General Hospital and Personal Use Devices; Reclassification of Sharps Needle Destruction Device, 51585-51589 [2017-24191]
Download as PDF
Federal Register / Vol. 82, No. 214 / Tuesday, November 7, 2017 / Proposed Rules
(Piper) Service Bulletin (SB) No. 1304, dated
August 23, 2017.
(h) Install Inspection Access Panels
If it is determined that no inspection access
panels are present during the inspection
required in paragraph (g) of this AD, within
the next 100 hours TIS after the effective date
of this AD or within the next 12 months after
the effective date of this AD, whichever
occurs first install inspection access panels
on the lower skin of the left wing and the
right wing using Piper SB No. 1304, dated
August 23, 2017.
(i) Inspect for Corrosion
Within the next 100 hours TIS after the
effective date of this AD or within the next
12 months after the effective date of this AD,
whichever occurs first, inspect the left and
the right main wing spar for any evidence of
corrosion using Part I of the Instructions
section of Piper SB No. 1304, dated August
23, 2017.
nshattuck on DSK9F9SC42PROD with PROPOSALS
(j) Corrective Actions
Before further flight after the inspection
required in paragraph (i) of this AD, if
evidence of corrosion is found, take all
necessary corrective actions to remove the
corrosion using Part I of the Instructions
section of Piper SB No. 1304, dated August
23, 2017, and/or make all necessary repairs
using Part II of the Instructions section of
Piper SB No. 1304, dated August 23, 2017.
(k) Alternative Methods of Compliance
(AMOCs)
(1) The Manager, Atlanta ACO Branch,
FAA, has the authority to approve AMOCs
for this AD, if requested using the procedures
found in 14 CFR 39.19. In accordance with
14 CFR 39.19, send your request to your
principal inspector or local Flight Standards
District Office, as appropriate. If sending
information directly to the manager of the
certification office, send it to the attention of
the person identified in paragraph (l)(1) of
this AD.
(2) Before using any approved AMOC,
notify your appropriate principal inspector,
or lacking a principal inspector, the manager
of the local flight standards district office/
certificate holding district office.
(3) For service information that contains
steps that are labeled as Required for
Compliance (RC), the provisions of
paragraphs (g) through (j) of this AD apply.
(i) The steps labeled as RC, including
substeps under an RC step and any figures
identified in an RC step, must be done to
comply with the AD. An AMOC is required
for any deviations to RC steps, including
substeps and identified figures.
(ii) Steps not labeled as RC may be
deviated from using accepted methods in
accordance with the operator’s maintenance
or inspection program without obtaining
approval of an AMOC, provided the RC steps,
including substeps and identified figures, can
still be done as specified, and the airplane
can be put back in an airworthy condition.
(l) Related Information
(1) For more information about this AD,
contact Dan McCully, Aerospace Engineer,
FAA, Atlanta ACO Branch, 1701 Columbia
VerDate Sep<11>2014
15:06 Nov 06, 2017
Jkt 244001
Avenue, College Park, Georgia 30337;
telephone: (404) 474–5548; fax: (404) 474–
5606; email: william.mccully@faa.gov.
(2) For service information identified in
this AD, contact Piper Aircraft, Inc., 2926
Piper Drive, Vero Beach, Florida 32960;
telephone: (772) 567–4361; Internet:
www.piper.com. You may review this
referenced service information at the FAA,
Policy and Innovation Division, 901 Locust,
Kansas City, Missouri 64106. For information
on the availability of this material at the
FAA, call (816) 329–4148.
Issued in Kansas City, Missouri, on
October 30, 2017.
Melvin J. Johnson,
Acting Deputy Director, Policy & Innovation
Division, Aircraft Certification Service.
[FR Doc. 2017–24083 Filed 11–6–17; 8:45 am]
BILLING CODE 4910–13–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA–2017–N–6216]
General Hospital and Personal Use
Devices; Reclassification of Sharps
Needle Destruction Device
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed order.
The Food and Drug
Administration (FDA) is issuing this
proposed order to reclassify the needle
destruction device, renaming the device
to ‘‘sharps needle destruction device,’’ a
postamendments class III device
(regulated under product code MTV),
into class II (special controls), subject to
premarket notification. FDA is also
identifying the proposed special
controls that the Agency believes are
necessary to provide a reasonable
assurance of safety and effectiveness of
the device. FDA is proposing this
reclassification on its own initiative
based on new information. If finalized,
this order will reclassify these types of
devices from class III to class II and
reduce regulatory burdens on industry
as these types of devices will no longer
be required to submit a premarket
approval application (PMA) but can
instead submit a less burdensome
premarket notification (510(k)) before
marketing their device.
DATES: Submit either electronic or
written comments on the proposed
order by January 8, 2018. Please see
section XI of this document for the
proposed effective date when the new
requirements apply and for the
SUMMARY:
PO 00000
Frm 00004
Fmt 4702
Sfmt 4702
51585
proposed effective date of a final order
based on this proposed order.
ADDRESSES: You may submit comments
as follows. Please note that late,
untimely filed comments will not be
considered. Electronic comments must
be submitted on or before January 8,
2018. The https://www.regulations.gov
electronic filing system will accept
comments until midnight Eastern Time
at the end of January 8, 2018. Comments
received by mail/hand delivery/courier
(for written/paper submissions) will be
considered timely if they are
postmarked or the delivery service
acceptance receipt is on or before that
date.
Electronic Submissions
Submit electronic comments in the
following way:
• Federal Rulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand delivery/Courier (for
written/paper submissions): Dockets
Management Staff (HFA–305), Food and
Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Dockets Management
Staff, FDA will post your comment, as
well as any attachments, except for
information submitted, marked and
identified, as confidential, if submitted
as detailed in ‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2017–N–6216 for ‘‘General Hospital and
Personal Use Devices; Reclassification
E:\FR\FM\07NOP1.SGM
07NOP1
51586
Federal Register / Vol. 82, No. 214 / Tuesday, November 7, 2017 / Proposed Rules
nshattuck on DSK9F9SC42PROD with PROPOSALS
of Sharps Needle Destruction Device.’’
