Medical Devices; Anesthesiology Devices; Classification of the Positive Airway Pressure Delivery System, 52964-52966 [2018-22840]
Download as PDF
52964
Federal Register / Vol. 83, No. 203 / Friday, October 19, 2018 / Rules and Regulations
publish notice of a rulemaking in the
Federal Register and provide an
opportunity for public comment. This
requirement does not apply, however, if
the agency ‘‘for good cause finds . . .
that notice and public procedure are
impracticable, unnecessary, or contrary
to the public interest.’’ 11 The APA also
generally requires that an agency
publish an adopted rule in the Federal
Register at least 30 days before it
becomes effective. This requirement
does not apply, however, if the agency
finds good cause for making the rule
effective sooner.12
Given the temporary nature of the
relief contemplated by the interim final
temporary rules and the significant and
immediate impact of Hurricane Michael
and its aftermath on issuers in affected
areas, as discussed above, the
Commission finds that good cause exists
to dispense with notice and comment as
impracticable and unnecessary, and to
act immediately to amend Rule 202 of
Regulation Crowdfunding and Rule 257
of Regulation A.13 Further, the interim
final temporary rules will not affect the
burden or cost estimates associated with
existing collections of information
under Regulation Crowdfunding and
Regulation A for purposes of the
Paperwork Reduction Act of 1995.14
V. Statutory Basis and Text of
Amendments
We are adopting amendments to Rule
202 of Regulation Crowdfunding and
Rule 257 of Regulation A under the
authority set forth in the Securities Act
(15 U.S.C. 77a et seq.), particularly,
Section 28 thereof.
PART 227—REGULATION
CROWDFUNDING, GENERAL RULES
AND REGULATIONS
1. The authority citation for part 227
is revised to read as follows:
■
Authority: 15 U.S.C. 77d, 77d–1, 77s, 77z–
3, 78c, 78o, 78q, 78w, 78mm, and Pub. L.
112–106, secs. 301–305, 126 Stat. 306 (2012).
2. Amend § 227.202 by adding
paragraph (d) to read as follows:
■
§ 227.202
Ongoing reporting requirements.
*
*
*
*
*
(d) Temporary relief from certain
reporting requirements. (1) An issuer
that is not able to meet a filing deadline
for any report or form required to be
filed by this section, 17 CFR
227.203(a)(3)), or 17 CFR 227.203(b)
during the period from and including
October 10, 2018 to and including
November 21, 2018 due to Hurricane
Michael and its aftermath shall be
deemed to have satisfied the filing
deadline for such report or form if the
issuer files such report or form with the
Commission on or before November 23,
2018.
(2) In any report or form filed
pursuant to paragraph (d)(1) of this
section, the issuer must disclose that it
is relying on this paragraph (d) and state
the reasons why, in good faith, it could
not file such report or form on a timely
basis.
List of Subjects
PART 230—GENERAL RULES AND
REGULATIONS, SECURITIES ACT OF
1933
17 CFR Part 227
■
Crowdfunding, Funding portals,
Intermediaries, Reporting and
recordkeeping requirements, Securities.
17 CFR Part 230
Reporting and recordkeeping
requirements, Securities.
11 5
U.S.C. 553(b)(3)(B).
U.S.C. 553(d)(3).
13 This finding also satisfies the requirements of
5 U.S.C. 808(2), allowing the interim final
temporary rules to become effective
notwithstanding the requirement of 5 U.S.C. 801 (if
a federal agency finds that notice and public
comment are impractical, unnecessary or contrary
to the public interest, a rule shall take effect at such
time as the federal agency promulgating the rule
determines). The interim final temporary rules also
do not require analysis under the Regulatory
Flexibility Act. See 5 U.S.C. 604(a) (requiring a final
regulatory flexibility analysis only for rules
required by the APA or other law to undergo notice
and comment).
14 44 U.S.C. 3501 et seq.
