Medical Devices; Gastroenterology-Urology Devices; Classification of the High Intensity Ultrasound System for Prostate Tissue Ablation, 45725-45727 [2017-21074]
Download as PDF
Federal Register / Vol. 82, No. 189 / Monday, October 2, 2017 / Rules and Regulations
17 CFR Part 230
PART 230—GENERAL RULES AND
REGULATIONS, SECURITIES ACT OF
1933
Reporting and recordkeeping
requirements, Securities.
In accordance with the foregoing, title
17, chapter II of the Code of Federal
Regulations is amended as follows:
Authority: 15 U.S.C. 77d, 77d–1, 77s, 77z–
3, 78c, 78o, 78q, 78w, 78mm, and Pub. L.
112–106, secs. 301–305, 126 Stat. 306 (2012).
2. Amend § 227.202 by adding
paragraph (c) to read as follows:
■
Ongoing reporting requirements.
jstallworth on DSKBBY8HB2PROD with RULES
15:21 Sep 29, 2017
Jkt 244001
21 CFR Part 876
Authority: 15 U.S.C. 77b, 77b note, 77c,
77d, 77f, 77g, 77h, 77j, 77r, 77s, 77z–3, 77sss,
78c, 78d, 78j, 78l, 78m, 78n, 78o, 78o–7 note,
78t, 78w, 78ll(d), 78mm, 80a–8, 80a–24, 80a–
28, 80a–29, 80a–30, and 80a–37, and Pub. L.
112–106, sec. 201(a), sec. 401, 126 Stat. 313
(2012), unless otherwise noted.
[Docket No. FDA–2017–N–5153]
*
*
*
*
4. Amend § 230.257 by adding
paragraph (f) to read as follows:
AGENCY:
HHS.
§ 230.257 Periodic and current reporting;
exit report.
SUMMARY:
*
*
*
*
*
(c) Temporary relief from certain
reporting requirements. (1) An issuer
that is not able to meet a filing deadline
for any report or form required to be
filed by this section (Rule 202), Rule
203(a)(3) (§ 227.203(a)(3)), or Rule
203(b) (§ 227.203(b)), as applicable:
(i) During the period from and
including August 25, 2017 to and
including October 26, 2017 due to
Hurricane Harvey and its aftermath
shall be deemed to have satisfied the
filing deadline for such report or form
if the issuer files such report or form
with the Commission on or before
October 27, 2017;
(ii) During the period from and
including September 6, 2017 to and
including November 7, 2017 due to
Hurricane Irma and its aftermath shall
be deemed to have satisfied the filing
deadline for such report or form if the
issuer files such report or form with the
Commission on or before November 8,
2017; or
(ii) During the period from and
including September 20, 2017 to and
including November 21, 2017 due to
Hurricane Maria and its aftermath shall
be deemed to have satisfied the filing
deadline for such report or form if the
issuer files such report or form with the
Commission on or before November 22,
2017.
(2) In any report or form filed
pursuant to paragraph (c)(1) of this
section, the issuer must disclose that it
is relying on this paragraph (c) (Rule
202(c) of Regulation Crowdfunding) and
state the reasons why, in good faith, it
could not file such report or form on a
timely basis.
VerDate Sep<11>2014
Food and Drug Administration
■
■
1. The authority citation for part 227
is revised to read as follows:
■
*
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
*
PART 227—REGULATION
CROWDFUNDING, GENERAL RULES
AND REGULATIONS
§ 227.202
3. The authority citation for part 230
continues to read in part as follows:
45725
*
*
*
*
(f) Temporary relief from ongoing
reporting requirements. (1) An issuer
that is not able to meet a filing deadline
for any report or form required to be
filed by Rule 252(f)(2)(i)
(§ 230.252(f)(2)(i)) or this section (Rule
257), as applicable:
(i) During the period from and
including August 25, 2017 to and
including October 26, 2017 due to
Hurricane Harvey and its aftermath
shall be deemed to have satisfied the
filing deadline for such report or form
if the issuer files such report or form
with the Commission on or before
October 27, 2017;
(ii) During the period from and
including September 6, 2017 to and
including November 7, 2017 due to
Hurricane Irma and its aftermath shall
be deemed to have satisfied the filing
deadline for such report or form if the
issuer files such report or form with the
Commission on or before November 8,
2017; or
(ii) During the period from and
including September 20, 2017 to and
including November 21, 2017 due to
Hurricane Maria and its aftermath shall
be deemed to have satisfied the filing
deadline for such report or form if the
issuer files such report or form with the
Commission on or before November 22,
2017.
