Effective Date of Requirement for
Premarket Approval for Automated
External Defibrillator Systems
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA or the Agency) is
issuing a final order to require the filing
of premarket approval applications
(PMA) for automated external
defibrillator (AED) systems, which
consist of an AED and those AED
accessories necessary for the AED to
detect and interpret an
electrocardiogram and deliver an
electrical shock (e.g., pad electrodes,
batteries, adapters, and hardware keys
for pediatric use).
DATES: This order is effective on January
29, 2015.
FOR FURTHER INFORMATION CONTACT:
Linda Ricci, Center for Devices and
Radiological Health, 10903 New
Hampshire Ave., Bldg. 66, Rm. 1314,
Silver Spring, MD 20993, 301–796–
6325, linda.ricci@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
SUMMARY:
tkelley on DSK3SPTVN1PROD with RULES
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act), as amended by the
Medical Device Amendments of 1976
(the 1976 amendments) (Pub. L. 94–
295), the Safe Medical Devices Act of
1990 (Pub. L. 101–629), the Food and
Drug Administration Modernization Act
of 1997 (Pub. L. 105–115), the Medical
Device User Fee and Modernization Act
of 2002 (Pub. L. 107–250), the Medical
Devices Technical Corrections Act (Pub.
L. 108–214), the Food and Drug
Administration Amendments Act of
2007 (Pub. L. 110–85), and the Food and
Drug Administration Safety and
Innovation Act (FDASIA) (Pub. L. 112–
144), among other amendments,
established a comprehensive system for
the regulation of medical devices
intended for human use. Section 513 of
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
the FD&C Act (21 U.S.C. 360c)
established three categories (classes) of
devices, reflecting the regulatory
controls needed to provide reasonable
assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Under section 513(d) of the FD&C Act
(21 U.S.C. 360c(d)), devices that were in
commercial distribution before the
enactment of the 1976 amendments,
May 28, 1976 (generally referred to as
‘‘preamendments devices’’), are
classified after FDA has: (1) Received a
recommendation from a device
classification panel (an FDA advisory
committee); (2) published the panel’s
recommendation for comment, along
with a proposed regulation classifying
the device; and (3) published a final
regulation classifying the device. FDA
has classified most preamendments
devices under these procedures.
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
‘‘postamendments devices’’) are
automatically classified by section
513(f) of the FD&C Act (21 U.S.C.
360c(f)) into class III without any FDA
rulemaking process. Those devices
remain in class III and require
premarket approval unless, and until,
the device is reclassified into class I or
II or FDA issues an order finding the
device to be substantially equivalent, in
accordance with section 513(i) of the
FD&C Act (21 U.S.C. 360c(i)), to a
predicate device that does not require
premarket approval. The Agency
determines whether new devices are
substantially equivalent to predicate
devices by means of premarket
notification procedures in section 510(k)
of the FD&C Act (21 U.S.C. 360(k)) and
part 807 (21 CFR part 807).
A preamendments device that has
been classified into class III and devices
found substantially equivalent by means
of premarket notification (510(k))
procedures to such a preamendments
device or to a device within that type
(both the preamendments and
substantially equivalent devices are
referred to as preamendments class III
devices) may be marketed without
submission of a premarket approval
application (PMA) until FDA issues a
final order under section 515(b) of the
FD&C Act (21 U.S.C. 360e(b)) requiring
premarket approval or until the device
is subsequently reclassified into class I
or class II. Section 515(b)(1) of the FD&C
Act (21 U.S.C. 360e(b)(1)) directs FDA to
issue an order requiring premarket
approval for a preamendments class III
device.
PO 00000
Frm 00027
Fmt 4700
Sfmt 4700
4783
Although, under the FD&C Act, the
manufacturer of a class III
preamendments device may respond to
the call for PMAs by filing a PMA or a
notice of completion of a product
development protocol (PDP), in
practice, the option of filing a notice of
completion of a PDP has not been used.
For simplicity, although corresponding
requirements for PDPs remain available
to manufacturers in response to a final
order under section 515(b) of the FD&C
Act (21 U.S.C. 360e(b)), this document
will refer only to the requirement for the
filing and receiving approval of a PMA.
On July 9, 2012, FDASIA was enacted.
Section 608(a) of FDASIA (126 Stat.
1056) amended section 513(e) of the
FD&C Act (21 U.S.C. 360c(e)), changing
the mechanism for reclassifying a device
from rulemaking to an administrative
order. Section 608(b) of FDASIA
amended section 515(b) of the FD&C Act
(21 U.S.C. 360e(b)) changing the
mechanism for requiring premarket
approval for a preamendments class III
device from rulemaking to an
administrative order.
Section 515(b)(1) of the FD&C Act sets
forth the process for issuing a final
order. Specifically, prior to the issuance
of a final order requiring premarket
approval for a preamendments class III
device, the following must occur: (1)
Publication of a proposed order in the
Federal Register; (2) a meeting of a
device classification panel described in
section 513(b) of the FD&C Act; and (3)
consideration of comments from all
affected stakeholders, including
patients, payers, and providers.
Section 515(b)(3) of the FD&C Act
provides that FDA shall, after the close
of the comment period on the proposed
order, consideration of any comments
received, and a meeting of a device
classification panel described in section
513(b) of the FD&C Act, issue a final
order to require premarket approval or
publish a document terminating the
proceeding together with the reasons for
such termination.
A preamendments class III device
may be commercially distributed
without a PMA until 90 days after FDA
issues a final order (a final rule issued
under section 515(b) of the FD&C Act
prior to the enactment of FDASIA is
considered to be a final order for
purposes of section 501(f) of the FD&C
Act (21 U.S.C. 351(f))) requiring
premarket approval for the device, or 30
months after final classification of the
device under section 513 of the FD&C
Act, whichever is later. For AED
systems, the later of these two time
periods is the 90-day period. Therefore,
section 501(f)(2)(B) of the FD&C Act (21
U.S.C. 351(f)(2)(B)) requires that a PMA
E:\FR\FM\29JAR1.SGM
29JAR1
4784
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
tkelley on DSK3SPTVN1PROD with RULES
for such devices be filed within 90 days
of the effective date of a final order.
However, for the reasons discussed
below, FDA does not intend to enforce
compliance with the 90-day deadline for
PMA submissions for currently
marketed AEDs and those AED
accessories identified in 21 CFR
870.5310(a) (see further discussion in
section V, ‘‘Implementation Strategy’’).
Also, a preamendments device subject
to the order process under section
515(b) of the FD&C Act (21 U.S.C. 360e)
is not required to have an approved
investigational device exemption (IDE)
(see part 812 (21 CFR part 812))
contemporaneous with its interstate
distribution until the date identified by
FDA in the final order requiring the
filing of a PMA for the device. At that
time, an IDE is required only if a PMA
has not been filed. If the manufacturer,
importer, or other sponsor of the device
submits an IDE application and FDA
approves it, the device may be
distributed for investigational use. If a
PMA is not filed by the later of the two
dates, and the device is not distributed
for investigational use under an IDE, the
device is deemed to be adulterated
within the meaning of section
501(f)(1)(A) of the FD&C Act (21 U.S.C.
351(f)(1)(A)), and subject to seizure and
condemnation under section 304 of the
FD&C Act (21 U.S.C. 334) if its
distribution continues. Other
enforcement actions include, but are not
limited to, the following: Shipment of
devices in interstate commerce may be
subject to injunction under section 302
of the FD&C Act (21 U.S.C. 332), and the
individuals responsible for such
shipment may be subject to prosecution
under section 303 of the FD&C Act (21
U.S.C. 333). FDA requests that
manufacturers take action to prevent the
further use of devices for which no PMA
has been filed.
II. Regulatory History of This Device
On January 25, 2011, the Circulatory
System Devices Panel (‘‘Panel’’)
recommended that AED systems be
classified as class III devices and subject
to premarket approval to provide
reasonable assurance of the safety and
effectiveness of the device (Ref. 1). The
Panel recommended that AED systems
be regulated as class III devices because,
among other things, they are lifesaving
devices. Furthermore, the problems
identified in adverse events in the
medical device reporting systems and
recalls related to AED systems indicated
these devices require more regulatory
oversight.
FDA published a proposed order to
require PMAs for AED systems in the
Federal Register of March 25, 2013 (78
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
FR 17890). FDA is now requiring PMAs
for AED systems, which include AED
accessories necessary for the
functionality of the AED (e.g., pad
electrodes, batteries, adapters, and
hardware keys for pediatric use)
(‘‘necessary AED accessories’’) (see
section IV, ‘‘The Final Order’’).
FDA received and has considered
comments on the AED systems
proposed order as discussed in section
III of this document.
III. Public Comments in Response to the
Proposed Order
In response to the March 25, 2013 (78
FR 17890) proposed order to maintain
the class III classification and require
premarket approval for AED systems,
FDA received 66 comments and one
petition for reclassification (see FDA–
2013–N–0234–0002).1 The comments
and the FDA’s responses to the
comments are summarized below.
Certain comments are grouped together
under a single number because the
subject matter of the comments is
similar. The number assigned to each
comment is purely for organizational
purposes and does not signify the
comment’s value or importance or the
order in which it was submitted.
(Comment 1) Many comments
indicated that AED systems have
already been demonstrated to be safe
and effective, and referenced literature
and studies supporting the reliability of
these devices and the value of AED
systems in treating sudden cardiac
arrest (SCA). The comments stated that
PMAs and associated increased
regulatory cost and review time is not
warranted and would hinder
innovation, increase device cost to
consumers, and reduce availability of
AED systems. The comments further
stated that it is widely recognized that
improvement in the survival rate from
SCA is due in large part to widespread
distribution of AED systems and
expressed concern that requiring PMAs
would limit availability of the devices.
(Response 1) FDA agrees that many
currently marketed AEDs have been
demonstrated to be effective in clinical
use and, when designed and
manufactured appropriately, AEDs can
be safe and effective. However, FDA
believes that there is insufficient
information to determine that general
and special controls would provide a
reasonable assurance of the safety and
effectiveness of these devices, which are
1 FDA will respond separately to the
reclassification petition and will address the issues
raised in that petition in its response. The
reclassification petition is available at https://
www.regulations.gov/#!documentDetail;D=FDA2013-N-0234-0002.
PO 00000
Frm 00028
Fmt 4700
Sfmt 4700
for a use in supporting or sustaining
human life (see section 513(a)(1)(C) of
the FD&C Act (21 U.S.C. 360c(a)(1)(C))).
Specifically, the postmarket information
on AEDs supports increased regulatory
review to ensure that device design and
manufacturing practices provide a
reasonable assurance of safety and
effectiveness. FDA acknowledges that
the PMA process may result in
increased regulatory cost to
manufacturers; however, FDA believes
that device quality will improve, which
will reduce costs associated with
postmarket actions including recalls.
FDA also agrees that continued efforts
to make safe and effective AED systems
available is in the interest of public
health, but disagrees that this call for
PMAs will limit device availability.
FDA believes that many manufacturers
of currently marketed AEDs already
have, or can reasonably obtain, the
necessary data to support a PMA, and
hence expects AED distribution to
continue to meet demand. Also, for the
reasons discussed below, FDA does not
intend to enforce compliance with the
90-day deadline for submission of PMAs
for currently marketed AEDs and
necessary AED accessories (for further
discussion see section V,
‘‘Implementation Strategy’’).
At the January 2011 Panel meeting,
the Panel discussed the impact of FDA
regulatory scrutiny on innovation.
Various Panel members agreed that the
appropriate focus should be on assuring
reliability of AEDs and that there was no
evidence presented to indicate that a
call for PMAs would unduly hinder
device innovation (Ref. 1). FDA notes
that previous significant innovations for
AED systems (e.g., new defibrillation
waveforms) have been supported by
clinical evidence in the 510(k) process
and that under the PMA process this
clinical evidence is not expected to
significantly change. As was mentioned
in the proposed order, FDA anticipates
that many AED manufacturers already
have sufficient clinical evidence to
support a PMA.
(Comment 2) Several comments noted
that AED system failures are often the
result of use error or improper
maintenance (e.g., expired batteries/
pads, periodic checks not performed,
etc.) and not of system failure or
malfunction. The comments stated that
efforts should be devoted to ensuring
appropriate public awareness, training
(particularly for lay users), and
maintenance to address these issues as
opposed to increasing premarket
regulatory review. One comment stated
that the proposed order should not be
finalized until all stakeholders, not only
device manufacturers, are engaged in an
E:\FR\FM\29JAR1.SGM
29JAR1
tkelley on DSK3SPTVN1PROD with RULES
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
integrated approach to increase the
likelihood that AED systems will be
available and functional when needed.
(Response 2) FDA agrees that AED
system training and maintenance are
important to help ensure AED system
availability and proper use and also
believes manufacturers and users are in
the best position to develop and
implement training and maintenance
materials. FDA supports ongoing
discussions and efforts to improve
training and maintenance, but disagrees
that these activities should delay
finalizing the requirement for PMAs for
these devices. Although we recognize
that there have been some medical
device reports (MDRs) associated with
use errors, the focus of FDA’s review of
MDRs and recalls of AED systems has
been related to problems with the
quality of these devices as related to
device design and manufacture and
FDA continues to believe that requiring
PMAs is appropriate.
