Cardiovascular Devices; Reclassification of External Pacemaker Pulse Generator Devices; Reclassification of Pacing System Analyzers, 54927-54936 [2014-21814]
Download as PDF
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA–2011–N–0650]
Cardiovascular Devices; Withdrawal of
Proposed Rule of Reclassification of
External Pacemaker Pulse Generator
Devices
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed rule; withdrawal.
The Food and Drug
Administration (FDA) is withdrawing
the proposed rule the Agency issued in
the Federal Register of October 17,
2011. In that document, FDA proposed
to reclassify the external pacemaker
pulse generator (EPPG) devices, a
preamendments class III device into
class II (special controls). In response to
the requirements under the Food and
Drug Administration Safety and
Innovation Act (FDASIA) and new
information received during a panel
meeting, FDA is withdrawing the
proposed rule and issuing a proposed
administrative order to reclassify
EPPGs.
SUMMARY:
The proposed rule is withdrawn
on September 15, 2014.
FOR FURTHER INFORMATION CONTACT:
Hina Pinto, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1652, Silver Spring,
MD 20993, 301–796–6351.
SUPPLEMENTARY INFORMATION:
DATES:
rmajette on DSK2TPTVN1PROD with PROPOSALS
I. Background—Regulatory Authorities
On October 17, 2011, FDA published
in the Federal Register (76 FR 64223) a
proposed rule proposing the
reclassification of external pacemaker
pulse generator (EPPG) devices from
class III to class II with special controls.
FDA identified special controls that the
Agency believed would provide
reasonable assurance of safety and
effectiveness for the device type. FDA
considered EPPGs in accordance with
the reserved criteria and determined
that the device type does require
premarket notification.
On July 9, 2012, FDASIA was enacted.
Section 608(a) of FDASIA (Pub. L. 112–
144) amended section 513(e) of the
Federal Food, Drug, and Cosmetic Act
(the FD&C Act) (21 U.S.C. 360c(e))
changing the process for reclassifying a
device from rulemaking to an
administrative order. Subsequent to the
publication of the proposed rule,
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
FDASIA’s amendments to section 513 of
the FD&C Act required FDA to hold a
classification panel (an FDA advisory
committee) meeting on the classification
of this device. On September 11, 2013,
a meeting of the Circulatory System
Devices Panel (the Panel) was held to
discuss whether EPPG devices should
be reclassified or remain in class III (Ref.
1). There was Panel consensus that
EPPG devices did not fit the regulatory
definition of a class III device. Coupled
with the rationale that special controls
could be established to reasonably
demonstrate an assurance of safety and
effectiveness, the Panel recommended
class II (special controls) for EPPG when
intended for cardiac rate control or
prophylactic arrhythmia prevention.
II. Withdrawal of the Proposed Rule
FDA provided an opportunity for
interested parties to comment on the
proposed rule for EPPG (76 FR 64223).
FDA received three comments to the
docket in response to the 2011 proposed
rule. These comments were received
and have been considered during the
presentations to the Panel and in
developing the proposed order. In
response to these comments and
findings at the Panel meeting, FDA is
withdrawing the proposed rule for these
devices and is issuing a proposed
administrative order.
III. Proposed Reclassification
Elsewhere in this issue of the Federal
Register, FDA is proposing in an order
to reclassify EPPG devices, currently a
preamendments class III device, into
class II (special controls). FDA
continues to review the merits of the
submissions for requests for
reclassification that meet the
requirements under 21 CFR 860.123,
submitted in response to the proposed
rule.
IV. Reference
The following reference has been
placed on display in the Division of
Dockets Management (HFA–305), Food
and Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852,
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and is available
electronically at https://
www.regulations.gov. (FDA has verified
the Web site address in this reference
section, but we are not responsible for
any subsequent changes to the Web site
after this document publishes in the
Federal Register.)
1. The panel transcript and other
meeting materials for the September 11,
2013, Circulatory System Devices Panel
are available on FDA’s Web site at
PO 00000
Frm 00011
Fmt 4702
Sfmt 4702
54927
https://www.fda.gov/
AdvisoryCommittees/
CommitteesMeetingMaterials/
MedicalDevices/
MedicalDevicesAdvisoryCommittee/
CirculatorySystemDevicesPanel/
ucm342357.htm.
Dated: September 8, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014–21816 Filed 9–12–14; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA–2011–N–0650]
Cardiovascular Devices;
Reclassification of External Pacemaker
Pulse Generator Devices;
Reclassification of Pacing System
Analyzers
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed order.
The Food and Drug
Administration (FDA) is proposing in
this administrative order to reclassify
the external pacemaker pulse generator
(EPPG) devices, a preamendments class
III device into class II (special controls),
and to amend the device identification
and reclassify the pacing system
analyzers (PSAs) into class II (special
controls). Specifically, single and dual
chamber PSAs, which are currently
classified with EPPG devices, and triple
chamber PSAs (TCPSAs), which are
postamendments class III devices, are
proposed to be reclassified to class II
devices. FDA is proposing this
reclassification based on new
information pertaining to the device.
This proposed action would implement
certain statutory requirements.
DATES: Submit either electronic or
written comments on the proposed
order by December 15, 2014. See section
XII for the effective date of any final
order that may publish based on this
proposed order.
ADDRESSES: You may submit comments,
identified by Docket No. FDA–2011–N–
0650, by any of the following methods:
SUMMARY:
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
E:\FR\FM\15SEP1.SGM
15SEP1
54928
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
rmajette on DSK2TPTVN1PROD with PROPOSALS
Written Submissions
Submit written submissions in the
following ways:
• Mail/Hand delivery/Courier (for
paper submissions): Division of Dockets
Management (HFA–305), Food and Drug
Administration, 5630 Fishers Lane, rm.
1061, Rockville, MD 20852.
Instructions: All submissions received
must include the Agency name and
Docket No. FDA–2011–N–0650 for this
order. All comments received may be
posted without change to https://
www.regulations.gov, including any
personal information provided. For
additional information on submitting
comments, see the ‘‘Comments’’ heading
of the SUPPLEMENTARY INFORMATION
section of this document.
Docket: For access to the docket to
read background documents or
comments received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Division of Dockets
Management, 5630 Fishers Lane, rm.
1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Hina Pinto, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, rm. 1652, Silver Spring,
MD 20993, 301–796–6351.
SUPPLEMENTARY INFORMATION:
I. Background—Regulatory Authorities
The Food, Drug, and Cosmetic Act
(the FD&C Act) as amended by the
Medical Device Amendments of 1976
(the 1976 amendments) (Pub. L. 94–
295), the Safe Medical Devices Act of
1990 (Pub. L. 101–629), the Food and
Drug Administration Modernization Act
of 1997 (Pub. L. 105–115), the Medical
Device User Fee and Modernization Act
of 2002 (Pub. L. 107–250), the Medical
Devices Technical Corrections Act of
2004 (Public Law 108–214), the Food
and Drug Administration Amendments
Act of 2007 (Pub. L. 110–85), and the
Food and Drug Administration Safety
and Innovation Act (FDASIA) (Pub. L.
112–144) establishes a comprehensive
system for the regulation of medical
devices intended for human use.
Section 513 of the FD&C Act (21 U.S.C.
360c) established three categories
(classes) of devices, reflecting the
regulatory controls needed to provide
reasonable assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Section 513(a)(1) of the FD&C Act
defines class II devices as those devices
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
for which the general controls by
themselves are insufficient to provide
reasonable assurance of safety and
effectiveness, but for which there is
sufficient information to establish
special controls to provide such
assurance.
Under section 513 of the FD&C Act,
devices that were in commercial
distribution before the enactment of the
1976 amendments, May 28, 1976
(generally referred to as preamendments
devices), are classified after FDA has: (1)
Received a recommendation from a
device classification panel (an FDA
advisory committee); (2) published the
panel’s recommendation for comment,
along with a proposed regulation
classifying the device; and (3) published
a final regulation classifying the device.
FDA has classified most
preamendments devices under these
procedures.
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices), are
automatically classified by section
513(f) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II or FDA issues an order finding the
device to be substantially equivalent, in
accordance with section 513(i) of the
FD&C Act, to a predicate device that
does not require premarket approval.
The Agency determines whether new
devices are substantially equivalent to
predicate devices by means of
premarket notification procedures in
section 510(k) of the FD&C Act (21
U.S.C. 360(k)) and part 807 (21 CFR part
807).
A preamendments device that has
been classified into class III and devices
found substantially equivalent by means
of premarket notification (510(k))
procedures to such a preamendments
device or to a device within that type
may be marketed without submission of
a premarket approval (PMA) application
until FDA issues a final order under
section 515(b) of the FD&C Act (21
U.S.C. 360e(b)) requiring premarket
approval or until the device is
subsequently reclassified into class I or
class II.
On July 9, 2012, FDASIA was enacted.
Section 608(a) of FDASIA amended the
device reclassification procedures under
section 513(e) of the FD&C Act,
changing the process for reclassifying a
device from rulemaking to an
administrative order. Prior to the
enactment of FDASIA, FDA published a
proposed rule under section 513(e) of
the FD&C Act proposing the
PO 00000
Frm 00012
Fmt 4702
Sfmt 4702
reclassification of EPPG devices (76 FR
64223, October 17, 2011). Three sets of
comments were received on the
proposed rule. The three sets of
comments submitted in response to the
proposed rule on EPPG devices will be
considered under this proposed
administrative order and do not need to
be resubmitted. FDA is issuing this
proposed administrative order to
comply with the new procedural
requirement created by FDASIA when
reclassifying a preamendments class III
device, as well as to reclassify a
postamendments class III device. Also,
as required by section 513(e) of the
FD&C Act for preamendment devices,
FDA convened a device classification
panel meeting which discussed the
proposed reclassification on September
11, 2013 (78 FR 49272). This action is
intended solely to fulfill the procedural
requirements for reclassification
implemented by FDASIA.
Section 513(e) of the FD&C Act
provides that FDA may, by
administrative order, reclassify a device
based upon ‘‘new information.’’ FDA
can initiate a reclassification under
section 513(e) or an interested person
may petition FDA to reclassify a
preamendments device. The term ‘‘new
information,’’ as used in section 513(e)
of the FD&C Act, includes information
developed as a result of a reevaluation
of the data before the Agency when the
device was originally classified, as well
as information not presented, not
available, or not developed at that time.
(See, e.g., Holland-Rantos Co. v. United
States Dep’t of Health, Educ., & Welfare,
587 F.2d 1173, 1174 n.1 (D.C. Cir. 1978);
Upjohn v. Finch, 422 F.2d 944 (6th Cir.
1970); Bell v. Goddard, 366 F.2d 177
(7th Cir. 1966).)
A postamendments device that has
been initially classified in class III
under section 513(f)(1) of the FD&C Act
may be reclassified later into class I or
class II under section 513(f)(3) of the
FD&C Act. Section 513(f)(3) provides
that FDA acting by administrative order
can reclassify the device into class I or
class II on its own initiative under
section 513(f)(1) of the FD&C Act, or in
response to the petition of the
manufacturer or importer of the device.
FDA’s regulations in 21 CFR 860.134 set
forth the procedures for the filing and
review of a petition for reclassification
of these class III devices. To change the
classification of the device, the
proposed new class must have sufficient
regulatory controls to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use.
Reevaluation of the data previously
before the Agency is an appropriate
E:\FR\FM\15SEP1.SGM
15SEP1
rmajette on DSK2TPTVN1PROD with PROPOSALS
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
basis for subsequent regulatory action
where the reevaluation is made in light
of newly available regulatory authority
(see Bell v. Goddard, supra, 366 F.2d at
181; Ethicon, Inc. v. FDA, 762 F.Supp.
382, 388–391 (D.D.C. 1991)), or in light
of changes in ‘‘medical science.’’ (See
Upjohn v. Flinch supra, 422 F.2d at
951.) Whether data before the Agency
are old or new data, the ‘‘new
information’’ to support reclassification
under section 513(e) must be ‘‘valid
scientific evidence,’’ as defined in
section 513(a)(3) of the FD&C Act and
§ 860.7(c)(2). (See, e.g., General Medical
Co. v. FDA, 770 F.2d 214 (D.C. Cir.
1985); Contact Lens Manufacturers
Association v. FDA, 766 F.2d 592 (D.C.
Cir. 1985), cert. denied, 474 U.S. 1062
(1986).
FDA relies upon ‘‘valid scientific
evidence’’ in the classification process
to determine the level of regulation for
devices. To be considered in the
reclassification process, the valid
scientific evidence upon which the
Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending premarket
approval application (PMA). (See
section 520(c) of the FD&C Act (21
U.S.C. 360j(c)).
