Radiology Devices; Reclassification of Medical Image Analyzers, 25598-25604 [2018-11880]
Download as PDF
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
(i) Terminating Action for Certain
Requirements of AD 2014–25–52
For airplanes with an AOA configuration
as identified in figure 1 to paragraph (i) of
(j) Terminating Action for Certain
Requirements of AD 2016–25–30
Accomplishment of the actions required by
paragraph (h) of this AD terminates the
requirements of paragraph (g) of AD 2016–
25–30 for that airplane.
(k) Parts Installation Prohibition
Installation of any software or hardware of
a version earlier than the one listed in table
1 to paragraphs (h) and (k) of this AD is
prohibited, as required by paragraphs (k)(1)
and (k)(2) of this AD, as applicable.
(1) For Group 1 airplanes: After
modification of an airplane as required by
paragraph (h) of this AD.
(2) For Group 2 airplanes: As of the
effective date of this AD.
sradovich on DSK3GMQ082PROD with PROPOSALS
(l) Other FAA AD Provisions
The following provisions also apply to this
AD:
(1) Alternative Methods of Compliance
(AMOCs): The Manager, International
Section, Transport Standards Branch, FAA,
has the authority to approve AMOCs for this
AD, if requested using the procedures found
in 14 CFR 39.19. In accordance with 14 CFR
39.19, send your request to your principal
inspector or local Flight Standards District
Office, as appropriate. If sending information
directly to the manager of the International
Branch, send it to the attention of the person
identified in paragraph (m)(2) of this AD.
Information may be emailed to: 9-ANM-116AMOC-REQUESTS@faa.gov. Before using
any approved AMOC, notify your appropriate
principal inspector, or lacking a principal
inspector, the manager of the local flight
standards district office/certificate holding
district office.
(2) Contacting the Manufacturer: For any
requirement in this AD to obtain corrective
actions from a manufacturer, the action must
be accomplished using a method approved
by the Manager, International Section,
Transport Standards Branch, FAA; or the
European Aviation Safety Agency (EASA); or
EASA Design Organization Approval (DOA).
If approved by the DOA, the approval must
include the DOA-authorized signature.
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
this AD, or as identified in paragraph (m)(2)
of AD 2016–12–15, Amendment 39–18564
(81 FR 40160, June 21, 2016) (‘‘AD 2016–12–
15’’), as applicable: Accomplishing the
upgrade required by paragraph (h) of this AD
terminates the requirements of paragraph (g)
of AD 2014–25–52, and the airplane flight
manual (AFM) procedure required by
paragraph (g) of AD 2014–25–52 may be
removed from the AFM.
(3) Required for Compliance (RC): If any
service information contains procedures or
tests that are identified as RC, those
procedures and tests must be done to comply
with this AD; any procedures or tests that are
not identified as RC are recommended. Those
procedures and tests that are not identified
as RC may be deviated from using accepted
methods in accordance with the operator’s
maintenance or inspection program without
obtaining approval of an AMOC, provided
the procedures and tests identified as RC can
be done and the airplane can be put back in
an airworthy condition. Any substitutions or
changes to procedures or tests identified as
RC require approval of an AMOC.
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
(m) Related Information
(1) Refer to Mandatory Continuing
Airworthiness Information (MCAI) EASA
Airworthiness Directive 2017–0246R1, dated
April 6, 2018, for related information. This
MCAI may be found in the AD docket on the
internet at https://www.regulations.gov by
searching for and locating Docket No. FAA–
2018–0498.
(2) For more information about this AD,
contact Vladimir Ulyanov, Aerospace
Engineer, International Section, Transport
Standards Branch, FAA, 2200 South 216th
St., Des Moines, WA 98198; telephone and
fax 206–231–3229.
(3) For service information identified in
this AD, contact Airbus SAS, Airworthiness
Office—EAL, 1 Rond Point Maurice Bellonte,
31707 Blagnac Cedex, France; telephone +33
5 61 93 36 96; fax +33 5 61 93 45 80; email
airworthiness.A330-A340@airbus.com;
internet https://www.airbus.com. You may
view this service information at the FAA,
Transport Standards Branch, 2200 South
216th St., Des Moines, WA. For information
on the availability of this material at the
FAA, call 206–231–3195.
Issued in Des Moines, Washington, on May
23, 2018.
James Cashdollar,
Acting Director, System Oversight Division,
Aircraft Certification Service.
[FR Doc. 2018–11700 Filed 6–1–18; 8:45 am]
BILLING CODE 4910–13–P
PO 00000
Frm 00016
Fmt 4702
Sfmt 4702
Food and Drug Administration
21 CFR Part 892
[Docket No. FDA–2018–N–1553]
Radiology Devices; Reclassification of
Medical Image Analyzers
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed order.
The Food and Drug
Administration (FDA or the Agency) is
issuing this proposed order to reclassify
medical image analyzers applied to
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection as postamendments
class III (premarket approval) devices
(regulated under product code MYN),
into class II (special controls), subject to
premarket notification. FDA is also
identifying the proposed special
controls that the Agency believes are
necessary to provide a reasonable
assurance of safety and effectiveness of
the device. These devices are intended
to direct the clinician’s attention to
portions of an image that may reveal
abnormalities during interpretation of
patient’s radiology images by the
clinician. If finalized, this order will
reclassify these types of devices from
class III to class II and reduce regulatory
burdens on industry as these types of
devices will no longer be required to
submit a premarket approval
application (PMA) but can instead
submit a less burdensome premarket
notification (510(k)) before marketing
their device.
SUMMARY:
E:\FR\FM\04JNP1.SGM
04JNP1
EP04JN18.001
25598
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
Submit either electronic or
written comments on the proposed
order by August 3, 2018. Please see
section X of this document for the
proposed effective date when the new
requirements apply and for the
proposed effective date of a final order
based on this proposed order.
ADDRESSES: You may submit comments
as follows: Please note that late,
untimely filed comments will not be
considered. Electronic comments must
be submitted on or before August 3,
2018. The https://www.regulations.gov
electronic filing system will accept
comments until midnight Eastern Time
at the end of August 3, 2018. Comments
received by mail/hand delivery/courier
(for written/paper submissions) will be
considered timely if they are
postmarked or the delivery service
acceptance receipt is on or before that
date.
DATES:
sradovich on DSK3GMQ082PROD with PROPOSALS
Electronic Submissions
Submit electronic comments in the
following way:
• Federal Rulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2018–N–1553 for ‘‘Radiology Devices;
Reclassification of Medical Image
Analyzers.’’ Received comments will be
placed in the docket and, except for
those submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Dockets Management Staff between 9
a.m. and 4 p.m., Monday through
Friday.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://www.gpo.gov/
fdsys/pkg/FR-2015-09-18/pdf/201523389.pdf.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Robert Ochs, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4312, Silver Spring,
MD 20993–0002, 301–796–6661,
Robert.Ochs@fda.hhs.gov.
PO 00000
Frm 00017
Fmt 4702
Sfmt 4702
25599
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act), as amended,
establishes a comprehensive system for
the regulation of medical devices
intended for human use. Section 513 of
the FD&C Act (21 U.S.C. 360c)
established three categories (classes) of
devices, reflecting the regulatory
controls needed to provide reasonable
assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices), are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II, or FDA issues an order finding
the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and 21 CFR
part 807.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or II
under section 513(f)(3) of the FD&C Act.
Section 513(f)(3) of the FD&C Act
provides that FDA acting by order can
reclassify the device into class I or II on
its own initiative, or in response to a
petition from the manufacturer or
importer of the device. To change the
classification of the device, the
proposed new class must have sufficient
regulatory controls to provide a
reasonable assurance of the safety and
effectiveness of the device for its
intended use.
Reevaluation of the data previously
before the Agency is an appropriate
basis for subsequent action where the
reevaluation is made in light of newly
available regulatory authority (see Bell
v. Goddard, 366 F.2d 177, 181 (7th Cir.
1966); Ethicon, Inc. v. FDA, 762 F.
