Ophthalmic Devices; Reclassification of Ultrasound Cyclodestructive Device, 48403-48408 [2018-20763]
Download as PDF
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
for the Board to solicit comment on the
effect of the proposal on small entities.
The Board will, if necessary, conduct a
final regulatory flexibility analysis after
consideration of comments received
during the public comment period.
1. Statement of the need for, and
objectives of, the proposed rule. Title X
of the Dodd-Frank Act transferred
rulemaking authority for the S.A.F.E.
Act and other enumerated consumer
financial protection laws from the Board
to the Bureau, effective July 21, 2011. In
December 2011, the Bureau issued an
Interim Final Rule to incorporate the
S.A.F.E. Act pursuant to the transfer of
rulemaking authority. Although the
Board retains authority to issue some
consumer financial protection rules, all
rulemaking authority under the S.A.F.E.
Act concerning mortgage loan originator
registration was transferred to the
Bureau. Consequently, the Board is
proposing to repeal its regulations that
incorporated the S.A.F.E. Act.
2. Small entities affected by the
proposed rule. Any entity that is
currently covered by the S.A.F.E. Act is
subject to the rules issued by the
Bureau, located in 12 CFR part 1007 and
1008. Therefore the Board’s repeal of its
regulations that incorporated the
S.A.F.E. Act would not affect any entity,
including small entities.
3. Recordkeeping, reporting, and
compliance requirements. The proposed
rule would repeal parts of the Board’s
regulations that incorporated the
S.A.F.E. Act, and would therefore not
impose any recordkeeping, reporting, or
compliance requirements on any
entities.
4. Other Federal Rules. The Board has
not identified any federal rules that
duplicate, overlap, or conflict with the
proposed repeal of the Board’s
regulations that incorporated the
S.A.F.E. Act.
5. Significant alternatives to the
proposed revisions. The Board is not
aware of any significant alternatives that
would further minimize the impact on
small entities of the proposed repeal,
but solicits comment on this approach.
daltland on DSKBBV9HB2PROD with PROPOSALS
III. Paperwork Reduction Act
In accordance with the Paperwork
Reduction Act (PRA) of 1995 (44 U.S.C.
3506; 5 CFR 1320 Appendix A.1), the
Board reviewed the rule under the
authority delegated to the Federal
Reserve by the Office of Management
and Budget (OMB). The proposed rule
contains no collections of information
under the PRA. See 44 U.S.C. 3502(3).
Accordingly, there is no paperwork
burden associated with the proposed
rule.
VerDate Sep<11>2014
16:45 Sep 24, 2018
Jkt 244001
List of Subjects
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
12 CFR Part 208
Accounting, Agriculture, Banks,
Banking, Confidential business
information, Consumer protection,
Crime, Currency, Insurance,
Investments, Mortgages, Reporting and
recordkeeping requirements, Securities.
12 CFR Part 211
Authority and Issuance
For the reasons set forth in the
preamble, the Board proposes to amend
chapter II of title 12 of the Code of
Federal Regulations as follows:
PART 208—MEMBERSHIP OF STATE
BANKING INSTITUTIONS IN THE
FEDERAL RESERVE SYSTEM
(REGULATION H)
1. The authority citation for part 208
continues to read as follows:
■
Authority: 12 U.S.C. 24, 36, 92a, 93a,
248(a), 248(c), 321–338a, 371d, 461, 481–486,
601, 611, 1814, 1816, 1818, 1820(d)(9),
1833(j), 1828(o), 1831, 1831o, 1831p–1,
1831r–1, 1831w, 1831x, 1835a, 1882, 2901–
2907, 3105, 3310, 3331–3351, 3905–3909,
and 5371; 15 U.S.C. 78b, 78I(b), 78l(i), 780–
4(c)(5), 78q, 78q–1, and 78w, 1681s, 1681w,
6801, and 6805; 31 U.S.C. 5318; 42 U.S.C.
4012a, 4104a, 4104b, 4106 and 4128.
Subpart I—[Removed and Reserved]
2. Remove and reserve subpart I,
consisting of §§ 208.101 through
208.105 and Appendix A.
■
PART 211—INTERNATIONAL
BANKING OPERATIONS
(REGULATION K)
3. The authority citation for part 211
continues to read as follows:
Authority: 12 U.S.C. 221 et seq., 1818,
1835a, 1841 et seq., 3101 et seq., 3901 et seq.,
and 5101 et seq.; 15 U.S.C. 1681s, 1681w,
6801 and 6805.
4. In § 211.24, remove paragraph (k).
By order of the Board of Governors of the
Federal Reserve System, September 20, 2018.
Ann Misback,
Secretary of the Board.
[FR Doc. 2018–20832 Filed 9–24–18; 8:45 am]
Fmt 4702
[Docket No. FDA–2018–N–3074]
Ophthalmic Devices; Reclassification
of Ultrasound Cyclodestructive Device
Food and Drug Administration,
HHS.
ACTION:
Proposed order.
The Food and Drug
Administration (FDA) is issuing this
proposed order to reclassify the
ultrasound cyclodestructive device, a
postamendments class III device
(regulated under product code LZR),
into class II (special controls), subject to
premarket notification. FDA is also
identifying the proposed special
controls that the Agency believes are
necessary to provide a reasonable
assurance of safety and effectiveness of
the device. FDA is proposing this
reclassification on its own initiative
based on new information. If finalized,
this order will reclassify these devices
from class III to class II (special
controls) and reduce regulatory burdens
as these types of devices will no longer
be required to submit a premarket
approval application (PMA) but can
instead submit a less burdensome
premarket notification (510(k)) before
marketing their device.
DATES: Submit either electronic or
written comments on the proposed
order by November 26, 2018. Please see
section XI for the proposed effective
date when the new requirements apply
and for the proposed effective date of a
final order based on this proposed
order.
SUMMARY:
You may submit comments
as follows. Please note that late,
untimely filed comments will not be
considered. Electronic comments must
be submitted on or before November 26,
2018. The https://www.regulations.gov
electronic filing system will accept
comments until midnight Eastern Time
at the end of November 26, 2018.
Comments received by mail/hand
delivery/courier (for written/paper
submissions) will be considered timely
if they are postmarked or the delivery
service acceptance receipt is on or
before that date.
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal:
https://www.regulations.gov. Follow the
BILLING CODE 6210–01–P
Frm 00002
21 CFR Part 886
ADDRESSES:
■
PO 00000
Food and Drug Administration
AGENCY:
Exports, Foreign banking, Holding
companies, Investments, Reporting and
recordkeeping requirements.
■
48403
Sfmt 4702
E:\FR\FM\25SEP1.SGM
25SEP1
48404
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
daltland on DSKBBV9HB2PROD with PROPOSALS
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand Delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2018–N–3074 for ‘‘Ophthalmic Devices;
Reclassification of Ultrasound
Cyclodestructive Device.’’ Received
comments, those filed in a timely
manner (see ADDRESSES), will be placed
in the docket and, except for those
submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Dockets Management Staff between 9
a.m. and 4 p.m., Monday through
Friday.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
VerDate Sep<11>2014
16:45 Sep 24, 2018
Jkt 244001
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://www.gpo.gov/
fdsys/pkg/FR-2015-09-18/pdf/201523389.pdf.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Hina Pinto, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1652, Silver Spring,
MD 20993, 301–796–6351, hina.pinto@
fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (FD&C Act), as amended, establishes
a comprehensive system for the
regulation of medical devices intended
for human use. Section 513 of the FD&C
Act (21 U.S.C. 360c) established three
categories (classes) of devices, reflecting
the regulatory controls needed to
provide reasonable assurance of their
safety and effectiveness. The three
categories of devices are class I (general
controls), class II (special controls), and
class III (premarket approval).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices) are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II, or FDA issues an order finding
PO 00000
Frm 00003
Fmt 4702
Sfmt 4702
the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and 21 CFR
part 807.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or class
II under section 513(f)(3) of the FD&C
Act. Section 513(f)(3) of the FD&C Act
provides that FDA acting by order can
reclassify the device into class I or class
II on its own initiative, or in response
to a petition from the manufacturer or
importer of the device. To change the
classification of the device, the
proposed new class must have sufficient
regulatory controls to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use.
