Medical Devices; Cardiovascular Devices; Classification of the Coronary Vascular Physiologic Simulation Software Device, 63671-63674 [2015-26658]
Download as PDF
Federal Register / Vol. 80, No. 203 / Wednesday, October 21, 2015 / Rules and Regulations
E. Effective Date and Congressional
Notification
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
31. This regulation is effective
December 21, 2015. The Commission
has determined, with concurrence of the
Administrator of the Office of
Information and Regulatory Affairs of
OMB, that this rule is not a ‘‘major rule’’
as defined in section 251 of the Small
Business Regulatory Enforcement
Fairness Act of 1996.41 This rule is
being submitted to the Senate, House,
and Government Accountability Office.
Food and Drug Administration
List of Subjects in 18 CFR Part 11
ACTION:
Electric power, Reporting and
recordkeeping requirements.
SUMMARY:
By the Commission.
Issued: October 15, 2015.
Kimberly D. Bose,
Secretary.
In consideration of the foregoing, the
Commission amends part 11, chapter I,
title 18, Code of Federal Regulations, as
follows:
PART 11—ANNUAL CHARGES UNDER
PART I OF THE FEDERAL POWER ACT
1. The authority citation for part 11
continues to read as follows:
■
Authority: 16 U.S.C. 792–828c; 42 U.S.C.
7101–7352.
2. Revise § 11.1(c)(5) to read as
follows:
■
§ 11.1
Costs of administration.
tkelley on DSK3SPTVN1PROD with RULES
*
*
*
*
*
(c) * * *
(5) For unconstructed projects, the
assessments begin on the date by which
the licensee or exemptee is required to
commence project construction, or as
that deadline may be extended, but in
no case longer than four years after the
issuance date of the license or
exemption. For constructed projects, the
assessments begin on the effective date
of the license or exemption, except for
any new capacity authorized therein.
The assessments for new authorized
capacity at licensed or exempted
projects begin on the date by which the
licensee or exemptee is required to
commence construction of the new
capacity. In the event that assessments
begin during a fiscal year, the charges
will be prorated.
*
*
*
*
*
[FR Doc. 2015–26726 Filed 10–20–15; 8:45 am]
BILLING CODE 6717–01–P
41 5
U.S.C. 804(2) (2012).
VerDate Sep<11>2014
16:24 Oct 20, 2015
Jkt 238001
21 CFR Part 870
[Docket No. FDA–2015–N–3387]
Medical Devices; Cardiovascular
Devices; Classification of the Coronary
Vascular Physiologic Simulation
Software Device
AGENCY:
Food and Drug Administration,
HHS.
Final order.
The Food and Drug
Administration (FDA) is classifying the
coronary vascular physiologic
simulation software device into class II
(special controls). The special controls
that will apply to the device are
identified in this order and will be part
of the codified language for the coronary
vascular physiologic simulation
software device’s classification. The
Agency is classifying the device into
class II (special controls) in order to
provide a reasonable assurance of safety
and effectiveness of the device.
DATES: This order is effective October
21, 2015. The classification was
applicable on November 26, 2014.
FOR FURTHER INFORMATION CONTACT:
Shawn Forrest, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1326, Silver Spring,
MD 20993–0002, 301–796–5554.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C.
360c(f)(1)), devices that were not in
commercial distribution before May 28,
1976 (the date of enactment of the
Medical Device Amendments of 1976),
generally referred to as postamendments
devices, are classified automatically by
statute into class III without any FDA
rulemaking process. These devices
remain in class III and require
premarket approval, unless and until
the device is classified or reclassified
into class I or II, or FDA issues an order
finding the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part
807 (21 CFR part 807) of the regulations.
PO 00000
Frm 00005
Fmt 4700
Sfmt 4700
63671
Section 513(f)(2) of the FD&C Act, as
amended by section 607 of the Food and
Drug Administration Safety and
Innovation Act (Pub. L. 112–144),
provides two procedures by which a
person may request FDA to classify a
device under the criteria set forth in
section 513(a)(1). Under the first
procedure, the person submits a
premarket notification under section
510(k) of the FD&C Act for a device that
has not previously been classified and,
within 30 days of receiving an order
classifying the device into class III
under section 513(f)(1) of the FD&C Act,
the person requests a classification
under section 513(f)(2). Under the
second procedure, rather than first
submitting a premarket notification
under section 510(k) of the FD&C Act
and then a request for classification
under the first procedure, the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence and requests a classification
under section 513(f)(2) of the FD&C Act.
If the person submits a request to
classify the device under this second
procedure, FDA may decline to
undertake the classification request if
FDA identifies a legally marketed device
that could provide a reasonable basis for
review of substantial equivalence with
the device or if FDA determines that the
device submitted is not of ‘‘lowmoderate risk’’ or that general controls
would be inadequate to control the risks
and special controls to mitigate the risks
cannot be developed.
