Medical Devices; Hematology and Pathology Devices; Classification of Lynch Syndrome Test Systems, 8355-8357 [2018-03924]
Download as PDF
daltland on DSKBBV9HB2PROD with RULES
Federal Register / Vol. 83, No. 39 / Tuesday, February 27, 2018 / Rules and Regulations
Background
Pursuant to the provisions of the
Convention on Cultural Property
Implementation Act (hereafter, the
Cultural Property Implementation Act
or the Act) (Pub. L. 97–446, 19 U.S.C.
2601 et seq.), which implements the
1970 United Nations Educational,
Scientific and Cultural Organization
(UNESCO) Convention on the Means of
Prohibiting and Preventing the Illicit
Import, Export and Transfer of
Ownership of Cultural Property
(hereinafter, the Convention), in U.S.
law, the United States may enter into an
international agreement with another
State Party to the Convention to impose
import restrictions on eligible
archaeological and ethnological
materials under procedures and
requirements prescribed by the Act.
Under the Act and applicable CBP
regulations (19 CFR 12.104g), the
restrictions are effective for no more
than five years beginning on the date on
which the agreement enters into force
with respect to the United States (19
U.S.C. 2602(b)). This period may be
extended for additional periods, not to
exceed five years, if it is determined that
the factors justifying the initial
agreement still pertain and no cause for
suspension of the agreement exists (19
U.S.C. 2602(e); 19 CFR 12.104g(a)).
On February 27, 2013, the United
States entered into a bilateral agreement
with the Government of Belize
concerning the imposition of import
restrictions on certain categories of
archaeological material originating in
Belize, pursuant to the Act. (The
agreement can be found online at
https://eca.state.gov/files/bureau/
bzmou2013.pdf.) On March 5, 2013,
CBP published CBP Dec. 13–05 in the
Federal Register (78 FR 14183), which
amended 19 CFR 12.104g(a) to reflect
the imposition of restrictions on this
material and included a list designating
the types of archaeological material
covered by the restrictions. These
restrictions were to be effective through
February 27, 2018.
On January 12, 2018, after reviewing
the findings and recommendations of
the Cultural Property Advisory
Committee, the Acting Assistant
Secretary for Educational and Cultural
Affairs, Department of State, concluding
that the cultural heritage of Belize
continues to be in jeopardy from pillage
of certain archaeological material, made
the necessary statutory determinations,
and decided to extend the agreement
with Belize for an additional five-year
period to February 27, 2023. Diplomatic
notes have been exchanged that reflect
the extension of the agreement.
VerDate Sep<11>2014
18:04 Feb 26, 2018
Jkt 244001
Accordingly, CBP is amending 19 CFR
12.104g(a) in order to reflect the
extension of the import restrictions
pursuant to the agreement.
The Designated List of Archaeological
Material originating in Belize covered
by these import restrictions is set forth
in CBP Dec. 13–05, which can be found
online at: https://eca.state.gov/files/
bureau/bz2013dlfrn.pdf.
The restrictions on the importation of
this archaeological material originating
in Belize are to continue in effect for an
additional five years. Importation of
such material continues to be restricted
unless the conditions set forth in 19
U.S.C. 2606 and 19 CFR 12.104c are
met.
Inapplicability of Notice and Delayed
Effective Date
This amendment involves a foreign
affairs function of the United States and
is, therefore, being made without notice
or public procedure (5 U.S.C. 553(a)(1)).
In addition, CBP has determined that
such notice or public procedure would
be impracticable and contrary to the
public interest because the action being
taken is essential to avoid interruption
of the application of the existing import
restrictions (5 U.S.C. 553(b)(B)). For the
same reason, a delayed effective date is
not required under 5 U.S.C 553(d)(3).
Regulatory Flexibility Act
Because no notice of proposed
rulemaking is required, the provisions
of the Regulatory Flexibility Act (5
U.S.C. 601 et seq.) do not apply.
Executive Orders 12866 and 13771
Because this rule involves a foreign
affairs function of the United States, it
is not subject to either Executive Order
12866 or Executive Order 13771.
Signing Authority
This regulation is being issued in
accordance with 19 CFR 0.1(a)(1).