Reevaluation of the data previously
before the Agency is an appropriate
basis for subsequent action where the
reevaluation is made in light of newly
available regulatory authority (see Bell
v. Goddard, 366 F.2d 177, 181 (7th Cir.
1966); Ethicon, Inc. v. FDA, 762 F.
Supp. 382, 388–391 (D.D.C. 1991)), or in
light of changes in ‘‘medical science’’
(Upjohn v. Finch, 422 F.2d 944, 951 (6th
Cir. 1970)). Whether data before the
Agency are old or new, the ‘‘new
information’’ to support reclassification
under section 513(f)(3) of the FD&C Act
must be ‘‘valid scientific evidence’’, as
defined in section 513(a)(3) of the FD&C
Act and 21 CFR 860.7(c)(2). (See, e.g.,
General Medical Co. v. FDA, 770 F.2d
214 (D.C. Cir. 1985); Contact Lens
Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.
1985), cert. denied, 474 U.S. 1062
(1986)).
FDA relies upon ‘‘valid scientific
evidence’’ in the classification process
to determine the level of regulation for
devices. To be considered in the
reclassification process, the ‘‘valid
scientific evidence’’ upon which the
Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA (see
section 520(c) of the FD&C Act (21
U.S.C. 360j(c)). Section 510(m) of the
FD&C Act provides that a class II device
may be exempted from the 510(k)
premarket notification requirements, if
the Agency determines that premarket
notification is not necessary to
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
reasonably assure the safety and
effectiveness of the device.
glaucoma population as class II devices,
subject to 510(k) requirements.
II. Regulatory History of the Devices
IV. Proposed Reclassification
On April 29, 2015, FDA published a
document in the Federal Register
entitled ‘‘Retrospective Review of
Premarket Approval Application
Devices; Striking the Balance Between
Premarket and Postmarket Data
Collection,’’ in which FDA announced
plans to consider reclassifying
ultrasound cyclodestructive devices
identified with the LZR product code
from class III to class II (80 FR 23798)
and requested comments. FDA received
no adverse comments regarding our
proposed intent for LZR.
In accordance with section 513(f)(3) of
the FD&C Act and 21 CFR part 860,
subpart C, FDA is proposing to
reclassify this postamendments class III
device into class II. FDA believes that
there is sufficient information available
to FDA through peer-reviewed literature
and knowledge of similar devices to
establish special controls that would
effectively mitigate the risks to health
identified in section V. Absent the
special controls identified in this
proposed order, general controls
applicable to the device are insufficient
to provide reasonable assurance of the
safety and effectiveness of the device.
FDA is proposing to create a separate
classification regulation for ultrasound
cyclodestructive devices that will be
reclassified from class III to II. Under
this proposed order, if finalized, the
ultrasound cyclodestructive devices will
be identified as a prescription device.
As such, the prescription device must
satisfy prescription labeling
requirements (see § 801.109 (21 CFR
801.109), Prescription devices).
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and § 801.5 (21 CFR 801.5), as
long as the conditions of § 801.109 are
met (referring to 21 U.S.C. 352(f)(1)). In
this proposed order, if finalized, the
Agency has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, will provide a reasonable
assurance of the safety and effectiveness
for ultrasound cyclodestructive devices.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
On June 30, 1988, FDA approved the
first and only ultrasound
cyclodestructive device through its
PMA process under section 515 of the
FD&C Act (21 U.S.C. 360e). On August
10, 1988 (53 FR 30101), FDA announced
a PMA order for Sonocare Inc.’s Model
CST–100 Therapeutic Ultrasound
System (Sonocare CST–100) (Ref. 1). As
of the date of issuance of this proposed
order, the Sonocare CST–100 is the only
PMA approved by FDA for this device
type.
daltland on DSKBBV9HB2PROD with PROPOSALS
III. Device Description
An ultrasound cyclodestructive
device is a postamendments device
classified into class III under section
513(f)(1) of the FD&C Act. An
ultrasound cyclodestructive device is
indicated for the treatment of refractory
glaucoma; it is intended for patients
who are refractory to or are poor
candidates for laser or surgical
treatment and fail to achieve target
intraocular pressures on maximally
tolerated drug therapy. The device is
designed to reduce intraocular pressure
by producing a series of lesions in the
ciliary body and trabecular meshwork
induced by high intensity focused
ultrasound (HIFU) energy. Different
technologies, such as laser
cyclodestruction to lower intraocular
pressure, have been studied since the
1970s (Refs. 2 and 3), which increases
FDA’s knowledge base for devices used
to treat this condition. As stated earlier
in section II, FDA has approved only
one ultrasound cyclodestructive device
through its PMA process under section
515 of the FD&C Act (Ref. 4). More
recently, reports in the literature
indicate that the HIFU technology has
been modified and currently studied in
Europe for treatment of refractory
glaucoma (Refs. 5 to 8). Based upon our
review experience and consistent with
the FD&C Act and FDA’s regulations,
FDA believes that these devices should
be reclassified from class III into class
II because there is sufficient information
to establish special controls that can
provide reasonable assurance of the
device’s safety and effectiveness.
Conventional refractory glaucoma
treatment modalities include
implantable aqueous shunts and valves,
trabeculectomy and other incisional
glaucoma surgeries, cyclocryotherapy,
as well as laser transcleral
cyclophotocoagulation. FDA currently
regulates all of the devices indicated for
these procedures in a refractory
VerDate Sep<11>2014
16:45 Sep 24, 2018
Jkt 244001
PO 00000
Frm 00004
Fmt 4702
Sfmt 4702
48405
is necessary for ultrasound
cyclodestructive devices to provide
reasonable assurance of safety and
effectiveness. Therefore, the Agency
does not intend to exempt these
proposed class II devices from 510(k)
requirements. Persons who intend to
market this type of device must submit
to FDA a 510(k) and receive clearance
prior to marketing the device.
This proposal, if finalized, will
decrease regulatory burden and will
reduce private costs and expenditures
required to comply with Federal
regulations. Specifically, regulated
industry will no longer have to submit
a PMA but can instead submit a 510(k)
to the Agency for review prior to
marketing their device. A 510(k) is a less
burdensome pathway to market a
device, which typically results in a
more timely premarket review
compared to a PMA and reduces the
regulatory burden in addition to
providing more timely access of these
types of devices to patients.
V. Risks to Health
After considering the information
available to FDA through the review
submission, peer-reviewed literature,
and knowledge of other technologies
indicated to treat the same refractory
glaucoma patient population (such as
aqueous shunt and cryotherapy), FDA
determined that the probable risks to
health associated with the use of
ultrasound cyclodestructive devices for
treatment of refractory glaucoma are as
follows:
• Thermal Injury. Exposure of the
ocular tissue to the HIFU energy causes
thermal damage of the tissue. The
misdirection or misalignment of the
beam may cause temperature elevation
in the non-target ocular tissues and
overall ocular tissue damage. Unsuitable
power and duration of the beam may
also result in temperature elevation,
which may cause corneoscleral lesions
including scleral thinning, corneal
ectasia and perforation, eyelid burns,
corneal burns, clouding of the cornea
(haze) and lens (cataract formation), and
retinal and choroidal lesions.