Section 513(e)(1) of the FD&C Act sets
forth the process for issuing a final
order. Specifically, prior to the issuance
of a final order reclassifying a device,
the following must occur: (1)
Publication of a proposed order in the
Federal Register; (2) a meeting of a
device classification panel described in
section 513(b) of the FD&C Act; and (3)
consideration of comments from all
affected stakeholders, including
patients, payors, and providers. In
addition, the proposed order must set
forth the proposed reclassification, and
a substantive summary of the valid
scientific evidence concerning the
proposed reclassification, including the
public health benefits of the use of the
device, and the nature and incidence (if
known) of the risk of the device. (See
section 513(e)(1)(A)(i) of the FD&C Act.)
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the premarket
notification requirements under section
510(k) of the FD&C Act if the Agency
determines that premarket notification
is not necessary to assure the safety and
effectiveness of the device. FDA has
determined that premarket notification
is necessary to reasonably assure the
safety and effectiveness of EPPG and
PSA devices.
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
II. Regulatory History of the Device
A. EPPG Devices
On March 9, 1979, FDA published a
proposed rule in the Federal Register
for classification of EPPG devices into
class III based on the recommendation
of the Cardiovascular Devices Panel (the
Panel) (44 FR 13284 at 13372). The
Panel meeting recommended EPPG
devices be classified into class III
because the device provided temporary
life support and that certain kinds of
failures could cause this device to emit
inappropriate electrical signals, which
could cause cardiac irregularities and
death. The Panel indicated that general
controls alone would not be sufficient
and that there was not enough
information to establish a performance
standard. Consequently, the Panel
believed that premarket approval was
necessary to reasonably assure the safety
and effectiveness of the device. In 1980,
FDA classified EPPG into class III under
§ 870.3600 (21 CFR 870.3600) after
receiving no comments on the proposed
rule (45 FR 7907, February 5, 1980). In
1987, FDA published a clarification by
inserting language in the codified
language stating that no effective date
had been established for the
requirement for premarket approval for
EPPG devices (52 FR 17732, May 11,
1987).
In 2009, FDA published an order in
the Federal Register under section
515(i) of the FD&C Act to call for
information on the remaining class III
510(k) preamendment devices,
including EPPG devices (74 FR 16214,
April 9, 2009). In response to that order,
FDA received two reclassification
petitions from one device manufacturer
who requested that EPPG devices be
reclassified into class II. The
manufacturers stated that safety and
effectiveness of these devices may be
assured by performance standards, the
intended use environment, postmarket
surveillance to include Medical Device
Reporting (MDRs), FDA inspections of
manufacturing facilities, and premarket
review of performance testing in a
510(k) submission. The manufacturers
specifically noted that the FDA
recognized consensus standard,
International Electrotechnical
Commission (IEC) 60601–2–31:
’Particular requirements for the basic
safety and essential performance of
external cardiac pacemakers with
internal power source’ provides
adequate design and testing parameters
for EPPG devices.
On October 17, 2011, FDA published
a proposed rule proposing the
reclassification of EPPG devices from
class III to class II (76 FR 64223) and
PO 00000
Frm 00013
Fmt 4702
Sfmt 4702
54929
announcing the availability of a draft
Special Controls Guidance Document
that, if finalized, would serve as a
special control, if FDA reclassified these
devices. FDA believed that the special
controls described in the draft special
controls guidance document entitled
‘‘Class II Special Controls Guidance
Document: External Pacemaker Pulse
Generator’’ would be sufficient to
mitigate the risks to health associated
with EPPG (Ref. 1).
The proposed rule provided for a
comment period that was open until
January 17, 2012. FDA received three
sets of comments. These comments
stated that: (1) FDA should retain EPPG
in class III, (2) FDA’s reclassification
proposed rule was not adequately
supported by new publicly valid
scientific evidence, (3) MDR data
showed that existing performance
standards are insufficient, (4) there were
no publicly available performance
standards that would apply to EPPG, (5)
FDA should convene an advisory
committee (the Panel) to seek a
recommendation on the classification of
EPPG, and (6) the recall process after
reclassification of EPPG would need to
be clarified. These comments were
considered by FDA in drafting this
proposed order.
B. PSA Devices
Single and dual chamber PSAs have
historically been classified with EPPG
devices. These devices combine the
functionality of a single or dual chamber
EPPG, which is currently class III and
the functionality of a pacemaker
electrode function tester, which is
regulated as a class II device (under
§ 870.3720 (21 CFR 870.3720)). Single
and dual chamber PSA devices have
been found substantially equivalent to
EPPG devices through the 510(k)
process. Triple chamber PSA (TCPSA)
devices have not been determined to be
substantially equivalent through the
510(k) process, and since this
technology was not on the market in
1976, TCPSAs have been reviewed
through the PMA process as
postamendment class III devices.
On July 9, 2012, FDASIA was enacted,
which amended the device
reclassification procedures under
sections 513 and 515 of the FD&C Act.
Accordingly, FDA is issuing a proposed
administrative order to comply with the
new procedural requirement created by
FDASIA when reclassifying a
preamendments class III device.
Further, FDA intends to codify the
proposed special controls within the
§ 870.3600 classification regulation for
EPPG and to create a separate
E:\FR\FM\15SEP1.SGM
15SEP1
54930
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
classification regulation for PSA
devices.
As explained further in section VII, a
meeting of the Circulatory System
Devices Panel (the 2013 Panel) took
place on September 11, 2013, to discuss
whether EPPG and TCPSA devices
should be reclassified or remain in class
III (Ref. 2). FDA included a discussion
of TCPSA devices in the 2013 Panel
discussion because the risks to health
and proposed special controls were very
similar to those being proposed for the
EPPG devices already under
consideration. The 2013 Panel
recommended that EPPG devices be
reclassified to class II with special
controls when intended for cardiac rate
control or prophylactic arrhythmia
prevention. The 2013 Panel also
recommended that TCPSA devices be
reclassified to class II with special
controls when intended for use during
the pulse generator implant procedure.
FDA is not aware of new information
that would provide a basis for a
different recommendation or finding.
rmajette on DSK2TPTVN1PROD with PROPOSALS
III. Device Description
A. EPPG Devices
An EPPG is a device that has a power
supply and electronic circuits that
produce a periodic electrical pulse to
stimulate the heart. This device, which
is used outside the body, is used as a
temporary substitute for the heart’s
intrinsic pacing system until a
permanent pacemaker can be implanted,
or to control irregular heartbeats in
patients following events such as
cardiac surgery or a myocardial
infarction. The device may have
adjustments for pacing rate, pulse
amplitude, pulse width (duration), Rwave sensitivity, and other pacing
variables.
An EPPG device is designed to be
used with cardiac pacing lead systems
for temporary atrial and/or ventricular
pacing. An EPPG system generally
includes the pulse generator, extension
cables, and adaptors which connect the
extension cable to the implanted pacing
lead and are critical to the functionality
of the EPPG system. The pacing leads
for use with EPPG may be for temporary
or permanent use for either tranvenous
or epicardial uses. The pacing leads are
not considered part of the EPPG device
because they have their own regulatory
designations (21 CFR 870.3680)
depending on their design and intended
use.
EPPG devices are used exclusively in
hospital environments with the patients
supervised by qualified medical
personnel. The electrical and heart
rhythm of patients are continuously
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
monitored using EPPG-independent
electrocardiogram (ECG) monitors
usually with alarm functions.
Independent ECG monitoring
requirements are identified in
international standards, such as IEC
60601–2–31 for device design.
FDA is also proposing in this order to
slightly modify the identification
language from the way it is presently
written in § 870.3600(a) to clarify that
these are prescription devices in
accordance with § 801.109 (21 CFR
801.109).
B. PSA Devices
A PSA combines the functionality of
a pacemaker electrode function tester
(under § 870.3720) and an EPPG. A
pacemaker electrode function tester is a
device that is connected to an implanted
pacemaker lead that supplies an
accurately calibrated, variable pacing
pulse for measuring the patient’s pacing
threshold and intracardiac R-wave
potential. A PSA can temporarily take
over pacing functions while
simultaneously testing one or more
implanted pacing leads. PSA devices
can be single, dual, or triple chamber,
translating into the measurement
capabilities/functionalities of the
device. Single chamber PSAs typically
measure pacing capture threshold,
whereas in the case of dual chamber
PSAs, the device can also measure
conduction times or intrinsic
atrioventricular delay. In the case of a
TCPSA, the device can also function as
a biventricular external pacemaker and
measure the intrinsic intra-ventricular
(VV) interval.
IV. Proposed Reclassification
A. EPPG Devices
FDA is proposing that EPPG devices
be reclassified from class III to class II.
In this proposed order, the Agency has
identified special controls under section
513(a)(1)(B) of the FD&C Act that,
together with general controls
(including prescription-use restrictions)
applicable to the devices, would
provide reasonable assurance of their
safety and effectiveness. Absent the
special controls identified in this
proposed order, general controls
applicable to the device are insufficient
to provide reasonable assurance of the
safety and effectiveness of the device.
Since the time of the 1979 classification,
new information about use and pacing
technology for this device has become
sufficiently available to establish special
controls. FDA believes that this new
information is sufficient to demonstrate
that the proposed special controls, when
finalized, can effectively mitigate the
PO 00000
Frm 00014
Fmt 4702
Sfmt 4702
risks to health identified in section V,
and that these special controls, together
with general controls, will provide a
reasonable assurance of safety and
effectiveness for EPPG devices.
B. PSA Devices
FDA is proposing to create a separate
classification regulation for PSA
devices, including single, dual, and
triple chamber PSA devices that will be
reclassified from class III to class II. In
this proposed order, the Agency has
identified special controls under section
513(a)(1)(B) of the FD&C Act that,
together with general controls
(including prescription-use restrictions)
applicable to the devices, would
provide reasonable assurance of their
safety and effectiveness. Absent the
special controls identified in this
proposed order, general controls
applicable to the device are insufficient
to provide reasonable assurance of the
safety and effectiveness of the device.
Since the 1979 classification of
temporary external pacing devices, new
information about device use and
pacing technology has become available
to establish special controls. FDA
believes that this new information is
sufficient to demonstrate that the
proposed special controls can
effectively mitigate the risks to health
identified in section V, and that these
special controls, together with general
controls, will provide a reasonable
assurance of safety and effectiveness for
PSA devices.
Section 510(m) of the FD&C Act
authorizes the Agency to exempt class II
devices from premarket notification
(510(k)) submission. FDA has
considered EPPG and PSA devices in
accordance with the reserved criteria set
forth in section 513(a) and determined
that both devices require premarket
notification (510(k) of the FD&C Act).
Therefore, the Agency does not intend
to exempt these proposed class II
devices from premarket notification
(510(k)) submission as provided under
section 510(m) of the FD&C Act.
V. Risks to Health
A. EPPG Devices
After considering available
information for the classification of
these devices, including the
recommendations of the advisory
committees (panels) for the
classification of these devices, FDA has
evaluated the risks to health associated
with the use of EPPG devices and
determined the following risks to health
are associated with its use:
• Failure to pace: Improper settings,
electromagnetic interference, or failure
E:\FR\FM\15SEP1.SGM
15SEP1
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
of mechanical/electrical components of
the device can prevent pacing of the
patient’s heart.
• Improperly high pacing rate:
Undersensing during demand pacing,
unintended asynchronous pacing, or
improper use of burst/overdrive pacing
can cause harmful acceleration of heart
rate or induce harmful arrhythmias such
as ventricular tachycardia.
• Improperly low pacing rate:
Oversensing or use error can result in
stimulation pulses being delivered at an
unwanted low pacing rate, which can
result in untreated symptomatic
bradycardia.
• Improper pacing leading to
unwanted stimulation: Pacing during
vulnerable periods of the cardiac cycle
or at higher than programmed
amplitude can induce arrhythmias.
• Micro/macro shock: Uncontrolled
leakage currents or patient auxiliary
currents can cause an electric shock
resulting in an arrhythmia or cardiac
tissue damage.
rmajette on DSK2TPTVN1PROD with PROPOSALS
B. PSA Devices
After considering available
information for the classification of
these devices, including the
recommendations of the advisory
committees (panels) for the
classification of these devices, FDA has
evaluated the risks to health associated
with the use of PSA devices and
determined the following risks to health
are associated with its use:
• Failure to pace: Improper settings,
electromagnetic interference, or failure
of mechanical/electrical components of
the device can prevent pacing of the
patient’s heart.
• Improperly high pacing rate:
Undersensing during demand pacing,
unintended asynchronous pacing, or
improper use of burst/overdrive pacing
can cause harmful acceleration of heart
rate or induce harmful arrhythmias such
as ventricular tachycardia.
• Improperly low pacing rate:
Oversensing or use error can result in
stimulation pulses being delivered at an
unwanted low pacing rate, which can
result in untreated symptomatic
bradycardia.
• Improper pacing leading to
unwanted stimulation: Pacing during
vulnerable periods of the cardiac cycle
or at higher than programmed
amplitude can induce arrhythmias. For
TCPSAs, this risk includes VV
dyssynchrony.
• Micro/macro shock: Uncontrolled
leakage currents or patient auxiliary
currents can cause an electric shock
resulting in an arrhythmia or cardiac
tissue damage.