Received comments, those filed in a
timely manner (see ADDRESSES), will be
placed in the docket and, except for
those submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Dockets Management Staff between 9
a.m. and 4 p.m., Monday through
Friday.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION.’’ The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Dockets Management
Staff. If you do not wish your name and
contact information to be made publicly
available, you can provide this
information on the cover sheet and not
in the body of your comments and you
must identify this information as
‘‘confidential.’’ Any information marked
as ‘‘confidential’’ will not be disclosed
except in accordance with 21 CFR 10.20
and other applicable disclosure law. For
more information about FDA’s posting
of comments to public dockets, see 80
FR 56469, September 18, 2015, or access
the information at: https://www.gpo.gov
/fdsys/pkg/FR-2015-09-18/pdf/201523389.pdf.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Dockets Management
Staff, 5630 Fishers Lane, Rm. 1061,
Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Christopher K. Dugard, Center for
Devices and Radiological Health, Food
and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 2561,
Silver Spring, MD 20993, 240–402–
6031, christopher.dugard@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
VerDate Sep<11>2014
15:06 Nov 06, 2017
Jkt 244001
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act), as amended,
establishes a comprehensive system for
the regulation of medical devices
intended for human use. Section 513 of
the FD&C Act (21 U.S.C. 360c)
established three categories (classes) of
devices, reflecting the regulatory
controls needed to provide reasonable
assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices) are
automatically classified by section
513(f)(1) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II, or FDA issues an order finding
the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and 21 CFR
part 807.
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified into class I or class
II under section 513(f)(3) of the FD&C
Act. Section 513(f)(3) of the FD&C Act
provides that FDA acting by order can
reclassify the device into class I or class
II on its own initiative, or in response
to a petition from the manufacturer or
importer of the device. To change the
classification of the device, the
proposed new class must have sufficient
regulatory controls to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use.
Reevaluation of the data previously
before the Agency is an appropriate
basis for subsequent action where the
reevaluation is made in light of newly
available regulatory authority (see Bell
v. Goddard, 366 F.2d 177, 181 (7th Cir.
1966); Ethicon, Inc. v. FDA, 762 F.
Supp. 382, 388–391 (D.D.C. 1991)) or in
light of changes in ‘‘medical science’’
(Upjohn v. Finch, 422 F.2d 944, 951 (6th
Cir. 1970)). Whether data before the
Agency are old or new, the ‘‘new
information’’ to support reclassification
under 513(f)(3) must be ‘‘valid scientific
PO 00000
Frm 00005
Fmt 4702
Sfmt 4702
evidence’’, as defined in section
513(a)(3) of the FD&C Act and 21 CFR
860.7(c)(2). (See, e.g., General Medical
Co. v. FDA, 770 F.2d 214 (D.C. Cir.
1985); Contact Lens Assoc. v. FDA, 766
F.2d 592 (D.C. Cir.1985), cert. denied,
474 U.S. 1062 (1986)).
FDA relies upon ‘‘valid scientific
evidence’’ in the classification process
to determine the level of regulation for
devices. To be considered in the
reclassification process, the ‘‘valid
scientific evidence’’ upon which the
Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA (see
section 520(c) of the FD&C Act (21
U.S.C. 360j(c)). Section 520(h)(4) of the
FD&C Act provides that FDA may use,
for reclassification of a device, certain
information in a PMA 6 years after the
application has been approved. This
includes information from clinical and
preclinical tests or studies that
demonstrate the safety or effectiveness
of the device, but does not include
descriptions of methods of manufacture
or product composition and other trade
secrets.
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the 510(k) premarket
notification requirements, if the Agency
determines that premarket notification
is not necessary to reasonably assure the
safety and effectiveness of the device.
II. Regulatory History of the Devices
On February 3, 1994, FDA issued a
Memorandum to manufacturers and
initial distributors of sharps containers
and destroyers used by health care
manufacturers to clarify the regulatory
status of sharps destroyer devices (Ref.
1).
On March 6, 1997, FDA approved its
first needle destruction device through
its PMA process under section 515 of
the FD&C Act (21 U.S.C. 360e). In the
June 11, 1997, Federal Register notice
(62 FR 31831), FDA announced a PMA
approval order for Millenium Medical
Supply’s Incorporated Needle-EaseTM
25011 device and the availability of the
Summary of Safety and Effectiveness
Data for the Device (SSED) (Ref. 2). As
of the date of issuance of this proposed
order, FDA has approved 18 original
PMAs for this device type.2
1 FDA approved a modified needle destruction
device on February 11, 1998. The device, as
modified, is marketed under the trade name NeedleEase®3500.
2 See PMA database for original PMAs regulated
under the product code MTV: https://www.access
data.fda.gov/scripts/cdrh/cfdocs/cfPMA/pma.cfm.
E:\FR\FM\07NOP1.SGM
07NOP1
Federal Register / Vol. 82, No. 214 / Tuesday, November 7, 2017 / Proposed Rules
On March 2, 2001, FDA finalized
guidance entitled, ‘‘Premarket Approval
Applications (PMA) for Sharps Needle
Destruction Devices,’’ describing the
Agency’s recommendations for
information to include in PMA
applications for sharps needle
destruction devices intended for use in
health care settings (Ref. 3).
nshattuck on DSK9F9SC42PROD with PROPOSALS
III. Device Description
A sharps needle destruction device is
a postamendments device classified into
class III under section 513(f)(1) of the
FD&C Act. A sharps needle destruction
device is a prescription device intended
for home use or in professional health
care facilities to destroy sharps or
needles used for medical purposes by
incineration or mechanical means.
Sharps needle destruction devices are
typically electrical devices that can
destruct sharps and/or needles in a
variety of methods (grinding,
incinerating, etc.) that can be either
portable or stationary. Some of these
devices may also employ software to
provide the user with greater control.
Please note these devices were
originally identified as needle
destruction devices (product code MTV)
in FDA SSEDs and product code
database; however, FDA believes the
identification of sharps needle
destruction device more accurately
describes this device type as it can be
used to destroy devices other than
needles (e.g., sharps).