• Physical Injury. Exposure of the
ocular tissue to the HIFU energy can
cause physical damage to the ocular
tissue due to cavitation or other
mechanical effects. These injuries could
be caused by the suboptimal selection of
the treatment parameters, misalignment/
displacement of the probe during the
treatment, device malfunction, or other
factors affecting stability of treatment.
For example, insonification of the
zonular fibers may cause elongation or
rupture of ligaments, which can lead to
a displacement of the lens.
E:\FR\FM\25SEP1.SGM
25SEP1
48406
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
• Post-treatment injury. Posttreatment injury from use of the device
may include intraocular inflammation
(e.g., iritis, uveitis), increased
intraocular pressure in the immediate
post-treatment period, ciliary body
hemorrhage, persistent or transient low
pressure, decreased visual acuity,
worsening glaucoma, phthisis, pain/
discomfort, corneal edema, hyphema,
retinal and choroidal complications, etc.
• Electrical shock. While in
operation, the device may discharge
electricity that could shock the user.
Electrical shock can be caused by use
error or device malfunction.
• Electromagnetic interference. While
in operation, electromagnetic
interferences from other devices
operated in the same environment may
cause the device to malfunction, which
could result in patient’s injury. In
addition, the device may interfere with
other electrically powered devices,
causing them to malfunction.
• Ocular irritation and corneal
infections. Inadequate biocompatibility
of the eye contact components of the
device can lead to irritation of the
ocular tissue. Inappropriately sterilized
or reprocessed eye contact components
of the device can lead to inflammation
and corneal infections.
daltland on DSKBBV9HB2PROD with PROPOSALS
VI. Summary of Reasons for
Reclassification
16:45 Sep 24, 2018
FDA believes that the identified
special controls, in addition to general
controls, are necessary to provide
reasonable assurance of safety and
effectiveness of these devices. Taking
into account the probable health
benefits of the use of the device and the
nature and known incidence of the risks
of the device, FDA, on its own initiative,
is proposing to reclassify this
postamendments class III device into
class II. FDA has considered and
analyzed the following information: An
inclusive search of the Agency’s
Manufacturer and User Facility Device
Experience (MAUDE) database, which
shows no adverse events for ultrasound
cyclodestructive device type; no recalls
have been received for this device type;
other technologies indicated to treat the
same refractory glaucoma patient
population, such as aqueous shunt and
cryotherapy, and currently regulated as
class II devices to compare the probable
risks (i.e., between the rate and severity
of the adverse events associated with
these class II technologies and
ultrasound cyclodestructive
procedures); and peer-reviewed
publications (Refs. 5 to 12) to identify
probable device risks (e.g., the types and
rates of adverse events) and mitigation
strategies.
VIII. Proposed Special Controls
FDA believes that the ultrasound
cyclodestructive devices for treatment of
refractory glaucoma should be
reclassified from class III to class II in
light of available information about the
effectiveness of these devices. There is
sufficient information to establish
special controls for ultrasound
cyclodestructive devices, in addition to
general controls, which can provide
reasonable assurance of safety and
effectiveness of the device, as general
controls themselves are insufficient to
provide reasonable assurance of its
safety and effectiveness. FDA believes
that the risks to health associated with
ultrasound cyclodestructive devices for
treatment of refractory glaucoma can be
mitigated with special controls and that
these mitigations will provide a
reasonable assurance of its safety and
effectiveness.
Based on a reconsideration of the
available information and data, FDA
believes that there is valid scientific
evidence of effectiveness for ultrasound
cyclodestructive devices to reduce
intraocular pressure intended for
treatment of refractory glaucoma using
ultrasound.
VerDate Sep<11>2014
VII. Summary of Data Upon Which the
Reclassification Is Based
Jkt 244001
FDA believes that the following
special controls, together with general
controls, are necessary and sufficient to
mitigate the risks to health described in
section V and provide a reasonable
assurance of safety and effectiveness for
ultrasound cyclodestructive devices.
• Non-clinical performance testing of
device features and characteristics will
demonstrate:
Æ The ability of the device to deposit
controllable HIFU energy to the target
area to evoke the required level of
thermal lesion.
Æ The design and geometry of the
HIFU transducer and the output
characteristics of the HIFU generator,
including operating frequency and
power, produce a small focal zone and
a steep transition of energy deposition
between the focal zone and the
untreated areas. In addition, the total
acoustic power radiated by the
transducer(s), spatial distribution of the
ultrasound field (including
compressional and rarefactional
pressure), and spatial peak, temporalaverage intensity will be evaluated. This
may be accomplished by demonstrating
compliance with the standard
International Electrotechnical
PO 00000
Frm 00005
Fmt 4702
Sfmt 4702
Commission (IEC) Technical
Specification (TS) 62556: Ultrasonics—
Field characterization—Specification
and measurement of field parameters for
high intensity therapeutic ultrasound
(HITU) transducers and systems.
Thermal and physical (due to potential
cavitation of gas bubbles) safety
analyses will also be evaluated.
Æ The appropriate alignment and
focusing of the ultrasound beam to the
target tissue to minimize unintended
damage to adjacent ocular tissues.
Æ The function of all safety features
built into the device, including the
energy monitoring system.
• Clinical performance data will
validate device performance and
characterize ocular tissue thermal
injuries, physical injury, and
postoperative adverse events by
establishing the treatment parameters
for which the device is safe and
effective.
• Electrical safety testing will
minimize the risk of electrical shock,
thermal or physical injury to the patient
and healthcare provider. This may be
accomplished by demonstrating
compliance with FDA-recognized
consensus standard American National
Standards Institute (ANSI)/Association
for the Advancement of Medical
Instrumentation (AAMI) 60601–1:
Medical electrical equipment, Part 1:
General requirements for basic safety
and essential performance.
• Electromagnetic compatibility
testing ensures that electromagnetic
interferences do not cause device
malfunction. It can also provide
assurance that electromagnetic
interferences generated by the device do
not affect the other devices operated in
the same environment. This may be
accomplished by demonstrating
compliance with FDA-recognized
consensus standard IEC 60601–1–2:
Medical electrical equipment, Part 1–2:
General requirements for safety. If the
device incorporates radiofrequency (RF)
wireless technology to perform medical
functions and/or communicates medical
data, testing will mitigate the risks
associated with interference or
degradation when using RF wireless
technology. This may be accomplished
by demonstrating compliance with
FDA-recognized consensus standard
AAMI TIR69: Risk Management of
Radio-frequency Wireless Coexistence
for Medical Devices and Systems (risk
assessment) and ANSI/Institute of
Electrical and Electronics Engineers
(IEEE) C63.27: American National
Standard for Evaluation of Wireless
Coexistence (coexistence testing).
• Software verification, validation,
and hazard analysis is necessary to
E:\FR\FM\25SEP1.SGM
25SEP1
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
mitigate the risks of thermal and
physical injury and ensures that
software performs as intended and
potential software malfunctions do not
impact the safety or effectiveness of the
device. If the device incorporates
internet network connectivity, testing
will demonstrate that cybersecurity
concerns are mitigated (e.g., data
integrity, unauthorized access, etc.).
• Biocompatibility evaluation can
help mitigate the risk of ocular irritation
and corneal infection by ensuring that
the patient-contacting components of
the device are safe for contact with skin
and ocular tissue.
