Medical Devices; General and Plastic Surgery Devices; Classification of the Surgical Smoke Precipitator, 4141-4143 [2018-01639]
Download as PDF
Federal Register / Vol. 83, No. 20 / Tuesday, January 30, 2018 / Rules and Regulations
rmajette on DSKBCKNHB2PROD with RULES
§ 870.1251 Temporary catheter for embolic
protection during transcatheter intracardiac
procedures.
(a) Identification. This device is a
single use percutaneous catheter system
that has (a) blood filter(s) at the distal
end. This device is indicated for use
while performing transcatheter
intracardiac procedures. The device is
used to filter blood in a manner that
may prevent embolic material
(thrombus/debris) from the transcatheter
intracardiac procedure from traveling
towards the cerebral circulation.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
(i) Simulated-use testing in a
clinically relevant bench anatomic
model to assess the following:
(A) Delivery, deployment, and
retrieval, including quantifying
deployment and retrieval forces, and
procedural time; and
(B) Device compatibility and lack of
interference with the transcatheter
intracardiac procedure and device.
(ii) Tensile strengths of joints and
components, tip flexibility, torque
strength, torque response, and kink
resistance.
(iii) Flow characteristics.
(A) The ability of the filter to not
impede blood flow.
(B) The amount of time the filter can
be deployed in position and/or retrieved
from its location without disrupting
blood flow.
(iv) Characterization and verification
of all dimensions.
(2) Animal testing must demonstrate
that the device performs as intended
under anticipated conditions of use. The
following performance characteristics
must be assessed:
(i) Delivery, deployment, and
retrieval, including quantifying
procedural time.
(ii) Device compatibility and lack of
interference with the transcatheter
intracardiac procedure and device.
(iii) Flow characteristics.
(A) The ability of the filter to not
impede blood flow.
(B) The amount of time the filter can
be deployed in position and/or retrieved
from its location without disrupting
blood flow.
(iv) Gross pathology and
histopathology assessing vascular injury
and downstream embolization.
(3) All patient contacting components
of the device must be demonstrated to
be biocompatible.
VerDate Sep<11>2014
15:14 Jan 29, 2018
Jkt 244001
(4) Performance data must
demonstrate the sterility of the device
components intended to be provided
sterile.
(5) Performance data must support the
shelf life of the device by demonstrating
continued sterility, package integrity,
and device functionality over the
identified shelf life.
(6) Labeling for the device must
include:
(i) Instructions for use;
(ii) Compatible transcatheter
intracardiac procedure devices;
(iii) A detailed summary of the
clinical testing conducted; and
(iv) A shelf life and storage
conditions.
(7) Clinical performance testing must
demonstrate:
(i) The ability to safely deliver,
deploy, and remove the device;
(ii) The ability of the device to filter
embolic material while not impeding
blood flow;
(iii) Secure positioning and stability
of the position throughout the
transcatheter intracardiac procedure;
and
(iv) Evaluation of all adverse events
including death, stroke, and vascular
injury.
Dated: January 24, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–01638 Filed 1–29–18; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 878
[Docket No. FDA–2017–N–6598]
Medical Devices; General and Plastic
Surgery Devices; Classification of the
Surgical Smoke Precipitator
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the surgical smoke
precipitator into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the surgical smoke
precipitator’s classification. We are
taking this action because we have
determined that classifying the device
into class II (special controls) will
provide a reasonable assurance of safety
SUMMARY:
PO 00000
Frm 00011
Fmt 4700
Sfmt 4700
4141
and effectiveness of the device. We
believe this action will also enhance
patients’ access to beneficial innovative
devices, in part by reducing regulatory
burdens.
DATES: This order is effective January
30, 2018. The classification was
applicable on December 20, 2016.
FOR FURTHER INFORMATION CONTACT:
Steven Elliott, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2565, Silver Spring,
MD 20993–0002, 301–796–5285,
steven.elliott@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the
surgical smoke precipitator as class II
(special controls), which we have
determined will provide a reasonable
assurance of safety and effectiveness. In
addition, we believe this action will
enhance patients’ access to beneficial
innovation, in part by reducing
regulatory burdens by placing the
device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
by means of the procedures for
premarket notification under section
510(k) of the FD&C Act and part 807 (21
U.S.C. 360(k) and 21 CFR part 807,
respectively).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
E:\FR\FM\30JAR1.SGM
30JAR1
4142
Federal Register / Vol. 83, No. 20 / Tuesday, January 30, 2018 / Rules and Regulations
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act. Although the device was
automatically placed within class III,
the De Novo classification is considered
to be the initial classification of the
device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or premarket
approval application in order to market
a substantially equivalent device (see 21
U.S.C. 360c(i), defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the less-burdensome 510(k) process,
when necessary, to market their device.
