Medical Devices; Neurological Devices; Classification of the Non-Electroencephalogram Physiological Signal Based Seizure Monitoring System, 50080-50082 [2017-23516]
Download as PDF
nlaroche on DSK9F9SC42PROD with RULES
50080
Federal Register / Vol. 82, No. 208 / Monday, October 30, 2017 / Rules and Regulations
of TREC levels and include samples
within the measuring range, samples
above and below the measuring range,
and samples very near above and below
the cutoff value. Multiple punches must
be obtained from each card for
demonstration of homogeneity of the
analyte across the dried blood spot.
Comparability of the test performance
for each filter paper must be
demonstrated. Stability and storage of
TREC DNA on each blood spot card
must be demonstrated. Results of the
lot-to-lot study must be summarized
providing the mean, standard deviation,
and percentage coefficient of variation
in a tabular format. Data must be
calculated for within-run, between-run,
within-lot, and between-lot. Data
demonstrating the concordance between
results across different filter papers
must be provided. Study acceptance
criteria must be provided and followed;
and
(I) If applicable, a thermocycler
reproducibility study must be
performed using thermocyclers from
three independent thermocyler
manufacturers. The sample panel must
consist of specimens with a range of
TREC levels and must include samples
within the measuring range, samples
above and below the measuring range,
and samples very near above and below
the cutoff value. The study must be
done using three filter paper lots and
conducted over five nonconsecutive
days. Results of the thermocycler
reproducibility study must be
summarized providing the mean,
standard deviation, and percentage
coefficient of variance in a tabular
format. Data must be calculated for the
within-run, between-run, within-lot,
between-lot, and between thermocycler
manufacturer study results. Study
acceptance criteria must be provided
and followed.
(iv) Identification of risk mitigation
elements used by your device, including
a description of all additional
procedures, methods, and practices
incorporated into the directions for use
that mitigate risks associated with
testing.
(2) Your § 809.10 compliant labeling
must include:
(i) A warning statement that reads
‘‘This test is not intended for diagnostic
use, preimplantation or prenatal testing,
or for screening of SCID-like syndromes,
such as DiGeorge syndrome or Omenn
syndrome. It is also not intended to
screen for less acute SCID syndromes,
such as leaky SCID or variant SCID.’’;
(ii) A warning statement that reads
‘‘Test results are intended to be used in
conjunction with other clinical and
diagnostic findings, consistent with
VerDate Sep<11>2014
13:41 Oct 27, 2017
Jkt 244001
professional standards of practice,
including confirmation by alternative
methods and clinical evaluation, as
appropriate.’’;
(iii) A description of the performance
studies listed in paragraph (b)(1)(iii) and
a summary of the results; and
(iv) A description of the filter paper
specifications required for the test.
Dated: October 24, 2017.
Anna K. Abram,
Deputy Commissioner for Policy, Planning,
Legislation, and Analysis.
[FR Doc. 2017–23496 Filed 10–27–17; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 876
[Docket No. FDA–2017–N–1609]
Medical Devices; GastroenterologyUrology Devices; Classification of the
Oral Removable Palatal Space
Occupying Device for Weight
Management and/or Weight Loss
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order; correction.
The Food and Drug
Administration (FDA) is correcting a
final order entitled ‘‘Medical Devices;
Gastroenterology-Urology Devices;
Classification of the Oral Removable
Palatal Space Occupying Device for
Weight Management and/or Weight
Loss’’ that appeared in the Federal
Register of July 28, 2017. The final order
was published with an incorrect
statement in the preamble about
whether FDA planned to exempt the
device from premarket notification
requirements. This document corrects
that error.
DATES: Effective October 30, 2017
FOR FURTHER INFORMATION CONTACT:
Mark Antonino, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. G208, Silver Spring,
MD 20993–0002, 240–402–9980,
mark.antonino@fda.hhs.gov.
