General Hospital and Personal Use Devices: Renaming of Pediatric Hospital Bed Classification and Designation of Special Controls for Pediatric Medical Crib; Classification of Medical Bassinet, 60809-60817 [2015-25627]
Download as PDF
60809
Proposed Rules
Federal Register
Vol. 80, No. 195
Thursday, October 8, 2015
This section of the FEDERAL REGISTER
contains notices to the public of the proposed
issuance of rules and regulations. The
purpose of these notices is to give interested
persons an opportunity to participate in the
rule making prior to the adoption of the final
rules.
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA–2015–N–0701]
General Hospital and Personal Use
Devices: Renaming of Pediatric
Hospital Bed Classification and
Designation of Special Controls for
Pediatric Medical Crib; Classification
of Medical Bassinet
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed rule.
The Food and Drug
Administration (FDA) is proposing to
rename pediatric hospital beds as
pediatric medical cribs and establish
special controls for these devices. FDA
is also proposing to establish a separate
classification regulation for medical
bassinets, previously under the
pediatric hospital bed classification
regulation, as a class II (special controls)
device. The proposed regulation for
both pediatric medical cribs and
medical bassinets would also include
the Consumer Product Safety
Commission’s (CPSC) mattress
flammability standards for the
mattresses intended for use with these
devices. In addition, this proposed rule
would require prescription use of
pediatric medical cribs and bassinets.
DATES: Submit either electronic or
written comments by December 7, 2015.
See section VII of this document for the
proposed effective date of a final rule
based on this proposed rule.
ADDRESSES: You may submit comments
as follows:
Lhorne on DSK5TPTVN1PROD with PROPOSALS
SUMMARY:
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Comments submitted electronically,
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
including attachments, to https://
www.regulations.gov will be posted to
the docket unchanged. Because your
comment will be made public, you are
solely responsible for ensuring that your
comment does not include any
confidential information that you or a
third party may not wish to be posted,
such as medical information, your or
anyone else’s Social Security number, or
confidential business information, such
as a manufacturing process. Please note
that if you include your name, contact
information, or other information that
identifies you in the body of your
comments, that information will be
posted on https://www.regulations.gov.
• If you want to submit a comment
with confidential information that you
do not wish to be made available to the
public, submit the comment as a
written/paper submission and in the
manner detailed (see ‘‘Written/Paper
Submissions’’ and ‘‘Instructions’’).
Written/Paper Submissions
Submit written/paper submissions as
follows:
• Mail/Hand delivery/Courier (for
written/paper submissions): Division of
Dockets Management (HFA–305), Food
and Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
• For written/paper comments
submitted to the Division of Dockets
Management, FDA will post your
comment, as well as any attachments,
except for information submitted,
marked and identified, as confidential,
if submitted as detailed in
‘‘Instructions.’’
Instructions: All submissions received
must include the Docket No. FDA–
2015–N–0701 for ‘‘General Hospital and
Personal Use Devices: Renaming of
Pediatric Hospital Bed Classification
and Designation of Special Controls for
Pediatric Medical Crib; Classification of
Medical Bassinet.’’ Received comments
will be placed in the docket and, except
for those submitted as ‘‘Confidential
Submissions,’’ publicly viewable at
https://www.regulations.gov or at the
Division of Dockets Management
between 9 a.m. and 4 p.m., Monday
through Friday.
• Confidential Submissions—To
submit a comment with confidential
information that you do not wish to be
made publicly available, submit your
comments only as a written/paper
submission. You should submit two
copies total. One copy will include the
PO 00000
Frm 00001
Fmt 4702
Sfmt 4702
information you claim to be confidential
with a heading or cover note that states
‘‘THIS DOCUMENT CONTAINS
CONFIDENTIAL INFORMATION’’. The
Agency will review this copy, including
the claimed confidential information, in
its consideration of comments. The
second copy, which will have the
claimed confidential information
redacted/blacked out, will be available
for public viewing and posted on
https://www.regulations.gov. Submit
both copies to the Division of Dockets
Management. If you do not wish your
name and contact information to be
made publicly available, you can
provide this information on the cover
sheet and not in the body of your
comments and you must identify this
information as ‘‘confidential.’’ Any
information marked as ‘‘confidential’’
will not be disclosed except in
accordance with 21 CFR 10.20 and other
applicable disclosure law. For more
information about FDA’s posting of
comments to public dockets, see 80 FR
56469, September 18, 2015, or access
the information at: https://www.fda.gov/
regulatoryinformation/dockets/
default.htm.
Docket: For access to the docket to
read background documents or the
electronic and written/paper comments
received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Division of Dockets
Management, 5630 Fishers Lane, Rm.
1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Michael J. Ryan, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1615, Silver Spring,
MD 20993–0002, 301–796–6283.
SUPPLEMENTARY INFORMATION:
I. Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C. 301 et
seq.), as amended by the Medical Device
Amendments of 1976 (the 1976
amendments) (Pub. L. 94–295), the Safe
Medical Device Amendments of 1990
(SMDA) (Pub. L. 101–629), the Food and
Drug Administration Modernization Act
of 1997 (Pub. L. 105–115), the Food and
Drug Administration Amendments Act
of 2007 (Pub. L. 110–85), and the FDA
Safety and Innovation Act (Pub. L. 112–
E:\FR\FM\08OCP1.SGM
08OCP1
Lhorne on DSK5TPTVN1PROD with PROPOSALS
60810
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
144), among other amendments,
establishes a comprehensive system for
the regulation of medical devices
intended for human use. Section 513 of
the FD&C Act (21 U.S.C. 360c)
establishes three categories (classes) of
devices, based on the regulatory
controls needed to provide reasonable
assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Most generic types of devices that
were on the market before May 28, 1976,
the date of the 1976 amendments
(generally referred to as preamendments
devices), have been classified by FDA
through the issuance of regulations in
accordance with the procedures set
forth in section 513(c) and (d) of the
FD&C Act into one of these three
regulatory classes. Devices introduced
into interstate commerce for the first
time on or after May 28, 1976 (generally
referred to as post-amendments
devices), are classified automatically by
statute (section 513(f) of the FD&C Act)
into class III without any FDA
rulemaking process. These devices
remain in class III and require
premarket approval, unless FDA
initiates one of the following
procedures: (1) FDA reclassifies the
device into class I or II; (2) FDA issues
an order classifying the device into class
I or II in accordance with section
513(f)(2) of the FD&C Act; or (3) FDA
issues an order finding the device to be
substantially equivalent, under section
513(i), to a predicate device that is
already legally marketed. The Agency
determines whether new devices are
substantially equivalent to predicate
devices through review of premarket
notifications under section 510(k) of the
FD&C Act (21 U.S.C. 360(k)). Section
510(k) of the FD&C Act and its
implementing regulations, codified in
Title 21 of the Code of Federal
Regulations (21 CFR part 807, subpart
E), require persons who intend to
market a new device that does not
require a premarket approval
application under section 515 of the
FD&C Act (21 U.S.C. 360e) to submit a
premarket notification report (510(k))
containing information that allows FDA
to determine whether the new device is
‘‘substantially equivalent’’ within the
meaning of section 513(i) of the FD&C
Act to a legally marketed device that
does not require premarket approval.
Section 513(a)(1)(B) of the FD&C Act
defines class II devices as those devices
for which the general controls in section
513(a)(1)(A) by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
but for which there is sufficient
information to establish special controls
to provide such assurance, including the
issuance of performance standards,
postmarket surveillance, patient
registries, development and
dissemination of guidelines,
recommendations, and any other
appropriate actions the Agency deems
necessary to provide such assurance
(see also 21 CFR 860.3(c)(2)).
Section 510(m)(2) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements on its own initiative or
upon petition of an interested person, if
FDA determines that a 510(k) is not
necessary to provide reasonable
assurance of the safety and effectiveness
of the device. Devices under the
pediatric hospital bed classification
regulation were exempted from
premarket notification, subject to certain
limitations, in accordance with section
510(m) of the FD&C Act (63 FR 59222
at 59229, November 3, 1998).
II. Regulatory History and Description
of the Devices
FDA classified pediatric hospital beds
(21 CFR 880.5140) as class II devices (45
FR 69678 at 69694, October 21, 1980),
and later exempted them from
premarket notification (510(k)), in a
final rule published in the Federal
Register of November 3, 1998 (63 FR
59222 at 59229). In § 880.5140, a
pediatric hospital bed is defined as ‘‘a
device intended for medical purposes
that consists of a bed or crib designed
for the use of a pediatric patient, with
fixed end rails and movable and
latchable side rails. The contour of the
bed surface may be adjustable.’’
A medical bassinet is a non-powered
device that consists of two components:
(1) A basket, the sleep or bed
component, which is typically made of
plastic and (2) a durable frame with
wheels, which holds the basket or bed
component (FDA refers to this
component as a ‘‘basket or bed
component’’ in this proposed rule). The
basket or bed component is a box-like
structure, generally made of a clear,
high-impact resistant plastic material,
with an open top and four stationary
walls to keep the baby in place. Medical
bassinets are typically used in hospital
settings for infants up to 5 months in
age. Medical bassinets currently fall
under the pediatric hospital bed
classification regulation.
III. Proposed Regulation
Pediatric medical cribs that meet the
definition of a device in section 201(h)
of the FD&C Act (21 U.S.C. 321(h)) are
regulated by FDA (referred to as
PO 00000
Frm 00002
Fmt 4702
Sfmt 4702
pediatric medical cribs or cribs intended
for medical purposes) (product code
FMS) and, if this rule is finalized, will
have to comply with the special controls
identified in the final regulation for
pediatric medical cribs. Cribs outside of
the device definition (referred to as cribs
for non-medical purposes) must meet
the CPSC’s regulations and guidelines.
A crib designed for the use of a pediatric
patient may meet the medical device
definition if it is intended for use in the
cure, mitigation or treatment of disease
(see section 201(h) of the FD&C Act).
In the Federal Register of December
28, 2010 (75 FR 81766), the CPSC issued
a final rule prohibiting the use of the
drop-side rail design for non-medical
cribs in consumer households as of June
28, 2011. Child care facilities, family
child care homes, and places of public
accommodation (e.g., hotels and motels)
had to comply with the rule as of
December 28, 2012. CPSC’s rule
establishes new standards for full-size
and non-full-size cribs used for nonmedical purposes, which effectively
prohibit the manufacture or sale of cribs
for non-medical purposes with a dropside rail design in households, child
care facilities, family child care homes,
and places of public accommodation.
This rule did not affect pediatric
medical cribs regulated by FDA, which
may contain a drop-side rail design that
includes movable and latchable side
and end rails.
Because drop-side rail cribs for nonmedical purposes and pediatric medical
cribs are regulated by different agencies,
CPSC consulted with FDA about the
impact their final rule could have on
settings, such as nursery schools and
day care centers, where pediatric
medical cribs with drop-side rails are
often used for pediatric patients after
they have been discharged from a health
care facility. CPSC, which regulates
consumer products, including drop-side
rail cribs not intended for medical
purposes, received reports of deaths of
children attributable to entrapment and/
or strangulation caused by the
malfunctioning of drop-side rail cribs.
Although drop-side cribs for nonmedical purposes are now prohibited,
there is still a need for pediatric medical
cribs with drop-side rails inside and
outside of traditional health care
settings. CPSC and FDA have heard
from medical device consumers and
health care providers that pediatric
medical cribs with drop-side rails are
extremely helpful for patient care in
hospital settings and even outside of
traditional health care settings, such as
day care centers caring for infants and
children with disabilities, because they
allow parents and care givers easy
E:\FR\FM\08OCP1.SGM
08OCP1
Lhorne on DSK5TPTVN1PROD with PROPOSALS
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
access to children to perform routine
and emergency medical procedures,
including, but not limited to, CPR,
blood collection, IV insertion,
respiratory care, and skin care. These
drop-side rail cribs also make it easier
for hospital staff to facilitate safe patient
transport and reduce the chance of
caregiver injury. Health care workers
have stated that they need to have
continued access to these medical cribs
with drop-side rails (Ref. 1). Therefore,
FDA is proposing to permit
manufacturers to continue to
manufacture and sell medical cribs with
the drop-side rail design in traditional
health care settings and to permit the
use of pediatric medical cribs with
drop-side rail designs outside of
traditional health care settings through
prescription use only (it is noted that
State child care licensing agencies are
generally responsible for overseeing day
care providers while FDA is responsible
for medical devices).
FDA is proposing to revise the
identification in § 880.5140 to include
only pediatric medical cribs, establish
special controls for this device, and
change the name of the classification
regulation from ‘‘pediatric hospital bed’’
to ‘‘pediatric medical crib.’’ The Agency
is taking these actions to clarify the
devices that fall under this particular
classification regulation and establish
special controls the Agency believes are
necessary for a reasonable assurance of
safety and effectiveness. In addition,
FDA is proposing that use of a pediatric
medical crib be restricted to
prescription use in accordance with 21
CFR 801.109. In order to use or
administer use of pediatric medical
cribs, authorization must be made by a
practitioner licensed by law through a
prescription for the device.