IV. Proposed Reclassification
As part of the Center for Devices and
Radiological Health’s 2014–2015
strategic priority ‘‘Strike the Right
Balance Between Premarket and
Postmarket Data Collection,’’ a
retrospective review of class III devices
subject to PMA was completed to
determine whether or not, based on our
current understanding of the
technology, reclassification may be
appropriate. On August 8, 2016, FDA
published a document in the Federal
Register entitled ‘‘Retrospective Review
of Premarket Approval Application
Devices; Striking the Balance Between
Premarket and Postmarket Data
Collection’’ in which FDA announced
plans to reclassify sharps needle
destruction devices identified with the
MTV product code from class III to class
II (81 FR 52445). FDA has found that
sufficient information exists to establish
special controls that, together with
general controls, can provide a
reasonable assurance of safety and
effectiveness for sharps needle
destruction devices.
In accordance with section 513(f)(3) of
the FD&C Act and 21 CFR part 860,
VerDate Sep<11>2014
15:06 Nov 06, 2017
Jkt 244001
subpart C, FDA is proposing to
reclassify this postamendments class III
device into class II. FDA believes that
there is sufficient information available
to FDA through FDA’s accumulated
experience with these devices from
review submissions, peer-reviewed
literature, and knowledge of similar
devices to establish special controls that
effectively mitigate the risks to health
identified in section V. Absent the
special controls identified in this
proposed order, general controls
applicable to the device are insufficient
to provide reasonable assurance of the
safety and effectiveness of the device.
FDA is proposing to create a separate
classification regulation for sharps
needle destruction devices that will be
reclassified from class III to II. Under
this proposed order, if finalized, the
sharps needle destruction devices will
be identified as a prescription device.
As such, the prescription device must
satisfy prescription labeling
requirements (see § 801.109 (21 CFR
801.109), Prescription devices).
Prescription devices are exempt from
the requirement for adequate directions
for use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and § 801.5 (21 CFR 801.5), as
long as the conditions of § 801.109 are
met. In this proposed order, if finalized,
the Agency has identified the special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls, will provide a reasonable
assurance of the safety and effectiveness
for sharps needle destruction devices.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
is necessary for sharps needle
destruction devices to provide
reasonable assurance of the safety and
effectiveness. Therefore, the Agency
does not intend to exempt these
proposed class II devices from 510(k)
requirements. Persons who intend to
market this type of device must submit
to FDA a 510(k) and receive clearance
prior to marketing the device.
V. Risks to Health
After considering the information
available to FDA through review
submissions, peer-reviewed literature,
and knowledge of similar devices, FDA
determined the probable risks to health
associated with the use of sharps needle
destruction devices are as follows:
PO 00000
Frm 00006
Fmt 4702
Sfmt 4702
51587
• Patient/user exposure to
environmental contaminants.
Destroying a used sharp or needle may
generate hazardous emissions from the
device that may result in infection or
respiratory problems for the patient
and/or user. Contamination of the
patient environment can occur through
emission of toxic fumes or infectious
aerosol when the device destroys a
sharp by incineration or mechanical
means, and may result from device
malfunction (mechanical and/or
software). The device may also become
contaminated through regular usage and
may cause cross-contamination.
• Patient/user burns as a result of
excessive heat discharge or spark
formation. Excessive heat or sparks may
be generated and discharged from the
device during destruction of sharps that
may burn the user.
• Electromagnetic interference. While
in operation, the device may interfere
with other electrically powered devices,
causing them to malfunction.
• Electrical shock. While in
operation, the device may discharge
electricity that could shock the user.
• Sharps injury. Incompletely
destroyed sharps, physical device
instability, device malfunctions, or use
error may pose a risk for a sharps injury
to the user.
VI. Summary of Reasons for
Reclassification
FDA believes that the sharps needle
destruction devices intended for home
use or in professional health care
facilities to reduce the incidence of
needlesticks by destroying sharps and/
or needles in a variety of methods
(grinding, incinerating, etc.) should be
reclassified from class III to class II in
light of new information about the
effectiveness of these devices. There is
sufficient information to establish
special controls for sharps needle
destruction devices, in addition to
general controls, which can provide
reasonable assurance of safety and
effectiveness of the device, as general
controls themselves are insufficient to
provide reasonable assurance of its
safety and effectiveness. FDA believes
that the risks to health associated with
sharps needle destruction devices
intended for home use or in professional
health care facilities to reduce the
incidence of needlesticks can be
mitigated with special controls and that
these mitigations will provide a
reasonable assurance of its safety and
effectiveness.
Based on a reconsideration of the
available information and data, FDA
believes that there is valid scientific
evidence of effectiveness for sharps
E:\FR\FM\07NOP1.SGM
07NOP1
51588
Federal Register / Vol. 82, No. 214 / Tuesday, November 7, 2017 / Proposed Rules
needle destruction devices to reduce the
incidence of needlesticks.
VII. Summary of Data Upon Which the
Reclassification Is Based
FDA believes that the identified
special controls, in addition to general
controls, are necessary to provide
reasonable assurance of safety and
effectiveness of these devices. Taking
into account the probable health
benefits of the use of the device and the
nature and known incidence of the risks
of the device, FDA, on its own initiative,
is proposing to reclassify this
postamendments class III device into
class II. FDA has considered and
analyzed the following information: An
inclusive search of the Agency’s
Manufacturer and User Facility Device
Experience (MAUDE) database, which
shows no adverse events for sharps
needle destruction devices; data
contained in PMAs approved 6 or more
years before the date of this proposal
(reviewed under section 520(h)(4) of the
FD&C Act, also known as the 6-year
rule); a review of sharps containers
regulated under 21 CFR 880.5570,
which have similar intended uses, but
different technology, and are currently
regulated as class II devices; and one
relevant article found from a literature
search that discussed the benefits and
the probable risks of these devices (Ref.
4).
nshattuck on DSK9F9SC42PROD with PROPOSALS
VIII. Proposed Special Controls
FDA believes that the following
special controls, together with general
controls, are sufficient to mitigate the
risks to health described in section V
and provide a reasonable assurance of
safety and effectiveness for sharps
needle destruction devices.
• Performance testing will
demonstrate:
Æ The device’s ability to contain or
ventilate aerosols or fumes from device
operation that may result in
environmental contamination and crosscontamination. Performance testing will
demonstrate that harmful fumes, such as
ozone, are not emitted by the device
during destruction of sharps needles.