List of Subjects in 19 CFR Part 12
Cultural property, Customs duties and
inspection, Imports, Prohibited
merchandise.
Amendment to CBP Regulations
For the reasons set forth above, part
12 of Title 19 of the Code of Federal
Regulations (19 CFR part 12), is
amended as set forth below.
PART 12—SPECIAL CLASSES OF
MERCHANDISE
1. The general authority citation for
part 12 and the specific authority
citation for § 12.104g continue to read as
follows:
■
PO 00000
Frm 00035
Fmt 4700
Sfmt 4700
8355
Authority: 5 U.S.C. 301; 19 U.S.C. 66, 1202
(General Note 3(i), Harmonized Tariff
Schedule of the United States (HTSUS)),
1624;
*
*
*
*
*
Sections 12.104 through 12.104i also issued
under 19 U.S.C. 2612;
*
*
*
§ 12.104g
*
*
[Amended]
2. In § 12.104g, the table in paragraph
(a) is amended in the entry for Belize by
adding the words ‘‘extended by ‘‘CBP
Dec. 18–02’’ after the words ‘‘CBP Dec.
13–05’’ in the column headed ‘‘Decision
No.’’.
■
Kevin K. McAleenan,
Acting Commissioner, U.S. Customs and
Border Protection.
Approved: February 21, 2018.
Timothy E. Skud,
Deputy Assistant Secretary of the Treasury.
[FR Doc. 2018–03946 Filed 2–26–18; 8:45 am]
BILLING CODE 9111–14–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 864
[Docket No. FDA 2018–N–0339]
Medical Devices; Hematology and
Pathology Devices; Classification of
Lynch Syndrome Test Systems
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying Lynch syndrome test systems
into class II (special controls). The
special controls that apply to the device
type are identified in this order and will
be part of the codified language for the
Lynch syndrome test systems’
classification. We are taking this action
because we have determined that
classifying the device into class II
(special controls) will provide a
reasonable assurance of safety and
effectiveness of the device. We believe
this action will also enhance patients’
access to beneficial innovative devices,
in part by reducing regulatory burdens.
DATES: This order is effective February
27, 2018. The classification was
applicable on October 27, 2017.
FOR FURTHER INFORMATION CONTACT:
Scott McFarland, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 4676, Silver Spring,
SUMMARY:
E:\FR\FM\27FER1.SGM
27FER1
8356
Federal Register / Vol. 83, No. 39 / Tuesday, February 27, 2018 / Rules and Regulations
MD 20993–0002, 301–796–5866,
Scott.McFarland@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified
Lynch syndrome test systems as class II
(special controls), which we have
determined will provide a reasonable
assurance of safety and effectiveness. In
addition, we believe this action will
enhance patients’ access to beneficial
innovation, in part by reducing
regulatory burdens by placing the
device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act to a
predicate device that does not require
premarket approval (see 21 U.S.C.
360c(i)). We determine whether a new
device is substantially equivalent to a
predicate by means of the procedures
for premarket notification under section
510(k) of the FD&C Act and part 807 (21
U.S.C. 360(k) and 21 CFR part 807,
respectively).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA is required to
classify the device by written order
within 120 days. The classification will
be according to the criteria under
section 513(a)(1) of the FD&C Act.
Although the device was automatically
within class III, the De Novo
classification is considered to be the
initial classification of the device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or premarket
approval application in order to market
a substantially equivalent device (see 21
U.S.C. 360c(i), defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the less-burdensome 510(k) process,
when necessary, to market their device.
II. De Novo Classification
On May 31, 2017, Ventana Medical
Systems, Inc. submitted a request for De
Novo classification of the Ventana MMR
IHC Panel. FDA reviewed the request in
order to classify the device under the
criteria for classification set forth in
section 513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on October 27, 2017, FDA
issued an order to the requester
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 864.1866. We
have named the generic type of device
Lynch syndrome test systems, and it is
identified as in vitro diagnostic tests for
use with tumor tissue to identify
previously diagnosed cancer patients at
risk for having Lynch syndrome.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in
table 1.
TABLE 1—LYNCH SYNDROME TEST SYSTEMS RISKS AND MITIGATION MEASURES
Identified risk
Mitigation measures
daltland on DSKBBV9HB2PROD with RULES
False positive test result .................