This rule also proposes to create a
separate regulation for medical bassinets
and establish special controls for this
device type to provide a reasonable
assurance of safety and effectiveness. In
addition, FDA is proposing that use of
medical bassinets be restricted to
prescription use in accordance with 21
CFR 801.109. In order to use or
administer use of medical bassinets,
authorization must be made by a
practitioner licensed by law through a
prescription for the device. FDA
proposes not to change the 510(k)
exempt status of pediatric medical cribs
and medical bassinets.
Devices currently under the pediatric
hospital bed classification regulation
include: Open pediatric medical cribs,
medical bassinets, pediatric cribs with
integrated air mattresses, youth beds,
pediatric stretchers, crib enclosure beds,
and cuddle-carrier infant beds. If this
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
proposed rule is finalized, devices that
do not meet the definition of ‘‘pediatric
medical crib’’ will be administratively
moved to more appropriate class II
regulations, and no longer be under the
revised pediatric hospital bed
classification regulation. At that time,
FDA proposes to send manufacturers of
the remaining pediatric hospital beds
notices identifying the new
classification regulation and product
code under which the device will be
classified.
If this proposed rule is finalized, FDA
intends to move the following medical
devices listed under § 880.5140 to
devices with similar intended uses and
class II regulations: Pediatric cribs with
integrated air mattresses to 21 CFR
890.5170, ‘‘Powered flotation therapy
bed;’’ youth beds to either 21 CFR
880.5100, ‘‘AC powered adjustable
hospital bed,’’ or 21 CFR 880.5120,
‘‘Manual adjustable hospital bed,’’
depending on whether they are powered
or not; pediatric stretchers to 21 CFR
880.6910, ‘‘Wheeled stretchers;’’ and
crib enclosure beds to 21 CFR 880.6760,
‘‘Protective restraint.’’ This action
would not have any substantive effect
on the current marketing status of the
devices. However, manufacturers of
these devices would need to refer to the
new regulation classification and
product code provided by the Agency in
future interactions with FDA.
As discussed in section IV, an
analysis of Medical Device Reports
(MDRs) submitted to the Manufacturer
and User Facility Device Experience
(MAUDE) database from January 1,
2005, to September 1, 2015, indicated
516 adverse events associated with
pediatric medical cribs including 15
serious injuries. The adverse events
associated with pediatric medical cribs
were assessed to better understand the
risks and establish the proposed special
controls for this device. FDA believes
that sufficient information is available
to establish special controls to provide
a reasonable assurance of the safety and
effectiveness of the device.
As discussed further in section VI,
FDA believes risks to health resulting
from use of these cribs would be
effectively mitigated by the special
controls proposed in this rule, and that
these controls, in combination with the
general controls, would provide a
reasonable assurance of safety and
effectiveness for pediatric medical cribs
for their intended use. Therefore, FDA
is proposing new safety requirements
and allowing medical cribs in homes
and day cares only when medically
necessary.
FDA is also taking this opportunity to
address adverse event reports pertaining
PO 00000
Frm 00003
Fmt 4702
Sfmt 4702
60811
to medical bassinets by proposing to
establish special controls for these
devices to provide a reasonable
assurance of safety and effectiveness.
FDA has received adverse events from
hospitals regarding incidents of medical
bassinet tipping and improper cleaning
of the basket or bed component that
caused cracks and crazing, which have
resulted in patient injury. The Agency is
proposing to separate medical bassinets
from other types of pediatric hospital
beds to allow for more targeted
postmarket surveillance of these
devices. FDA believes the special
controls it is proposing here, in
combination with the general controls,
would provide a reasonable assurance of
safety and effectiveness for medical
bassinets.
IV. Risks to Health
A. Pediatric Medical Crib
Between January 1, 2005, and
September 1, 2015, FDA received 516
adverse event reports, or MDRs,
associated with open pediatric medical
cribs, through the Agency’s MAUDE
database. There were 15 adverse event
reports of serious injuries including 6
reports of entrapment, which were
predominantly extremity entrapments of
legs or arms. The majority of MDRs for
medical cribs were for malfunctions
such as drop-side rails not latching or
lowering, brakes not holding, wheels or
casters breaking, and where applicable,
scales not reading correct weights.
These malfunctions (501 reports) were
not associated with any adverse health
effects. After considering available
information, FDA determined that the
following risks to health are associated
with the use of pediatric medical cribs:
• Injury resulting from mechanical or
structural failure of the device—
Mechanical or structural failure of the
crib can result in failure of load-bearing
components such as the wheels or
casters, or failure of the latches or other
locking mechanisms that secure the
sides of the crib. These failures can
result in injuries, as demonstrated by
the MDRs received in FDA’s database.
• Pinching, laceration, splinters, and
foreign body ingestion—Depending on
the material of the pediatric crib, certain
cribs may peel or crack and may expose
pediatric patients to substances or
materials that may be toxic or may cause
abrasions or lacerations if the surface of
the crib material is compromised.
• Entrapment, falls, and
strangulation—Pediatric medical cribs
may cause entrapment of patient limbs
if the width of the side rails are not
correct and if there are gaps between the
mattress and crib frame that are larger
E:\FR\FM\08OCP1.SGM
08OCP1
60812
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
than the width of two fingers.
Depending on the height requirements
of the rails a pediatric patient may
escape or fall from the crib. The term
‘‘entrapment’’ refers to circumstances
where a patient is caught, trapped, or
entangled in the space in or about the
bed rail, mattress, or hospital bed frame.
• Burns—Certain flammable materials
used in the construction of pediatric
medical cribs may allow for the spread
of fire, which may result in serious
injuries. Fires can spread easily in
hospital rooms with a patient using
oxygen. The free-flowing oxygen can
intensify a fire, which can rapidly
spread to flammable objects in the room,
including crib mattresses.
• Use error—Use error may contribute
to or exacerbate any of the previously
mentioned adverse events. For example,
a user may be unaware that a side rail
did not latch, or may fail to properly
maintain a pediatric medical crib.
Therefore, adequate instructions for use
and user education are essential to safe
device operation.
B. Medical Bassinet
Between January 1, 2005, and
September 1, 2015, FDA received 40
adverse event reports associated with
this device type. The most common
MDRs for medical bassinets include
reports of malfunctions such as casters
or wheels not working, which have
caused tipping, and broken bassinet
base components, such as doors and
drawers, or collapse or breakage of
utility shelves or chart holders. There
are also reports of the plastic sleep
basket or bed component crazing
(cracking), resulting in sharp edges and
cuts to hospital personnel.
FDA has considered the available
information and determined that the
following risks to health are associated
with medical bassinets:
• Injury resulting from mechanical or
structural failure of the device—
Mechanical or structural failure of the
bassinet can result in failure of loadbearing components such as the wheels
or casters, or failure of the latches or
other locking mechanisms that secure
the drawers of the bassinet. These
failures can result in injuries,
particularly if the bassinet tips over, as
demonstrated by the MDRs received in
FDA’s database.
• Burns—Certain flammable materials
used in the construction of pediatric
medical bassinets may allow for the
spread of fire, which may result in
serious injuries. Fires can spread easily
in hospital rooms with a patient using
oxygen. The free-flowing oxygen can
intensify a fire, which can rapidly
spread to flammable objects in the room,
including bassinet mattresses.
• Crazing or cracking of basket or bed
component—The basket or bed
component of the bassinet that the
pediatric patient is placed in may craze
or crack due to improper care or
handling, such as cleaning the plastic
material of the basket or bed component
with inappropriate cleaning solutions.
Crazing or cracking may result in
injuries such as cuts.
• Use error—Use error may contribute
to, or exacerbate, any of the previously
mentioned risks. For example, a user
may accidentally leave a door or drawer
in the base component of the bassinet
open or place too much weight in a
drawer or on a shelf, which may present
a tipping hazard. Also, a user may fail
to properly maintain a medical bassinet.
V. Establishment of Special Controls
Under section 513(a)(1)(B) of the
FD&C Act, as amended by the SMDA,
class II devices are defined as devices
for which general controls by
themselves are insufficient to provide
reasonable assurance of safety and
effectiveness, but for which there is
sufficient information to establish
special controls to provide such
assurance. Special controls may include
the issuance of performance standards,
postmarket surveillance, patient
registries, development and
dissemination of guidelines,
recommendations, and other
appropriate actions the Agency deems
necessary to provide such assurance
(see also § 860.3(c)(2)).
Under this authority, FDA is
proposing to establish special controls
for pediatric medical cribs (§ 880.5140)
and pediatric medical bassinets (§ 880.
5145). The Agency believes that the
applicable special controls, together
with the general controls, would
provide reasonable assurance of the
safety and effectiveness of these devices.
VI. Proposed Special Controls
A. Pediatric Medical Crib
FDA consulted with health care
providers in children’s hospitals,
registered nurses in pediatric units,
biomedical engineers, and technicians,
and analyzed the associated adverse
events with pediatric medical cribs.
Specifically, FDA consulted with
MedSun hospitals regarding their 2011
survey on clinicians’ experiences with
pediatric medical cribs with drop-side
rails used in MedSun’s hospitals (Ref.
1). The MedSun survey summary
highlights the clinical perspective and
the importance of this device in medical
and health care settings. The most
common issues and concerns in the
survey were the lack of understanding
of side rail operation and the need for
reinforcing patient safety when the side
rails are raised or lowered. Many
respondents of the survey suggested
further improvements for pediatric
medical cribs, for instance, improved
labeling, specific distance between slats
and emergency releases on side rails for
faster access to pediatric patients. The
adverse events identified in the MedSun
survey are similar to the MDRs FDA has
received on this device. FDA believes
that the special controls proposed in
this proposed rule, in combination with
the general controls, would provide a
reasonable assurance of safety and
effectiveness for pediatric medical cribs
their intended use.
TABLE 1—HEALTH RISKS AND MITIGATION MEASURES FOR PEDIATRIC MEDICAL CRIB
Identified risks to health
Mitigation measures
Injury Resulting from Mechanical or Structural Failure of the Device .....
Lhorne on DSK5TPTVN1PROD with PROPOSALS
Pinching, Lacerations, Splinters, and Foreign Body Ingestion ................
Entrapment, Falls, and Strangulation .......................................................
Burns ........................................................................................................
Use Error ..................................................................................................
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
PO 00000
Frm 00004
Fmt 4702
Performance Testing.
Design Testing.
Appropriate Materials Free From Surface Defects.
Labeling.
Performance Testing.
Rail and End Panel Design.
Side Rail Spacing and Safety Features.
Appropriate Fitting of Mattress.
CPSC’s Mattress Flammability Standard.
Labeling.
Sfmt 4702
E:\FR\FM\08OCP1.SGM
08OCP1
Lhorne on DSK5TPTVN1PROD with PROPOSALS
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
As provided in Table 1, the Agency
believes the following special controls,
in combination with the general
controls, would effectively mitigate the
identified risks to health and provide
reasonable assurance of the safety and
effectiveness of the device:
1. Design and performance testing
must be conducted to ensure the
mechanical and structural stability of
the crib under expected conditions of
use, including the security of latches
and other locking mechanisms when
engaged. These requirements are
derived from sections 6.2 and 6.3 of
ASTM (formerly the American Society
for Testing and Materials) International
Standard F1169–13, entitled ‘‘Standard
Consumer Safety Specification for FullSize Baby Cribs’’ (Ref. 2) and sections
5.7 and 6.3 of ASTM International
Standard F2710–13, entitled ‘‘Standard
Consumer Safety Performance
Specification for Commercial Cribs’’
(Ref. 3), which was developed with
input from crib manufacturers.
2. To reduce possible injury of
pinching, lacerations, and crushing, the
crib shall be designed and constructed
in a manner that eliminates hardware
accessible to a child within the crib.
This requirement is derived from
section 5.10 of ASTM International
Standard F1169–13, entitled ‘‘Standard
Consumer Safety Specification for FullSize Baby Cribs’’ (Ref. 2), which was
developed with input from crib
manufacturers. Also, materials used
shall be appropriate for the conditions
of use, allow for proper sanitation, and
free from surface defects of the device
that could result in injuries.
3. To reduce the risk of head and limb
entrapment, the distance between side
rail components (such as slats, spindles,
corner posts, and rods) shall be
designed to reduce potential entrapment
of pediatric patients and the distance
between such components shall not
exceed 23⁄8 inches (6 centimeters) apart.
In addition, the rails and end panels of
a crib must be of a height to mitigate the
possibility of falls and/or escapes by the
patient. These requirements are derived
from sections 5.7.2 and 5.8.1 of ASTM
International Standard F1169–13,
entitled ‘‘Standard Consumer Safety
Specification for Full-Size Baby Cribs’’
(Ref. 2), which was developed with
input from crib manufacturers.
4. To reduce the risk of head and limb
entrapment, no gap shall exist between
the edge of the bottom rail and the top
of the mattress surface and the mattress
must fit tightly around all four sides of
the crib. These requirements are derived
from section 5.9 of ASTM International
Standard F1169–13, entitled ‘‘Standard
Consumer Safety Specification for FullSize Baby Cribs’’ (Ref. 2), which was
developed with input from crib
manufacturers.
