Medical Devices; Clinical Chemistry and Clinical Toxicology Devices; Classification of Hemoglobin A1c Test System, 50549-50551 [2014-20022]
Download as PDF
Federal Register / Vol. 79, No. 164 / Monday, August 25, 2014 / Rules and Regulations
requirements of the Paperwork Reduction
Act unless that collection of information
displays a current valid OMB Control
Number. The OMB Control Number for this
information collection is 2120–0056. Public
reporting for this collection of information is
estimated to be approximately 5 minutes per
response, including the time for reviewing
instructions, completing and reviewing the
collection of information. All responses to
this collection of information are mandatory.
Comments concerning the accuracy of this
burden and suggestions for reducing the
burden should be directed to the FAA at: 800
Independence Ave. SW., Washington, DC
20591, Attn: Information Collection
Clearance Officer, AES–200.
(o) Alternative Methods of Compliance
(AMOCs)
(1) The Manager, Fort Worth Airplane
Certification Office, FAA, has the authority to
approve AMOCs for this AD, if requested
using the procedures found in 14 CFR 39.19.
In accordance with 14 CFR 39.19, send your
request to your principal inspector or local
Flight Standards District Office, as
appropriate. If sending information directly
to the manager of the ACO, send it to the
attention of the person identified in
paragraph (p) of this AD.
(2) Before using any approved AMOC,
notify your appropriate principal inspector,
or lacking a principal inspector, the manager
of the local flight standards district office/
certificate holding district office.
(3) AMOCs approved for AD 2010–17–18
R1 are approved as AMOCs for this AD.
emcdonald on DSK67QTVN1PROD with RULES
(p) Related Information
For more information about this AD,
contact Andrew McAnaul, Aerospace
Engineer, ASW–150 (c/o San Antonio MIDO),
10100 Reunion Place, Suite 650, San
Antonio, Texas 78216; phone: (210) 308–
3365; fax: (210) 308–3370; email:
andrew.mcanaul@faa.gov.
(q) Material Incorporated by Reference
(1) The Director of the Federal Register
approved the incorporation by reference
(IBR) of the service information listed in this
paragraph under 5 U.S.C. 552(a) and 1 CFR
part 51.
(2) You must use this service information
as applicable to do the actions required by
this AD, unless the AD specifies otherwise.
(3) The following service information was
approved for IBR on September 9, 2010 (75
FR 52255, August 25, 2010).
(i) Snow Engineering Co. Service Letter
#80GG, dated December 21, 2005.
(ii) Snow Engineering Co. Service Letter
#284, dated October 4, 2009.
(iii) Snow Engineering Co. Service Letter
#281, dated August 1, 2009.
(iv) Snow Engineering Co. Service Letter
#245, dated April 25, 2005.
(v) Snow Engineering Co. Drawing Number
20995, Sheet 2, Rev. D., dated November 25,
2005.
(vi) Snow Engineering Co. Drawing
Number 20995, Sheet 3, dated November 25,
2005.
(vii) Snow Engineering Co. Drawing
Number 20975, Sheet 4, Rev. A., dated
January 7, 2009
VerDate Mar<15>2010
16:10 Aug 22, 2014
Jkt 232001
(4) The following service information was
approved for IBR on April 21, 2006 (71 FR
19994, April 19, 2006).
(i) Snow Engineering Co. Process
Specification #197, page 1, revised June 4,
2002; pages 2 through 4, dated February 23,
2001; and page 5, dated May 3, 2002.
(ii) Snow Engineering Co. Service Letter
#240, dated September 30, 2004.
(5) For Air Tractor, Inc. service information
identified in this AD, contact Air Tractor,
Inc., P.O. Box 485, Olney, Texas 76374;
telephone: (940) 564–5616; fax: (940) 564–
5612; email: airmail@airtractor.com; Internet:
www.airtractor.com.
(6) You may view this service information
at FAA, Small Airplane Directorate, 901
Locust, Kansas City, Missouri 64106. For
information on the availability of this
material at the FAA, call 816–329–4148.