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
• Misdiagnosis: If the zero or
calibration of the device is inaccurate or
unstable the device may generate
inaccurate diagnostic data. If inaccurate
diagnostic data are used in managing
the patient, the physician may prescribe
a course of treatment that places the
patient at risk unnecessarily.
VI. Summary of Reasons for
Reclassification
FDA believes that EPPG devices and
PSA devices should be reclassified from
class III to class II because special
controls, in addition to general controls,
can be established to provide reasonable
assurance of the safety and effectiveness
of the devices, and because general
controls themselves are insufficient to
provide reasonable assurance of their
safety and effectiveness. In addition,
there is now sufficient information to
establish special controls to provide
such assurance. FDA also believes that
TCPSA devices—as a subset of PSA
devices—can achieve a reasonable
assurance of safety and effectiveness
using the same special controls
proposed for EPPG and PSA devices
with the addition of limiting use to the
duration of the implant procedure in
order to mitigate the risk of unwanted
interventricular stimulation leading to
arrhythmia (captured as misdiagnosis
and improper pacing leading to
unwanted stimulation in the list of risks
to health in section V).
VII. Summary of Data Upon Which the
Reclassification Is Based
A. EPPG Devices
FDA believes that the identified
special controls, in addition to general
controls, are necessary to provide
reasonable assurance of safety and
effectiveness of these devices.
Therefore, in accordance with sections
513(e) and 515(i) of the FD&C Act and
§ 860.130, based on new information
with respect to the device and taking
into account the public health benefit of
the use of the device and the nature and
known incidence of the risk of the
device, FDA, on its own initiative, is
proposing to reclassify this
preamendments class III device into
class II. The Agency has identified
special controls that would provide
reasonable assurance of their safety and
effectiveness. EPPG are prescription
devices restricted to patient use only
upon the authorization of a practitioner
licensed by law to administer or use the
device. Since 1979 when FDA classified
EPPG devices into class III, sufficient
evidence has been developed to support
a reclassification to class II with the
establishment of special controls. FDA
PO 00000
Frm 00015
Fmt 4702
Sfmt 4702
54931
has been reviewing these devices for
many years and their risks are well
known. The risks to health are
identified in section V, and FDA
believes these risks can be adequately
mitigated by special controls.
EPPG devices that use temporary
cardiac pacing for the purposes of rate
control or treatment of bradycardia use
mature technology with wellestablished evidence of effectiveness
(Refs. 3, 8, 9, 11, 13). A review of 14
clinical studies published over four
decades shows that temporary external
pacing is generally safe and has an
electrophysiologic as well as
hemodynamic benefit when used as
indicated (Refs. 3 to 17).
The low frequency of serious adverse
events as evidenced through FDA’s
Manufacturer and User Facility Device
Experience (MAUDE) database, the low
rate of postmarket recalls, the
established scientific evidence to
support pacing for specific indications,
the hospital use environment, and
FDA’s review experience with these
devices, all support the reclassification
of these devices to class II. In addition,
several key performance standards (such
as IEC 60601–1 and IEC 60601–2–31)
that address various aspects of design
and performance have been developed
and used to support marketing
applications since the original
classification. In light of these
considerations, FDA has tentatively
concluded that the identified special
controls, in addition to general controls,
provide reasonable assurance of the
safety and effectiveness of EPPG
devices.
FDA’s presentation to the 2013 Panel
included a summary of the available
safety and effectiveness information for
EPPG devices, including adverse event
reports from FDA’s MAUDE database
and available literature. Based on the
available scientific literature, which
supports that use of EPPG devices may
be beneficial for patients needing
temporary atrial and/or ventricular
pacing, FDA recommended to the 2013
Panel that EPPG devices be reclassified
to class II (special controls). The 2013
Panel discussed and made
recommendations regarding the
regulatory classification of EPPG
devices to either reconfirm to class III
(subject to premarket approval
application) or reclassify to class II
(subject to special controls). The 2013
Panel agreed with FDA’s conclusion
that the available scientific evidence is
adequate to support the safety and
effectiveness of EPPG devices. The 2013
Panel also acknowledged that EPPG
devices are life-supporting devices and
provided the following rationale per
E:\FR\FM\15SEP1.SGM
15SEP1
54932
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
rmajette on DSK2TPTVN1PROD with PROPOSALS
§ 860.93 for recommending that EPPG
devices be reclassified to class II: (1)
These devices are used exclusively in
the hospital environment where backup
monitoring is available, hazards can be
recognized and treated immediately,
and where there is a reasonable
expectation that users are adequately
trained; (2) there is sufficient clinical
experience that attests to the benefit of
the device; and (3) the recommended
special controls will mitigate the health
risks associated with the device.
The 2013 Panel also agreed with the
identified risks to health presented at
the meeting; however, it recommended
that FDA consider rewording some of
the language for clarity and also to
ensure that certain hazards, such as
asynchronous pacing and arrhythmia
induction, are included in the risks to
health. FDA agrees with the 2013
Panel’s recommendations and modified
the risks to health accordingly as
outlined in section V. The 2013 Panel
also agreed with FDA’s proposed special
controls outlined in section VIII;
however, the 2013 Panel further
recommended that FDA add labeling
requirements for proper training, proper
maintenance of the device, and remedial
actions for failures due to lead
connections. FDA agrees with the 2013
Panel and the proposed special controls
have been modified to reflect more
specific labeling requirements.
The 2013 Panel transcript and other
meeting materials are available on
FDA’s Web site (Ref. 2).
B. PSA Devices
FDA believes that the identified
special controls, in addition to general
controls, are necessary to provide
reasonable assurance of safety and
effectiveness of these devices.
Therefore, in accordance with sections
513(e) and 515(i) of the FD&C Act and
§ 860.130, based on new information
with respect to the device and taking
into account the public health benefit of
the use of the device and the nature and
known incidence of the risks of the
device, FDA, on its own initiative, is
proposing to reclassify these class III
devices into class II. The Agency has
identified special controls that would
provide reasonable assurance of their
safety and effectiveness.
Single and dual chamber PSA devices
combine the functions of a pacemaker
electrode function tester (class II) and an
EPPG device (proposed class II). No new
risks have been identified from the
combination of these devices and the
2013 Panel likewise did not identify
new concerns with regulating single and
dual chamber PSAs in a manner
consistent with EPPG devices. The low
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
frequency of serious adverse events as
evidenced through FDA’s MAUDE
database, the low rate of postmarket
recalls, the established scientific
evidence to support pacing for specific
indications, the hospital use
environment, and FDA’s review
experience with these devices, all
supports the reclassification of these
devices to class II. These devices are
prescription devices restricted to patient
use only upon the authorization of a
practitioner licensed by law to
administer or use the device.
Sufficient evidence has been
developed to support a reclassification
of single and dual chamber PSA
devices, to class II with special controls.
FDA has been reviewing these devices
for many years and their risks are well
known. The risks to health are
identified in section V, and FDA
believes these risks can be adequately
mitigated by general and special
controls.
Sufficient evidence has also been
developed to support a reclassification
of TCPSA devices. FDA has not
identified any additional risks to the
patient due to the availability to pace
three chambers in terms of failure to
pace or improper pacing rate during the
implant procedure. The longer-term
hemodynamic issues associated with
biventricular pacing are not relevant to
the acute implant procedure. The use of
TCPSAs is limited by labeling to use
only during implant of a pacemaker or
implantable cardioverter defibrillator
(ICD). Accordingly, the proposed special
controls for TCPSA devices contain the
same requirements as EPPG devices
with the addition of labeling that
indicates TCPSA use only during the
implant procedure.
FDA’s presentation to the 2013 Panel
included a summary of the available
safety and effectiveness information for
TCPSA devices, including adverse event
reports from FDA’s MAUDE database
and a search of the available literature.
The searches did not identify any safety
issues for this device type.
Based on the available evidence, FDA
recommended to the 2013 Panel that
TCPSA devices be reclassified to class II
(special controls). The 2013 Panel
discussed and made recommendations
regarding the regulatory classification of
TCPSA devices to either reconfirm to
class III (subject to premarket approval
application) or reclassify to class II
(subject to special controls) as directed
by section 513(e) of the FD&C Act. The
2013 Panel agreed with FDA’s
conclusion that the available scientific
evidence is adequate to support the
safety and effectiveness of TCPSA
devices and reclassify them to class II.
PO 00000
Frm 00016
Fmt 4702
Sfmt 4702
The 2013 Panel also acknowledged that
TCPSA devices are life-supporting
devices and provided the following
rationale per § 860.93 for recommending
that TCPSA devices be reclassified to
class II: (1) These devices are used only
during the implant procedure where
backup monitoring is continuous,
hazards can be recognized and treated
immediately, and where there is a
reasonable expectation that users are
adequately trained; (2) these devices are
not intended to provide the long-term
hemodynamic benefit of biventricular
pacing or cardiac resynchronization
therapy; and (3) the recommended
special controls will mitigate the health
risks associated with the device.
The 2013 Panel also agreed with the
identified risks to health presented at
the meeting; however, the 2013 Panel
recommended that FDA consider the
same modifications as recommended for
EPPG devices. FDA agrees with the 2013
Panel’s recommendations and modified
the risks to health accordingly as
outlined in section V. The 2013 Panel
also agreed with FDA’s proposed special
controls outlined in section VIII;
however, the 2013 Panel further
recommended that FDA add labeling
requirements for proper training, proper
maintenance of the device, and remedial
actions for failures due to lead
connections. FDA agrees with the 2013
Panel.
The 2013 Panel transcript and other
meeting materials are available on
FDA’s Web site (Ref. 2).
VIII. Proposed Special Controls
A. EPPG Devices
FDA believes that the following
special controls, together with general
controls (including applicable
prescription-use restrictions and
continuing 510(k) notification
requirements), are sufficient to mitigate
the risks to health described in section
V for EPPG devices:
1. Appropriate analysis/testing must
validate electromagnetic compatibility
(EMC) within a hospital environment.
2. Electrical bench testing must
demonstrate device safety during
intended use. This must include testing
with the specific power source (i.e.,
battery power, AC mains connections,
or both).
3. Non-clinical performance testing
data must demonstrate the performance
characteristics of the device. Testing
must include the following:
• Testing must demonstrate the
accuracy of monitoring functions,
alarms, measurement features,
therapeutic features, and all adjustable
E:\FR\FM\15SEP1.SGM
15SEP1
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
or programmable parameters as
identified in labeling;
• mechanical bench testing of
material strength must demonstrate that
the device and connection cables will
withstand forces or conditions
encountered during use;
• simulated use analysis/testing must
demonstrate adequate user interface for
adjustable parameters, performance of
alarms, display screens, interface with
external devices (e.g. data storage,
printing), and indicator(s) functionality
under intended use conditions; and
• methods and instructions for
cleaning the pulse generator and
connection cables must be validated.
4. Appropriate software verification,
validation, and hazard analysis must be
performed.
5. Labeling must include the
following:
• The labeling must clearly state that
these devices are intended for use in a
hospital environment and under the
supervision of a clinician trained in its
use;
• connector terminals should be
clearly, unambiguously marked on the
outside of the EPPG device. The
markings should identify positive (+)
and negative (¥) polarities. Dual
chamber devices should clearly identify
atrial and ventricular terminals;
54933
• the labeling must list all pacing
modes available in the device;
• labeling must include a detailed
description of any special capabilities
(e.g., overdrive pacing or automatic
mode switching); and
• appropriate electromagnetic
compatibility information must be
included.
Table 1 shows how FDA believes that
the risks to health identified in section
V can be mitigated by the proposed
special controls.
TABLE 1—HEALTH RISKS AND MITIGATION MEASURES FOR EPPG DEVICES
Identified risk
Mitigation measures
Failure to Pace ........................................
Improper High Rate Pacing .....................
Pacing at an Improperly Low Rate ..........
Improper Pacing Leading to Unwanted
Stimulation.
Micro/Macro Shock ..................................
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards
Labeling.
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards
Labeling.
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards
Labeling.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards
Labeling.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Labeling.
In addition, under § 801.109, the sale,
distribution, and use of EPPG devices
are restricted to prescription use.
Prescription use restrictions are a type
of general control in section
513(a)(1)(A)(i) of the FD&C Act. Also,
under § 807.81, the device would
continue to be subject to 510(k)
notification requirements.
rmajette on DSK2TPTVN1PROD with PROPOSALS
B. PSA Devices
FDA believes that the following
special controls, together with general
controls (including applicable
prescription-use restrictions and
continuing 510(k) notification
requirements), are sufficient to mitigate
the risks to health described in section
V for single, dual, and triple chamber
PSA devices:
1. Appropriate analysis/testing must
validate EMC within a hospital
environment.
VerDate Mar<15>2010
17:17 Sep 12, 2014
Jkt 232001
Analysis.
Analysis.
Analysis.
Analysis.
2. Electrical bench testing must
demonstrate device safety during
intended use. This must include testing
with the specific power source (i.e.,
battery power, AC mains connections,
or both).
