001
The device is assigned the generic
name medical image analyzer, and
identified as a medical image analyzers,
including CADe devices, for
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection, is a prescription device
that is intended to identify, mark,
highlight, or in any other manner direct
the clinicians’ attention to portions of a
radiology image that may reveal
abnormalities during interpretation of
patient radiology images by the
clinicians. This device incorporates
pattern recognition and data analysis
capabilities and operates on previously
acquired medical images. This device is
not intended to replace the review by a
qualified radiologist and is not intended
to be used for triage or to recommend
diagnosis.
Under this final order, the medical
image analyzer is a prescription use
only device under § 801.109 (21 CFR
801.109). Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352(f)(1)) and 21
CFR 801.5, as long as the conditions of
§ 801.109 are met (referring to 21 U.S.C.
352(f)(1)). Under 21 CFR 807.81, the
device continues to be subject to 510(k)
requirements.
IV. Codification of Orders
Prior to the amendments in the Food
and Drug Administration Safety and
Innovation Act (FDASIA), section 513(e)
of the FD&C Act provided for FDA to
issue regulations to reclassify devices.
Although section 513(e) as amended
requires FDA to issue final orders rather
than regulations, it also provides for
FDA to revoke previously issued
regulations by order. FDA will continue
to codify classifications and
reclassifications in the Code of Federal
Regulations (CFR). Changes resulting
from final orders will appear in the CFR
as changes to codified classification
determinations or as newly codified
orders. Therefore, under section
513(e)(1)(A)(i), as amended by FDASIA,
PO 00000
Frm 00009
Fmt 4700
Sfmt 4700
3547
in this final order, we are proposing to
codify the classification of medical
image analyzers, applied to CADe
devices for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection in the
new § 892.2070, under which medical
image analyzers would be reclassified
into class II.
V. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
VI. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved FDA collections. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3521). The collections of information in
part 807, subpart E have been approved
under OMB control number 0910–0120;
the collections of information in 21 CFR
part 801 have been approved under
OMB control number 0910–0485; and
the collections of information in part
820 have been approved under OMB
control number 0910–0073.
List of Subjects in 21 CFR Part 892
Medical devices, Radiation
protection, X-rays.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 892 is
amended as follows:
PART 892—RADIOLOGY DEVICES
1. The authority citation for part 892
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
E:\FR\FM\22JAR1.SGM
22JAR1
3548
Federal Register / Vol. 85, No. 14 / Wednesday, January 22, 2020 / Rules and Regulations
2. Add § 892.2070 to subpart B to read
as follows:
■
§ 892.2070
Medical image analyzer.
khammond on DSKJM1Z7X2PROD with RULES
(a) Identification. Medical image
analyzers, including computer-assisted/
aided detection (CADe) devices for
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection, is a prescription device
that is intended to identify, mark,
highlight, or in any other manner direct
the clinicians’ attention to portions of a
radiology image that may reveal
abnormalities during interpretation of
patient radiology images by the
clinicians. This device incorporates
pattern recognition and data analysis
capabilities and operates on previously
acquired medical images. This device is
not intended to replace the review by a
qualified radiologist, and is not
intended to be used for triage, or to
recommend diagnosis.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Design verification and validation
must include:
(i) A detailed description of the image
analysis algorithms including a
description of the algorithm inputs and
outputs, each major component or
block, and algorithm limitations.
(ii) A detailed description of prespecified performance testing methods
and dataset(s) used to assess whether
the device will improve reader
performance as intended and to
characterize the standalone device
performance. Performance testing
includes one or more standalone tests,
side-by-side comparisons, or a reader
study, as applicable.
(iii) Results from performance testing
that demonstrate that the device
improves reader performance in the
intended use population when used in
accordance with the instructions for
use. The performance assessment must
be based on appropriate diagnostic
accuracy measures (e.g., receiver
operator characteristic plot, sensitivity,
specificity, predictive value, and
diagnostic likelihood ratio). The test
dataset must contain a sufficient
number of cases from important cohorts
(e.g., subsets defined by clinically
VerDate Sep<11>2014
15:46 Jan 21, 2020
Jkt 250001
relevant confounders, effect modifiers,
concomitant diseases, and subsets
defined by image acquisition
characteristics) such that the
performance estimates and confidence
intervals of the device for these
individual subsets can be characterized
for the intended use population and
imaging equipment.