Supp. 382, 388–391 (D.D.C. 1991)), or in
light of changes in ‘‘medical science’’
(Upjohn v. Finch, 422 F.2d 944, 951 (6th
Cir. 1970)). Whether data before the
Agency are old or new, the ‘‘new
information’’ to support reclassification
E:\FR\FM\04JNP1.SGM
04JNP1
25600
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
sradovich on DSK3GMQ082PROD with PROPOSALS
under 513(f)(3) must be ‘‘valid scientific
evidence’’, as defined in section
513(a)(3) of the FD&C Act and 21 CFR
860.7(c)(2). (See, e.g., General Medical
Co. v. FDA, 770 F.2d 214 (D.C. Cir.
1985); Contact Lens Mfrs. Assoc. v. FDA,
766 F.2d 592 (DC Cir.1985), cert.
denied, 474 U.S. 1062 (1986)). FDA
relies upon ‘‘valid scientific evidence’’
in the classification process to
determine the level of regulation for
devices. To be considered in the
reclassification process, the ‘‘valid
scientific evidence’’ upon which the
Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA (see
section 520(c) of the FD&C Act (21
U.S.C. 360j(c)).
In accordance with section 513(f)(3) of
the FD&C Act, the Agency is proposing
to reclassify medical image analyzers
applied to mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection from class III into class
II on the basis that there is sufficient
information to establish special
controls, in addition to general controls,
to provide reasonable assurance of the
safety and effectiveness of the device.
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the 510(k) premarket
notification requirements, if the Agency
determines that premarket notification
is not necessary to reasonably assure the
safety and effectiveness of the device.
II. Regulatory History of the Devices
This proposed order covers medical
image analyzers including computerassisted/aided detection (CADe) devices
for mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection that are assigned
product code MYN. These
postamendments devices are currently
regulated as class III devices under
section 513(f)(1) of the FD&C Act. FDA
has experience reviewing and analyzing
data and information for medical image
analyzers since premarket approval of
the first device for these uses in 1998.
On June 26, 1998, the Center for Devices
and Radiological Health (CDRH)
approved the first CADe device
included in this reclassification order.
In the December 30, 1998, Federal
Register notice (63 FR 71930), FDA
announced a PMA approval order for R2
Technology, Inc. M 1000 Image Checker
and the availability of the summary of
safety and effectiveness data for the
device. Since 1998, 11 devices have
received premarket approval for the
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
analysis of several modalities, including
mammography, ultrasound, as well as
chest and dental radiographs. Based
upon our review experience and
consistent with the FD&C Act and
FDA’s regulations, FDA believes that
these devices should be reclassified
from class III into class II because there
is sufficient information to establish
special controls that can provide
reasonable assurance of the device’s
safety and effectiveness.
This proposed order does not apply to
medical image analyzers/CADe devices
currently classified under § 892.2050
(21 CFR 892.2050), Picture archiving
and communication system. FDA has
regulated other CADe devices intended
to aid lung nodule and colon polyp
detection from computed tomography
images as class II devices under
§ 892.2050, Picture archiving and
communication system and assigned the
following product codes:
• NWE (Colon Computed
Tomography System, Computer-Aided
Detection);
• OEB (Lung Computed Tomography
System, Computer-Aided Detection);
• OMJ (Chest X-Ray Computer Aided
Detection).
There have been no recalls for class II
CADe devices. As of the date of this
proposal, FDA has received three recalls
for class III devices and one Medical
Device Report (MDR), however, in the
past 10 years only one recall for the
class III devices has been received due
to distribution of the CADe device
without PMA approval. None of these
recalls were classified as a Class I recall.
There were also no MDRs related to
either the class III medical image
analyzers or class II CADe devices in the
past 10 years. This evidence suggests
that the safety profiles for existing class
III CADe devices are similar to the class
II CADe, and consequently that our
regulatory controls applied should be
similar.
III. Device Description
This proposed order applies to
medical image analyzers including
CADe devices for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection that
are currently regulated as class III
devices as postamendment devices.
These devices are intended to identify,
mark, highlight, or in any other manner
direct the clinicians’ attention to
portions of a radiology image that may
reveal abnormalities during
interpretation of patient radiology
images by the clinicians. These devices
incorporate pattern recognition and data
analysis capabilities and operate on
PO 00000
Frm 00018
Fmt 4702
Sfmt 4702
previously acquired radiology images,
including mammography, radiograph,
and ultrasound. These devices are not
intended to replace the review by a
qualified radiologist or to be used for
triage. Furthermore, these devices are
not intended to recommend diagnosis of
any diseases.
IV. Proposed Reclassification
The Radiological Devices Panel (the
Panel) convened on March 4–5, 2008
(Ref. 1) and discussed issues relating to
how medical image analyzers including
CADe devices are used in clinical
decisionmaking, how the performance
of the devices should be evaluated, and
the information needed to determine
whether the device provides a
reasonable assurance of its safety and
effectiveness. Additional discussions
were held regarding medical image
analyzers for mammography and
radiograph applications. Following the
2008 Panel Meeting, FDA convened a
second meeting of the Panel on
November 18, 2009. The 2009 Panel
Meeting was asked to discuss two
proposed draft guidances for the
evaluation of medical image analyzers
and the Agency’s regulatory strategy for
these devices (Ref. 2). Subsequently, the
two draft guidance documents were
finalized by FDA and were made public
on July 3, 2012 (Refs. 3 and 4). The
guidance document entitled ‘‘Clinical
Performance Assessment:
Considerations for Computer-Assisted
Detection Devices Applied to Radiology
Images and Radiology Device Data—
Premarket Approval (PMA) and
Premarket Notification [510(k)]
Submissions’’ provides guidance
regarding clinical performance
assessment studies for CADe applied to
radiology images and radiology device
data. The guidance document entitled
‘‘Computer-Assisted Detection Devices
Applied to Radiology Images and
Radiology Device Data—Premarket
Notification [510(k)] Submissions’’
provides guidance regarding premarket
notification (510(k)) submissions for
CADe applied to radiology images and
radiology device data. These guidance
documents describe clinical and nonclinical methods to evaluate the safety
and effectiveness of CADe devices,
including medical image analyzers
covered by this proposed order. In
addition to the two guidance
documents, the Panel’s discussion
regarding the benefits and risks of
medical image analyzers that were
discussed at the 2008 and 2009 Panel
meetings have been taken into
consideration by the Agency when
developing the proposed special
E:\FR\FM\04JNP1.SGM
04JNP1
sradovich on DSK3GMQ082PROD with PROPOSALS
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
controls provided in this proposed order
below.
Since publication of these guidance
documents, the Agency has gained
considerable experience in reviewing
medical image analyzers using the
methods described in the
aforementioned guidance documents.
Further, as part CDRH’s 2014–2015
strategic priority ‘‘Strike the Right
Balance Between Premarket and
Postmarket Data Collection,’’ a
retrospective review of class III devices
subject to a PMA was completed to
determine whether or not, based on our
current understanding of the
technology, reclassification may be
appropriate. During this retrospective
review, FDA determined that sufficient
information exists such that the risks of
false positive and false negative results,
misuse, and device failure can be
mitigated, to establish special controls
that, together with general controls, can
provide a reasonable assurance of the
safety and effectiveness of medical
image analyzers and therefore proposes
these devices be reclassified from class
III to class II. On April 29, 2015, FDA
published a notice in the Federal
Register entitled ‘‘Retrospective Review
of Premarket Approval Application
Devices; Striking the Balance Between
Premarket and Postmarket Data
Collection’’ in which FDA announced
plans to consider reclassifying medical
image analyzers identified with the
MYN product code from class III to class
II (80 FR 23798). No adverse comments
were received regarding our proposed
intent for MYN.