False negative test result ................
General controls; Special controls (1) and (2) (21 CFR 864.1866(b)(1) and (2)).
General controls; Special control (1) and (2) (21 CFR 864.1866(b)(1) and(2)).
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
VerDate Sep<11>2014
18:04 Feb 26, 2018
Jkt 244001
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k).
PO 00000
Frm 00036
Fmt 4700
Sfmt 4700
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
E:\FR\FM\27FER1.SGM
27FER1
Federal Register / Vol. 83, No. 39 / Tuesday, February 27, 2018 / Rules and Regulations
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations and
guidance. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in the
guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in part 814, subparts A
through E, regarding premarket
approval, have been approved under
OMB control number 0910–0231; the
collections of information in part 807,
subpart E, regarding premarket
notification submissions, have been
approved under OMB control number
0910–0120; and the collections of
information in 21 CFR parts 801 and
809, regarding labeling, have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 864
Blood, Medical devices, Packaging
and containers.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 864 is
amended as follows:
PART 864—HEMATOLOGY AND
PATHOLOGY DEVICES
1. The authority citation for part 864
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 864.1866 to subpart B to read
as follows:
■
daltland on DSKBBV9HB2PROD with RULES
§ 864.1866
Lynch syndrome test systems.
(a) Identification. Lynch syndrome
test systems are in vitro diagnostic tests
for use with tumor tissue to identify
previously diagnosed cancer patients at
risk for having Lynch syndrome.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Premarket notification
submissions must include the following
information, as appropriate:
(i) A detailed description of all test
components, including all provided
reagents, and required but not provided,
ancillary reagents.
VerDate Sep<11>2014
18:04 Feb 26, 2018
Jkt 244001
(ii) A detailed description of
instrumentation and equipment,
including illustrations or photographs of
non-standard equipment or manuals.
(iii) Detailed documentation of the
device software, including, but not
limited to, standalone software
applications and hardware-based
devices that incorporate software.
(iv) A detailed description of quality
controls including appropriate positive
and negative controls that are
recommended or provided.
(v) Detailed specifications for sample
collection, processing, and storage.
(vi) A detailed description of
methodology and assay procedure.
(vii) A description of the assay cut-off
(i.e., the medical decision point between
positive and negative results) or other
relevant criteria that distinguishes
positive and negative results, or ordinal
classes of marker expression, including
the rationale for the chosen cut-off or
other relevant criteria and results
supporting validation of the cut-off.
(viii) Detailed specification of the
criteria for test result interpretation and
reporting.
(ix) Detailed information
demonstrating the performance
characteristics of the device, including:
(A) Data from an appropriate study
demonstrating clinical accuracy using
well-characterized clinical specimens
representative of the intended use
population (i.e., concordance to
Deoxyribonucleic Acid (DNA)
sequencing results of the Lynch
syndrome associated genes or method
comparison to the predicate device
using samples with known alterations in
genes representative of Lynch
syndrome). Pre-specified acceptance
criteria must be provided and followed.
(B) Appropriate device
reproducibility data investigating all
sources of variance (e.g., for distributed
tests, data generated using a minimum
of three sites, of which at least two sites
must be external sites). Each site must
perform testing over a minimum of 5
nonconsecutive days evaluating a
sample panel that spans the claimed
measuring range, and includes the
clinical threshold. Pre-specified
acceptance criteria must be provided
and followed.
(C) Data demonstrating reader
reproducibility, both within-reader and
between-reader, assessed by three
readers over 3 nonconsecutive days at
each site, including a 2 week washout
period between reads, as appropriate.
(D) Device precision data using
clinical samples spanning the
measuring range and controls to
evaluate the within-lot, between-lot,
PO 00000
Frm 00037
Fmt 4700
Sfmt 4700
8357
within-run, between run, and total
variation.
(E) Analytical specificity studies
including as appropriate, western blots,
peptide inhibition, testing in normal
tissues and neoplastic tissues,
interference by endogenous and
exogenous substances, and crossreactivity and cross contamination
testing.