5. To reduce flammability and the risk
of burns, the mattress for the crib shall
meet the CPSC Standard for the
Flammability of Mattresses and Mattress
Pads and its Standard for the
Flammability (Open Flame) of Mattress
Sets, 16 CFR parts 1632 and 1633,
respectively. This proposed special
control would clarify for manufacturers
the standards necessary for mattresses
intended to be used with pediatric
medical cribs to prevent the spread of
fires that can easily occur in hospital
rooms with a patient using oxygen. The
free-flowing oxygen can intensify a fire,
which can rapidly spread to most of the
flammable objects in the room
especially mattresses. The consumer
standards for flammability of mattresses
in 16 CFR parts 1632 and 1633 are also
accepted by the Joint Commission
(formerly the Joint Commission on
Accreditation of Healthcare
Organizations). CPSC’s mattress and
mattress pad flammability standard
under 16 CFR part 1632 addresses
mattress fires ignited by open flame
sources, including matches, candles,
lighters, and other related scenarios. It
prescribes a test to determine the
ignition resistance of a mattress or a
mattress pad when exposed to a lighted
cigarette. CPSC’s standard for the
flammability of mattress sets under 16
CFR part 1633 is a broader standard
designed to reduce deaths and injuries
caused by mattress fires, particularly
those fires ignited by, among others
things, oxygen use or electrical
equipment sources that may occur in a
patient’s room. In addition, CPSC’s
regulations require that manufacturers
meet an established fire safety
performance standard, based on ASTM
E2187–09, entitled ‘‘International’s
Standard Test Method for Measuring the
Ignition Strength of Cigarettes’’ (Ref. 4),
which was developed with input from
crib manufacturers.
6. To reduce flammability and the risk
of burns, the labeling must bear all
information required pursuant to the
CPSC Standard for the Flammability of
Mattresses and Mattress Pads and its
Standard for the Flammability (Open
Flame) of Mattress Sets, 16 CFR parts
1632 and 1633, respectively.
7. To reduce the risk of use error,
which may result in mechanical or
structural failure of the crib due to
inadequate care or maintenance,
pediatric medical crib labeling must
include adequate instructions for users
to care for and maintain their crib.
These requirements are derived from
sections 5.18 of ASTM International
Standard F1169–13, entitled ‘‘Standard
Consumer Safety Specification for FullSize Baby Cribs’’ (Ref. 2).
FDA believes that the special controls
proposed in this rule would provide a
reasonable assurance of safety and
effectiveness of pediatric medical cribs
in their intended use. The ASTM and
CPSC standards noted above apply to all
mattresses and mattress pads intended
or promoted for sleeping upon, as
defined in 16 CFR 1632.1(a), including
medical mattresses that are regulated by
FDA as an accessory to medical beds.
Therefore, FDA anticipates that
manufacturers would be able to meet
the requirements imposed by the
proposed special controls in this
proposed rule without undue burden.
FDA invites comments on this
conclusion, including comments
regarding the types of performance
testing manufacturers conduct for
pediatric medical cribs, particularly to
ensure the performance of medical crib
latches on drop-side rails.
In addition, FDA is proposing to
restrict these devices to prescription use
under section 520(e) of the FD&C Act
(see § 801.109 (prescription devices)). In
order to use or administer use of
pediatric medical cribs, authorization
must be made by a practitioner licensed
by law.
B. Medical Bassinet
Table 2 lists the risks to health FDA
has identified for Medical Bassinets, as
described in the Risks to Health, section
IV of this proposed rule, along with the
corresponding proposed mitigation
measures for each risk.
TABLE 2—HEALTH RISKS AND MITIGATION MEASURES FOR MEDICAL BASSINET
Identified risks to health
Mitigation measures
Injury Resulting from Mechanical or Structural Failure of the Device .....
Burns ........................................................................................................
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
PO 00000
Frm 00005
Fmt 4702
Performance Testing.
Labeling.
CPSC’s Mattress Flammability Standard.
Sfmt 4702
60813
E:\FR\FM\08OCP1.SGM
08OCP1
60814
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
TABLE 2—HEALTH RISKS AND MITIGATION MEASURES FOR MEDICAL BASSINET—Continued
Identified risks to health
Mitigation measures
Crazing or Cracking of Basket or Bed Component .................................
Use Error ..................................................................................................
The Agency believes the following
special controls, in combination with
the general controls, would effectively
mitigate the identified risks to health
and provide reasonable assurance of the
safety and effectiveness of medical
bassinets:
1. To mitigate crazing, cracking, and
deterioration of the basket or bed
component of the device, the
manufacturer must conduct
performance testing to determine
material compatibility with cleansing
products labeled to clean the device.
2. To reduce flammability and the risk
of burns, the bassinet shall meet CPSC’s
Standard for the Flammability of
Mattresses and Mattress Pads and its
Standard for the Flammability (Open
Flame) of Mattress Sets, 16 CFR parts
1632 and 1633, respectively.
3. To reduce the risk of injury
resulting from mechanical or structural
failure of the device, and particularly,
device tipping that can result from those
failures; manufacturers shall conduct
performance testing to ensure the
mechanical and structural stability of
the bassinet under expected use
conditions, including transport of
patients in the bassinet.
4. To reduce the risk of use error,
specifically error that may result in
bassinet tipping, FDA proposes that
manufacturers shall have a label on the
front of the bassinet cabinet with the
following warning statement:
Lhorne on DSK5TPTVN1PROD with PROPOSALS
WARNING: To avoid tipping hazards of
this device, make sure that the basket or bed
component sits firmly in the base and that all
doors, drawers, and casters are secure.
The label must be affixed to the front
of the bassinet base cabinet and the text
shall be in letters not less than 10
millimeters in height.
FDA believes this warning is
necessary because even if performance
testing demonstrates that a bassinet does
not present a tipping hazard under
expected use conditions, users may
exceed these expected use conditions,
particularly during transport of a patient
in the bassinet.
5. To reduce the risk of use error,
which may result in mechanical or
structural failure of the bassinet due to
inadequate care or maintenance,
medical bassinet labeling must include
adequate instructions for users to care
for and maintain the bassinet.
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
Performance Testing.
Labeling.
Labeling.
FDA believes that the special controls
proposed in this rule would provide a
reasonable assurance of safety and
effectiveness of medical bassinets in
their intended use. The CPSC standards
noted previously apply to all mattresses
and mattress pads intended or promoted
for sleeping upon, as defined in 16 CFR
1632.1(a), including medical mattresses
that are regulated by FDA as an
accessory to medical beds. Therefore,
FDA believes most manufacturers are
already complying with the proposed
special control for mattress flammability
set forth in this proposed rule. FDA
invites comments on the types of
performance testing manufacturers
conduct for medical bassinets.
In addition, FDA is proposing to
restrict these devices to prescription use
under section 520(e) of the FD&C Act
(see § 801.109 (Prescription devices)). In
order to use or administer use of
medical bassinets, authorization must
be made by a practitioner licensed by
law.
VII. Proposed Effective Date
FDA proposes that any final rule
based on this proposal become effective
60 days after its publication in the
Federal Register.
VIII. Environmental Impact, No
Significant Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
IX. Economic Analysis of Impacts
FDA has examined the impacts of the
proposed rule under Executive Order
12866, Executive Order 13563, the
Regulatory Flexibility Act (5 U.S.C.
601–612) and the Unfunded Mandates
Reform Act of 1995 (Pub. L. 104–4).
Executive Orders 12866 and 13563
direct Agencies to assess all costs and
benefits of available regulatory
alternatives and, when regulation is
necessary, to select regulatory
approaches that maximize net benefits
(including potential economic,
environmental, public health and safety,
and other advantages; distributive
PO 00000
Frm 00006
Fmt 4702
Sfmt 4702
impacts; and equity). The Agency
believes that this proposed rule is not a
significant regulatory action as defined
by Executive Order 12866.
The Regulatory Flexibility Act
requires Agencies to analyze regulatory
options that would minimize any
significant impact of a rule on small
entities. Because the expected costs
associated with this rule are expected to
be modest, we propose to certify that
this rule would not have a significant
economic impact on a substantial
number of small entities.
Section 202(a) of the Unfunded
Mandates Reform Act of 1995 requires
that Agencies prepare a written
statement, which includes an
assessment of anticipated costs and
benefits, before proposing ‘‘any rule that
includes any Federal mandate that may
result in the expenditure by State, local,
and tribal governments, in the aggregate,
or by the private sector, of $100,000,000
or more (adjusted annually for inflation)
in any one year.’’ The current threshold
after adjustment for inflation is $144
million, using the most current (2014)
Implicit Price Deflator for the Gross
Domestic Product. We do not expect
this proposed rule to result in any 1year expenditure that would meet or
exceed this amount.
A. Objective of the Rule
Pediatric hospital beds are classified
as class II, 510(k) exempt medical
devices intended for the treatment, care,
or diagnosis of diseases or illnesses of
pediatric patients. In this proposed rule,
FDA proposes to amend § 880.5140 by
revising the identification and
establishing special controls for
pediatric medical cribs. This rule would
also change the name of the
classification regulation from ‘‘pediatric
hospital bed’’ to ‘‘pediatric medical
crib,’’ and place medical bassinets,
previously under the pediatric hospital
beds classification regulation, as a
separate class II, 510(k) exempt device,
subject to its own special controls.
Pediatric medical cribs used in health
care settings contain a drop-side rail
design that includes movable and
latchable side and end rails. As stated
previously, the CPSC issued a final rule
prohibiting the use of the drop-side rail
design for non-medical cribs in
consumer households as of June 28,
E:\FR\FM\08OCP1.SGM
08OCP1
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
Lhorne on DSK5TPTVN1PROD with PROPOSALS
2011 (December 28, 2010, 75 FR 81766).
CPSC’s rule establishes new standards
for full-size and non-full-size cribs used
for non-medical purposes, which
effectively prohibit the manufacture or
sale of cribs for non-medical purposes
with a drop-side rail design in
households, child care facilities, family
child care homes, and places of public
accommodation. The compliance date
for this same rule for child care
facilities, family child care homes, and
places of public accommodation was
December 28, 2012 (75 FR 81766).
CPSC’s rule was established in response
to infant deaths that occurred when the
side rail of a crib used for non-medical
purposes detached or disengaged. In
contrast, there have been no deaths and
a few serious injuries reported to FDA
associated with pediatric medical cribs,
which are made of more durable
materials and construction in
comparison to cribs used for nonmedical purposes. Additionally, FDA
has determined that drop-side rails are
essential for patient care in hospital
settings and even outside of traditional
health care settings to allow parents and
care givers easy access to their patients
in order to perform both routine and
emergency medical procedures. To
address concerns raised by CPSC’s
reports on consumer drop-side rail cribs
and account for the medical need of this
device, FDA is proposing to establish
special controls and require a
prescription for this device when used
outside of traditional health care setting.
Thus, FDA has determined that cribs
with drop-side rail designs may remain
on the market when intended for
medical use, and can be used outside of
traditional health care settings through
prescription use only. In addition, this
proposed rule adds specific special
controls for medical bassinets to
mitigate health risks such as tipping of
the device and crazing of the plastic
basket or bed component. Regulation of
this device under a separate regulation
would also allow for more targeted post
market surveillance for this device.
B. Benefits
FDA’s Registration and Listing
database identifies 38 manufacturers of
medical cribs and bassinets. FDA has
reviewed the safety standards of several
large pediatric crib and bassinet
manufacturers in order to determine the
compliance burden associated with the
proposed special controls. The Agency
concludes that many of the special
controls proposed by this rule are
consistent with current industry
practice among many medical crib and
bassinet manufacturers. The proposed
special control that is not currently
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
practiced is the warning labeling
requirements for medical bassinets. For
new pediatric medical crib and bassinet
manufacturers entering the market or
manufacturers that may not be currently
following the practices required by the
proposed special controls, if this
proposed rule is finalized, its special
controls will clarify safety standards
and minimize the risk of injury to
pediatric patients.
The beneficial features of medical
bassinets are portability, ease of
cleaning, and, when it is made of a clear
material, the ability to see the baby from
all sides. The proposed special controls
would require bassinet manufacturers to
place labels on their devices warning
against device tipping. This requirement
would apply to new bassinets; bassinets
that have already been sold would not
be required to add the new labels to
their devices. The warning label is
intended to prevent tipping of the
device, which may be caused by
unlatched drawers, dislodged wheels, or
too much weight on the shelves. The
Agency has not received any reports of
death or serious injury related to
medical bassinets, although there have
been a small number of reports of
malfunctioning casters, which may
cause device tipping. The benefits of the
new warning label are not readily
quantifiable, but it is expected to reduce
the risk of the bassinet from tipping and
thus, reduce potential injury to pediatric
patients.
The provision allowing for the
medical cribs outside of traditional
health care settings would benefit
pediatric patients who require the
specialized care provided by these
devices outside of traditional health
care settings. Due to the CPSC rule
regarding cribs used for non-medical
purposes, discussed previously,
consumers and child care facilities are
restricted from using cribs with a dropside rail design. If this proposed rule is
finalized, it will allow consumers and
child care facilities to utilize the
pediatric medical cribs if they are
prescribed by a health care professional.
