Medical Devices; Ophthalmic Devices; Classification of the Oral Electronic Vision Aid, 57090-57092 [2015-24026]
Download as PDF
57090
Federal Register / Vol. 80, No. 183 / Tuesday, September 22, 2015 / Rules and Regulations
(2) Contacting the Manufacturer: For any
requirement in this AD to obtain corrective
actions from a manufacturer, the action must
be accomplished using a method approved
by the Manager, International Branch, ANM–
116, Transport Airplane Directorate, FAA; or
EASA; or Airbus’s EASA DOA. If approved
by the DOA, the approval must include the
DOA-authorized signature.
Lhorne on DSK5TPTVN1PROD with RULES
(r) Related Information
(1) Refer to Mandatory Continuing
Airworthiness Information (MCAI) EASA
Airworthiness Directive 2014–0137, dated
May 28, 2014, for related information. This
MCAI may be found in the AD docket on the
Internet at https://www.regulations.gov by
searching for and locating Docket No. FAA–
2014–0753.
(2) Airbus Service information identified in
this AD that is not incorporated by reference
is available at the addresses specified in
paragraphs (s)(6) and (s)(8) of this AD.
(s) Material Incorporated by Reference
(1) The Director of the Federal Register
approved the incorporation by reference
(IBR) of the service information listed in this
paragraph under 5 U.S.C. 552(a) and 1 CFR
part 51.
(2) You must use this service information
as applicable to do the actions required by
this AD, unless this AD specifies otherwise.
(3) The following service information was
approved for IBR on October 27, 2015.
(i) Airbus Service Bulletin A320–27–1223,
dated September 3, 2013.
(ii) UTC Aerospace Systems Service
Bulletin 31075–27–22, dated July 2, 2013.
(4) The following service information was
approved for IBR on October 21, 2011 (76 FR
57630, September 16, 2011).
(i) Airbus Service Bulletin A320–27A1186,
Revision 07, including Appendices 1, 2, 3, 4,
5, and 6, dated March 2, 2011.
(ii) Reserved.
(5) The following service information was
approved for IBR on September 22, 2009 (74
FR 41611, August 18, 2009).
(i) Airbus All Operators Telex A320–
27A1186, Revision 04, dated April 3, 2009.
The document number and issue date of
Airbus AOT A320–27A1186, Revision 04,
dated April 3, 2009, are specified only on the
first page of the AOT.
(ii) Reserved.
(6) For Airbus service information
identified in this AD, contact Airbus,
Airworthiness Office—EIAS, 1 Rond Point
Maurice Bellonte, 31707 Blagnac Cedex,
France; telephone +33 5 61 93 36 96; fax +33
5 61 93 44 51; email account.airworth-eas@
airbus.com; Internet https://www.airbus.com.
(7) For UTC service information identified
in this AD, contact UTC Aerospace Systems;
Roger Dangremont; telephone +01 34 32 63
28; email roger.dangrement@goodrich.com.
(8) You may view this service information
at the FAA, Transport Airplane Directorate,
1601 Lind Avenue SW., Renton, WA. For
information on the availability of this
material at the FAA, call 425–227–1221.
(9) You may view this service information
that is incorporated by reference at the
National Archives and Records
Administration (NARA). For information on
VerDate Sep<11>2014
13:50 Sep 21, 2015
Jkt 235001
the availability of this material at NARA, call
202–741–6030, or go to: https://
www.archives.gov/federal-register/cfr/ibrlocations.html.
Issued in Renton, Washington, on
September 11, 2015.
Michael Kaszycki,
Acting Manager, Transport Airplane
Directorate, Aircraft Certification Service.
[FR Doc. 2015–23541 Filed 9–21–15; 8:45 am]
BILLING CODE 4910–13–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA–2015–N–3044]
Medical Devices; Ophthalmic Devices;
Classification of the Oral Electronic
Vision Aid
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA) is classifying the
oral electronic vision aid into class II
(special controls). The special controls
that will apply to the device are
identified in this order and will be part
of the codified language for the oral
electronic vision aid’s classification.
The Agency is classifying the device
into class II (special controls) in order
to provide a reasonable assurance of
safety and effectiveness of the device.
DATES: This order is effective September
22, 2015. The classification was
applicable on June 18, 2015.