(F) Device analytical sensitivity data
generated by testing an adequate
number of samples from individuals
with the target condition such that
prevalence of the biomarker in the target
population is established.
(G) Device stability data, including
real-time stability and in-use stability,
and stability evaluating various storage
times, temperatures, and freeze-thaw
conditions, as appropriate.
(H) The staining performance criteria
assessed must include overall staining
acceptability, background staining
acceptability, and morphology
acceptability, as appropriate.
(I) Appropriate training requirements
for users, including interpretation
manual, as applicable.
(J) Identification of risk mitigation
elements used by the device, including
a description of all additional
procedures, methods, and practices
incorporated into the instructions for
use that mitigate risks associated with
testing.
(2) The device’s § 809.10(b) of this
chapter compliant labeling must include
a detailed description of the protocol,
including the information described in
paragraphs (b)(1)(i) through (viii) of this
section, as appropriate, and a detailed
description of the performance studies
performed and the summary of the
results, including those that relate to
paragraph (b)(1)(ix) of this section, as
appropriate.
Dated: February 21, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–03924 Filed 2–26–18; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Part 165
[Docket Number USCG–2018–0074]
RIN 1625–AA00
Safety Zone; Wando Terminal Crane
Movement; Charleston, SC
AGENCY:
E:\FR\FM\27FER1.SGM
Coast Guard, DHS.
27FER1
Agencies
[Federal Register Volume 83, Number 39 (Tuesday, February 27, 2018)]
[Rules and Regulations]
[Pages 8355-8357]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2018-03924]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 864
[Docket No. FDA 2018-N-0339]
Medical Devices; Hematology and Pathology Devices; Classification
of Lynch Syndrome Test Systems
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
Lynch syndrome test systems into class II (special controls). The
special controls that apply to the device type are identified in this
order and will be part of the codified language for the Lynch syndrome
test systems' classification. We are taking this action because we have
determined that classifying the device into class II (special controls)
will provide a reasonable assurance of safety and effectiveness of the
device. We believe this action will also enhance patients' access to
beneficial innovative devices, in part by reducing regulatory burdens.
DATES: This order is effective February 27, 2018. The classification
was applicable on October 27, 2017.
FOR FURTHER INFORMATION CONTACT: Scott McFarland, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 4676, Silver Spring,
[[Page 8356]]
MD 20993-0002, 301-796-5866, [email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified Lynch syndrome test systems as
class II (special controls), which we have determined will provide a
reasonable assurance of safety and effectiveness. In addition, we
believe this action will enhance patients' access to beneficial
innovation, in part by reducing regulatory burdens by placing the
device into a lower device class than the automatic class III
assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act to a predicate device that does not require
premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new
device is substantially equivalent to a predicate by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act and part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA is required
to classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically within
class III, the De Novo classification is considered to be the initial
classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application in order to market a substantially
equivalent device (see 21 U.S.C. 360c(i), defining ``substantial
equivalence''). Instead, sponsors can use the less-burdensome 510(k)
process, when necessary, to market their device.
II. De Novo Classification
On May 31, 2017, Ventana Medical Systems, Inc. submitted a request
for De Novo classification of the Ventana MMR IHC Panel. FDA reviewed
the request in order to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on October 27, 2017, FDA issued an order to the
requester classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 864.1866. We have named
the generic type of device Lynch syndrome test systems, and it is
identified as in vitro diagnostic tests for use with tumor tissue to
identify previously diagnosed cancer patients at risk for having Lynch
syndrome.
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Lynch Syndrome Test Systems Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risk Mitigation measures
------------------------------------------------------------------------
False positive test result........ General controls; Special controls
(1) and (2) (21 CFR 864.1866(b)(1)
and (2)).
False negative test result........ General controls; Special control
(1) and (2) (21 CFR 864.1866(b)(1)
and(2)).
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k).
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment
[[Page 8357]]
nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations and guidance. These collections of information are subject
to review by the Office of Management and Budget (OMB) under the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections
of information in the guidance document ``De Novo Classification
Process (Evaluation of Automatic Class III Designation)'' have been
approved under OMB control number 0910-0844; the collections of
information in part 814, subparts A through E, regarding premarket
approval, have been approved under OMB control number 0910-0231; the
collections of information in part 807, subpart E, regarding premarket
notification submissions, have been approved under OMB control number
0910-0120; and the collections of information in 21 CFR parts 801 and
809, regarding labeling, have been approved under OMB control number
0910-0485.