(2) In any report or form filed
pursuant to paragraph (f)(1) of this
section, the issuer must disclose that it
is relying on this paragraph (f) (Rule
257(f) of Regulation A) and state the
reasons why, in good faith, it could not
file such report or form on a timely
basis.
By the Commission.
Dated: September 27, 2017.
Brent J. Fields,
Secretary.
[FR Doc. 2017–21094 Filed 9–28–17; 4:15 pm]
BILLING CODE 8011–01–P
PO 00000
Frm 00047
Fmt 4700
Sfmt 4700
Medical Devices; GastroenterologyUrology Devices; Classification of the
High Intensity Ultrasound System for
Prostate Tissue Ablation
ACTION:
Food and Drug Administration,
Final order.
The Food and Drug
Administration (FDA or we) is
classifying the high intensity ultrasound
system for prostate tissue ablation into
class II (special controls). The special
controls that apply to the device type
are identified in this order and will be
part of the codified language for the
high intensity ultrasound system for
prostate tissue ablation’s classification.
We are taking this action because we
have determined that classifying the
device into class II (special controls)
will provide a reasonable assurance of
safety and effectiveness of the device.
We believe this action will also enhance
patients’ access to beneficial innovative
devices, in part by reducing regulatory
burdens.
DATES: This order is effective October 2,
2017. The classification was applicable
on October 9, 2015.
FOR FURTHER INFORMATION CONTACT: John
Baxley, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. G210, Silver Spring,
MD 20993–0002, 301–796–6549,
john.baxley@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the
high intensity ultrasound system for
prostate tissue ablation as class II
(special controls), which we have
determined will provide a reasonable
assurance of safety and effectiveness. In
addition, we believe this action will
enhance patients’ access to beneficial
innovation, in part by reducing
regulatory burdens by placing the
device into a lower device class than the
automatic class III assignment.
The automatic assignment of class III
occurs by operation of law and without
any action by FDA, regardless of the
level of risk posed by the new device.
Any device that was not in commercial
distribution before May 28, 1976, is
automatically classified as, and remains
E:\FR\FM\02OCR1.SGM
02OCR1
45726
Federal Register / Vol. 82, No. 189 / Monday, October 2, 2017 / Rules and Regulations
within, class III and requires premarket
approval unless and until FDA takes an
action to classify or reclassify the device
(see 21 U.S.C. 360c(f)(1)). We refer to
these devices as ‘‘postamendments
devices’’ because they were not in
commercial distribution prior to the
date of enactment of the Medical Device
Amendments of 1976, which amended
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act).
FDA may take a variety of actions in
appropriate circumstances to classify or
reclassify a device into class I or II. We
may issue an order finding a new device
to be substantially equivalent under
section 513(i) of the FD&C Act to a
predicate device that does not require
premarket approval (see 21 U.S.C.
360c(i)). We determine whether a new
device is substantially equivalent to a
predicate by means of the procedures
for premarket notification under section
510(k) of the FD&C Act and part 807 (21
U.S.C. 360(k) and 21 CFR part 807,
respectively).
FDA may also classify a device
through ‘‘De Novo’’ classification, a
common name for the process
authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and
Drug Administration Modernization Act
of 1997 established the first procedure
for De Novo classification
(Pub. L. 105–115). Section 607 of the
Food and Drug Administration Safety
and Innovation Act modified the De
Novo application process by adding a
second procedure (Pub. L. 112–144). A
device sponsor may utilize either
procedure for De Novo classification.
Under the first procedure, the person
submits a 510(k) for a device that has
not previously been classified. After
receiving an order from FDA classifying
the device into class III under section
513(f)(1) of the FD&C Act, the person
then requests a classification under
section 513(f)(2).
Under the second procedure, rather
than first submitting a 510(k) and then
a request for classification, if the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence, that person requests a
classification under section 513(f)(2) of
the FD&C Act.
Under either procedure for De Novo
classification, FDA shall classify the
device by written order within 120 days.
The classification will be according to
the criteria under section 513(a)(1) of
the FD&C Act. Although the device was
automatically within class III, the De
Novo classification is considered to be
the initial classification of the device.