12 5
amozie on DSK3GDR082PROD with RULES
In accordance with the foregoing, title
17, chapter II of the Code of Federal
Regulations is amended as follows:
VerDate Sep<11>2014
16:11 Oct 18, 2018
Jkt 247001
3. The authority citation for part 230
continues to read in part as follows:
Authority: 15 U.S.C. 77b, 77b note, 77c,
77d, 77f, 77g, 77h, 77j, 77r, 77s, 77z–3, 77sss,
78c, 78d, 78j, 78l, 78m, 78n, 78o, 78o–7 note,
78t, 78w, 78ll(d), 78mm, 80a–8, 80a–24, 80a–
28, 80a–29, 80a–30, and 80a–37, and Pub. L.
112–106, sec. 201(a), sec. 401, 126 Stat. 313
(2012), unless otherwise noted.
*
*
*
*
*
4. Amend § 230.257 by adding
paragraph (g) to read as follows:
■
§ 230.257 Periodic and current reporting;
exit report.
(g) Temporary relief from ongoing
reporting requirements. (1) An issuer
that is not able to meet a filing deadline
for any report or form required to be
filed by 17 CFR 230.252(f)(2)(i) or this
section during the period from and
including October 10, 2018 to and
including November 21, 2018 due to
Hurricane Michael and its aftermath
PO 00000
Frm 00022
Fmt 4700
Sfmt 4700
shall be deemed to have satisfied the
filing deadline for such report or form
if the issuer files such report or form
with the Commission on or before
November 23, 2018.
(2) In any report or form filed
pursuant to paragraph (g)(1) of this
section, the issuer must disclose that it
is relying on this paragraph (g) and state
the reasons why, in good faith, it could
not file such report or form on a timely
basis.
By the Commission.
Dated: October 16, 2018.
Brent J. Fields,
Secretary.
[FR Doc. 2018–22930 Filed 10–17–18; 4:15 pm]
BILLING CODE 8011–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 868
[Docket No. FDA–2018–N–3684]
Medical Devices; Anesthesiology
Devices; Classification of the Positive
Airway Pressure Delivery System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the positive airway pressure
delivery system into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the positive
airway pressure delivery system’s
classification. We are taking this action
because we have determined that
classifying the device into class II
(special controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices,
in part by reducing regulatory burdens.
DATES: This order is effective October
19, 2018. The classification was
applicable on June 5, 2018.
FOR FURTHER INFORMATION CONTACT:
Deepika Arora Lakhani, Center for
Devices and Radiological Health, Food
and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 2543,
Silver Spring, MD 20993–0002, 301–
796–4042, Deepika.Lakhani@
fda.hhs.gov.
SUMMARY:
SUPPLEMENTARY INFORMATION:
E:\FR\FM\19OCR1.SGM
19OCR1
Federal Register / Vol. 83, No. 203 / Friday, October 19, 2018 / Rules and Regulations
I. Background
Upon request, FDA has classified the
positive airway pressure delivery
system as class II (special controls),
which we have determined will provide
a reasonable assurance of safety and
effectiveness. In addition, we believe
this action will enhance patients’ access
to beneficial innovation, in part by
reducing regulatory burdens by placing
the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act to a
predicate device that does not require
premarket approval (see 21 U.S.C.
360c(i)). We determine whether a new
device is substantially equivalent to a
predicate by means of the procedures
for premarket notification under section
510(k) of the FD&C Act and 21 CFR part
807 (21 U.S.C. 360(k) and part 807,
respectively).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act (21 U.S.C. 360c(f)(2)). Section
207 of the Food and Drug
Administration Modernization Act of
1997 established the first procedure for
De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act (21 U.S.C. 360c(a)(1)).
Although the device was automatically
within class III, the De Novo
classification is considered to be the
initial classification of the device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or PMA in order to
market a substantially equivalent device
(see 21 U.S.C. 360c(i), defining
‘‘substantial equivalence’’). Instead,
sponsors can use the less-burdensome
510(k) process, when necessary, to
market their device.