In accordance with section 513(f)(3) of
the FD&C Act and 21 CFR part 860,
subpart C, FDA is proposing to
reclassify postamendments medical
image analyzers, including CADe
devices for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection, from
class III into class II. FDA believes that
there is sufficient information to
establish special controls, in addition to
general controls, that would effectively
mitigate the risks to health identified in
section V and provide reasonable
assurance of the safety and effectiveness
of these devices. Absent the special
controls identified in this proposed
order, general controls applicable to the
device are insufficient to provide
reasonable assurance of the safety and
effectiveness of the device.
FDA is proposing to create a separate
classification regulation for medical
image analyzer devices that will be
reclassified from class III to II. Under
this proposed order, if finalized, the
medical image analyzer devices will be
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
identified as a prescription device. As
such, the prescription device must
satisfy prescription labeling
requirements (see § 801.109 (21 CFR
801.109), Prescription devices).
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352) and § 801.5 (21 CFR 801.5), as long
as the conditions of § 801.109 are met.
In this proposed order, if finalized, the
Agency has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, will provide a reasonable
assurance of the safety and effectiveness
for medical image analyzer devices.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
is necessary to provide reasonable
assurance of safety and effectiveness
and, therefore, does not intend to
exempt these proposed class II devices
from the premarket notification
requirements. Persons who intend to
market this type of device must submit
to FDA a 510(k) and receive clearance
prior to marketing the device.
This proposal, if finalized, will
decrease regulatory burden on the
medical device industry and will reduce
private costs and expenditures required
to comply with Federal Regulations.
Specifically, regulated industry will no
longer have to submit a PMA but can
instead submit a 510(k) to the Agency
for review prior to marketing their
device. A 510(k) is a less-burdensome
pathway to market a device which
typically results in a more timely
premarket review compared to a PMA
and reduces the regulatory burden on
industry in addition to providing more
timely access of these types of devices
to patients.
V. Risks to Health
From the Panel discussions on March
4–5, 2008, and November 18, 2009,
along with the peer-reviewed literature
(Refs. 5–8) and FDA’s experiences over
the years in reviewing submissions for
these devices and similar devices, FDA
determined the probable risks to health
associated with medical image analyzers
including CADe devices for
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection are as follows: (1) False
PO 00000
Frm 00019
Fmt 4702
Sfmt 4702
25601
positive results may result in
complications, such as incorrect
management of the patient with possible
adverse effects, and unnecessary
additional radiology imaging and/or
invasive procedures, such as biopsy; (2)
false negative results could result in
complications, including incorrect
diagnosis and delay in disease
management; (3) the device could be
misused to analyze images from an
unintended patient population or on
images acquired with incompatible
imaging hardware or incompatible
image acquisition parameters, resulting
in possibly lower device performance;
(4) the device could be misused by not
following the appropriate reading
protocol, which may lead to lower
sensitivity; and (5) device failure could
result in the absence or delay of device
output, or incorrect device output,
which could likewise lead to inaccurate
patient assessment.
VI. Summary of the Reasons for
Reclassification
After considering the information
above, FDA has determined that all
class III medical image analyzers
currently approved by FDA should be
reclassified into class II on the basis that
special controls, in addition to general
controls, can be established to provide
reasonable assurance of the safety and
effectiveness of the device. FDA
believes that the risks to health
associated with medical image analyzers
applied to mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection can be mitigated with
special controls and that these
mitigations will provide a reasonable
assurance of its safety and effectiveness.
FDA’s reasons for reclassification of
these devices are as follows:
• The risk of false positive results and
false negative results can be mitigated
by demonstrating, through clinical
performance assessment (e.g., reader
studies), that reader performance
improves when using the medical image
analyzer. In instances where a medical
image analyzer has the same intended
use but has different technological
characteristics compared to the legally
marketed device (predicate), a
performance comparison of the
predicate and new device evaluating
with the same assessment process on
the same dataset that is representative of
the intended population may be
sufficient to demonstrate device safety
and effectiveness. The risk of false
positive results and false negative
results can be further mitigated by
special controls that require sufficient
information in labeling to provide
E:\FR\FM\04JNP1.SGM
04JNP1
25602
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
detailed instructions for use to the user
and inform the user of the expected
device performance on a dataset
representative of the intended
population.
• The risk associated with misuse of
the medical image analyzers on an
unintended population can be mitigated
by specifying in the labeling and
indications for use of the device the
intended patient population for which
the device has been demonstrated to be
effective. This risk can be further
mitigated by special controls that
require informing intended users in the
labeling of foreseeable situations in
which the device is likely to fail or not
to operate at its expected performance
level.
• The risk associated with misuse of
the medical image analyzer on images
acquired from unintended image
acquisition hardware or image
acquisition parameters can be mitigated
by special controls that require
including in the device labeling
specifications for compatible imaging
hardware and imaging protocols.
• The risk resulting from not
following the intended reading protocol
can be mitigated by including in the
labeling the indications for use of the
device, by providing adequate
instructions for use including a
description of the intended reading
protocol, and by special controls
requiring that the device labeling
provide a detailed description of user
training that addresses appropriate
reading protocols for the device.
• The risk of device failure can be
mitigated by requiring design
verification and validation testing, and
special controls that require device
operating instructions. This risk can be
further mitigated by special controls
that require informing users in the
labeling of foreseeable situations in
which the device is likely to fail or not
to operate at its expected performance
level.
VII. Proposed Special Controls
FDA believes that the following
special controls, in addition to general
controls, are sufficient to mitigate the
risks to health described in section V
and provide a reasonable assurance of
safety and effectiveness for these
medical image analyzers:
• Design verification and validation
must include detailed descriptions of
image analysis algorithms, detailed
descriptions of study protocols and
datasets, results from performance
testing demonstrating the device
improves reader performance in the
intended use population, standalone
performance testing protocols and
results, and appropriate software
documentation. Performance testing
ensures that the risk of false positive
and false negative results is reduced.
• Labeling for the device must
include detailed descriptions of the
following: patient population, the
intended reading protocol, the intended
user and user training, device inputs
and outputs, compatible imaging
hardware and imaging protocols. In
addition, the labeling for the device
must also include applicable warnings,
limitations, precautions, device
operating instructions, and a detailed
summary of the performance testing.
Detailed instructions for use and
expected device performance on a
dataset representative of the intended
population in labeling helps minimize
the risk of false positive and false
negative results. Labeling ensures
proper use of the device, including
warnings to inform users of foreseeable
situations in which the device is likely
to fail or not to operate at its expected
performance level.
Table 1 shows how FDA believes the
special controls set forth in the
proposed order will mitigate each of the
risks to health described in section V.
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR MEDICAL IMAGE ANALYZERS
Identified risk to health
Mitigation measures/21 CFR section
False positive results ................................................................................................
Special controls 1 (21 CFR 892.2070(b)(1)) and 2 (21 CFR
892.2070(b)(2)).
Special controls 1 (21 CFR 892.2070(b)(1)) and 2 (21 CFR
892.2070(b)(2)).
Special control 2 (21 CFR 892.2070(b)(2)).
False negative results ...............................................................................................
sradovich on DSK3GMQ082PROD with PROPOSALS
Device misuse (analyzing images from an unintended patient population, images
acquired with incompatible imaging hardware, or incompatible image acquisition parameters) resulting in possibly lower device performance.
Device misuse (not following the appropriate reading protocol) which may lead to
lower sensitivity.
Device failure ............................................................................................................
In addition, FDA is proposing to limit
these devices to prescription use under
§ 801.109. Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act and § 801.5, as long as the
conditions of § 801.109 are met
(referring to 21 U.S.C. 352(f)(1)). Under
§ 807.81, the device would continue to
be subject to 510(k) notification
requirements.
If this proposed order is finalized,
medical image analyzers including
CADe devices for mammography breast
cancer, ultrasound breast lesions,
radiograph lung nodules, and
radiograph dental caries detection will
be reclassified into class II. The
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
Special control 2 (21 CFR 892.2070(b)(2)).