We believe this De Novo classification
will enhance patients’ access to
beneficial innovation, in part by
reducing regulatory burdens. When FDA
classifies a device into class I or II via
the De Novo process, the device can
serve as a predicate for future devices of
that type, including for 510(k)s (see 21
U.S.C. 360c(f)(2)(B)(i)). As a result, other
device sponsors do not have to submit
a De Novo request or premarket
approval application in order to market
a substantially equivalent device (see 21
U.S.C. 360c(i), defining ‘‘substantial
equivalence’’). Instead, sponsors can use
the less burdensome 510(k) process,
when necessary, to market their device.
II. De Novo Classification
On March 23, 2015, SonaCare
Medical, LLC submitted a request for De
Novo classification of the
Sonablate® 450. FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1) of the FD&C
Act. We classify devices into class II if
general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls that, in
combination with the general controls,
provide reasonable assurance of the
safety and effectiveness of the device for
its intended use (see 21 U.S.C.
360c(a)(1)(B)). After review of the
information submitted in the request,
we determined that the device can be
classified into class II with the
establishment of special controls. FDA
has determined that these special
controls, in addition to general controls,
will provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on October 9, 2015, FDA
issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 876.4340. We
have named the generic type of device
high intensity ultrasound system for
prostate tissue ablation, and it is
identified as a prescription device that
transmits high intensity therapeutic
ultrasound energy into the prostate to
thermally ablate a defined, targeted
volume of tissue, performed under
imaging guidance. This classification
does not include devices that are
intended for the treatment of any
specific prostate disease and does not
include devices that are intended to
ablate non-prostatic tissues/organs.
FDA has identified the following risks
to health associated specifically with
this type of device and the measures
required to mitigate these risks in
table 1.
TABLE 1—HIGH INTENSITY ULTRASOUND SYSTEM FOR PROSTATE TISSUE ABLATION RISKS AND MITIGATION MEASURES
jstallworth on DSKBBY8HB2PROD with RULES
Identified risks
Mitigation measures
Thermal injury from high intensity ultrasound exposure to non-target tissue:
• Erectile dysfunction
• Urinary incontinence
• Rectal fistula
• Osteomyelitis pubis
Thermal injury from high intensity ultrasound exposure to target tissue:
• Urethral stricture
• Bladder neck contracture
• Urinary retention
• Tissue debris/obstruction
• Voiding dysfunction
• Dysuria
• Hematuria
• Ejaculation disorder
Mechanical injury from unintentional movement of ultrasound components:
• Patient rectal injury
• Operator hand injury
Non-clinical performance testing; Software verification, validation, and
hazard analysis; In vivo testing; Clinical testing; Labeling; and Physician training.
VerDate Sep<11>2014
15:21 Sep 29, 2017
Jkt 244001
PO 00000
Frm 00048
Fmt 4700
Clinical testing, Labeling, and Physician training.
Software verification, validation, and hazard analysis; Clinical testing;
Labeling; and Physician training.
Sfmt 4700
E:\FR\FM\02OCR1.SGM
02OCR1
Federal Register / Vol. 82, No. 189 / Monday, October 2, 2017 / Rules and Regulations
45727
TABLE 1—HIGH INTENSITY ULTRASOUND SYSTEM FOR PROSTATE TISSUE ABLATION RISKS AND MITIGATION MEASURES—
Continued
Identified risks
Mitigation measures
Infection ....................................................................................................
Sterilization validation, Reprocessing validation, Shelf life validation,
and Labeling.
Electrical safety testing, Electromagnetic compatibility testing, and Labeling.
Biocompatibility testing.
Electrical shock/electromagnetic interference ..........................................
Adverse tissue reaction ............................................................................
FDA has determined that special
controls, in combination with the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness. In order for
a device to fall within this classification,
and thus avoid automatic classification
in class III, it would have to comply
with the special controls named in this
final order. The necessary special
controls appear in the regulation
codified by this order. This device is
subject to premarket notification
requirements under section 510(k).
At the time of classification, high
intensity ultrasound systems for
prostate tissue ablation are for
prescription use only. Prescription
devices are exempt from the
requirement for adequate directions for
use for the layperson under section
502(f)(1) of the FD&C Act (21 U.S.C.