52965
II. De Novo Classification
On December 14, 2017, FRESCA
Medical submitted a request for De
Novo classification of the CURVE TM
Positive Airway Pressure System. FDA
reviewed the request in order to classify
the device under the criteria for
classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on June 5, 2018, FDA
issued an order to the requester
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 868.5273. We
have named the generic type of device
positive airway pressure delivery
system, and it is identified as a
prescription noninvasive ventilatory
device that delivers expiratory positive
airway pressure for patients suffering
from obstructive sleep apnea. The
system also provides positive airway
pressure during incipient apnea. The
system may include a dedicated flow
generator and a patient interface.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
TABLE 1—POSITIVE AIRWAY PRESSURE DELIVERY SYSTEM RISKS AND MITIGATION MEASURES
amozie on DSK3GDR082PROD with RULES
Identified risks
Mitigation measures
Adverse tissue reaction ............................................................................
Electromagnetic interference with other devices .....................................
Infection ....................................................................................................
Device software failure leading to ineffective treatment ..........................
Device hardware failure/malfunction leading to high airway pressure,
carbon dioxide rebreathing or ineffective treatment.
Electrical shock injury or thermal injury ...................................................
Use error leading to ineffective therapy or patient injury .........................
VerDate Sep<11>2014
16:11 Oct 18, 2018
Jkt 247001
PO 00000
Frm 00023
Fmt 4700
Biocompatibility evaluation, and Labeling.
Electromagnetic compatibility testing, and Labeling.
Reprocessing validation, and Labeling.
Software verification, validation, and hazard analysis.
Non-clinical performance testing, and Labeling.
Electrical safety, thermal safety, and mechanical testing; Software
verification, validation, and hazard analysis; and Labeling.
Labeling.
Sfmt 4700
E:\FR\FM\19OCR1.SGM
19OCR1
52966
Federal Register / Vol. 83, No. 203 / Friday, October 19, 2018 / Rules and Regulations
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. In order for
a device to fall within this classification,
and thus avoid automatic classification
in class III, it would have to comply
with the special controls named in this
final order. The necessary special
controls appear in the regulation
codified by this order. This device is
subject to premarket notification
requirements under section 510(k) of the
FD&C Act.
At the time of classification, positive
airway pressure delivery systems are for
prescription use only. Prescription
devices are exempt from the
requirement for adequate directions for
use for the layperson under section
502(f)(1) of the FD&C Act and 21 CFR
801.5, as long as the conditions of 21
CFR 801.109 are met (referring to 21
U.S.C. 352(f)(1)).
amozie on DSK3GDR082PROD with RULES
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collection of
information in 21 CFR part 820,
regarding quality system regulation,
have been approved under OMB control
number 0910–0073; the collections of
information in part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120; and
the collections of information in part 21
CFR part 801, regarding labeling, have
been approved under OMB control
number 0910–0485.
VerDate Sep<11>2014
16:11 Oct 18, 2018
Jkt 247001
List of Subjects in 21 CFR Part 868
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, part 868 is amended
as follows:
PART 868—ANESTHESIOLOGY
DEVICES
1. The authority citation for part 868
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 868.5273 to subpart F to read
as follows:
■
§ 868.5273 Positive airway pressure
delivery system.
(a) Identification. A positive airway
pressure delivery system is a
prescription noninvasive ventilatory
device that delivers expiratory positive
airway pressure for patients suffering
from obstructive sleep apnea. The
system also provides positive airway
pressure during incipient apnea. The
system may include a dedicated flow
generator and a patient interface.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(2) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use, including the
following:
(i) Waveform testing must simulate
breathing conditions and evaluate
pressure and airflow response over a
range and combination of high and low
breath rates and tidal volumes.
(ii) Use life testing must demonstrate
adequate device performance over the
labeled use life of the device.
(iii) Device integrity testing must
demonstrate that the device can
withstand typical forces expected
during use.
(iv) Carbon dioxide rebreathing
testing must be performed.
(v) System flow rate, maximum
expiratory pressure, inhalation pressure,
and intra-mask static pressure testing
must be performed.
(vi) Air bolus testing must
demonstrate that the device can
withstand worst-case scenario air
pressures.
(vii) Maximum limited pressure
testing of the flow generator in single
fault condition must be performed.
(viii) Maximum output temperature
testing of delivered gas, if humidified,
must be performed.
PO 00000
Frm 00024
Fmt 4700
Sfmt 4700
(3) Performance data must validate
reprocessing instructions for any
reusable components of the device.