3. Non-clinical performance testing
data must demonstrate the performance
characteristics of the device. Testing
must include the following:
• Testing must demonstrate the
accuracy of monitoring functions,
alarms, measurement features,
therapeutic features, and all adjustable
or programmable parameters as
identified in labeling;
• mechanical bench testing of
material strength must demonstrate that
the device and connection cables will
withstand forces or conditions
encountered during use;
• simulated use analysis/testing must
demonstrate adequate user interface for
PO 00000
Frm 00017
Fmt 4702
Sfmt 4702
adjustable parameters, performance of
alarms, display screens, interface with
external devices (e.g. data storage,
printing), and indicator(s) functionality
under intended use conditions; and
• methods and instructions for
cleaning the pulse generator and
connection cables must be validated.
4. Appropriate software verification,
validation, and hazard analysis must be
performed.
5. Labeling must include the
following:
• The labeling must clearly state that
these devices are intended for use in a
hospital environment and under the
supervision of a clinician trained in its
use;
• connector terminals should be
clearly, unambiguously marked on the
outside of the EPPG device. The
markings should identify positive (+)
and negative (¥) polarities. Dual
E:\FR\FM\15SEP1.SGM
15SEP1
54934
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
chamber devices should clearly identify
atrial and ventricular terminals. Triple
chamber devices should clearly identify
atrial, right ventricular, and left
ventricular terminals;
• the labeling must list all pacing
modes available in the device;
• labeling must include a detailed
description of any special capabilities
(e.g., overdrive pacing or automatic
mode switching);
• labeling must limit the use of
external pacing to the implant
procedure; and
• appropriate electromagnetic
compatibility information must be
included.
Table 2 shows how FDA believes that
the risks to health identified in section
V can be mitigated by the proposed
special controls.
TABLE 2—HEALTH RISKS AND MITIGATION MEASURES FOR PSA DEVICES
Identified risk
Mitigation measures
Misdiagnosis ............................................
Failure to Pace ........................................
Improper High Rate Pacing .....................
Pacing at an Improperly Low Rate ..........
Improper Pacing Leading to Unwanted
Stimulation.
Micro/Macro Shock ..................................
Non-Clinical Performance Evaluation.
Labeling.
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards Analysis.
Labeling.
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards Analysis.
Labeling.
Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards Analysis.
Labeling.
Non-Clinical Performance Evaluation.
Software Verification, Validation & Hazards Analysis.
Labeling.
Electrical Safety Testing.
Non-Clinical Performance Evaluation.
Labeling.
In addition, under § 801.109, the sale,
distribution, and use of these single and
dual chamber PSA devices are restricted
to prescription use. Prescription use
restrictions are a type of general controls
in section 513(a)(1)(A)(i) of the FD&C
Act. Also, under § 807.81, the device
would continue to be subject to 510(k)
notification requirements.
IX. Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
rmajette on DSK2TPTVN1PROD with PROPOSALS
X. Paperwork Reduction Act of 1995
This proposed order refers to
currently approved collections of
information found in FDA regulations.
These collections of information are
subject to review by the Office of
Management and Budget (OMB) under
the Paperwork Reduction Act of 1995
(44 U.S.C. 3501–3520). The collections
of information in part 807, subpart E,
VerDate Mar<15>2010
17:17 Sep 12, 2014
Jkt 232001
have been approved under OMB control
number 0910–0120; the collections of
information in 21 CFR part 814, subpart
B, have been approved under OMB
control number 0910–0231; and the
collections of information under 21 CFR
part 801 have been approved under
OMB control number 0910–0485.
XI. Codification of Orders
Prior to the amendments by FDASIA,
section 513(e) of the FD&C Act provided
for FDA to issue regulations to reclassify
devices. Although section 513(e) as
amended authorizes FDA to issue orders
rather than regulations, FDASIA also
provides for FDA to revoke previously
issued regulations by order. FDA will
continue to codify classifications and
reclassifications in the Code of Federal
Regulations (CFR). Changes resulting
from final orders will appear in the CFR
as changes to codified classification
determinations or as newly codified
orders. Therefore, under section
513(e)(1)(A)(i) of the FD&C Act, as
amended by FDASIA, in this proposed
order we are proposing to: (1) Revoke
the requirements in § 870.3600 related
to the classification of EPPG devices as
PO 00000
Frm 00018
Fmt 4702
Sfmt 4702
class III devices and to codify the
reclassification of EPPG devices into
class II (special controls) and (2) codify
the reclassification of PSA devices into
class II (special controls).
XII. Proposed Effective Date
FDA is proposing that any final order
based on this proposed order become
effective on the date of its publication
in the Federal Register or at a later date
if stated in the final order.
XIII. Comments
Comments already submitted to the
docket (FDA–2011–N–0650) have been
officially noted and do not need to be
resubmitted. FDA will consider
previous docket comments in issuing
any final orders for these devices.
Interested persons may submit either
electronic comments regarding this
document or the associated guidance to
https://www.regulations.gov or written
comments to the Division of Dockets
Management (see ADDRESSES). It is only
necessary to send one set of comments.
Identify comments with the docket
number found in brackets in the
heading of this document. Received
E:\FR\FM\15SEP1.SGM
15SEP1
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
comments may be seen in the Division
of Dockets Management between 9 a.m.
and 4 p.m., Monday through Friday, and
will be posted to the docket at https://
www.regulations.gov.
rmajette on DSK2TPTVN1PROD with PROPOSALS
XIV. References
The following references have been
placed on display in the Division of
Dockets Management (see ADDRESSES),
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and are available
electronically at https://
www.regulations.gov. (FDA has verified
all the Web site addresses in this
reference section, but we are not
responsible for any subsequent changes
to the Web sites after this document
publishes in the Federal Register.)
1. Class II Special Controls Draft Guidance
Document: External Pacemaker Pulse
Generator, available at https://
www.fda.gov/downloads/Medical
Devices/DeviceRegulationandGuidance/
GuidanceDocuments/UCM275703.pdf.
2. The panel transcript and other meeting
materials for the September 11, 2013,
Circulatory System Devices Panel are
available on FDA’s Web site at https://
www.fda.gov/AdvisoryCommittees/
CommitteesMeetingMaterials/Medical
Devices/MedicalDevicesAdvisory
Committee/CirculatorySystemDevices
Panel/ucm342357.htm.
3. ‘‘ACC/AHA Guidelines for the
Management of Patients with STElevation Myocardial Infarction,’’
Circulation, 110:e82-e292, 2004.
4. Beland, M. J., P. S. Hesslein, C. D. Finlay,
et al., ‘‘Noninvasive Transcutaneous
Cardiac Pacing in Children,’’ Pacing and
Clinical Electrophysiology, 10:1262–
1270, 1987.
5. Braun, M. U., T. Rauwolf, M. Bock, et al.,
‘‘Percutaneous Lead Implantation
Connected to an External Device in
Stimulation-Dependent Patients With
Systemic Infection—A Prospective and
Controlled Study,’’ Pacing and Clinical
Electrophysiology, August; 29(8):875–
879, 2006.
6. Ceresnak, S. R., R. H. Pass, T. J. Starc, et
al., ‘‘Predictors for Hemodynamic
Improvement With Temporary Pacing
After Pediatric Cardiac Surgery,’’ Journal
of Thoracic Cardiovascular Surgery,
January; 141(1):183–187, 2011.
7. Dunn, D. L. and J. J. Gregory, ‘‘Noninvasive
Temporary Pacing: Experience in a
Community Hospital,’’ Heart and Lung,
January; 18(1):23–28, 1989.
8. Eberhardt, F., T. Hanke, M. Heringlake, et
al., ‘‘Feasibility of Temporary
Biventricular Pacing in Patients With
Reduced Left Ventricular Function After
Coronary Artery Bypass Grafting,’’
Pacing and Clinical Electrophysiology,
January; 30 Suppl 1:S50–53, 2007.
9. Ferguson, T. B. Jr. and J. L. Cox,
‘‘Temporary External DDD Pacing After
Cardiac Operations,’’ The Annals of
Thoracic Surgery, May; 51(5):723–732,
1991.
10. Halldorsson, A. O., W. T. Vigneswaran,
F. J. Podbielski, and D. M. Evans,
VerDate Mar<15>2010
17:17 Sep 12, 2014
Jkt 232001
‘‘Electrophysiological and Clinical
Comparison of Two Temporary Pacing
Leads Following Cardiac Surgery,’’
Pacing and Clinical Electrophysiology,
August; 22(8):1221–1225, 1999.
11. Hindman, M. C., G. S. Wagner, M. JaRo,
et al., ‘‘The Clinical Significance of
Bundle Branch Block Complicating
Acute Myocardial Infarction. Indications
for Temporary and Permanent
Pacemaker Insertion,’’ Circulation,
October; 58(4):689–699, 1978.
12. Janousek J., P. Vojtovic, and R. A.
Gebauer, ‘‘Use of a Modified,
Commercially Available Temporary
Pacemaker for R Wave Synchronized
Atrial Pacing in Postoperative Junctional
Ectopic Tachycardia,’’ Pacing and
Clinical Electrophysiology, February;
26(2 Pt. 1):579–586, 2003.
13. Kratz, J. and J. Tyler, ‘‘Clinical Experience
with a New DDD External Pacemaker,’’
Pacing and Clinical Electrophysiology,
December; 16(12):2227–2234, 1993.
14. Nimetz, A. A., S. J. Shubrooks, A. M.
Hutter, and R. W. DeSanctis, ‘‘The
Significance of Bundle Branch Block
During Acute Myocardial Infarction,’’
American Heart Journal, October;
90(4):439–444, 1975.
15. Pinneri, F., S. Frea, K. Najd, et al.,
‘‘Echocardiography-Guided Versus
Fluoroscopy-Guided Temporary Pacing
in the Emergency Setting: An
Observational Study,’’ Journal of
Cardiovascular Medicine (Hagerstown),
March; 14(3):242–246, 2013.
16. Siddons, H., ‘‘Transvenous Long-Term
Pacing With an External Pacemaker.
What Are the Risks?’’, Pacing and
Clinical Electrophysiology, April;
1(2):163–165, 1978.
¨
¨
¨
17. Voigtlander T., B, Nowak, P. Barenfanger,
et al., ‘‘Feasibility and Sensing
Thresholds of Temporary Single-Lead
VDD Pacing in Intensive Care,’’
American Journal of Cardiology, May 15;
79(10):1360–1363, 1997.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 870 be amended as follows:
PART 870—CARDIOVASCULAR
DEVICES
1. The authority citation for 21 CFR
part 870 continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Section 870.3600 is revised to read
as follows:
■
§ 870.3600
generator.
External pacemaker pulse
(a) Identification. An external
pacemaker pulse generator (EPPG) is a
prescription device that has a power
supply and electronic circuits that
produce a periodic electrical pulse to
stimulate the heart. This device, which
PO 00000
Frm 00019
Fmt 4702
Sfmt 4702
54935
is used outside the body, is used as a
temporary substitute for the heart’s
intrinsic pacing system until a
permanent pacemaker can be implanted,
or to control irregular heartbeats in
patients following cardiac surgery or a
myocardial infarction. The device may
have adjustments for impulse strength,
duration, R-wave sensitivity, and other
pacing variables.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Appropriate analysis/testing must
validate electromagnetic compatibility
(EMC) within a hospital environment.
(2) Electrical bench testing must
demonstrate device safety during
intended use. This must include testing
with the specific power source (i.e.,
battery power, AC mains connections,
or both).
(3) Non-clinical performance testing
data must demonstrate the performance
characteristics of the device. Testing
must include the following:
(i) Testing must demonstrate the
accuracy of monitoring functions,
alarms, measurement features,
therapeutic features, and all adjustable
or programmable parameters as
identified in labeling;
(ii) Mechanical bench testing of
material strength must demonstrate that
the device and connection cables will
withstand forces or conditions
encountered during use;
(iii) Simulated use analysis/testing
must demonstrate adequate user
interface for adjustable parameters,
performance of alarms, display screens,
interface with external devices (e.g. data
storage, printing), and indicator(s)
functionality under intended use
conditions; and
(iv) Methods and instructions for
cleaning the pulse generator and
connection cables must be validated.
(4) Appropriate software verification,
validation, and hazard analysis must be
performed.
(5) Labeling must include the
following:
(i) The labeling must clearly state that
these devices are intended for use in a
hospital environment and under the
supervision of a clinician trained in
their use; and
(ii) Connector terminals should be
clearly, unambiguously marked on the
outside of the EPPG device. The
markings should identify positive (+)
and negative (-) polarities. Dual chamber
devices should clearly identify atrial
and ventricular terminals; and
(iii) The labeling must list all pacing
modes available in the device;
E:\FR\FM\15SEP1.SGM
15SEP1
54936
Federal Register / Vol. 79, No. 178 / Monday, September 15, 2014 / Proposed Rules
(iv) Labeling must include a detailed
description of any special capabilities
(e.g., overdrive pacing or automatic
mode switching); and
(v) Appropriate electromagnetic
compatibility information must be
included.