352(f)(1)) and 21 CFR 801.5, as long as
the conditions of 21 CFR 801.109 are
met.
jstallworth on DSKBBY8HB2PROD with RULES
III. Analysis of Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IV. Paperwork Reduction Act of 1995
This final administrative order
establishes special controls that refer to
previously approved collections of
information found in other FDA
regulations. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in part 807,
subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120, and the collections of
information in part 801, regarding
labeling, have been approved under
OMB control number 0910–0485.
List of Subjects in 21 CFR Part 876
Medical devices.
VerDate Sep<11>2014
15:21 Sep 29, 2017
Jkt 244001
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 876 is
amended as follows:
PART 876—GASTROENTEROLOGY–
UROLOGY DEVICES
1. The authority citation for part 876
continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 360l, 371.
2. Add § 876.4340 to subpart E to read
as follows:
■
§ 876.4340 High intensity ultrasound
system for prostate tissue ablation.
(a) Identification. A high intensity
ultrasound system for prostate tissue
ablation is a prescription device that
transmits high intensity therapeutic
ultrasound energy into the prostate to
thermally ablate a defined, targeted
volume of tissue, performed under
imaging guidance. This classification
does not include devices that are
intended for the treatment of any
specific prostate disease and does not
include devices that are intended to
ablate non-prostatic tissues/organs.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Non-clinical performance data
must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
(i) Characterization of acoustic
pressure and power output at clinically
relevant levels;
(ii) Measurement of targeting accuracy
and reproducibility of high intensity
ultrasound output;
(iii) Ultrasound-induced heating
verification testing at target and nontarget tissues;
(iv) Electrical safety testing; and
(v) Electromagnetic compatibility
testing.
(2) Software verification, validation,
and hazard analysis must be performed.
(3) The elements of the device that
may contact the patient’s mucosal tissue
must be demonstrated to be
biocompatible.
PO 00000
Frm 00049
Fmt 4700
Sfmt 9990
(4) Performance data must
demonstrate the sterility of the device
components that contact the patient’s
mucosal tissue.
(5) Performance data must support
shelf life by demonstrating continued
sterility of the device or the sterile
components, package integrity, and
device functionality over the identified
shelf life.
(6) Performance data must support the
instructions for reprocessing all reusable
components.
(7) In vivo testing must demonstrate
that the device thermally ablates
targeted tissue in a controlled manner
without thermal injury to adjacent, nontarget tissues.
(8) Clinical testing must document the
adverse event profile, provide evidence
of prostatic ablation, and demonstrate
that the device performs as intended
under anticipated conditions of use.
(9) Training must be provided so that
upon completion of the training
program, the physician can:
(i) Use all safety features of the
device;
(ii) Accurately target the high
intensity ultrasound energy within the
desired region of the prostate; and
(iii) Perform the ablation procedure in
a manner that minimizes damage to
non-target tissues.
(10) Labeling must include:
(i) A section that summarizes the
clinical testing results, including the
adverse event profile and evidence of
prostate ablation achieved; and
(ii) An expiration date or shelf life for
single use components.
Dated: September 25, 2017.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2017–21074 Filed 9–29–17; 8:45 am]
BILLING CODE 4164–01–P
E:\FR\FM\02OCR1.SGM
02OCR1
Agencies
[Federal Register Volume 82, Number 189 (Monday, October 2, 2017)]
[Rules and Regulations]
[Pages 45725-45727]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2017-21074]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 876
[Docket No. FDA-2017-N-5153]
Medical Devices; Gastroenterology-Urology Devices; Classification
of the High Intensity Ultrasound System for Prostate Tissue Ablation
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or we) is classifying
the high intensity ultrasound system for prostate tissue ablation into
class II (special controls). The special controls that apply to the
device type are identified in this order and will be part of the
codified language for the high intensity ultrasound system for prostate
tissue ablation's classification. We are taking this action because we
have determined that classifying the device into class II (special
controls) will provide a reasonable assurance of safety and
effectiveness of the device. We believe this action will also enhance
patients' access to beneficial innovative devices, in part by reducing
regulatory burdens.
DATES: This order is effective October 2, 2017. The classification was
applicable on October 9, 2015.
FOR FURTHER INFORMATION CONTACT: John Baxley, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. G210, Silver Spring, MD 20993-0002, 301-796-6549,
john.baxley@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
Upon request, FDA has classified the high intensity ultrasound
system for prostate tissue ablation as class II (special controls),
which we have determined will provide a reasonable assurance of safety
and effectiveness. In addition, we believe this action will enhance
patients' access to beneficial innovation, in part by reducing
regulatory burdens by placing the device into a lower device class than
the automatic class III assignment.