• Sterilization validation, for devices
provided sterile, and/or cleaning
validation, for devices or components
that are reusable, help mitigate the risk
of inflammation and corneal infections
(e.g., keratitis).
• The labeling will also include
necessary information to ensure safe
and effective use of the ultrasound
cyclodestructive device and minimize
probability of the ocular treatmentrelated adverse events. Labeling needs
to include sufficient information that
will help the patient and healthcare
provider make an informed decision
regarding treatment-related adverse
effects of the ultrasound
cyclodestructive treatment. For
example, the labeling needs to include
information regarding the most common
reported treatment-related injuries,
which may include intraocular
inflammation (e.g., iritis, uveitis) and
increased intraocular pressure in the
immediate post-treatment period.
Adverse event information related to
ciliary body hemorrhage, persistent low
pressure, decreased visual acuity,
worsening glaucoma, phthisis, pain/
discomfort, transient low pressure,
corneal edema, hyphema, retinal
complications (such as cystoid macula
edema), and choroidal effusion or
detachment need to be discussed. The
labeling will mitigate the risk associated
with the intraoperative events, such as
pain/discomfort, and postoperative
adverse events by providing appropriate
clinical information along with
mitigation strategies (e.g., retrobulbar or
peribulbar anesthesia). Specifically,
device labeling must include:
Æ Appropriate warnings and
precautions to ensure safe and effective
48407
use of the device and minimize
potential device malfunctions and user
errors.
Æ A summary of the clinical
evaluation pertinent to use of the
device, including study outcomes and
adverse events.
Æ Information regarding procedure
parameters, proper positioning of the
HIFU transducer and its coupling with
the eye, and typical course of treatment
to ensure the user can safely operate the
device.
Æ Validated reprocessing instructions
to ensure the safe use of reusable device
components.
Æ Safety information regarding
electrical safety and electromagnetic
compatibility to minimize risks to the
patient and users.
Table 1 shows how FDA believes the
risks to health identified and described
in section V will be mitigated by the
proposed special controls. This
reclassification order and the identified
special controls, if finalized, would
provide sufficient detail regarding
FDA’s requirements to reasonably
assure safety and effectiveness of
ultrasound cyclodestructive devices.
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR ULTRASOUND CYCLODESTRUCTIVE DEVICES
Identified risk to health
Mitigation measures
Thermal Injury .....................................................
Non-clinical performance testing, Clinical performance data, Electrical Safety, Electromagnetic
compatibility, Software verification, validation, and hazard analysis, Labeling.
Non-clinical performance testing, Clinical performance data, Electrical safety, Electromagnetic
compatibility, Software verification, validation, and hazard analysis, Labeling.
Non-clinical performance testing, Clinical performance data, Software verification, validation,
and hazard analysis, Labeling.
Electrical safety, Labeling.
Electromagnetic compatibility, Labeling.
Biocompatibility evaluation, Sterility/reprocessing validation, Labeling.
Physical Injury .....................................................
Post-treatment Injury ..........................................
daltland on DSKBBV9HB2PROD with PROPOSALS
Electrical Shock ..................................................
Electromagnetic Interference ..............................
Ocular Irritation and Corneal Infections ..............
In addition, FDA is proposing to limit
these devices to prescription use under
§ 801.109. Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act and § 801.5, as long as the
conditions of § 801.109 are met
(referring to 21 U.S.C. 352(f)(1)). Under
21 CFR 807.81, the device would
continue to be subject to 510(k)
requirements.
This reclassification order and the
identified special controls, if finalized,
would provide sufficient detail
regarding FDA’s requirements to
reasonably assure safety and
effectiveness of ultrasound
cyclodestructive devices for the
treatment of refractory glaucoma. As
discussed below, the reclassification
will be codified in 21 CFR 886.5350.
FDA believes that adherence to the
VerDate Sep<11>2014
16:45 Sep 24, 2018
Jkt 244001
proposed special controls, in addition to
the general controls, is necessary to
provide a reasonable assurance of the
safety and effectiveness of the devices.
IX. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this
proposed order contains no new
collections of information. Therefore,
clearance by the Office of Management
and Budget (OMB) under the Paperwork
Reduction Act of 1995 (PRA) (44 U.S.C.
3501–3520) is not required. This
PO 00000
Frm 00006
Fmt 4702
Sfmt 4702
proposed order refers to previously
approved collections of information
found in FDA regulations. These
collections of information are subject to
review by OMB under the PRA. The
collections of information in 21 CFR
part 807, subpart E have been approved
under OMB control number 0910–0120
and the collections of information under
21 CFR part 801 have been approved
under OMB control number 0910–0485.
XI. Proposed Effective Date
FDA proposes that any final order
based on this proposal become effective
30 days after the date of its publication
in the Federal Register.
XII. References
The following references are on
display at the Dockets Management Staff
(see ADDRESSES), and are available for
viewing by interested persons between
E:\FR\FM\25SEP1.SGM
25SEP1
48408
Federal Register / Vol. 83, No. 186 / Tuesday, September 25, 2018 / Proposed Rules
daltland on DSKBBV9HB2PROD with PROPOSALS
9 a.m. and 4 p.m., Monday through
Friday; these are not available
electronically at https://
www.regulations.gov as they are
copyrighted or are available through the
website address. FDA has verified the
website addresses, as of the date this
document publishes in the Federal
Register, but websites are subject to
change over time.
1. PMA Database Reference for Sonocare.
Available at https://www.accessdata.
fda.gov/scripts/cdrh/cfdocs/cfpma/pma.
cfm?ID=319101.
2. Shields, M.B., ‘‘Transcleral Diode
Cyclophotocoagulation,’’ in Chen, T.C.
(ed.), ‘‘Surgical Techniques in
Ophthalmology: Glaucoma Surgery.’’
Saunders Elsevier, 2008.
3. Chen, T.C., L.R. Pasquale, D.S. Walton, et
al., ‘‘Diode Laser Transcleral
Cyclophotocoagulation.’’ International
Ophthalmology Clinics, 39(1):169–176,
1999.
4. Muratore, R.A., ‘‘History of the Sonocare
CST–100: The First FDA-Approved HIFU
Device.’’ Therapeutic Ultrasound,
809:508–512, 2006.
5. Melamed, S., M. Goldenfeld, D. Cotlear, et
al., ‘‘High-Intensity Focused Ultrasound
Treatment in Refractory Glaucoma
Patients: Results at 1 Year of Prospective
Clinical Study.’’ European Journal of
Ophthalmology, 25(6):483–489, 2015.
6. Aptel, F., T. Charrel, C. Labon, et al.,
‘‘Miniaturized High-Intensity Focused
Ultrasound Device in Patients with
Glaucoma: A Clinical Pilot Study.’’
Investigative Ophthalmology and Visual
Science, 52 (12):8747–8753, 2011.
7. Aptel, F. and C. Lafon, ‘‘Therapeutic
Applications of Ultrasound in
Ophthalmology.’’ International Journal
of Hyperthermia, 28(4):405–418, 2012.
8. Mastropasqua, R., V. Fasanella, A.
Mastropasqua, et al., ‘‘High-Intensity
Focused Ultrasound Circular
Cyclocoagulation in Glaucoma: A Step
Forward for Cyclodestruction?’’ Journal
of Ophthalmology, (2017), 1–14, 2017.
9. Burgess, SE, R.H. Silverman, D.J. Coleman,
et al., ‘‘Treatment of Glaucoma with High
Intensity Focused Ultrasound.’’
Ophthalmology, 93:831–838, 1986.
10. Coleman, D.J., F.L. Lizzi, J. Driller, et al.,
‘‘Therapeutic Ultrasound in the
Treatment of Glaucoma. II. Clinical
Applications.’’ Ophthalmology, 92:347–
353, 1985.