SUPPLEMENTARY INFORMATION: In the
Federal Register of July 28, 2017 (82 FR
35067), FDA published the final order
‘‘Medical Devices; GastroenterologyUrology Devices; Classification of the
Oral Removable Palatal Space
Occupying Device for Weight
Management and/or Weight Loss.’’ The
final order published with an incorrect
statement in the preamble about
SUMMARY:
PO 00000
Frm 00022
Fmt 4700
Sfmt 4700
whether FDA planned to exempt the
device from premarket notification
requirements under section 510(k) of the
FD&C Act.
In the Federal Register of July 28,
2017, (82 FR 35067), the following
correction is made: On page 35069, in
the first column, the first paragraph is
corrected as follows:
‘‘Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k), if
FDA determines that premarket
notification is not necessary to provide
reasonable assurance of the safety and
effectiveness of the device. For this type
of device, FDA has determined that
premarket notification is necessary to
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, this device type is not
exempt from premarket notification
requirements. Persons who intend to
market this type of device must submit
to FDA a premarket notification, prior to
marketing the device, which contains
information about the oral removable
palatal space occupying device for
weight management and/or weight loss
they intend to market.’’
Dated: October 24, 2017.
Anna K. Abram,
Deputy Commissioner for Policy, Planning,
Legislation, and Analysis.
[FR Doc. 2017–23490 Filed 10–27–17; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 882
[Docket No. FDA–2017–N–5934]
Medical Devices; Neurological
Devices; Classification of the NonElectroencephalogram Physiological
Signal Based Seizure Monitoring
System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the nonelectroencephalogram (non-EEG)
physiological signal based seizure
monitoring system into class II (special
controls). The special controls that
apply to the device type are identified
in this order and will be part of the
codified language for the non-EEG
physiological signal based seizure
monitoring system’s classification. We
SUMMARY:
E:\FR\FM\30OCR1.SGM
30OCR1
Federal Register / Vol. 82, No. 208 / Monday, October 30, 2017 / Rules and Regulations
are taking this action because we have
determined that classifying the device
into class II (special controls) will
provide a reasonable assurance of safety
and effectiveness of the device. We
believe this action will also enhance
patients’ access to beneficial innovative
devices, in part by reducing regulatory
burdens.
DATES: This order is effective October
30, 2017. The classification was
applicable on February 16, 2017.
FOR FURTHER INFORMATION CONTACT:
Xiaorui Tang, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2609, Silver Spring,
MD 20993–0002, 301–796–6500,
xiaorui.tang@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the
non-EEG physiological signal based
seizure monitoring system as class II
(special controls), which we have
determined will provide a reasonable
assurance of safety and effectiveness. In
addition, we believe this action will
enhance patients’ access to beneficial
innovation, in part by reducing
regulatory burdens by placing the
device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act (21
U.S.C. 360c(i)) to a predicate device that
does not require premarket approval.
We determine whether a new device is
substantially equivalent to a predicate
by means of the procedures for
premarket notification under section
510(k) of the FD&C Act and part 807 (21
U.S.C. 360(k) and 21 CFR part 807,
respectively).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification (Pub. L. 105–
115). Section 607 of the Food and Drug
Administration Safety and Innovation
Act modified the De Novo application
process by adding a second procedure
(Pub. L. 112–144). A device sponsor
may utilize either procedure for De
Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act. Although the device was
automatically placed within class III,
the De Novo classification is considered
to be the initial classification of the
device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
50081
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or premarket
approval application in order to market
a substantially equivalent device (see 21
U.S.C. 360c(i), defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the less-burdensome 510(k) process,
when necessary, to market their device.