Æ Excessive heat or sparks are not
generated during device operation that
may injure users or patients through
characterization of the heat dissipation
profile from the heat source to the
enclosure surface, and the point of
contact between the held syringe and
the user. Performance testing will
ensure the heat generated through
normal operation of the device will not
harm users or patients or affect circuit
performance and useful life of the
device.
VerDate Sep<11>2014
15:06 Nov 06, 2017
Jkt 244001
Æ Complete destruction of sharps
intended to be destructed to mitigate
user injuries from incomplete sharps
destruction by conducting performance
testing such as simulated use
demonstrating complete destruction of
the sharps and/or needles intended to
be destroyed.
Æ Mitigation of injuries from device
instability through characterization of
the vibrations and movement generated
by the device to ensure device stability
in the use environment.
• Validation of cleaning and
disinfection instructions to demonstrate
that the device can be safely and
effectively reprocessed after use to
minimize the risk of patient/user crosscontamination.
• Performance testing ensures
electromagnetic compatibility with
other devices under conditions which
are consistent with the intended
environment of device use.
• Electrical safety testing ensures the
risk of shock to the patient/user is
minimized.
• Software hazard analysis, as well as
software verification and validation,
ensures that software performs as
intended and potential software
malfunctions do not impact the
performance of the device.
• Labeling to ensure proper use of the
device, including warnings of the
generation of excessive heat, potential
for needle stick injuries, instructions for
reprocessing, and instructions for
installation (e.g., on a stable surface,
adequate ventilation).
Table 1 shows how FDA believes
these special controls will mitigate each
risk to health described in section V.
TABLE 1—RISKS TO HEALTH AND MITIGATION MEASURES FOR SHARPS
NEEDLE DESTRUCTION DEVICES—
Continued
Identified risk to
health
Mitigation measures
Sharps injury .............
Performance testing.
Software verification,
validation, and hazard analysis.
Labeling.
In addition, FDA is proposing to limit
these devices to prescription use under
§ 801.109. Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act and § 801.5, as long as the
conditions of § 801.109 are met
(referring to 21 U.S.C. 352(f)(1)). Under
21 CFR 807.81, the device would
continue to be subject to 510(k)
notification requirements. FDA does not
believe that clinical data is necessary to
mitigate the identified risks to health for
sharps needle destruction devices. FDA
may request clinical data to evaluate
substantial equivalence when a
manufacturer includes new indications
for use, such as indications for disease
prevention or organism destruction.
This reclassification order and the
identified special controls, if finalized,
would provide sufficient detail
regarding FDA’s requirements to
reasonably assure safety and
effectiveness of sharps needle
destruction devices. FDA intends to
withdraw the final guidance entitled,
‘‘Premarket Approval Applications
(PMA) for Sharps Needle Destruction
Devices; Final Guidance for Industry
and FDA’’ issued in 2001 upon
finalization of this proposed
TABLE 1—RISKS TO HEALTH AND MITI- reclassification order (Ref. 3).
GATION MEASURES FOR SHARPS IX. Analysis of Environmental Impact
NEEDLE DESTRUCTION DEVICES
Identified risk to
health
Mitigation measures
Patient/user exposure
to environmental
contaminants.
Performance testing.
Reprocessing validation.
Software verification,
validation, and hazard analysis.
Labeling.
Performance testing.
Labeling.
Electrical Safety Testing.
Labeling.
Electromagnetic
Compatibility
(EMC) Testing.
Labeling.
Patient/user burns .....
Electrical shock .........
Electromagnetic interference.
PO 00000
Frm 00007
Fmt 4702
Sfmt 4702
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this
proposed order contains no new
collections of information. Therefore,
clearance by the Office of Management
and Budget (OMB) under the Paperwork
Reduction Act of 1995 (PRA) (44 U.S.C.
3501–3520) is not required. This
proposed order refers to previously
approved collections of information
E:\FR\FM\07NOP1.SGM
07NOP1
Federal Register / Vol. 82, No. 214 / Tuesday, November 7, 2017 / Proposed Rules
found in FDA regulations. These
collections of information are subject to
review by OMB under the PRA. The
collections of information in 21 CFR
part 807, subpart E have been approved
under OMB control number 0910–0120
and the collections of information under
21 CFR part 801 have been approved
under OMB control number 0910–0485.
XI. Proposed Effective Date
FDA proposes that any final order
based on this proposal become effective
30 days after the date of its publication
in the Federal Register.
XII. References
The following references are on
display in the Dockets Management
Staff (see ADDRESSES) and are available
for viewing by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday; they are also available
electronically at https://
www.regulations.gov. FDA has verified
the Web site addresses, as of the date
this document publishes in the Federal
Register, but Web sites are subject to
change over time.
1. FDA memorandum ‘‘To Manufacturers
and Initial Distributors of Sharps Containers
and Destroyers Used by Health Care
Professionals,’’ February 3, 1994, available at:
https://www.fda.gov/downloads/medical
devices/deviceregulationandguidance/
guidancedocuments/ucm070679.pdf.
2. FDA ‘‘Premarket Approval of Millenium
Medical Supply Incorporated Needle-EaseTM
2501–ACTION,’’ March 6, 1997, available at:
https://www.accessdata.fda.gov/cdrh_docs/
pdf/p960044.pdf.
3. ‘‘Premarket Approval Applications
(PMA) for Sharps Needle Destruction
Devices; Final Guidance for Industry and
FDA,’’ March 2, 2001, available at: https://
www.fda.gov/downloads/MedicalDevices/
DeviceRegulationandGuidance/Guidance
Documents/ucm073601.pdf.
4. Tamplin S.A., D. Davidson, B. Powis,
and Z. O’Leary, ‘‘Issues and Options for the
Safe Destruction and Disposal of Used
Injection Materials,’’ Waste Management, vol.
25, pp. 655–665, 2005.
nshattuck on DSK9F9SC42PROD with PROPOSALS
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act (21 U.S.C. 321
et seq., as amended) and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 880 be amended as follows:
PART 880—GENERAL HOSPITAL AND
PERSONAL USE DEVICES
1. The authority citation for part 880
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
VerDate Sep<11>2014
15:06 Nov 06, 2017
Jkt 244001
2. Add § 880.6210 to subpart G to read
as follows:
■
§ 880.6210
device.