(Comment 3) Several comments stated
that special controls, including
performance testing to industry
standards, device labeling, guidance
documents, human factors analysis and
design, summary of field actions and
mitigations to address Quality System
(QS) concerns, risk management, and
post-market surveillance were sufficient
to regulate AED systems as class II
devices under the existing 510(k)
regulatory regime. One comment
indicated that several of the regulatory
controls identified by FDA as consistent
with PMA requirements—such as premarket inspections, review of changes
that could significantly affect the safety
or effectiveness of the device, and
postmarket surveillance—could also be
conducted under the 510(k) regime.
Other comments supported FDA’s
proposal to maintain the devices in
class III and agreed that the
manufacturing controls, premarket
review requirements, and assessment of
lay use are best managed under the
PMA process.
(Response 3) FDA disagrees that there
is sufficient information to determine
that general and special controls would
provide a reasonable assurance of safety
and effectiveness of these devices given
safety concerns related to the
manufacturing processes and design
changes, problems which FDA
considered in determining that PMAs
are warranted (see section 513(a)(1)(C)
of the FD&C Act (21 U.S.C.
360c(a)(1)(C))). FDA does not generally
conduct preclearance inspections under
the 510(k) process because such
information is not required in a 510(k)
submission under the FD&C Act or FDA
regulations. Further, under section
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
513(f)(5) of the FD&C Act (21 U.S.C.
360c(f)(5)), FDA may not withhold a
510(k) ‘‘because of a failure to comply
with any provision of this Act unrelated
to a substantial equivalence decision,
including a finding that the facility in
which the device is manufactured is not
in compliance with good manufacturing
requirements as set forth in regulations
of the Secretary under section 520(f)
(other than a finding that there is a
substantial likelihood that the failure to
comply with such regulations will
potentially present a serious risk to
human health).’’ In contrast, under
section 515(c)(1)(C) of the FD&C Act (21
U.S.C. 360e(c)(1)(C)), a PMA must
include ‘‘a full description of the
methods used in, and the facilities and
controls used for, the manufacturing,
processing, and when relevant, packing
and installation of, such device.’’
Moreover, many of the design and
manufacturing changes that have led to
AED system recalls were not required to
be reported to FDA under the 510(k)
process. If these changes had been
reported prior to implementation, as
would be required in the PMA regime,
these recalls may have been avoided.
FDA continues to believe that the
necessary regulatory controls for AED
systems are consistent with the PMA
review process, and that the 510(k)
process does not provide sufficient
regulatory oversight for these devices.
Similarly, FDA’s oversight of
postmarket changes to devices is very
different in the 510(k) context as
compared to the PMA context. Under 21
CFR 807.81, FDA requires 510(k)s for a
change to a device only when the
change ‘‘could significantly affect the
safety or effectiveness of the device, e.g.,
a significant change or modification in
design, material, chemical composition,
energy source, or manufacturing
process. ’’ In contrast, under 21 CFR
814.39, FDA requires PMA supplements
(including 30-day notices) for any
change to a PMA-approved device that
affects safety or effectiveness. These
differences in authorities, among the
other reasons discussed above, warrant
regulation of AEDs in class III.
(Comment 4) A few comments
indicated that existing AED and AED
accessory manufacturers are already
subject to the QS regulation (21 CFR
820) and manufacturing quality would
not be measurably improved as a result
of requiring PMAs. One comment noted
that specific expectations under the QS
regulation for design controls,
purchasing controls, and other issues
identified by FDA as problematic for
AEDs could be addressed by special
controls and other regulations, and
AEDs could remain in class II. One
PO 00000
Frm 00029
Fmt 4700
Sfmt 4700
4785
comment further stated that such
concerns could be managed via
postmarket controls, which are available
under the 510(k) regime, such as
submission of a summary of recent field
actions and related design mitigations.
(Response 4) FDA disagrees with the
comments. FDA acknowledges that AED
and AED accessory manufacturers are
already subject to the QS regulation and
that QS requirements result in
rigorously designed and manufactured
devices and resultant quality
improvements. By requiring premarket
review of QS processes as well as device
changes for AEDs, FDA believes the
PMA process will provide a reasonable
assurance of safety and effectiveness
(see Response 3 above).
(Comment 5) One comment stated
that certain AED accessories,
specifically electrodes, cables, and
adapters, are well-understood devices
and that their classification into class III
is not warranted. The comment stated
that these accessories could be
adequately regulated in class II with
special controls, as is already the case
when these accessories are used with
manual defibrillators. The comment
recommended special controls,
including the following: Performance
testing, usability evaluation, labeling,
biocompatibility, and readiness for use.
Two comments stated that because AED
accessories often have identical designs
and the same intended use as
accessories used with class II manual
defibrillators, FDA should not perform
duplicative reviews under both the
510(k) and PMA regimes and that PMA
review should be required only when
use of the accessory with an AED results
in a change in intended use or design.
(Response 5) Accessories necessary
for an AED to detect and interpret an
electrocardiogram and deliver an
electrical shock (e.g., battery, pad
electrode, adapter, and hardware keys
for pediatric use) are necessary for AED
system functionality. Failure of these
necessary accessories leads to the same
negative outcomes as a failure of the
AED itself; e.g., an AED not ready for
use because of a faulty battery is unable
to detect heart rhythm abnormalities
and/or deliver a defibrillation shock to
a victim of SCA. FDA’s review of
adverse events and recalls has shown
that problems with AED accessories
have occurred during clinical use. As
such, FDA continues to believe that the
same regulatory oversight is warranted
for certain critical accessories (i.e.,
batteries, pad electrodes, adapters, and
hardware keys for pediatric use) as for
the AEDs with which they are used. As
discussed in the response to Comment
3 above, FDA does not believe that
E:\FR\FM\29JAR1.SGM
29JAR1
tkelley on DSK3SPTVN1PROD with RULES
4786
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
adequate regulatory controls are
available under the 510(k) process, and
hence PMAs are necessary to provide a
reasonable assurance of safety and
effectiveness.
(Comment 6) Several comments
questioned the validity of FDA’s data
regarding adverse events associated
with AED failures. One comment noted
that FDA provided no data in the
proposed order on the frequency of
adverse events or relationship of
number of events to total distribution
and use of AEDs. The comment
requested additional information from
FDA to support the validity of the MDR
analysis presented at the 2011 Panel and
relied upon to support the proposed
order. A few comments presented
alternate analyses of MDR data that
suggested that MDRs for AEDs are not
increasing. One comment presented an
analysis that showed no statistically
significant increase in the rate of
adverse event reports over the time
period of 2007 to 2011. Two comments
stated that a majority of AED MDRs
reported to FDA resulted from self-test
errors—which are reported as
malfunction MDRs because they could
cause or contribute to a death or serious
injury but do not represent device
failures in clinical use. The comments
contended that any analysis of MDRs
should focus instead on actual use
adverse events, which would represent
a small subset of the overall MDRs. One
comment stated that self-test related
events are representative of an effective
design risk mitigation strategy being
employed for AEDs and that because
AEDs are often in standby for a large
percentage of time, self-test detection of
problems before use should not be
included in the overall assessment of
the benefit-risk profile for AEDs. Two
comments requested further guidance
from FDA on MDR reporting
expectations for AEDs.
(Response 6) Although FDA requires
manufacturers to submit an MDR when
their device may have caused or
contributed to a death, serious injury, or
in certain situations when their device
has malfunctioned, FDA acknowledges
that there are limitations on the review
of MDR data, including the fact that
FDA typically does not have complete
information on the number of devices in
distribution from which to calculate
adverse events rates. These limitations
were discussed at the 2011 Panel
meeting. FDA has previously stated that
fatality statistics and injury statistics
from MDRs should be considered in
light of underreporting (58 FR 61952 at
61972, November 23, 1993). In addition,
FDA notes that the evaluation of MDR
data for AEDs was focused on
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
manufacturing and design concerns and
was not aimed at developing specific
failure rates. Moreover, FDA believes
that the large number of devices in
distribution and the life-saving nature of
these devices combined with the steady
rate of MDRs support a call for PMAs to
help ensure that these devices are
adequately designed and manufactured
so that they are available when needed.
FDA disagrees that evaluation of
adverse events should focus only on
those events that occur during clinical
use. Although some distributed AEDs
may seldom be used, this does not
reduce the importance that they are safe
and effective when needed. FDA
acknowledges the importance of AED
self-test features and recognizes that
many self-test failures are not indicative
of issues with overall device quality.
FDA believes, however, that some selftest failures signal significant quality
problems arising from device design or
manufacturing issues and are
appropriately considered as adverse
events if recurrence of such failures
could, for example, render the device
unavailable for use when needed. FDA
also recognizes that some MDRs may
eventually be found to be the result of
problems not associated with the
device; however, this concern is
applicable to all devices subject to
adverse event reporting requirements
and FDA does not believe such reports
unduly influence overall reporting
numbers.
FDA also notes that our review of
available information, as presented at
the January 2011 Panel meeting,
included data on voluntary corrections
and removals (i.e., ‘‘recalls’’) of AEDs
pursuant to section 519(g) of the FD&C
Act (21 U.S.C. 360i(g)). Recalls are
conducted ‘‘(A) to reduce a risk to
health posed by the device, or (B) to
remedy a violation of this Act caused by
the device which may present a risk to
health,’’ and as such may reflect safety
concerns for AEDs (section 519(g)(1) of
FD&C Act (21 U.S.C. 360i(g)(1)). Since
the January 2011 Panel meeting, over 40
additional class I or class II recalls have
been conducted by AED manufacturers
and have impacted over 2 million
distributed AEDs (Ref. 2). The root
cause of these recalls has been
attributed to a variety of causes, with
design controls, purchasing controls,
and receiving acceptance activities
being the most common. FDA continues
to believe that the recall data reinforces
the overall conclusion regarding the
inadequacy of regulatory controls for
AED systems under the 510(k) process.
Additional guidance on MDR
requirements for AEDs is beyond the
scope of this document; however, FDA
PO 00000
Frm 00030
Fmt 4700
Sfmt 4700
intends to continue efforts to clarify
medical device reporting expectations
and manufacturers who have questions
regarding their reporting obligations
should contact FDA.
(Comment 7) Several comments
responded to FDA’s request for feedback
regarding whether 15 months is
sufficient to allow companies to collect
information necessary to support
submission of a PMA. Two comments
stated that this issue was dependent on
the data expected by FDA and that FDA
should provide more guidance in this
respect. One comment requested
clarification on what clinical data is
known to FDA that would support a
PMA because it is critical that AED
manufacturers understand the type and
amount of data that will be required.
One comment stated that it is unclear
what FDA’s expectations would be for
clinical trials of new AEDs or the need
for clinical trials for AED accessories
given available less burdensome
methods for obtaining performance data
on accessories. Another comment
requested clarification on whether AED
manufacturers would be expected to retest and re-validate older AED models to
currently recognized standards. One
comment requested clarification on
when marketing materials for AEDs
would need to comply with 21 CFR
801.109.
One comment suggested that the 15month period should be extended to 30
months, which the commenter claimed
would be consistent with section
501(f)(2)(B) of the FD&C Act (21 U.S.C.
351(f)(2)(B). One commenter requested
clarification regarding whether the 15
months started at the 90th day after a
final order was issued and another
comment indicated that 15 months
should be sufficient, but that the 15
months should not include FDA’s 180day PMA review time. One comment
suggested that FDA require PMAs 90
days after the final order.
(Response 7) The data required to
support premarket approval will vary by
device and the specific data
requirements. FDA is aware of clinical
study information that can be leveraged
for AEDs from both published studies
and clinical data previously submitted
to FDA under the 510(k) process, and,
as was stated in the proposed order,
FDA believes that many AED
accessories ‘‘may need to submit nonclinical performance testing with
confirmatory animal studies in order to
support independent PMA approval’’
(78 FR 17894, March 25, 2013).
Performance testing of AEDs must be
provided in a PMA to support a
reasonable assurance of safety and
effectiveness. Although retesting older
E:\FR\FM\29JAR1.SGM
29JAR1
tkelley on DSK3SPTVN1PROD with RULES
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
AED models to currently recognized
standards is one way to meet the
performance testing requirements,
compliance with such standards is
voluntary and manufacturers may
submit a justification for how other
testing conducted on their devices
provides equivalent assurances of safety
and effectiveness. FDA encourages
manufacturers to proactively engage
FDA via the pre-submission process to
discuss the specific data needed for
their PMAs (Ref. 3). FDA notes that
existing prescription AEDs are already
subject to 21 CFR 801.109, and will
remain so after this call for PMAs. FDA
review of AED PMAs will include
review of the associated AED labeling to
ensure such device labeling complies
with regulatory requirements.
FDA notes that the 30 months
discussed in section 501(f)(2)(B) of the
FD&C Act (21 U.S.C. 351(f)(2)(B))
references the date from initial
classification of a device into class III.