(7) You may view this service information
that is incorporated by reference at the
National Archives and Records
Administration (NARA). For information on
the availability of this material at NARA, call
202–741–6030, or go to: https://
www.archives.gov/federal-register/cfr/ibrlocations.html.
Issued in Kansas City, Missouri, on August
7, 2014.
Earl Lawrence,
Manager, Small Airplane Directorate, Aircraft
Certification Service.
[FR Doc. 2014–20098 Filed 8–22–14; 8:45 am]
BILLING CODE 4910–13–C
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 862
[Docket No. FDA–2014–N–1112]
Medical Devices; Clinical Chemistry
and Clinical Toxicology Devices;
Classification of Hemoglobin A1c Test
System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA) is classifying
hemoglobin A1c test system into class II
(special controls). The special controls
that will apply to this device are
identified in this order and will be part
of the codified language for the
hemoglobin A1c test system
classification. The Agency is classifying
the device into class II (special controls)
in order to provide a reasonable
assurance of safety and effectiveness of
the device.
DATES: This order is effective September
24, 2014. The classification was
applicable May 23, 2013.
SUMMARY:
PO 00000
Frm 00013
Fmt 4700
Sfmt 4700
50549
FOR FURTHER INFORMATION CONTACT:
Meshaun Payne, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 5650, Silver Spring,
MD 20993–0002, 301–796–6668.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C.
360c(f)(1)), devices that were not in
commercial distribution before May 28,
1976 (the date of enactment of the
Medical Device Amendments of 1976),
generally referred to as postamendments
devices, are classified automatically by
statute into class III without any FDA
rulemaking process. These devices
remain in class III and require
premarket approval, unless and until
the device is classified or reclassified
into class I or II, or FDA issues an order
finding the device to be substantially
equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate
device that does not require premarket
approval. The Agency determines
whether new devices are substantially
equivalent to predicate devices by
means of premarket notification
procedures in section 510(k) of the
FD&C Act (21 U.S.C. 360(k)) and part
807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act, as
amended by section 607 of the Food and
Drug Administration Safety and
Innovation Act (Public Law 112–144),
provides two procedures by which a
person may request FDA to classify a
device under the criteria set forth in
section 513(a)(1). Under the first
procedure, the person submits a
premarket notification under section
510(k) of the FD&C Act for a device that
has not previously been classified and,
within 30 days of receiving an order
classifying the device into class III
under section 513(f)(1) of the FD&C Act,
the person requests a classification
under section 513(f)(2). Under the
second procedure, rather than first
submitting a premarket notification
under section 510(k) of the FD&C Act
and then a request for classification
under the first procedure, the person
determines that there is no legally
marketed device upon which to base a
determination of substantial
equivalence and requests a classification
under section 513(f)(2) of the FD&C Act.
If the person submits a request to
classify the device under this second
procedure, FDA may decline to
undertake the classification request if
FDA identifies a legally marketed device
that could provide a reasonable basis for
E:\FR\FM\25AUR1.SGM
25AUR1
50550
Federal Register / Vol. 79, No. 164 / Monday, August 25, 2014 / Rules and Regulations
review of substantial equivalence with
the device or if FDA determines that the
device submitted is not of ‘‘lowmoderate risk’’ or that general controls
would be inadequate to control the risks
and special controls to mitigate the risks
cannot be developed.
In response to a request to classify a
device under either procedure provided
by section 513(f)(2) of the FD&C Act,
FDA will classify the device by written
order within 120 days. This
classification will be the initial
classification of the device.
In accordance with section 513(f)(1) of
the FD&C Act, FDA issued an order on
March 14, 2013, classifying the COBAS
INTEGRA 800 Tina-quant HbA1cDx
Gen.2 assay into class III, because it was
not substantially equivalent to a device
that was introduced or delivered for
introduction into interstate commerce
for commercial distribution before May
28, 1976, or a device which was
subsequently reclassified into class I or
class II. On March 29, 2013, Roche
Diagnostics Corporation submitted a
request for classification of the COBAS
INTEGRA 800 Tina-quant HbA1cDx
Gen.2 assay under section 513(f)(2) of
the FD&C Act. The manufacturer
recommended that the device be
classified into class II.