11. Valtot, F., J. Kopel, and J. Haut,
‘‘Treatment of Glaucoma with High
Intensity Focused Ultrasound.’’
International Ophthalmology,13:167–
170, 1989.
12. Silverman, R.H., B. Vogelsang, M.J.
Rondeau, et al., ‘‘Therapeutic Ultrasound
for the Treatment of Glaucoma.’’
VerDate Sep<11>2014
16:45 Sep 24, 2018
Jkt 244001
American Journal of Ophthalmology,
111:327–337, 1991.
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods
and services.
Therefore, under the Federal Food,
Drug, and Cosmetic Act (21 U.S.C. 321
et seq., as amended) and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 886 be amended as follows:
PART 886—OPHTHALMIC DEVICES
1. The authority citation for part 886
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 886.5350 to subpart F to read
as follows:
■
§ 886.5350
device.
Ultrasound cyclodestructive
(a) Identification. An ultrasound
cyclodestructive device is a prescription
device to reduce intraocular pressure by
producing a series of lesions in the
ciliary body and trabecular meshwork
induced by high intensity focused
ultrasound (HIFU) energy and that is
intended for treatment of refractory
glaucoma.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The clinical performance data
must demonstrate an adequate safety
profile and an appropriate reduction in
intraocular pressure in patients with
refractory glaucoma and capture any
adverse events observed during clinical
use.
(2) Non-clinical performance testing
of device features and characteristics
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
(i) Ultrasound field characteristics,
which must include the total acoustic
power radiated by the transducer(s), the
spatial distribution of the ultrasound
field (including compressional and
rarefactional pressure), and spatial-peak,
temporal-average intensity;
(ii) Thermal and physical safety
characterization of the device; and
(iii) Simulated use testing to validate
that the device performs as intended
under anticipated conditions of use,
PO 00000
Frm 00007
Fmt 4702
Sfmt 9990
including eye movements and
positioning error.
(3) Analysis/testing must demonstrate
electrical safety in the appropriate useenvironment.
(4) Analysis/testing must demonstrate
electromagnetic compatibility (EMC),
including wireless coexistence (if
applicable) in the appropriate useenvironment.
(5) Software verification, validation,
and hazard analysis must be performed
commensurate with the level of concern
of the device.
(6) The patient-contacting
components must be demonstrated to be
biocompatible.
(7) Performance data must
demonstrate sterility of all patientcontacting components labeled as
sterile. If the device contains reusable
eye-contact components, the validation
tests must demonstrate adequate
cleaning/reprocessing of these
components.
(8) Labeling must include:
(i) A detailed description of the
patient population for which the device
is indicated for use, as well as warnings,
and precautions regarding potential for
device malfunction and use-error
pertinent to use of the device.
(ii) A detailed summary of the clinical
testing, including study outcomes and
adverse events.
(iii) Information on how the device
operates and the typical course of
treatment.
(iv) Description of all main
components of the device including
HIFU generator, transducer(s), and
controls. The labeling must include
technical specification of the device
including, but not limited to, treatment
frequency, total acoustic power
delivered by transducer, treatment
duration, treatment zone, site targeting,
power requirements, weight, and
physical dimensions of the device.
(v) Where appropriate, validated
methods and instructions for
reprocessing of any reusable
components.
(vi) Safe-use conditions for electrical
safety and electromagnetic
compatibility.
Dated: September 18, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–20763 Filed 9–24–18; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\25SEP1.SGM
25SEP1
Agencies
[Federal Register Volume 83, Number 186 (Tuesday, September 25, 2018)]
[Proposed Rules]
[Pages 48403-48408]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2018-20763]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA-2018-N-3074]
Ophthalmic Devices; Reclassification of Ultrasound
Cyclodestructive Device
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is issuing this
proposed order to reclassify the ultrasound cyclodestructive device, a
postamendments class III device (regulated under product code LZR),
into class II (special controls), subject to premarket notification.
FDA is also identifying the proposed special controls that the Agency
believes are necessary to provide a reasonable assurance of safety and
effectiveness of the device. FDA is proposing this reclassification on
its own initiative based on new information. If finalized, this order
will reclassify these devices from class III to class II (special
controls) and reduce regulatory burdens as these types of devices will
no longer be required to submit a premarket approval application (PMA)
but can instead submit a less burdensome premarket notification
(510(k)) before marketing their device.
DATES: Submit either electronic or written comments on the proposed
order by November 26, 2018. Please see section XI for the proposed
effective date when the new requirements apply and for the proposed
effective date of a final order based on this proposed order.
ADDRESSES: You may submit comments as follows. Please note that late,
untimely filed comments will not be considered. Electronic comments
must be submitted on or before November 26, 2018. The https://www.regulations.gov electronic filing system will accept comments until
midnight Eastern Time at the end of November 26, 2018. Comments
received by mail/hand delivery/courier (for written/paper submissions)
will be considered timely if they are postmarked or the delivery
service acceptance receipt is on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the
[[Page 48404]]
instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand Delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2018-N-3074 for ``Ophthalmic Devices; Reclassification of
Ultrasound Cyclodestructive Device.'' Received comments, those filed in
a timely manner (see ADDRESSES), will be placed in the docket and,
except for those submitted as ``Confidential Submissions,'' publicly
viewable at https://www.regulations.gov or at the Dockets Management
Staff between 9 a.m. and 4 p.m., Monday through Friday.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.gpo.gov/fdsys/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Dockets Management Staff, 5630 Fishers Lane,
Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Hina Pinto, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1652, Silver Spring, MD 20993, 301-796-6351,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended,
establishes a comprehensive system for the regulation of medical
devices intended for human use. Section 513 of the FD&C Act (21 U.S.C.
360c) established three categories (classes) of devices, reflecting the
regulatory controls needed to provide reasonable assurance of their
safety and effectiveness. The three categories of devices are class I
(general controls), class II (special controls), and class III
(premarket approval).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices) are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until, the device
is reclassified into class I or II, or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and 21 CFR part 807.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or class II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA acting by order can
reclassify the device into class I or class II on its own initiative,
or in response to a petition from the manufacturer or importer of the
device. To change the classification of the device, the proposed new
class must have sufficient regulatory controls to provide reasonable
assurance of the safety and effectiveness of the device for its
intended use.
Reevaluation of the data previously before the Agency is an
appropriate basis for subsequent action where the reevaluation is made
in light of newly available regulatory authority (see Bell v. Goddard,
366 F.2d 177, 181 (7th Cir. 1966); Ethicon, Inc. v. FDA, 762 F. Supp.
382, 388-391 (D.D.C. 1991)), or in light of changes in ``medical
science'' (Upjohn v. Finch, 422 F.2d 944, 951 (6th Cir. 1970)). Whether
data before the Agency are old or new, the ``new information'' to
support reclassification under section 513(f)(3) of the FD&C Act must
be ``valid scientific evidence'', as defined in section 513(a)(3) of
the FD&C Act and 21 CFR 860.7(c)(2). (See, e.g., General Medical Co. v.
FDA, 770 F.2d 214 (D.C. Cir. 1985); Contact Lens Assoc. v. FDA, 766
F.2d 592 (D.C. Cir. 1985), cert. denied, 474 U.S. 1062 (1986)).
FDA relies upon ``valid scientific evidence'' in the classification
process to determine the level of regulation for devices. To be
considered in the reclassification process, the ``valid scientific
evidence'' upon which the Agency relies must be publicly available.
Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
PMA (see section 520(c) of the FD&C Act (21 U.S.C. 360j(c)). Section
510(m) of the FD&C Act provides that a class II device may be exempted
from the 510(k) premarket notification requirements, if the Agency
determines that premarket notification is not necessary to
[[Page 48405]]
reasonably assure the safety and effectiveness of the device.
II. Regulatory History of the Devices
On June 30, 1988, FDA approved the first and only ultrasound
cyclodestructive device through its PMA process under section 515 of
the FD&C Act (21 U.S.C. 360e). On August 10, 1988 (53 FR 30101), FDA
announced a PMA order for Sonocare Inc.'s Model CST-100 Therapeutic
Ultrasound System (Sonocare CST-100) (Ref. 1). As of the date of
issuance of this proposed order, the Sonocare CST-100 is the only PMA
approved by FDA for this device type.
III. Device Description
An ultrasound cyclodestructive device is a postamendments device
classified into class III under section 513(f)(1) of the FD&C Act. An
ultrasound cyclodestructive device is indicated for the treatment of
refractory glaucoma; it is intended for patients who are refractory to
or are poor candidates for laser or surgical treatment and fail to
achieve target intraocular pressures on maximally tolerated drug
therapy. The device is designed to reduce intraocular pressure by
producing a series of lesions in the ciliary body and trabecular
meshwork induced by high intensity focused ultrasound (HIFU) energy.
Different technologies, such as laser cyclodestruction to lower
intraocular pressure, have been studied since the 1970s (Refs. 2 and
3), which increases FDA's knowledge base for devices used to treat this
condition. As stated earlier in section II, FDA has approved only one
ultrasound cyclodestructive device through its PMA process under
section 515 of the FD&C Act (Ref. 4). More recently, reports in the
literature indicate that the HIFU technology has been modified and
currently studied in Europe for treatment of refractory glaucoma (Refs.
5 to 8). Based upon our review experience and consistent with the FD&C
Act and FDA's regulations, FDA believes that these devices should be
reclassified from class III into class II because there is sufficient
information to establish special controls that can provide reasonable
assurance of the device's safety and effectiveness.
Conventional refractory glaucoma treatment modalities include
implantable aqueous shunts and valves, trabeculectomy and other
incisional glaucoma surgeries, cyclocryotherapy, as well as laser
transcleral cyclophotocoagulation. FDA currently regulates all of the
devices indicated for these procedures in a refractory glaucoma
population as class II devices, subject to 510(k) requirements.
IV. Proposed Reclassification
On April 29, 2015, FDA published a document in the Federal Register
entitled ``Retrospective Review of Premarket Approval Application
Devices; Striking the Balance Between Premarket and Postmarket Data
Collection,'' in which FDA announced plans to consider reclassifying
ultrasound cyclodestructive devices identified with the LZR product
code from class III to class II (80 FR 23798) and requested comments.
FDA received no adverse comments regarding our proposed intent for LZR.
In accordance with section 513(f)(3) of the FD&C Act and 21 CFR
part 860, subpart C, FDA is proposing to reclassify this postamendments
class III device into class II. FDA believes that there is sufficient
information available to FDA through peer-reviewed literature and
knowledge of similar devices to establish special controls that would
effectively mitigate the risks to health identified in section V.
Absent the special controls identified in this proposed order, general
controls applicable to the device are insufficient to provide
reasonable assurance of the safety and effectiveness of the device.
FDA is proposing to create a separate classification regulation for
ultrasound cyclodestructive devices that will be reclassified from
class III to II. Under this proposed order, if finalized, the
ultrasound cyclodestructive devices will be identified as a
prescription device. As such, the prescription device must satisfy
prescription labeling requirements (see Sec. 801.109 (21 CFR 801.109),
Prescription devices). Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and Sec. 801.5
(21 CFR 801.5), as long as the conditions of Sec. 801.109 are met
(referring to 21 U.S.C. 352(f)(1)). In this proposed order, if
finalized, the Agency has identified the special controls under section
513(a)(1)(B) of the FD&C Act that, together with general controls, will
provide a reasonable assurance of the safety and effectiveness for
ultrasound cyclodestructive devices.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary for ultrasound
cyclodestructive devices to provide reasonable assurance of safety and
effectiveness. Therefore, the Agency does not intend to exempt these
proposed class II devices from 510(k) requirements. Persons who intend
to market this type of device must submit to FDA a 510(k) and receive
clearance prior to marketing the device.
This proposal, if finalized, will decrease regulatory burden and
will reduce private costs and expenditures required to comply with
Federal regulations. Specifically, regulated industry will no longer
have to submit a PMA but can instead submit a 510(k) to the Agency for
review prior to marketing their device. A 510(k) is a less burdensome
pathway to market a device, which typically results in a more timely
premarket review compared to a PMA and reduces the regulatory burden in
addition to providing more timely access of these types of devices to
patients.
V. Risks to Health
After considering the information available to FDA through the
review submission, peer-reviewed literature, and knowledge of other
technologies indicated to treat the same refractory glaucoma patient
population (such as aqueous shunt and cryotherapy), FDA determined that
the probable risks to health associated with the use of ultrasound
cyclodestructive devices for treatment of refractory glaucoma are as
follows:
Thermal Injury. Exposure of the ocular tissue to the HIFU
energy causes thermal damage of the tissue. The misdirection or
misalignment of the beam may cause temperature elevation in the non-
target ocular tissues and overall ocular tissue damage. Unsuitable
power and duration of the beam may also result in temperature
elevation, which may cause corneoscleral lesions including scleral
thinning, corneal ectasia and perforation, eyelid burns, corneal burns,
clouding of the cornea (haze) and lens (cataract formation), and
retinal and choroidal lesions.
Physical Injury. Exposure of the ocular tissue to the HIFU
energy can cause physical damage to the ocular tissue due to cavitation
or other mechanical effects. These injuries could be caused by the
suboptimal selection of the treatment parameters, misalignment/
displacement of the probe during the treatment, device malfunction, or
other factors affecting stability of treatment. For example,
insonification of the zonular fibers may cause elongation or rupture of
ligaments, which can lead to a displacement of the lens.
[[Page 48406]]
Post-treatment injury. Post-treatment injury from use of
the device may include intraocular inflammation (e.g., iritis,
uveitis), increased intraocular pressure in the immediate post-
treatment period, ciliary body hemorrhage, persistent or transient low
pressure, decreased visual acuity, worsening glaucoma, phthisis, pain/
discomfort, corneal edema, hyphema, retinal and choroidal
complications, etc.
Electrical shock. While in operation, the device may
discharge electricity that could shock the user. Electrical shock can
be caused by use error or device malfunction.
Electromagnetic interference. While in operation,
electromagnetic interferences from other devices operated in the same
environment may cause the device to malfunction, which could result in
patient's injury. In addition, the device may interfere with other
electrically powered devices, causing them to malfunction.
Ocular irritation and corneal infections. Inadequate
biocompatibility of the eye contact components of the device can lead
to irritation of the ocular tissue. Inappropriately sterilized or
reprocessed eye contact components of the device can lead to
inflammation and corneal infections.
VI. Summary of Reasons for Reclassification
FDA believes that the ultrasound cyclodestructive devices for
treatment of refractory glaucoma should be reclassified from class III
to class II in light of available information about the effectiveness
of these devices. There is sufficient information to establish special
controls for ultrasound cyclodestructive devices, in addition to
general controls, which can provide reasonable assurance of safety and
effectiveness of the device, as general controls themselves are
insufficient to provide reasonable assurance of its safety and
effectiveness. FDA believes that the risks to health associated with
ultrasound cyclodestructive devices for treatment of refractory
glaucoma can be mitigated with special controls and that these
mitigations will provide a reasonable assurance of its safety and
effectiveness.