In response to a request to classify a
device under either procedure provided
by section 513(f)(2) of the FD&C Act,
FDA will classify the device by written
order within 120 days. This
classification will be the initial
classification of the device.
On November 6, 2013, HeartFlow, Inc.
submitted a request for classification of
the FFRCT v.1.4 under section 513(f)(2)
of the FD&C Act. The manufacturer
recommended that the device be
classified into class II (Ref. 1).
In accordance with section 513(f)(2) of
the FD&C Act, FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1). FDA classifies
devices into class II if general controls
by themselves are insufficient to
provide reasonable assurance of safety
and effectiveness, but there is sufficient
information to establish special controls
to provide reasonable assurance of the
safety and effectiveness of the device for
its intended use. After review of the
information submitted in the request,
FDA determined that the device can be
classified into class II with the
E:\FR\FM\21OCR1.SGM
21OCR1
63672
Federal Register / Vol. 80, No. 203 / Wednesday, October 21, 2015 / Rules and Regulations
tkelley on DSK3SPTVN1PROD with RULES
establishment of special controls. FDA
believes these special controls, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on November 26, 2014,
FDA issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 870.1415.
Following the effective date of this
final classification order, any firm
submitting a premarket notification
(510(k)) for a coronary vascular
physiologic simulation software device
will need to comply with the special
controls named in this final order.
The device is assigned the generic
name coronary vascular physiologic
simulation software device, and it is
identified as a prescription device that
provides simulated functional
assessment of blood flow in the
coronary vascular system using data
extracted from medical device imaging
to solve algorithms and yield simulated
metrics of physiologic information (e.g.,
blood flow, coronary flow reserve,
fractional flow reserve, myocardial
perfusion). A coronary vascular
physiologic simulation software device
is intended to generate results for use
and review by a qualified clinician.
FDA has identified the following risks
to health associated with this type of
device, as well as the mitigation
measures required to mitigate these
risks, in table 1.
Æ Statistical methods must consider
TABLE 1—CORONARY VASCULAR
PHYSIOLOGIC SIMULATION SOFT- the predefined endpoints;
D Estimates of probabilities of
WARE DEVICE RISKS AND MITIGAincorrect results must be provided for
TION MEASURES—Continued
Identified risk
Failure to properly interpret device results leads to incorrect patient management.
Mitigation measure
Human factors testing.
Labeling.
FDA believes that the following
special controls, in combination with
the general controls, address these risks
to health and provide reasonable
assurance of safety and effectiveness:
• Adequate software verification and
validation based on comprehensive
hazard analysis with identification of
appropriate mitigations, must be
performed including:
Æ Full characterization of the
technical parameters of the software,
including any proprietary algorithm(s)
used to model the vascular anatomy;
D Adequate description of the
expected impact of all applicable image
acquisition hardware features and
characteristics on performance and any
associated minimum specifications;
Æ Adequate consideration of privacy
and security issues in the system design;
and
D Adequate mitigation of the impact
of failure of any subsystem components
(e.g., signal detection and analysis, data
storage, system communications and
cybersecurity) with respect to incorrect
patient reports and operator failures.
• Adequate non-clinical performance
TABLE 1—CORONARY VASCULAR
testing must be provided to demonstrate
PHYSIOLOGIC SIMULATION SOFT- the validity of computational modeling
WARE DEVICE RISKS AND MITIGA- methods for flow measurement.
TION MEASURES
• Clinical data supporting the
proposed intended use must be
Identified risk
Mitigation measure
provided, including the following:
Æ Output measure(s) must be
False negative results Software verification,
compared to a clinically acceptable
improperly indivalidation, and haz- method and must adequately represent
cating diseased
ard analysis.
the simulated measure(s) the device
vessel as low prob- Non-clinical performprovides in an accurate and
ability for significant
ance testing.
reproducible manner;
disease leads to
Æ Clinical utility of the device
delay of further
measurement accuracy must be
evaluation/treatdemonstrated by comparison to that of
ment.
other available diagnostic tests (e.g.,
False positive results
Clinical testing.
improperly indifrom literature analysis);
cating diseased
Æ Statistical performance of the
vessel as high
device within clinical risk strata (e.g.,
probability for sigage, relevant comorbidities, disease
nificant disease
stability) must be reported;
leads to incorrect
Æ The dataset must be adequately
patient managerepresentative of the intended use
ment.
population for the device (e.g., patients,
Delayed delivery of
Consistency (repeatrange of vessel sizes, imaging device
results leading to
ability/reproducmodels). Any selection criteria or
delay of further
ibility) evaluation.
limitations of the samples must be fully
evaluation/treatLabeling.
described and justified;
ment.