(iv) Appropriate software
documentation (e.g., device hazard
analysis; software requirements
specification document; software design
specification document; traceability
analysis; description of verification and
validation activities including system
level test protocol, pass/fail criteria, and
results; and cybersecurity).
(2) Labeling must include the
following:
(i) A detailed description of the
patient population for which the device
is indicated for use.
(ii) A detailed description of the
intended reading protocol.
(iii) A detailed description of the
intended user and user training that
addresses appropriate reading protocols
for the device.
(iv) A detailed description of the
device inputs and outputs.
(v) A detailed description of
compatible imaging hardware and
imaging protocols.
(vi) Discussion of warnings,
precautions, and limitations must
include situations in which the device
may fail or may not operate at its
expected performance level (e.g., poor
image quality or for certain
subpopulations), as applicable.
(vii) Device operating instructions.
(viii) A detailed summary of the
performance testing, including: test
methods, dataset characteristics, results,
and a summary of sub-analyses on case
distributions stratified by relevant
confounders, such as lesion and organ
characteristics, disease stages, and
imaging equipment.
Dated: January 9, 2020.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2020–00494 Filed 1–21–20; 8:45 am]
BILLING CODE 4164–01–P
PO 00000
Frm 00010
Fmt 4700
Sfmt 4700
DEPARTMENT OF HOMELAND
SECURITY
Federal Emergency Management
Agency
44 CFR Part 64
[Docket ID FEMA–2020–0005; Internal
Agency Docket No. FEMA–8613]
Suspension of Community Eligibility
Federal Emergency
Management Agency, DHS.
ACTION: Final rule.
AGENCY:
This rule identifies
communities where the sale of flood
insurance has been authorized under
the National Flood Insurance Program
(NFIP) that are scheduled for
suspension on the effective dates listed
within this rule because of
noncompliance with the floodplain
management requirements of the
program. If the Federal Emergency
Management Agency (FEMA) receives
documentation that the community has
adopted the required floodplain
management measures prior to the
effective suspension date given in this
rule, the suspension will not occur and
a notice of this will be provided by
publication in the Federal Register on a
subsequent date. Also, information
identifying the current participation
status of a community can be obtained
from FEMA’s Community Status Book
(CSB). The CSB is available at https://
www.fema.gov/national-floodinsurance-program-community-statusbook.
EFFECTIVE DATES: The effective date of
each community’s scheduled
suspension is the third date (‘‘Susp.’’)
listed in the third column of the
following tables.
FOR FURTHER INFORMATION: If you want
to determine whether a particular
community was suspended on the
suspension date or for further
information, contact Adrienne L.
Sheldon, PE, CFM, Federal Insurance
and Mitigation Administration, Federal
Emergency Management Agency, 400 C
Street SW, Washington, DC 20472, (202)
212–3966.
SUMMARY:
E:\FR\FM\22JAR1.SGM
22JAR1
Agencies
[Federal Register Volume 85, Number 14 (Wednesday, January 22, 2020)]
[Rules and Regulations]
[Pages 3545-3548]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2020-00494]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 892
[Docket No. FDA-2018-N-1553]
Radiology Devices; Reclassification of Medical Image Analyzers
AGENCY: Food and Drug Administration, HHS.
ACTION: Final amendment; final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or the Agency) is
issuing a final order to reclassify medical image analyzers applied to
mammography breast cancer, ultrasound breast lesions, radiograph lung
nodules, and radiograph dental caries detection, postamendments class
III devices (regulated under product code MYN), into class II (special
controls), subject to premarket notification. These devices are
intended to direct the clinician's attention to portions of an image
that may reveal abnormalities during interpretation of patient
radiology images by the clinician. FDA is also identifying the special
controls that the Agency believes are necessary to provide a reasonable
assurance of safety and effectiveness of the device type.
DATES: This order is effective February 21, 2020.
FOR FURTHER INFORMATION CONTACT: Robert Ochs, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 3680, Silver Spring, MD 20993-0002, 301-796-6661,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended,
establishes a comprehensive system for the regulation of medical
devices intended for human use. Section 513 of the FD&C Act (21 U.S.C.