FOR FURTHER INFORMATION CONTACT:
Dexiu Shi, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2246, Silver Spring,
MD, 20993–0002, 301–796–6470,
dexiu.shi@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
In accordance with section 513(f)(1) of
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C.
360c(f)(1)), devices that were not in
commercial distribution before May 28,
1976 (the date of enactment of the
Medical Device Amendments of 1976),
generally referred to as postamendments
devices, are classified automatically by
statute into class III without any FDA
rulemaking process. These devices
remain in class III and require
premarket approval, unless and until
the device is classified or reclassified
PO 00000
Frm 00022
Fmt 4700
Sfmt 4700
into class I or II, or FDA issues an order
finding the device to be substantially
equivalent, in accordance with section
513(i), to a predicate device that does
not require premarket approval. The
Agency determines whether new
devices are substantially equivalent to
predicate devices by means of
premarket notification procedures in
section 510(k) of the FD&C Act (21
U.S.C. 360(k)) and part 807 (21 CFR part
807) of the regulations.
Section 513(f)(2) of the FD&C Act (21
U.S.C. 360c(f)(2)), as amended by
section 607 of the Food and Drug
Administration Safety and Innovation
Act (Pub. L. 112–144), provides two
procedures by which a person may
request FDA to classify a device under
the criteria set forth in section 513(a)(1).
Under the first procedure, the person
submits a premarket notification under
section 510(k) of the FD&C Act for a
device that has not previously been
classified and, within 30 days of
receiving an order classifying the device
into class III under section 513(f)(1), the
person requests a classification under
section 513(f)(2). Under the second
procedure, rather than first submitting a
premarket notification under section
510(k) and then a request for
classification under the first procedure,
the person determines that there is no
legally marketed device upon which to
base a determination of substantial
equivalence and requests a classification
under section 513(f)(2) of the FD&C Act.
If the person submits a request to
classify the device under this second
procedure, FDA may decline to
undertake the classification request if
FDA identifies a legally marketed device
that could provide a reasonable basis for
review of substantial equivalence with
the device or if FDA determines that the
device submitted is not of ‘‘lowmoderate risk’’ or that general controls
would be inadequate to control the risks
and special controls to mitigate the risks
cannot be developed.
In response to a request to classify a
device under either procedure provided
by section 513(f)(2) of the FD&C Act,
FDA will classify the device by written
order within 120 days. This
classification will be the initial
classification of the device.
On August 7, 2013, Wicab Inc.,
submitted a request for classification of
the BrainPort V100 under section
513(f)(2) of the FD&C Act. The
manufacturer recommended that the
device be classified into class II (Ref. 1).
In accordance with section 513(f)(2) of
the FD&C Act, FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1). FDA classifies
E:\FR\FM\22SER1.SGM
22SER1
Federal Register / Vol. 80, No. 183 / Tuesday, September 22, 2015 / Rules and Regulations
devices into class II if general controls
by themselves are insufficient to
provide reasonable assurance of safety
and effectiveness, but there is sufficient
information to establish special controls
to provide reasonable assurance of the
safety and effectiveness of the device for
its intended use. After review of the
information submitted in the request,
FDA determined that the device can be
classified into class II with the
establishment of special controls. FDA
believes these special controls, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on June 18, 2015, FDA
issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 886.5905.
Following the effective date of this
final classification order, any firm
submitting a premarket notification
(510(k)) for an oral electronic vision aid
will need to comply with the special
controls named in this final order. The
device is assigned the generic name oral
electronic vision aid, and it is identified
57091
as a battery-powered prescription device
that contains an electrode stimulation
array to generate electrotactile
stimulation patterns that are derived
from digital object images captured by a
camera. It is intended to aid profoundly
blind patients in orientation, mobility,
and object recognition as an adjunctive
device to other assistive methods such
as a white cane or a guide dog.
FDA has identified the following risks
to health associated specifically with
this type of device, as well as the
mitigation measures required to mitigate
these risks in table 1.
TABLE 1—ORAL ELECTRONIC VISION AID RISKS AND MITIGATION MEASURES
Identified risk
Mitigation method
Irritation, Discomfort or Adverse Events Involving the Mouth, Tongue,
or Gums.
Adverse Tissue Reaction .........................................................................
Clinical Testing.
Labeling.
Biocompatibility Testing.
Labeling.
Non-Clinical Performance Testing.
Clinical Testing.
Labeling.