Based on a reconsideration of the available information and data,
FDA believes that there is valid scientific evidence of effectiveness
for ultrasound cyclodestructive devices to reduce intraocular pressure
intended for treatment of refractory glaucoma using ultrasound.
VII. Summary of Data Upon Which the Reclassification Is Based
FDA believes that the identified special controls, in addition to
general controls, are necessary to provide reasonable assurance of
safety and effectiveness of these devices. Taking into account the
probable health benefits of the use of the device and the nature and
known incidence of the risks of the device, FDA, on its own initiative,
is proposing to reclassify this postamendments class III device into
class II. FDA has considered and analyzed the following information: An
inclusive search of the Agency's Manufacturer and User Facility Device
Experience (MAUDE) database, which shows no adverse events for
ultrasound cyclodestructive device type; no recalls have been received
for this device type; other technologies indicated to treat the same
refractory glaucoma patient population, such as aqueous shunt and
cryotherapy, and currently regulated as class II devices to compare the
probable risks (i.e., between the rate and severity of the adverse
events associated with these class II technologies and ultrasound
cyclodestructive procedures); and peer-reviewed publications (Refs. 5
to 12) to identify probable device risks (e.g., the types and rates of
adverse events) and mitigation strategies.
VIII. Proposed Special Controls
FDA believes that the following special controls, together with
general controls, are necessary and sufficient to mitigate the risks to
health described in section V and provide a reasonable assurance of
safety and effectiveness for ultrasound cyclodestructive devices.
Non-clinical performance testing of device features and
characteristics will demonstrate:
[cir] The ability of the device to deposit controllable HIFU energy
to the target area to evoke the required level of thermal lesion.
[cir] The design and geometry of the HIFU transducer and the output
characteristics of the HIFU generator, including operating frequency
and power, produce a small focal zone and a steep transition of energy
deposition between the focal zone and the untreated areas. In addition,
the total acoustic power radiated by the transducer(s), spatial
distribution of the ultrasound field (including compressional and
rarefactional pressure), and spatial peak, temporal-average intensity
will be evaluated. This may be accomplished by demonstrating compliance
with the standard International Electrotechnical Commission (IEC)
Technical Specification (TS) 62556: Ultrasonics--Field
characterization--Specification and measurement of field parameters for
high intensity therapeutic ultrasound (HITU) transducers and systems.
Thermal and physical (due to potential cavitation of gas bubbles)
safety analyses will also be evaluated.
[cir] The appropriate alignment and focusing of the ultrasound beam
to the target tissue to minimize unintended damage to adjacent ocular
tissues.
[cir] The function of all safety features built into the device,
including the energy monitoring system.
Clinical performance data will validate device performance
and characterize ocular tissue thermal injuries, physical injury, and
postoperative adverse events by establishing the treatment parameters
for which the device is safe and effective.
Electrical safety testing will minimize the risk of
electrical shock, thermal or physical injury to the patient and
healthcare provider. This may be accomplished by demonstrating
compliance with FDA-recognized consensus standard American National
Standards Institute (ANSI)/Association for the Advancement of Medical
Instrumentation (AAMI) 60601-1: Medical electrical equipment, Part 1:
General requirements for basic safety and essential performance.
Electromagnetic compatibility testing ensures that
electromagnetic interferences do not cause device malfunction. It can
also provide assurance that electromagnetic interferences generated by
the device do not affect the other devices operated in the same
environment. This may be accomplished by demonstrating compliance with
FDA-recognized consensus standard IEC 60601-1-2: Medical electrical
equipment, Part 1-2: General requirements for safety. If the device
incorporates radiofrequency (RF) wireless technology to perform medical
functions and/or communicates medical data, testing will mitigate the
risks associated with interference or degradation when using RF
wireless technology. This may be accomplished by demonstrating
compliance with FDA-recognized consensus standard AAMI TIR69: Risk
Management of Radio-frequency Wireless Coexistence for Medical Devices
and Systems (risk assessment) and ANSI/Institute of Electrical and
Electronics Engineers (IEEE) C63.27: American National Standard for
Evaluation of Wireless Coexistence (coexistence testing).
Software verification, validation, and hazard analysis is
necessary to
[[Page 48407]]
mitigate the risks of thermal and physical injury and ensures that
software performs as intended and potential software malfunctions do
not impact the safety or effectiveness of the device. If the device
incorporates internet network connectivity, testing will demonstrate
that cybersecurity concerns are mitigated (e.g., data integrity,
unauthorized access, etc.).
Biocompatibility evaluation can help mitigate the risk of
ocular irritation and corneal infection by ensuring that the patient-
contacting components of the device are safe for contact with skin and
ocular tissue.
Sterilization validation, for devices provided sterile,
and/or cleaning validation, for devices or components that are
reusable, help mitigate the risk of inflammation and corneal infections
(e.g., keratitis).
The labeling will also include necessary information to
ensure safe and effective use of the ultrasound cyclodestructive device
and minimize probability of the ocular treatment-related adverse
events. Labeling needs to include sufficient information that will help
the patient and healthcare provider make an informed decision regarding
treatment-related adverse effects of the ultrasound cyclodestructive
treatment. For example, the labeling needs to include information
regarding the most common reported treatment-related injuries, which
may include intraocular inflammation (e.g., iritis, uveitis) and
increased intraocular pressure in the immediate post-treatment period.
Adverse event information related to ciliary body hemorrhage,
persistent low pressure, decreased visual acuity, worsening glaucoma,
phthisis, pain/discomfort, transient low pressure, corneal edema,
hyphema, retinal complications (such as cystoid macula edema), and
choroidal effusion or detachment need to be discussed. The labeling
will mitigate the risk associated with the intraoperative events, such
as pain/discomfort, and postoperative adverse events by providing
appropriate clinical information along with mitigation strategies
(e.g., retrobulbar or peribulbar anesthesia). Specifically, device
labeling must include:
[cir] Appropriate warnings and precautions to ensure safe and
effective use of the device and minimize potential device malfunctions
and user errors.
[cir] A summary of the clinical evaluation pertinent to use of the
device, including study outcomes and adverse events.
[cir] Information regarding procedure parameters, proper
positioning of the HIFU transducer and its coupling with the eye, and
typical course of treatment to ensure the user can safely operate the
device.
[cir] Validated reprocessing instructions to ensure the safe use of
reusable device components.
[cir] Safety information regarding electrical safety and
electromagnetic compatibility to minimize risks to the patient and
users.
Table 1 shows how FDA believes the risks to health identified and
described in section V will be mitigated by the proposed special
controls. This reclassification order and the identified special
controls, if finalized, would provide sufficient detail regarding FDA's
requirements to reasonably assure safety and effectiveness of
ultrasound cyclodestructive devices.
Table 1--Risks to Health and Mitigation Measures for Ultrasound
Cyclodestructive Devices
------------------------------------------------------------------------
Identified risk to health Mitigation measures
------------------------------------------------------------------------
Thermal Injury............... Non-clinical performance testing,
Clinical performance data, Electrical
Safety, Electromagnetic compatibility,
Software verification, validation, and
hazard analysis, Labeling.
Physical Injury.............. Non-clinical performance testing,
Clinical performance data, Electrical
safety, Electromagnetic compatibility,
Software verification, validation, and
hazard analysis, Labeling.
Post-treatment Injury........ Non-clinical performance testing,
Clinical performance data, Software
verification, validation, and hazard
analysis, Labeling.
Electrical Shock............. Electrical safety, Labeling.