In accordance with section 513(f)(2) of
the FD&C Act, FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1). FDA classifies
devices into class II if general controls
by themselves are insufficient to
provide reasonable assurance of safety
and effectiveness, but there is sufficient
information to establish special controls
to provide reasonable assurance of the
safety and effectiveness of the device for
its intended use. After review of the
information submitted in the de novo
request, FDA determined that the device
can be classified into class II with the
establishment of special controls. FDA
believes these special controls, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on May 23, 2013, FDA
issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding § 862.1373.
Following the effective date of this
final classification administrative order,
any firm submitting a premarket
notification (510(k)) for a hemoglobin
A1c test system will need to comply
with the special controls named in the
final administrative order.
The device is assigned the generic
name hemoglobin A1c test system, and
it is identified as a device used to
measure the percentage concentration of
hemoglobin A1c in blood. Measurement
of hemoglobin A1c is used as an aid in
the diagnosis of diabetes mellitus and as
an aid in the identification of patients
at risk for developing diabetes mellitus.
FDA has identified the following risks
to health associated with this type of
device and the measures required to
mitigate these risks in table 1:
TABLE 1—IDENTIFIED RISKS AND REQUIRED MITIGATIONS
Identified risks
Required mitigations
emcdonald on DSK67QTVN1PROD with RULES
False negative result due to inadequate device performance ..............................................................................
False positive result due to inadequate device performance ...............................................................................
Use of the test for patients with hemoglobin variants that may interfere with the test system, and lead to incorrect results.
FDA believes that the following
special controls, in addition to the
general controls, address these risks to
health and provide reasonable assurance
of safety and effectiveness:
1. The device must have initial and
annual standardization verification by a
certifying glycohemoglobin
standardization organization deemed
acceptable by FDA.
2. The premarket notification
submission must include performance
testing to evaluate precision, accuracy,
linearity and interference, including the
following:
• Performance testing of device
precision must, at a minimum, use
blood samples with concentrations near
5.0 percent, 6.5 percent, 8.0 percent and
12 percent hemoglobin A1c. This testing
must evaluate precision over a
minimum of 20 days using at least three
lots of the device and three instruments,
as applicable.
• Performance testing of device
accuracy must include a minimum of
120 blood samples that span the
measuring interval of the device and
compare results of the new device to
results of a standardized test method.
VerDate Mar<15>2010
16:10 Aug 22, 2014
Jkt 232001
Results must demonstrate little or no
bias versus the standardized method.
• Total error of the new device must
be evaluated using single measurements
by the new device compared to results
of the standardized test method, and
this evaluation must demonstrate a total
error less than or equal to 6 percent.
• Performance testing must
demonstrate that there is little to no
interference from common hemoglobin
variants, including Hemoglobin C,
Hemoglobin D, Hemoglobin E,
Hemoglobin A2, and Hemoglobin S.
3. When assay interference from
Hemoglobin F or interference with other
hemoglobin variants with low frequency
in the population is observed, a warning
statement must be placed in a black box
and must appear in all labeling material
for these devices describing the
interference and any affected
populations.
Hemoglobin A1c test system devices
are prescription devices restricted to
patient use only upon the authorization
of a practitioner licensed by law to
administer or use the device. (See
section 520(e) of the FD&C Act (21
U.S.C. 360j(e)) and 21 CFR 801.109
(Prescription devices).). Prescription-use
PO 00000
Frm 00014
Fmt 4700
Sfmt 4700
Special controls (1) and (2).
Special controls (1) and (2).
Special control (3).
restrictions are a type of general control
as defined in section 513(a)(1)(A)(i) of
the FD&C Act.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) if
FDA determines that premarket
notification is not necessary to provide
reasonable assurance of the safety and
effectiveness of the device. For this type
of device, FDA has determined that
premarket notification is necessary to
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, this device type is not
exempt from premarket notification
requirements. Persons who intend to
market this type of device must submit
to FDA a premarket notification, prior to
marketing the device, which contains
information about the hemoglobin A1c
test system they intend to market.
II. Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
E:\FR\FM\25AUR1.SGM
25AUR1
Federal Register / Vol. 79, No. 164 / Monday, August 25, 2014 / Rules and Regulations
nor an environmental impact statement
is required.
III. Paperwork Reduction Act of 1995
This final administrative order
establishes special controls that refer to
previously approved collections of
information found in other FDA
regulations. These collections of
information are subject to review by the
Office of Management and Budget
(OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501–3520). The
collections of information in part 807,
subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120; the collections of
information in 21 CFR part 820 have
been approved under OMB control
number 0910–0073; and the collections
of information in 21 CFR parts 801 and
809 regarding labeling have been
approved under OMB control number
0910–0485.
List of Subjects in 21 CFR Part 862
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 862 is
amended as follows:
PART 862—CLINICAL CHEMISTRY
AND CLINICAL TOXICOLOGY
DEVICES
1. The authority citation for 21 CFR
part 862 continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Add § 862.1373 to subpart B to read
as follows:
■
emcdonald on DSK67QTVN1PROD with RULES
§ 862.1373
Dated: August 19, 2014.
Peter Lurie,
Associate Commissioner for Policy and
Planning.
16:10 Aug 22, 2014
Jkt 232001
DATES:
BILLING CODE 4164–01–P
This rule is effective August 25,
2014.
FOR FURTHER INFORMATION CONTACT:
Aaron Josephson, Center for Devices
and Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 5449, Silver Spring,
MD 20993–0002, 301–796–5178.
FDA is
amending certain regulations in parts
862, 864, 866, and 872 (21 CFR parts
862, 864, 866, and 872). This action
updates certain Web site addresses that
have been changed due to recent FDA
Web site changes.
Publication of this document
constitutes final action of these changes
under the Administrative Procedure Act
(5 U.S.C. 553). These amendments are
merely correcting nonsubstantive errors.
FDA therefore, for good cause, finds
under 5 U.S.C. 553(b)(3)(B) and (d)(3)
that notice and public comment are
unnecessary.
FDA has determined under 21 CFR
25.30(i) that this final rule is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required. In addition, FDA has
determined that this final rule contains
no collection of information. Therefore,
clearance by the Office of Management
and Budget under the Paperwork
Reduction Act of 1995 is not required.
SUPPLEMENTARY INFORMATION:
List of Subjects
21 CFR Part 862
[FR Doc. 2014–20022 Filed 8–22–14; 8:45 am]
Medical devices.
21 CFR Part 864
Hemoglobin A1c test system.
(a) Identification. A hemoglobin A1c
test system is a device used to measure
the percentage concentration of
hemoglobin A1c in blood. Measurement
of hemoglobin A1c is used as an aid in
the diagnosis of diabetes mellitus and as
an aid in the identification of patients
at risk for developing diabetes mellitus.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) The device must have initial and
annual standardization verification by a
certifying glycohemoglobin
standardization organization deemed
acceptable by FDA.
(2) The premarket notification
submission must include performance
testing to evaluate precision, accuracy,
linearity, and interference, including the
following:
(i) Performance testing of device
precision must, at a minimum, use
VerDate Mar<15>2010
blood samples with concentrations near
5.0 percent, 6.5 percent, 8.0 percent,
and 12 percent hemoglobin A1c. This
testing must evaluate precision over a
minimum of 20 days using at least three
lots of the device and three instruments,
as applicable.
(ii) Performance testing of device
accuracy must include a minimum of
120 blood samples that span the
measuring interval of the device and
compare results of the new device to
results of a standardized test method.
Results must demonstrate little or no
bias versus the standardized method.
(iii) Total error of the new device
must be evaluated using single
measurements by the new device
compared to results of the standardized
test method, and this evaluation must
demonstrate a total error less than or
equal to 6 percent.
