Medical Devices; Clinical Chemistry and Clinical Toxicology Devices; Classification of the Continuous Glucose Monitor Data Management System, 57816-57818 [2019-23471]
Download as PDF
57816
Federal Register / Vol. 84, No. 209 / Tuesday, October 29, 2019 / Rules and Regulations
(f) Compliance
Comply with this AD within the
compliance times specified, unless already
done.
(g) Required Actions
(1) Except for those engines identified in
paragraph (g)(2) of this AD, borescope inspect
the LPC R1 for damage and cracks at the
locations in paragraph (g)(1)(iv) of this AD as
follows:
(i) For engines that have accumulated
fewer than 300 flight cycles since new (CSN),
inspect within 50 flight cycles from
September 26, 2019 (the effective date of AD
2019–19–11).
(ii) For engines that have accumulated
fewer than 300 flight cycles since installation
of v2.11.7 or v2.11.8 electronic engine
control (EEC) software, inspect within 50
flight cycles from the effective date of this
AD.
(iii) Thereafter, at intervals not to exceed
50 flight cycles until the engine accumulates
300 flight CSN or accumulates 300 flight
cycles since the installation of v2.11.7 or
v2.11.8 EEC software, whichever occurs later,
repeat this borescope inspection for damage
and cracks at the locations in paragraph
(g)(1)(iv) of this AD.
(iv) Perform the borescope inspection
required by paragraphs (g)(1)(i) through (iii)
of this AD at the following locations:
(A) the blades tips;
(B) the leading edge;
(C) the leading edge fillet to rotor platform
radius; and
(D) the airfoil convex side root fillet to
rotor platform radius.
(2) For all affected PW model turbofan
engines installed as a ‘‘zero time spare,’’
except for PW1519G, PW1521GA and
PW1919G model turbofan engines, within 15
flight cycles from the effective date of this
AD, and thereafter at intervals not to exceed
15 flight cycles until the engine accumulates
300 flight CSN, perform the borescope
inspections required by paragraph (g)(1) of
this AD.
(3) As the result of the inspections required
by paragraphs (g)(1) and (2) of this AD, before
further flight, remove and replace the LPC if:
(i) there is damage on an LPC R1 that
exceeds serviceable limits; or
(ii) there is any crack in the LPC R1.
Note 1 to paragraph (g): Guidance on
determining serviceable limits can be found
in PW Service Bulletin (SB) PW1000G–A–
72–00–0125–00A–930A–D, Issue No. 002,
dated October 22, 2019, and PW SB
PW1000G–A–72–00–0075–00B–930A–D,
Issue No. 003, dated October 22, 2019.
(h) Definition
For the purpose of this AD, a ‘‘zero time
spare’’ is an engine that had zero flight hours
time-in-service when it was installed on an
airplane after the airplane had entered
service.
(i) Alternative Methods of Compliance
(AMOCs)
(1) The Manager, ECO Branch, FAA, has
the authority to approve AMOCs for this AD,
if requested using the procedures found in 14
CFR 39.19. In accordance with 14 CFR 39.19,
VerDate Sep<11>2014
17:42 Oct 28, 2019
Jkt 250001
send your request to your principal inspector
or local Flight Standards District Office, as
appropriate. If sending information directly
to the manager of the certification office,
send it to the attention of the person
identified in paragraph (j) of this AD. You
may email your request to: ANE–AD–AMOC@
faa.gov.
(2) Before using any approved AMOC,
notify your appropriate principal inspector,
or lacking a principal inspector, the manager
of the local flight standards district office/
certificate holding district office.
(j) Related Information
For more information about this AD,
contact Kevin M. Clark, Aerospace Engineer,
ECO Branch, FAA, 1200 District Avenue,
Burlington, MA 01803; phone: 781–238–
7088; fax: 781–238–7199; email:
kevin.m.clark@faa.gov.
(k) Material Incorporated by Reference
None.
Issued in Burlington, Massachusetts, on
October 25, 2019.
Karen M. Grant,
Acting Manager, Engine & Propeller
Standards Branch, Aircraft Certification
Service.