II. De Novo Classification
On November 10, 2014, Brain
Sentinel, Inc., submitted a request for
De Novo classification of the Brain
Sentinel Monitoring and Alerting
System. FDA reviewed the request in
order to classify the device under the
criteria for classification set forth in
section 513(a)(1) of the FD&C Act. We
classify devices into class II if general
controls by themselves are insufficient
to provide reasonable assurance of
safety and effectiveness, but there is
sufficient information to establish
special controls that, in combination
with the general controls, provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to general controls,
will provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on February 16, 2017, FDA
issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 882.1580. We
have named the generic type of device
non-EEG physiological signal based
seizure monitoring system, and it is
identified as a noninvasive prescription
device that collects physiological
signals other than EEG to identify
physiological signals that may be
associated with a seizure.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in
table 1.
nlaroche on DSK9F9SC42PROD with RULES
TABLE 1—NON-EEG PHYSIOLOGICAL SIGNAL BASED SEIZURE MONITORING SYSTEM RISKS AND MITIGATION MEASURES
Identified risks
Mitigation measures
Adverse tissue reaction ............................................................................
Equipment malfunction leading to injury to users (shock, burn) ..............
Interference with or from other electrical devices ....................................
VerDate Sep<11>2014
13:41 Oct 27, 2017
Jkt 244001
PO 00000
Frm 00023
Fmt 4700
Biocompatibility evaluation.
Electrical safety, thermal, and mechanical testing;
Electromagnetic compatibility testing; and
Labeling.
Electromagnetic compatibility testing.
Sfmt 4700
E:\FR\FM\30OCR1.SGM
30OCR1
50082
Federal Register / Vol. 82, No. 208 / Monday, October 30, 2017 / Rules and Regulations
TABLE 1—NON-EEG PHYSIOLOGICAL SIGNAL BASED SEIZURE MONITORING SYSTEM RISKS AND MITIGATION MEASURES—
Continued
Identified risks
Mitigation measures
Incorrect alerts, including:
• Missing a seizure—device fails to identify physiological signal
that is associated with a seizure; or.
• False alarm—device mistakenly identifies a physiological signal
as being associated with a seizure.
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of the safety and effectiveness. In order
for a device to fall within this
classification, and thus avoid automatic
classification in class III, it would have
to comply with the special controls
named in this final order. The necessary
special controls appear in the regulation
codified by this order. This device is
subject to premarket notification
requirements under section 510(k) of the
FD&C Act.
At the time of classification, non-EEG
physiological signal based seizure
monitoring systems are for prescription
use only. Prescription devices are
exempt from the requirement for
adequate directions for use for the
layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352(f)(1)) and 21
CFR 801.5, as long as the conditions of
21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
nlaroche on DSK9F9SC42PROD with RULES
IV. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
part 807, subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120, and the collections of
information in 21 CFR part 801,
regarding labeling have been approved
under OMB control number 0910–0485.
List of Subjects in 21 CFR Part 882
Medical devices.
VerDate Sep<11>2014
13:41 Oct 27, 2017
Jkt 244001
Clinical performance testing;
Non-clinical performance testing;
Software verification, validation, and hazard analysis;
Labeling; and
Training.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 882 is
amended as follows:
PART 882—NEUROLOGICAL DEVICES
1. The authority citation for part 882
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 882.1580 to subpart B to read
as follows:
■
§ 882.1580 Non-electroencephalogram
(EEG) physiological signal based seizure
monitoring system.
(a) Identification. A nonelectroencephalogram (non-EEG)
physiological signal based seizure
monitoring system is a noninvasive
prescription device that collects
physiological signals other than EEG to
identify physiological signals that may
be associated with a seizure.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The technical parameters of the
device, hardware and software, must be
fully characterized and include the
following information:
(i) Hardware specifications must be
provided. Appropriate verification,
validation, and hazard analysis must be
performed.
(ii) Software, including any
proprietary algorithm(s) used by the
device to achieve its intended use, must
be described in detail in the Software
Requirements Specification (SRS) and
Software Design Specification (SDS).
Appropriate software verification,
validation, and hazard analysis must be
performed.
(2) The patient-contacting
components of the device must be
demonstrated to be biocompatible.
(3) The device must be designed and
tested for electrical, thermal, and
mechanical safety and electromagnetic
compatibility (EMC).