■ 3. In Subpart D, add § 870.3605 to
read as follows:
rmajette on DSK2TPTVN1PROD with PROPOSALS
§ 870.3605
Pacing system analyzer.
(a) Identification. A pacing system
analyzer (PSA) is a prescription device
that combines the functionality of a
pacemaker electrode function tester
(§ 870.3720) and an external pacemaker
pulse generator (EPPG) (§ 870.3600). It is
connected to a pacemaker lead and uses
a power supply and electronic circuits
to supply an accurately calibrated,
variable pacing pulse for measuring the
patient’s pacing threshold and
intracardiac R-wave potential. A PSA
may be a single, dual, or triple chamber
system and can simultaneously deliver
pacing therapy while testing one or
more implanted pacing leads.
(b) Classification. Class II (special
controls) for PSAs. The special controls
for this device are:
(1) Appropriate analysis/testing must
validate electromagnetic compatibility
(EMC) within a hospital environment.
(2) Electrical bench testing must
demonstrate device safety during
intended use. This must include testing
with the specific power source (i.e.,
battery power, AC mains connections,
or both).
(3) Non-clinical performance testing
data must demonstrate the performance
characteristics of the device. Testing
must include the following:
(i) Testing must demonstrate the
accuracy of monitoring functions,
alarms, measurement features,
therapeutic features, and all adjustable
or programmable parameters as
identified in labeling;
(ii) Mechanical bench testing of
material strength must demonstrate that
the device and connection cables will
withstand forces or conditions
encountered during use;
(iii) Simulated use analysis/testing
must demonstrate adequate user
interface for adjustable parameters,
performance of alarms, display screens,
interface with external devices (e.g. data
storage, printing), and indicator(s)
functionality under intended use
conditions; and
(iv) Methods and instructions for
cleaning the pulse generator and
connection cables must be validated.
(4) Appropriate software verification,
validation, and hazard analysis must be
performed.
(5) Labeling must include the
following:
VerDate Mar<15>2010
15:36 Sep 12, 2014
Jkt 232001
(i) The labeling must clearly state that
these devices are intended for use in a
hospital environment and under the
supervision of a clinician trained in
their use;
(ii) Connector terminals should be
clearly, unambiguously marked on the
outside of the EPPG. The markings
should identify positive (+) and negative
(-) polarities. Dual chamber devices
should clearly identify atrial and
ventricular terminals. Triple chamber
devices should clearly identify atrial,
right ventricular, and left ventricular
terminals;
(iii) The labeling must list all pacing
modes available in the device;
(iv) Labeling must include a detailed
description of any special capabilities
(e.g., overdrive pacing or automatic
mode switching);
(v) Labeling must limit the use of
external pacing to the implant
procedure; and
(vi) Appropriate electromagnetic
compatibility information must be
included.
Dated: September 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014–21814 Filed 9–12–14; 8:45 am]
BILLING CODE 4164–01–P
DEPARTMENT OF THE INTERIOR
Bureau of Indian Affairs
25 CFR Part 41
[145A2100DD.AADD001000.A0E501010.
999900]
RIN 1076–AF08
Grants to Tribally Controlled Colleges
´
and Universities, Dine College, and
Tribally Controlled Postsecondary
Career and Technical Institutions
Bureau of Indian Affairs,
Interior.
ACTION: Notice of tribal consultation
sessions.
AGENCY:
The Bureau of Indian
Education provides financial and
technical assistance to tribally
controlled colleges and universities and
´
Dine College. In collaboration with the
American Indian Higher Education
Consortium, we have prepared a
discussion draft that updates the
policies and procedures for
administration and oversight of these
assistance programs and revises
regulatory language to conform to
statutory amendments. This notice
announces tribal consultation sessions
SUMMARY:
PO 00000
Frm 00020
Fmt 4702
Sfmt 4702
and a comment period on the
preliminary discussion draft.
DATES: Comments must be received by
November 15, 2014. See the
SUPPLEMENTARY INFORMATION section of
this notice for dates of the tribal
consultation sessions.
ADDRESSES: See the SUPPLEMENTARY
INFORMATION section of this notice for
locations of the tribal consultation
sessions and the Web site where the
preliminary discussion draft is
available. You may submit comments by
either of the following methods:
—Federal Rulemaking Portal: https://
www.regulations.gov. This rule is listed
under the agency name ‘‘Bureau of
Indian Affairs’’ and Docket ID ‘‘BIA–
2011–0002.’’
—Mail or Hand-Delivery: Ms. Juanita
Mendoza, Program Analyst, Bureau of
Indian Education, U.S. Department of
the Interior, 1951 Constitution Ave.
NW., MS 312, Washington, DC 20240.
Include ‘‘1076–AF08’’ on the cover of
the submission.
FOR FURTHER INFORMATION CONTACT: Ms.
Juanita Mendoza, Program Analyst,
Bureau of Indian Education, U.S.
Department of the Interior, 1951
Constitution Ave. NW., MS 312,
Washington, DC 20240; or email to
juanita.mendoza@bie.edu.
SUPPLEMENTARY INFORMATION: The BIE
supports and encourages the
establishment, operation, and
improvement of tribally controlled
colleges and universities (TCUs) to
ensure continued and expanded
educational opportunities for Indian
students. The TCUs are both integral
and essential to their communities,
creating environments that foster
American Indian culture, languages, and
traditions. The TCUs serve a variety of
people from young adults to senior
citizens. The TCUs offer 358 total
programs, including apprenticeships,
diplomas, certificates, and degrees.
These programs include 181 associate
degree programs at 23 TCUs, 40
bachelor’s degree programs at 11 TCUs,
and 5 master’s degree programs at 2
TCUs.
The BIE is revising the regulations at
25 CFR Part 41 and has prepared a
preliminary discussion draft. Subpart B
of the preliminary discussion draft
concerns financial and technical
assistance to tribal colleges and
universities funded under the Tribally
Controlled Colleges and Universities
Assistance Act of 1978, as amended (25
U.S.C. 1801 et seq.). Subpart B does not
´
concern financial assistance to Dine
College or to tribally controlled
postsecondary career and technical
institutions. Subpart C of the
E:\FR\FM\15SEP1.SGM
15SEP1
Agencies
[Federal Register Volume 79, Number 178 (Monday, September 15, 2014)]
[Proposed Rules]
[Pages 54927-54936]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2014-21814]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 870
[Docket No. FDA-2011-N-0650]
Cardiovascular Devices; Reclassification of External Pacemaker
Pulse Generator Devices; Reclassification of Pacing System Analyzers
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is proposing in this
administrative order to reclassify the external pacemaker pulse
generator (EPPG) devices, a preamendments class III device into class
II (special controls), and to amend the device identification and
reclassify the pacing system analyzers (PSAs) into class II (special
controls). Specifically, single and dual chamber PSAs, which are
currently classified with EPPG devices, and triple chamber PSAs
(TCPSAs), which are postamendments class III devices, are proposed to
be reclassified to class II devices. FDA is proposing this
reclassification based on new information pertaining to the device.
This proposed action would implement certain statutory requirements.
DATES: Submit either electronic or written comments on the proposed
order by December 15, 2014. See section XII for the effective date of
any final order that may publish based on this proposed order.
ADDRESSES: You may submit comments, identified by Docket No. FDA-2011-
N-0650, by any of the following methods:
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments.
[[Page 54928]]
Written Submissions
Submit written submissions in the following ways:
Mail/Hand delivery/Courier (for paper submissions):
Division of Dockets Management (HFA-305), Food and Drug Administration,
5630 Fishers Lane, rm. 1061, Rockville, MD 20852.
Instructions: All submissions received must include the Agency name
and Docket No. FDA-2011-N-0650 for this order. All comments received
may be posted without change to https://www.regulations.gov, including
any personal information provided. For additional information on
submitting comments, see the ``Comments'' heading of the SUPPLEMENTARY
INFORMATION section of this document.
Docket: For access to the docket to read background documents or
comments received, go to https://www.regulations.gov and insert the
docket number, found in brackets in the heading of this document, into
the ``Search'' box and follow the prompts and/or go to the Division of
Dockets Management, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Hina Pinto, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, rm. 1652, Silver Spring, MD 20993, 301-796-6351.
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Food, Drug, and Cosmetic Act (the FD&C Act) as amended by the
Medical Device Amendments of 1976 (the 1976 amendments) (Pub. L. 94-
295), the Safe Medical Devices Act of 1990 (Pub. L. 101-629), the Food
and Drug Administration Modernization Act of 1997 (Pub. L. 105-115),
the Medical Device User Fee and Modernization Act of 2002 (Pub. L. 107-
250), the Medical Devices Technical Corrections Act of 2004 (Public Law
108-214), the Food and Drug Administration Amendments Act of 2007 (Pub.
L. 110-85), and the Food and Drug Administration Safety and Innovation
Act (FDASIA) (Pub. L. 112-144) establishes a comprehensive system for
the regulation of medical devices intended for human use. Section 513
of the FD&C Act (21 U.S.C. 360c) established three categories (classes)
of devices, reflecting the regulatory controls needed to provide
reasonable assurance of their safety and effectiveness. The three
categories of devices are class I (general controls), class II (special
controls), and class III (premarket approval).
Section 513(a)(1) of the FD&C Act defines class II devices as those
devices for which the general controls by themselves are insufficient
to provide reasonable assurance of safety and effectiveness, but for
which there is sufficient information to establish special controls to
provide such assurance.
Under section 513 of the FD&C Act, devices that were in commercial
distribution before the enactment of the 1976 amendments, May 28, 1976
(generally referred to as preamendments devices), are classified after
FDA has: (1) Received a recommendation from a device classification
panel (an FDA advisory committee); (2) published the panel's
recommendation for comment, along with a proposed regulation
classifying the device; and (3) published a final regulation
classifying the device. FDA has classified most preamendments devices
under these procedures.
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices), are
automatically classified by section 513(f) of the FD&C Act into class
III without any FDA rulemaking process. Those devices remain in class
III and require premarket approval unless, and until, the device is
reclassified into class I or II or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and part 807 (21 CFR part 807).
A preamendments device that has been classified into class III and
devices found substantially equivalent by means of premarket
notification (510(k)) procedures to such a preamendments device or to a
device within that type may be marketed without submission of a
premarket approval (PMA) application until FDA issues a final order
under section 515(b) of the FD&C Act (21 U.S.C. 360e(b)) requiring
premarket approval or until the device is subsequently reclassified
into class I or class II.
On July 9, 2012, FDASIA was enacted. Section 608(a) of FDASIA
amended the device reclassification procedures under section 513(e) of
the FD&C Act, changing the process for reclassifying a device from
rulemaking to an administrative order. Prior to the enactment of
FDASIA, FDA published a proposed rule under section 513(e) of the FD&C
Act proposing the reclassification of EPPG devices (76 FR 64223,
October 17, 2011). Three sets of comments were received on the proposed
rule. The three sets of comments submitted in response to the proposed
rule on EPPG devices will be considered under this proposed
administrative order and do not need to be resubmitted. FDA is issuing
this proposed administrative order to comply with the new procedural
requirement created by FDASIA when reclassifying a preamendments class
III device, as well as to reclassify a postamendments class III device.
Also, as required by section 513(e) of the FD&C Act for preamendment
devices, FDA convened a device classification panel meeting which
discussed the proposed reclassification on September 11, 2013 (78 FR
49272). This action is intended solely to fulfill the procedural
requirements for reclassification implemented by FDASIA.
Section 513(e) of the FD&C Act provides that FDA may, by
administrative order, reclassify a device based upon ``new
information.'' FDA can initiate a reclassification under section 513(e)
or an interested person may petition FDA to reclassify a preamendments
device. The term ``new information,'' as used in section 513(e) of the
FD&C Act, includes information developed as a result of a reevaluation
of the data before the Agency when the device was originally
classified, as well as information not presented, not available, or not
developed at that time. (See, e.g., Holland-Rantos Co. v. United States
Dep't of Health, Educ., & Welfare, 587 F.2d 1173, 1174 n.1 (D.C. Cir.
1978); Upjohn v. Finch, 422 F.2d 944 (6th Cir. 1970); Bell v. Goddard,
366 F.2d 177 (7th Cir. 1966).)
A postamendments device that has been initially classified in class
III under section 513(f)(1) of the FD&C Act may be reclassified later
into class I or class II under section 513(f)(3) of the FD&C Act.
Section 513(f)(3) provides that FDA acting by administrative order can
reclassify the device into class I or class II on its own initiative
under section 513(f)(1) of the FD&C Act, or in response to the petition
of the manufacturer or importer of the device. FDA's regulations in 21
CFR 860.134 set forth the procedures for the filing and review of a
petition for reclassification of these class III devices. To change the
classification of the device, the proposed new class must have
sufficient regulatory controls to provide reasonable assurance of the
safety and effectiveness of the device for its intended use.