[FR Doc. 2019–23715 Filed 10–25–19; 4:15 pm]
BILLING CODE 4910–13–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 862
[Docket No. FDA–2019–N–2484]
Medical Devices; Clinical Chemistry
and Clinical Toxicology Devices;
Classification of the Continuous
Glucose Monitor Data Management
System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the continuous glucose
monitor data management system into
class I (general controls). We are taking
this action because we have determined
that classifying the device into class I
(general controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices,
in part by reducing regulatory burdens.
DATES: This order is effective October
29, 2019. The classification was
applicable on August 19, 2014.
FOR FURTHER INFORMATION CONTACT:
Ryan Lubert, Center for Devices and
Radiological Health, Food and Drug
SUMMARY:
PO 00000
Frm 00004
Fmt 4700
Sfmt 4700
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4545, Silver Spring,
MD 20993–0002, 240–402–6357,
ryan.lubert@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the
continuous glucose monitor data
management system as class I (general
controls), which we have determined
will provide a reasonable assurance of
safety and effectiveness. In addition, we
believe this action will enhance
patients’ access to beneficial innovation,
in part by reducing regulatory burdens
by placing the device into a lower
device class than the automatic class III
assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
by means of the procedures for
premarket notification under section
510(k) of the FD&C Act (21 U.S.C. 360(k)
and part 807 (21 CFR part 807).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
E:\FR\FM\29OCR1.SGM
29OCR1
Federal Register / Vol. 84, No. 209 / Tuesday, October 29, 2019 / Rules and Regulations
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act. Although the device was
automatically within class III, the De
Novo classification is considered to be
the initial classification of the device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or premarket
approval application in order to market
a substantially equivalent device (see 21
U.S.C. 360c(i), defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the 510(k) process, when necessary, to
market their device.
II. De Novo Classification
On April 22, 2014, DEXCOM, Inc.,
submitted a request for De Novo
classification of the STUDIO on the
Cloud Data Management Software. FDA
reviewed the request in order to classify
the device under the criteria for
classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class I if
general controls are sufficient to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
57817
information submitted in the request,
we determined that the device can be
classified into class I. FDA has
determined that general controls will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on August 19, 2014, FDA
issued an order to the requestor
classifying the device into class I. FDA
is codifying the classification of the
device by adding 21 CFR 862.2120. We
have named the generic type of device
continuous glucose monitor data
management system, and it is identified
as an electronic device intended to
acquire, process, and correlate
retrospective data from a continuous
glucose monitoring device. This device
is intended to be used by patients or
their healthcare providers when
determining therapeutic strategies. A
continuous glucose monitor data
management system is not a drug dose
calculator and does not provide
treatment recommendations.
FDA has identified the following risks
to health associated specifically with
this type of device in table 1.
TABLE 1—CONTINUOUS GLUCOSE MONITOR DATA MANAGEMENT SYSTEM RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
Device malfunction (e.g., incorrect data analysis, etc.) ......................................................................
Section 510(l)(1) of the FD&C Act
provides that a device within a type that
has been classified into class I under
section 513 of the FD&C Act is exempt
from premarket notification under
section 510(k), unless the device is of
substantial importance in preventing
impairment of human health or presents
a potentially unreasonable risk of illness
or injury (21 U.S.C. 360(l)(1)). Devices
within this type are exempt from the
premarket notification requirements
under section 510(k), subject to the
limitations of exemptions in 21 CFR
862.9.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order refers to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
VerDate Sep<11>2014
17:42 Oct 28, 2019
Jkt 250001
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; and
the collections of information in 21 CFR
part 820, regarding the quality system
regulation, including recordkeeping for
design controls, have been approved
under OMB control number 0910–0073.
List of Subjects in 21 CFR Part 862
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 862 is
amended as follows:
PO 00000
Frm 00005
Fmt 4700
Sfmt 4700
General controls, including design controls.
PART 862—CLINICAL CHEMISTRY
AND CLINICAL TOXICOLOGY
DEVICES
1. The authority citation for part 862
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 862.2120 to subpart C to read
as follows:
■
§ 862.2120 Continuous glucose monitor
data management system.