VerDate Sep<11>2014
16:24 Oct 20, 2015
Jkt 238001
PO 00000
Frm 00006
Fmt 4700
Sfmt 4700
each endpoint;
D Where multiple samples from the
same patient are used, statistical
analysis must not assume statistical
independence without adequate
justification;
D The report must provide
appropriate confidence intervals for
each performance metric;
Æ Sensitivity and specificity must be
characterized across the range of
available measurements;
Æ Agreement of the simulated
measure(s) with clinically acceptable
measure(s) must be assessed across the
full range of measurements;
Æ Comparison of the measurement
performance must be provided across
the range of intended image acquisition
hardware; and
Æ If the device uses a cutoff threshold
or operates across a spectrum of disease,
it must be established prior to validation
and it must be justified as to how it was
determined and clinically validated.
• Adequate validation must be
performed and controls implemented to
characterize and ensure consistency
(i.e., repeatability and reproducibility)
of measurement outputs;
Æ Acceptable incoming image quality
control measures and the resulting
image rejection rate for the clinical data
must be specified;
Æ Data must be provided within the
clinical validation study or using
equivalent datasets demonstrating the
consistency (i.e., repeatability and
reproducibility) of the output that is
representative of the range of data
quality likely to be encountered in the
intended use population and relevant
use conditions in the intended use
environment;
D Testing must be performed using
multiple operators meeting planned
qualification criteria and using the
procedure that will be implemented in
the production use of the device; and
D The factors (e.g., medical imaging
data set, operator) must be identified
regarding which were held constant and
which were varied during the
evaluation, and a description must be
provided for the computations and
statistical analyses used to evaluate the
data.
• Human factors evaluation and
validation must be provided to
demonstrate adequate performance of
the user interface to allow for users to
accurately measure intended
parameters, particularly where
parameter settings that have impact on
E:\FR\FM\21OCR1.SGM
21OCR1
tkelley on DSK3SPTVN1PROD with RULES
Federal Register / Vol. 80, No. 203 / Wednesday, October 21, 2015 / Rules and Regulations
measurements require significant user
intervention.
• Device labeling must be provided
that adequately describes the following:
Æ The device’s intended use,
including the type of imaging data used,
what the device measures and outputs
to the user, whether the measure is
qualitative or quantitative, the clinical
indications for which it is to be used,
and the specific population for which
the device use is intended;
Æ Appropriate warnings specifying
the intended patient population,
identifying anatomy and image
acquisition factors that may impact
measurement results, and providing
cautionary guidance for interpretation of
the provided measurements;
Æ Key assumptions made in the
calculation and determination of
simulated measurements;
Æ The measurement performance of
the device for all presented parameters,
with appropriate confidence intervals,
and the supporting evidence for this
performance. Per-vessel clinical
performance, including where
applicable localized performance
according to vessel and segment, must
be included as well as a characterization
of the measurement error across the
expected range of measurement for key
parameters based on the clinical data;
Æ A detailed description of the
patients studied in the clinical
validation (e.g., age, gender, race or
ethnicity, clinical stability, current
treatment regimen) as well as
procedural details of the clinical study
(e.g., scanner representation, calcium
scores, use of beta-blockers or nitrates);
and
Æ Where significant human interface
is necessary for accurate analysis,
adequately detailed description of the
analysis procedure using the device and
any data features that could affect
accuracy of results.
A coronary vascular physiologic
simulation software device is not safe
for use except under the supervision of
a practitioner licensed by law to direct
the use of the device. As such, the
device is a prescription device and must
satisfy prescription labeling
requirements (see 21 CFR 801.109,
Prescription devices).
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
is necessary to provide reasonable
VerDate Sep<11>2014
16:24 Oct 20, 2015
Jkt 238001
assurance of the safety and effectiveness
of the device. Therefore, this device
type is not exempt from premarket
notification requirements. Persons who
intend to market this type of device
must submit to FDA a premarket
notification, prior to marketing the
device, which contains information
about the coronary vascular physiologic
simulation software device they intend
to market.
II. Environmental Impact, No
Significant Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
III. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
part 807, subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120, and the collections of
information in 21 CFR part 801,
regarding labeling have been approved
under OMB control number 0910–0485.
IV. Reference
The following reference has been
placed on display in the Division of
Dockets Management (HFA–305), Food
and Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852,
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and is available
electronically at https://
www.regulations.gov.
1. DEN130045: De Novo Request from
HeartFlow, Inc., dated November 1,
2013.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 870 is
amended as follows:
PART 870—CARDIOVASCULAR
DEVICES
1. The authority citation for 21 CFR
part 870 continues to read as follows:
■
PO 00000
Frm 00007
Fmt 4700
Sfmt 4700
63673
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Add § 870.1415 to subpart B to read
as follows:
■
§ 870.1415 Coronary vascular physiologic
simulation software device.