Special control 2 (21 CFR 892.2070(b)(2)).
reclassification will be codified in
§ 892.2070. FDA believes that adherence
to the proposed special controls, in
addition to the general controls, is
necessary to provide a reasonable
assurance of the safety and effectiveness
of the devices. FDA intends to update
the guidance document entitled
‘‘Clinical Performance Assessment:
Considerations for Computer-Assisted
Detection Devices Applied to Radiology
Images and Radiology Device Data—
Premarket Approval (PMA) and
Premarket Notification [510(k)]
Submissions’’ to make it consistent with
this reclassification upon finalization of
this proposed reclassification order.
PO 00000
Frm 00020
Fmt 4702
Sfmt 4702
VIII. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IX. Paperwork Reduction Act of 1995
FDA tentatively concludes that this
proposed order contains no new
collections of information. Therefore,
clearance by the Office of Management
and Budget (OMB) under the Paperwork
Reduction Act of 1995 (PRA) (44 U.S.C.
3501–3520) is not required. This
E:\FR\FM\04JNP1.SGM
04JNP1
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
proposed order refers to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by OMB under the PRA. The
collections of information in 21 CFR
part 807, subpart E have been approved
under OMB control number 0910–0120
and the collections of information in 21
CFR part 801 have been approved under
OMB control number 0910–0485.
X. Proposed Effective Date
FDA proposes that any final order
based on this proposed order become
effective 30 days after its date of
publication in the Federal Register.
sradovich on DSK3GMQ082PROD with PROPOSALS
XI. References
The following references are on
display in the Dockets Management
Staff (see ADDRESSES) and are available
for viewing by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday; they are also available
electronically at https://
www.regulations.gov. FDA has verified
the website addresses, as of the date this
document publishes in the Federal
Register, but websites are subject to
change over time.
1. Transcript of the FDA Radiological Devices
Panel Meeting, March 4–5, 2008
(available at: https://www.accessdata.
fda.gov/scripts/cdrh/cfdocs/cfAdvisory/
results.cfm?panel=24&searchtype=
1&month=0&year=&maxrows=10).
2. Transcript of the FDA Radiological Devices
Panel Meeting, November 18, 2009
(available at: https://wayback.archiveit.org/7993/20170404002254/https://
www.fda.gov/downloads/Advisory
Committees/CommitteesMeeting
Materials/MedicalDevices/Medical
DevicesAdvisoryCommittee/Radiological
DevicesPanel/UCM197419.pdf).
3. ‘‘Guidance for Industry and Food and Drug
Administration Staff—ComputerAssisted Detection Devices Applied to
Radiology Images and Radiology Device
Data—Premarket Notification [510(k)]
Submissions,’’ issued July 3, 2012
(https://www.fda.gov/downloads/
MedicalDevices/DeviceRegulationand
Guidance/GuidanceDocuments/
cm187294.pdf).
4. ‘‘Guidance for Industry and FDA Staff—
Clinical Performance Assessment:
Considerations for Computer-Assisted
Detection Devices Applied to Radiology
Images and Radiology Device Data—
Premarket Approval (PMA) and
Premarket Notification [510(k)]
Submissions,’’ issued July 3, 2012
(https://www.fda.gov/downloads/
MedicalDevices/Device
RegulationandGuidance/Guidance
Documents/ucm187315.pdf).
´
5. Dromain, C., B. Boyer, R. Ferre, et al.,
‘‘Computed-Aided Diagnosis (CAD) in
the Detection of Breast Cancer,’’
European Journal of Radiology, 82(3):
417–423 (2013).
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
6. Fenton, J.J., G. Xing, J.G. Elmore, et al.,
‘‘Short-Term Outcomes of Screening
Mammography Using Computer-Aided
Detection: A Population-Based Study of
Medicare Enrollees,’’ Annals of Internal
Medicine, 158: 580–587 (2013).
7. Gur, D., J.H. Sumkin, H.E. Rockette, et al.,
‘‘Changes in Breast Cancer Detection and
Mammography Recall Rates After the
Introduction of a Computer-Aided
Detection System,’’ Journal of the
National Cancer Institute, 96: 185–190
(2004).
8. Noble M., W. Bruening, S. Uhl, and K.
Schoelles, ‘‘Computer-Aided Detection
Mammography for Breast Cancer
Screening: Systematic Review and MetaAnalysis,’’ Archives of Gynecology and
Obstetrics, 279(6): 881–90 (2009).
List of Subjects in 21 CFR Part 892
Radiology devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 892 be amended as follows:
PART 892—RADIOLOGY DEVICES
1. The authority citation for part 892
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 892.2070 to subpart B to read
as follows:
■
§ 892.2070
Medical image analyzer.
(a) Identification. Medical image
analyzers, including computer-assisted/
aided detection (CADe) devices for
mammography breast cancer,
ultrasound breast lesions, radiograph
lung nodules, and radiograph dental
caries detection, is a prescription device
that is intended to identify, mark,
highlight, or in any other manner direct
the clinicians’ attention to portions of a
radiology image that may reveal
abnormalities during interpretation of
patient radiology images by the
clinicians. This device incorporates
pattern recognition and data analysis
capabilities and operates on previously
acquired medical images. This device is
not intended to replace the review by a
qualified radiologist, and is not
intended to be used for triage, or to
recommend diagnosis.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Design verification and validation
must include:
(i) A detailed description of the image
analysis algorithms including a
description of the algorithm inputs and
outputs, each major component or
block, and algorithm limitations.
(ii) A detailed description of prespecified performance testing methods
PO 00000
Frm 00021
Fmt 4702
Sfmt 4702
25603
and dataset(s) used to assess whether
the device will improve reader
performance as intended and to
characterize the standalone device
performance. Performance testing
includes one or more standalone tests,
side-by-side comparisons, or a reader
study, as applicable.
(iii) Results from performance testing
that demonstrate that the device
improves reader performance in the
intended use population when used in
accordance with the instructions for
use. The performance assessment must
be based on appropriate diagnostic
accuracy measures (e.g., receiver
operator characteristic plot, sensitivity,
specificity, predictive value, and
diagnostic likelihood ratio). The test
dataset must contain a sufficient
number of cases from important cohorts
(e.g., subsets defined by clinically
relevant confounders, effect modifiers,
concomitant diseases, and subsets
defined by image acquisition
characteristics) such that the
performance estimates and confidence
intervals of the device for these
individual subsets can be characterized
for the intended use population and
imaging equipment.
(iv) Appropriate software
documentation (e.g., device hazard
analysis; software requirements
specification document; software design
specification document; traceability
analysis; description of verification and
validation activities including system
level test protocol, pass/fail criteria, and
results; and cybersecurity).
(2) Labeling must include the
following:
(i) A detailed description of the
patient population for which the device
is indicated for use.
(ii) A detailed description of the
intended reading protocol.
(iii) A detailed description of the
intended user and user training that
addresses appropriate reading protocols
for the device.
(iv) A detailed description of the
device inputs and outputs.
(v) A detailed description of
compatible imaging hardware and
imaging protocols.
(vi) Discussion of warnings,
precautions, and limitations must
include situations in which the device
may fail or may not operate at its
expected performance level (e.g., poor
image quality or for certain
subpopulations), as applicable.
(vii) Device operating instructions.
(viii) A detailed summary of the
performance testing, including: test
methods, dataset characteristics, results,
and a summary of sub-analyses on case
distributions stratified by relevant
E:\FR\FM\04JNP1.SGM
04JNP1
25604
Federal Register / Vol. 83, No. 107 / Monday, June 4, 2018 / Proposed Rules
confounders, such as lesion and organ
characteristics, disease stages, and
imaging equipment.
Dated: May 29, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–11880 Filed 6–1–18; 8:45 am]
BILLING CODE 4164–01–P
ENVIRONMENTAL PROTECTION
AGENCY
40 CFR Part 52
[EPA–R04–OAR–2018–0073; FRL–9978–
92—Region 4]
Air Plan Approval; SC; Regional Haze
Plan and Prong 4 (Visibility) for the
2012 PM2.5, 2010 NO2, 2010 SO2, and
2008 Ozone NAAQS
Environmental Protection
Agency (EPA).
ACTION: Proposed rule.