Electromagnetic Interference. Electromagnetic compatibility, Labeling.
Ocular Irritation and Corneal Biocompatibility evaluation, Sterility/
Infections. reprocessing validation, Labeling.
------------------------------------------------------------------------
In addition, FDA is proposing to limit these devices to
prescription use under Sec. 801.109. Prescription devices are exempt
from the requirement for adequate directions for use for the layperson
under section 502(f)(1) of the FD&C Act and Sec. 801.5, as long as the
conditions of Sec. 801.109 are met (referring to 21 U.S.C. 352(f)(1)).
Under 21 CFR 807.81, the device would continue to be subject to 510(k)
requirements.
This reclassification order and the identified special controls, if
finalized, would provide sufficient detail regarding FDA's requirements
to reasonably assure safety and effectiveness of ultrasound
cyclodestructive devices for the treatment of refractory glaucoma. As
discussed below, the reclassification will be codified in 21 CFR
886.5350. FDA believes that adherence to the proposed special controls,
in addition to the general controls, is necessary to provide a
reasonable assurance of the safety and effectiveness of the devices.
IX. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this proposed order contains no new
collections of information. Therefore, clearance by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(PRA) (44 U.S.C. 3501-3520) is not required. This proposed order refers
to previously approved collections of information found in FDA
regulations. These collections of information are subject to review by
OMB under the PRA. The collections of information in 21 CFR part 807,
subpart E have been approved under OMB control number 0910-0120 and the
collections of information under 21 CFR part 801 have been approved
under OMB control number 0910-0485.
XI. Proposed Effective Date
FDA proposes that any final order based on this proposal become
effective 30 days after the date of its publication in the Federal
Register.
XII. References
The following references are on display at the Dockets Management
Staff (see ADDRESSES), and are available for viewing by interested
persons between
[[Page 48408]]
9 a.m. and 4 p.m., Monday through Friday; these are not available
electronically at https://www.regulations.gov as they are copyrighted
or are available through the website address. FDA has verified the
website addresses, as of the date this document publishes in the
Federal Register, but websites are subject to change over time.
1. PMA Database Reference for Sonocare. Available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma.cfm?ID=319101.
2. Shields, M.B., ``Transcleral Diode Cyclophotocoagulation,'' in
Chen, T.C. (ed.), ``Surgical Techniques in Ophthalmology: Glaucoma
Surgery.'' Saunders Elsevier, 2008.
3. Chen, T.C., L.R. Pasquale, D.S. Walton, et al., ``Diode Laser
Transcleral Cyclophotocoagulation.'' International Ophthalmology
Clinics, 39(1):169-176, 1999.
4. Muratore, R.A., ``History of the Sonocare CST-100: The First FDA-
Approved HIFU Device.'' Therapeutic Ultrasound, 809:508-512, 2006.
5. Melamed, S., M. Goldenfeld, D. Cotlear, et al., ``High-Intensity
Focused Ultrasound Treatment in Refractory Glaucoma Patients:
Results at 1 Year of Prospective Clinical Study.'' European Journal
of Ophthalmology, 25(6):483-489, 2015.
6. Aptel, F., T. Charrel, C. Labon, et al., ``Miniaturized High-
Intensity Focused Ultrasound Device in Patients with Glaucoma: A
Clinical Pilot Study.'' Investigative Ophthalmology and Visual
Science, 52 (12):8747-8753, 2011.
7. Aptel, F. and C. Lafon, ``Therapeutic Applications of Ultrasound
in Ophthalmology.'' International Journal of Hyperthermia,
28(4):405-418, 2012.
8. Mastropasqua, R., V. Fasanella, A. Mastropasqua, et al., ``High-
Intensity Focused Ultrasound Circular Cyclocoagulation in Glaucoma:
A Step Forward for Cyclodestruction?'' Journal of Ophthalmology,
(2017), 1-14, 2017.
9. Burgess, SE, R.H. Silverman, D.J. Coleman, et al., ``Treatment of
Glaucoma with High Intensity Focused Ultrasound.'' Ophthalmology,
93:831-838, 1986.
10. Coleman, D.J., F.L. Lizzi, J. Driller, et al., ``Therapeutic
Ultrasound in the Treatment of Glaucoma. II. Clinical
Applications.'' Ophthalmology, 92:347-353, 1985.
11. Valtot, F., J. Kopel, and J. Haut, ``Treatment of Glaucoma with
High Intensity Focused Ultrasound.'' International
Ophthalmology,13:167-170, 1989.
12. Silverman, R.H., B. Vogelsang, M.J. Rondeau, et al.,
``Therapeutic Ultrasound for the Treatment of Glaucoma.'' American
Journal of Ophthalmology, 111:327-337, 1991.
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods and services.
Therefore, under the Federal Food, Drug, and Cosmetic Act (21
U.S.C. 321 et seq., as amended) and under authority delegated to the
Commissioner of Food and Drugs, it is proposed that 21 CFR part 886 be
amended as follows:
PART 886--OPHTHALMIC DEVICES
0
1. The authority citation for part 886 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 886.5350 to subpart F to read as follows:
Sec. 886.5350 Ultrasound cyclodestructive device.
(a) Identification. An ultrasound cyclodestructive device is a
prescription device to reduce intraocular pressure by producing a
series of lesions in the ciliary body and trabecular meshwork induced
by high intensity focused ultrasound (HIFU) energy and that is intended
for treatment of refractory glaucoma.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The clinical performance data must demonstrate an adequate
safety profile and an appropriate reduction in intraocular pressure in
patients with refractory glaucoma and capture any adverse events
observed during clinical use.
(2) Non-clinical performance testing of device features and
characteristics must demonstrate that the device performs as intended
under anticipated conditions of use. The following performance
characteristics must be tested:
(i) Ultrasound field characteristics, which must include the total
acoustic power radiated by the transducer(s), the spatial distribution
of the ultrasound field (including compressional and rarefactional
pressure), and spatial-peak, temporal-average intensity;
(ii) Thermal and physical safety characterization of the device;
and
(iii) Simulated use testing to validate that the device performs as
intended under anticipated conditions of use, including eye movements
and positioning error.
(3) Analysis/testing must demonstrate electrical safety in the
appropriate use-environment.
(4) Analysis/testing must demonstrate electromagnetic compatibility
(EMC), including wireless coexistence (if applicable) in the
appropriate use-environment.
(5) Software verification, validation, and hazard analysis must be
performed commensurate with the level of concern of the device.
(6) The patient-contacting components must be demonstrated to be
biocompatible.
(7) Performance data must demonstrate sterility of all patient-
contacting components labeled as sterile. If the device contains
reusable eye-contact components, the validation tests must demonstrate
adequate cleaning/reprocessing of these components.
(8) Labeling must include:
(i) A detailed description of the patient population for which the
device is indicated for use, as well as warnings, and precautions
regarding potential for device malfunction and use-error pertinent to
use of the device.
(ii) A detailed summary of the clinical testing, including study
outcomes and adverse events.
(iii) Information on how the device operates and the typical course
of treatment.
(iv) Description of all main components of the device including
HIFU generator, transducer(s), and controls. The labeling must include
technical specification of the device including, but not limited to,
treatment frequency, total acoustic power delivered by transducer,
treatment duration, treatment zone, site targeting, power requirements,
weight, and physical dimensions of the device.
(v) Where appropriate, validated methods and instructions for
reprocessing of any reusable components.
(vi) Safe-use conditions for electrical safety and electromagnetic
compatibility.
Dated: September 18, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-20763 Filed 9-24-18; 8:45 am]
BILLING CODE 4164-01-P