(a) Identification. A coronary vascular
physiologic simulation software device
is a prescription device that provides
simulated functional assessment of
blood flow in the coronary vascular
system using data extracted from
medical device imaging to solve
algorithms and yield simulated metrics
of physiologic information (e.g., blood
flow, coronary flow reserve, fractional
flow reserve, myocardial perfusion). A
coronary vascular physiologic
simulation software device is intended
to generate results for use and review by
a qualified clinician.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Adequate software verification and
validation based on comprehensive
hazard analysis, with identification of
appropriate mitigations, must be
performed, including:
(i) Full characterization of the
technical parameters of the software,
including:
(A) Any proprietary algorithm(s) used
to model the vascular anatomy; and
(B) Adequate description of the
expected impact of all applicable image
acquisition hardware features and
characteristics on performance and any
associated minimum specifications;
(ii) Adequate consideration of privacy
and security issues in the system design;
and
(iii) Adequate mitigation of the impact
of failure of any subsystem components
(e.g., signal detection and analysis, data
storage, system communications and
cybersecurity) with respect to incorrect
patient reports and operator failures.
(2) Adequate non-clinical
performance testing must be provided to
demonstrate the validity of
computational modeling methods for
flow measurement; and
(3) Clinical data supporting the
proposed intended use must be
provided, including the following:
(i) Output measure(s) must be
compared to a clinically acceptable
method and must adequately represent
the simulated measure(s) the device
provides in an accurate and
reproducible manner;
(ii) Clinical utility of the device
measurement accuracy must be
demonstrated by comparison to that of
other available diagnostic tests (e.g.,
from literature analysis);
E:\FR\FM\21OCR1.SGM
21OCR1
tkelley on DSK3SPTVN1PROD with RULES
63674
Federal Register / Vol. 80, No. 203 / Wednesday, October 21, 2015 / Rules and Regulations
(iii) Statistical performance of the
device within clinical risk strata (e.g.,
age, relevant comorbidities, disease
stability) must be reported;
(iv) The dataset must be adequately
representative of the intended use
population for the device (e.g., patients,
range of vessel sizes, imaging device
models). Any selection criteria or
limitations of the samples must be fully
described and justified;
(v) Statistical methods must consider
the predefined endpoints:
(A) Estimates of probabilities of
incorrect results must be provided for
each endpoint,
(B) Where multiple samples from the
same patient are used, statistical
analysis must not assume statistical
independence without adequate
justification, and
(C) The report must provide
appropriate confidence intervals for
each performance metric;
(vi) Sensitivity and specificity must
be characterized across the range of
available measurements;
(vii) Agreement of the simulated
measure(s) with clinically acceptable
measure(s) must be assessed across the
full range of measurements;
(viii) Comparison of the measurement
performance must be provided across
the range of intended image acquisition
hardware; and
(ix) If the device uses a cutoff
threshold or operates across a spectrum
of disease, it must be established prior
to validation, and it must be justified as
to how it was determined and clinically
validated;
(4) Adequate validation must be
performed and controls implemented to
characterize and ensure consistency
(i.e., repeatability and reproducibility)
of measurement outputs:
(i) Acceptable incoming image quality
control measures and the resulting
image rejection rate for the clinical data
must be specified, and
(ii) Data must be provided within the
clinical validation study or using
equivalent datasets demonstrating the
consistency (i.e., repeatability and
reproducibility) of the output that is
representative of the range of data
quality likely to be encountered in the
intended use population and relevant
use conditions in the intended use
environment;
(A) Testing must be performed using
multiple operators meeting planned
qualification criteria and using the
procedure that will be implemented in
the production use of the device, and
(B) The factors (e.g., medical imaging
dataset, operator) must be identified
regarding which were held constant and
which were varied during the
VerDate Sep<11>2014
16:24 Oct 20, 2015
Jkt 238001
evaluation, and a description must be
provided for the computations and
statistical analyses used to evaluate the
data;
(5) Human factors evaluation and
validation must be provided to
demonstrate adequate performance of
the user interface to allow for users to
accurately measure intended
parameters, particularly where
parameter settings that have impact on
measurements require significant user
intervention; and
(6) Device labeling must be provided
that adequately describes the following:
(i) The device’s intended use,
including the type of imaging data used,
what the device measures and outputs
to the user, whether the measure is
qualitative or quantitative, the clinical
indications for which it is to be used,
and the specific population for which
the device use is intended;
(ii) Appropriate warnings specifying
the intended patient population,
identifying anatomy and image
acquisition factors that may impact
measurement results, and providing
cautionary guidance for interpretation of
the provided measurements;
(iii) Key assumptions made in the
calculation and determination of
simulated measurements;
(iv) The measurement performance of
the device for all presented parameters,
with appropriate confidence intervals,
and the supporting evidence for this
performance. Per-vessel clinical
performance, including where
applicable localized performance
according to vessel and segment, must
be included as well as a characterization
of the measurement error across the
expected range of measurement for key
parameters based on the clinical data;
(v) A detailed description of the
patients studied in the clinical
validation (e.g., age, gender, race or
ethnicity, clinical stability, current
treatment regimen) as well as
procedural details of the clinical study
(e.g., scanner representation, calcium
scores, use of beta-blockers or nitrates);
and
(vi) Where significant human
interface is necessary for accurate
analysis, adequately detailed
description of the analysis procedure
using the device and any data features
that could affect accuracy of results.