360c) established three classes of devices, reflecting the regulatory
controls needed to provide reasonable assurance of their safety and
effectiveness. The three classes of devices are class I (general
controls), class II (special controls), and class III (premarket
approval).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices), are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until, the device
is reclassified into class I or II, or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or II under section 513(f)(3). Section 513(f)(3) of the FD&C
Act provides that FDA acting by order can reclassify the device into
class I or II on its own initiative, or in response to a petition from
the manufacturer or importer of the device. To change the
classification of the device, the proposed new class must have
sufficient regulatory controls to provide a reasonable assurance of the
safety and effectiveness of the device for its intended use.
On June 4, 2018 (83 FR 25598), FDA published in the Federal
Register a proposed order to reclassify the device type from class III
to class II, subject to premarket notification. The comment period on
the proposed order closed on August 3, 2018.
II. Comments on the Proposed Order
In response to the June 4, 2018, proposed order (83 FR 25598), FDA
received two comments including from
[[Page 3546]]
a healthcare professional in the medical device industry and a
professional society by the close of the comment period, each
containing one or more comments on one or more issues. We describe and
respond to the comments in this section of the document. The order of
response to the commenters is purely for organizational purposes and
does not signify the comment's value or importance nor the order in
which comments were received.
(Comment 1) One commenter supported the proposed reclassification
and requested that the Agency provide examples to further clarify what
devices may fall within the scope of the proposed reclassification
action. The commenter offered three use cases as examples: (1) A
concurrent-read software device that shows computer assisted or aided
detection (CADe) marks (on the regions of interest) intended to draw
the clinicians' attention during their standard review workflow, (2) a
software device that provides quantitative measures of disease risk,
and (3) a software device that suggests prioritization of the cases in
a review list/worklist for a clinician's reading session.
(Response 1) FDA agrees that providing examples as part of the
preamble would provide further clarity in the scope of the
reclassification order. FDA stated in Section II of the proposed order
(83 FR 25598) that, if finalized, the reclassification would cover
medical image analyzers including CADe devices for mammography breast
cancer, ultrasound breast lesions, radiograph lung nodules, and
radiograph dental caries detection that are assigned product code MYN.
Therefore, if example number 1 from the commenter above provided is a
CADe device intended for use in the clinical applications noted above,
it likely falls within the scope of this reclassification order. FDA
believes that examples number 2 and number 3 identified in Comment 1
would not be appropriately classified as medical image analyzers or
CADe devices. Therefore, these two examples are device types not
covered by this reclassification. Specifically, FDA believes that
example number 2 is likely a computer-aided diagnosis device, which FDA
has classified separately under 21 CFR 892.2060 Radiological computer-
assisted diagnostic (CADx) software for lesions suspicious for cancer,
and example number 3 is likely a computer-aided triage device, which
FDA has classified separately under 21 CFR 892.2080, Radiological
Computer Aided Triage and Notification Software.
FDA believes that the identification language in Sec. 892.2070(a)
(21 CFR 892.2070) of the classification regulation as described in the
proposed order adequately identifies the types of devices that would be
affected by this reclassification action and declines to further modify
the identification language for medical image analyzers.
(Comment 2) The commenter generally supported the proposed
reclassification and proposed special controls but recommended the
following additional special controls to ensure the safety and
effectiveness of the devices:
Collect postmarket data: The commenter recommends that FDA
consider adding special controls that require the collection of
postmarket data when necessary to ensure that the device performance is
acceptable in clinical use and to allow for a comparison to premarket
data submitted for 510(k) clearance.
Employ periodic retraining: The commenter recommends that
FDA consider adding special controls for periodic retraining of
clinicians to ensure that they understand the use of the device
according to device labeling and the appropriate reading protocol to
follow.
Implement quality assurance requirements: The commenter
notes that there are no quality assurance requirements in the proposed
special controls to assure the medical image analyzer does not fail or
to ensure that it operates at its expected performance level.
Accordingly, the commenter asks FDA to consider adding special controls
for a well-defined device-specific quality assurance process.
(Response 2) FDA disagrees with this comment. The Agency believes
that the special controls, as identified in the proposed order,
together with general controls, are sufficient to provide reasonable
assurance of safety and effectiveness of medical image analyzers. FDA
does not believe that a requirement for collecting postmarket data as a
special control is necessary to provide reasonable assurance of the
safety and effectiveness for medical image analyzers devices. FDA has
nearly 20 years of experience regulating medical image analyzers as
class III devices considered within the scope of this reclassification,
and postmarket studies have not been required during this period.