(4) Clinical performance testing must
demonstrate the ability of the device to
function as an assessment aid for
PO 00000
Frm 00024
Fmt 4700
Sfmt 4700
monitoring for seizure-related activity in
the intended population and for the
intended use setting. Performance
measurements must include positive
percent agreement and false alarm rate.
(5) Training must be provided for
intended users that includes
information regarding the proper use of
the device and factors that may affect
the collection of the physiologic data.
(6) The labeling must include health
care professional labeling and patientcaregiver labeling. The health care
professional and the patient-caregiver
labeling must include the following
information:
(i) A detailed summary of the clinical
performance testing, including any
adverse events and complications.
(ii) Any instructions technicians and
clinicians should convey to patients and
caregivers regarding the proper use of
the device and factors that may affect
the collection of the physiologic data.
(iii) Instructions to technicians and
clinicians regarding how to set the
device threshold to achieve the
intended performance of the device.
Dated: October 24, 2017.
Lauren Silvis,
Chief of Staff.
[FR Doc. 2017–23516 Filed 10–27–17; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Part 117
[Docket No. USCG 2017–0162]
RIN 1625–AA09
Drawbridge Operation Regulation;
Nanticoke River, Seaford, DE
Coast Guard, DHS.
Final rule.
AGENCY:
ACTION:
The Coast Guard is modifying
the operating schedule that governs the
SR 13 Bridge across the Nanticoke
River, mile 39.6, in Seaford, Delaware
(DE). This modification will require the
SUMMARY:
E:\FR\FM\30OCR1.SGM
30OCR1
Agencies
[Federal Register Volume 82, Number 208 (Monday, October 30, 2017)]
[Rules and Regulations]
[Pages 50080-50082]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2017-23516]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 882
[Docket No. FDA-2017-N-5934]
Medical Devices; Neurological Devices; Classification of the Non-
Electroencephalogram Physiological Signal Based Seizure Monitoring
System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the non-electroencephalogram (non-EEG) physiological signal based
seizure monitoring system into class II (special controls). The special
controls that apply to the device type are identified in this order and
will be part of the codified language for the non-EEG physiological
signal based seizure monitoring system's classification. We
[[Page 50081]]
are taking this action because we have determined that classifying the
device into class II (special controls) will provide a reasonable
assurance of safety and effectiveness of the device. We believe this
action will also enhance patients' access to beneficial innovative
devices, in part by reducing regulatory burdens.
DATES: This order is effective October 30, 2017. The classification was
applicable on February 16, 2017.
FOR FURTHER INFORMATION CONTACT: Xiaorui Tang, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2609, Silver Spring, MD 20993-0002, 301-796-6500,
[email protected].
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the non-EEG physiological signal
based seizure monitoring system as class II (special controls), which
we have determined will provide a reasonable assurance of safety and
effectiveness. In addition, we believe this action will enhance
patients' access to beneficial innovation, in part by reducing
regulatory burdens by placing the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (the FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act (21 U.S.C. 360c(i)) to a predicate device that
does not require premarket approval. We determine whether a new device
is substantially equivalent to a predicate by means of the procedures
for premarket notification under section 510(k) of the FD&C Act and
part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically placed
within class III, the De Novo classification is considered to be the
initial classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application in order to market a substantially
equivalent device (see 21 U.S.C. 360c(i), defining ``substantial
equivalence''). Instead, sponsors can use the less-burdensome 510(k)
process, when necessary, to market their device.