Dated: October 14, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015–26658 Filed 10–20–15; 8:45 am]
BILLING CODE 4164–01–P
PO 00000
Frm 00008
Fmt 4700
Sfmt 4700
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
33 CFR Parts 100, 117, 147, and 165
[USCG–2015–0242]
Quarterly Listings; Safety Zones,
Security Zones, Special Local
Regulations, Drawbridge Operation
Regulations and Regulated Navigation
Areas
Coast Guard, DHS.
Notice of expired temporary
rules issued.
AGENCY:
ACTION:
This document provides
notice of substantive rules issued by the
Coast Guard that were made temporarily
effective between January 2015 and
March 2015 but expired before they
could be published in the Federal
Register. This notice lists temporary
safety zones, security zones, special
local regulations, drawbridge operation
regulations and regulated navigation
areas, all of limited duration and for
which timely publication in the Federal
Register was not possible.
DATES: This document lists temporary
Coast Guard rules that became effective
between January 2015 and March 2015
and were terminated before they could
be published in the Federal Register.
ADDRESSES: Temporary rules listed in
this document may be viewed online,
under their respective docket numbers,
using the Federal eRulemaking Portal at
https://www.regulations.gov.
FOR FURTHER INFORMATION CONTACT: For
questions on this notice contact Yeoman
First Class Maria Fiorella Villanueva,
Office of Regulations and
Administrative Law, telephone (202)
372–3862.
SUPPLEMENTARY INFORMATION: Coast
Guard District Commanders and
Captains of the Port (COTP) must be
immediately responsive to the safety
and security needs within their
jurisdiction; therefore, District
Commanders and COTPs have been
delegated the authority to issue certain
local regulations. Safety zones may be
established for safety or environmental
purposes. A safety zone may be
stationary and described by fixed limits
or it may be described as a zone around
a vessel in motion. Security zones limit
access to prevent injury or damage to
vessels, ports, or waterfront facilities.
Special local regulations are issued to
enhance the safety of participants and
spectators at regattas and other marine
events. Drawbridge operation
regulations authorize changes to
SUMMARY:
E:\FR\FM\21OCR1.SGM
21OCR1
Agencies
[Federal Register Volume 80, Number 203 (Wednesday, October 21, 2015)]
[Rules and Regulations]
[Pages 63671-63674]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2015-26658]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA-2015-N-3387]
Medical Devices; Cardiovascular Devices; Classification of the
Coronary Vascular Physiologic Simulation Software Device
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
coronary vascular physiologic simulation software device into class II
(special controls). The special controls that will apply to the device
are identified in this order and will be part of the codified language
for the coronary vascular physiologic simulation software device's
classification. The Agency is classifying the device into class II
(special controls) in order to provide a reasonable assurance of safety
and effectiveness of the device.
DATES: This order is effective October 21, 2015. The classification was
applicable on November 26, 2014.
FOR FURTHER INFORMATION CONTACT: Shawn Forrest, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1326, Silver Spring, MD 20993-0002, 301-796-5554.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of the Federal Food, Drug, and
Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were
not in commercial distribution before May 28, 1976 (the date of
enactment of the Medical Device Amendments of 1976), generally referred
to as postamendments devices, are classified automatically by statute
into class III without any FDA rulemaking process. These devices remain
in class III and require premarket approval, unless and until the
device is classified or reclassified into class I or II, or FDA issues
an order finding the device to be substantially equivalent, in
accordance with section 513(i) of the FD&C Act, to a predicate device
that does not require premarket approval. The Agency determines whether
new devices are substantially equivalent to predicate devices by means
of premarket notification procedures in section 510(k) of the FD&C Act
(21 U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act, as amended by section 607 of the
Food and Drug Administration Safety and Innovation Act (Pub. L. 112-
144), provides two procedures by which a person may request FDA to
classify a device under the criteria set forth in section 513(a)(1).