(4) Performance data must
demonstrate the electrical, thermal, and
mechanical safety and the
electromagnetic compatibility of the
device.
(5) Software verification, validation,
and hazard analysis must be performed.
(6) Labeling must include the
following:
(i) Therapy pressure range;
(ii) Use life and replacement schedule
for all components;
(iii) Cleaning instructions; and
(iv) Instructions for assembly and
connection of device components.
Dated: October 16, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–22840 Filed 10–18–18; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 878
[Docket No. FDA–2018–N–3696]
Medical Devices; General and Plastic
Surgery Devices; Classification of the
Wound Autofluorescence Imaging
Device
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the wound autofluorescence
imaging device into class I. We are
taking this action because we have
determined that classifying the device
into class I will provide a reasonable
assurance of safety and effectiveness of
the device. We believe this action will
also enhance patients’ access to
beneficial innovative devices, in part by
reducing regulatory burdens.
DATES: This order is effective October
19, 2018. The classification was
applicable on July 31, 2018.
FOR FURTHER INFORMATION CONTACT:
Yasaman Ardeshirpour, Center for
Devices and Radiological Health, Food
and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. G455,
Silver Spring, MD, 20993–0002, 240–
402–3706, Yasaman.Ardeshirpour@
fda.hhs.gov.
SUMMARY:
SUPPLEMENTARY INFORMATION:
E:\FR\FM\19OCR1.SGM
19OCR1
Agencies
[Federal Register Volume 83, Number 203 (Friday, October 19, 2018)]
[Rules and Regulations]
[Pages 52964-52966]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2018-22840]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 868
[Docket No. FDA-2018-N-3684]
Medical Devices; Anesthesiology Devices; Classification of the
Positive Airway Pressure Delivery System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the positive airway pressure delivery system into class II (special
controls). The special controls that apply to the device type are
identified in this order and will be part of the codified language for
the positive airway pressure delivery system's classification. We are
taking this action because we have determined that classifying the
device into class II (special controls) will provide a reasonable
assurance of safety and effectiveness of the device. We believe this
action will also enhance patients' access to beneficial innovative
devices, in part by reducing regulatory burdens.
DATES: This order is effective October 19, 2018. The classification was
applicable on June 5, 2018.
FOR FURTHER INFORMATION CONTACT: Deepika Arora Lakhani, Center for
Devices and Radiological Health, Food and Drug Administration, 10903
New Hampshire Ave., Bldg. 66, Rm. 2543, Silver Spring, MD 20993-0002,
301-796-4042, [email protected].
SUPPLEMENTARY INFORMATION:
[[Page 52965]]
I. Background
Upon request, FDA has classified the positive airway pressure
delivery system as class II (special controls), which we have
determined will provide a reasonable assurance of safety and
effectiveness. In addition, we believe this action will enhance
patients' access to beneficial innovation, in part by reducing
regulatory burdens by placing the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act to a predicate device that does not require
premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new
device is substantially equivalent to a predicate by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act and 21 CFR part 807 (21 U.S.C. 360(k) and part 807, respectively).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act (21 U.S.C. 360c(f)(2)). Section 207 of the Food and Drug
Administration Modernization Act of 1997 established the first
procedure for De Novo classification (Pub. L. 105-115). Section 607 of
the Food and Drug Administration Safety and Innovation Act modified the
De Novo application process by adding a second procedure (Pub. L. 112-
144). A device sponsor may utilize either procedure for De Novo
classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act (21 U.S.C. 360c(a)(1)). Although the device
was automatically within class III, the De Novo classification is
considered to be the initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or PMA in order to market a substantially equivalent device (see 21
U.S.C. 360c(i), defining ``substantial equivalence''). Instead,
sponsors can use the less-burdensome 510(k) process, when necessary, to
market their device.
II. De Novo Classification
On December 14, 2017, FRESCA Medical submitted a request for De
Novo classification of the CURVE TM Positive Airway Pressure
System. FDA reviewed the request in order to classify the device under
the criteria for classification set forth in section 513(a)(1) of the
FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on June 5, 2018, FDA issued an order to the requester
classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 868.5273. We have named
the generic type of device positive airway pressure delivery system,
and it is identified as a prescription noninvasive ventilatory device
that delivers expiratory positive airway pressure for patients
suffering from obstructive sleep apnea. The system also provides
positive airway pressure during incipient apnea. The system may include
a dedicated flow generator and a patient interface.