AGENCY:
The Environmental Protection
Agency (EPA) is proposing to take the
following four actions regarding the
South Carolina State Implementation
Plan (SIP): Approve the portion of South
Carolina’s September 5, 2017, SIP
submittal seeking to change reliance
from the Clean Air Interstate Rule
(CAIR) to the Cross-State Air Pollution
Rule (CSAPR) for certain regional haze
requirements; convert EPA’s limited
approval/limited disapproval of South
Carolina’s regional haze plan to a full
approval; remove EPA’s Federal
Implementation Plan (FIP) for South
Carolina, which replaced reliance on
CAIR with reliance on CSAPR to
address the deficiencies identified in
the limited disapproval of South
Carolina’s regional haze plan; and
convert the conditional approvals of the
visibility prong of South Carolina’s
infrastructure SIP submittals for the
2012 Fine Particulate Matter (PM2.5),
2010 Nitrogen Dioxide (NO2), 2010
Sulfur Dioxide (SO2), and 2008 8-hour
Ozone National Ambient Air Quality
Standards (NAAQS) to full approvals.
DATES: Comments must be received on
or before July 5, 2018.
ADDRESSES: Submit your comments,
identified by Docket ID No. EPA–R04–
OAR–2018–0073 at https://
www.regulations.gov. Follow the online
instructions for submitting comments.
Once submitted, comments cannot be
edited or removed from Regulations.gov.
EPA may publish any comment received
to its public docket. Do not submit
electronically any information you
consider to be Confidential Business
Information (CBI) or other information
sradovich on DSK3GMQ082PROD with PROPOSALS
SUMMARY:
VerDate Sep<11>2014
16:57 Jun 01, 2018
Jkt 244001
whose disclosure is restricted by statute.
Multimedia submissions (audio, video,
etc.) must be accompanied by a written
comment. The written comment is
considered the official comment and
should include discussion of all points
you wish to make. EPA will generally
not consider comments or comment
contents located outside of the primary
submission (i.e., on the web, cloud, or
other file sharing system). For
additional submission methods, the full
EPA public comment policy,
information about CBI or multimedia
submissions, and general guidance on
making effective comments, please visit
https://www2.epa.gov/dockets/
commenting-epa-dockets.
FOR FURTHER INFORMATION CONTACT:
Michele Notarianni, Air Regulatory
Management Section, Air Planning and
Implementation Branch, Air, Pesticides
and Toxics Management Division, U.S.
Environmental Protection Agency,
Region 4, 61 Forsyth Street SW, Atlanta,
Georgia 30303–8960. Ms. Notarianni can
be reached by telephone at (404) 562–
9031 or via electronic mail at
notarianni.michele@epa.gov.
SUPPLEMENTARY INFORMATION:
I. Background
A. Regional Haze Plans and Their
Relationship With CAIR and CSAPR
Section 169A(b)(2)(A) of the Clean Air
Act (CAA or Act) requires states to
submit regional haze plans that contain
such measures as may be necessary to
make reasonable progress towards the
natural visibility goal, including a
requirement that certain categories of
existing major stationary sources built
between 1962 and 1977 procure, install,
and operate Best Available Retrofit
Technology (BART) as determined by
the state. Under the Regional Haze Rule
(RHR), states are directed to conduct
BART determinations for such ‘‘BARTeligible’’ sources that may be
anticipated to cause or contribute to any
visibility impairment in a Class I area.
Rather than requiring source-specific
BART controls, states also have the
flexibility to adopt an emissions trading
program or other alternative program as
long as the alternative provides greater
reasonable progress towards improving
visibility than BART. See 40 CFR
51.308(e)(2). EPA provided states with
this flexibility in the RHR, adopted in
1999, and further refined the criteria for
assessing whether an alternative
program provides for greater reasonable
progress in two subsequent
rulemakings. See 64 FR 35714 (July 1,
1999); 70 FR 39104 (July 6, 2005); 71 FR
60612 (October 13, 2006).
PO 00000
Frm 00022
Fmt 4702
Sfmt 4702
EPA demonstrated that CAIR would
achieve greater reasonable progress than
BART in revisions to the regional haze
program made in 2005.1 See 70 FR 39104
(July 6, 2005). In those revisions, EPA
amended its regulations to provide that
states participating in the CAIR cap-andtrade programs pursuant to an EPAapproved CAIR SIP or states that remain
subject to a CAIR FIP need not require
affected BART-eligible electric
generating units (EGUs) to install,
operate, and maintain BART for
emissions of SO2 and nitrogen oxides
(NOX). As a result of EPA’s
determination that CAIR was ‘‘betterthan-BART,’’ a number of states in the
CAIR region, including South Carolina,
relied on the CAIR cap-and-trade
programs as an alternative to BART for
EGU emissions of SO2 and NOX in
designing their regional haze plans.
These states also relied on CAIR as an
element of a long-term strategy (LTS) for
achieving their reasonable progress
goals (RPGs) for their regional haze
programs. However, in 2008, the United
States Court of Appeals for the District
of Columbia Circuit (D.C. Circuit)
remanded CAIR to EPA without vacatur
to preserve the environmental benefits
provided by CAIR. North Carolina v.
EPA, 550 F.3d 1176, 1178 (DC Cir.
2008). On August 8, 2011 (76 FR 48208),
acting on the D.C. Circuit’s remand, EPA
promulgated CSAPR to replace CAIR
and issued FIPs to implement the rule
in CSAPR-subject states.2
Implementation of CSAPR was
scheduled to begin on January 1, 2012,
when CSAPR would have superseded
the CAIR program.
Due to the D.C. Circuit’s 2008 ruling
that CAIR was ‘‘fatally flawed’’ and its
resulting status as a temporary measure
following that ruling, EPA could not
fully approve regional haze plans to the
extent that they relied on CAIR to satisfy
the BART requirement and the
1 CAIR created regional cap-and-trade programs to
reduce SO2 and NOX emissions in 27 eastern states
(and the District of Columbia), including South
Carolina, that contributed to downwind
nonattainment or interfered with maintenance of
the 1997 8-hour ozone NAAQS or the 1997 PM2.5
NAAQS.
2 CSAPR requires 28 eastern states to limit their
statewide emissions of SO2 and/or NOX in order to
mitigate transported air pollution unlawfully
impacting other states’ ability to attain or maintain
four NAAQS: The 1997 ozone NAAQS, the 1997
annual PM2.5 NAAQS, the 2006 24-hour PM2.5
NAAQS, and the 2008 8-hour ozone NAAQS. The
CSAPR emissions limitations are defined in terms
of maximum statewide ‘‘budgets’’ for emissions of
annual SO2, annual NOX, and/or ozone-season NOX
by each covered state’s large EGUs. The CSAPR
state budgets are implemented in two phases of
generally increasing stringency, with the Phase 1
budgets applying to emissions in 2015 and 2016
and the Phase 2 budgets applying to emissions in
2017 and later years.
E:\FR\FM\04JNP1.SGM
04JNP1
Agencies
[Federal Register Volume 83, Number 107 (Monday, June 4, 2018)]
[Proposed Rules]
[Pages 25598-25604]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2018-11880]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 892
[Docket No. FDA-2018-N-1553]
Radiology Devices; Reclassification of Medical Image Analyzers
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or the Agency) is
issuing this proposed order to reclassify medical image analyzers
applied to mammography breast cancer, ultrasound breast lesions,
radiograph lung nodules, and radiograph dental caries detection as
postamendments class III (premarket approval) devices (regulated under
product code MYN), into class II (special controls), subject to
premarket notification. FDA is also identifying the proposed special
controls that the Agency believes are necessary to provide a reasonable
assurance of safety and effectiveness of the device. These devices are
intended to direct the clinician's attention to portions of an image
that may reveal abnormalities during interpretation of patient's
radiology images by the clinician. If finalized, this order will
reclassify these types of devices from class III to class II and reduce
regulatory burdens on industry as these types of devices will no longer
be required to submit a premarket approval application (PMA) but can
instead submit a less burdensome premarket notification (510(k)) before
marketing their device.