The automatic assignment of class III occurs by operation of law
and without any action by FDA, regardless of the level of risk posed by
the new device. Any device that was not in commercial distribution
before May 28, 1976, is automatically classified as, and remains
[[Page 45726]]
within, class III and requires premarket approval unless and until FDA
takes an action to classify or reclassify the device (see 21 U.S.C.
360c(f)(1)). We refer to these devices as ``postamendments devices''
because they were not in commercial distribution prior to the date of
enactment of the Medical Device Amendments of 1976, which amended the
Federal Food, Drug, and Cosmetic Act (the FD&C Act).
FDA may take a variety of actions in appropriate circumstances to
classify or reclassify a device into class I or II. We may issue an
order finding a new device to be substantially equivalent under section
513(i) of the FD&C Act to a predicate device that does not require
premarket approval (see 21 U.S.C. 360c(i)). We determine whether a new
device is substantially equivalent to a predicate by means of the
procedures for premarket notification under section 510(k) of the FD&C
Act and part 807 (21 U.S.C. 360(k) and 21 CFR part 807, respectively).
FDA may also classify a device through ``De Novo'' classification,
a common name for the process authorized under section 513(f)(2) of the
FD&C Act. Section 207 of the Food and Drug Administration Modernization
Act of 1997 established the first procedure for De Novo classification
(Pub. L. 105-115). Section 607 of the Food and Drug Administration
Safety and Innovation Act modified the De Novo application process by
adding a second procedure (Pub. L. 112-144). A device sponsor may
utilize either procedure for De Novo classification.
Under the first procedure, the person submits a 510(k) for a device
that has not previously been classified. After receiving an order from
FDA classifying the device into class III under section 513(f)(1) of
the FD&C Act, the person then requests a classification under section
513(f)(2).
Under the second procedure, rather than first submitting a 510(k)
and then a request for classification, if the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence, that person requests a classification under
section 513(f)(2) of the FD&C Act.
Under either procedure for De Novo classification, FDA shall
classify the device by written order within 120 days. The
classification will be according to the criteria under section
513(a)(1) of the FD&C Act. Although the device was automatically within
class III, the De Novo classification is considered to be the initial
classification of the device.
We believe this De Novo classification will enhance patients'
access to beneficial innovation, in part by reducing regulatory
burdens. When FDA classifies a device into class I or II via the De
Novo process, the device can serve as a predicate for future devices of
that type, including for 510(k)s (see 21 U.S.C. 360c(f)(2)(B)(i)). As a
result, other device sponsors do not have to submit a De Novo request
or premarket approval application in order to market a substantially
equivalent device (see 21 U.S.C. 360c(i), defining ``substantial
equivalence''). Instead, sponsors can use the less burdensome 510(k)
process, when necessary, to market their device.
II. De Novo Classification
On March 23, 2015, SonaCare Medical, LLC submitted a request for De
Novo classification of the Sonablate[supreg] 450. FDA reviewed the
request in order to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act. We
classify devices into class II if general controls by themselves are
insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls that, in combination with the general controls, provide
reasonable assurance of the safety and effectiveness of the device for
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the
information submitted in the request, we determined that the device can
be classified into class II with the establishment of special controls.
FDA has determined that these special controls, in addition to general
controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on October 9, 2015, FDA issued an order to the requestor
classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 876.4340. We have named
the generic type of device high intensity ultrasound system for
prostate tissue ablation, and it is identified as a prescription device
that transmits high intensity therapeutic ultrasound energy into the
prostate to thermally ablate a defined, targeted volume of tissue,
performed under imaging guidance. This classification does not include
devices that are intended for the treatment of any specific prostate
disease and does not include devices that are intended to ablate non-
prostatic tissues/organs.
FDA has identified the following risks to health associated
specifically with this type of device and the measures required to
mitigate these risks in table 1.
Table 1--High Intensity Ultrasound System for Prostate Tissue Ablation
Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risks Mitigation measures
------------------------------------------------------------------------
Thermal injury from high intensity Non-clinical performance
ultrasound exposure to non-target testing; Software
tissue: verification, validation, and
Erectile dysfunction hazard analysis; In vivo
Urinary incontinence testing; Clinical testing;
Rectal fistula Labeling; and Physician
Osteomyelitis pubis training.