The special controls regarding the
mechanical structure of pediatric
medical cribs are intended to minimize
the risk of injury, including entrapment
or strangulation of pediatric patients.
The spacing specifications of the side
rail components are designed to prevent
head or neck entrapment and
strangulation incidents in which infants
may slip between the openings of the
slats, and the performance testing
requirements are designed to ensure the
side rail latches of pediatric medical
cribs will perform as intended and
remain secure when the latches are
PO 00000
Frm 00007
Fmt 4702
Sfmt 4702
60815
engaged. The special control requiring
specific height of the rails and end
panels may prevent falls and/or escapes
by the patient. Also, by having pediatric
medical crib manufacturers use
materials that are appropriate for the
conditions of use and allow for proper
sanitation, these special controls may
help mitigate surface defects that can
cause injury to the patient.
Additionally, the mattress size
standards for cribs and bassinets are
intended to reduce the risk of significant
gaps between the mattress and the
device structure, which could
potentially create an entrapment hazard.
The flammability standard is intended
to reduce deaths and injuries related to
mattress fires, particularly those
initially ignited by open flame sources
such as lighters, candles, and matches.
Although the practices proposed in
these special controls are believed to be
followed by almost all manufacturers of
products currently on the market, the
proposed special controls would
reinforce safety standards for such
manufacturers and ensure that other
manufacturers and manufacturers of
new products adhere to the same safety
standards.
C. Costs
The economic impact of the proposed
regulation is determined primarily by
whether manufacturers currently
comply with the proposed special
controls. As stated previously, the
special controls that are not currently
practiced by industry, of which FDA is
aware, are the bassinet warning labeling
and the performance testing
requirements. FDA is also aware that
many manufacturers of pediatric
medical cribs and medical bassinets
registered with the FDA currently
conform to the risk mitigations and
structural requirements that are being
proposed as special controls, and thus
conforming to these special controls, if
finalized, would not result in an
increase in cost to pediatric medical crib
manufacturers and only cause a small
increase in cost for medical bassinet
manufactures. Additionally, the
renaming of pediatric medical cribs and
redesignation in the CFR for medical
bassinets and the remaining devices
under the pediatric hospital bed
classification are administrative in
nature, and are not expected to result in
any cost burdens.
The new warning labeling
requirements for medical bassinets will
apply to manufacturers of new bassinets
only. FDA does not expect bassinets that
are currently on the market to be
relabeled. If manufacturers of new
bassinets add labels to the devices at the
E:\FR\FM\08OCP1.SGM
08OCP1
Lhorne on DSK5TPTVN1PROD with PROPOSALS
60816
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
time of production, the cost burden to
manufacturers would be minimized.
Although we do not have direct
estimates of labeling costs for these
devices, the best estimate of these costs
is derived from FDA’s labeling cost
model. Because FDA would require
specific language and format of the
labels, we consider this to be a minor
labeling change that would not require
label design, market tests, or analytical
tests. Labeling costs would include
labor and material, and are estimated to
be, on average, approximately $140 per
unit. Then we use the number of live
births per year as reported by the Center
for Disease Control and Prevention in
order to determine the number of
bassinets produced per year for medical
use (Ref. 5). Using an estimate of 4
million births per year and 11,000 births
per day, we estimate that each birth
requires an average hospital stay of 3
days. This yields a total supply of
approximately 33,000 medical bassinets
in the United States. Given an average
yearly replacement rate of 20 percent for
all medical bassinets, we estimate that
approximately 6,600 new bassinets will
be produced annually. Applying the
$140 per unit labeling cost yields a total
yearly cost of $924,000 associated with
the new bassinet warning label
requirement.
The special controls require
performance testing for medical
bassinets to reduce the risk of crazing of
the plastic basket or bed component. We
assume that the performance testing
may be conducted as an extension to
current product testing and may be
performed at the same testing facilities
currently utilized by bassinet
manufacturers. FDA projects that a
maximum of an additional week of
testing would be required. The costs
associated with the performance testing
include the labor costs of mechanical
engineers, who typically perform these
tests. The mean 2012 hourly wage for
mechanical engineers is $40.75, as
reported by the Occupational
Employment Statistics provided by the
Bureau of Labor Statistics (Ref. 6).
Applying a multiplier of 1.45 to adjust
for benefits, hourly labor costs are
estimated to be approximately $59.
Assuming a 40-hour work week, the
total maximum estimated cost for each
manufacturer to perform these
additional tests is approximately $2,360.
It is uncertain the exact number of
manufacturers that do not currently
conduct performance testing and would
therefore be required to extend current
testing practices. However, given the
relatively small number of medical
bassinet manufacturers, FDA anticipates
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
that even the upper-bound total cost
would be modest.
The prescription use of pediatric
medical cribs outside of traditional
health care settings may potentially
increase Medicaid spending for eligible
pediatric patients. According to our
review of Healthcare Common
Procedure Coding System billing codes
for the Medicaid program, currently,
States typically offer Medicaid coverage
for prescribed rental or purchase of
hospital beds and pediatric cribs (Ref.
7). We estimate the number of
additional prescriptions for pediatric
medical cribs to be filled annually as a
result of this proposed rule would be
less than 100. Medicaid expenditure on
pediatric medical cribs is estimated to
be on average $2,500 per device. This
yields a maximum annual total cost of
$250,000.
Although it is unlikely that these
devices would require physical
modification to meet the standards
proposed by the special controls in this
proposed rule, there may be
manufacturers on the market of which
we are unaware that do not conform to
the requirements proposed in the
special controls. The proposed special
controls could have a significant impact
on firms that are not currently in
compliance with the special controls, as
their products may require
modifications. The special control that
may cause additional costs for
manufacturers is the special control
concerning the mechanical structure of
pediatric medical cribs. We are not able
to estimate the actual compliance costs
for manufacturers of pediatric medical
cribs because such costs may vary by
firm size and the amount of
modification required. Alternatively, we
provide an estimate of the modification
cost by using aggregate industry market
price information and cost data. The
costs associated with these
modifications may include the costs
associated with product design and
testing, labor, material, and production.
We use data from the Annual Survey of
Manufacturers to calculate aggregate
labor and materials costs as a percentage
of total sales for manufacturers
represented by North American Industry
Classification System code 339113 (Ref.
8). The data indicate that labor and
materials represent approximately 45
percent of total sales. Allowing market
price to represent per unit revenue at
the firm level, we estimate the cost of
modification to be approximately 45
percent of the average price of a
pediatric medical crib. After surveying
market prices of pediatric medical cribs,
we estimate an average per unit price of
$2,500. This yields an average cost of
PO 00000
Frm 00008
Fmt 4702
Sfmt 4702
approximately $1,125 to modify a
pediatric medical crib to be in
compliance with the proposed special
controls.
FDA invites comments on the
compliance of manufacturers with the
special controls, including the
performance testing, mechanical
structure, flammability requirements,
and bassinet labeling requirements, as
well as cost information if modifications
are required.
X. Paperwork Reduction Act of 1995
The proposed rule refers to previously
approved collections of information
found in FDA regulations. These
collections of information are subject to
review by the Office of Management
Budget (OMB) and the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information,
regarding premarket notification
submissions (21 CFR part 807, subpart
E), are approved under OMB control
number 0910–0120. The collections of
information, regarding labeling (21 CFR
part 801), including prescription device
labeling and adequate directions for use,
are approved under OMB control
number 0910–0485. The collections of
information regarding current good
manufacturing practice quality systems
(21 CFR part 820), including design
controls (as referenced in proposed
§ 880.5140(b)(1) and proposed
§ 880.5145(b)(1) and (b)(3) of this
document), are approved under OMB
control number 0910–0073. The
collections of information in 16 CFR
1632 and 1633, regarding mattress
flammability, are approved under OMB
control number 3041–0014.
In addition, FDA concludes that the
warning label for bassinets does not
constitute a ‘‘collection of information’’
under the PRA. Rather, the labeling
statement is ‘‘public disclosure(s) of
information originally supplied by the
Federal government to the recipient for
the purpose of disclosure to the public.’’
(5 CFR 1320.3(c)(2)).
XI. References
The following references have been
placed on display in the Division of
Dockets Management (see ADDRESSES)
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and are available
electronically at https://
www.regulations.gov. (FDA has verified
the Web site addresses, but FDA is not
responsible for any subsequent changes
to the Web sites after this document
publishes in the Federal Register.)
1. MedSun: Newsletter #66: ‘‘Pediatric
Hospital Cribs: MedSun Small Sample
Survey Summary’’ (November 2011),
E:\FR\FM\08OCP1.SGM
08OCP1
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 / Proposed Rules
available at https://www.fda.gov/
downloads/MedicalDevices/Safety/
MedSunMedicalProductSafetyNetwork/
Newsletters/UCM422131.pdf.
2. ASTM International (formerly the
American Society for Testing and
Materials), ‘‘Standard Consumer Safety
Specifications for Full-Size Baby Cribs,’’
Designation: F1169–13, available at
https://www.astm.org/Standards/
F1169.htm, 2013.
3. ASTM International (formerly the
American Society for Testing and
Materials), ‘‘Standard Consumer Safety
Performance Specification for
Commercial Cribs,’’ Designation: F2710–
13, available at https://www.astm.org/
Standards/F2710.htm, 2013.
4. ASTM International, ‘‘Standard Test
Method for Measuring the Ignition
Strength of Cigarettes,’’ Designation:
E2187–09 Standard, available at https://
www.astm.org/Standards/E2187.htm.
5. Centers for Disease Control and Prevention
(CDC), National Vital Statistics System,
Birth Data, available at https://
www.cdc.gov/nchs/births.htm.
6. Occupational Employment Statistics
provided by the Bureau of Labor
Statistics, May 2012 Occupational
Employment Statistics, available at
https://www.bls.gov/oes/2012/may/oes_
stru.htm.
7. Centers for Medicare & Medicaid Services,
HCPCS 2015 Code: E0300, available at
https://www.cms.gov/Medicare/Coding/
MedHCPCSGenInfo/
index.html?redirect=/medhcpcsgeninfo/.
8. U.S. Census Bureau, 2013 Annual Survey
of Manufacturers, available at https://
www.census.gov/manufacturing/asm/
index.html.
2. Revise § 880.5140 to read as
follows:
premarket notification procedures in
subpart E of part 807 of this chapter
subject to § 880.9. The special controls
for this device are:
(1) Crib design and performance
testing shall demonstrate the
mechanical and structural stability of
the crib under expected conditions of
use, including the security of latches
and other locking mechanisms when
engaged;
(2) Materials used shall be appropriate
for the conditions of use, allow for
proper sanitation and free from surface
defects that could result in injuries;
(3) Rails and end panels shall be
designed taking into account the crib’s
height at its lowest point to the top of
the mattress to prevent patient falls and/
or escape. Hardware and fasteners shall
be designed and constructed to
eliminate mechanical hazards to the
patient;
(4) The distance between components
of the side rail (such as slats, spindles,
and corner posts) shall not be greater
than 23⁄8 inches (6 centimeters (cm))
apart at any point. Side rails shall
contain safety features for locking and
adjust the lowest position of the crib to
a height that shall be 20 inches (51 cm)
above the top of the mattress;
(5) The device shall not have a gap
between the bottom of the rail and the
top surface of the mattress and the
mattress pad must fit tightly around all
four sides of the crib;
(6) The mattress for the crib shall
meet the Consumer Product Safety
Commission (CPSC) Standard for the
Flammability of Mattresses and Mattress
Pads and Standard for the Flammability
(Open Flame) of Mattress Sets, 16 CFR
parts 1632 and 1633, respectively;
(7) The labeling must bear all
information required pursuant to the
CPSC Standard for the Flammability of
Mattresses and Mattress Pads and
Standard for the Flammability (Open
Flame) of Mattress Sets, 16 CFR parts
1632 and 1633, respectively; and
(8) Pediatric medical crib labeling
must include adequate instructions for
users to care for and maintain their crib.
■ 3. Add § 880.5145 to subpart F to read
as follows:
§ 880.5140
§ 880.5145
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 880 be amended as follows:
PART 880—GENERAL HOSPITAL AND
PERSONAL USE DEVICES
1. The authority citation for 21 CFR
part 880 continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
Lhorne on DSK5TPTVN1PROD with PROPOSALS
■
Pediatric medical crib.
(a) Identification. A pediatric medical
crib is a prescription device intended
for medical purposes for use with a
pediatric patient that consists of an
open crib, fixed-end rails, movable and
latchable side rail components, and
possibly an accompanying mattress. The
contour of the crib surface may be
adjustable.
(b) Classification. Class II (special
controls). The device is exempt from the
VerDate Sep<11>2014
14:48 Oct 07, 2015
Jkt 238001
Medical bassinet.
(a) Identification. A medical bassinet
is a prescription device that is a small
bed intended for use with pediatric
patients, generally from birth to
approximately 5 months of age. It is
intended for medical purposes for use in
a nursery, labor and delivery unit, or
patient room, but may also be used
outside of traditional health care
settings. A medical bassinet is a nonpowered device that consists of two
PO 00000
Frm 00009
Fmt 4702
Sfmt 9990
60817
components: The plastic basket or bed
component and a durable frame with
wheels, which holds the basket or bed
component. The basket or bed
component is a box-like structure,
generally made of a clear, high impactresistant plastic material, with an open
top and four stationary walls to hold the
pediatric patient. The frame can include
drawers, shelving or cabinetry that
provides space to hold baby care items.