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Positive Airway Pressure Delivery System Risks and Mitigation
Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Adverse tissue reaction................ Biocompatibility evaluation,
and Labeling.
Electromagnetic interference with other Electromagnetic compatibility
devices. testing, and Labeling.
Infection.............................. Reprocessing validation, and
Labeling.
Device software failure leading to Software verification,
ineffective treatment. validation, and hazard
analysis.
Device hardware failure/malfunction Non-clinical performance
leading to high airway pressure, testing, and Labeling.
carbon dioxide rebreathing or
ineffective treatment.
Electrical shock injury or thermal Electrical safety, thermal
injury. safety, and mechanical
testing; Software
verification, validation, and
hazard analysis; and Labeling.
Use error leading to ineffective Labeling.
therapy or patient injury.
------------------------------------------------------------------------
[[Page 52966]]
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. In order for a device to fall
within this classification, and thus avoid automatic classification in
class III, it would have to comply with the special controls named in
this final order. The necessary special controls appear in the
regulation codified by this order. This device is subject to premarket
notification requirements under section 510(k) of the FD&C Act.
At the time of classification, positive airway pressure delivery
systems are for prescription use only. Prescription devices are exempt
from the requirement for adequate directions for use for the layperson
under section 502(f)(1) of the FD&C Act and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are met (referring to 21 U.S.C.
352(f)(1)).
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections
of information in the guidance document ``De Novo Classification
Process (Evaluation of Automatic Class III Designation)'' have been
approved under OMB control number 0910-0844; the collections of
information in 21 CFR part 814, subparts A through E, regarding
premarket approval, have been approved under OMB control number 0910-
0231; the collection of information in 21 CFR part 820, regarding
quality system regulation, have been approved under OMB control number
0910-0073; the collections of information in part 807, subpart E,
regarding premarket notification submissions, have been approved under
OMB control number 0910-0120; and the collections of information in
part 21 CFR part 801, regarding labeling, have been approved under OMB
control number 0910-0485.
List of Subjects in 21 CFR Part 868
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, part 868 is
amended as follows:
PART 868--ANESTHESIOLOGY DEVICES
0
1. The authority citation for part 868 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 868.5273 to subpart F to read as follows:
Sec. 868.5273 Positive airway pressure delivery system.
(a) Identification. A positive airway pressure delivery system is a
prescription noninvasive ventilatory device that delivers expiratory
positive airway pressure for patients suffering from obstructive sleep
apnea. The system also provides positive airway pressure during
incipient apnea. The system may include a dedicated flow generator and
a patient interface.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The patient-contacting components of the device must be
demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use,
including the following:
(i) Waveform testing must simulate breathing conditions and
evaluate pressure and airflow response over a range and combination of
high and low breath rates and tidal volumes.
(ii) Use life testing must demonstrate adequate device performance
over the labeled use life of the device.
(iii) Device integrity testing must demonstrate that the device can
withstand typical forces expected during use.
(iv) Carbon dioxide rebreathing testing must be performed.
(v) System flow rate, maximum expiratory pressure, inhalation
pressure, and intra-mask static pressure testing must be performed.
(vi) Air bolus testing must demonstrate that the device can
withstand worst-case scenario air pressures.
(vii) Maximum limited pressure testing of the flow generator in
single fault condition must be performed.
(viii) Maximum output temperature testing of delivered gas, if
humidified, must be performed.
(3) Performance data must validate reprocessing instructions for
any reusable components of the device.
(4) Performance data must demonstrate the electrical, thermal, and
mechanical safety and the electromagnetic compatibility of the device.
(5) Software verification, validation, and hazard analysis must be
performed.
(6) Labeling must include the following:
(i) Therapy pressure range;
(ii) Use life and replacement schedule for all components;
(iii) Cleaning instructions; and
(iv) Instructions for assembly and connection of device components.
Dated: October 16, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-22840 Filed 10-18-18; 8:45 am]
BILLING CODE 4164-01-P