(a) Identification. A continuous
glucose monitor data management
system is an electronic device intended
to acquire, process, and correlate
retrospective data from a continuous
glucose monitoring device. This device
is intended to be used by patients or
their healthcare providers when
determining therapeutic strategies. A
continuous glucose monitor data
management system is not a drug dose
calculator and does not provide
treatment recommendations.
(b) Classification. Class I (general
controls). The device is exempt from the
premarket notification procedures in
subpart E of part 807 of this chapter,
subject to the limitations in § 862.9.
E:\FR\FM\29OCR1.SGM
29OCR1
57818
Federal Register / Vol. 84, No. 209 / Tuesday, October 29, 2019 / Rules and Regulations
Dated: October 23, 2019.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2019–23471 Filed 10–28–19; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Part 165
[Docket Number USCG–2019–0857]
RIN 1625–AA00
Safety Zone; Naval Training
Operations, U.S. Naval Magazine
Indian Island, WA
AGENCY:
ACTION:
Coast Guard, DHS.
Temporary final rule.
The Coast Guard is
establishing a temporary safety zone for
navigable waters within a 500-yard
radius of Walan Point, Indian Island,
WA. This safety zone is needed to
protect personnel, vessels, and the
marine environment from potential
hazards due to naval training
operations. Entry of vessels or persons
into this zone is prohibited unless
specifically authorized by the Captain of
the Port Puget Sound.
SUMMARY:
This rule is effective from 8 a.m.
on October 30, 2019, to 4 p.m. on
October 31, 2019, and will be subject to
enforcement each of these days from 8
a.m. to 4 p.m.
DATES:
To view documents
mentioned in this preamble as being
available in the docket, go to https://
www.regulations.gov, type USCG–2019–
0857 in the ‘‘SEARCH’’ box and click
‘‘SEARCH.’’ Click on Open Docket
Folder on the line associated with this
rule.
ADDRESSES:
If
you have questions on this rule, call or
email Lieutenant Ellie Wu, Sector Puget
Sound Waterways Management
Division, U.S. Coast Guard; telephone
(206) 217–6051, email
SectorPugetSoundWWM@uscg.mil.
FOR FURTHER INFORMATION CONTACT:
SUPPLEMENTARY INFORMATION:
I. Table of Abbreviations
17:42 Oct 28, 2019
The Coast Guard is issuing this
temporary rule without prior notice and
opportunity to comment pursuant to
authority under section 4(a) of the
Administrative Procedure Act (APA) (5
U.S.C. 553(b)). This provision
authorizes an agency to issue a rule
without prior notice and opportunity to
comment when the agency for good
cause finds that those procedures are
‘‘impracticable, unnecessary, or contrary
to the public interest.’’ Under 5 U.S.C.
553(b)(B), the Coast Guard finds that
good cause exists for not publishing a
notice of proposed rulemaking (NPRM)
with respect to this rule because issuing
an NPRM is impracticable. The Coast
Guard received notification of these
naval training operations from the U.S.
Navy on October 7, 2019, and we must
take action by October 30, 2019, to
protect the public from potential
hazards implicated by these training
operations. Delaying issuance of this
temporary final rule to publish an
NPRM and consider comments in
response to the NPRM is impracticable,
because the safety zone must be in place
for the operation, which begins on
October 30, 2019.
Under 5 U.S.C. 553(d)(3), the Coast
Guard finds that good cause exists for
making this rule effective less than 30
days after publication in the Federal
Register. Delaying the effective date of
this rule would be impracticable
because of the danger associated with
these training operations, which may
include but is not limited to high-speed
maneuvers, simulated attacks, and the
firing of blank ammunition. This rule
must be effective starting October 30,
2019, to protect vessels, personnel, and
the marine environment from potential
hazards associated with these training
operations.