AEDs have been classified as class III for
more than 30 months, and hence this
statutory provision has expired. FDA
also acknowledges that it is in the
interest of public health to ensure the
availability of AEDs because they are
life-saving devices and their clinical use
is well-established. After consideration
of the comments, FDA continues to
believe that the proposed 15 months for
filing a PMA (Ref. 4) strikes an
appropriate balance between the need to
ensure continued availability of AEDs
for the public health reasons stated
above and the implementation of PMA
requirements to ensure the safety and
effectiveness of AEDs.
For currently marketed AEDs, FDA
does not intend to enforce compliance
with the 90-day deadline by which
PMAs must be submitted for 15 months
after that deadline (i.e., 18 months after
the effective date of the final order), as
long as a notice of intent to file a PMA
is submitted within 90 days of the
effective date of the final order (see
section V, ‘‘Implementation Strategy’’).
Even if a notice of intent and PMA are
submitted by these dates, manufacturers
must cease distribution of devices upon
receiving a not approvable or denial
decision rendered on a PMA. To resume
distribution, these manufacturers must
receive PMA approval for their devices.
Moreover, for currently marketed
necessary AED accessories, FDA does
not intend to enforce compliance with
the 90-day deadline by which PMAs
must be submitted for 57 months after
that deadline (i.e., 5 years after the
effective date of the final order) (see
section V, ‘‘Implementation Strategy’’).
Continued availability of necessary AED
accessories, including consumable
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
accessory items (e.g., pad electrodes)
and accessories with limited useful life
(e.g., batteries), is critical to ensuring the
safety and efficacy of currently
marketed AEDs during the time while
PMAs for those AEDs are being pursued.
In addition, the continued availability of
necessary accessories for ‘‘legacy
devices’’—individual AEDs that have
been distributed and are currently in
use (e.g., in public facilities, etc.) and
for which the manufacturer is not
seeking PMA approval for that AED
model—ensures the availability of
functional legacy AEDs until they are
replaced with PMA-approved AEDs.
(Comment 8) One commenter stated
that FDA did not have a legal basis for
continuing with finalization of a call for
PMAs for AED systems because FDA
failed to convene a panel as is required
under FDASIA prior to issuing a final
order. The commenter stated that FDA
may not rely on the 2011 pre-FDASIA
Panel because that Panel meeting was
related to reclassifications under section
515(i) of the FD&C Act (21 U.S.C.
360e(i)) and not related to calls for
PMAs under section 515(b) of the FD&C
Act (21 U.S.C. 360e(b)). The commenter
further contended that the 2011 Panel
neither considered new information
contained in a reclassification petition
submitted to FDA in 2009 nor
adequately discussed the
appropriateness of class II special
controls.
(Response 8) FDA disagrees with the
comment that FDA does not have a legal
basis to finalize an order calling for
PMAs for AED systems. Pursuant to
FDASIA, the amendments to section
515(b) of the FD&C Act require, in
relevant part, that issuance of an
administrative order calling for PMAs
for a preamendments device be
preceded by a proposed order and a
meeting of a classification panel. As
amended, this section of the FD&C Act
does not prescribe when these two
events (the panel and proposed order)
must occur in relation to each other.
More importantly, FDA believes that the
Panel’s deliberations and
recommendations remain relevant and
fully satisfy the requirements in section
515(b) of FD&C Act.
FDA disagrees with the comment that
the Panel did not consider new
information contained in the 2009
reclassification petition. A
representative from the petitioner was
present at the meeting and provided
comments on the reclassification
petition during the Panel meeting (Ref.
1). In addition, the petitioner was given
an opportunity to explain the
petitioner’s reasons for why AEDs
should be class II devices, including a
PO 00000
Frm 00031
Fmt 4700
Sfmt 4700
4787
discussion of the special controls
described in the reclassification
petition. Therefore, the Panel heard the
petitioner’s arguments and these
arguments were available for the Panel’s
consideration when it made its
recommendation.
(Comment 9) One commenter objected
to FDA’s use of the term ‘‘diagnose’’ in
the proposed order to describe the
functionality of AEDs (78 FR 17893,
March 25, 2013), and stated that AEDs
sense shockable rhythms and are not
diagnostic devices.
(Response 9) FDA disagrees that these
devices do not perform diagnostic
functions. AEDs analyze and interpret
ECG data to produce an assessment as
to whether a shock should be delivered;
while FDA does believe that AEDs have
diagnostic functions, we note that the
regulatory identification for the device
in 21 CFR 870.5310(a), as finalized in
the order, does not use the term
diagnose, and instead describes the
function of the device as ‘‘analyzes’’ and
‘‘interprets.’’
(Comment 10) One commenter stated
that FDA’s proposal to allow
manufacturers to ‘‘bundle’’ several AED
models under a single PMA is
inconsistent with the PMA regulatory
paradigm, which relies on a device-bydevice assessment. The comment points
to FDA’s guidance on bundling, which
states that ‘‘[g]enerally, [manufacturers]
should not bundle differing generic
device types in a single PMA
submission because of the substantially
different pre-clinical and clinical data
needed to support each of the devices’’
(Ref. 5).
(Response 10) FDA disagrees with the
comment. Different AED models can be
included in one PMA if they are the
same generic device type. Because
shock advisory algorithms and
defibrillation waveforms will likely be
common across various models from a
given manufacturer of devices, FDA
expects the clinical data needed to
support devices within an appropriately
bundled AED PMA to be the same.
However, because of the differences in
device labeling and user requirements
between professional and lay use
devices, FDA continues to believe that
separate PMAs should be submitted for
a manufacturer’s professional use versus
lay use devices. FDA believes this
approach is least burdensome to
manufacturers and is consistent with
the bundling guidance, which states that
‘‘[b]undling is appropriate for devices
that present scientific and regulatory
issues that can most efficiently be
addressed during one review’’ (Ref. 5).
(Comment 11) One comment
requested clarification on whether
E:\FR\FM\29JAR1.SGM
29JAR1
tkelley on DSK3SPTVN1PROD with RULES
4788
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
separate PMAs are required for AEDs
and the associated AED accessories
when a company manufacturers both for
use together. Two comments requested
additional clarification on whether
accessories not specified in the
proposed order (such as
electrocardiograph modules and
electrodes, training pads/batteries,
protective carrying cases, Bluetooth
modules, hardware keys or specialized
pads to reduce energy for pediatric use,
self-testers, SpO2/blood pressure
monitoring devices, cardiopulmonary
resuscitation (CPR) aids, medical device
data systems (MDDS), etc.) would
require PMAs. One comment suggested
that AED accessories that are already
510(k) cleared should not be subject to
premarket approval by virtue of being
used with an AED.
(Response 11) In response to this
comment, FDA has revised the
identification language to clarify that
AED accessories regulated under 21 CFR
870.5310 are ‘‘those accessories
necessary for the AED to detect and
interpret an electrocardiogram and
deliver an electrical shock (e.g., battery,
pad electrode, adapter, and hardware
keys for pediatric use).’’ Manufacturers
of accessory devices that are not
addressed by the final order and are not
already the subject of an existing
classification regulation should contact
FDA.
Under the final order, manufacturers
must submit PMAs for accessories that
are necessary for operation of the AED
system (e.g., accessories necessary to
allow the AED to detect or interpret an
electrocardiogram or deliver a
defibrillation shock). These AED
accessories include batteries, pad
electrodes (including reduced energy
pads for pediatric use), adapters, and
hardware keys for pediatric use. In
response to this comment, FDA has
added ‘‘hardware keys for pediatric use’’
to the identification. Necessary AED
accessories that are for use with a
specific AED should be included in that
PMA for the AED system as a whole.
Alternatively, necessary AED
accessories, including those
manufactured by a third party, may be
submitted in their own PMAs.
Accessories that are not necessary for
the functionality of the AED are not
addressed by the final order. Currently
marketed AED accessories that are not
addressed by the final order, such as
SpO2/blood pressure monitoring
devices, ECG modules and testers, CPR
aids, and MDDS, may be subject to other
regulations and will continue to be
subject to those existing regulations.
Training accessories such as training
pads and batteries for training-only
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
AEDs are not currently subject to any
additional regulations, and will not
become so as a result of the final order.
(Comment 12) One comment
requested clarification regarding AEDs
being considered adulterated 90 days
after the effective date of a final order
in the absence of submission of a
statement of intent to submit a PMA or
the submission of a full PMA. The
comment questioned whether devices
legally distributed prior to the 90th day
could remain in distribution. Another
comment requested clarification on
whether manufacturers could continue
to provide consumable accessory items
(such as batteries and pads) for
previously distributed devices even if a
PMA will not be submitted for that AED
model. Two comments requested
clarification on how and whether
manufacturers would be allowed to
distribute components required for field
servicing of a device, including
refurbished replacement devices, before
PMAs are submitted for the devices.
(Response 12) Under the final order
(see section IV, ‘‘the Final Order’’) and
section 501(f)(2)(B) of the FD&C Act,
PMAs must be submitted within 90 days
after the effective date of the final order
for currently marketed AED systems. If
a PMA is not submitted, the devices are
adulterated. However, for the reasons
discussed above, for currently marketed
AEDs, FDA does not intend to enforce
compliance with the 90-day deadline by
which PMAs must be submitted for 15
months after that deadline (i.e., 18
months after the effective date of the
final order), as long as a notice of intent
to file a PMA is submitted within 90
days of the effective date of the final
order (see section V, ‘‘Implementation
Strategy’’). For currently marketed
necessary AED accessories, FDA also
does not intend to enforce compliance
with the 90-day deadline by which
PMAs must be submitted for 57 months
after that deadline (i.e., 5 years after the
effective date of the final order) (see
section V, ‘‘Implementation Strategy).
This intention applies to necessary AED
accessories regardless of whether a PMA
is being or has been sought for the AED
model.
(Comment 13) One comment
indicated that premarket review of
medical devices such as AEDs should
include review of the software
embedded into the devices, including
review of software verification and
validation documentation. The
comment noted that such review should
also occur for software modifications
and software developed for maintenance
of the devices, including self-test
functions. The comment relayed the
importance of having reviewers with
PO 00000
Frm 00032
Fmt 4700
Sfmt 4700
adequate training, expertise, and
experience.
(Response 13) FDA agrees with the
comment. Review of AEDs under the
510(k) process has included a detailed
review of software documentation
supporting premarket submissions by
appropriately trained and experienced
FDA reviewers. The PMA review will
also involve a review of software
documentation and will be conducted
by trained and experienced FDA
reviewers.
(Comment 14) One comment
suggested an alternative regulatory
approach whereby AEDs for medical
professional use be reclassified into
class II and public access defibrillators
used by laypersons remain in class III
with PMAs required. The comment
stated that professional use devices have
advanced functionality and are operated
by skilled and trained professionals,
which lessens the chance of human
factor errors and increases the
likelihood that the user will be able to
recognize and troubleshoot any
malfunctions. The comment stated that
such users can rely on past experience
and other means of attempting to rescue
a patient, whereas lay users are often
fully reliant on the AED. Two comments
also indicated that professional use
devices are typically manual
defibrillators with additional
functionality, including AED, and that
the proposed order would create an
inconsistent system whereby the same
hardware if used only for manual
defibrillation would be class II, but by
virtue of configuring the device to
include AED functionality would
become a PMA class III product.
(Response 14) FDA disagrees with the
comments and believes that the
submission of PMAs is warranted
regardless of the intended user of the
device. FDA does not believe that there
is sufficient information to establish
special controls to provide a reasonable
assurance of safety and effectiveness of
the device regardless of the training of
AED users.
FDA recognizes that some manual
defibrillators and AEDs share common
hardware and software platforms, and
hence devices with similar or identical
platforms may receive different
regulatory review based on the
configuration. For the reasons
previously stated, however, FDA
continues to believe AED systems
should be class III devices. FDA also
notes that the performance and other
data needed to support safety and
effectiveness for hardware and software
platforms for both types of devices
would be nearly identical; the difference
would be related to the amount of
E:\FR\FM\29JAR1.SGM
29JAR1
tkelley on DSK3SPTVN1PROD with RULES
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
information that must be submitted to
FDA. For a PMA, more information on
the design controls process is required
to be submitted whereas for a 510(k)
submission, some information may not
need to be submitted and instead can
reside within the company’s overall
quality system and associated design
documentation. Such situations of
different regulatory processes have
occurred in other product areas
including contact lenses (daily-wear are
typically class II, whereas extended
wear are class III) and ablation devices
(general surgical use are class II,
whereas use for treatment of atrial
fibrillation is class III), and FDA does
not believe this changes the overall
rationale supporting the need for PMAs.
(Comment 15) Two comments noted
that there are numerous companies that
refurbish and/or resell AEDs. The
comments requested that FDA include
AED resellers and refurbishers in their
consideration of regulatory strategy.
(Response 15) Regardless of the
supplier, the introduction or delivery
for introduction into interstate
commerce of any device that is
adulterated is a prohibited act under
section 301 of the FD&C Act (21 U.S.C.