The wheels or casters allow the bassinet
to transport the baby throughout the
care setting.
(b) Classification. Class II (special
controls). The device is exempt from the
premarket notification procedures in
subpart E of part 807 of this chapter
subject to § 880.9. The special controls
for this device are:
(1) The manufacturer must conduct
performance testing to determine
material compatibility with cleansing
products labeled to clean the device.
Testing must demonstrate that the
cleaning instructions provided by the
manufacturer do not cause crazing,
cracking, or deterioration of the device;
(2) The mattress for the device shall
meet the Consumer Product Safety
Commission Standard for the
Flammability of Mattresses and Mattress
Pads and Standard for the Flammability
(Open Flame) of Mattress Sets, 16 CFR
parts 1632 and 1633, respectively;
(3) Manufacturers shall conduct
performance testing to ensure the
mechanical and structural stability of
the bassinet under expected use
conditions, including transport of
patients in the bassinet. Testing must
demonstrate that failures such as wheel
or caster breakage do not occur, and that
the device does not present a tipping
hazard due to any mechanical failures,
under expected use conditions;
(4) Each device must have affixed a
label on the front of the bassinet cabinet
with the following language in text of at
least 10 millimeters in height:
WARNING: To avoid tipping hazards of
this device, make sure that the basket or bed
component sits firmly in the base and that all
doors, drawers, and casters are secure.
(5) Labeling must include adequate
instructions for users to care for and
maintain their bassinet.
Dated: October 2, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015–25627 Filed 10–7–15; 8:45 a.m.]
BILLING CODE 4164–01–P
E:\FR\FM\08OCP1.SGM
08OCP1
Agencies
[Federal Register Volume 80, Number 195 (Thursday, October 8, 2015)]
[Proposed Rules]
[Pages 60809-60817]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2015-25627]
========================================================================
Proposed Rules
Federal Register
________________________________________________________________________
This section of the FEDERAL REGISTER contains notices to the public of
the proposed issuance of rules and regulations. The purpose of these
notices is to give interested persons an opportunity to participate in
the rule making prior to the adoption of the final rules.
========================================================================
Federal Register / Vol. 80, No. 195 / Thursday, October 8, 2015 /
Proposed Rules
[[Page 60809]]
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA-2015-N-0701]
General Hospital and Personal Use Devices: Renaming of Pediatric
Hospital Bed Classification and Designation of Special Controls for
Pediatric Medical Crib; Classification of Medical Bassinet
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed rule.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is proposing to rename
pediatric hospital beds as pediatric medical cribs and establish
special controls for these devices. FDA is also proposing to establish
a separate classification regulation for medical bassinets, previously
under the pediatric hospital bed classification regulation, as a class
II (special controls) device. The proposed regulation for both
pediatric medical cribs and medical bassinets would also include the
Consumer Product Safety Commission's (CPSC) mattress flammability
standards for the mattresses intended for use with these devices. In
addition, this proposed rule would require prescription use of
pediatric medical cribs and bassinets.
DATES: Submit either electronic or written comments by December 7,
2015. See section VII of this document for the proposed effective date
of a final rule based on this proposed rule.
ADDRESSES: You may submit comments as follows:
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments. Comments submitted
electronically, including attachments, to https://www.regulations.gov
will be posted to the docket unchanged. Because your comment will be
made public, you are solely responsible for ensuring that your comment
does not include any confidential information that you or a third party
may not wish to be posted, such as medical information, your or anyone
else's Social Security number, or confidential business information,
such as a manufacturing process. Please note that if you include your
name, contact information, or other information that identifies you in
the body of your comments, that information will be posted on https://www.regulations.gov.
If you want to submit a comment with confidential
information that you do not wish to be made available to the public,
submit the comment as a written/paper submission and in the manner
detailed (see ``Written/Paper Submissions'' and ``Instructions'').
Written/Paper Submissions
Submit written/paper submissions as follows:
Mail/Hand delivery/Courier (for written/paper
submissions): Division of Dockets Management (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
For written/paper comments submitted to the Division of
Dockets Management, FDA will post your comment, as well as any
attachments, except for information submitted, marked and identified,
as confidential, if submitted as detailed in ``Instructions.''
Instructions: All submissions received must include the Docket No.
FDA-2015-N-0701 for ``General Hospital and Personal Use Devices:
Renaming of Pediatric Hospital Bed Classification and Designation of
Special Controls for Pediatric Medical Crib; Classification of Medical
Bassinet.'' Received comments will be placed in the docket and, except
for those submitted as ``Confidential Submissions,'' publicly viewable
at https://www.regulations.gov or at the Division of Dockets Management
between 9 a.m. and 4 p.m., Monday through Friday.
Confidential Submissions--To submit a comment with
confidential information that you do not wish to be made publicly
available, submit your comments only as a written/paper submission. You
should submit two copies total. One copy will include the information
you claim to be confidential with a heading or cover note that states
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION''. The Agency will
review this copy, including the claimed confidential information, in
its consideration of comments. The second copy, which will have the
claimed confidential information redacted/blacked out, will be
available for public viewing and posted on https://www.regulations.gov.
Submit both copies to the Division of Dockets Management. If you do not
wish your name and contact information to be made publicly available,
you can provide this information on the cover sheet and not in the body
of your comments and you must identify this information as
``confidential.'' Any information marked as ``confidential'' will not
be disclosed except in accordance with 21 CFR 10.20 and other
applicable disclosure law. For more information about FDA's posting of
comments to public dockets, see 80 FR 56469, September 18, 2015, or
access the information at: https://www.fda.gov/regulatoryinformation/dockets/default.htm.
Docket: For access to the docket to read background documents or
the electronic and written/paper comments received, go to https://www.regulations.gov and insert the docket number, found in brackets in
the heading of this document, into the ``Search'' box and follow the
prompts and/or go to the Division of Dockets Management, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Michael J. Ryan, Center for Devices
and Radiological Health, Food and Drug Administration, 10903 New
Hampshire Ave., Bldg. 66, Rm. 1615, Silver Spring, MD 20993-0002, 301-
796-6283.
SUPPLEMENTARY INFORMATION:
I. Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C.
301 et seq.), as amended by the Medical Device Amendments of 1976 (the
1976 amendments) (Pub. L. 94-295), the Safe Medical Device Amendments
of 1990 (SMDA) (Pub. L. 101-629), the Food and Drug Administration
Modernization Act of 1997 (Pub. L. 105-115), the Food and Drug
Administration Amendments Act of 2007 (Pub. L. 110-85), and the FDA
Safety and Innovation Act (Pub. L. 112-
[[Page 60810]]
144), among other amendments, establishes a comprehensive system for
the regulation of medical devices intended for human use. Section 513
of the FD&C Act (21 U.S.C. 360c) establishes three categories (classes)
of devices, based on the regulatory controls needed to provide
reasonable assurance of their safety and effectiveness. The three
categories of devices are class I (general controls), class II (special
controls), and class III (premarket approval).
Most generic types of devices that were on the market before May
28, 1976, the date of the 1976 amendments (generally referred to as
preamendments devices), have been classified by FDA through the
issuance of regulations in accordance with the procedures set forth in
section 513(c) and (d) of the FD&C Act into one of these three
regulatory classes. Devices introduced into interstate commerce for the
first time on or after May 28, 1976 (generally referred to as post-
amendments devices), are classified automatically by statute (section
513(f) of the FD&C Act) into class III without any FDA rulemaking
process. These devices remain in class III and require premarket
approval, unless FDA initiates one of the following procedures: (1) FDA
reclassifies the device into class I or II; (2) FDA issues an order
classifying the device into class I or II in accordance with section
513(f)(2) of the FD&C Act; or (3) FDA issues an order finding the
device to be substantially equivalent, under section 513(i), to a
predicate device that is already legally marketed. The Agency
determines whether new devices are substantially equivalent to
predicate devices through review of premarket notifications under
section 510(k) of the FD&C Act (21 U.S.C. 360(k)). Section 510(k) of
the FD&C Act and its implementing regulations, codified in Title 21 of
the Code of Federal Regulations (21 CFR part 807, subpart E), require
persons who intend to market a new device that does not require a
premarket approval application under section 515 of the FD&C Act (21
U.S.C. 360e) to submit a premarket notification report (510(k))
containing information that allows FDA to determine whether the new
device is ``substantially equivalent'' within the meaning of section
513(i) of the FD&C Act to a legally marketed device that does not
require premarket approval.
Section 513(a)(1)(B) of the FD&C Act defines class II devices as
those devices for which the general controls in section 513(a)(1)(A) by
themselves are insufficient to provide reasonable assurance of safety
and effectiveness, but for which there is sufficient information to
establish special controls to provide such assurance, including the
issuance of performance standards, postmarket surveillance, patient
registries, development and dissemination of guidelines,
recommendations, and any other appropriate actions the Agency deems
necessary to provide such assurance (see also 21 CFR 860.3(c)(2)).
Section 510(m)(2) of the FD&C Act provides that FDA may exempt a
class II device from the premarket notification requirements on its own
initiative or upon petition of an interested person, if FDA determines
that a 510(k) is not necessary to provide reasonable assurance of the
safety and effectiveness of the device. Devices under the pediatric
hospital bed classification regulation were exempted from premarket
notification, subject to certain limitations, in accordance with
section 510(m) of the FD&C Act (63 FR 59222 at 59229, November 3,
1998).
II. Regulatory History and Description of the Devices
FDA classified pediatric hospital beds (21 CFR 880.5140) as class
II devices (45 FR 69678 at 69694, October 21, 1980), and later exempted
them from premarket notification (510(k)), in a final rule published in
the Federal Register of November 3, 1998 (63 FR 59222 at 59229). In
Sec. 880.5140, a pediatric hospital bed is defined as ``a device
intended for medical purposes that consists of a bed or crib designed
for the use of a pediatric patient, with fixed end rails and movable
and latchable side rails. The contour of the bed surface may be
adjustable.''
A medical bassinet is a non-powered device that consists of two
components: (1) A basket, the sleep or bed component, which is
typically made of plastic and (2) a durable frame with wheels, which
holds the basket or bed component (FDA refers to this component as a
``basket or bed component'' in this proposed rule). The basket or bed
component is a box-like structure, generally made of a clear, high-
impact resistant plastic material, with an open top and four stationary
walls to keep the baby in place. Medical bassinets are typically used
in hospital settings for infants up to 5 months in age. Medical
bassinets currently fall under the pediatric hospital bed
classification regulation.
III. Proposed Regulation
Pediatric medical cribs that meet the definition of a device in
section 201(h) of the FD&C Act (21 U.S.C. 321(h)) are regulated by FDA
(referred to as pediatric medical cribs or cribs intended for medical
purposes) (product code FMS) and, if this rule is finalized, will have
to comply with the special controls identified in the final regulation
for pediatric medical cribs. Cribs outside of the device definition
(referred to as cribs for non-medical purposes) must meet the CPSC's
regulations and guidelines. A crib designed for the use of a pediatric
patient may meet the medical device definition if it is intended for
use in the cure, mitigation or treatment of disease (see section 201(h)
of the FD&C Act).
In the Federal Register of December 28, 2010 (75 FR 81766), the
CPSC issued a final rule prohibiting the use of the drop-side rail
design for non-medical cribs in consumer households as of June 28,
2011. Child care facilities, family child care homes, and places of
public accommodation (e.g., hotels and motels) had to comply with the
rule as of December 28, 2012. CPSC's rule establishes new standards for
full-size and non-full-size cribs used for non-medical purposes, which
effectively prohibit the manufacture or sale of cribs for non-medical
purposes with a drop-side rail design in households, child care
facilities, family child care homes, and places of public
accommodation. This rule did not affect pediatric medical cribs
regulated by FDA, which may contain a drop-side rail design that
includes movable and latchable side and end rails.
Because drop-side rail cribs for non-medical purposes and pediatric
medical cribs are regulated by different agencies, CPSC consulted with
FDA about the impact their final rule could have on settings, such as
nursery schools and day care centers, where pediatric medical cribs
with drop-side rails are often used for pediatric patients after they
have been discharged from a health care facility. CPSC, which regulates
consumer products, including drop-side rail cribs not intended for
medical purposes, received reports of deaths of children attributable
to entrapment and/or strangulation caused by the malfunctioning of
drop-side rail cribs.
Although drop-side cribs for non-medical purposes are now
prohibited, there is still a need for pediatric medical cribs with
drop-side rails inside and outside of traditional health care settings.