II. De Novo Classification
On November 10, 2014, Brain Sentinel, Inc., submitted a request for
De Novo classification of the Brain Sentinel Monitoring and Alerting
System. FDA reviewed the request in order to classify the device under
the criteria for classification set forth in section 513(a)(1) of the
FD&C Act. We classify devices into class II if general controls by
themselves are insufficient to provide reasonable assurance of safety
and effectiveness, but there is sufficient information to establish
special controls that, in combination with the general controls,
provide reasonable assurance of the safety and effectiveness of the
device for its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review
of the information submitted in the request, we determined that the
device can be classified into class II with the establishment of
special controls. FDA has determined that these special controls, in
addition to general controls, will provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on February 16, 2017, FDA issued an order to the
requestor classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 882.1580. We have named
the generic type of device non-EEG physiological signal based seizure
monitoring system, and it is identified as a noninvasive prescription
device that collects physiological signals other than EEG to identify
physiological signals that may be associated with a seizure.
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--Non-EEG Physiological Signal Based Seizure Monitoring System
Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Adverse tissue reaction................ Biocompatibility evaluation.
Equipment malfunction leading to injury Electrical safety, thermal, and
to users (shock, burn). mechanical testing;
Electromagnetic compatibility
testing; and
Labeling.
Interference with or from other Electromagnetic compatibility
electrical devices. testing.
[[Page 50082]]
Incorrect alerts, including: ...............................
Missing a seizure--device Clinical performance testing;
fails to identify physiological Non-clinical performance
signal that is associated with a testing;
seizure; or. Software verification,
False alarm--device validation, and hazard
mistakenly identifies a analysis;
physiological signal as being Labeling; and
associated with a seizure. Training.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of the safety and effectiveness. In order for a device to
fall within this classification, and thus avoid automatic
classification in class III, it would have to comply with the special
controls named in this final order. The necessary special controls
appear in the regulation codified by this order. This device is subject
to premarket notification requirements under section 510(k) of the FD&C
Act.
At the time of classification, non-EEG physiological signal based
seizure monitoring systems are for prescription use only. Prescription
devices are exempt from the requirement for adequate directions for use
for the layperson under section 502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as the conditions of 21 CFR
801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in
part 807, subpart E, regarding premarket notification submissions have
been approved under OMB control number 0910-0120, and the collections
of information in 21 CFR part 801, regarding labeling have been
approved under OMB control number 0910-0485.
List of Subjects in 21 CFR Part 882
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
882 is amended as follows:
PART 882--NEUROLOGICAL DEVICES
0
1. The authority citation for part 882 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 882.1580 to subpart B to read as follows:
Sec. 882.1580 Non-electroencephalogram (EEG) physiological signal
based seizure monitoring system.
(a) Identification. A non-electroencephalogram (non-EEG)
physiological signal based seizure monitoring system is a noninvasive
prescription device that collects physiological signals other than EEG
to identify physiological signals that may be associated with a
seizure.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The technical parameters of the device, hardware and software,
must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Appropriate
verification, validation, and hazard analysis must be performed.
(ii) Software, including any proprietary algorithm(s) used by the
device to achieve its intended use, must be described in detail in the
Software Requirements Specification (SRS) and Software Design
Specification (SDS). Appropriate software verification, validation, and
hazard analysis must be performed.
(2) The patient-contacting components of the device must be
demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal,
and mechanical safety and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the ability of
the device to function as an assessment aid for monitoring for seizure-
related activity in the intended population and for the intended use
setting. Performance measurements must include positive percent
agreement and false alarm rate.
(5) Training must be provided for intended users that includes
information regarding the proper use of the device and factors that may
affect the collection of the physiologic data.
(6) The labeling must include health care professional labeling and
patient-caregiver labeling. The health care professional and the
patient-caregiver labeling must include the following information:
(i) A detailed summary of the clinical performance testing,
including any adverse events and complications.
(ii) Any instructions technicians and clinicians should convey to
patients and caregivers regarding the proper use of the device and
factors that may affect the collection of the physiologic data.
(iii) Instructions to technicians and clinicians regarding how to
set the device threshold to achieve the intended performance of the
device.
Dated: October 24, 2017.
Lauren Silvis,
Chief of Staff.
[FR Doc. 2017-23516 Filed 10-27-17; 8:45 am]
BILLING CODE 4164-01-P