Software Verification, Validation, and Hazard Analysis.
Clinical Testing.
Healthcare Professional Training.
Patient Training.
Labeling.
Electromagnetic Compatibility and Electromagnetic Interference Testing.
Wireless Coexistence Testing.
Labeling.
Electrical Safety Testing.
Labeling.
Unit (Hardware) Malfunction, Functional Reliability .................................
Software Malfunction ................................................................................
Use Error ..................................................................................................
Interference with Other Devices ...............................................................
Lhorne on DSK5TPTVN1PROD with RULES
Electrical Shock ........................................................................................
FDA believes that the following
special controls, in combination with
the general controls, address these risks
to health and provide reasonable
assurance of the safety and
effectiveness:
• Clinical performance testing must
demonstrate an acceptable adverse
event profile, including adverse events
involving the mouth, tongue, and gums
and demonstrate the effect of the
stimulation to provide clinically
meaningful outcomes. The clinical
performance testing must also
investigate the anticipated conditions of
use, including potential use error,
intended environment of use, and
duration of use.
• Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use, including simulated
moisture ingress, device durability, and
battery reliability.
• Software verification, validation,
and hazard analysis must be performed.
• Analysis/testing must validate
electromagnetic compatibility.
• Analysis/testing must validate
electrical safety.
VerDate Sep<11>2014
13:50 Sep 21, 2015
Jkt 235001
• Analysis/testing must assess and
validate wireless coexistence concerns.
• Any elements of the device that
contact the patient must be
demonstrated to be biocompatible.
• Training must include elements to
ensure that the healthcare provider and
user can identify the safe environments
for device use, use all safety features of
the device, and operate the device in the
intended environment of use.
• Labeling for the trainer and user
must include a summary of the clinical
testing including adverse events
encountered under use conditions,
summary of study outcomes and
endpoints, and information pertinent to
use of the device including the
conditions under which the device was
studied (e.g., level of supervision or
assistance, and environment of use).
Oral electronic vision aid devices are
prescription devices restricted to patient
use only upon the authorization of a
practitioner licensed by law to
administer or use the device; see 21 CFR
801.109 (Prescription devices).
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
PO 00000
Frm 00023
Fmt 4700
Sfmt 4700
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
is necessary to provide reasonable
assurance of the safety and effectiveness
of the device. Therefore, this device
type is not exempt from premarket
notification requirements. Persons who
intend to market this type of device
must submit to FDA a premarket
notification, prior to marketing the
device, which contains information
about the oral electronic vision aid they
intend to market.
II. Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
E:\FR\FM\22SER1.SGM
22SER1
57092
Federal Register / Vol. 80, No. 183 / Tuesday, September 22, 2015 / Rules and Regulations
III. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
part 807, subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120, and the collections of
information in 21 CFR part 801,
regarding labeling have been approved
under OMB control number 0910–0485.
IV. Reference
The following reference has been
placed on display in the Division of
Dockets Management (HFA–305), Food
and Drug Administration, 5630 Fishers
Lane, rm. 1061, Rockville, MD 20852,
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and is available
electronically at https://
www.regulations.gov.
1. DEN130039: De Novo Request per 513(f)(2)
from Wicab Inc., dated August 7, 2013.
List of Subjects in 21 CFR Part 886
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 886 is
amended as follows:
PART 886—OPHTHALMIC DEVICES
1. The authority citation for 21 CFR
part 886 continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Add § 886.5905 to subpart F to read
as follows:
■
Lhorne on DSK5TPTVN1PROD with RULES
§ 886.5905
13:50 Sep 21, 2015
Jkt 235001
In this rule, BOEM amends its
existing regulations by: Updating
address locations; removing an outdated
Web site address and correcting a form
number; changing the term ‘‘Associate
Director’’ to ‘‘Deputy Director’’ in the
regulations; and other housekeeping
changes, such as removing reference to
a URL hyperlink for a Web page that no
longer exists.
SUMMARY:
This rule is effective September
22, 2015.
DATES:
FOR FURTHER INFORMATION CONTACT:
Robert Samuels, Office of Policy,
Regulation and Analysis, BOEM, 45600
Woodland Road, Sterling, VA 20166;
email: robert.samuels@boem.gov.