331) (see Comment 12). FDA encourages
refurbishers and resellers who have
questions about the continued
distribution of AEDs to contact FDA via
the pre-submission process.
(Comment 16) One comment
requested clarification of the process for
modifications of currently marketed
AEDs (and notifying FDA of such
modifications) during the 90-day period
after the final order is issued. One
comment stated that given the nature of
commercial, electrical and mechanical
components used in AEDs, an extended
transition period without the ability to
implement changes would not be
tenable and would result in
unavailability of devices. One comment
requested clarification on 510(k)
submissions accepted for review, but for
which no decision had been rendered,
prior to the effective date of a final order
calling for PMAs.
(Response 16) Under 21 CFR
870.5310, as amended, all new AED and
necessary AED accessories must have an
approved PMA in effect before being
placed in commercial distribution. We
recommend that manufacturers of
currently marketed AEDs contact FDA
regarding implementation of any
changes necessary for their AEDs in
order to address safety concerns or to
support ongoing distribution while
PMA approval is being sought. FDA
understands that issues may arise
relating to part obsolescence or changes
necessary to reduce a risk to health
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
posed by a currently marketed AED that
is not functioning properly.
All other changes need to be
accounted for in a PMA. Moreover, all
new AED and necessary AED
accessories must have an approved
PMA in effect before being placed in
commercial distribution.
(Comment 17) One comment objected
to the comparisons made by FDA at the
2011 Panel meeting between implanted
cardioverter defibrillators (ICDs) (PMA
devices) and AEDs. The comment noted
the number of commercial components
(e.g., components supplied to multiple
industries for a variety of uses) in order
to maintain affordable price-points for
AEDs and the potentially burdensome
PMA supplements that would be
necessary to support incremental
changes in manufacturing for AEDs. The
comment further contended that
purchased component-related recalls for
AEDs have largely been a result of latent
component failures and that FDA’s
examples at the 2011 Panel meeting of
QS concerns related to changes to
purchased components or device design
would not have been averted by the
stricter premarket regulatory oversight
via PMA supplements.
(Response 17) FDA acknowledges that
more stringent regulatory oversight via
the PMA process may not completely
eliminate AED recalls. FDA also
recognizes that AEDs typically contain
commercial components and
manufacturers will need to submit PMA
supplements for changes to these
components, as well as changes to
suppliers and manufacturing processes.
Use of commercial components in PMA
devices is not uncommon and changes
at the component level may have led to
some AED recalls and adverse events,
providing further support for increased
regulatory review. FDA continues to
believe that these failures and the need
for careful consideration and adequate
verification and validation of such
changes support more rigorous review
under the PMA process.
(Comment 18) One comment
requested clarification on activities
during the time after a notice of intent
to file is submitted, including whether
FDA will place additional postmarket
approval requirements on previously
distributed products as allowed under
21 CFR 814.82. The comment further
asked whether IDEs would be required
for design changes (e.g., would an IDE
be required to conduct human factors/
usability studies).
(Response 18) FDA will consider the
need for postapproval requirements in
the context of each manufacturer’s PMA
submission and the devices in
distribution. FDA does not intend to
PO 00000
Frm 00033
Fmt 4700
Sfmt 4700
4789
exempt manufacturers from the IDE
requirements and hence any study
which meets the IDE requirements must
be conducted in accordance with the
requirements of 21 CFR parts 50 and
812. There will be no extended period
for filing an IDE and studies may not be
initiated without appropriate IDE
approvals. Manufacturers who have
questions regarding whether an IDE is
needed for a particular AED study are
encouraged to interact with FDA via the
pre-submission process.
IV. The Final Order
FDA is adopting its findings as
published in the preamble of the
proposed order (78 FR 17890, March 25,
2013) and issuing this final order to
require the filing of a PMA for AED
systems under 515(b) of the FD&C Act
(21 U.S.C. 360e(b)). An AED system
consists of an AED and those
accessories necessary for the AED to
detect and interpret an
electrocardiogram and deliver an
electrical shock (e.g., battery, pad
electrode, adapter, and hardware keys
for pediatric use). An AED system
analyzes the patient’s
electrocardiogram, interprets the cardiac
rhythm, and automatically delivers an
electrical shock (fully automated AED),
or advises the user to deliver the shock
(semi-automated or shock advisory
AED) to treat ventricular fibrillation or
pulseless ventricular tachycardia. Under
section 515(b)(1)(A) of the FD&C Act (21
U.S.C. 360e(b)(1)(A)), PMAs for AED
systems are required to be filed on or
before 90 days after the effective date of
a final order. This final order will revise
21 CFR part 870.
V. Implementation Strategy
Based on comments on the proposed
order, we are clarifying FDA’s
intentions regarding enforcing
compliance with the final order (see
section IV, ‘‘The Final Order’’) and
section 501(f)(2)(B) of the FD&C Act (21
U.S.C. 351(f)(2)(B)).
A. Currently Marketed AEDs
Under the final order and section
501(f)(2)(B) (21 U.S.C. 351(f)(2)(B)),
PMAs for currently marketed AEDs are
required to be filed on or before 90 days
after the effective date of a final order.
However, for currently marketed AEDs,
FDA does not intend to enforce
compliance with this 90-day deadline
for 15 months after that deadline (i.e., 18
months after the effective date of the
final order), as long as notice of intent
to file a PMA is submitted within 90
days of the effective date of the final
order. The notification of the intent to
file a PMA submission must include a
E:\FR\FM\29JAR1.SGM
29JAR1
4790
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
list of all model numbers for which a
manufacturer plans to seek marketing
approval through a PMA.
In conducting any clinical studies,
AEDs may be distributed for
investigational use if the requirements
of the IDE regulations (21 CFR part 812)
are met. There will be no extended
period for filing an IDE nor exemption
from IDE requirements, and studies may
not be initiated without appropriate IDE
approvals, where necessary.
B. Currently Marketed Necessary AED
Accessories
Under the final order and section
501(f)(2)(B) (21 U.S.C. 351(f)(2)(B)),
PMAs for currently marketed necessary
AED accessories are required to be filed
on or before 90 days after the effective
date of this final order. However, for
currently marketed necessary AED
accessories, FDA does not intend to
enforce compliance with this 90-day
deadline for 57 months after the
deadline (i.e., 5 years after the effective
date of the final order). Currently
marketed necessary AED accessory
manufacturers are not required to file an
intent-to-file by the 90-day deadline.
After the effective date of the final
order, new AEDs and necessary AED
accessories must have approved PMAs
to be legally marketed. The following
tables show the regulatory timetable for
currently marketed AEDs and necessary
AED accessories.
TABLE 1—CURRENTLY MARKETED AEDS
Timetable for which FDA does not intend to
enforce compliance
(time after effective date of order)
Distribution period
(time after effective date of order)
Intent to File a PMA ......
90 days ............................................................
File a PMA ....................
18 months ........................................................
Devices included in an intent to file: 18 months.
Devices not included in intent to file: 90 days.
Until a not approvable decision or denial decision is issued; can continue distribution if an approval order is issued.
TABLE 2—CURRENTLY MARKETED NECESSARY AED ACCESSORIES
Timetable for which FDA does not intend to
enforce compliance
(time after effective date of order)
Intent to File a PMA ......
File a PMA ....................
Distribution period
(time after effective date of order)
N/A ...................................................................
60 months ........................................................
N/A.
Until a not approvable decision or denial decision is issued; can continue distribution if an approval order is issued.
VI. Environmental Impact
The Agency has determined under 21
CFR 25.30 (h) that this action is of a
type that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
tkelley on DSK3SPTVN1PROD with RULES
VII. Paperwork Reduction Act of 1995
The final order refers to previously
approved collections of information
found in FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
21 CFR part 812 have been approved
under OMB control number 0910–0078;
the collections of information in 21 CFR
part 807, subpart E, have been approved
under OMB control number 0910–0120;
the collections of information in 21 CFR
part 814, subpart B, have been approved
under OMB control number 0910–0231;
and the collections of information under
21 CFR part 801 have been approved
under OMB control number 0910–0485.
VIII. Codification of Orders
Prior to the amendments by FDASIA,
section 515(b) of the FD&C Act (21
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
U.S.C. 360e(b)) provided for FDA to
issue regulations to require PMA
approval for preamendments devices or
devices found substantially equivalent
to preamendments devices. Section
515(b) of the FD&C Act (21 U.S.C.
360e(b)), as amended by FDASIA,
provides for FDA to require PMA
approval for such devices by issuing a
final order, following the issuance of a
proposed order in the Federal Register.
FDA will continue to codify the
requirement for a PMA approval in the
Code of Federal Regulations. Therefore,
under section 515(b)(1)(A) of the FD&C
Act (21 U.S.C. 360e(b)(1)(A)), as
amended by FDASIA, in this final order,
we are requiring PMA approval for AED
systems and we are making the language
in 21 CFR 870.5310 consistent with the
final version of this order.
IX. References
The following references have been
placed on display in the Division of
Dockets Management (see ADDRESSES)
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday. (FDA has verified the
Web site address, but FDA is not
responsible for any subsequent changes
to the Web site after this document
publishes in the Federal Register.)
PO 00000
Frm 00034
Fmt 4700
Sfmt 4700
1. Meeting Materials for 515(i) Regulatory
Classification of Automated External
Defibrillator Systems, January 25, 2011,
available at https://www.fda.gov/
AdvisoryCommittees/
CommitteesMeetingMaterials/
MedicalDevices/
MedicalDevicesAdvisoryCommittee/
CirculatorySystemDevicesPanel/
ucm240575.htm.
2. U.S. Food and Drug Administration,
Medical Device Recalls Database, available at
https://www.accessdata.fda.gov/scripts/cdrh/
cfdocs/cfRES/res.cfm.
3. Requests for Feedback on Medical
Device Submissions: The Pre-Submission
Program and Meetings with Food and Drug
Administration Staff, Guidance for Industry
and Food and Drug Administration Staff,
February 18, 2014, available at https://
www.fda.gov/downloads/MedicalDevices/
DeviceRegulationandGuidance/
GuidanceDocuments/UCM311176.pdf.
4. Acceptance and Filing Reviews for
Premarket Approval Applications (PMAs),
Guidance for Industry and Food and Drug
Administration Staff, December 31, 2012,
available at https://www.fda.gov/downloads/
MedicalDevices/
DeviceRegulationandGuidance/
GuidanceDocuments/UCM313368.pdf.
5. Guidance for Industry and FDA Staff:
Bundling Multiple Devices or Multiple
Indications in a Single Submission, June 22,
2007, available at https://www.fda.gov/
medicaldevices/
deviceregulationandguidance/
guidancedocuments/ucm089731.htm.
E:\FR\FM\29JAR1.SGM
29JAR1
Federal Register / Vol. 80, No. 19 / Thursday, January 29, 2015 / Rules and Regulations
Dated: January 23, 2015.
Leslie Kux,
Associate Commissioner for Policy.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 870 is
amended as follows:
[FR Doc. 2015–01619 Filed 1–28–15; 8:45 am]
PART 870—CARDIOVASCULAR
DEVICES
Internal Revenue Service
BILLING CODE 4164–01–P
DEPARTMENT OF THE TREASURY
26 CFR Part 1
1. The authority citation for 21 CFR
part 870 continues to read as follows:
■
[TD 9709]
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Section 870.5310 is amended by
revising the section heading and
paragraphs (a) and (c) to read as follows:
■
tkelley on DSK3SPTVN1PROD with RULES
§ 870.5310
system.
Automated external defibrillator
(a) Identification. An automated
external defibrillator (AED) system
consists of an AED and those
accessories necessary for the AED to
detect and interpret an
electrocardiogram and deliver an
electrical shock (e.g., battery, pad
electrode, adapter, and hardware key for
pediatric use). An AED system analyzes
the patient’s electrocardiogram,
interprets the cardiac rhythm, and
automatically delivers an electrical
shock (fully automated AED), or advises
the user to deliver the shock (semiautomated or shock advisory AED) to
treat ventricular fibrillation or pulseless
ventricular tachycardia.
*
*
*
*
*
(c) Date PMA or notice of completion
of PDP is required. A PMA will be
required to be submitted to the Food
and Drug Administration by April 29,
2015, for any AED that was in
commercial distribution before May 28,
1976, or that has, by April 29, 2015,
been found to be substantially
equivalent to any AED that was in
commercial distribution before May 28,
1976. A PMA will be required to be
submitted to the Food and Drug
Administration by April 29, 2015, for
any AED accessory described in
paragraph (a) that was in commercial
distribution before May 28, 1976, or that
has, by April 29, 2015, been found to be
substantially equivalent to any AED
accessory described in paragraph (a)
that was in commercial distribution
before May 28, 1976. Any other AED
and AED accessory described in
paragraph (a), shall have an approved
PMA or declared completed PDP in
effect before being placed in commercial
distribution.
VerDate Sep<11>2014
16:36 Jan 28, 2015
Jkt 235001
RIN 1545–BK64
Application for Recognition as a
501(c)(29) Organization
Internal Revenue Service (IRS),
Treasury.