Sharps needle destruction
Frm 00008
Dated: November 2, 2017.
Anna K. Abram,
Deputy Commissioner for Policy, Planning,
Legislation, and Analysis.
[FR Doc. 2017–24191 Filed 11–6–17; 8:45 am]
(a) Identification. A sharps needle
destruction device is a prescription
device that is intended to destroy
needles or sharps used for medical
purposes by incineration or mechanical
means.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Performance testing must
demonstrate the following during
operation of the device:
(i) The device safely contains or
ventilates aerosols or fumes from device
operation.
(ii) Excessive heat or sparks are not
generated that may injure users or
patients.
(iii) Simulated use testing must
demonstrate sharps and/or needles are
completely destroyed using a range of
types and sizes of sharps sufficient to
represent actual use.
(iv) Simulated use testing must
demonstrate that the device is
physically stable on the surface for
which it is intended to be mounted to
ensure the risk of harm to the patient/
user as a result of the device falling is
minimized.
(2) Validation of cleaning and
disinfection instructions must
demonstrate that the device can be
safely and effectively reprocessed after
use per the recommended cleaning and
disinfection protocol in the instructions
for use.
(3) Analysis and/or testing must
validate electromagnetic compatibility
(EMC) and electrical safety, including
the safety of any battery used in the
device, under conditions which are
consistent with the intended
environment of device use.
(4) Software verification, validation,
and hazard analysis must be performed.
(5) Labeling must include:
(i) A clear description of the device
and its technological features;
(ii) How the device is to be used,
including validated cleaning and
disinfection instructions;
(iii) Relevant precautions and
warnings based on performance and inuse testing to ensure proper use of the
device; and
(iv) Instructions to install device in
adequately ventilated area and stable
area.
PO 00000
51589
Fmt 4702
Sfmt 4702
BILLING CODE 4164–01–P
LIBRARY OF CONGRESS
Copyright Royalty Board
37 CFR Part 381
[Docket No. 16–CRB–0002–PBR (2018–
2022)]
Determination of Rates and Terms for
Public Broadcasting (PB III)
Copyright Royalty Board (CRB),
Library of Congress.
ACTION: Proposed rule.
AGENCY:
The Copyright Royalty Judges
solicit comments on proposed rates and
terms for use of certain works in
connection with noncommercial
broadcasting for the period commencing
January 1, 2018, and ending on
December 31, 2022.
DATES: Comments and objections, if any,
are due on or before November 27, 2017.
ADDRESSES: You may submit comments
and objections, identified by docket
number 16–CRB–0002–PBR (2018–
2022), by any of the following methods:
CRB’s electronic filing application:
Submit comments online in eCRB at
https://app.crb.gov/.
U.S. mail: Copyright Royalty Board,
P.O. Box 70977, Washington, DC 20024–
0977; or
Overnight service (only USPS Express
Mail is acceptable): Copyright Royalty
Board, P.O. Box 70977, Washington, DC
20024–0977; or
Commercial courier: Address package
to: Copyright Royalty Board, Library of
Congress, James Madison Memorial
Building, LM–403, 101 Independence
Avenue SE., Washington, DC 20559–
6000. Deliver to: Congressional Courier
Acceptance Site, 2nd Street NE. and D
Street NE., Washington, DC; or
Hand delivery: Library of Congress,
James Madison Memorial Building, LM–
401, 101 Independence Avenue SE.,
Washington, DC 20559–6000.
Instructions: Unless submitting
online, commenters must submit an
original, five paper copies, and an
electronic version on a CD. All
submissions must include the CRB’s
name and docket number. All
submissions will be posted without
change to eCRB on https://www.crb.gov
including any personal information
provided.
SUMMARY:
E:\FR\FM\07NOP1.SGM
07NOP1
Agencies
[Federal Register Volume 82, Number 214 (Tuesday, November 7, 2017)]
[Proposed Rules]
[Pages 51585-51589]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2017-24191]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA-2017-N-6216]
General Hospital and Personal Use Devices; Reclassification of
Sharps Needle Destruction Device
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is issuing this
proposed order to reclassify the needle destruction device, renaming
the device to ``sharps needle destruction device,'' a postamendments
class III device (regulated under product code MTV), into class II
(special controls), subject to premarket notification. FDA is also
identifying the proposed special controls that the Agency believes are
necessary to provide a reasonable assurance of safety and effectiveness
of the device. FDA is proposing this reclassification on its own
initiative based on new information. If finalized, this order will
reclassify these types of devices from class III to class II and reduce
regulatory burdens on industry as these types of devices will no longer
be required to submit a premarket approval application (PMA) but can
instead submit a less burdensome premarket notification (510(k)) before
marketing their device.
DATES: Submit either electronic or written comments on the proposed
order by January 8, 2018. Please see section XI of this document for
the proposed effective date when the new requirements apply and for the
proposed effective date of a final order based on this proposed order.
ADDRESSES: You may submit comments as follows. Please note that late,
untimely filed comments will not be considered. Electronic comments
must be submitted on or before January 8, 2018. The https://www.regulations.gov electronic filing system will accept comments until
midnight Eastern Time at the end of January 8, 2018. Comments received
by mail/hand delivery/courier (for written/paper submissions) will be
considered timely if they are postmarked or the delivery service
acceptance receipt is on or before that date.