II. De Novo Classification
On May 26, 2015, Alesi Surgical
submitted a request for De Novo
classification of the UltravisionTM
Visual Field Clearing System. FDA
reviewed the request in order to classify
the device under the criteria for
classification set forth in section
513(a)(1) of the FD&C Act.
We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to the general
controls, will provide reasonable
assurance of the safety and effectiveness
of the device.
Therefore, on December 20, 2016,
FDA issued an order to the requester
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 878.5050. We
have named the generic type of device
surgical smoke precipitator, and it is
identified as a prescription device
intended for clearance of the visual field
by precipitation of surgical smoke and
other aerosolized particulate matter
created during laparoscopic surgery.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in table
1.
TABLE 1—SURGICAL SMOKE PRECIPITATOR RISKS AND MITIGATION MEASURES
Mitigation measures
Electrical shock .........................................................................................
Electromagnetic interference with other devices .....................................
Infection ....................................................................................................
Adverse tissue reaction ............................................................................
Tissue injury .............................................................................................
rmajette on DSKBCKNHB2PROD with RULES
Identified risks
Electrical safety testing and Labeling.
Electromagnetic compatibility testing and Labeling.
Sterilization validation, Shelf-life validation, and Labeling.
Biocompatibility evaluation.
Animal testing; Software verification, validation, and hazard analysis;
and Labeling.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. For a device
to fall within this classification, and
thus avoid automatic classification in
class III, it would have to comply with
the special controls named in this final
order. The necessary special controls
appear in the regulation codified by this
order. This device is subject to
premarket notification requirements
under section 510(k) of the FD&C Act.
At the time of classification, surgical
smoke precipitators are for prescription
use only. Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352(f)(1)) and 21
CFR 801.5, as long as the conditions of
VerDate Sep<11>2014
15:14 Jan 29, 2018
Jkt 244001
21 CFR 801.109 are met (referring to 21
U.S.C. 352(f)(1)).
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
PO 00000
Frm 00012
Fmt 4700
Sfmt 4700
the guidance document ‘‘De Novo
Classification Process (Evaluation of
Automatic Class III Designation)’’ have
been approved under OMB control
number 0910–0844; the collections of
information in 21 CFR part 814,
subparts A through E, regarding
premarket approval, have been
approved under OMB control number
0910–0231; the collections of
information part 807, subpart E,
regarding premarket notification
submissions, have been approved under
OMB control number 0910–0120, and
the collections of information in 21 CFR
part 801, regarding labeling, have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 878
Medical devices.
E:\FR\FM\30JAR1.SGM
30JAR1
Federal Register / Vol. 83, No. 20 / Tuesday, January 30, 2018 / Rules and Regulations
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 878 is
amended as follows:
PART 878—GENERAL AND PLASTIC
SURGERY DEVICES
(iii) A shelf life.
PART 51—PASSPORTS
Dated: January 24, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018–01639 Filed 1–29–18; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF STATE
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
22 CFR Part 51
2. Add § 878.5050 to subpart F to read
as follows:
[Public Notice: 9867]
rmajette on DSKBCKNHB2PROD with RULES
§ 878.5050
Surgical smoke precipitator.
15:14 Jan 29, 2018
Jkt 244001
[FR Doc. 2018–01708 Filed 1–29–18; 8:45 am]
BILLING CODE 4710–06–P
RIN 1400–AE01
(a) Identification. A surgical smoke
precipitator is a prescription device
intended for clearance of the visual field
by precipitation of surgical smoke and
other aerosolized particulate matter
created during laparoscopic surgery.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Adverse tissue reaction must be
mitigated through the following:
(i) Chemical characterization and
toxicological risk assessment of the
treated surgical smoke.
(ii) Demonstration that the elements
of the device that may contact the
patient are biocompatible.
(2) Electrical safety and
electromagnetic compatibility testing
must demonstrate that the device
performs as intended.
(3) Software verification, validation,
and hazard analysis must be performed.
(4) Performance data must
demonstrate the sterility of the patient
contacting components of the device.
(5) Performance data must support the
shelf life of the sterile components of
the device by demonstrating continued
functionality, sterility, and package
integrity over the identified shelf life.
(6) Animal simulated-use testing must
demonstrate that the device performs as
intended under anticipated conditions
of use. The following performance
characteristics must be tested:
(i) Device must be demonstrated to be
effectively inserted, positioned, and
removed from the site of use.
(ii) Device must be demonstrated to
precipitate surgical smoke particulates
to clear the visual field for laparoscopic
surgeries.
(iii) Device must be demonstrated to
be non-damaging to the site of use and
animal subject.