ACTION: Final regulations and removal of
temporary regulations.
AGENCY:
This document contains final
regulations authorizing the IRS to
prescribe the procedures by which
certain entities may apply to the IRS for
recognition of exemption from Federal
income tax. These regulations affect
qualified nonprofit health insurance
issuers participating in the Consumer
Operated and Oriented Plan program
established by the Centers for Medicare
and Medicaid Services that seek
exemption from federal income tax
under the Internal Revenue Code.
DATES: Effective date: These regulations
are effective on January 29, 2015.
Applicability date: For date of
applicability, see § 1.501(c)(29)–1(c).
FOR FURTHER INFORMATION CONTACT:
¨
Martin Schaffer, (202) 317–5800 (not a
toll-free number).
SUPPLEMENTARY INFORMATION:
SUMMARY:
Background
Section 501(c)(29) of the Internal
Revenue Code (Code) provides
requirements for tax exemption under
section 501(a) for qualified nonprofit
health insurance issuers (QNHIIs).
Section 501(c)(29) was added to the
Code by section 1322(h)(1) of the Patient
Protection and Affordable Care Act,
Public Law 111–148 (March 23, 2010)
(Affordable Care Act).
Section 1322 of the Affordable Care
Act directs the Centers for Medicare and
Medicaid Services (CMS) to establish
the Consumer Operated and Oriented
Plan (CO–OP) program. The purpose of
the CO–OP program is to foster the
creation of member-governed QNHIIs
that will operate with a strong consumer
focus and offer qualified health plans in
the individual and small group markets.
CMS provides loans and repayable
PO 00000
Frm 00035
Fmt 4700
Sfmt 4700
4791
grants (collectively, loans) to
organizations applying to become
QNHIIs to help cover start-up costs and
meet any solvency requirements in
States in which the organization is
licensed to issue qualified health plans.
For each loan, CMS issues a Notice of
Award and Loan Agreement to the
QNHII. The appropriate officer of the
QNHII or of the QNHII’s board of
directors must sign and return the loan
agreement to CMS. On December 13,
2011, CMS issued final regulations
implementing the CO–OP program at 76
FR 77392.
The CMS final regulations define a
QNHII as an entity that, within specified
time frames, satisfies or can reasonably
be expected to satisfy the standards in
section 1322(c) of the Affordable Care
Act and in the CMS final regulations.
The entity will constitute a QNHII until
such time as CMS determines the entity
does not satisfy or cannot reasonably be
expected to satisfy these standards.
Section 1322(c) of the Affordable Care
Act imposes a number of requirements,
including that a QNHII be organized as
a nonprofit member corporation under
State law and that substantially all its
activities consist of the issuance of
qualified health plans in the individual
and small group markets in each State
in which it is licensed to issue such
plans.
Section 501(c)(29)(A) of the Code
provides that a QNHII (within the
meaning of section 1322(c) of the
Affordable Care Act) which has received
a loan or grant under the CO–OP
program may be recognized as exempt
from taxation under section 501(a), but
only for periods for which the
organization is in compliance with the
requirements of section 1322 of the
Affordable Care Act and any loan or
grant agreement with the Secretary of
Health and Human Services. Section
501(c)(29)(B) provides that a QNHII will
not qualify for tax-exemption unless it
meets four additional requirements.
First, the QNHII must give notice to the
Secretary of the Treasury, in such
manner as the Secretary may by
regulations prescribe, that it is applying
for recognition of exemption as an
organization described in section
501(c)(29). Second, no part of the
QNHII’s net earnings may inure to the
benefit of any private shareholder or
individual, except to the extent
permitted by section 1322(c)(4) of the
Affordable Care Act (which requires that
any profits be used to lower premiums,
to improve benefits, or for other
programs intended to improve the
quality of health care delivered to the
organization’s members). Third, no
substantial part of the QNHII’s activities
E:\FR\FM\29JAR1.SGM
29JAR1
Agencies
[Federal Register Volume 80, Number 19 (Thursday, January 29, 2015)]
[Rules and Regulations]
[Pages 4783-4791]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2015-01619]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA-2013-N-0234]
Effective Date of Requirement for Premarket Approval for
Automated External Defibrillator Systems
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or the Agency) is
issuing a final order to require the filing of premarket approval
applications (PMA) for automated external defibrillator (AED) systems,
which consist of an AED and those AED accessories necessary for the AED
to detect and interpret an electrocardiogram and deliver an electrical
shock (e.g., pad electrodes, batteries, adapters, and hardware keys for
pediatric use).
DATES: This order is effective on January 29, 2015.
FOR FURTHER INFORMATION CONTACT: Linda Ricci, Center for Devices and
Radiological Health, 10903 New Hampshire Ave., Bldg. 66, Rm. 1314,
Silver Spring, MD 20993, 301-796-6325, linda.ricci@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act), as amended
by the Medical Device Amendments of 1976 (the 1976 amendments) (Pub. L.
94-295), the Safe Medical Devices Act of 1990 (Pub. L. 101-629), the
Food and Drug Administration Modernization Act of 1997 (Pub. L. 105-
115), the Medical Device User Fee and Modernization Act of 2002 (Pub.
L. 107-250), the Medical Devices Technical Corrections Act (Pub. L.
108-214), the Food and Drug Administration Amendments Act of 2007 (Pub.
L. 110-85), and the Food and Drug Administration Safety and Innovation
Act (FDASIA) (Pub. L. 112-144), among other amendments, established a
comprehensive system for the regulation of medical devices intended for
human use. Section 513 of the FD&C Act (21 U.S.C. 360c) established
three categories (classes) of devices, reflecting the regulatory
controls needed to provide reasonable assurance of their safety and
effectiveness. The three categories of devices are class I (general
controls), class II (special controls), and class III (premarket
approval).
Under section 513(d) of the FD&C Act (21 U.S.C. 360c(d)), devices
that were in commercial distribution before the enactment of the 1976
amendments, May 28, 1976 (generally referred to as ``preamendments
devices''), are classified after FDA has: (1) Received a recommendation
from a device classification panel (an FDA advisory committee); (2)
published the panel's recommendation for comment, along with a proposed
regulation classifying the device; and (3) published a final regulation
classifying the device. FDA has classified most preamendments devices
under these procedures.
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as ``postamendments devices'') are
automatically classified by section 513(f) of the FD&C Act (21 U.S.C.
360c(f)) into class III without any FDA rulemaking process. Those
devices remain in class III and require premarket approval unless, and
until, the device is reclassified into class I or II or FDA issues an
order finding the device to be substantially equivalent, in accordance
with section 513(i) of the FD&C Act (21 U.S.C. 360c(i)), to a predicate
device that does not require premarket approval. The Agency determines
whether new devices are substantially equivalent to predicate devices
by means of premarket notification procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
A preamendments device that has been classified into class III and
devices found substantially equivalent by means of premarket
notification (510(k)) procedures to such a preamendments device or to a
device within that type (both the preamendments and substantially
equivalent devices are referred to as preamendments class III devices)
may be marketed without submission of a premarket approval application
(PMA) until FDA issues a final order under section 515(b) of the FD&C
Act (21 U.S.C. 360e(b)) requiring premarket approval or until the
device is subsequently reclassified into class I or class II. Section
515(b)(1) of the FD&C Act (21 U.S.C. 360e(b)(1)) directs FDA to issue
an order requiring premarket approval for a preamendments class III
device.
Although, under the FD&C Act, the manufacturer of a class III
preamendments device may respond to the call for PMAs by filing a PMA
or a notice of completion of a product development protocol (PDP), in
practice, the option of filing a notice of completion of a PDP has not
been used. For simplicity, although corresponding requirements for PDPs
remain available to manufacturers in response to a final order under
section 515(b) of the FD&C Act (21 U.S.C. 360e(b)), this document will
refer only to the requirement for the filing and receiving approval of
a PMA.
On July 9, 2012, FDASIA was enacted. Section 608(a) of FDASIA (126
Stat. 1056) amended section 513(e) of the FD&C Act (21 U.S.C. 360c(e)),
changing the mechanism for reclassifying a device from rulemaking to an
administrative order. Section 608(b) of FDASIA amended section 515(b)
of the FD&C Act (21 U.S.C. 360e(b)) changing the mechanism for
requiring premarket approval for a preamendments class III device from
rulemaking to an administrative order.
Section 515(b)(1) of the FD&C Act sets forth the process for
issuing a final order. Specifically, prior to the issuance of a final
order requiring premarket approval for a preamendments class III
device, the following must occur: (1) Publication of a proposed order
in the Federal Register; (2) a meeting of a device classification panel
described in section 513(b) of the FD&C Act; and (3) consideration of
comments from all affected stakeholders, including patients, payers,
and providers.
Section 515(b)(3) of the FD&C Act provides that FDA shall, after
the close of the comment period on the proposed order, consideration of
any comments received, and a meeting of a device classification panel
described in section 513(b) of the FD&C Act, issue a final order to
require premarket approval or publish a document terminating the
proceeding together with the reasons for such termination.
A preamendments class III device may be commercially distributed
without a PMA until 90 days after FDA issues a final order (a final
rule issued under section 515(b) of the FD&C Act prior to the enactment
of FDASIA is considered to be a final order for purposes of section
501(f) of the FD&C Act (21 U.S.C. 351(f))) requiring premarket approval
for the device, or 30 months after final classification of the device
under section 513 of the FD&C Act, whichever is later. For AED systems,
the later of these two time periods is the 90-day period. Therefore,
section 501(f)(2)(B) of the FD&C Act (21 U.S.C. 351(f)(2)(B)) requires
that a PMA
[[Page 4784]]
for such devices be filed within 90 days of the effective date of a
final order. However, for the reasons discussed below, FDA does not
intend to enforce compliance with the 90-day deadline for PMA
submissions for currently marketed AEDs and those AED accessories
identified in 21 CFR 870.5310(a) (see further discussion in section V,
``Implementation Strategy'').
Also, a preamendments device subject to the order process under
section 515(b) of the FD&C Act (21 U.S.C. 360e) is not required to have
an approved investigational device exemption (IDE) (see part 812 (21
CFR part 812)) contemporaneous with its interstate distribution until
the date identified by FDA in the final order requiring the filing of a
PMA for the device. At that time, an IDE is required only if a PMA has
not been filed. If the manufacturer, importer, or other sponsor of the
device submits an IDE application and FDA approves it, the device may
be distributed for investigational use. If a PMA is not filed by the
later of the two dates, and the device is not distributed for
investigational use under an IDE, the device is deemed to be
adulterated within the meaning of section 501(f)(1)(A) of the FD&C Act
(21 U.S.C. 351(f)(1)(A)), and subject to seizure and condemnation under
section 304 of the FD&C Act (21 U.S.C. 334) if its distribution
continues. Other enforcement actions include, but are not limited to,
the following: Shipment of devices in interstate commerce may be
subject to injunction under section 302 of the FD&C Act (21 U.S.C.
332), and the individuals responsible for such shipment may be subject
to prosecution under section 303 of the FD&C Act (21 U.S.C. 333). FDA
requests that manufacturers take action to prevent the further use of
devices for which no PMA has been filed.
II. Regulatory History of This Device
On January 25, 2011, the Circulatory System Devices Panel
(``Panel'') recommended that AED systems be classified as class III
devices and subject to premarket approval to provide reasonable
assurance of the safety and effectiveness of the device (Ref. 1). The
Panel recommended that AED systems be regulated as class III devices
because, among other things, they are lifesaving devices. Furthermore,
the problems identified in adverse events in the medical device
reporting systems and recalls related to AED systems indicated these
devices require more regulatory oversight.
FDA published a proposed order to require PMAs for AED systems in
the Federal Register of March 25, 2013 (78 FR 17890). FDA is now
requiring PMAs for AED systems, which include AED accessories necessary
for the functionality of the AED (e.g., pad electrodes, batteries,
adapters, and hardware keys for pediatric use) (``necessary AED
accessories'') (see section IV, ``The Final Order'').
FDA received and has considered comments on the AED systems
proposed order as discussed in section III of this document.
III. Public Comments in Response to the Proposed Order
In response to the March 25, 2013 (78 FR 17890) proposed order to
maintain the class III classification and require premarket approval
for AED systems, FDA received 66 comments and one petition for
reclassification (see FDA-2013-N-0234-0002).\1\ The comments and the
FDA's responses to the comments are summarized below. Certain comments
are grouped together under a single number because the subject matter
of the comments is similar. The number assigned to each comment is
purely for organizational purposes and does not signify the comment's
value or importance or the order in which it was submitted.
---------------------------------------------------------------------------
\1\ FDA will respond separately to the reclassification petition
and will address the issues raised in that petition in its response.
The reclassification petition is available at https://www.regulations.gov/#!documentDetail;D=FDA-2013-N-0234-0002.