CPSC and FDA have heard from medical device consumers and health care
providers that pediatric medical cribs with drop-side rails are
extremely helpful for patient care in hospital settings and even
outside of traditional health care settings, such as day care centers
caring for infants and children with disabilities, because they allow
parents and care givers easy
[[Page 60811]]
access to children to perform routine and emergency medical procedures,
including, but not limited to, CPR, blood collection, IV insertion,
respiratory care, and skin care. These drop-side rail cribs also make
it easier for hospital staff to facilitate safe patient transport and
reduce the chance of caregiver injury. Health care workers have stated
that they need to have continued access to these medical cribs with
drop-side rails (Ref. 1). Therefore, FDA is proposing to permit
manufacturers to continue to manufacture and sell medical cribs with
the drop-side rail design in traditional health care settings and to
permit the use of pediatric medical cribs with drop-side rail designs
outside of traditional health care settings through prescription use
only (it is noted that State child care licensing agencies are
generally responsible for overseeing day care providers while FDA is
responsible for medical devices).
FDA is proposing to revise the identification in Sec. 880.5140 to
include only pediatric medical cribs, establish special controls for
this device, and change the name of the classification regulation from
``pediatric hospital bed'' to ``pediatric medical crib.'' The Agency is
taking these actions to clarify the devices that fall under this
particular classification regulation and establish special controls the
Agency believes are necessary for a reasonable assurance of safety and
effectiveness. In addition, FDA is proposing that use of a pediatric
medical crib be restricted to prescription use in accordance with 21
CFR 801.109. In order to use or administer use of pediatric medical
cribs, authorization must be made by a practitioner licensed by law
through a prescription for the device.
This rule also proposes to create a separate regulation for medical
bassinets and establish special controls for this device type to
provide a reasonable assurance of safety and effectiveness. In
addition, FDA is proposing that use of medical bassinets be restricted
to prescription use in accordance with 21 CFR 801.109. In order to use
or administer use of medical bassinets, authorization must be made by a
practitioner licensed by law through a prescription for the device. FDA
proposes not to change the 510(k) exempt status of pediatric medical
cribs and medical bassinets.
Devices currently under the pediatric hospital bed classification
regulation include: Open pediatric medical cribs, medical bassinets,
pediatric cribs with integrated air mattresses, youth beds, pediatric
stretchers, crib enclosure beds, and cuddle-carrier infant beds. If
this proposed rule is finalized, devices that do not meet the
definition of ``pediatric medical crib'' will be administratively moved
to more appropriate class II regulations, and no longer be under the
revised pediatric hospital bed classification regulation. At that time,
FDA proposes to send manufacturers of the remaining pediatric hospital
beds notices identifying the new classification regulation and product
code under which the device will be classified.
If this proposed rule is finalized, FDA intends to move the
following medical devices listed under Sec. 880.5140 to devices with
similar intended uses and class II regulations: Pediatric cribs with
integrated air mattresses to 21 CFR 890.5170, ``Powered flotation
therapy bed;'' youth beds to either 21 CFR 880.5100, ``AC powered
adjustable hospital bed,'' or 21 CFR 880.5120, ``Manual adjustable
hospital bed,'' depending on whether they are powered or not; pediatric
stretchers to 21 CFR 880.6910, ``Wheeled stretchers;'' and crib
enclosure beds to 21 CFR 880.6760, ``Protective restraint.'' This
action would not have any substantive effect on the current marketing
status of the devices. However, manufacturers of these devices would
need to refer to the new regulation classification and product code
provided by the Agency in future interactions with FDA.
As discussed in section IV, an analysis of Medical Device Reports
(MDRs) submitted to the Manufacturer and User Facility Device
Experience (MAUDE) database from January 1, 2005, to September 1, 2015,
indicated 516 adverse events associated with pediatric medical cribs
including 15 serious injuries. The adverse events associated with
pediatric medical cribs were assessed to better understand the risks
and establish the proposed special controls for this device. FDA
believes that sufficient information is available to establish special
controls to provide a reasonable assurance of the safety and
effectiveness of the device.
As discussed further in section VI, FDA believes risks to health
resulting from use of these cribs would be effectively mitigated by the
special controls proposed in this rule, and that these controls, in
combination with the general controls, would provide a reasonable
assurance of safety and effectiveness for pediatric medical cribs for
their intended use. Therefore, FDA is proposing new safety requirements
and allowing medical cribs in homes and day cares only when medically
necessary.
FDA is also taking this opportunity to address adverse event
reports pertaining to medical bassinets by proposing to establish
special controls for these devices to provide a reasonable assurance of
safety and effectiveness. FDA has received adverse events from
hospitals regarding incidents of medical bassinet tipping and improper
cleaning of the basket or bed component that caused cracks and crazing,
which have resulted in patient injury. The Agency is proposing to
separate medical bassinets from other types of pediatric hospital beds
to allow for more targeted postmarket surveillance of these devices.
FDA believes the special controls it is proposing here, in combination
with the general controls, would provide a reasonable assurance of
safety and effectiveness for medical bassinets.
IV. Risks to Health
A. Pediatric Medical Crib
Between January 1, 2005, and September 1, 2015, FDA received 516
adverse event reports, or MDRs, associated with open pediatric medical
cribs, through the Agency's MAUDE database. There were 15 adverse event
reports of serious injuries including 6 reports of entrapment, which
were predominantly extremity entrapments of legs or arms. The majority
of MDRs for medical cribs were for malfunctions such as drop-side rails
not latching or lowering, brakes not holding, wheels or casters
breaking, and where applicable, scales not reading correct weights.
These malfunctions (501 reports) were not associated with any adverse
health effects. After considering available information, FDA determined
that the following risks to health are associated with the use of
pediatric medical cribs:
Injury resulting from mechanical or structural failure of
the device--Mechanical or structural failure of the crib can result in
failure of load-bearing components such as the wheels or casters, or
failure of the latches or other locking mechanisms that secure the
sides of the crib. These failures can result in injuries, as
demonstrated by the MDRs received in FDA's database.
Pinching, laceration, splinters, and foreign body
ingestion--Depending on the material of the pediatric crib, certain
cribs may peel or crack and may expose pediatric patients to substances
or materials that may be toxic or may cause abrasions or lacerations if
the surface of the crib material is compromised.
Entrapment, falls, and strangulation--Pediatric medical
cribs may cause entrapment of patient limbs if the width of the side
rails are not correct and if there are gaps between the mattress and
crib frame that are larger
[[Page 60812]]
than the width of two fingers. Depending on the height requirements of
the rails a pediatric patient may escape or fall from the crib. The
term ``entrapment'' refers to circumstances where a patient is caught,
trapped, or entangled in the space in or about the bed rail, mattress,
or hospital bed frame.
Burns--Certain flammable materials used in the
construction of pediatric medical cribs may allow for the spread of
fire, which may result in serious injuries. Fires can spread easily in
hospital rooms with a patient using oxygen. The free-flowing oxygen can
intensify a fire, which can rapidly spread to flammable objects in the
room, including crib mattresses.
Use error--Use error may contribute to or exacerbate any
of the previously mentioned adverse events. For example, a user may be
unaware that a side rail did not latch, or may fail to properly
maintain a pediatric medical crib. Therefore, adequate instructions for
use and user education are essential to safe device operation.
B. Medical Bassinet
Between January 1, 2005, and September 1, 2015, FDA received 40
adverse event reports associated with this device type. The most common
MDRs for medical bassinets include reports of malfunctions such as
casters or wheels not working, which have caused tipping, and broken
bassinet base components, such as doors and drawers, or collapse or
breakage of utility shelves or chart holders. There are also reports of
the plastic sleep basket or bed component crazing (cracking), resulting
in sharp edges and cuts to hospital personnel.
FDA has considered the available information and determined that
the following risks to health are associated with medical bassinets:
Injury resulting from mechanical or structural failure of
the device--Mechanical or structural failure of the bassinet can result
in failure of load-bearing components such as the wheels or casters, or
failure of the latches or other locking mechanisms that secure the
drawers of the bassinet. These failures can result in injuries,
particularly if the bassinet tips over, as demonstrated by the MDRs
received in FDA's database.
Burns--Certain flammable materials used in the
construction of pediatric medical bassinets may allow for the spread of
fire, which may result in serious injuries. Fires can spread easily in
hospital rooms with a patient using oxygen. The free-flowing oxygen can
intensify a fire, which can rapidly spread to flammable objects in the
room, including bassinet mattresses.
Crazing or cracking of basket or bed component--The basket
or bed component of the bassinet that the pediatric patient is placed
in may craze or crack due to improper care or handling, such as
cleaning the plastic material of the basket or bed component with
inappropriate cleaning solutions. Crazing or cracking may result in
injuries such as cuts.
Use error--Use error may contribute to, or exacerbate, any
of the previously mentioned risks. For example, a user may accidentally
leave a door or drawer in the base component of the bassinet open or
place too much weight in a drawer or on a shelf, which may present a
tipping hazard. Also, a user may fail to properly maintain a medical
bassinet.
V. Establishment of Special Controls
Under section 513(a)(1)(B) of the FD&C Act, as amended by the SMDA,
class II devices are defined as devices for which general controls by
themselves are insufficient to provide reasonable assurance of safety
and effectiveness, but for which there is sufficient information to
establish special controls to provide such assurance. Special controls
may include the issuance of performance standards, postmarket
surveillance, patient registries, development and dissemination of
guidelines, recommendations, and other appropriate actions the Agency
deems necessary to provide such assurance (see also Sec. 860.3(c)(2)).
Under this authority, FDA is proposing to establish special
controls for pediatric medical cribs (Sec. 880.5140) and pediatric
medical bassinets (Sec. 880. 5145). The Agency believes that the
applicable special controls, together with the general controls, would
provide reasonable assurance of the safety and effectiveness of these
devices.
VI. Proposed Special Controls
A. Pediatric Medical Crib
FDA consulted with health care providers in children's hospitals,
registered nurses in pediatric units, biomedical engineers, and
technicians, and analyzed the associated adverse events with pediatric
medical cribs. Specifically, FDA consulted with MedSun hospitals
regarding their 2011 survey on clinicians' experiences with pediatric
medical cribs with drop-side rails used in MedSun's hospitals (Ref. 1).
The MedSun survey summary highlights the clinical perspective and the
importance of this device in medical and health care settings. The most
common issues and concerns in the survey were the lack of understanding
of side rail operation and the need for reinforcing patient safety when
the side rails are raised or lowered. Many respondents of the survey
suggested further improvements for pediatric medical cribs, for
instance, improved labeling, specific distance between slats and
emergency releases on side rails for faster access to pediatric
patients. The adverse events identified in the MedSun survey are
similar to the MDRs FDA has received on this device. FDA believes that
the special controls proposed in this proposed rule, in combination
with the general controls, would provide a reasonable assurance of
safety and effectiveness for pediatric medical cribs their intended
use.
Table 1--Health Risks and Mitigation Measures for Pediatric Medical Crib
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Injury Resulting from Mechanical or Performance Testing.
Structural Failure of the Device. Design Testing.
Pinching, Lacerations, Splinters, and Appropriate Materials Free From
Foreign Body Ingestion. Surface Defects.
Labeling.
Entrapment, Falls, and Strangulation... Performance Testing.
Rail and End Panel Design.
Side Rail Spacing and Safety
Features.
Appropriate Fitting of
Mattress.
Burns.................................. CPSC's Mattress Flammability
Standard.
Use Error.............................. Labeling.
------------------------------------------------------------------------
[[Page 60813]]
As provided in Table 1, the Agency believes the following special
controls, in combination with the general controls, would effectively
mitigate the identified risks to health and provide reasonable
assurance of the safety and effectiveness of the device:
1. Design and performance testing must be conducted to ensure the
mechanical and structural stability of the crib under expected
conditions of use, including the security of latches and other locking
mechanisms when engaged. These requirements are derived from sections
6.2 and 6.3 of ASTM (formerly the American Society for Testing and
Materials) International Standard F1169-13, entitled ``Standard
Consumer Safety Specification for Full-Size Baby Cribs'' (Ref. 2) and
sections 5.7 and 6.3 of ASTM International Standard F2710-13, entitled
``Standard Consumer Safety Performance Specification for Commercial
Cribs'' (Ref. 3), which was developed with input from crib
manufacturers.
2. To reduce possible injury of pinching, lacerations, and
crushing, the crib shall be designed and constructed in a manner that
eliminates hardware accessible to a child within the crib. This
requirement is derived from section 5.10 of ASTM International Standard
F1169-13, entitled ``Standard Consumer Safety Specification for Full-
Size Baby Cribs'' (Ref. 2), which was developed with input from crib
manufacturers. Also, materials used shall be appropriate for the
conditions of use, allow for proper sanitation, and free from surface
defects of the device that could result in injuries.
3. To reduce the risk of head and limb entrapment, the distance
between side rail components (such as slats, spindles, corner posts,
and rods) shall be designed to reduce potential entrapment of pediatric
patients and the distance between such components shall not exceed 2\3/
8\ inches (6 centimeters) apart. In addition, the rails and end panels
of a crib must be of a height to mitigate the possibility of falls and/
or escapes by the patient. These requirements are derived from sections
5.7.2 and 5.8.1 of ASTM International Standard F1169-13, entitled
``Standard Consumer Safety Specification for Full-Size Baby Cribs''
(Ref. 2), which was developed with input from crib manufacturers.