Reevaluation of the data previously before the Agency is an
appropriate
[[Page 54929]]
basis for subsequent regulatory action where the reevaluation is made
in light of newly available regulatory authority (see Bell v. Goddard,
supra, 366 F.2d at 181; Ethicon, Inc. v. FDA, 762 F.Supp. 382, 388-391
(D.D.C. 1991)), or in light of changes in ``medical science.'' (See
Upjohn v. Flinch supra, 422 F.2d at 951.) Whether data before the
Agency are old or new data, the ``new information'' to support
reclassification under section 513(e) must be ``valid scientific
evidence,'' as defined in section 513(a)(3) of the FD&C Act and Sec.
860.7(c)(2). (See, e.g., General Medical Co. v. FDA, 770 F.2d 214 (D.C.
Cir. 1985); Contact Lens Manufacturers Association v. FDA, 766 F.2d 592
(D.C. Cir. 1985), cert. denied, 474 U.S. 1062 (1986).
FDA relies upon ``valid scientific evidence'' in the classification
process to determine the level of regulation for devices. To be
considered in the reclassification process, the valid scientific
evidence upon which the Agency relies must be publicly available.
Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
premarket approval application (PMA). (See section 520(c) of the FD&C
Act (21 U.S.C. 360j(c)).
Section 513(e)(1) of the FD&C Act sets forth the process for
issuing a final order. Specifically, prior to the issuance of a final
order reclassifying a device, the following must occur: (1) Publication
of a proposed order in the Federal Register; (2) a meeting of a device
classification panel described in section 513(b) of the FD&C Act; and
(3) consideration of comments from all affected stakeholders, including
patients, payors, and providers. In addition, the proposed order must
set forth the proposed reclassification, and a substantive summary of
the valid scientific evidence concerning the proposed reclassification,
including the public health benefits of the use of the device, and the
nature and incidence (if known) of the risk of the device. (See section
513(e)(1)(A)(i) of the FD&C Act.)
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the premarket notification requirements under section
510(k) of the FD&C Act if the Agency determines that premarket
notification is not necessary to assure the safety and effectiveness of
the device. FDA has determined that premarket notification is necessary
to reasonably assure the safety and effectiveness of EPPG and PSA
devices.
II. Regulatory History of the Device
A. EPPG Devices
On March 9, 1979, FDA published a proposed rule in the Federal
Register for classification of EPPG devices into class III based on the
recommendation of the Cardiovascular Devices Panel (the Panel) (44 FR
13284 at 13372). The Panel meeting recommended EPPG devices be
classified into class III because the device provided temporary life
support and that certain kinds of failures could cause this device to
emit inappropriate electrical signals, which could cause cardiac
irregularities and death. The Panel indicated that general controls
alone would not be sufficient and that there was not enough information
to establish a performance standard. Consequently, the Panel believed
that premarket approval was necessary to reasonably assure the safety
and effectiveness of the device. In 1980, FDA classified EPPG into
class III under Sec. 870.3600 (21 CFR 870.3600) after receiving no
comments on the proposed rule (45 FR 7907, February 5, 1980). In 1987,
FDA published a clarification by inserting language in the codified
language stating that no effective date had been established for the
requirement for premarket approval for EPPG devices (52 FR 17732, May
11, 1987).
In 2009, FDA published an order in the Federal Register under
section 515(i) of the FD&C Act to call for information on the remaining
class III 510(k) preamendment devices, including EPPG devices (74 FR
16214, April 9, 2009). In response to that order, FDA received two
reclassification petitions from one device manufacturer who requested
that EPPG devices be reclassified into class II. The manufacturers
stated that safety and effectiveness of these devices may be assured by
performance standards, the intended use environment, postmarket
surveillance to include Medical Device Reporting (MDRs), FDA
inspections of manufacturing facilities, and premarket review of
performance testing in a 510(k) submission. The manufacturers
specifically noted that the FDA recognized consensus standard,
International Electrotechnical Commission (IEC) 60601-2-31: 'Particular
requirements for the basic safety and essential performance of external
cardiac pacemakers with internal power source' provides adequate design
and testing parameters for EPPG devices.
On October 17, 2011, FDA published a proposed rule proposing the
reclassification of EPPG devices from class III to class II (76 FR
64223) and announcing the availability of a draft Special Controls
Guidance Document that, if finalized, would serve as a special control,
if FDA reclassified these devices. FDA believed that the special
controls described in the draft special controls guidance document
entitled ``Class II Special Controls Guidance Document: External
Pacemaker Pulse Generator'' would be sufficient to mitigate the risks
to health associated with EPPG (Ref. 1).
The proposed rule provided for a comment period that was open until
January 17, 2012. FDA received three sets of comments. These comments
stated that: (1) FDA should retain EPPG in class III, (2) FDA's
reclassification proposed rule was not adequately supported by new
publicly valid scientific evidence, (3) MDR data showed that existing
performance standards are insufficient, (4) there were no publicly
available performance standards that would apply to EPPG, (5) FDA
should convene an advisory committee (the Panel) to seek a
recommendation on the classification of EPPG, and (6) the recall
process after reclassification of EPPG would need to be clarified.
These comments were considered by FDA in drafting this proposed order.
B. PSA Devices
Single and dual chamber PSAs have historically been classified with
EPPG devices. These devices combine the functionality of a single or
dual chamber EPPG, which is currently class III and the functionality
of a pacemaker electrode function tester, which is regulated as a class
II device (under Sec. 870.3720 (21 CFR 870.3720)). Single and dual
chamber PSA devices have been found substantially equivalent to EPPG
devices through the 510(k) process. Triple chamber PSA (TCPSA) devices
have not been determined to be substantially equivalent through the
510(k) process, and since this technology was not on the market in
1976, TCPSAs have been reviewed through the PMA process as
postamendment class III devices.
On July 9, 2012, FDASIA was enacted, which amended the device
reclassification procedures under sections 513 and 515 of the FD&C Act.
Accordingly, FDA is issuing a proposed administrative order to comply
with the new procedural requirement created by FDASIA when
reclassifying a preamendments class III device. Further, FDA intends to
codify the proposed special controls within the Sec. 870.3600
classification regulation for EPPG and to create a separate
[[Page 54930]]
classification regulation for PSA devices.
As explained further in section VII, a meeting of the Circulatory
System Devices Panel (the 2013 Panel) took place on September 11, 2013,
to discuss whether EPPG and TCPSA devices should be reclassified or
remain in class III (Ref. 2). FDA included a discussion of TCPSA
devices in the 2013 Panel discussion because the risks to health and
proposed special controls were very similar to those being proposed for
the EPPG devices already under consideration. The 2013 Panel
recommended that EPPG devices be reclassified to class II with special
controls when intended for cardiac rate control or prophylactic
arrhythmia prevention. The 2013 Panel also recommended that TCPSA
devices be reclassified to class II with special controls when intended
for use during the pulse generator implant procedure. FDA is not aware
of new information that would provide a basis for a different
recommendation or finding.
III. Device Description
A. EPPG Devices
An EPPG is a device that has a power supply and electronic circuits
that produce a periodic electrical pulse to stimulate the heart. This
device, which is used outside the body, is used as a temporary
substitute for the heart's intrinsic pacing system until a permanent
pacemaker can be implanted, or to control irregular heartbeats in
patients following events such as cardiac surgery or a myocardial
infarction. The device may have adjustments for pacing rate, pulse
amplitude, pulse width (duration), R-wave sensitivity, and other pacing
variables.
An EPPG device is designed to be used with cardiac pacing lead
systems for temporary atrial and/or ventricular pacing. An EPPG system
generally includes the pulse generator, extension cables, and adaptors
which connect the extension cable to the implanted pacing lead and are
critical to the functionality of the EPPG system. The pacing leads for
use with EPPG may be for temporary or permanent use for either
tranvenous or epicardial uses. The pacing leads are not considered part
of the EPPG device because they have their own regulatory designations
(21 CFR 870.3680) depending on their design and intended use.
EPPG devices are used exclusively in hospital environments with the
patients supervised by qualified medical personnel. The electrical and
heart rhythm of patients are continuously monitored using EPPG-
independent electrocardiogram (ECG) monitors usually with alarm
functions. Independent ECG monitoring requirements are identified in
international standards, such as IEC 60601-2-31 for device design.
FDA is also proposing in this order to slightly modify the
identification language from the way it is presently written in Sec.
870.3600(a) to clarify that these are prescription devices in
accordance with Sec. 801.109 (21 CFR 801.109).
B. PSA Devices
A PSA combines the functionality of a pacemaker electrode function
tester (under Sec. 870.3720) and an EPPG. A pacemaker electrode
function tester is a device that is connected to an implanted pacemaker
lead that supplies an accurately calibrated, variable pacing pulse for
measuring the patient's pacing threshold and intracardiac R-wave
potential. A PSA can temporarily take over pacing functions while
simultaneously testing one or more implanted pacing leads. PSA devices
can be single, dual, or triple chamber, translating into the
measurement capabilities/functionalities of the device. Single chamber
PSAs typically measure pacing capture threshold, whereas in the case of
dual chamber PSAs, the device can also measure conduction times or
intrinsic atrioventricular delay. In the case of a TCPSA, the device
can also function as a biventricular external pacemaker and measure the
intrinsic intra-ventricular (VV) interval.
IV. Proposed Reclassification
A. EPPG Devices
FDA is proposing that EPPG devices be reclassified from class III
to class II. In this proposed order, the Agency has identified special
controls under section 513(a)(1)(B) of the FD&C Act that, together with
general controls (including prescription-use restrictions) applicable
to the devices, would provide reasonable assurance of their safety and
effectiveness. Absent the special controls identified in this proposed
order, general controls applicable to the device are insufficient to
provide reasonable assurance of the safety and effectiveness of the
device. Since the time of the 1979 classification, new information
about use and pacing technology for this device has become sufficiently
available to establish special controls. FDA believes that this new
information is sufficient to demonstrate that the proposed special
controls, when finalized, can effectively mitigate the risks to health
identified in section V, and that these special controls, together with
general controls, will provide a reasonable assurance of safety and
effectiveness for EPPG devices.
B. PSA Devices
FDA is proposing to create a separate classification regulation for
PSA devices, including single, dual, and triple chamber PSA devices
that will be reclassified from class III to class II. In this proposed
order, the Agency has identified special controls under section
513(a)(1)(B) of the FD&C Act that, together with general controls
(including prescription-use restrictions) applicable to the devices,
would provide reasonable assurance of their safety and effectiveness.
Absent the special controls identified in this proposed order, general
controls applicable to the device are insufficient to provide
reasonable assurance of the safety and effectiveness of the device.
Since the 1979 classification of temporary external pacing devices, new
information about device use and pacing technology has become available
to establish special controls. FDA believes that this new information
is sufficient to demonstrate that the proposed special controls can
effectively mitigate the risks to health identified in section V, and
that these special controls, together with general controls, will
provide a reasonable assurance of safety and effectiveness for PSA
devices.
Section 510(m) of the FD&C Act authorizes the Agency to exempt
class II devices from premarket notification (510(k)) submission. FDA
has considered EPPG and PSA devices in accordance with the reserved
criteria set forth in section 513(a) and determined that both devices
require premarket notification (510(k) of the FD&C Act). Therefore, the
Agency does not intend to exempt these proposed class II devices from
premarket notification (510(k)) submission as provided under section
510(m) of the FD&C Act.
V. Risks to Health
A. EPPG Devices
After considering available information for the classification of
these devices, including the recommendations of the advisory committees
(panels) for the classification of these devices, FDA has evaluated the
risks to health associated with the use of EPPG devices and determined
the following risks to health are associated with its use:
Failure to pace: Improper settings, electromagnetic
interference, or failure
[[Page 54931]]
of mechanical/electrical components of the device can prevent pacing of
the patient's heart.
Improperly high pacing rate: Undersensing during demand
pacing, unintended asynchronous pacing, or improper use of burst/
overdrive pacing can cause harmful acceleration of heart rate or induce
harmful arrhythmias such as ventricular tachycardia.
Improperly low pacing rate: Oversensing or use error can
result in stimulation pulses being delivered at an unwanted low pacing
rate, which can result in untreated symptomatic bradycardia.
Improper pacing leading to unwanted stimulation: Pacing
during vulnerable periods of the cardiac cycle or at higher than
programmed amplitude can induce arrhythmias.
Micro/macro shock: Uncontrolled leakage currents or
patient auxiliary currents can cause an electric shock resulting in an
arrhythmia or cardiac tissue damage.
B. PSA Devices
After considering available information for the classification of
these devices, including the recommendations of the advisory committees
(panels) for the classification of these devices, FDA has evaluated the
risks to health associated with the use of PSA devices and determined
the following risks to health are associated with its use:
Failure to pace: Improper settings, electromagnetic
interference, or failure of mechanical/electrical components of the
device can prevent pacing of the patient's heart.
Improperly high pacing rate: Undersensing during demand
pacing, unintended asynchronous pacing, or improper use of burst/
overdrive pacing can cause harmful acceleration of heart rate or induce
harmful arrhythmias such as ventricular tachycardia.