[[Page 25599]]
DATES: Submit either electronic or written comments on the proposed
order by August 3, 2018. Please see section X of this document for the
proposed effective date when the new requirements apply and for the
proposed effective date of a final order based on this proposed order.
ADDRESSES: You may submit comments as follows: Please note that late,
untimely filed comments will not be considered. Electronic comments
must be submitted on or before August 3, 2018. The https://www.regulations.gov electronic filing system will accept comments until
midnight Eastern Time at the end of August 3, 2018. Comments received
by mail/hand delivery/courier (for written/paper submissions) will be
considered timely if they are postmarked or the delivery service
acceptance receipt is on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal Rulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2018-N-1553 for ``Radiology Devices; Reclassification of Medical
Image Analyzers.'' Received comments will be placed in the docket and,
except for those submitted as ``Confidential Submissions,'' publicly
viewable at https://www.regulations.gov or at the Dockets Management
Staff between 9 a.m. and 4 p.m., Monday through Friday.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.gpo.gov/fdsys/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Dockets Management Staff, 5630 Fishers Lane,
Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Robert Ochs, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4312, Silver Spring, MD 20993-0002, 301-796-6661,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act), as
amended, establishes a comprehensive system for the regulation of
medical devices intended for human use. Section 513 of the FD&C Act (21
U.S.C. 360c) established three categories (classes) of devices,
reflecting the regulatory controls needed to provide reasonable
assurance of their safety and effectiveness. The three categories of
devices are class I (general controls), class II (special controls),
and class III (premarket approval).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices), are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until, the device
is reclassified into class I or II, or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and 21 CFR part 807.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA acting by order can
reclassify the device into class I or II on its own initiative, or in
response to a petition from the manufacturer or importer of the device.
To change the classification of the device, the proposed new class must
have sufficient regulatory controls to provide a reasonable assurance
of the safety and effectiveness of the device for its intended use.
Reevaluation of the data previously before the Agency is an
appropriate basis for subsequent action where the reevaluation is made
in light of newly available regulatory authority (see Bell v. Goddard,
366 F.2d 177, 181 (7th Cir. 1966); Ethicon, Inc. v. FDA, 762 F. Supp.
382, 388-391 (D.D.C. 1991)), or in light of changes in ``medical
science'' (Upjohn v. Finch, 422 F.2d 944, 951 (6th Cir. 1970)). Whether
data before the Agency are old or new, the ``new information'' to
support reclassification
[[Page 25600]]
under 513(f)(3) must be ``valid scientific evidence'', as defined in
section 513(a)(3) of the FD&C Act and 21 CFR 860.7(c)(2). (See, e.g.,
General Medical Co. v. FDA, 770 F.2d 214 (D.C. Cir. 1985); Contact Lens
Mfrs. Assoc. v. FDA, 766 F.2d 592 (DC Cir.1985), cert. denied, 474 U.S.
1062 (1986)). FDA relies upon ``valid scientific evidence'' in the
classification process to determine the level of regulation for
devices. To be considered in the reclassification process, the ``valid
scientific evidence'' upon which the Agency relies must be publicly
available. Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
PMA (see section 520(c) of the FD&C Act (21 U.S.C. 360j(c)).
In accordance with section 513(f)(3) of the FD&C Act, the Agency is
proposing to reclassify medical image analyzers applied to mammography
breast cancer, ultrasound breast lesions, radiograph lung nodules, and
radiograph dental caries detection from class III into class II on the
basis that there is sufficient information to establish special
controls, in addition to general controls, to provide reasonable
assurance of the safety and effectiveness of the device.
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the 510(k) premarket notification requirements, if the
Agency determines that premarket notification is not necessary to
reasonably assure the safety and effectiveness of the device.
II. Regulatory History of the Devices
This proposed order covers medical image analyzers including
computer-assisted/aided detection (CADe) devices for mammography breast
cancer, ultrasound breast lesions, radiograph lung nodules, and
radiograph dental caries detection that are assigned product code MYN.
These postamendments devices are currently regulated as class III
devices under section 513(f)(1) of the FD&C Act. FDA has experience
reviewing and analyzing data and information for medical image
analyzers since premarket approval of the first device for these uses
in 1998. On June 26, 1998, the Center for Devices and Radiological
Health (CDRH) approved the first CADe device included in this
reclassification order. In the December 30, 1998, Federal Register
notice (63 FR 71930), FDA announced a PMA approval order for R2
Technology, Inc. M 1000 Image Checker and the availability of the
summary of safety and effectiveness data for the device. Since 1998, 11
devices have received premarket approval for the analysis of several
modalities, including mammography, ultrasound, as well as chest and
dental radiographs. Based upon our review experience and consistent
with the FD&C Act and FDA's regulations, FDA believes that these
devices should be reclassified from class III into class II because
there is sufficient information to establish special controls that can
provide reasonable assurance of the device's safety and effectiveness.
This proposed order does not apply to medical image analyzers/CADe
devices currently classified under Sec. 892.2050 (21 CFR 892.2050),
Picture archiving and communication system. FDA has regulated other
CADe devices intended to aid lung nodule and colon polyp detection from
computed tomography images as class II devices under Sec. 892.2050,
Picture archiving and communication system and assigned the following
product codes:
NWE (Colon Computed Tomography System, Computer-Aided
Detection);
OEB (Lung Computed Tomography System, Computer-Aided
Detection);
OMJ (Chest X-Ray Computer Aided Detection).
There have been no recalls for class II CADe devices. As of the
date of this proposal, FDA has received three recalls for class III
devices and one Medical Device Report (MDR), however, in the past 10
years only one recall for the class III devices has been received due
to distribution of the CADe device without PMA approval. None of these
recalls were classified as a Class I recall. There were also no MDRs
related to either the class III medical image analyzers or class II
CADe devices in the past 10 years. This evidence suggests that the
safety profiles for existing class III CADe devices are similar to the
class II CADe, and consequently that our regulatory controls applied
should be similar.
III. Device Description
This proposed order applies to medical image analyzers including
CADe devices for mammography breast cancer, ultrasound breast lesions,
radiograph lung nodules, and radiograph dental caries detection that
are currently regulated as class III devices as postamendment devices.
These devices are intended to identify, mark, highlight, or in any
other manner direct the clinicians' attention to portions of a
radiology image that may reveal abnormalities during interpretation of
patient radiology images by the clinicians. These devices incorporate
pattern recognition and data analysis capabilities and operate on
previously acquired radiology images, including mammography,
radiograph, and ultrasound. These devices are not intended to replace
the review by a qualified radiologist or to be used for triage.
Furthermore, these devices are not intended to recommend diagnosis of
any diseases.
IV. Proposed Reclassification
The Radiological Devices Panel (the Panel) convened on March 4-5,
2008 (Ref. 1) and discussed issues relating to how medical image
analyzers including CADe devices are used in clinical decisionmaking,
how the performance of the devices should be evaluated, and the
information needed to determine whether the device provides a
reasonable assurance of its safety and effectiveness. Additional
discussions were held regarding medical image analyzers for mammography
and radiograph applications. Following the 2008 Panel Meeting, FDA
convened a second meeting of the Panel on November 18, 2009. The 2009
Panel Meeting was asked to discuss two proposed draft guidances for the
evaluation of medical image analyzers and the Agency's regulatory
strategy for these devices (Ref. 2). Subsequently, the two draft
guidance documents were finalized by FDA and were made public on July
3, 2012 (Refs. 3 and 4). The guidance document entitled ``Clinical
Performance Assessment: Considerations for Computer-Assisted Detection
Devices Applied to Radiology Images and Radiology Device Data--
Premarket Approval (PMA) and Premarket Notification [510(k)]
Submissions'' provides guidance regarding clinical performance
assessment studies for CADe applied to radiology images and radiology
device data. The guidance document entitled ``Computer-Assisted
Detection Devices Applied to Radiology Images and Radiology Device
Data--Premarket Notification [510(k)] Submissions'' provides guidance
regarding premarket notification (510(k)) submissions for CADe applied
to radiology images and radiology device data. These guidance documents
describe clinical and non-clinical methods to evaluate the safety and
effectiveness of CADe devices, including medical image analyzers
covered by this proposed order. In addition to the two guidance
documents, the Panel's discussion regarding the benefits and risks of
medical image analyzers that were discussed at the 2008 and 2009 Panel
meetings have been taken into consideration by the Agency when
developing the proposed special
[[Page 25601]]
controls provided in this proposed order below.