Electronic Submissions
Submit electronic comments in the following way:
Federal Rulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand delivery/Courier (for written/paper
submissions): Dockets Management Staff (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Dockets
Management Staff, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2017-N-6216 for ``General Hospital and Personal Use Devices;
Reclassification
[[Page 51586]]
of Sharps Needle Destruction Device.'' Received comments, those filed
in a timely manner (see ADDRESSES), will be placed in the docket and,
except for those submitted as ``Confidential Submissions,'' publicly
viewable at https://www.regulations.gov or at the Dockets Management
Staff between 9 a.m. and 4 p.m., Monday through Friday.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Dockets Management Staff. If you do not wish
your name and contact information to be made publicly available, you
can provide this information on the cover sheet and not in the body of
your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.gpo.gov/fdsys/pkg/FR-2015-09-18/pdf/2015-23389.pdf.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Dockets Management Staff, 5630 Fishers Lane,
Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Christopher K. Dugard, Center for
Devices and Radiological Health, Food and Drug Administration, 10903
New Hampshire Ave., Bldg. 66, Rm. 2561, Silver Spring, MD 20993, 240-
402-6031, christopher.dugard@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act), as
amended, establishes a comprehensive system for the regulation of
medical devices intended for human use. Section 513 of the FD&C Act (21
U.S.C. 360c) established three categories (classes) of devices,
reflecting the regulatory controls needed to provide reasonable
assurance of their safety and effectiveness. The three categories of
devices are class I (general controls), class II (special controls),
and class III (premarket approval).
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices) are
automatically classified by section 513(f)(1) of the FD&C Act into
class III without any FDA rulemaking process. Those devices remain in
class III and require premarket approval unless, and until, the device
is reclassified into class I or II, or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and 21 CFR part 807.
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified into
class I or class II under section 513(f)(3) of the FD&C Act. Section
513(f)(3) of the FD&C Act provides that FDA acting by order can
reclassify the device into class I or class II on its own initiative,
or in response to a petition from the manufacturer or importer of the
device. To change the classification of the device, the proposed new
class must have sufficient regulatory controls to provide reasonable
assurance of the safety and effectiveness of the device for its
intended use.
Reevaluation of the data previously before the Agency is an
appropriate basis for subsequent action where the reevaluation is made
in light of newly available regulatory authority (see Bell v. Goddard,
366 F.2d 177, 181 (7th Cir. 1966); Ethicon, Inc. v. FDA, 762 F. Supp.
382, 388-391 (D.D.C. 1991)) or in light of changes in ``medical
science'' (Upjohn v. Finch, 422 F.2d 944, 951 (6th Cir. 1970)). Whether
data before the Agency are old or new, the ``new information'' to
support reclassification under 513(f)(3) must be ``valid scientific
evidence'', as defined in section 513(a)(3) of the FD&C Act and 21 CFR
860.7(c)(2). (See, e.g., General Medical Co. v. FDA, 770 F.2d 214 (D.C.
Cir. 1985); Contact Lens Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.1985),
cert. denied, 474 U.S. 1062 (1986)).
FDA relies upon ``valid scientific evidence'' in the classification
process to determine the level of regulation for devices. To be
considered in the reclassification process, the ``valid scientific
evidence'' upon which the Agency relies must be publicly available.
Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
PMA (see section 520(c) of the FD&C Act (21 U.S.C. 360j(c)). Section
520(h)(4) of the FD&C Act provides that FDA may use, for
reclassification of a device, certain information in a PMA 6 years
after the application has been approved. This includes information from
clinical and preclinical tests or studies that demonstrate the safety
or effectiveness of the device, but does not include descriptions of
methods of manufacture or product composition and other trade secrets.
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the 510(k) premarket notification requirements, if the
Agency determines that premarket notification is not necessary to
reasonably assure the safety and effectiveness of the device.
II. Regulatory History of the Devices
On February 3, 1994, FDA issued a Memorandum to manufacturers and
initial distributors of sharps containers and destroyers used by health
care manufacturers to clarify the regulatory status of sharps destroyer
devices (Ref. 1).
On March 6, 1997, FDA approved its first needle destruction device
through its PMA process under section 515 of the FD&C Act (21 U.S.C.
360e). In the June 11, 1997, Federal Register notice (62 FR 31831), FDA
announced a PMA approval order for Millenium Medical Supply's
Incorporated Needle-EaseTM 2501\1\ device and the
availability of the Summary of Safety and Effectiveness Data for the
Device (SSED) (Ref. 2). As of the date of issuance of this proposed
order, FDA has approved 18 original PMAs for this device type.\2\
---------------------------------------------------------------------------
\1\ FDA approved a modified needle destruction device on
February 11, 1998. The device, as modified, is marketed under the
trade name Needle-Ease[supreg]3500.
\2\ See PMA database for original PMAs regulated under the
product code MTV: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMA/pma.cfm.
---------------------------------------------------------------------------
[[Page 51587]]
On March 2, 2001, FDA finalized guidance entitled, ``Premarket
Approval Applications (PMA) for Sharps Needle Destruction Devices,''
describing the Agency's recommendations for information to include in
PMA applications for sharps needle destruction devices intended for use
in health care settings (Ref. 3).
III. Device Description
A sharps needle destruction device is a postamendments device
classified into class III under section 513(f)(1) of the FD&C Act. A
sharps needle destruction device is a prescription device intended for
home use or in professional health care facilities to destroy sharps or
needles used for medical purposes by incineration or mechanical means.
Sharps needle destruction devices are typically electrical devices that
can destruct sharps and/or needles in a variety of methods (grinding,
incinerating, etc.) that can be either portable or stationary. Some of
these devices may also employ software to provide the user with greater
control. Please note these devices were originally identified as needle
destruction devices (product code MTV) in FDA SSEDs and product code
database; however, FDA believes the identification of sharps needle
destruction device more accurately describes this device type as it can
be used to destroy devices other than needles (e.g., sharps).
IV. Proposed Reclassification
As part of the Center for Devices and Radiological Health's 2014-
2015 strategic priority ``Strike the Right Balance Between Premarket
and Postmarket Data Collection,'' a retrospective review of class III
devices subject to PMA was completed to determine whether or not, based
on our current understanding of the technology, reclassification may be
appropriate. On August 8, 2016, FDA published a document in the Federal
Register entitled ``Retrospective Review of Premarket Approval
Application Devices; Striking the Balance Between Premarket and
Postmarket Data Collection'' in which FDA announced plans to reclassify
sharps needle destruction devices identified with the MTV product code
from class III to class II (81 FR 52445). FDA has found that sufficient
information exists to establish special controls that, together with
general controls, can provide a reasonable assurance of safety and
effectiveness for sharps needle destruction devices.