SUPPLEMENTARY INFORMATION:
I. Rulemaking Procedure
This rule pertains solely to
administrative changes. It makes no
changes to the substantive legal rights,
obligations, or interests of affected
parties. This rule, therefore, is a ‘‘rule of
agency organization, procedure or
practice’’ and is, therefore, exempt from
the notice-and-comment requirements
of 5 U.S.C. 553 under 5 U.S.C.
553(b)(A).
II. Overview of the Direct Final Rule
Bureau of Ocean Energy Management
In early 2015, many of BOEM’s
headquarters’ offices moved from
Herndon, Virginia to Sterling, Virginia.
References in the 30 CFR part 550
regulations to the Herndon, Virginia
location are updated in this rule to
reflect the Sterling, Virginia location.
This rule also updates other addresses
in 30 CFR part 519. Also, the existing
regulations contain references to the
title ‘‘Associate Director,’’ which is a
remnant of BOEM’s predecessor agency,
the Minerals Management Service. This
rule changes ‘‘Associate Director’’ to
‘‘Deputy Director’’ in the current
regulations. This rule also makes other
housekeeping changes, such as
removing reference to a URL hyperlink
for a Web page that no longer exists.
30 CFR Parts 519, 550, 551, 553, 556,
560, 580, 581, 582, and 585
III. Section-by-Section Analysis of
Direct Final Rule
[Docket No. BOEM–2015–0060; MMAA
104000]
30 CFR Part 519 (Distribution and
Disbursement of Royalties, Rentals, and
Bonuses)
Dated: September 16, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015–24026 Filed 9–21–15; 8:45 am]
BILLING CODE 4164–01–P
Oral electronic vision aid.
(a) Identification. An oral electronic
vision aid is a battery-powered
prescription device that contains an
electrode stimulation array to generate
electrotactile stimulation patterns that
are derived from digital object images
captured by a camera. It is intended to
aid profoundly blind patients in
orientation, mobility, and object
recognition as an adjunctive device to
other assistive methods such as a white
cane or a guide dog.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Clinical performance testing must
demonstrate an acceptable adverse
event profile, including adverse events
VerDate Sep<11>2014
involving the mouth, tongue, and gums
and demonstrate the effect of the
stimulation to provide clinically
meaningful outcomes. The clinical
performance testing must also
investigate the anticipated conditions of
use, including potential use error,
intended environment of use, and
duration of use.
(2) Non-clinical performance testing
must demonstrate that the device
performs as intended under anticipated
conditions of use, including simulated
moisture ingress, device durability, and
battery reliability.
(3) Software verification, validation,
and hazard analysis must be performed.
(4) Analysis/testing must validate
electromagnetic compatibility.
(5) Analysis/testing must validate
electrical safety.
(6) Analysis/testing must assess and
validate wireless coexistence concerns.
(7) Any elements of the device that
contact the patient must be
demonstrated to be biocompatible.
(8) Training must include elements to
ensure that the healthcare provider and
user can identify the safe environments
for device use, use all safety features of
the device, and operate the device in the
intended environment of use.
(9) Labeling for the trainer and user
must include a summary of the clinical
testing including adverse events
encountered under use conditions,
summary of study outcomes and
endpoints, and information pertinent to
use of the device including the
conditions under which the device was
studied (e.g., level of supervision or
assistance, and environment of use).
DEPARTMENT OF THE INTERIOR
RIN 1010–AD94
Updating Addresses and Contact
Information in the Bureau of Ocean
Energy Management’s Regulations
Bureau of Ocean Energy
Management (BOEM), Interior.
ACTION: Direct final rule.
AGENCY:
PO 00000
Frm 00024
Fmt 4700
Sfmt 4700
Section 519.410 What does this
subpart contain?
Section 519.410(b) contains contact
information for the Office of Natural
Resources Revenue Financial
Management Program Manager. The
Direct Final Rule updates the address
and phone number.
E:\FR\FM\22SER1.SGM
22SER1
Agencies
[Federal Register Volume 80, Number 183 (Tuesday, September 22, 2015)]
[Rules and Regulations]
[Pages 57090-57092]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2015-24026]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA-2015-N-3044]
Medical Devices; Ophthalmic Devices; Classification of the Oral
Electronic Vision Aid
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the oral
electronic vision aid into class II (special controls). The special
controls that will apply to the device are identified in this order and
will be part of the codified language for the oral electronic vision
aid's classification. The Agency is classifying the device into class
II (special controls) in order to provide a reasonable assurance of
safety and effectiveness of the device.