List of Subjects in 21 CFR Part 864
Blood, Medical devices, Packaging and containers.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
864 is amended as follows:
PART 864--HEMATOLOGY AND PATHOLOGY DEVICES
0
1. The authority citation for part 864 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 864.1866 to subpart B to read as follows:
Sec. 864.1866 Lynch syndrome test systems.
(a) Identification. Lynch syndrome test systems are in vitro
diagnostic tests for use with tumor tissue to identify previously
diagnosed cancer patients at risk for having Lynch syndrome.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Premarket notification submissions must include the following
information, as appropriate:
(i) A detailed description of all test components, including all
provided reagents, and required but not provided, ancillary reagents.
(ii) A detailed description of instrumentation and equipment,
including illustrations or photographs of non-standard equipment or
manuals.
(iii) Detailed documentation of the device software, including, but
not limited to, standalone software applications and hardware-based
devices that incorporate software.
(iv) A detailed description of quality controls including
appropriate positive and negative controls that are recommended or
provided.
(v) Detailed specifications for sample collection, processing, and
storage.
(vi) A detailed description of methodology and assay procedure.
(vii) A description of the assay cut-off (i.e., the medical
decision point between positive and negative results) or other relevant
criteria that distinguishes positive and negative results, or ordinal
classes of marker expression, including the rationale for the chosen
cut-off or other relevant criteria and results supporting validation of
the cut-off.
(viii) Detailed specification of the criteria for test result
interpretation and reporting.
(ix) Detailed information demonstrating the performance
characteristics of the device, including:
(A) Data from an appropriate study demonstrating clinical accuracy
using well-characterized clinical specimens representative of the
intended use population (i.e., concordance to Deoxyribonucleic Acid
(DNA) sequencing results of the Lynch syndrome associated genes or
method comparison to the predicate device using samples with known
alterations in genes representative of Lynch syndrome). Pre-specified
acceptance criteria must be provided and followed.
(B) Appropriate device reproducibility data investigating all
sources of variance (e.g., for distributed tests, data generated using
a minimum of three sites, of which at least two sites must be external
sites). Each site must perform testing over a minimum of 5
nonconsecutive days evaluating a sample panel that spans the claimed
measuring range, and includes the clinical threshold. Pre-specified
acceptance criteria must be provided and followed.
(C) Data demonstrating reader reproducibility, both within-reader
and between-reader, assessed by three readers over 3 nonconsecutive
days at each site, including a 2 week washout period between reads, as
appropriate.
(D) Device precision data using clinical samples spanning the
measuring range and controls to evaluate the within-lot, between-lot,
within-run, between run, and total variation.
(E) Analytical specificity studies including as appropriate,
western blots, peptide inhibition, testing in normal tissues and
neoplastic tissues, interference by endogenous and exogenous
substances, and cross-reactivity and cross contamination testing.
(F) Device analytical sensitivity data generated by testing an
adequate number of samples from individuals with the target condition
such that prevalence of the biomarker in the target population is
established.
(G) Device stability data, including real-time stability and in-use
stability, and stability evaluating various storage times,
temperatures, and freeze-thaw conditions, as appropriate.
(H) The staining performance criteria assessed must include overall
staining acceptability, background staining acceptability, and
morphology acceptability, as appropriate.
(I) Appropriate training requirements for users, including
interpretation manual, as applicable.
(J) Identification of risk mitigation elements used by the device,
including a description of all additional procedures, methods, and
practices incorporated into the instructions for use that mitigate
risks associated with testing.
(2) The device's Sec. 809.10(b) of this chapter compliant labeling
must include a detailed description of the protocol, including the
information described in paragraphs (b)(1)(i) through (viii) of this
section, as appropriate, and a detailed description of the performance
studies performed and the summary of the results, including those that
relate to paragraph (b)(1)(ix) of this section, as appropriate.
Dated: February 21, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-03924 Filed 2-26-18; 8:45 am]
BILLING CODE 4164-01-P