In accordance with section 513(f)(3) of the FD&C Act and 21 CFR
part 860, subpart C, FDA is proposing to reclassify this postamendments
class III device into class II. FDA believes that there is sufficient
information available to FDA through FDA's accumulated experience with
these devices from review submissions, peer-reviewed literature, and
knowledge of similar devices to establish special controls that
effectively mitigate the risks to health identified in section V.
Absent the special controls identified in this proposed order, general
controls applicable to the device are insufficient to provide
reasonable assurance of the safety and effectiveness of the device.
FDA is proposing to create a separate classification regulation for
sharps needle destruction devices that will be reclassified from class
III to II. Under this proposed order, if finalized, the sharps needle
destruction devices will be identified as a prescription device. As
such, the prescription device must satisfy prescription labeling
requirements (see Sec. 801.109 (21 CFR 801.109), Prescription
devices). Prescription devices are exempt from the requirement for
adequate directions for use for the layperson under section 502(f)(1)
of the FD&C Act (21 U.S.C. 352(f)(1)) and Sec. 801.5 (21 CFR 801.5),
as long as the conditions of Sec. 801.109 are met. In this proposed
order, if finalized, the Agency has identified the special controls
under section 513(a)(1)(B) of the FD&C Act that, together with general
controls, will provide a reasonable assurance of the safety and
effectiveness for sharps needle destruction devices.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act if FDA determines that premarket notification is
not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary for sharps needle
destruction devices to provide reasonable assurance of the safety and
effectiveness. Therefore, the Agency does not intend to exempt these
proposed class II devices from 510(k) requirements. Persons who intend
to market this type of device must submit to FDA a 510(k) and receive
clearance prior to marketing the device.
V. Risks to Health
After considering the information available to FDA through review
submissions, peer-reviewed literature, and knowledge of similar
devices, FDA determined the probable risks to health associated with
the use of sharps needle destruction devices are as follows:
Patient/user exposure to environmental contaminants.
Destroying a used sharp or needle may generate hazardous emissions from
the device that may result in infection or respiratory problems for the
patient and/or user. Contamination of the patient environment can occur
through emission of toxic fumes or infectious aerosol when the device
destroys a sharp by incineration or mechanical means, and may result
from device malfunction (mechanical and/or software). The device may
also become contaminated through regular usage and may cause cross-
contamination.
Patient/user burns as a result of excessive heat discharge
or spark formation. Excessive heat or sparks may be generated and
discharged from the device during destruction of sharps that may burn
the user.
Electromagnetic interference. While in operation, the
device may interfere with other electrically powered devices, causing
them to malfunction.
Electrical shock. While in operation, the device may
discharge electricity that could shock the user.
Sharps injury. Incompletely destroyed sharps, physical
device instability, device malfunctions, or use error may pose a risk
for a sharps injury to the user.
VI. Summary of Reasons for Reclassification
FDA believes that the sharps needle destruction devices intended
for home use or in professional health care facilities to reduce the
incidence of needlesticks by destroying sharps and/or needles in a
variety of methods (grinding, incinerating, etc.) should be
reclassified from class III to class II in light of new information
about the effectiveness of these devices. There is sufficient
information to establish special controls for sharps needle destruction
devices, in addition to general controls, which can provide reasonable
assurance of safety and effectiveness of the device, as general
controls themselves are insufficient to provide reasonable assurance of
its safety and effectiveness. FDA believes that the risks to health
associated with sharps needle destruction devices intended for home use
or in professional health care facilities to reduce the incidence of
needlesticks can be mitigated with special controls and that these
mitigations will provide a reasonable assurance of its safety and
effectiveness.
Based on a reconsideration of the available information and data,
FDA believes that there is valid scientific evidence of effectiveness
for sharps
[[Page 51588]]
needle destruction devices to reduce the incidence of needlesticks.
VII. Summary of Data Upon Which the Reclassification Is Based
FDA believes that the identified special controls, in addition to
general controls, are necessary to provide reasonable assurance of
safety and effectiveness of these devices. Taking into account the
probable health benefits of the use of the device and the nature and
known incidence of the risks of the device, FDA, on its own initiative,
is proposing to reclassify this postamendments class III device into
class II. FDA has considered and analyzed the following information: An
inclusive search of the Agency's Manufacturer and User Facility Device
Experience (MAUDE) database, which shows no adverse events for sharps
needle destruction devices; data contained in PMAs approved 6 or more
years before the date of this proposal (reviewed under section
520(h)(4) of the FD&C Act, also known as the 6-year rule); a review of
sharps containers regulated under 21 CFR 880.5570, which have similar
intended uses, but different technology, and are currently regulated as
class II devices; and one relevant article found from a literature
search that discussed the benefits and the probable risks of these
devices (Ref. 4).
VIII. Proposed Special Controls
FDA believes that the following special controls, together with
general controls, are sufficient to mitigate the risks to health
described in section V and provide a reasonable assurance of safety and
effectiveness for sharps needle destruction devices.
Performance testing will demonstrate:
[cir] The device's ability to contain or ventilate aerosols or
fumes from device operation that may result in environmental
contamination and cross-contamination. Performance testing will
demonstrate that harmful fumes, such as ozone, are not emitted by the
device during destruction of sharps needles.
[cir] Excessive heat or sparks are not generated during device
operation that may injure users or patients through characterization of
the heat dissipation profile from the heat source to the enclosure
surface, and the point of contact between the held syringe and the
user. Performance testing will ensure the heat generated through normal
operation of the device will not harm users or patients or affect
circuit performance and useful life of the device.
[cir] Complete destruction of sharps intended to be destructed to
mitigate user injuries from incomplete sharps destruction by conducting
performance testing such as simulated use demonstrating complete
destruction of the sharps and/or needles intended to be destroyed.
[cir] Mitigation of injuries from device instability through
characterization of the vibrations and movement generated by the device
to ensure device stability in the use environment.
Validation of cleaning and disinfection instructions to
demonstrate that the device can be safely and effectively reprocessed
after use to minimize the risk of patient/user cross-contamination.
Performance testing ensures electromagnetic compatibility
with other devices under conditions which are consistent with the
intended environment of device use.
Electrical safety testing ensures the risk of shock to the
patient/user is minimized.