Under the first procedure, the person submits a premarket notification
under section 510(k) of the FD&C Act for a device that has not
previously been classified and, within 30 days of receiving an order
classifying the device into class III under section 513(f)(1) of the
FD&C Act, the person requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a premarket
notification under section 510(k) of the FD&C Act and then a request
for classification under the first procedure, the person determines
that there is no legally marketed device upon which to base a
determination of substantial equivalence and requests a classification
under section 513(f)(2) of the FD&C Act. If the person submits a
request to classify the device under this second procedure, FDA may
decline to undertake the classification request if FDA identifies a
legally marketed device that could provide a reasonable basis for
review of substantial equivalence with the device or if FDA determines
that the device submitted is not of ``low-moderate risk'' or that
general controls would be inadequate to control the risks and special
controls to mitigate the risks cannot be developed.
In response to a request to classify a device under either
procedure provided by section 513(f)(2) of the FD&C Act, FDA will
classify the device by written order within 120 days. This
classification will be the initial classification of the device.
On November 6, 2013, HeartFlow, Inc. submitted a request for
classification of the FFRCT v.1.4 under section 513(f)(2) of
the FD&C Act. The manufacturer recommended that the device be
classified into class II (Ref. 1).
In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed
the request in order to classify the device under the criteria for
classification set forth in section 513(a)(1). FDA classifies devices
into class II if general controls by themselves are insufficient to
provide reasonable assurance of safety and effectiveness, but there is
sufficient information to establish special controls to provide
reasonable assurance of the safety and effectiveness of the device for
its intended use. After review of the information submitted in the
request, FDA determined that the device can be classified into class II
with the
[[Page 63672]]
establishment of special controls. FDA believes these special controls,
in addition to general controls, will provide reasonable assurance of
the safety and effectiveness of the device.
Therefore, on November 26, 2014, FDA issued an order to the
requestor classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 870.1415.
Following the effective date of this final classification order,
any firm submitting a premarket notification (510(k)) for a coronary
vascular physiologic simulation software device will need to comply
with the special controls named in this final order.
The device is assigned the generic name coronary vascular
physiologic simulation software device, and it is identified as a
prescription device that provides simulated functional assessment of
blood flow in the coronary vascular system using data extracted from
medical device imaging to solve algorithms and yield simulated metrics
of physiologic information (e.g., blood flow, coronary flow reserve,
fractional flow reserve, myocardial perfusion). A coronary vascular
physiologic simulation software device is intended to generate results
for use and review by a qualified clinician.
FDA has identified the following risks to health associated with
this type of device, as well as the mitigation measures required to
mitigate these risks, in table 1.
Table 1--Coronary Vascular Physiologic Simulation Software Device Risks
and Mitigation Measures
------------------------------------------------------------------------
Identified risk Mitigation measure
------------------------------------------------------------------------
False negative results improperly Software verification,
indicating diseased vessel as low validation, and hazard
probability for significant disease leads analysis.
to delay of further evaluation/treatment. Non-clinical performance
testing.
False positive results improperly Clinical testing.
indicating diseased vessel as high
probability for significant disease leads
to incorrect patient management.
Delayed delivery of results leading to Consistency (repeatability/
delay of further evaluation/treatment. reproducibility)
evaluation.
Labeling.
Failure to properly interpret device Human factors testing.
results leads to incorrect patient Labeling.
management.
------------------------------------------------------------------------
FDA believes that the following special controls, in combination
with the general controls, address these risks to health and provide
reasonable assurance of safety and effectiveness:
Adequate software verification and validation based on
comprehensive hazard analysis with identification of appropriate
mitigations, must be performed including:
[cir] Full characterization of the technical parameters of the
software, including any proprietary algorithm(s) used to model the
vascular anatomy;
[ssquf] Adequate description of the expected impact of all
applicable image acquisition hardware features and characteristics on
performance and any associated minimum specifications;
[cir] Adequate consideration of privacy and security issues in the
system design; and
[ssquf] Adequate mitigation of the impact of failure of any
subsystem components (e.g., signal detection and analysis, data
storage, system communications and cybersecurity) with respect to
incorrect patient reports and operator failures.
Adequate non-clinical performance testing must be provided
to demonstrate the validity of computational modeling methods for flow
measurement.