Additionally, there have not been signals observed during this time
that would suggest postmarket studies were necessary. As stated in the
proposed order, in the past 10 years, there have been no medical device
reports related to these CADe devices. FDA has only classified one
recall for these CADe devices due to distribution of the CADe device
without prior premarket application approval. Further, FDA still
maintains the authority to require manufacturers to conduct postmarket
surveillance of class II devices under section 522 of the FD&C Act (21
U.S.C. 360l) when certain conditions are met.
FDA does not believe that an additional special control regarding
user retraining is necessary to provide reasonable assurance of the
safety and effectiveness of medical image analyzers. The Agency
believes that it has identified the risk of misuse by a physician and
that the measures described in this final order will be effective in
mitigating this probable risk to health, mainly Sec. 892.2070(b)(2)
that requires the labeling of these devices to include a detailed
description of the intended user and user training that addresses
appropriate reading protocols for the device.
While FDA agrees that appropriate quality assurance (QA) approaches
are helpful in ensuring that a medical image analyzer consistently
operates at its expected performance level, FDA disagrees with the need
to write specific QA requirements for the end-users into the special
controls. FDA believes specifying end-user QA requirements may be
redundant with the special controls already identified in this final
order (see specifically Sec. 892.2070(b)(1) and (2)), in combination
with general controls, including especially the Quality System (QS)
Regulations under part 820 (21 CFR part 820). In addition, because many
different approaches to address QA exist, coupled with the general and
special controls requirements that are already applicable to this
device type, FDA does not believe that defining specific QA
requirements in the special controls is necessary to reasonably assure
safety, and FDA believes manufacturers should consider the need for,
and most appropriate methods, to evaluate potential changes in CADe
performance over time. This information should address the general
controls as well as special controls (see specifically Sec.
892.2070(b)(2)), including providing user training, identifying
compatible hardware, identifying limitations, and providing operating
instructions.
III. The Final Order
Based on the information discussed in the preamble to the proposed
order (83 FR 25598, June 4, 2018), the comments received for the
proposed order, prior panel discussions, and FDA's experiences over the
years in reviewing
[[Page 3547]]
these device types, FDA concludes that special controls, in conjunction
with general controls, will provide reasonable assurance of the safety
and effectiveness of medical image analyzers. FDA is adopting its
findings under section 513(f)(3) of the FD&C Act, as published in the
preamble to the proposed order (83 FR 25598). FDA is issuing this final
order to reclassify medical image analyzers, including CADe devices,
for mammography breast cancer, ultrasound breast lesions, radiograph
lung nodules, and radiograph dental caries detection from class III to
class II, and establishing special controls by revising 21 CFR part
892.\1\ In this final order, the Agency has identified the special
controls under section 513(a)(1)(B) of the FD&C Act that, together with
general controls, provide reasonable assurance of the safety and
effectiveness of these devices.
---------------------------------------------------------------------------
\1\ FDA notes that the ``ACTION'' caption for this final order
is styled as ``Final amendment; final order,'' rather than ``Final
order.'' Beginning in December 2019, this editorial change was made
to indicate that the document ``amends'' the Code of Federal
Regulations. The change was made in accordance with the Office of
Federal Register's (OFR) interpretations of the Federal Register Act
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. FDA has determined that premarket
notification is necessary to provide reasonable assurance of safety and
effectiveness of medical image analyzers, and therefore, this device
type is not exempt from premarket notification requirements. Persons
who intend to market a new CADe device must obtain clearance of a
premarket notification and demonstrate compliance with the special
controls included in this final order, prior to marketing the device.
The device is assigned the generic name medical image analyzer, and
identified as a medical image analyzers, including CADe devices, for
mammography breast cancer, ultrasound breast lesions, radiograph lung
nodules, and radiograph dental caries detection, is a prescription
device that is intended to identify, mark, highlight, or in any other
manner direct the clinicians' attention to portions of a radiology
image that may reveal abnormalities during interpretation of patient
radiology images by the clinicians. This device incorporates pattern
recognition and data analysis capabilities and operates on previously
acquired medical images. This device is not intended to replace the
review by a qualified radiologist and is not intended to be used for
triage or to recommend diagnosis.