4. To reduce the risk of head and limb entrapment, no gap shall
exist between the edge of the bottom rail and the top of the mattress
surface and the mattress must fit tightly around all four sides of the
crib. These requirements are derived from section 5.9 of ASTM
International Standard F1169-13, entitled ``Standard Consumer Safety
Specification for Full-Size Baby Cribs'' (Ref. 2), which was developed
with input from crib manufacturers.
5. To reduce flammability and the risk of burns, the mattress for
the crib shall meet the CPSC Standard for the Flammability of
Mattresses and Mattress Pads and its Standard for the Flammability
(Open Flame) of Mattress Sets, 16 CFR parts 1632 and 1633,
respectively. This proposed special control would clarify for
manufacturers the standards necessary for mattresses intended to be
used with pediatric medical cribs to prevent the spread of fires that
can easily occur in hospital rooms with a patient using oxygen. The
free-flowing oxygen can intensify a fire, which can rapidly spread to
most of the flammable objects in the room especially mattresses. The
consumer standards for flammability of mattresses in 16 CFR parts 1632
and 1633 are also accepted by the Joint Commission (formerly the Joint
Commission on Accreditation of Healthcare Organizations). CPSC's
mattress and mattress pad flammability standard under 16 CFR part 1632
addresses mattress fires ignited by open flame sources, including
matches, candles, lighters, and other related scenarios. It prescribes
a test to determine the ignition resistance of a mattress or a mattress
pad when exposed to a lighted cigarette. CPSC's standard for the
flammability of mattress sets under 16 CFR part 1633 is a broader
standard designed to reduce deaths and injuries caused by mattress
fires, particularly those fires ignited by, among others things, oxygen
use or electrical equipment sources that may occur in a patient's room.
In addition, CPSC's regulations require that manufacturers meet an
established fire safety performance standard, based on ASTM E2187-09,
entitled ``International's Standard Test Method for Measuring the
Ignition Strength of Cigarettes'' (Ref. 4), which was developed with
input from crib manufacturers.
6. To reduce flammability and the risk of burns, the labeling must
bear all information required pursuant to the CPSC Standard for the
Flammability of Mattresses and Mattress Pads and its Standard for the
Flammability (Open Flame) of Mattress Sets, 16 CFR parts 1632 and 1633,
respectively.
7. To reduce the risk of use error, which may result in mechanical
or structural failure of the crib due to inadequate care or
maintenance, pediatric medical crib labeling must include adequate
instructions for users to care for and maintain their crib. These
requirements are derived from sections 5.18 of ASTM International
Standard F1169-13, entitled ``Standard Consumer Safety Specification
for Full-Size Baby Cribs'' (Ref. 2).
FDA believes that the special controls proposed in this rule would
provide a reasonable assurance of safety and effectiveness of pediatric
medical cribs in their intended use. The ASTM and CPSC standards noted
above apply to all mattresses and mattress pads intended or promoted
for sleeping upon, as defined in 16 CFR 1632.1(a), including medical
mattresses that are regulated by FDA as an accessory to medical beds.
Therefore, FDA anticipates that manufacturers would be able to meet the
requirements imposed by the proposed special controls in this proposed
rule without undue burden. FDA invites comments on this conclusion,
including comments regarding the types of performance testing
manufacturers conduct for pediatric medical cribs, particularly to
ensure the performance of medical crib latches on drop-side rails.
In addition, FDA is proposing to restrict these devices to
prescription use under section 520(e) of the FD&C Act (see Sec.
801.109 (prescription devices)). In order to use or administer use of
pediatric medical cribs, authorization must be made by a practitioner
licensed by law.
B. Medical Bassinet
Table 2 lists the risks to health FDA has identified for Medical
Bassinets, as described in the Risks to Health, section IV of this
proposed rule, along with the corresponding proposed mitigation
measures for each risk.
Table 2--Health Risks and Mitigation Measures for Medical Bassinet
------------------------------------------------------------------------
Identified risks to health Mitigation measures
------------------------------------------------------------------------
Injury Resulting from Mechanical or Performance Testing.
Structural Failure of the Device.
Labeling.
Burns.................................. CPSC's Mattress Flammability
Standard.
[[Page 60814]]
Crazing or Cracking of Basket or Bed Performance Testing.
Component.
Labeling.
Use Error.............................. Labeling.
------------------------------------------------------------------------
The Agency believes the following special controls, in combination
with the general controls, would effectively mitigate the identified
risks to health and provide reasonable assurance of the safety and
effectiveness of medical bassinets:
1. To mitigate crazing, cracking, and deterioration of the basket
or bed component of the device, the manufacturer must conduct
performance testing to determine material compatibility with cleansing
products labeled to clean the device.
2. To reduce flammability and the risk of burns, the bassinet shall
meet CPSC's Standard for the Flammability of Mattresses and Mattress
Pads and its Standard for the Flammability (Open Flame) of Mattress
Sets, 16 CFR parts 1632 and 1633, respectively.
3. To reduce the risk of injury resulting from mechanical or
structural failure of the device, and particularly, device tipping that
can result from those failures; manufacturers shall conduct performance
testing to ensure the mechanical and structural stability of the
bassinet under expected use conditions, including transport of patients
in the bassinet.
4. To reduce the risk of use error, specifically error that may
result in bassinet tipping, FDA proposes that manufacturers shall have
a label on the front of the bassinet cabinet with the following warning
statement:
WARNING: To avoid tipping hazards of this device, make sure that
the basket or bed component sits firmly in the base and that all
doors, drawers, and casters are secure.
The label must be affixed to the front of the bassinet base cabinet
and the text shall be in letters not less than 10 millimeters in
height.
FDA believes this warning is necessary because even if performance
testing demonstrates that a bassinet does not present a tipping hazard
under expected use conditions, users may exceed these expected use
conditions, particularly during transport of a patient in the bassinet.
5. To reduce the risk of use error, which may result in mechanical
or structural failure of the bassinet due to inadequate care or
maintenance, medical bassinet labeling must include adequate
instructions for users to care for and maintain the bassinet.
FDA believes that the special controls proposed in this rule would
provide a reasonable assurance of safety and effectiveness of medical
bassinets in their intended use. The CPSC standards noted previously
apply to all mattresses and mattress pads intended or promoted for
sleeping upon, as defined in 16 CFR 1632.1(a), including medical
mattresses that are regulated by FDA as an accessory to medical beds.
Therefore, FDA believes most manufacturers are already complying with
the proposed special control for mattress flammability set forth in
this proposed rule. FDA invites comments on the types of performance
testing manufacturers conduct for medical bassinets.
In addition, FDA is proposing to restrict these devices to
prescription use under section 520(e) of the FD&C Act (see Sec.
801.109 (Prescription devices)). In order to use or administer use of
medical bassinets, authorization must be made by a practitioner
licensed by law.
VII. Proposed Effective Date
FDA proposes that any final rule based on this proposal become
effective 60 days after its publication in the Federal Register.
VIII. Environmental Impact, No Significant Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
IX. Economic Analysis of Impacts
FDA has examined the impacts of the proposed rule under Executive
Order 12866, Executive Order 13563, the Regulatory Flexibility Act (5
U.S.C. 601-612) and the Unfunded Mandates Reform Act of 1995 (Pub. L.
104-4). Executive Orders 12866 and 13563 direct Agencies to assess all
costs and benefits of available regulatory alternatives and, when
regulation is necessary, to select regulatory approaches that maximize
net benefits (including potential economic, environmental, public
health and safety, and other advantages; distributive impacts; and
equity). The Agency believes that this proposed rule is not a
significant regulatory action as defined by Executive Order 12866.
The Regulatory Flexibility Act requires Agencies to analyze
regulatory options that would minimize any significant impact of a rule
on small entities. Because the expected costs associated with this rule
are expected to be modest, we propose to certify that this rule would
not have a significant economic impact on a substantial number of small
entities.
Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires
that Agencies prepare a written statement, which includes an assessment
of anticipated costs and benefits, before proposing ``any rule that
includes any Federal mandate that may result in the expenditure by
State, local, and tribal governments, in the aggregate, or by the
private sector, of $100,000,000 or more (adjusted annually for
inflation) in any one year.'' The current threshold after adjustment
for inflation is $144 million, using the most current (2014) Implicit
Price Deflator for the Gross Domestic Product. We do not expect this
proposed rule to result in any 1-year expenditure that would meet or
exceed this amount.
A. Objective of the Rule
Pediatric hospital beds are classified as class II, 510(k) exempt
medical devices intended for the treatment, care, or diagnosis of
diseases or illnesses of pediatric patients. In this proposed rule, FDA
proposes to amend Sec. 880.5140 by revising the identification and
establishing special controls for pediatric medical cribs. This rule
would also change the name of the classification regulation from
``pediatric hospital bed'' to ``pediatric medical crib,'' and place
medical bassinets, previously under the pediatric hospital beds
classification regulation, as a separate class II, 510(k) exempt
device, subject to its own special controls.
Pediatric medical cribs used in health care settings contain a
drop-side rail design that includes movable and latchable side and end
rails. As stated previously, the CPSC issued a final rule prohibiting
the use of the drop-side rail design for non-medical cribs in consumer
households as of June 28,
[[Page 60815]]
2011 (December 28, 2010, 75 FR 81766). CPSC's rule establishes new
standards for full-size and non-full-size cribs used for non-medical
purposes, which effectively prohibit the manufacture or sale of cribs
for non-medical purposes with a drop-side rail design in households,
child care facilities, family child care homes, and places of public
accommodation. The compliance date for this same rule for child care
facilities, family child care homes, and places of public accommodation
was December 28, 2012 (75 FR 81766). CPSC's rule was established in
response to infant deaths that occurred when the side rail of a crib
used for non-medical purposes detached or disengaged. In contrast,
there have been no deaths and a few serious injuries reported to FDA
associated with pediatric medical cribs, which are made of more durable
materials and construction in comparison to cribs used for non-medical
purposes. Additionally, FDA has determined that drop-side rails are
essential for patient care in hospital settings and even outside of
traditional health care settings to allow parents and care givers easy
access to their patients in order to perform both routine and emergency
medical procedures. To address concerns raised by CPSC's reports on
consumer drop-side rail cribs and account for the medical need of this
device, FDA is proposing to establish special controls and require a
prescription for this device when used outside of traditional health
care setting. Thus, FDA has determined that cribs with drop-side rail
designs may remain on the market when intended for medical use, and can
be used outside of traditional health care settings through
prescription use only. In addition, this proposed rule adds specific
special controls for medical bassinets to mitigate health risks such as
tipping of the device and crazing of the plastic basket or bed
component. Regulation of this device under a separate regulation would
also allow for more targeted post market surveillance for this device.
B. Benefits
FDA's Registration and Listing database identifies 38 manufacturers
of medical cribs and bassinets. FDA has reviewed the safety standards
of several large pediatric crib and bassinet manufacturers in order to
determine the compliance burden associated with the proposed special
controls. The Agency concludes that many of the special controls
proposed by this rule are consistent with current industry practice
among many medical crib and bassinet manufacturers. The proposed
special control that is not currently practiced is the warning labeling
requirements for medical bassinets. For new pediatric medical crib and
bassinet manufacturers entering the market or manufacturers that may
not be currently following the practices required by the proposed
special controls, if this proposed rule is finalized, its special
controls will clarify safety standards and minimize the risk of injury
to pediatric patients.
The beneficial features of medical bassinets are portability, ease
of cleaning, and, when it is made of a clear material, the ability to
see the baby from all sides. The proposed special controls would
require bassinet manufacturers to place labels on their devices warning
against device tipping. This requirement would apply to new bassinets;
bassinets that have already been sold would not be required to add the
new labels to their devices. The warning label is intended to prevent
tipping of the device, which may be caused by unlatched drawers,
dislodged wheels, or too much weight on the shelves. The Agency has not
received any reports of death or serious injury related to medical
bassinets, although there have been a small number of reports of
malfunctioning casters, which may cause device tipping. The benefits of
the new warning label are not readily quantifiable, but it is expected
to reduce the risk of the bassinet from tipping and thus, reduce
potential injury to pediatric patients.
The provision allowing for the medical cribs outside of traditional
health care settings would benefit pediatric patients who require the
specialized care provided by these devices outside of traditional
health care settings. Due to the CPSC rule regarding cribs used for
non-medical purposes, discussed previously, consumers and child care
facilities are restricted from using cribs with a drop-side rail
design. If this proposed rule is finalized, it will allow consumers and
child care facilities to utilize the pediatric medical cribs if they
are prescribed by a health care professional.
The special controls regarding the mechanical structure of
pediatric medical cribs are intended to minimize the risk of injury,
including entrapment or strangulation of pediatric patients. The
spacing specifications of the side rail components are designed to
prevent head or neck entrapment and strangulation incidents in which
infants may slip between the openings of the slats, and the performance
testing requirements are designed to ensure the side rail latches of
pediatric medical cribs will perform as intended and remain secure when
the latches are engaged. The special control requiring specific height
of the rails and end panels may prevent falls and/or escapes by the
patient. Also, by having pediatric medical crib manufacturers use
materials that are appropriate for the conditions of use and allow for
proper sanitation, these special controls may help mitigate surface
defects that can cause injury to the patient.