(7) Labeling must identify the
following:
(i) Detailed instructions for use.
(ii) Electrical safety and
electromagnetic compatibility
information.
VerDate Sep<11>2014
Accordingly, the interim final rule
amending 22 CFR part 51, which was
published at 81 FR 67156 on September
30, 2016, is adopted as a final rule
without change.
■
Carl C. Risch,
Assistant Secretary, Consular Affairs.
1. The authority citation for part 878
continues to read as follows:
■
■
4143
DEPARTMENT OF HOMELAND
SECURITY
Passports: Service Passports
Coast Guard
Department of State.
Final rule.
AGENCY:
33 CFR Part 117
ACTION:
[Docket No. USCG–2018–0025]
This rule finalizes the interim
final rule from the Department of State
that established a new service passport,
which may be approved for certain nonpersonal services contractors who travel
abroad in support of and pursuant to a
contract with the U.S. government. The
Department received no public
comments in response to the rule.
DATES: Effective January 30, 2018.
FOR FURTHER INFORMATION CONTACT:
Sitara Kedilaya, Attorney-Adviser,
PassportRules@state.gov, (202) 485–
6500.
SUMMARY:
On
September 30, 2016, the Department
published an interim final rule
amending 22 CFR part 51, to create a
‘‘service passport’’ that would be used
by non-personal services contractors to
carry out critical security, maintenance
and other functions on behalf of the U.S.
government. As noted in the interim
final rule, the Department estimates that
this rulemaking will affect
approximately 1,000 non-personal
services contractors per year. Further
information concerning the rationale for
this rule can be found in the interim
final rule.
The Department provided 60 days for
the public to comment on this rule. This
period expired on November 29, 2016.
The Department received no public
comments.
The Regulatory Findings included
with the interim final rule are
incorporated herein. This rule is not an
E.O. 13771 regulatory action because it
is not significant under E.O. 12866.
SUPPLEMENTARY INFORMATION:
List of Subjects in 22 CFR Part 51
Administrative practice and
procedure, Drug traffic control,
Passports and visas, Reporting and
recordkeeping requirements.
PO 00000
Frm 00013
Fmt 4700
Sfmt 4700
Drawbridge Operation Regulation;
Willamette River, Portland, OR
Coast Guard, DHS.
Notice of deviation from
drawbridge regulation.
AGENCY:
ACTION:
The Coast Guard has issued a
temporary deviation from the operating
schedule that governs the Broadway
Bridge across the Willamette River, mile
11.7, at Portland, OR. The deviation is
necessary to make adjustments to new
equipment. This deviation allows the
bridge to operate the double bascule
span one side at a time, single leaf.
DATES: This deviation is effective
without actual notice from January 30,
2018, to 11:59 p.m. on February 23,
2018. For the purposes of enforcement,
actual notice will be used from 1 a.m.
on January 27, 2018, through January
30, 2018.
ADDRESSES: The docket for this
deviation, USCG–2018–0025, is
available at https://www.regulations.gov.
Type the docket number in the
‘‘SEARCH’’ box and click ‘‘SEARCH.’’
Click on Open Docket Folder on the line
associated with this deviation.
FOR FURTHER INFORMATION CONTACT: If
you have questions on this temporary
deviation, call or email Mr. Steven
Fischer, Bridge Administrator,
Thirteenth Coast Guard District;
telephone 206–220–7282, email
Steven.M.Fischer@uscg.mil.
SUPPLEMENTARY INFORMATION:
Multnomah County, the bridge owner,
requested the Broadway Bridge be
authorized to open half the span in
single leaf mode to make adjustments to
newly installed equipment. The
Broadway Bridge crosses the Willamette
River at mile 11.7, and provides 90 feet
of vertical clearance above Columbia
SUMMARY:
E:\FR\FM\30JAR1.SGM
30JAR1
Agencies
[Federal Register Volume 83, Number 20 (Tuesday, January 30, 2018)]
[Rules and Regulations]
[Pages 4141-4143]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2018-01639]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 878
[Docket No. FDA-2017-N-6598]
Medical Devices; General and Plastic Surgery Devices;
Classification of the Surgical Smoke Precipitator
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the surgical smoke precipitator into class II (special controls). The
special controls that apply to the device type are identified in this
order and will be part of the codified language for the surgical smoke
precipitator's classification. We are taking this action because we
have determined that classifying the device into class II (special
controls) will provide a reasonable assurance of safety and
effectiveness of the device. We believe this action will also enhance
patients' access to beneficial innovative devices, in part by reducing
regulatory burdens.
DATES: This order is effective January 30, 2018. The classification was
applicable on December 20, 2016.