DATES: This order is effective September 22, 2015. The classification
was applicable on June 18, 2015.
FOR FURTHER INFORMATION CONTACT: Dexiu Shi, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2246, Silver Spring, MD, 20993-0002, 301-796-6470,
dexiu.shi@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of the Federal Food, Drug, and
Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were
not in commercial distribution before May 28, 1976 (the date of
enactment of the Medical Device Amendments of 1976), generally referred
to as postamendments devices, are classified automatically by statute
into class III without any FDA rulemaking process. These devices remain
in class III and require premarket approval, unless and until the
device is classified or reclassified into class I or II, or FDA issues
an order finding the device to be substantially equivalent, in
accordance with section 513(i), to a predicate device that does not
require premarket approval. The Agency determines whether new devices
are substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act (21 U.S.C. 360c(f)(2)), as
amended by section 607 of the Food and Drug Administration Safety and
Innovation Act (Pub. L. 112-144), provides two procedures by which a
person may request FDA to classify a device under the criteria set
forth in section 513(a)(1). Under the first procedure, the person
submits a premarket notification under section 510(k) of the FD&C Act
for a device that has not previously been classified and, within 30
days of receiving an order classifying the device into class III under
section 513(f)(1), the person requests a classification under section
513(f)(2). Under the second procedure, rather than first submitting a
premarket notification under section 510(k) and then a request for
classification under the first procedure, the person determines that
there is no legally marketed device upon which to base a determination
of substantial equivalence and requests a classification under section
513(f)(2) of the FD&C Act. If the person submits a request to classify
the device under this second procedure, FDA may decline to undertake
the classification request if FDA identifies a legally marketed device
that could provide a reasonable basis for review of substantial
equivalence with the device or if FDA determines that the device
submitted is not of ``low-moderate risk'' or that general controls
would be inadequate to control the risks and special controls to
mitigate the risks cannot be developed.
In response to a request to classify a device under either
procedure provided by section 513(f)(2) of the FD&C Act, FDA will
classify the device by written order within 120 days. This
classification will be the initial classification of the device.
On August 7, 2013, Wicab Inc., submitted a request for
classification of the BrainPort V100 under section 513(f)(2) of the
FD&C Act. The manufacturer recommended that the device be classified
into class II (Ref. 1).
In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed
the request in order to classify the device under the criteria for
classification set forth in section 513(a)(1). FDA classifies
[[Page 57091]]
devices into class II if general controls by themselves are
insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls to provide reasonable assurance of the safety and
effectiveness of the device for its intended use. After review of the
information submitted in the request, FDA determined that the device
can be classified into class II with the establishment of special
controls. FDA believes these special controls, in addition to general
controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on June 18, 2015, FDA issued an order to the requestor
classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 886.5905.
Following the effective date of this final classification order,
any firm submitting a premarket notification (510(k)) for an oral
electronic vision aid will need to comply with the special controls
named in this final order. The device is assigned the generic name oral
electronic vision aid, and it is identified as a battery-powered
prescription device that contains an electrode stimulation array to
generate electrotactile stimulation patterns that are derived from
digital object images captured by a camera. It is intended to aid
profoundly blind patients in orientation, mobility, and object
recognition as an adjunctive device to other assistive methods such as
a white cane or a guide dog.
FDA has identified the following risks to health associated
specifically with this type of device, as well as the mitigation
measures required to mitigate these risks in table 1.
Table 1--Oral Electronic Vision Aid Risks and Mitigation Measures
------------------------------------------------------------------------
Identified risk Mitigation method
------------------------------------------------------------------------
Irritation, Discomfort or Adverse Clinical Testing.
Events Involving the Mouth, Tongue, or Labeling.
Gums.
Adverse Tissue Reaction................ Biocompatibility Testing.
Labeling.
Unit (Hardware) Malfunction, Functional Non-Clinical Performance
Reliability. Testing.
Clinical Testing.
Labeling.
Software Malfunction................... Software Verification,
Validation, and Hazard
Analysis.
Use Error.............................. Clinical Testing.
Healthcare Professional
Training.
Patient Training.
Labeling.
Interference with Other Devices........ Electromagnetic Compatibility
and Electromagnetic
Interference Testing.
Wireless Coexistence Testing.
Labeling.