Additionally, the mattress size standards for cribs and bassinets
are intended to reduce the risk of significant gaps between the
mattress and the device structure, which could potentially create an
entrapment hazard. The flammability standard is intended to reduce
deaths and injuries related to mattress fires, particularly those
initially ignited by open flame sources such as lighters, candles, and
matches. Although the practices proposed in these special controls are
believed to be followed by almost all manufacturers of products
currently on the market, the proposed special controls would reinforce
safety standards for such manufacturers and ensure that other
manufacturers and manufacturers of new products adhere to the same
safety standards.
C. Costs
The economic impact of the proposed regulation is determined
primarily by whether manufacturers currently comply with the proposed
special controls. As stated previously, the special controls that are
not currently practiced by industry, of which FDA is aware, are the
bassinet warning labeling and the performance testing requirements. FDA
is also aware that many manufacturers of pediatric medical cribs and
medical bassinets registered with the FDA currently conform to the risk
mitigations and structural requirements that are being proposed as
special controls, and thus conforming to these special controls, if
finalized, would not result in an increase in cost to pediatric medical
crib manufacturers and only cause a small increase in cost for medical
bassinet manufactures. Additionally, the renaming of pediatric medical
cribs and redesignation in the CFR for medical bassinets and the
remaining devices under the pediatric hospital bed classification are
administrative in nature, and are not expected to result in any cost
burdens.
The new warning labeling requirements for medical bassinets will
apply to manufacturers of new bassinets only. FDA does not expect
bassinets that are currently on the market to be relabeled. If
manufacturers of new bassinets add labels to the devices at the
[[Page 60816]]
time of production, the cost burden to manufacturers would be
minimized. Although we do not have direct estimates of labeling costs
for these devices, the best estimate of these costs is derived from
FDA's labeling cost model. Because FDA would require specific language
and format of the labels, we consider this to be a minor labeling
change that would not require label design, market tests, or analytical
tests. Labeling costs would include labor and material, and are
estimated to be, on average, approximately $140 per unit. Then we use
the number of live births per year as reported by the Center for
Disease Control and Prevention in order to determine the number of
bassinets produced per year for medical use (Ref. 5). Using an estimate
of 4 million births per year and 11,000 births per day, we estimate
that each birth requires an average hospital stay of 3 days. This
yields a total supply of approximately 33,000 medical bassinets in the
United States. Given an average yearly replacement rate of 20 percent
for all medical bassinets, we estimate that approximately 6,600 new
bassinets will be produced annually. Applying the $140 per unit
labeling cost yields a total yearly cost of $924,000 associated with
the new bassinet warning label requirement.
The special controls require performance testing for medical
bassinets to reduce the risk of crazing of the plastic basket or bed
component. We assume that the performance testing may be conducted as
an extension to current product testing and may be performed at the
same testing facilities currently utilized by bassinet manufacturers.
FDA projects that a maximum of an additional week of testing would be
required. The costs associated with the performance testing include the
labor costs of mechanical engineers, who typically perform these tests.
The mean 2012 hourly wage for mechanical engineers is $40.75, as
reported by the Occupational Employment Statistics provided by the
Bureau of Labor Statistics (Ref. 6). Applying a multiplier of 1.45 to
adjust for benefits, hourly labor costs are estimated to be
approximately $59. Assuming a 40-hour work week, the total maximum
estimated cost for each manufacturer to perform these additional tests
is approximately $2,360. It is uncertain the exact number of
manufacturers that do not currently conduct performance testing and
would therefore be required to extend current testing practices.
However, given the relatively small number of medical bassinet
manufacturers, FDA anticipates that even the upper-bound total cost
would be modest.
The prescription use of pediatric medical cribs outside of
traditional health care settings may potentially increase Medicaid
spending for eligible pediatric patients. According to our review of
Healthcare Common Procedure Coding System billing codes for the
Medicaid program, currently, States typically offer Medicaid coverage
for prescribed rental or purchase of hospital beds and pediatric cribs
(Ref. 7). We estimate the number of additional prescriptions for
pediatric medical cribs to be filled annually as a result of this
proposed rule would be less than 100. Medicaid expenditure on pediatric
medical cribs is estimated to be on average $2,500 per device. This
yields a maximum annual total cost of $250,000.
Although it is unlikely that these devices would require physical
modification to meet the standards proposed by the special controls in
this proposed rule, there may be manufacturers on the market of which
we are unaware that do not conform to the requirements proposed in the
special controls. The proposed special controls could have a
significant impact on firms that are not currently in compliance with
the special controls, as their products may require modifications. The
special control that may cause additional costs for manufacturers is
the special control concerning the mechanical structure of pediatric
medical cribs. We are not able to estimate the actual compliance costs
for manufacturers of pediatric medical cribs because such costs may
vary by firm size and the amount of modification required.
Alternatively, we provide an estimate of the modification cost by using
aggregate industry market price information and cost data. The costs
associated with these modifications may include the costs associated
with product design and testing, labor, material, and production. We
use data from the Annual Survey of Manufacturers to calculate aggregate
labor and materials costs as a percentage of total sales for
manufacturers represented by North American Industry Classification
System code 339113 (Ref. 8). The data indicate that labor and materials
represent approximately 45 percent of total sales. Allowing market
price to represent per unit revenue at the firm level, we estimate the
cost of modification to be approximately 45 percent of the average
price of a pediatric medical crib. After surveying market prices of
pediatric medical cribs, we estimate an average per unit price of
$2,500. This yields an average cost of approximately $1,125 to modify a
pediatric medical crib to be in compliance with the proposed special
controls.
FDA invites comments on the compliance of manufacturers with the
special controls, including the performance testing, mechanical
structure, flammability requirements, and bassinet labeling
requirements, as well as cost information if modifications are
required.
X. Paperwork Reduction Act of 1995
The proposed rule refers to previously approved collections of
information found in FDA regulations. These collections of information
are subject to review by the Office of Management Budget (OMB) and the
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections
of information, regarding premarket notification submissions (21 CFR
part 807, subpart E), are approved under OMB control number 0910-0120.
The collections of information, regarding labeling (21 CFR part 801),
including prescription device labeling and adequate directions for use,
are approved under OMB control number 0910-0485. The collections of
information regarding current good manufacturing practice quality
systems (21 CFR part 820), including design controls (as referenced in
proposed Sec. 880.5140(b)(1) and proposed Sec. 880.5145(b)(1) and
(b)(3) of this document), are approved under OMB control number 0910-
0073. The collections of information in 16 CFR 1632 and 1633, regarding
mattress flammability, are approved under OMB control number 3041-0014.
In addition, FDA concludes that the warning label for bassinets
does not constitute a ``collection of information'' under the PRA.
Rather, the labeling statement is ``public disclosure(s) of information
originally supplied by the Federal government to the recipient for the
purpose of disclosure to the public.'' (5 CFR 1320.3(c)(2)).
XI. References
The following references have been placed on display in the
Division of Dockets Management (see ADDRESSES) and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday,
and are available electronically at https://www.regulations.gov. (FDA
has verified the Web site addresses, but FDA is not responsible for any
subsequent changes to the Web sites after this document publishes in
the Federal Register.)
1. MedSun: Newsletter #66: ``Pediatric Hospital Cribs: MedSun Small
Sample Survey Summary'' (November 2011),
[[Page 60817]]
available at https://www.fda.gov/downloads/MedicalDevices/Safety/MedSunMedicalProductSafetyNetwork/Newsletters/UCM422131.pdf.
2. ASTM International (formerly the American Society for Testing and
Materials), ``Standard Consumer Safety Specifications for Full-Size
Baby Cribs,'' Designation: F1169-13, available at https://www.astm.org/Standards/F1169.htm, 2013.
3. ASTM International (formerly the American Society for Testing and
Materials), ``Standard Consumer Safety Performance Specification for
Commercial Cribs,'' Designation: F2710-13, available at https://www.astm.org/Standards/F2710.htm, 2013.
4. ASTM International, ``Standard Test Method for Measuring the
Ignition Strength of Cigarettes,'' Designation: E2187-09 Standard,
available at https://www.astm.org/Standards/E2187.htm.
5. Centers for Disease Control and Prevention (CDC), National Vital
Statistics System, Birth Data, available at https://www.cdc.gov/nchs/births.htm.
6. Occupational Employment Statistics provided by the Bureau of
Labor Statistics, May 2012 Occupational Employment Statistics,
available at https://www.bls.gov/oes/2012/may/oes_stru.htm.
7. Centers for Medicare & Medicaid Services, HCPCS 2015 Code: E0300,
available at https://www.cms.gov/Medicare/Coding/MedHCPCSGenInfo/?redirect=/medhcpcsgeninfo/.
8. U.S. Census Bureau, 2013 Annual Survey of Manufacturers,
available at https://www.census.gov/manufacturing/asm/.
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 880 be amended as follows:
PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES
0
1. The authority citation for 21 CFR part 880 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Revise Sec. 880.5140 to read as follows:
Sec. 880.5140 Pediatric medical crib.
(a) Identification. A pediatric medical crib is a prescription
device intended for medical purposes for use with a pediatric patient
that consists of an open crib, fixed-end rails, movable and latchable
side rail components, and possibly an accompanying mattress. The
contour of the crib surface may be adjustable.
(b) Classification. Class II (special controls). The device is
exempt from the premarket notification procedures in subpart E of part
807 of this chapter subject to Sec. 880.9. The special controls for
this device are:
(1) Crib design and performance testing shall demonstrate the
mechanical and structural stability of the crib under expected
conditions of use, including the security of latches and other locking
mechanisms when engaged;
(2) Materials used shall be appropriate for the conditions of use,
allow for proper sanitation and free from surface defects that could
result in injuries;
(3) Rails and end panels shall be designed taking into account the
crib's height at its lowest point to the top of the mattress to prevent
patient falls and/or escape. Hardware and fasteners shall be designed
and constructed to eliminate mechanical hazards to the patient;
(4) The distance between components of the side rail (such as
slats, spindles, and corner posts) shall not be greater than 2\3/8\
inches (6 centimeters (cm)) apart at any point. Side rails shall
contain safety features for locking and adjust the lowest position of
the crib to a height that shall be 20 inches (51 cm) above the top of
the mattress;
(5) The device shall not have a gap between the bottom of the rail
and the top surface of the mattress and the mattress pad must fit
tightly around all four sides of the crib;
(6) The mattress for the crib shall meet the Consumer Product
Safety Commission (CPSC) Standard for the Flammability of Mattresses
and Mattress Pads and Standard for the Flammability (Open Flame) of
Mattress Sets, 16 CFR parts 1632 and 1633, respectively;
(7) The labeling must bear all information required pursuant to the
CPSC Standard for the Flammability of Mattresses and Mattress Pads and
Standard for the Flammability (Open Flame) of Mattress Sets, 16 CFR
parts 1632 and 1633, respectively; and
(8) Pediatric medical crib labeling must include adequate
instructions for users to care for and maintain their crib.
0
3. Add Sec. 880.5145 to subpart F to read as follows:
Sec. 880.5145 Medical bassinet.
(a) Identification. A medical bassinet is a prescription device
that is a small bed intended for use with pediatric patients, generally
from birth to approximately 5 months of age. It is intended for medical
purposes for use in a nursery, labor and delivery unit, or patient
room, but may also be used outside of traditional health care settings.
A medical bassinet is a non-powered device that consists of two
components: The plastic basket or bed component and a durable frame
with wheels, which holds the basket or bed component. The basket or bed
component is a box-like structure, generally made of a clear, high
impact-resistant plastic material, with an open top and four stationary
walls to hold the pediatric patient. The frame can include drawers,
shelving or cabinetry that provides space to hold baby care items. The
wheels or casters allow the bassinet to transport the baby throughout
the care setting.
(b) Classification. Class II (special controls). The device is
exempt from the premarket notification procedures in subpart E of part
807 of this chapter subject to Sec. 880.9. The special controls for
this device are:
(1) The manufacturer must conduct performance testing to determine
material compatibility with cleansing products labeled to clean the
device. Testing must demonstrate that the cleaning instructions
provided by the manufacturer do not cause crazing, cracking, or
deterioration of the device;
(2) The mattress for the device shall meet the Consumer Product
Safety Commission Standard for the Flammability of Mattresses and
Mattress Pads and Standard for the Flammability (Open Flame) of
Mattress Sets, 16 CFR parts 1632 and 1633, respectively;
(3) Manufacturers shall conduct performance testing to ensure the
mechanical and structural stability of the bassinet under expected use
conditions, including transport of patients in the bassinet. Testing
must demonstrate that failures such as wheel or caster breakage do not
occur, and that the device does not present a tipping hazard due to any
mechanical failures, under expected use conditions;
(4) Each device must have affixed a label on the front of the
bassinet cabinet with the following language in text of at least 10
millimeters in height:
WARNING: To avoid tipping hazards of this device, make sure that
the basket or bed component sits firmly in the base and that all
doors, drawers, and casters are secure.
(5) Labeling must include adequate instructions for users to care
for and maintain their bassinet.
Dated: October 2, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015-25627 Filed 10-7-15; 8:45 a.m.]