Under this final order, the medical image analyzer is a
prescription use only device under Sec. 801.109 (21 CFR 801.109).
Prescription devices are exempt from the requirement for adequate
directions for use for the layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the
conditions of Sec. 801.109 are met (referring to 21 U.S.C. 352(f)(1)).
Under 21 CFR 807.81, the device continues to be subject to 510(k)
requirements.
IV. Codification of Orders
Prior to the amendments in the Food and Drug Administration Safety
and Innovation Act (FDASIA), section 513(e) of the FD&C Act provided
for FDA to issue regulations to reclassify devices. Although section
513(e) as amended requires FDA to issue final orders rather than
regulations, it also provides for FDA to revoke previously issued
regulations by order. FDA will continue to codify classifications and
reclassifications in the Code of Federal Regulations (CFR). Changes
resulting from final orders will appear in the CFR as changes to
codified classification determinations or as newly codified orders.
Therefore, under section 513(e)(1)(A)(i), as amended by FDASIA, in this
final order, we are proposing to codify the classification of medical
image analyzers, applied to CADe devices for mammography breast cancer,
ultrasound breast lesions, radiograph lung nodules, and radiograph
dental caries detection in the new Sec. 892.2070, under which medical
image analyzers would be reclassified into class II.
V. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
VI. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved FDA collections. These collections of information
are subject to review by the Office of Management and Budget (OMB)
under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The
collections of information in part 807, subpart E have been approved
under OMB control number 0910-0120; the collections of information in
21 CFR part 801 have been approved under OMB control number 0910-0485;
and the collections of information in part 820 have been approved under
OMB control number 0910-0073.
List of Subjects in 21 CFR Part 892
Medical devices, Radiation protection, X-rays.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
892 is amended as follows:
PART 892--RADIOLOGY DEVICES
0
1. The authority citation for part 892 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
[[Page 3548]]
0
2. Add Sec. 892.2070 to subpart B to read as follows:
Sec. 892.2070 Medical image analyzer.
(a) Identification. Medical image analyzers, including computer-
assisted/aided detection (CADe) devices for mammography breast cancer,
ultrasound breast lesions, radiograph lung nodules, and radiograph
dental caries detection, is a prescription device that is intended to
identify, mark, highlight, or in any other manner direct the
clinicians' attention to portions of a radiology image that may reveal
abnormalities during interpretation of patient radiology images by the
clinicians. This device incorporates pattern recognition and data
analysis capabilities and operates on previously acquired medical
images. This device is not intended to replace the review by a
qualified radiologist, and is not intended to be used for triage, or to
recommend diagnosis.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms
including a description of the algorithm inputs and outputs, each major
component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing
methods and dataset(s) used to assess whether the device will improve
reader performance as intended and to characterize the standalone
device performance. Performance testing includes one or more standalone
tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the
device improves reader performance in the intended use population when
used in accordance with the instructions for use. The performance
assessment must be based on appropriate diagnostic accuracy measures
(e.g., receiver operator characteristic plot, sensitivity, specificity,
predictive value, and diagnostic likelihood ratio). The test dataset
must contain a sufficient number of cases from important cohorts (e.g.,
subsets defined by clinically relevant confounders, effect modifiers,
concomitant diseases, and subsets defined by image acquisition
characteristics) such that the performance estimates and confidence
intervals of the device for these individual subsets can be
characterized for the intended use population and imaging equipment.
(iv) Appropriate software documentation (e.g., device hazard
analysis; software requirements specification document; software design
specification document; traceability analysis; description of
verification and validation activities including system level test
protocol, pass/fail criteria, and results; and cybersecurity).
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the
device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training
that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and
imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must
include situations in which the device may fail or may not operate at
its expected performance level (e.g., poor image quality or for certain
subpopulations), as applicable.
(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including:
test methods, dataset characteristics, results, and a summary of sub-
analyses on case distributions stratified by relevant confounders, such
as lesion and organ characteristics, disease stages, and imaging
equipment.
Dated: January 9, 2020.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2020-00494 Filed 1-21-20; 8:45 am]
BILLING CODE 4164-01-P