Software hazard analysis, as well as software verification
and validation, ensures that software performs as intended and
potential software malfunctions do not impact the performance of the
device.
Labeling to ensure proper use of the device, including
warnings of the generation of excessive heat, potential for needle
stick injuries, instructions for reprocessing, and instructions for
installation (e.g., on a stable surface, adequate ventilation).
Table 1 shows how FDA believes these special controls will mitigate
each risk to health described in section V.
Table 1--Risks to Health and Mitigation Measures for Sharps Needle
Destruction Devices
------------------------------------------------------------------------
Identified risk to health Mitigation measures
------------------------------------------------------------------------
Patient/user exposure to environmental Performance testing.
contaminants. Reprocessing validation.
Software verification,
validation, and hazard
analysis.
Labeling.
Patient/user burns........................ Performance testing.
Labeling.
Electrical shock.......................... Electrical Safety Testing.
Labeling.
Electromagnetic interference.............. Electromagnetic
Compatibility (EMC)
Testing.
Labeling.
Sharps injury............................. Performance testing.
Software verification,
validation, and hazard
analysis.
Labeling.
------------------------------------------------------------------------
In addition, FDA is proposing to limit these devices to
prescription use under Sec. 801.109. Prescription devices are exempt
from the requirement for adequate directions for use for the layperson
under section 502(f)(1) of the FD&C Act and Sec. 801.5, as long as the
conditions of Sec. 801.109 are met (referring to 21 U.S.C. 352(f)(1)).
Under 21 CFR 807.81, the device would continue to be subject to 510(k)
notification requirements. FDA does not believe that clinical data is
necessary to mitigate the identified risks to health for sharps needle
destruction devices. FDA may request clinical data to evaluate
substantial equivalence when a manufacturer includes new indications
for use, such as indications for disease prevention or organism
destruction.
This reclassification order and the identified special controls, if
finalized, would provide sufficient detail regarding FDA's requirements
to reasonably assure safety and effectiveness of sharps needle
destruction devices. FDA intends to withdraw the final guidance
entitled, ``Premarket Approval Applications (PMA) for Sharps Needle
Destruction Devices; Final Guidance for Industry and FDA'' issued in
2001 upon finalization of this proposed reclassification order (Ref.
3).
IX. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
X. Paperwork Reduction Act of 1995
FDA tentatively concludes that this proposed order contains no new
collections of information. Therefore, clearance by the Office of
Management and Budget (OMB) under the Paperwork Reduction Act of 1995
(PRA) (44 U.S.C. 3501-3520) is not required. This proposed order refers
to previously approved collections of information
[[Page 51589]]
found in FDA regulations. These collections of information are subject
to review by OMB under the PRA. The collections of information in 21
CFR part 807, subpart E have been approved under OMB control number
0910-0120 and the collections of information under 21 CFR part 801 have
been approved under OMB control number 0910-0485.
XI. Proposed Effective Date
FDA proposes that any final order based on this proposal become
effective 30 days after the date of its publication in the Federal
Register.
XII. References
The following references are on display in the Dockets Management
Staff (see ADDRESSES) and are available for viewing by interested
persons between 9 a.m. and 4 p.m., Monday through Friday; they are also
available electronically at https://www.regulations.gov. FDA has
verified the Web site addresses, as of the date this document publishes
in the Federal Register, but Web sites are subject to change over time.
1. FDA memorandum ``To Manufacturers and Initial Distributors of
Sharps Containers and Destroyers Used by Health Care
Professionals,'' February 3, 1994, available at: https://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm070679.pdf.
2. FDA ``Premarket Approval of Millenium Medical Supply
Incorporated Needle-EaseTM 2501-ACTION,'' March 6, 1997,
available at: https://www.accessdata.fda.gov/cdrh_docs/pdf/p960044.pdf.
3. ``Premarket Approval Applications (PMA) for Sharps Needle
Destruction Devices; Final Guidance for Industry and FDA,'' March 2,
2001, available at: https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm073601.pdf.
4. Tamplin S.A., D. Davidson, B. Powis, and Z. O'Leary, ``Issues
and Options for the Safe Destruction and Disposal of Used Injection
Materials,'' Waste Management, vol. 25, pp. 655-665, 2005.
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act (21
U.S.C. 321 et seq., as amended) and under authority delegated to the
Commissioner of Food and Drugs, it is proposed that 21 CFR part 880 be
amended as follows:
PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES
0
1. The authority citation for part 880 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Add Sec. 880.6210 to subpart G to read as follows:
Sec. 880.6210 Sharps needle destruction device.
(a) Identification. A sharps needle destruction device is a
prescription device that is intended to destroy needles or sharps used
for medical purposes by incineration or mechanical means.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Performance testing must demonstrate the following during
operation of the device:
(i) The device safely contains or ventilates aerosols or fumes from
device operation.
(ii) Excessive heat or sparks are not generated that may injure
users or patients.
(iii) Simulated use testing must demonstrate sharps and/or needles
are completely destroyed using a range of types and sizes of sharps
sufficient to represent actual use.
(iv) Simulated use testing must demonstrate that the device is
physically stable on the surface for which it is intended to be mounted
to ensure the risk of harm to the patient/user as a result of the
device falling is minimized.
(2) Validation of cleaning and disinfection instructions must
demonstrate that the device can be safely and effectively reprocessed
after use per the recommended cleaning and disinfection protocol in the
instructions for use.
(3) Analysis and/or testing must validate electromagnetic
compatibility (EMC) and electrical safety, including the safety of any
battery used in the device, under conditions which are consistent with
the intended environment of device use.
(4) Software verification, validation, and hazard analysis must be
performed.
(5) Labeling must include:
(i) A clear description of the device and its technological
features;
(ii) How the device is to be used, including validated cleaning and
disinfection instructions;
(iii) Relevant precautions and warnings based on performance and
in-use testing to ensure proper use of the device; and
(iv) Instructions to install device in adequately ventilated area
and stable area.
Dated: November 2, 2017.
Anna K. Abram,
Deputy Commissioner for Policy, Planning, Legislation, and Analysis.
[FR Doc. 2017-24191 Filed 11-6-17; 8:45 am]
BILLING CODE 4164-01-P