Improperly low pacing rate: Oversensing or use error can
result in stimulation pulses being delivered at an unwanted low pacing
rate, which can result in untreated symptomatic bradycardia.
Improper pacing leading to unwanted stimulation: Pacing
during vulnerable periods of the cardiac cycle or at higher than
programmed amplitude can induce arrhythmias. For TCPSAs, this risk
includes VV dyssynchrony.
Micro/macro shock: Uncontrolled leakage currents or
patient auxiliary currents can cause an electric shock resulting in an
arrhythmia or cardiac tissue damage.
Misdiagnosis: If the zero or calibration of the device is
inaccurate or unstable the device may generate inaccurate diagnostic
data. If inaccurate diagnostic data are used in managing the patient,
the physician may prescribe a course of treatment that places the
patient at risk unnecessarily.
VI. Summary of Reasons for Reclassification
FDA believes that EPPG devices and PSA devices should be
reclassified from class III to class II because special controls, in
addition to general controls, can be established to provide reasonable
assurance of the safety and effectiveness of the devices, and because
general controls themselves are insufficient to provide reasonable
assurance of their safety and effectiveness. In addition, there is now
sufficient information to establish special controls to provide such
assurance. FDA also believes that TCPSA devices--as a subset of PSA
devices--can achieve a reasonable assurance of safety and effectiveness
using the same special controls proposed for EPPG and PSA devices with
the addition of limiting use to the duration of the implant procedure
in order to mitigate the risk of unwanted interventricular stimulation
leading to arrhythmia (captured as misdiagnosis and improper pacing
leading to unwanted stimulation in the list of risks to health in
section V).
VII. Summary of Data Upon Which the Reclassification Is Based
A. EPPG Devices
FDA believes that the identified special controls, in addition to
general controls, are necessary to provide reasonable assurance of
safety and effectiveness of these devices. Therefore, in accordance
with sections 513(e) and 515(i) of the FD&C Act and Sec. 860.130,
based on new information with respect to the device and taking into
account the public health benefit of the use of the device and the
nature and known incidence of the risk of the device, FDA, on its own
initiative, is proposing to reclassify this preamendments class III
device into class II. The Agency has identified special controls that
would provide reasonable assurance of their safety and effectiveness.
EPPG are prescription devices restricted to patient use only upon the
authorization of a practitioner licensed by law to administer or use
the device. Since 1979 when FDA classified EPPG devices into class III,
sufficient evidence has been developed to support a reclassification to
class II with the establishment of special controls. FDA has been
reviewing these devices for many years and their risks are well known.
The risks to health are identified in section V, and FDA believes these
risks can be adequately mitigated by special controls.
EPPG devices that use temporary cardiac pacing for the purposes of
rate control or treatment of bradycardia use mature technology with
well-established evidence of effectiveness (Refs. 3, 8, 9, 11, 13). A
review of 14 clinical studies published over four decades shows that
temporary external pacing is generally safe and has an
electrophysiologic as well as hemodynamic benefit when used as
indicated (Refs. 3 to 17).
The low frequency of serious adverse events as evidenced through
FDA's Manufacturer and User Facility Device Experience (MAUDE)
database, the low rate of postmarket recalls, the established
scientific evidence to support pacing for specific indications, the
hospital use environment, and FDA's review experience with these
devices, all support the reclassification of these devices to class II.
In addition, several key performance standards (such as IEC 60601-1 and
IEC 60601-2-31) that address various aspects of design and performance
have been developed and used to support marketing applications since
the original classification. In light of these considerations, FDA has
tentatively concluded that the identified special controls, in addition
to general controls, provide reasonable assurance of the safety and
effectiveness of EPPG devices.
FDA's presentation to the 2013 Panel included a summary of the
available safety and effectiveness information for EPPG devices,
including adverse event reports from FDA's MAUDE database and available
literature. Based on the available scientific literature, which
supports that use of EPPG devices may be beneficial for patients
needing temporary atrial and/or ventricular pacing, FDA recommended to
the 2013 Panel that EPPG devices be reclassified to class II (special
controls). The 2013 Panel discussed and made recommendations regarding
the regulatory classification of EPPG devices to either reconfirm to
class III (subject to premarket approval application) or reclassify to
class II (subject to special controls). The 2013 Panel agreed with
FDA's conclusion that the available scientific evidence is adequate to
support the safety and effectiveness of EPPG devices. The 2013 Panel
also acknowledged that EPPG devices are life-supporting devices and
provided the following rationale per
[[Page 54932]]
Sec. 860.93 for recommending that EPPG devices be reclassified to
class II: (1) These devices are used exclusively in the hospital
environment where backup monitoring is available, hazards can be
recognized and treated immediately, and where there is a reasonable
expectation that users are adequately trained; (2) there is sufficient
clinical experience that attests to the benefit of the device; and (3)
the recommended special controls will mitigate the health risks
associated with the device.
The 2013 Panel also agreed with the identified risks to health
presented at the meeting; however, it recommended that FDA consider
rewording some of the language for clarity and also to ensure that
certain hazards, such as asynchronous pacing and arrhythmia induction,
are included in the risks to health. FDA agrees with the 2013 Panel's
recommendations and modified the risks to health accordingly as
outlined in section V. The 2013 Panel also agreed with FDA's proposed
special controls outlined in section VIII; however, the 2013 Panel
further recommended that FDA add labeling requirements for proper
training, proper maintenance of the device, and remedial actions for
failures due to lead connections. FDA agrees with the 2013 Panel and
the proposed special controls have been modified to reflect more
specific labeling requirements.
The 2013 Panel transcript and other meeting materials are available
on FDA's Web site (Ref. 2).
B. PSA Devices
FDA believes that the identified special controls, in addition to
general controls, are necessary to provide reasonable assurance of
safety and effectiveness of these devices. Therefore, in accordance
with sections 513(e) and 515(i) of the FD&C Act and Sec. 860.130,
based on new information with respect to the device and taking into
account the public health benefit of the use of the device and the
nature and known incidence of the risks of the device, FDA, on its own
initiative, is proposing to reclassify these class III devices into
class II. The Agency has identified special controls that would provide
reasonable assurance of their safety and effectiveness.
Single and dual chamber PSA devices combine the functions of a
pacemaker electrode function tester (class II) and an EPPG device
(proposed class II). No new risks have been identified from the
combination of these devices and the 2013 Panel likewise did not
identify new concerns with regulating single and dual chamber PSAs in a
manner consistent with EPPG devices. The low frequency of serious
adverse events as evidenced through FDA's MAUDE database, the low rate
of postmarket recalls, the established scientific evidence to support
pacing for specific indications, the hospital use environment, and
FDA's review experience with these devices, all supports the
reclassification of these devices to class II. These devices are
prescription devices restricted to patient use only upon the
authorization of a practitioner licensed by law to administer or use
the device.
Sufficient evidence has been developed to support a
reclassification of single and dual chamber PSA devices, to class II
with special controls. FDA has been reviewing these devices for many
years and their risks are well known. The risks to health are
identified in section V, and FDA believes these risks can be adequately
mitigated by general and special controls.
Sufficient evidence has also been developed to support a
reclassification of TCPSA devices. FDA has not identified any
additional risks to the patient due to the availability to pace three
chambers in terms of failure to pace or improper pacing rate during the
implant procedure. The longer-term hemodynamic issues associated with
biventricular pacing are not relevant to the acute implant procedure.
The use of TCPSAs is limited by labeling to use only during implant of
a pacemaker or implantable cardioverter defibrillator (ICD).
Accordingly, the proposed special controls for TCPSA devices contain
the same requirements as EPPG devices with the addition of labeling
that indicates TCPSA use only during the implant procedure.
FDA's presentation to the 2013 Panel included a summary of the
available safety and effectiveness information for TCPSA devices,
including adverse event reports from FDA's MAUDE database and a search
of the available literature. The searches did not identify any safety
issues for this device type.
Based on the available evidence, FDA recommended to the 2013 Panel
that TCPSA devices be reclassified to class II (special controls). The
2013 Panel discussed and made recommendations regarding the regulatory
classification of TCPSA devices to either reconfirm to class III
(subject to premarket approval application) or reclassify to class II
(subject to special controls) as directed by section 513(e) of the FD&C
Act. The 2013 Panel agreed with FDA's conclusion that the available
scientific evidence is adequate to support the safety and effectiveness
of TCPSA devices and reclassify them to class II. The 2013 Panel also
acknowledged that TCPSA devices are life-supporting devices and
provided the following rationale per Sec. 860.93 for recommending that
TCPSA devices be reclassified to class II: (1) These devices are used
only during the implant procedure where backup monitoring is
continuous, hazards can be recognized and treated immediately, and
where there is a reasonable expectation that users are adequately
trained; (2) these devices are not intended to provide the long-term
hemodynamic benefit of biventricular pacing or cardiac
resynchronization therapy; and (3) the recommended special controls
will mitigate the health risks associated with the device.
The 2013 Panel also agreed with the identified risks to health
presented at the meeting; however, the 2013 Panel recommended that FDA
consider the same modifications as recommended for EPPG devices. FDA
agrees with the 2013 Panel's recommendations and modified the risks to
health accordingly as outlined in section V. The 2013 Panel also agreed
with FDA's proposed special controls outlined in section VIII; however,
the 2013 Panel further recommended that FDA add labeling requirements
for proper training, proper maintenance of the device, and remedial
actions for failures due to lead connections. FDA agrees with the 2013
Panel.
The 2013 Panel transcript and other meeting materials are available
on FDA's Web site (Ref. 2).
VIII. Proposed Special Controls
A. EPPG Devices
FDA believes that the following special controls, together with
general controls (including applicable prescription-use restrictions
and continuing 510(k) notification requirements), are sufficient to
mitigate the risks to health described in section V for EPPG devices:
1. Appropriate analysis/testing must validate electromagnetic
compatibility (EMC) within a hospital environment.
2. Electrical bench testing must demonstrate device safety during
intended use. This must include testing with the specific power source
(i.e., battery power, AC mains connections, or both).
3. Non-clinical performance testing data must demonstrate the
performance characteristics of the device. Testing must include the
following:
Testing must demonstrate the accuracy of monitoring
functions, alarms, measurement features, therapeutic features, and all
adjustable
[[Page 54933]]
or programmable parameters as identified in labeling;
mechanical bench testing of material strength must
demonstrate that the device and connection cables will withstand forces
or conditions encountered during use;
simulated use analysis/testing must demonstrate adequate
user interface for adjustable parameters, performance of alarms,
display screens, interface with external devices (e.g. data storage,
printing), and indicator(s) functionality under intended use
conditions; and
methods and instructions for cleaning the pulse generator
and connection cables must be validated.
4. Appropriate software verification, validation, and hazard
analysis must be performed.
5. Labeling must include the following:
The labeling must clearly state that these devices are
intended for use in a hospital environment and under the supervision of
a clinician trained in its use;
connector terminals should be clearly, unambiguously
marked on the outside of the EPPG device. The markings should identify
positive (+) and negative (-) polarities. Dual chamber devices should
clearly identify atrial and ventricular terminals;
the labeling must list all pacing modes available in the
device;
labeling must include a detailed description of any
special capabilities (e.g., overdrive pacing or automatic mode
switching); and
appropriate electromagnetic compatibility information must
be included.
Table 1 shows how FDA believes that the risks to health identified
in section V can be mitigated by the proposed special controls.
Table 1--Health Risks and Mitigation Measures for EPPG Devices
------------------------------------------------------------------------
Identified risk Mitigation measures
------------------------------------------------------------------------
Failure to Pace...................... Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Improper High Rate Pacing............ Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Pacing at an Improperly Low Rate..... Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Improper Pacing Leading to Unwanted Non-Clinical Performance
Stimulation. Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Micro/Macro Shock.................... Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Labeling.
------------------------------------------------------------------------
In addition, under Sec. 801.109, the sale, distribution, and use
of EPPG devices are restricted to prescription use. Prescription use
restrictions are a type of general control in section 513(a)(1)(A)(i)
of the FD&C Act. Also, under Sec. 807.81, the device would continue to
be subject to 510(k) notification requirements.
B. PSA Devices
FDA believes that the following special controls, together with
general controls (including applicable prescription-use restrictions
and continuing 510(k) notification requirements), are sufficient to
mitigate the risks to health described in section V for single, dual,
and triple chamber PSA devices:
1. Appropriate analysis/testing must validate EMC within a hospital
environment.
2. Electrical bench testing must demonstrate device safety during
intended use. This must include testing with the specific power source
(i.e., battery power, AC mains connections, or both).
3. Non-clinical performance testing data must demonstrate the
performance characteristics of the device. Testing must include the
following:
Testing must demonstrate the accuracy of monitoring
functions, alarms, measurement features, therapeutic features, and all
adjustable or programmable parameters as identified in labeling;
mechanical bench testing of material strength must
demonstrate that the device and connection cables will withstand forces
or conditions encountered during use;
simulated use analysis/testing must demonstrate adequate
user interface for adjustable parameters, performance of alarms,
display screens, interface with external devices (e.g. data storage,
printing), and indicator(s) functionality under intended use
conditions; and
methods and instructions for cleaning the pulse generator
and connection cables must be validated.