Since publication of these guidance documents, the Agency has
gained considerable experience in reviewing medical image analyzers
using the methods described in the aforementioned guidance documents.
Further, as part CDRH's 2014-2015 strategic priority ``Strike the Right
Balance Between Premarket and Postmarket Data Collection,'' a
retrospective review of class III devices subject to a PMA was
completed to determine whether or not, based on our current
understanding of the technology, reclassification may be appropriate.
During this retrospective review, FDA determined that sufficient
information exists such that the risks of false positive and false
negative results, misuse, and device failure can be mitigated, to
establish special controls that, together with general controls, can
provide a reasonable assurance of the safety and effectiveness of
medical image analyzers and therefore proposes these devices be
reclassified from class III to class II. On April 29, 2015, FDA
published a notice in the Federal Register entitled ``Retrospective
Review of Premarket Approval Application Devices; Striking the Balance
Between Premarket and Postmarket Data Collection'' in which FDA
announced plans to consider reclassifying medical image analyzers
identified with the MYN product code from class III to class II (80 FR
23798). No adverse comments were received regarding our proposed intent
for MYN.
In accordance with section 513(f)(3) of the FD&C Act and 21 CFR
part 860, subpart C, FDA is proposing to reclassify postamendments
medical image analyzers, including CADe devices for mammography breast
cancer, ultrasound breast lesions, radiograph lung nodules, and
radiograph dental caries detection, from class III into class II. FDA
believes that there is sufficient information to establish special
controls, in addition to general controls, that would effectively
mitigate the risks to health identified in section V and provide
reasonable assurance of the safety and effectiveness of these devices.
Absent the special controls identified in this proposed order, general
controls applicable to the device are insufficient to provide
reasonable assurance of the safety and effectiveness of the device.
FDA is proposing to create a separate classification regulation for
medical image analyzer devices that will be reclassified from class III
to II. Under this proposed order, if finalized, the medical image
analyzer devices will be identified as a prescription device. As such,
the prescription device must satisfy prescription labeling requirements
(see Sec. 801.109 (21 CFR 801.109), Prescription devices).
Prescription devices are exempt from the requirement for adequate
directions for use for the layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352) and Sec. 801.5 (21 CFR 801.5), as long as the
conditions of Sec. 801.109 are met. In this proposed order, if
finalized, the Agency has identified the special controls under section
513(a)(1)(B) of the FD&C Act that, together with general controls, will
provide a reasonable assurance of the safety and effectiveness for
medical image analyzer devices.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary to provide
reasonable assurance of safety and effectiveness and, therefore, does
not intend to exempt these proposed class II devices from the premarket
notification requirements. Persons who intend to market this type of
device must submit to FDA a 510(k) and receive clearance prior to
marketing the device.
This proposal, if finalized, will decrease regulatory burden on the
medical device industry and will reduce private costs and expenditures
required to comply with Federal Regulations. Specifically, regulated
industry will no longer have to submit a PMA but can instead submit a
510(k) to the Agency for review prior to marketing their device. A
510(k) is a less-burdensome pathway to market a device which typically
results in a more timely premarket review compared to a PMA and reduces
the regulatory burden on industry in addition to providing more timely
access of these types of devices to patients.
V. Risks to Health
From the Panel discussions on March 4-5, 2008, and November 18,
2009, along with the peer-reviewed literature (Refs. 5-8) and FDA's
experiences over the years in reviewing submissions for these devices
and similar devices, FDA determined the probable risks to health
associated with medical image analyzers including CADe devices for
mammography breast cancer, ultrasound breast lesions, radiograph lung
nodules, and radiograph dental caries detection are as follows: (1)
False positive results may result in complications, such as incorrect
management of the patient with possible adverse effects, and
unnecessary additional radiology imaging and/or invasive procedures,
such as biopsy; (2) false negative results could result in
complications, including incorrect diagnosis and delay in disease
management; (3) the device could be misused to analyze images from an
unintended patient population or on images acquired with incompatible
imaging hardware or incompatible image acquisition parameters,
resulting in possibly lower device performance; (4) the device could be
misused by not following the appropriate reading protocol, which may
lead to lower sensitivity; and (5) device failure could result in the
absence or delay of device output, or incorrect device output, which
could likewise lead to inaccurate patient assessment.
VI. Summary of the Reasons for Reclassification
After considering the information above, FDA has determined that
all class III medical image analyzers currently approved by FDA should
be reclassified into class II on the basis that special controls, in
addition to general controls, can be established to provide reasonable
assurance of the safety and effectiveness of the device. FDA believes
that the risks to health associated with medical image analyzers
applied to mammography breast cancer, ultrasound breast lesions,
radiograph lung nodules, and radiograph dental caries detection can be
mitigated with special controls and that these mitigations will provide
a reasonable assurance of its safety and effectiveness. FDA's reasons
for reclassification of these devices are as follows:
The risk of false positive results and false negative
results can be mitigated by demonstrating, through clinical performance
assessment (e.g., reader studies), that reader performance improves
when using the medical image analyzer. In instances where a medical
image analyzer has the same intended use but has different
technological characteristics compared to the legally marketed device
(predicate), a performance comparison of the predicate and new device
evaluating with the same assessment process on the same dataset that is
representative of the intended population may be sufficient to
demonstrate device safety and effectiveness. The risk of false positive
results and false negative results can be further mitigated by special
controls that require sufficient information in labeling to provide
[[Page 25602]]
detailed instructions for use to the user and inform the user of the
expected device performance on a dataset representative of the intended
population.
The risk associated with misuse of the medical image
analyzers on an unintended population can be mitigated by specifying in
the labeling and indications for use of the device the intended patient
population for which the device has been demonstrated to be effective.
This risk can be further mitigated by special controls that require
informing intended users in the labeling of foreseeable situations in
which the device is likely to fail or not to operate at its expected
performance level.
The risk associated with misuse of the medical image
analyzer on images acquired from unintended image acquisition hardware
or image acquisition parameters can be mitigated by special controls
that require including in the device labeling specifications for
compatible imaging hardware and imaging protocols.
The risk resulting from not following the intended reading
protocol can be mitigated by including in the labeling the indications
for use of the device, by providing adequate instructions for use
including a description of the intended reading protocol, and by
special controls requiring that the device labeling provide a detailed
description of user training that addresses appropriate reading
protocols for the device.
The risk of device failure can be mitigated by requiring
design verification and validation testing, and special controls that
require device operating instructions. This risk can be further
mitigated by special controls that require informing users in the
labeling of foreseeable situations in which the device is likely to
fail or not to operate at its expected performance level.
VII. Proposed Special Controls
FDA believes that the following special controls, in addition to
general controls, are sufficient to mitigate the risks to health
described in section V and provide a reasonable assurance of safety and
effectiveness for these medical image analyzers:
Design verification and validation must include detailed
descriptions of image analysis algorithms, detailed descriptions of
study protocols and datasets, results from performance testing
demonstrating the device improves reader performance in the intended
use population, standalone performance testing protocols and results,
and appropriate software documentation. Performance testing ensures
that the risk of false positive and false negative results is reduced.
Labeling for the device must include detailed descriptions
of the following: patient population, the intended reading protocol,
the intended user and user training, device inputs and outputs,
compatible imaging hardware and imaging protocols. In addition, the
labeling for the device must also include applicable warnings,
limitations, precautions, device operating instructions, and a detailed
summary of the performance testing. Detailed instructions for use and
expected device performance on a dataset representative of the intended
population in labeling helps minimize the risk of false positive and
false negative results. Labeling ensures proper use of the device,
including warnings to inform users of foreseeable situations in which
the device is likely to fail or not to operate at its expected
performance level.