Clinical data supporting the proposed intended use must be
provided, including the following:
[cir] Output measure(s) must be compared to a clinically acceptable
method and must adequately represent the simulated measure(s) the
device provides in an accurate and reproducible manner;
[cir] Clinical utility of the device measurement accuracy must be
demonstrated by comparison to that of other available diagnostic tests
(e.g., from literature analysis);
[cir] Statistical performance of the device within clinical risk
strata (e.g., age, relevant comorbidities, disease stability) must be
reported;
[cir] The dataset must be adequately representative of the intended
use population for the device (e.g., patients, range of vessel sizes,
imaging device models). Any selection criteria or limitations of the
samples must be fully described and justified;
[cir] Statistical methods must consider the predefined endpoints;
[ssquf] Estimates of probabilities of incorrect results must be
provided for each endpoint;
[ssquf] Where multiple samples from the same patient are used,
statistical analysis must not assume statistical independence without
adequate justification;
[ssquf] The report must provide appropriate confidence intervals
for each performance metric;
[cir] Sensitivity and specificity must be characterized across the
range of available measurements;
[cir] Agreement of the simulated measure(s) with clinically
acceptable measure(s) must be assessed across the full range of
measurements;
[cir] Comparison of the measurement performance must be provided
across the range of intended image acquisition hardware; and
[cir] If the device uses a cutoff threshold or operates across a
spectrum of disease, it must be established prior to validation and it
must be justified as to how it was determined and clinically validated.
Adequate validation must be performed and controls
implemented to characterize and ensure consistency (i.e., repeatability
and reproducibility) of measurement outputs;
[cir] Acceptable incoming image quality control measures and the
resulting image rejection rate for the clinical data must be specified;
[cir] Data must be provided within the clinical validation study or
using equivalent datasets demonstrating the consistency (i.e.,
repeatability and reproducibility) of the output that is representative
of the range of data quality likely to be encountered in the intended
use population and relevant use conditions in the intended use
environment;
[ssquf] Testing must be performed using multiple operators meeting
planned qualification criteria and using the procedure that will be
implemented in the production use of the device; and
[ssquf] The factors (e.g., medical imaging data set, operator) must
be identified regarding which were held constant and which were varied
during the evaluation, and a description must be provided for the
computations and statistical analyses used to evaluate the data.
Human factors evaluation and validation must be provided
to demonstrate adequate performance of the user interface to allow for
users to accurately measure intended parameters, particularly where
parameter settings that have impact on
[[Page 63673]]
measurements require significant user intervention.
Device labeling must be provided that adequately describes
the following:
[cir] The device's intended use, including the type of imaging data
used, what the device measures and outputs to the user, whether the
measure is qualitative or quantitative, the clinical indications for
which it is to be used, and the specific population for which the
device use is intended;
[cir] Appropriate warnings specifying the intended patient
population, identifying anatomy and image acquisition factors that may
impact measurement results, and providing cautionary guidance for
interpretation of the provided measurements;
[cir] Key assumptions made in the calculation and determination of
simulated measurements;
[cir] The measurement performance of the device for all presented
parameters, with appropriate confidence intervals, and the supporting
evidence for this performance. Per-vessel clinical performance,
including where applicable localized performance according to vessel
and segment, must be included as well as a characterization of the
measurement error across the expected range of measurement for key
parameters based on the clinical data;
[cir] A detailed description of the patients studied in the
clinical validation (e.g., age, gender, race or ethnicity, clinical
stability, current treatment regimen) as well as procedural details of
the clinical study (e.g., scanner representation, calcium scores, use
of beta-blockers or nitrates); and
[cir] Where significant human interface is necessary for accurate
analysis, adequately detailed description of the analysis procedure
using the device and any data features that could affect accuracy of
results.
A coronary vascular physiologic simulation software device is not
safe for use except under the supervision of a practitioner licensed by
law to direct the use of the device. As such, the device is a
prescription device and must satisfy prescription labeling requirements
(see 21 CFR 801.109, Prescription devices).
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act if FDA determines that premarket notification is
not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary to provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, this device type is not exempt from premarket notification
requirements. Persons who intend to market this type of device must
submit to FDA a premarket notification, prior to marketing the device,
which contains information about the coronary vascular physiologic
simulation software device they intend to market.
II. Environmental Impact, No Significant Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
III. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in
part 807, subpart E, regarding premarket notification submissions have
been approved under OMB control number 0910-0120, and the collections
of information in 21 CFR part 801, regarding labeling have been
approved under OMB control number 0910-0485.
IV. Reference
The following reference has been placed on display in the Division
of Dockets Management (HFA-305), Food and Drug Administration, 5630
Fishers Lane, Rm. 1061, Rockville, MD 20852, and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday,
and is available electronically at https://www.regulations.gov.
1. DEN130045: De Novo Request from HeartFlow, Inc., dated November
1, 2013.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
870 is amended as follows:
PART 870--CARDIOVASCULAR DEVICES
0
1. The authority citation for 21 CFR part 870 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Add Sec. 870.1415 to subpart B to read as follows:
Sec. 870.1415 Coronary vascular physiologic simulation software
device.