4. Appropriate software verification, validation, and hazard
analysis must be performed.
5. Labeling must include the following:
The labeling must clearly state that these devices are
intended for use in a hospital environment and under the supervision of
a clinician trained in its use;
connector terminals should be clearly, unambiguously
marked on the outside of the EPPG device. The markings should identify
positive (+) and negative (-) polarities. Dual
[[Page 54934]]
chamber devices should clearly identify atrial and ventricular
terminals. Triple chamber devices should clearly identify atrial, right
ventricular, and left ventricular terminals;
the labeling must list all pacing modes available in the
device;
labeling must include a detailed description of any
special capabilities (e.g., overdrive pacing or automatic mode
switching);
labeling must limit the use of external pacing to the
implant procedure; and
appropriate electromagnetic compatibility information must
be included.
Table 2 shows how FDA believes that the risks to health identified
in section V can be mitigated by the proposed special controls.
Table 2--Health Risks and Mitigation Measures for PSA Devices
------------------------------------------------------------------------
Identified risk Mitigation measures
------------------------------------------------------------------------
Misdiagnosis......................... Non-Clinical Performance
Evaluation.
Labeling.
Failure to Pace...................... Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Improper High Rate Pacing............ Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Pacing at an Improperly Low Rate..... Use Environment.
EMC Testing.
Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Improper Pacing Leading to Unwanted Non-Clinical Performance
Stimulation. Evaluation.
Software Verification, Validation
& Hazards Analysis.
Labeling.
Micro/Macro Shock.................... Electrical Safety Testing.
Non-Clinical Performance
Evaluation.
Labeling.
------------------------------------------------------------------------
In addition, under Sec. 801.109, the sale, distribution, and use
of these single and dual chamber PSA devices are restricted to
prescription use. Prescription use restrictions are a type of general
controls in section 513(a)(1)(A)(i) of the FD&C Act. Also, under Sec.
807.81, the device would continue to be subject to 510(k) notification
requirements.
IX. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
X. Paperwork Reduction Act of 1995
This proposed order refers to currently approved collections of
information found in FDA regulations. These collections of information
are subject to review by the Office of Management and Budget (OMB)
under the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The
collections of information in part 807, subpart E, have been approved
under OMB control number 0910-0120; the collections of information in
21 CFR part 814, subpart B, have been approved under OMB control number
0910-0231; and the collections of information under 21 CFR part 801
have been approved under OMB control number 0910-0485.
XI. Codification of Orders
Prior to the amendments by FDASIA, section 513(e) of the FD&C Act
provided for FDA to issue regulations to reclassify devices. Although
section 513(e) as amended authorizes FDA to issue orders rather than
regulations, FDASIA also provides for FDA to revoke previously issued
regulations by order. FDA will continue to codify classifications and
reclassifications in the Code of Federal Regulations (CFR). Changes
resulting from final orders will appear in the CFR as changes to
codified classification determinations or as newly codified orders.
Therefore, under section 513(e)(1)(A)(i) of the FD&C Act, as amended by
FDASIA, in this proposed order we are proposing to: (1) Revoke the
requirements in Sec. 870.3600 related to the classification of EPPG
devices as class III devices and to codify the reclassification of EPPG
devices into class II (special controls) and (2) codify the
reclassification of PSA devices into class II (special controls).
XII. Proposed Effective Date
FDA is proposing that any final order based on this proposed order
become effective on the date of its publication in the Federal Register
or at a later date if stated in the final order.
XIII. Comments
Comments already submitted to the docket (FDA-2011-N-0650) have
been officially noted and do not need to be resubmitted. FDA will
consider previous docket comments in issuing any final orders for these
devices. Interested persons may submit either electronic comments
regarding this document or the associated guidance to https://www.regulations.gov or written comments to the Division of Dockets
Management (see ADDRESSES). It is only necessary to send one set of
comments. Identify comments with the docket number found in brackets in
the heading of this document. Received
[[Page 54935]]
comments may be seen in the Division of Dockets Management between 9
a.m. and 4 p.m., Monday through Friday, and will be posted to the
docket at https://www.regulations.gov.
XIV. References
The following references have been placed on display in the
Division of Dockets Management (see ADDRESSES), and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday,
and are available electronically at https://www.regulations.gov. (FDA
has verified all the Web site addresses in this reference section, but
we are not responsible for any subsequent changes to the Web sites
after this document publishes in the Federal Register.)
1. Class II Special Controls Draft Guidance Document: External
Pacemaker Pulse Generator, available at https://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM275703.pdf.
2. The panel transcript and other meeting materials for the
September 11, 2013, Circulatory System Devices Panel are available
on FDA's Web site at https://www.fda.gov/AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/MedicalDevicesAdvisoryCommittee/CirculatorySystemDevicesPanel/ucm342357.htm.
3. ``ACC/AHA Guidelines for the Management of Patients with ST-
Elevation Myocardial Infarction,'' Circulation, 110:e82-e292, 2004.
4. Beland, M. J., P. S. Hesslein, C. D. Finlay, et al.,
``Noninvasive Transcutaneous Cardiac Pacing in Children,'' Pacing
and Clinical Electrophysiology, 10:1262-1270, 1987.
5. Braun, M. U., T. Rauwolf, M. Bock, et al., ``Percutaneous Lead
Implantation Connected to an External Device in Stimulation-
Dependent Patients With Systemic Infection--A Prospective and
Controlled Study,'' Pacing and Clinical Electrophysiology, August;
29(8):875-879, 2006.
6. Ceresnak, S. R., R. H. Pass, T. J. Starc, et al., ``Predictors
for Hemodynamic Improvement With Temporary Pacing After Pediatric
Cardiac Surgery,'' Journal of Thoracic Cardiovascular Surgery,
January; 141(1):183-187, 2011.
7. Dunn, D. L. and J. J. Gregory, ``Noninvasive Temporary Pacing:
Experience in a Community Hospital,'' Heart and Lung, January;
18(1):23-28, 1989.
8. Eberhardt, F., T. Hanke, M. Heringlake, et al., ``Feasibility of
Temporary Biventricular Pacing in Patients With Reduced Left
Ventricular Function After Coronary Artery Bypass Grafting,'' Pacing
and Clinical Electrophysiology, January; 30 Suppl 1:S50-53, 2007.
9. Ferguson, T. B. Jr. and J. L. Cox, ``Temporary External DDD
Pacing After Cardiac Operations,'' The Annals of Thoracic Surgery,
May; 51(5):723-732, 1991.
10. Halldorsson, A. O., W. T. Vigneswaran, F. J. Podbielski, and D.
M. Evans, ``Electrophysiological and Clinical Comparison of Two
Temporary Pacing Leads Following Cardiac Surgery,'' Pacing and
Clinical Electrophysiology, August; 22(8):1221-1225, 1999.
11. Hindman, M. C., G. S. Wagner, M. JaRo, et al., ``The Clinical
Significance of Bundle Branch Block Complicating Acute Myocardial
Infarction. Indications for Temporary and Permanent Pacemaker
Insertion,'' Circulation, October; 58(4):689-699, 1978.
12. Janousek J., P. Vojtovic, and R. A. Gebauer, ``Use of a
Modified, Commercially Available Temporary Pacemaker for R Wave
Synchronized Atrial Pacing in Postoperative Junctional Ectopic
Tachycardia,'' Pacing and Clinical Electrophysiology, February; 26(2
Pt. 1):579-586, 2003.
13. Kratz, J. and J. Tyler, ``Clinical Experience with a New DDD
External Pacemaker,'' Pacing and Clinical Electrophysiology,
December; 16(12):2227-2234, 1993.
14. Nimetz, A. A., S. J. Shubrooks, A. M. Hutter, and R. W.
DeSanctis, ``The Significance of Bundle Branch Block During Acute
Myocardial Infarction,'' American Heart Journal, October; 90(4):439-
444, 1975.
15. Pinneri, F., S. Frea, K. Najd, et al., ``Echocardiography-Guided
Versus Fluoroscopy-Guided Temporary Pacing in the Emergency Setting:
An Observational Study,'' Journal of Cardiovascular Medicine
(Hagerstown), March; 14(3):242-246, 2013.
16. Siddons, H., ``Transvenous Long-Term Pacing With an External
Pacemaker. What Are the Risks?'', Pacing and Clinical
Electrophysiology, April; 1(2):163-165, 1978.
17. Voigtl[auml]nder T., B, Nowak, P. B[auml]renf[auml]nger, et al.,
``Feasibility and Sensing Thresholds of Temporary Single-Lead VDD
Pacing in Intensive Care,'' American Journal of Cardiology, May 15;
79(10):1360-1363, 1997.
List of Subjects in 21 CFR Part 870
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 870 be amended as follows:
PART 870--CARDIOVASCULAR DEVICES
0
1. The authority citation for 21 CFR part 870 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Section 870.3600 is revised to read as follows:
Sec. 870.3600 External pacemaker pulse generator.
(a) Identification. An external pacemaker pulse generator (EPPG) is
a prescription device that has a power supply and electronic circuits
that produce a periodic electrical pulse to stimulate the heart. This
device, which is used outside the body, is used as a temporary
substitute for the heart's intrinsic pacing system until a permanent
pacemaker can be implanted, or to control irregular heartbeats in
patients following cardiac surgery or a myocardial infarction. The
device may have adjustments for impulse strength, duration, R-wave
sensitivity, and other pacing variables.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Appropriate analysis/testing must validate electromagnetic
compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during
intended use. This must include testing with the specific power source
(i.e., battery power, AC mains connections, or both).
(3) Non-clinical performance testing data must demonstrate the
performance characteristics of the device. Testing must include the
following:
(i) Testing must demonstrate the accuracy of monitoring functions,
alarms, measurement features, therapeutic features, and all adjustable
or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate
that the device and connection cables will withstand forces or
conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user
interface for adjustable parameters, performance of alarms, display
screens, interface with external devices (e.g. data storage, printing),
and indicator(s) functionality under intended use conditions; and
(iv) Methods and instructions for cleaning the pulse generator and
connection cables must be validated.
(4) Appropriate software verification, validation, and hazard
analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended
for use in a hospital environment and under the supervision of a
clinician trained in their use; and
(ii) Connector terminals should be clearly, unambiguously marked on
the outside of the EPPG device. The markings should identify positive
(+) and negative (-) polarities. Dual chamber devices should clearly
identify atrial and ventricular terminals; and
(iii) The labeling must list all pacing modes available in the
device;
[[Page 54936]]
(iv) Labeling must include a detailed description of any special
capabilities (e.g., overdrive pacing or automatic mode switching); and
(v) Appropriate electromagnetic compatibility information must be
included.
0
3. In Subpart D, add Sec. 870.3605 to read as follows:
Sec. 870.3605 Pacing system analyzer.
(a) Identification. A pacing system analyzer (PSA) is a
prescription device that combines the functionality of a pacemaker
electrode function tester (Sec. 870.3720) and an external pacemaker
pulse generator (EPPG) (Sec. 870.3600). It is connected to a pacemaker
lead and uses a power supply and electronic circuits to supply an
accurately calibrated, variable pacing pulse for measuring the
patient's pacing threshold and intracardiac R-wave potential. A PSA may
be a single, dual, or triple chamber system and can simultaneously
deliver pacing therapy while testing one or more implanted pacing
leads.
(b) Classification. Class II (special controls) for PSAs. The
special controls for this device are:
(1) Appropriate analysis/testing must validate electromagnetic
compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during
intended use. This must include testing with the specific power source
(i.e., battery power, AC mains connections, or both).
(3) Non-clinical performance testing data must demonstrate the
performance characteristics of the device. Testing must include the
following:
(i) Testing must demonstrate the accuracy of monitoring functions,
alarms, measurement features, therapeutic features, and all adjustable
or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate
that the device and connection cables will withstand forces or
conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user
interface for adjustable parameters, performance of alarms, display
screens, interface with external devices (e.g. data storage, printing),
and indicator(s) functionality under intended use conditions; and
(iv) Methods and instructions for cleaning the pulse generator and
connection cables must be validated.
(4) Appropriate software verification, validation, and hazard
analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended
for use in a hospital environment and under the supervision of a
clinician trained in their use;
(ii) Connector terminals should be clearly, unambiguously marked on
the outside of the EPPG. The markings should identify positive (+) and
negative (-) polarities. Dual chamber devices should clearly identify
atrial and ventricular terminals. Triple chamber devices should clearly
identify atrial, right ventricular, and left ventricular terminals;
(iii) The labeling must list all pacing modes available in the
device;
(iv) Labeling must include a detailed description of any special
capabilities (e.g., overdrive pacing or automatic mode switching);
(v) Labeling must limit the use of external pacing to the implant
procedure; and
(vi) Appropriate electromagnetic compatibility information must be
included.
Dated: September 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014-21814 Filed 9-12-14; 8:45 am]
BILLING CODE 4164-01-P