Table 1 shows how FDA believes the special controls set forth in
the proposed order will mitigate each of the risks to health described
in section V.
Table 1--Risks to Health and Mitigation Measures for Medical Image
Analyzers
------------------------------------------------------------------------
Mitigation measures/21 CFR
Identified risk to health section
------------------------------------------------------------------------
False positive results.................... Special controls 1 (21 CFR
892.2070(b)(1)) and 2 (21
CFR 892.2070(b)(2)).
False negative results.................... Special controls 1 (21 CFR
892.2070(b)(1)) and 2 (21
CFR 892.2070(b)(2)).
Device misuse (analyzing images from an Special control 2 (21 CFR
unintended patient population, images 892.2070(b)(2)).
acquired with incompatible imaging
hardware, or incompatible image
acquisition parameters) resulting in
possibly lower device performance.
Device misuse (not following the Special control 2 (21 CFR
appropriate reading protocol) which may 892.2070(b)(2)).
lead to lower sensitivity.
Device failure............................ Special control 2 (21 CFR
892.2070(b)(2)).
------------------------------------------------------------------------
In addition, FDA is proposing to limit these devices to
prescription use under Sec. 801.109. Prescription devices are exempt
from the requirement for adequate directions for use for the layperson
under section 502(f)(1) of the FD&C Act and Sec. 801.5, as long as the
conditions of Sec. 801.109 are met (referring to 21 U.S.C. 352(f)(1)).
Under Sec. 807.81, the device would continue to be subject to 510(k)
notification requirements.
If this proposed order is finalized, medical image analyzers
including CADe devices for mammography breast cancer, ultrasound breast
lesions, radiograph lung nodules, and radiograph dental caries
detection will be reclassified into class II. The reclassification will
be codified in Sec. 892.2070. FDA believes that adherence to the
proposed special controls, in addition to the general controls, is
necessary to provide a reasonable assurance of the safety and
effectiveness of the devices. FDA intends to update the guidance
document entitled ``Clinical Performance Assessment: Considerations for
Computer-Assisted Detection Devices Applied to Radiology Images and
Radiology Device Data--Premarket Approval (PMA) and Premarket
Notification [510(k)] Submissions'' to make it consistent with this
reclassification upon finalization of this proposed reclassification
order.
VIII. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IX. Paperwork Reduction Act of 1995
FDA tentatively concludes that this proposed order contains no new
collections of information. Therefore, clearance by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(PRA) (44 U.S.C. 3501-3520) is not required. This
[[Page 25603]]
proposed order refers to previously approved collections of information
found in other FDA regulations. These collections of information are
subject to review by OMB under the PRA. The collections of information
in 21 CFR part 807, subpart E have been approved under OMB control
number 0910-0120 and the collections of information in 21 CFR part 801
have been approved under OMB control number 0910-0485.
X. Proposed Effective Date
FDA proposes that any final order based on this proposed order
become effective 30 days after its date of publication in the Federal
Register.
XI. References
The following references are on display in the Dockets Management
Staff (see ADDRESSES) and are available for viewing by interested
persons between 9 a.m. and 4 p.m., Monday through Friday; they are also
available electronically at https://www.regulations.gov. FDA has
verified the website addresses, as of the date this document publishes
in the Federal Register, but websites are subject to change over time.
1. Transcript of the FDA Radiological Devices Panel Meeting, March
4-5, 2008 (available at: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfAdvisory/results.cfm?panel=24&searchtype=1&month=0&year=&maxrows=10).
2. Transcript of the FDA Radiological Devices Panel Meeting,
November 18, 2009 (available at: https://wayback.archive-it.org/7993/20170404002254/https://www.fda.gov/downloads/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/MedicalDevicesAdvisoryCommittee/RadiologicalDevicesPanel/UCM197419.pdf).
3. ``Guidance for Industry and Food and Drug Administration Staff--
Computer-Assisted Detection Devices Applied to Radiology Images and
Radiology Device Data--Premarket Notification [510(k)]
Submissions,'' issued July 3, 2012 (https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/cm187294.pdf).
4. ``Guidance for Industry and FDA Staff--Clinical Performance
Assessment: Considerations for Computer-Assisted Detection Devices
Applied to Radiology Images and Radiology Device Data--Premarket
Approval (PMA) and Premarket Notification [510(k)] Submissions,''
issued July 3, 2012 (https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm187315.pdf).
5. Dromain, C., B. Boyer, R. Ferr[eacute], et al., ``Computed-Aided
Diagnosis (CAD) in the Detection of Breast Cancer,'' European
Journal of Radiology, 82(3): 417-423 (2013).
6. Fenton, J.J., G. Xing, J.G. Elmore, et al., ``Short-Term Outcomes
of Screening Mammography Using Computer-Aided Detection: A
Population-Based Study of Medicare Enrollees,'' Annals of Internal
Medicine, 158: 580-587 (2013).
7. Gur, D., J.H. Sumkin, H.E. Rockette, et al., ``Changes in Breast
Cancer Detection and Mammography Recall Rates After the Introduction
of a Computer-Aided Detection System,'' Journal of the National
Cancer Institute, 96: 185-190 (2004).
8. Noble M., W. Bruening, S. Uhl, and K. Schoelles, ``Computer-Aided
Detection Mammography for Breast Cancer Screening: Systematic Review
and Meta-Analysis,'' Archives of Gynecology and Obstetrics, 279(6):
881-90 (2009).
List of Subjects in 21 CFR Part 892
Radiology devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 892 be amended as follows:
PART 892--RADIOLOGY DEVICES
0
1. The authority citation for part 892 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 892.2070 to subpart B to read as follows:
Sec. 892.2070 Medical image analyzer.
(a) Identification. Medical image analyzers, including computer-
assisted/aided detection (CADe) devices for mammography breast cancer,
ultrasound breast lesions, radiograph lung nodules, and radiograph
dental caries detection, is a prescription device that is intended to
identify, mark, highlight, or in any other manner direct the
clinicians' attention to portions of a radiology image that may reveal
abnormalities during interpretation of patient radiology images by the
clinicians. This device incorporates pattern recognition and data
analysis capabilities and operates on previously acquired medical
images. This device is not intended to replace the review by a
qualified radiologist, and is not intended to be used for triage, or to
recommend diagnosis.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms
including a description of the algorithm inputs and outputs, each major
component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing
methods and dataset(s) used to assess whether the device will improve
reader performance as intended and to characterize the standalone
device performance. Performance testing includes one or more standalone
tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the
device improves reader performance in the intended use population when
used in accordance with the instructions for use. The performance
assessment must be based on appropriate diagnostic accuracy measures
(e.g., receiver operator characteristic plot, sensitivity, specificity,
predictive value, and diagnostic likelihood ratio). The test dataset
must contain a sufficient number of cases from important cohorts (e.g.,
subsets defined by clinically relevant confounders, effect modifiers,
concomitant diseases, and subsets defined by image acquisition
characteristics) such that the performance estimates and confidence
intervals of the device for these individual subsets can be
characterized for the intended use population and imaging equipment.
(iv) Appropriate software documentation (e.g., device hazard
analysis; software requirements specification document; software design
specification document; traceability analysis; description of
verification and validation activities including system level test
protocol, pass/fail criteria, and results; and cybersecurity).
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the
device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training
that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and
imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must
include situations in which the device may fail or may not operate at
its expected performance level (e.g., poor image quality or for certain
subpopulations), as applicable.
(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including:
test methods, dataset characteristics, results, and a summary of sub-
analyses on case distributions stratified by relevant
[[Page 25604]]
confounders, such as lesion and organ characteristics, disease stages,
and imaging equipment.
Dated: May 29, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-11880 Filed 6-1-18; 8:45 am]
BILLING CODE 4164-01-P