III. Legal Authority and Need for Rule
The Coast Guard is issuing this rule
under authority in 46 U.S.C. 70034
(formerly 33 U.S.C. 1231). The Captain
of the Port Puget Sound (COTP) has
determined that potential hazards exist
with this naval training operation. This
rule is needed to protect personnel,
vessels, and the marine environment in
the navigable waters within the safety
zone from potential hazards posed by
the naval training operation.
IV. Discussion of the Rule
CFR Code of Federal Regulations
DHS Department of Homeland Security
FR Federal Register
NPRM Notice of proposed rulemaking
§ Section
U.S.C. United States Code
VerDate Sep<11>2014
II. Background Information and
Regulatory History
Jkt 250001
This rule establishes a safety zone
regulation from 10 a.m. on October 30,
2019, to 4 p.m. on October 31, 2019.
This regulation will only be subject to
enforcement for the following 8-hour
PO 00000
Frm 00006
Fmt 4700
Sfmt 4700
period each of these 2 days: 8 a.m. to 4
p.m.
The safety zone will cover navigable
waters within a 500-yard radius of
Walan Point, Indian Island.
The duration of this regulation is
intended to protect personnel, vessels,
and the marine environment in these
navigable waters while naval training
operations are taking place. No vessel or
person will be permitted to enter the
safety zone without obtaining
permission from the COTP or a
designated representative.
V. Regulatory Analyses
We developed this rule after
considering numerous statutes and
Executive orders related to rulemaking.
Below we summarize our analyses
based on a number of these statutes and
Executive orders, and we discuss First
Amendment rights of protestors.
A. Regulatory Planning and Review
Executive Orders 12866 and 13563
direct agencies to assess the costs and
benefits of available regulatory
alternatives and, if regulation is
necessary, to select regulatory
approaches that maximize net benefits.
Executive Order 13771 directs agencies
to control regulatory costs through a
budgeting process. This rule has not
been designated a ‘‘significant
regulatory action,’’ under Executive
Order 12866. Accordingly, this rule has
not been reviewed by the Office of
Management and Budget (OMB), and
pursuant to OMB guidance it is exempt
from the requirements of Executive
Order 13771.
This regulatory action determination
is based on the size, location, duration
of the safety zone. Vessel traffic will be
able to safely transit around this safety
zone which would impact a small
designated area of the waterway on the
western side of U.S. Naval Magazine
Indian Island. Moreover, the Coast
Guard will issue a Broadcast Notice to
Mariners via VHF–FM marine channel
16 about the zone, and the rule will
allow vessels to seek permission to enter
the zone.
B. Impact on Small Entities
The Regulatory Flexibility Act of
1980, 5 U.S.C. 601–612, as amended,
requires Federal agencies to consider
the potential impact of regulations on
small entities during rulemaking. The
term ‘‘small entities’’ comprises small
businesses, not-for-profit organizations
that are independently owned and
operated and are not dominant in their
fields, and governmental jurisdictions
with populations of less than 50,000.
The Coast Guard certifies under 5 U.S.C.
E:\FR\FM\29OCR1.SGM
29OCR1
Agencies
[Federal Register Volume 84, Number 209 (Tuesday, October 29, 2019)]
[Rules and Regulations]
[Pages 57816-57818]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2019-23471]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 862
[Docket No. FDA-2019-N-2484]
Medical Devices; Clinical Chemistry and Clinical Toxicology
Devices; Classification of the Continuous Glucose Monitor Data
Management System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the continuous glucose monitor data management system into class I
(general controls). We are taking this action because we have
determined that classifying the device into class I (general controls)
will provide a reasonable assurance of safety and effectiveness of the
device. We believe this action will also enhance patients' access to
beneficial innovative devices, in part by reducing regulatory burdens.
DATES: This order is effective October 29, 2019. The classification was
applicable on August 19, 2014.