Electrical Shock....................... Electrical Safety Testing.
Labeling.
------------------------------------------------------------------------
FDA believes that the following special controls, in combination
with the general controls, address these risks to health and provide
reasonable assurance of the safety and effectiveness:
Clinical performance testing must demonstrate an
acceptable adverse event profile, including adverse events involving
the mouth, tongue, and gums and demonstrate the effect of the
stimulation to provide clinically meaningful outcomes. The clinical
performance testing must also investigate the anticipated conditions of
use, including potential use error, intended environment of use, and
duration of use.
Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use,
including simulated moisture ingress, device durability, and battery
reliability.
Software verification, validation, and hazard analysis
must be performed.
Analysis/testing must validate electromagnetic
compatibility.
Analysis/testing must validate electrical safety.
Analysis/testing must assess and validate wireless
coexistence concerns.
Any elements of the device that contact the patient must
be demonstrated to be biocompatible.
Training must include elements to ensure that the
healthcare provider and user can identify the safe environments for
device use, use all safety features of the device, and operate the
device in the intended environment of use.
Labeling for the trainer and user must include a summary
of the clinical testing including adverse events encountered under use
conditions, summary of study outcomes and endpoints, and information
pertinent to use of the device including the conditions under which the
device was studied (e.g., level of supervision or assistance, and
environment of use).
Oral electronic vision aid devices are prescription devices
restricted to patient use only upon the authorization of a practitioner
licensed by law to administer or use the device; see 21 CFR 801.109
(Prescription devices).
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary to provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, this device type is not exempt from premarket notification
requirements. Persons who intend to market this type of device must
submit to FDA a premarket notification, prior to marketing the device,
which contains information about the oral electronic vision aid they
intend to market.
II. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
[[Page 57092]]
III. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in
part 807, subpart E, regarding premarket notification submissions have
been approved under OMB control number 0910-0120, and the collections
of information in 21 CFR part 801, regarding labeling have been
approved under OMB control number 0910-0485.
IV. Reference
The following reference has been placed on display in the Division
of Dockets Management (HFA-305), Food and Drug Administration, 5630
Fishers Lane, rm. 1061, Rockville, MD 20852, and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday,
and is available electronically at https://www.regulations.gov.
1. DEN130039: De Novo Request per 513(f)(2) from Wicab Inc., dated
August 7, 2013.
List of Subjects in 21 CFR Part 886
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
886 is amended as follows:
PART 886--OPHTHALMIC DEVICES
0
1. The authority citation for 21 CFR part 886 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Add Sec. 886.5905 to subpart F to read as follows:
Sec. 886.5905 Oral electronic vision aid.
(a) Identification. An oral electronic vision aid is a battery-
powered prescription device that contains an electrode stimulation
array to generate electrotactile stimulation patterns that are derived
from digital object images captured by a camera. It is intended to aid
profoundly blind patients in orientation, mobility, and object
recognition as an adjunctive device to other assistive methods such as
a white cane or a guide dog.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Clinical performance testing must demonstrate an acceptable
adverse event profile, including adverse events involving the mouth,
tongue, and gums and demonstrate the effect of the stimulation to
provide clinically meaningful outcomes. The clinical performance
testing must also investigate the anticipated conditions of use,
including potential use error, intended environment of use, and
duration of use.
(2) Non-clinical performance testing must demonstrate that the
device performs as intended under anticipated conditions of use,
including simulated moisture ingress, device durability, and battery
reliability.
(3) Software verification, validation, and hazard analysis must be
performed.
(4) Analysis/testing must validate electromagnetic compatibility.
(5) Analysis/testing must validate electrical safety.
(6) Analysis/testing must assess and validate wireless coexistence
concerns.
(7) Any elements of the device that contact the patient must be
demonstrated to be biocompatible.
(8) Training must include elements to ensure that the healthcare
provider and user can identify the safe environments for device use,
use all safety features of the device, and operate the device in the
intended environment of use.
(9) Labeling for the trainer and user must include a summary of the
clinical testing including adverse events encountered under use
conditions, summary of study outcomes and endpoints, and information
pertinent to use of the device including the conditions under which the
device was studied (e.g., level of supervision or assistance, and
environment of use).
Dated: September 16, 2015.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2015-24026 Filed 9-21-15; 8:45 am]
BILLING CODE 4164-01-P