FOR FURTHER INFORMATION CONTACT: Steven Elliott, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2565, Silver Spring, MD 20993-0002, 301-796-5285,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the surgical smoke precipitator as
class II (special controls), which we have determined will provide a
reasonable assurance of safety and effectiveness. In addition, we
believe this action will enhance patients' access to beneficial
innovation, in part by reducing regulatory burdens by placing the
device into a lower device class than the automatic class III
assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate by means of the procedures
for premarket notification under section 510(k) of the FD&C Act and
part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act
[[Page 4142]]
of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application in order to market a substantially
equivalent device (see 21 U.S.C. 360c(i), defining ``substantial
equivalence''). Instead, sponsors can use the less-burdensome 510(k)
process, when necessary, to market their device.
II. De Novo Classification
On May 26, 2015, Alesi Surgical submitted a request for De Novo
classification of the UltravisionTM Visual Field Clearing
System. FDA reviewed the request in order to classify the device under
the criteria for classification set forth in section 513(a)(1) of the
FD&C Act.
We classify devices into class II if general controls by themselves
are insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to the
general controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on December 20, 2016, FDA issued an order to the
requester classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 878.5050. We have named
the generic type of device surgical smoke precipitator, and it is
identified as a prescription device intended for clearance of the
visual field by precipitation of surgical smoke and other aerosolized
particulate matter created during laparoscopic surgery.
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Surgical Smoke Precipitator Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Electrical shock....................... Electrical safety testing and
Labeling.
Electromagnetic interference with other Electromagnetic compatibility
devices. testing and Labeling.
Infection.............................. Sterilization validation, Shelf-
life validation, and Labeling.
Adverse tissue reaction................ Biocompatibility evaluation.
Tissue injury.......................... Animal testing; Software
verification, validation, and
hazard analysis; and Labeling.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. For a device to fall within this
classification, and thus avoid automatic classification in class III,
it would have to comply with the special controls named in this final
order. The necessary special controls appear in the regulation codified
by this order. This device is subject to premarket notification
requirements under section 510(k) of the FD&C Act.
At the time of classification, surgical smoke precipitators are for
prescription use only. Prescription devices are exempt from the
requirement for adequate directions for use for the layperson under
section 502(f)(1) of the FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR
801.5, as long as the conditions of 21 CFR 801.109 are met (referring
to 21 U.S.C. 352(f)(1)).
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in
the guidance document ``De Novo Classification Process (Evaluation of
Automatic Class III Designation)'' have been approved under OMB control
number 0910-0844; the collections of information in 21 CFR part 814,
subparts A through E, regarding premarket approval, have been approved
under OMB control number 0910-0231; the collections of information part
807, subpart E, regarding premarket notification submissions, have been
approved under OMB control number 0910-0120, and the collections of
information in 21 CFR part 801, regarding labeling, have been approved
under OMB control number 0910-0485.
List of Subjects in 21 CFR Part 878
Medical devices.
[[Page 4143]]
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
878 is amended as follows:
PART 878--GENERAL AND PLASTIC SURGERY DEVICES
0
1. The authority citation for part 878 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 878.5050 to subpart F to read as follows:
Sec. 878.5050 Surgical smoke precipitator.
(a) Identification. A surgical smoke precipitator is a prescription
device intended for clearance of the visual field by precipitation of
surgical smoke and other aerosolized particulate matter created during
laparoscopic surgery.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Adverse tissue reaction must be mitigated through the
following:
(i) Chemical characterization and toxicological risk assessment of
the treated surgical smoke.
(ii) Demonstration that the elements of the device that may contact
the patient are biocompatible.
(2) Electrical safety and electromagnetic compatibility testing
must demonstrate that the device performs as intended.
(3) Software verification, validation, and hazard analysis must be
performed.
(4) Performance data must demonstrate the sterility of the patient
contacting components of the device.
(5) Performance data must support the shelf life of the sterile
components of the device by demonstrating continued functionality,
sterility, and package integrity over the identified shelf life.
(6) Animal simulated-use testing must demonstrate that the device
performs as intended under anticipated conditions of use. The following
performance characteristics must be tested:
(i) Device must be demonstrated to be effectively inserted,
positioned, and removed from the site of use.
(ii) Device must be demonstrated to precipitate surgical smoke
particulates to clear the visual field for laparoscopic surgeries.
(iii) Device must be demonstrated to be non-damaging to the site of
use and animal subject.
(7) Labeling must identify the following:
(i) Detailed instructions for use.
(ii) Electrical safety and electromagnetic compatibility
information.
(iii) A shelf life.
Dated: January 24, 2018.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2018-01639 Filed 1-29-18; 8:45 am]
BILLING CODE 4164-01-P