FOR FURTHER INFORMATION CONTACT: Ryan Lubert, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4545, Silver Spring, MD 20993-0002, 240-402-6357,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the continuous glucose monitor
data management system as class I (general controls), which we have
determined will provide a reasonable assurance of safety and
effectiveness. In addition, we believe this action will enhance
patients' access to beneficial innovation, in part by reducing
regulatory burdens by placing the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate by means of the procedures
for premarket notification under section 510(k) of the FD&C Act (21
U.S.C. 360(k) and part 807 (21 CFR part 807).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After
[[Page 57817]]
receiving an order from FDA classifying the device into class III under
section 513(f)(1) of the FD&C Act, the person then requests a
classification under section 513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically within
class III, the De Novo classification is considered to be the initial
classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application in order to market a substantially
equivalent device (see 21 U.S.C. 360c(i), defining ``substantial
equivalence''). Instead, sponsors can use the 510(k) process, when
necessary, to market their device.
II. De Novo Classification
On April 22, 2014, DEXCOM, Inc., submitted a request for De Novo
classification of the STUDIO on the Cloud Data Management Software. FDA
reviewed the request in order to classify the device under the criteria
for classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class I if general controls are sufficient
to provide reasonable assurance of the safety and effectiveness of the
device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review
of the information submitted in the request, we determined that the
device can be classified into class I. FDA has determined that general
controls will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on August 19, 2014, FDA issued an order to the requestor
classifying the device into class I. FDA is codifying the
classification of the device by adding 21 CFR 862.2120. We have named
the generic type of device continuous glucose monitor data management
system, and it is identified as an electronic device intended to
acquire, process, and correlate retrospective data from a continuous
glucose monitoring device. This device is intended to be used by
patients or their healthcare providers when determining therapeutic
strategies. A continuous glucose monitor data management system is not
a drug dose calculator and does not provide treatment recommendations.
FDA has identified the following risks to health associated
specifically with this type of device in table 1.
Table 1--Continuous Glucose Monitor Data Management System Risks and Mitigation Measures
----------------------------------------------------------------------------------------------------------------
Identified risks Mitigation measures
----------------------------------------------------------------------------------------------------------------
Device malfunction (e.g., incorrect data General controls, including design controls.
analysis, etc.).
----------------------------------------------------------------------------------------------------------------
Section 510(l)(1) of the FD&C Act provides that a device within a
type that has been classified into class I under section 513 of the
FD&C Act is exempt from premarket notification under section 510(k),
unless the device is of substantial importance in preventing impairment
of human health or presents a potentially unreasonable risk of illness
or injury (21 U.S.C. 360(l)(1)). Devices within this type are exempt
from the premarket notification requirements under section 510(k),
subject to the limitations of exemptions in 21 CFR 862.9.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order refers to previously approved collections of
information found in other FDA regulations and guidance. These
collections of information are subject to review by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(44 U.S.C. 3501-3520). The collections of information in the guidance
document ``De Novo Classification Process (Evaluation of Automatic
Class III Designation)'' have been approved under OMB control number
0910-0844; the collections of information in 21 CFR part 814, subparts
A through E, regarding premarket approval, have been approved under OMB
control number 0910-0231; the collections of information in part 807,
subpart E, regarding premarket notification submissions, have been
approved under OMB control number 0910-0120; and the collections of
information in 21 CFR part 820, regarding the quality system
regulation, including recordkeeping for design controls, have been
approved under OMB control number 0910-0073.
List of Subjects in 21 CFR Part 862
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
862 is amended as follows:
PART 862--CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES
0
1. The authority citation for part 862 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 862.2120 to subpart C to read as follows:
Sec. 862.2120 Continuous glucose monitor data management system.
(a) Identification. A continuous glucose monitor data management
system is an electronic device intended to acquire, process, and
correlate retrospective data from a continuous glucose monitoring
device. This device is intended to be used by patients or their
healthcare providers when determining therapeutic strategies. A
continuous glucose monitor data management system is not a drug dose
calculator and does not provide treatment recommendations.
(b) Classification. Class I (general controls). The device is
exempt from the premarket notification procedures in subpart E of part
807 of this chapter, subject to the limitations in Sec. 862.9.
[[Page 57818]]
Dated: October 23, 2019.
Lowell J. Schiller,
Principal Associate Commissioner for Policy.
[FR Doc. 2019-23471 Filed 10-28-19; 8:45 am]
BILLING CODE 4164-01-P