(a) Identification. A coronary vascular physiologic simulation
software device is a prescription device that provides simulated
functional assessment of blood flow in the coronary vascular system
using data extracted from medical device imaging to solve algorithms
and yield simulated metrics of physiologic information (e.g., blood
flow, coronary flow reserve, fractional flow reserve, myocardial
perfusion). A coronary vascular physiologic simulation software device
is intended to generate results for use and review by a qualified
clinician.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Adequate software verification and validation based on
comprehensive hazard analysis, with identification of appropriate
mitigations, must be performed, including:
(i) Full characterization of the technical parameters of the
software, including:
(A) Any proprietary algorithm(s) used to model the vascular
anatomy; and
(B) Adequate description of the expected impact of all applicable
image acquisition hardware features and characteristics on performance
and any associated minimum specifications;
(ii) Adequate consideration of privacy and security issues in the
system design; and
(iii) Adequate mitigation of the impact of failure of any subsystem
components (e.g., signal detection and analysis, data storage, system
communications and cybersecurity) with respect to incorrect patient
reports and operator failures.
(2) Adequate non-clinical performance testing must be provided to
demonstrate the validity of computational modeling methods for flow
measurement; and
(3) Clinical data supporting the proposed intended use must be
provided, including the following:
(i) Output measure(s) must be compared to a clinically acceptable
method and must adequately represent the simulated measure(s) the
device provides in an accurate and reproducible manner;
(ii) Clinical utility of the device measurement accuracy must be
demonstrated by comparison to that of other available diagnostic tests
(e.g., from literature analysis);
[[Page 63674]]
(iii) Statistical performance of the device within clinical risk
strata (e.g., age, relevant comorbidities, disease stability) must be
reported;
(iv) The dataset must be adequately representative of the intended
use population for the device (e.g., patients, range of vessel sizes,
imaging device models). Any selection criteria or limitations of the
samples must be fully described and justified;
(v) Statistical methods must consider the predefined endpoints:
(A) Estimates of probabilities of incorrect results must be
provided for each endpoint,
(B) Where multiple samples from the same patient are used,
statistical analysis must not assume statistical independence without
adequate justification, and
(C) The report must provide appropriate confidence intervals for
each performance metric;
(vi) Sensitivity and specificity must be characterized across the
range of available measurements;
(vii) Agreement of the simulated measure(s) with clinically
acceptable measure(s) must be assessed across the full range of
measurements;
(viii) Comparison of the measurement performance must be provided
across the range of intended image acquisition hardware; and
(ix) If the device uses a cutoff threshold or operates across a
spectrum of disease, it must be established prior to validation, and it
must be justified as to how it was determined and clinically validated;
(4) Adequate validation must be performed and controls implemented
to characterize and ensure consistency (i.e., repeatability and
reproducibility) of measurement outputs:
(i) Acceptable incoming image quality control measures and the
resulting image rejection rate for the clinical data must be specified,
and
(ii) Data must be provided within the clinical validation study or
using equivalent datasets demonstrating the consistency (i.e.,
repeatability and reproducibility) of the output that is representative
of the range of data quality likely to be encountered in the intended
use population and relevant use conditions in the intended use
environment;
(A) Testing must be performed using multiple operators meeting
planned qualification criteria and using the procedure that will be
implemented in the production use of the device, and
(B) The factors (e.g., medical imaging dataset, operator) must be
identified regarding which were held constant and which were varied
during the evaluation, and a description must be provided for the
computations and statistical analyses used to evaluate the data;
(5) Human factors evaluation and validation must be provided to
demonstrate adequate performance of the user interface to allow for
users to accurately measure intended parameters, particularly where
parameter settings that have impact on measurements require significant
user intervention; and
(6) Device labeling must be provided that adequately describes the
following:
(i) The device's intended use, including the type of imaging data
used, what the device measures and outputs to the user, whether the
measure is qualitative or quantitative, the clinical indications for
which it is to be used, and the specific population for which the
device use is intended;
(ii) Appropriate warnings specifying the intended patient
population, identifying anatomy and image acquisition factors that may
impact measurement results, and providing cautionary guidance for
interpretation of the provided measurements;
(iii) Key assumptions made in the calculation and determination of
simulated measurements;
(iv) The measurement performance of the device for all presented
parameters, with appropriate confidence intervals, and the supporting
evidence for this performance. Per-vessel clinical performance,
including where applicable localized performance according to vessel
and segment, must be included as well as a characterization of the
measurement error across the expected range of measurement for key
parameters based on the clinical data;
(v) A detailed description of the patients studied in the clinical
validation (e.g., age, gender, race or ethnicity, clinical stability,
current treatment regimen) as well as procedural details of the
clinical study (e.g., scanner representation, calcium scores, use of
beta-blockers or nitrates); and
(vi) Where significant human interface is necessary for accurate
analysis, adequately detailed description of the analysis procedure
using the device and any data features that could affect accuracy of
results.
Dated: October 14, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015-26658 Filed 10-20-15; 8:45 am]
BILLING CODE 4164-01-P