---------------------------------------------------------------------------
(Comment 1) Many comments indicated that AED systems have already
been demonstrated to be safe and effective, and referenced literature
and studies supporting the reliability of these devices and the value
of AED systems in treating sudden cardiac arrest (SCA). The comments
stated that PMAs and associated increased regulatory cost and review
time is not warranted and would hinder innovation, increase device cost
to consumers, and reduce availability of AED systems. The comments
further stated that it is widely recognized that improvement in the
survival rate from SCA is due in large part to widespread distribution
of AED systems and expressed concern that requiring PMAs would limit
availability of the devices.
(Response 1) FDA agrees that many currently marketed AEDs have been
demonstrated to be effective in clinical use and, when designed and
manufactured appropriately, AEDs can be safe and effective. However,
FDA believes that there is insufficient information to determine that
general and special controls would provide a reasonable assurance of
the safety and effectiveness of these devices, which are for a use in
supporting or sustaining human life (see section 513(a)(1)(C) of the
FD&C Act (21 U.S.C. 360c(a)(1)(C))). Specifically, the postmarket
information on AEDs supports increased regulatory review to ensure that
device design and manufacturing practices provide a reasonable
assurance of safety and effectiveness. FDA acknowledges that the PMA
process may result in increased regulatory cost to manufacturers;
however, FDA believes that device quality will improve, which will
reduce costs associated with postmarket actions including recalls.
FDA also agrees that continued efforts to make safe and effective
AED systems available is in the interest of public health, but
disagrees that this call for PMAs will limit device availability. FDA
believes that many manufacturers of currently marketed AEDs already
have, or can reasonably obtain, the necessary data to support a PMA,
and hence expects AED distribution to continue to meet demand. Also,
for the reasons discussed below, FDA does not intend to enforce
compliance with the 90-day deadline for submission of PMAs for
currently marketed AEDs and necessary AED accessories (for further
discussion see section V, ``Implementation Strategy'').
At the January 2011 Panel meeting, the Panel discussed the impact
of FDA regulatory scrutiny on innovation. Various Panel members agreed
that the appropriate focus should be on assuring reliability of AEDs
and that there was no evidence presented to indicate that a call for
PMAs would unduly hinder device innovation (Ref. 1). FDA notes that
previous significant innovations for AED systems (e.g., new
defibrillation waveforms) have been supported by clinical evidence in
the 510(k) process and that under the PMA process this clinical
evidence is not expected to significantly change. As was mentioned in
the proposed order, FDA anticipates that many AED manufacturers already
have sufficient clinical evidence to support a PMA.
(Comment 2) Several comments noted that AED system failures are
often the result of use error or improper maintenance (e.g., expired
batteries/pads, periodic checks not performed, etc.) and not of system
failure or malfunction. The comments stated that efforts should be
devoted to ensuring appropriate public awareness, training
(particularly for lay users), and maintenance to address these issues
as opposed to increasing premarket regulatory review. One comment
stated that the proposed order should not be finalized until all
stakeholders, not only device manufacturers, are engaged in an
[[Page 4785]]
integrated approach to increase the likelihood that AED systems will be
available and functional when needed.
(Response 2) FDA agrees that AED system training and maintenance
are important to help ensure AED system availability and proper use and
also believes manufacturers and users are in the best position to
develop and implement training and maintenance materials. FDA supports
ongoing discussions and efforts to improve training and maintenance,
but disagrees that these activities should delay finalizing the
requirement for PMAs for these devices. Although we recognize that
there have been some medical device reports (MDRs) associated with use
errors, the focus of FDA's review of MDRs and recalls of AED systems
has been related to problems with the quality of these devices as
related to device design and manufacture and FDA continues to believe
that requiring PMAs is appropriate.
(Comment 3) Several comments stated that special controls,
including performance testing to industry standards, device labeling,
guidance documents, human factors analysis and design, summary of field
actions and mitigations to address Quality System (QS) concerns, risk
management, and post-market surveillance were sufficient to regulate
AED systems as class II devices under the existing 510(k) regulatory
regime. One comment indicated that several of the regulatory controls
identified by FDA as consistent with PMA requirements--such as pre-
market inspections, review of changes that could significantly affect
the safety or effectiveness of the device, and postmarket
surveillance--could also be conducted under the 510(k) regime. Other
comments supported FDA's proposal to maintain the devices in class III
and agreed that the manufacturing controls, premarket review
requirements, and assessment of lay use are best managed under the PMA
process.
(Response 3) FDA disagrees that there is sufficient information to
determine that general and special controls would provide a reasonable
assurance of safety and effectiveness of these devices given safety
concerns related to the manufacturing processes and design changes,
problems which FDA considered in determining that PMAs are warranted
(see section 513(a)(1)(C) of the FD&C Act (21 U.S.C. 360c(a)(1)(C))).
FDA does not generally conduct preclearance inspections under the
510(k) process because such information is not required in a 510(k)
submission under the FD&C Act or FDA regulations. Further, under
section 513(f)(5) of the FD&C Act (21 U.S.C. 360c(f)(5)), FDA may not
withhold a 510(k) ``because of a failure to comply with any provision
of this Act unrelated to a substantial equivalence decision, including
a finding that the facility in which the device is manufactured is not
in compliance with good manufacturing requirements as set forth in
regulations of the Secretary under section 520(f) (other than a finding
that there is a substantial likelihood that the failure to comply with
such regulations will potentially present a serious risk to human
health).'' In contrast, under section 515(c)(1)(C) of the FD&C Act (21
U.S.C. 360e(c)(1)(C)), a PMA must include ``a full description of the
methods used in, and the facilities and controls used for, the
manufacturing, processing, and when relevant, packing and installation
of, such device.'' Moreover, many of the design and manufacturing
changes that have led to AED system recalls were not required to be
reported to FDA under the 510(k) process. If these changes had been
reported prior to implementation, as would be required in the PMA
regime, these recalls may have been avoided. FDA continues to believe
that the necessary regulatory controls for AED systems are consistent
with the PMA review process, and that the 510(k) process does not
provide sufficient regulatory oversight for these devices.
Similarly, FDA's oversight of postmarket changes to devices is very
different in the 510(k) context as compared to the PMA context. Under
21 CFR 807.81, FDA requires 510(k)s for a change to a device only when
the change ``could significantly affect the safety or effectiveness of
the device, e.g., a significant change or modification in design,
material, chemical composition, energy source, or manufacturing
process. '' In contrast, under 21 CFR 814.39, FDA requires PMA
supplements (including 30-day notices) for any change to a PMA-approved
device that affects safety or effectiveness. These differences in
authorities, among the other reasons discussed above, warrant
regulation of AEDs in class III.
(Comment 4) A few comments indicated that existing AED and AED
accessory manufacturers are already subject to the QS regulation (21
CFR 820) and manufacturing quality would not be measurably improved as
a result of requiring PMAs. One comment noted that specific
expectations under the QS regulation for design controls, purchasing
controls, and other issues identified by FDA as problematic for AEDs
could be addressed by special controls and other regulations, and AEDs
could remain in class II. One comment further stated that such concerns
could be managed via postmarket controls, which are available under the
510(k) regime, such as submission of a summary of recent field actions
and related design mitigations.
(Response 4) FDA disagrees with the comments. FDA acknowledges that
AED and AED accessory manufacturers are already subject to the QS
regulation and that QS requirements result in rigorously designed and
manufactured devices and resultant quality improvements. By requiring
premarket review of QS processes as well as device changes for AEDs,
FDA believes the PMA process will provide a reasonable assurance of
safety and effectiveness (see Response 3 above).
(Comment 5) One comment stated that certain AED accessories,
specifically electrodes, cables, and adapters, are well-understood
devices and that their classification into class III is not warranted.
The comment stated that these accessories could be adequately regulated
in class II with special controls, as is already the case when these
accessories are used with manual defibrillators. The comment
recommended special controls, including the following: Performance
testing, usability evaluation, labeling, biocompatibility, and
readiness for use. Two comments stated that because AED accessories
often have identical designs and the same intended use as accessories
used with class II manual defibrillators, FDA should not perform
duplicative reviews under both the 510(k) and PMA regimes and that PMA
review should be required only when use of the accessory with an AED
results in a change in intended use or design.
(Response 5) Accessories necessary for an AED to detect and
interpret an electrocardiogram and deliver an electrical shock (e.g.,
battery, pad electrode, adapter, and hardware keys for pediatric use)
are necessary for AED system functionality. Failure of these necessary
accessories leads to the same negative outcomes as a failure of the AED
itself; e.g., an AED not ready for use because of a faulty battery is
unable to detect heart rhythm abnormalities and/or deliver a
defibrillation shock to a victim of SCA. FDA's review of adverse events
and recalls has shown that problems with AED accessories have occurred
during clinical use. As such, FDA continues to believe that the same
regulatory oversight is warranted for certain critical accessories
(i.e., batteries, pad electrodes, adapters, and hardware keys for
pediatric use) as for the AEDs with which they are used. As discussed
in the response to Comment 3 above, FDA does not believe that
[[Page 4786]]
adequate regulatory controls are available under the 510(k) process,
and hence PMAs are necessary to provide a reasonable assurance of
safety and effectiveness.
(Comment 6) Several comments questioned the validity of FDA's data
regarding adverse events associated with AED failures. One comment
noted that FDA provided no data in the proposed order on the frequency
of adverse events or relationship of number of events to total
distribution and use of AEDs. The comment requested additional
information from FDA to support the validity of the MDR analysis
presented at the 2011 Panel and relied upon to support the proposed
order. A few comments presented alternate analyses of MDR data that
suggested that MDRs for AEDs are not increasing. One comment presented
an analysis that showed no statistically significant increase in the
rate of adverse event reports over the time period of 2007 to 2011. Two
comments stated that a majority of AED MDRs reported to FDA resulted
from self-test errors--which are reported as malfunction MDRs because
they could cause or contribute to a death or serious injury but do not
represent device failures in clinical use. The comments contended that
any analysis of MDRs should focus instead on actual use adverse events,
which would represent a small subset of the overall MDRs. One comment
stated that self-test related events are representative of an effective
design risk mitigation strategy being employed for AEDs and that
because AEDs are often in standby for a large percentage of time, self-
test detection of problems before use should not be included in the
overall assessment of the benefit-risk profile for AEDs. Two comments
requested further guidance from FDA on MDR reporting expectations for
AEDs.
(Response 6) Although FDA requires manufacturers to submit an MDR
when their device may have caused or contributed to a death, serious
injury, or in certain situations when their device has malfunctioned,
FDA acknowledges that there are limitations on the review of MDR data,
including the fact that FDA typically does not have complete
information on the number of devices in distribution from which to
calculate adverse events rates. These limitations were discussed at the
2011 Panel meeting. FDA has previously stated that fatality statistics
and injury statistics from MDRs should be considered in light of
underreporting (58 FR 61952 at 61972, November 23, 1993). In addition,
FDA notes that the evaluation of MDR data for AEDs was focused on
manufacturing and design concerns and was not aimed at developing
specific failure rates. Moreover, FDA believes that the large number of
devices in distribution and the life-saving nature of these devices
combined with the steady rate of MDRs support a call for PMAs to help
ensure that these devices are adequately designed and manufactured so
that they are available when needed.
FDA disagrees that evaluation of adverse events should focus only
on those events that occur during clinical use. Although some
distributed AEDs may seldom be used, this does not reduce the
importance that they are safe and effective when needed. FDA
acknowledges the importance of AED self-test features and recognizes
that many self-test failures are not indicative of issues with overall
device quality. FDA believes, however, that some self-test failures
signal significant quality problems arising from device design or
manufacturing issues and are appropriately considered as adverse events
if recurrence of such failures could, for example, render the device
unavailable for use when needed. FDA also recognizes that some MDRs may
eventually be found to be the result of problems not associated with
the device; however, this concern is applicable to all devices subject
to adverse event reporting requirements and FDA does not believe such
reports unduly influence overall reporting numbers.
FDA also notes that our review of available information, as
presented at the January 2011 Panel meeting, included data on voluntary
corrections and removals (i.e., ``recalls'') of AEDs pursuant to
section 519(g) of the FD&C Act (21 U.S.C. 360i(g)). Recalls are
conducted ``(A) to reduce a risk to health posed by the device, or (B)
to remedy a violation of this Act caused by the device which may
present a risk to health,'' and as such may reflect safety concerns for
AEDs (section 519(g)(1) of FD&C Act (21 U.S.C. 360i(g)(1)). Since the
January 2011 Panel meeting, over 40 additional class I or class II
recalls have been conducted by AED manufacturers and have impacted over
2 million distributed AEDs (Ref. 2). The root cause of these recalls
has been attributed to a variety of causes, with design controls,
purchasing controls, and receiving acceptance activities being the most
common. FDA continues to believe that the recall data reinforces the
overall conclusion regarding the inadequacy of regulatory controls for
AED systems under the 510(k) process.
Additional guidance on MDR requirements for AEDs is beyond the
scope of this document; however, FDA intends to continue efforts to
clarify medical device reporting expectations and manufacturers who
have questions regarding their reporting obligations should contact
FDA.