Thermal injury from high intensity Clinical testing, Labeling, and
ultrasound exposure to target tissue: Physician training.
Urethral stricture
Bladder neck contracture
Urinary retention
Tissue debris/obstruction
Voiding dysfunction
Dysuria
Hematuria
Ejaculation disorder
Mechanical injury from unintentional Software verification,
movement of ultrasound components: validation, and hazard
Patient rectal injury analysis; Clinical testing;
Operator hand injury Labeling; and Physician
training.
[[Page 45727]]
Infection.............................. Sterilization validation,
Reprocessing validation, Shelf
life validation, and Labeling.
Electrical shock/electromagnetic Electrical safety testing,
interference. Electromagnetic compatibility
testing, and Labeling.
Adverse tissue reaction................ Biocompatibility testing.
------------------------------------------------------------------------
FDA has determined that special controls, in combination with the
general controls, address these risks to health and provide reasonable
assurance of safety and effectiveness. In order for a device to fall
within this classification, and thus avoid automatic classification in
class III, it would have to comply with the special controls named in
this final order. The necessary special controls appear in the
regulation codified by this order. This device is subject to premarket
notification requirements under section 510(k).
At the time of classification, high intensity ultrasound systems
for prostate tissue ablation are for prescription use only.
Prescription devices are exempt from the requirement for adequate
directions for use for the layperson under section 502(f)(1) of the
FD&C Act (21 U.S.C. 352(f)(1)) and 21 CFR 801.5, as long as the
conditions of 21 CFR 801.109 are met.
III. Analysis of Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IV. Paperwork Reduction Act of 1995
This final administrative order establishes special controls that
refer to previously approved collections of information found in other
FDA regulations. These collections of information are subject to review
by the Office of Management and Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of
information in part 807, subpart E, regarding premarket notification
submissions have been approved under OMB control number 0910-0120, and
the collections of information in part 801, regarding labeling, have
been approved under OMB control number 0910-0485.
List of Subjects in 21 CFR Part 876
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
876 is amended as follows:
PART 876--GASTROENTEROLOGY-UROLOGY DEVICES
0
1. The authority citation for part 876 continues to read as follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.
0
2. Add Sec. 876.4340 to subpart E to read as follows:
Sec. 876.4340 High intensity ultrasound system for prostate tissue
ablation.
(a) Identification. A high intensity ultrasound system for prostate
tissue ablation is a prescription device that transmits high intensity
therapeutic ultrasound energy into the prostate to thermally ablate a
defined, targeted volume of tissue, performed under imaging guidance.
This classification does not include devices that are intended for the
treatment of any specific prostate disease and does not include devices
that are intended to ablate non-prostatic tissues/organs.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Non-clinical performance data must demonstrate that the device
performs as intended under anticipated conditions of use. The following
performance characteristics must be tested:
(i) Characterization of acoustic pressure and power output at
clinically relevant levels;
(ii) Measurement of targeting accuracy and reproducibility of high
intensity ultrasound output;
(iii) Ultrasound-induced heating verification testing at target and
non-target tissues;
(iv) Electrical safety testing; and
(v) Electromagnetic compatibility testing.
(2) Software verification, validation, and hazard analysis must be
performed.
(3) The elements of the device that may contact the patient's
mucosal tissue must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of the device
components that contact the patient's mucosal tissue.
(5) Performance data must support shelf life by demonstrating
continued sterility of the device or the sterile components, package
integrity, and device functionality over the identified shelf life.
(6) Performance data must support the instructions for reprocessing
all reusable components.
(7) In vivo testing must demonstrate that the device thermally
ablates targeted tissue in a controlled manner without thermal injury
to adjacent, non-target tissues.
(8) Clinical testing must document the adverse event profile,
provide evidence of prostatic ablation, and demonstrate that the device
performs as intended under anticipated conditions of use.
(9) Training must be provided so that upon completion of the
training program, the physician can:
(i) Use all safety features of the device;
(ii) Accurately target the high intensity ultrasound energy within
the desired region of the prostate; and
(iii) Perform the ablation procedure in a manner that minimizes
damage to non-target tissues.
(10) Labeling must include:
(i) A section that summarizes the clinical testing results,
including the adverse event profile and evidence of prostate ablation
achieved; and
(ii) An expiration date or shelf life for single use components.
Dated: September 25, 2017.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2017-21074 Filed 9-29-17; 8:45 am]
BILLING CODE 4164-01-P