(Comment 7) Several comments responded to FDA's request for
feedback regarding whether 15 months is sufficient to allow companies
to collect information necessary to support submission of a PMA. Two
comments stated that this issue was dependent on the data expected by
FDA and that FDA should provide more guidance in this respect. One
comment requested clarification on what clinical data is known to FDA
that would support a PMA because it is critical that AED manufacturers
understand the type and amount of data that will be required. One
comment stated that it is unclear what FDA's expectations would be for
clinical trials of new AEDs or the need for clinical trials for AED
accessories given available less burdensome methods for obtaining
performance data on accessories. Another comment requested
clarification on whether AED manufacturers would be expected to re-test
and re-validate older AED models to currently recognized standards. One
comment requested clarification on when marketing materials for AEDs
would need to comply with 21 CFR 801.109.
One comment suggested that the 15-month period should be extended
to 30 months, which the commenter claimed would be consistent with
section 501(f)(2)(B) of the FD&C Act (21 U.S.C. 351(f)(2)(B). One
commenter requested clarification regarding whether the 15 months
started at the 90th day after a final order was issued and another
comment indicated that 15 months should be sufficient, but that the 15
months should not include FDA's 180-day PMA review time. One comment
suggested that FDA require PMAs 90 days after the final order.
(Response 7) The data required to support premarket approval will
vary by device and the specific data requirements. FDA is aware of
clinical study information that can be leveraged for AEDs from both
published studies and clinical data previously submitted to FDA under
the 510(k) process, and, as was stated in the proposed order, FDA
believes that many AED accessories ``may need to submit non-clinical
performance testing with confirmatory animal studies in order to
support independent PMA approval'' (78 FR 17894, March 25, 2013).
Performance testing of AEDs must be provided in a PMA to support a
reasonable assurance of safety and effectiveness. Although retesting
older
[[Page 4787]]
AED models to currently recognized standards is one way to meet the
performance testing requirements, compliance with such standards is
voluntary and manufacturers may submit a justification for how other
testing conducted on their devices provides equivalent assurances of
safety and effectiveness. FDA encourages manufacturers to proactively
engage FDA via the pre-submission process to discuss the specific data
needed for their PMAs (Ref. 3). FDA notes that existing prescription
AEDs are already subject to 21 CFR 801.109, and will remain so after
this call for PMAs. FDA review of AED PMAs will include review of the
associated AED labeling to ensure such device labeling complies with
regulatory requirements.
FDA notes that the 30 months discussed in section 501(f)(2)(B) of
the FD&C Act (21 U.S.C. 351(f)(2)(B)) references the date from initial
classification of a device into class III. AEDs have been classified as
class III for more than 30 months, and hence this statutory provision
has expired. FDA also acknowledges that it is in the interest of public
health to ensure the availability of AEDs because they are life-saving
devices and their clinical use is well-established. After consideration
of the comments, FDA continues to believe that the proposed 15 months
for filing a PMA (Ref. 4) strikes an appropriate balance between the
need to ensure continued availability of AEDs for the public health
reasons stated above and the implementation of PMA requirements to
ensure the safety and effectiveness of AEDs.
For currently marketed AEDs, FDA does not intend to enforce
compliance with the 90-day deadline by which PMAs must be submitted for
15 months after that deadline (i.e., 18 months after the effective date
of the final order), as long as a notice of intent to file a PMA is
submitted within 90 days of the effective date of the final order (see
section V, ``Implementation Strategy''). Even if a notice of intent and
PMA are submitted by these dates, manufacturers must cease distribution
of devices upon receiving a not approvable or denial decision rendered
on a PMA. To resume distribution, these manufacturers must receive PMA
approval for their devices.
Moreover, for currently marketed necessary AED accessories, FDA
does not intend to enforce compliance with the 90-day deadline by which
PMAs must be submitted for 57 months after that deadline (i.e., 5 years
after the effective date of the final order) (see section V,
``Implementation Strategy''). Continued availability of necessary AED
accessories, including consumable accessory items (e.g., pad
electrodes) and accessories with limited useful life (e.g., batteries),
is critical to ensuring the safety and efficacy of currently marketed
AEDs during the time while PMAs for those AEDs are being pursued. In
addition, the continued availability of necessary accessories for
``legacy devices''--individual AEDs that have been distributed and are
currently in use (e.g., in public facilities, etc.) and for which the
manufacturer is not seeking PMA approval for that AED model--ensures
the availability of functional legacy AEDs until they are replaced with
PMA-approved AEDs.
(Comment 8) One commenter stated that FDA did not have a legal
basis for continuing with finalization of a call for PMAs for AED
systems because FDA failed to convene a panel as is required under
FDASIA prior to issuing a final order. The commenter stated that FDA
may not rely on the 2011 pre-FDASIA Panel because that Panel meeting
was related to reclassifications under section 515(i) of the FD&C Act
(21 U.S.C. 360e(i)) and not related to calls for PMAs under section
515(b) of the FD&C Act (21 U.S.C. 360e(b)). The commenter further
contended that the 2011 Panel neither considered new information
contained in a reclassification petition submitted to FDA in 2009 nor
adequately discussed the appropriateness of class II special controls.
(Response 8) FDA disagrees with the comment that FDA does not have
a legal basis to finalize an order calling for PMAs for AED systems.
Pursuant to FDASIA, the amendments to section 515(b) of the FD&C Act
require, in relevant part, that issuance of an administrative order
calling for PMAs for a preamendments device be preceded by a proposed
order and a meeting of a classification panel. As amended, this section
of the FD&C Act does not prescribe when these two events (the panel and
proposed order) must occur in relation to each other. More importantly,
FDA believes that the Panel's deliberations and recommendations remain
relevant and fully satisfy the requirements in section 515(b) of FD&C
Act.
FDA disagrees with the comment that the Panel did not consider new
information contained in the 2009 reclassification petition. A
representative from the petitioner was present at the meeting and
provided comments on the reclassification petition during the Panel
meeting (Ref. 1). In addition, the petitioner was given an opportunity
to explain the petitioner's reasons for why AEDs should be class II
devices, including a discussion of the special controls described in
the reclassification petition. Therefore, the Panel heard the
petitioner's arguments and these arguments were available for the
Panel's consideration when it made its recommendation.
(Comment 9) One commenter objected to FDA's use of the term
``diagnose'' in the proposed order to describe the functionality of
AEDs (78 FR 17893, March 25, 2013), and stated that AEDs sense
shockable rhythms and are not diagnostic devices.
(Response 9) FDA disagrees that these devices do not perform
diagnostic functions. AEDs analyze and interpret ECG data to produce an
assessment as to whether a shock should be delivered; while FDA does
believe that AEDs have diagnostic functions, we note that the
regulatory identification for the device in 21 CFR 870.5310(a), as
finalized in the order, does not use the term diagnose, and instead
describes the function of the device as ``analyzes'' and
``interprets.''
(Comment 10) One commenter stated that FDA's proposal to allow
manufacturers to ``bundle'' several AED models under a single PMA is
inconsistent with the PMA regulatory paradigm, which relies on a
device-by-device assessment. The comment points to FDA's guidance on
bundling, which states that ``[g]enerally, [manufacturers] should not
bundle differing generic device types in a single PMA submission
because of the substantially different pre-clinical and clinical data
needed to support each of the devices'' (Ref. 5).
(Response 10) FDA disagrees with the comment. Different AED models
can be included in one PMA if they are the same generic device type.
Because shock advisory algorithms and defibrillation waveforms will
likely be common across various models from a given manufacturer of
devices, FDA expects the clinical data needed to support devices within
an appropriately bundled AED PMA to be the same. However, because of
the differences in device labeling and user requirements between
professional and lay use devices, FDA continues to believe that
separate PMAs should be submitted for a manufacturer's professional use
versus lay use devices. FDA believes this approach is least burdensome
to manufacturers and is consistent with the bundling guidance, which
states that ``[b]undling is appropriate for devices that present
scientific and regulatory issues that can most efficiently be addressed
during one review'' (Ref. 5).
(Comment 11) One comment requested clarification on whether
[[Page 4788]]
separate PMAs are required for AEDs and the associated AED accessories
when a company manufacturers both for use together. Two comments
requested additional clarification on whether accessories not specified
in the proposed order (such as electrocardiograph modules and
electrodes, training pads/batteries, protective carrying cases,
Bluetooth modules, hardware keys or specialized pads to reduce energy
for pediatric use, self-testers, SpO2/blood pressure
monitoring devices, cardiopulmonary resuscitation (CPR) aids, medical
device data systems (MDDS), etc.) would require PMAs. One comment
suggested that AED accessories that are already 510(k) cleared should
not be subject to premarket approval by virtue of being used with an
AED.
(Response 11) In response to this comment, FDA has revised the
identification language to clarify that AED accessories regulated under
21 CFR 870.5310 are ``those accessories necessary for the AED to detect
and interpret an electrocardiogram and deliver an electrical shock
(e.g., battery, pad electrode, adapter, and hardware keys for pediatric
use).'' Manufacturers of accessory devices that are not addressed by
the final order and are not already the subject of an existing
classification regulation should contact FDA.
Under the final order, manufacturers must submit PMAs for
accessories that are necessary for operation of the AED system (e.g.,
accessories necessary to allow the AED to detect or interpret an
electrocardiogram or deliver a defibrillation shock). These AED
accessories include batteries, pad electrodes (including reduced energy
pads for pediatric use), adapters, and hardware keys for pediatric use.
In response to this comment, FDA has added ``hardware keys for
pediatric use'' to the identification. Necessary AED accessories that
are for use with a specific AED should be included in that PMA for the
AED system as a whole. Alternatively, necessary AED accessories,
including those manufactured by a third party, may be submitted in
their own PMAs.
Accessories that are not necessary for the functionality of the AED
are not addressed by the final order. Currently marketed AED
accessories that are not addressed by the final order, such as
SpO2/blood pressure monitoring devices, ECG modules and
testers, CPR aids, and MDDS, may be subject to other regulations and
will continue to be subject to those existing regulations. Training
accessories such as training pads and batteries for training-only AEDs
are not currently subject to any additional regulations, and will not
become so as a result of the final order.
(Comment 12) One comment requested clarification regarding AEDs
being considered adulterated 90 days after the effective date of a
final order in the absence of submission of a statement of intent to
submit a PMA or the submission of a full PMA. The comment questioned
whether devices legally distributed prior to the 90th day could remain
in distribution. Another comment requested clarification on whether
manufacturers could continue to provide consumable accessory items
(such as batteries and pads) for previously distributed devices even if
a PMA will not be submitted for that AED model. Two comments requested
clarification on how and whether manufacturers would be allowed to
distribute components required for field servicing of a device,
including refurbished replacement devices, before PMAs are submitted
for the devices.
(Response 12) Under the final order (see section IV, ``the Final
Order'') and section 501(f)(2)(B) of the FD&C Act, PMAs must be
submitted within 90 days after the effective date of the final order
for currently marketed AED systems. If a PMA is not submitted, the
devices are adulterated. However, for the reasons discussed above, for
currently marketed AEDs, FDA does not intend to enforce compliance with
the 90-day deadline by which PMAs must be submitted for 15 months after
that deadline (i.e., 18 months after the effective date of the final
order), as long as a notice of intent to file a PMA is submitted within
90 days of the effective date of the final order (see section V,
``Implementation Strategy''). For currently marketed necessary AED
accessories, FDA also does not intend to enforce compliance with the
90-day deadline by which PMAs must be submitted for 57 months after
that deadline (i.e., 5 years after the effective date of the final
order) (see section V, ``Implementation Strategy). This intention
applies to necessary AED accessories regardless of whether a PMA is
being or has been sought for the AED model.
(Comment 13) One comment indicated that premarket review of medical
devices such as AEDs should include review of the software embedded
into the devices, including review of software verification and
validation documentation. The comment noted that such review should
also occur for software modifications and software developed for
maintenance of the devices, including self-test functions. The comment
relayed the importance of having reviewers with adequate training,
expertise, and experience.
(Response 13) FDA agrees with the comment. Review of AEDs under the
510(k) process has included a detailed review of software documentation
supporting premarket submissions by appropriately trained and
experienced FDA reviewers. The PMA review will also involve a review of
software documentation and will be conducted by trained and experienced
FDA reviewers.
(Comment 14) One comment suggested an alternative regulatory
approach whereby AEDs for medical professional use be reclassified into
class II and public access defibrillators used by laypersons remain in
class III with PMAs required. The comment stated that professional use
devices have advanced functionality and are operated by skilled and
trained professionals, which lessens the chance of human factor errors
and increases the likelihood that the user will be able to recognize
and troubleshoot any malfunctions. The comment stated that such users
can rely on past experience and other means of attempting to rescue a
patient, whereas lay users are often fully reliant on the AED. Two
comments also indicated that professional use devices are typically
manual defibrillators with additional functionality, including AED, and
that the proposed order would create an inconsistent system whereby the
same hardware if used only for manual defibrillation would be class II,
but by virtue of configuring the device to include AED functionality
would become a PMA class III product.
(Response 14) FDA disagrees with the comments and believes that the
submission of PMAs is warranted regardless of the intended user of the
device. FDA does not believe that there is sufficient information to
establish special controls to provide a reasonable assurance of safety
and effectiveness of the device regardless of the training of AED
users.
FDA recognizes that some manual defibrillators and AEDs share
common hardware and software platforms, and hence devices with similar
or identical platforms may receive different regulatory review based on
the configuration. For the reasons previously stated, however, FDA
continues to believe AED systems should be class III devices. FDA also
notes that the performance and other data needed to support safety and
effectiveness for hardware and software platforms for both types of
devices would be nearly identical; the difference would be related to
the amount of
[[Page 4789]]
information that must be submitted to FDA. For a PMA, more information
on the design controls process is required to be submitted whereas for
a 510(k) submission, some information may not need to be submitted and
instead can reside within the company's overall quality system and
associated design documentation. Such situations of different
regulatory processes have occurred in other product areas including
contact lenses (daily-wear are typically class II, whereas extended
wear are class III) and ablation devices (general surgical use are
class II, whereas use for treatment of atrial fibrillation is class
III), and FDA does not believe this changes the overall rationale
supporting the need for PMAs.
(Comment 15) Two comments noted that there are numerous companies
that refurbish and/or resell AEDs. The comments requested that FDA
include AED resellers and refurbishers in their consideration of
regulatory strategy.
(Response 15) Regardless of the supplier, the introduction or
delivery for introduction into interstate commerce of any device that
is adulterated is a prohibited act under section 301 of the FD&C Act
(21 U.S.C. 331) (see Comment 12). FDA encourages refurbishers and
resellers who have questions about the continued distribution of AEDs
to contact FDA via the pre-submission process.
(Comment 16) One comment requested clarification of the process for
modifications of currently marketed AEDs (and notifying FDA of such
modifications) during the 90-day period after the final order is
issued. One comment stated that given the nature of commercial,
electrical and mechanical components used in AEDs, an extended
transition period without the ability to implement changes would not be
tenable and would result in unavailability of devices. One comment
requested clarification on 510(k) submissions accepted for review, but
for which no decision had been rendered, prior to the effective date of
a final order calling for PMAs.
(Response 16) Under 21 CFR 870.5310, as amended, all new AED and
necessary AED accessories must have an approved PMA in effect before
being placed in commercial distribution. We recommend that
manufacturers of currently marketed AEDs contact FDA regarding
implementation of any changes necessary for their AEDs in order to
address safety concerns or to support ongoing distribution while PMA
approval is being sought. FDA understands that issues may arise
relating to part obsolescence or changes necessary to reduce a risk to
health posed by a currently marketed AED that is not functioning
properly.
All other changes need to be accounted for in a PMA. Moreover, all
new AED and necessary AED accessories must have an approved PMA in
effect before being placed in commercial distribution.
(Comment 17) One comment objected to the comparisons made by FDA at
the 2011 Panel meeting between implanted cardioverter defibrillators
(ICDs) (PMA devices) and AEDs. The comment noted the number of
commercial components (e.g., components supplied to multiple industries
for a variety of uses) in order to maintain affordable price-points for
AEDs and the potentially burdensome PMA supplements that would be
necessary to support incremental changes in manufacturing for AEDs. The
comment further contended that purchased component-related recalls for
AEDs have largely been a result of latent component failures and that
FDA's examples at the 2011 Panel meeting of QS concerns related to
changes to purchased components or device design would not have been
averted by the stricter premarket regulatory oversight via PMA
supplements.
(Response 17) FDA acknowledges that more stringent regulatory
oversight via the PMA process may not completely eliminate AED recalls.
FDA also recognizes that AEDs typically contain commercial components
and manufacturers will need to submit PMA supplements for changes to
these components, as well as changes to suppliers and manufacturing
processes. Use of commercial components in PMA devices is not uncommon
and changes at the component level may have led to some AED recalls and
adverse events, providing further support for increased regulatory
review. FDA continues to believe that these failures and the need for
careful consideration and adequate verification and validation of such
changes support more rigorous review under the PMA process.
(Comment 18) One comment requested clarification on activities
during the time after a notice of intent to file is submitted,
including whether FDA will place additional postmarket approval
requirements on previously distributed products as allowed under 21 CFR
814.82. The comment further asked whether IDEs would be required for
design changes (e.g., would an IDE be required to conduct human
factors/usability studies).
(Response 18) FDA will consider the need for postapproval
requirements in the context of each manufacturer's PMA submission and
the devices in distribution. FDA does not intend to exempt
manufacturers from the IDE requirements and hence any study which meets
the IDE requirements must be conducted in accordance with the
requirements of 21 CFR parts 50 and 812. There will be no extended
period for filing an IDE and studies may not be initiated without
appropriate IDE approvals. Manufacturers who have questions regarding
whether an IDE is needed for a particular AED study are encouraged to
interact with FDA via the pre-submission process.
IV. The Final Order
FDA is adopting its findings as published in the preamble of the
proposed order (78 FR 17890, March 25, 2013) and issuing this final
order to require the filing of a PMA for AED systems under 515(b) of
the FD&C Act (21 U.S.C. 360e(b)). An AED system consists of an AED and
those accessories necessary for the AED to detect and interpret an
electrocardiogram and deliver an electrical shock (e.g., battery, pad
electrode, adapter, and hardware keys for pediatric use). An AED system
analyzes the patient's electrocardiogram, interprets the cardiac
rhythm, and automatically delivers an electrical shock (fully automated
AED), or advises the user to deliver the shock (semi-automated or shock
advisory AED) to treat ventricular fibrillation or pulseless
ventricular tachycardia. Under section 515(b)(1)(A) of the FD&C Act (21
U.S.C. 360e(b)(1)(A)), PMAs for AED systems are required to be filed on
or before 90 days after the effective date of a final order. This final
order will revise 21 CFR part 870.
V. Implementation Strategy
Based on comments on the proposed order, we are clarifying FDA's
intentions regarding enforcing compliance with the final order (see
section IV, ``The Final Order'') and section 501(f)(2)(B) of the FD&C
Act (21 U.S.C. 351(f)(2)(B)).
A. Currently Marketed AEDs
Under the final order and section 501(f)(2)(B) (21 U.S.C.
351(f)(2)(B)), PMAs for currently marketed AEDs are required to be
filed on or before 90 days after the effective date of a final order.
However, for currently marketed AEDs, FDA does not intend to enforce
compliance with this 90-day deadline for 15 months after that deadline
(i.e., 18 months after the effective date of the final order), as long
as notice of intent to file a PMA is submitted within 90 days of the
effective date of the final order. The notification of the intent to
file a PMA submission must include a
[[Page 4790]]
list of all model numbers for which a manufacturer plans to seek
marketing approval through a PMA.
In conducting any clinical studies, AEDs may be distributed for
investigational use if the requirements of the IDE regulations (21 CFR
part 812) are met. There will be no extended period for filing an IDE
nor exemption from IDE requirements, and studies may not be initiated
without appropriate IDE approvals, where necessary.
B. Currently Marketed Necessary AED Accessories
Under the final order and section 501(f)(2)(B) (21 U.S.C.
351(f)(2)(B)), PMAs for currently marketed necessary AED accessories
are required to be filed on or before 90 days after the effective date
of this final order. However, for currently marketed necessary AED
accessories, FDA does not intend to enforce compliance with this 90-day
deadline for 57 months after the deadline (i.e., 5 years after the
effective date of the final order). Currently marketed necessary AED
accessory manufacturers are not required to file an intent-to-file by
the 90-day deadline.
After the effective date of the final order, new AEDs and necessary
AED accessories must have approved PMAs to be legally marketed. The
following tables show the regulatory timetable for currently marketed
AEDs and necessary AED accessories.
Table 1--Currently Marketed AEDs
----------------------------------------------------------------------------------------------------------------
Timetable for which FDA does
not intend to enforce Distribution period (time after
compliance (time after effective date of order)
effective date of order)
----------------------------------------------------------------------------------------------------------------
Intent to File a PMA.................. 90 days....................... Devices included in an intent to file:
18 months.
Devices not included in intent to file:
90 days.
File a PMA............................ 18 months..................... Until a not approvable decision or
denial decision is issued; can continue
distribution if an approval order is
issued.
----------------------------------------------------------------------------------------------------------------
Table 2--Currently Marketed Necessary AED Accessories
----------------------------------------------------------------------------------------------------------------
Timetable for which FDA does
not intend to enforce Distribution period (time after
compliance (time after effective date of order)
effective date of order)
----------------------------------------------------------------------------------------------------------------
Intent to File a PMA.................. N/A........................... N/A.
File a PMA............................ 60 months..................... Until a not approvable decision or
denial decision is issued; can continue
distribution if an approval order is
issued.
----------------------------------------------------------------------------------------------------------------
VI. Environmental Impact
The Agency has determined under 21 CFR 25.30 (h) that this action
is of a type that does not individually or cumulatively have a
significant effect on the human environment. Therefore, neither an
environmental assessment nor an environmental impact statement is
required.
VII. Paperwork Reduction Act of 1995
The final order refers to previously approved collections of
information found in FDA regulations. These collections of information
are subject to review by the Office of Management and Budget (OMB)
under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The
collections of information in 21 CFR part 812 have been approved under
OMB control number 0910-0078; the collections of information in 21 CFR
part 807, subpart E, have been approved under OMB control number 0910-
0120; the collections of information in 21 CFR part 814, subpart B,
have been approved under OMB control number 0910-0231; and the
collections of information under 21 CFR part 801 have been approved
under OMB control number 0910-0485.
VIII. Codification of Orders
Prior to the amendments by FDASIA, section 515(b) of the FD&C Act
(21 U.S.C. 360e(b)) provided for FDA to issue regulations to require
PMA approval for preamendments devices or devices found substantially
equivalent to preamendments devices. Section 515(b) of the FD&C Act (21
U.S.C. 360e(b)), as amended by FDASIA, provides for FDA to require PMA
approval for such devices by issuing a final order, following the
issuance of a proposed order in the Federal Register. FDA will continue
to codify the requirement for a PMA approval in the Code of Federal
Regulations. Therefore, under section 515(b)(1)(A) of the FD&C Act (21
U.S.C. 360e(b)(1)(A)), as amended by FDASIA, in this final order, we
are requiring PMA approval for AED systems and we are making the
language in 21 CFR 870.5310 consistent with the final version of this
order.
IX. References
The following references have been placed on display in the
Division of Dockets Management (see ADDRESSES) and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday.
(FDA has verified the Web site address, but FDA is not responsible for
any subsequent changes to the Web site after this document publishes in
the Federal Register.)
1. Meeting Materials for 515(i) Regulatory Classification of
Automated External Defibrillator Systems, January 25, 2011,
available at https://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/MedicalDevicesAdvisoryCommittee/CirculatorySystemDevicesPanel/ucm240575.htm.
2. U.S. Food and Drug Administration, Medical Device Recalls
Database, available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm.
3. Requests for Feedback on Medical Device Submissions: The Pre-
Submission Program and Meetings with Food and Drug Administration
Staff, Guidance for Industry and Food and Drug Administration Staff,
February 18, 2014, available at https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM311176.pdf.
4. Acceptance and Filing Reviews for Premarket Approval
Applications (PMAs), Guidance for Industry and Food and Drug
Administration Staff, December 31, 2012, available at https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM313368.pdf.
5. Guidance for Industry and FDA Staff: Bundling Multiple
Devices or Multiple Indications in a Single Submission, June 22,
2007, available at https://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm089731.htm.
[[Page 4791]]
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
870 is amended as follows:
PART 870--CARDIOVASCULAR DEVICES
0
1. The authority citation for 21 CFR part 870 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Section 870.5310 is amended by revising the section heading and
paragraphs (a) and (c) to read as follows:
Sec. 870.5310 Automated external defibrillator system.
(a) Identification. An automated external defibrillator (AED)
system consists of an AED and those accessories necessary for the AED
to detect and interpret an electrocardiogram and deliver an electrical
shock (e.g., battery, pad electrode, adapter, and hardware key for
pediatric use). An AED system analyzes the patient's electrocardiogram,
interprets the cardiac rhythm, and automatically delivers an electrical
shock (fully automated AED), or advises the user to deliver the shock
(semi-automated or shock advisory AED) to treat ventricular
fibrillation or pulseless ventricular tachycardia.
* * * * *
(c) Date PMA or notice of completion of PDP is required. A PMA will
be required to be submitted to the Food and Drug Administration by
April 29, 2015, for any AED that was in commercial distribution before
May 28, 1976, or that has, by April 29, 2015, been found to be
substantially equivalent to any AED that was in commercial distribution
before May 28, 1976. A PMA will be required to be submitted to the Food
and Drug Administration by April 29, 2015, for any AED accessory
described in paragraph (a) that was in commercial distribution before
May 28, 1976, or that has, by April 29, 2015, been found to be
substantially equivalent to any AED accessory described in paragraph
(a) that was in commercial distribution before May 28, 1976. Any other
AED and AED accessory described in paragraph (a), shall have an
approved PMA or declared completed PDP in effect before being placed in
commercial distribution.