(iv) Performance testing must
demonstrate that there is little to no
interference from common hemoglobin
variants, including Hemoglobin C,
Hemoglobin D, Hemoglobin E,
Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from
Hemoglobin F or interference with other
hemoglobin variants with low frequency
in the population is observed, a warning
statement must be placed in a black box
and must appear in all labeling material
for these devices describing the
interference and any affected
populations.
50551
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Blood, Medical devices, Packaging
and containers.
21 CFR Part 866
Food and Drug Administration
21 CFR Parts 862, 864, 866, and 872
Biologics, Laboratories, Medical
devices.
[Docket No. FDA–2014–N–0011]
21 CFR Part 872
Medical Devices; Technical
Amendment
AGENCY:
Food and Drug Administration,
HHS.
Final rule; technical
amendment.
ACTION:
The Food and Drug
Administration (FDA) is amending
certain medical device regulations to
correct minor errors in the Code of
Federal Regulations (CFR). This action
is editorial in nature and is intended to
correct outdated Web site addresses.
SUMMARY:
PO 00000
Frm 00015
Fmt 4700
Sfmt 4700
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act, and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR Chapter I is
amended as follows:
PARTS 862, 864, 866, AND 872
[AMENDED]
1. The authority citation for parts 862,
864, 866, and 872 continues to read as
follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
E:\FR\FM\25AUR1.SGM
25AUR1
Agencies
[Federal Register Volume 79, Number 164 (Monday, August 25, 2014)]
[Rules and Regulations]
[Pages 50549-50551]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2014-20022]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 862
[Docket No. FDA-2014-N-1112]
Medical Devices; Clinical Chemistry and Clinical Toxicology
Devices; Classification of Hemoglobin A1c Test System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying
hemoglobin A1c test system into class II (special controls). The
special controls that will apply to this device are identified in this
order and will be part of the codified language for the hemoglobin A1c
test system classification. The Agency is classifying the device into
class II (special controls) in order to provide a reasonable assurance
of safety and effectiveness of the device.
DATES: This order is effective September 24, 2014. The classification
was applicable May 23, 2013.
FOR FURTHER INFORMATION CONTACT: Meshaun Payne, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 5650, Silver Spring, MD 20993-0002, 301-796-6668.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of the Federal Food, Drug, and
Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were
not in commercial distribution before May 28, 1976 (the date of
enactment of the Medical Device Amendments of 1976), generally referred
to as postamendments devices, are classified automatically by statute
into class III without any FDA rulemaking process. These devices remain
in class III and require premarket approval, unless and until the
device is classified or reclassified into class I or II, or FDA issues
an order finding the device to be substantially equivalent, in
accordance with section 513(i) of the FD&C Act, to a predicate device
that does not require premarket approval. The Agency determines whether
new devices are substantially equivalent to predicate devices by means
of premarket notification procedures in section 510(k) of the FD&C Act
(21 U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act, as amended by section 607 of the
Food and Drug Administration Safety and Innovation Act (Public Law 112-
144), provides two procedures by which a person may request FDA to
classify a device under the criteria set forth in section 513(a)(1).
Under the first procedure, the person submits a premarket notification
under section 510(k) of the FD&C Act for a device that has not
previously been classified and, within 30 days of receiving an order
classifying the device into class III under section 513(f)(1) of the
FD&C Act, the person requests a classification under section 513(f)(2).
Under the second procedure, rather than first submitting a premarket
notification under section 510(k) of the FD&C Act and then a request
for classification under the first procedure, the person determines
that there is no legally marketed device upon which to base a
determination of substantial equivalence and requests a classification
under section 513(f)(2) of the FD&C Act. If the person submits a
request to classify the device under this second procedure, FDA may
decline to undertake the classification request if FDA identifies a
legally marketed device that could provide a reasonable basis for
[[Page 50550]]
review of substantial equivalence with the device or if FDA determines
that the device submitted is not of ``low-moderate risk'' or that
general controls would be inadequate to control the risks and special
controls to mitigate the risks cannot be developed.
In response to a request to classify a device under either
procedure provided by section 513(f)(2) of the FD&C Act, FDA will
classify the device by written order within 120 days. This
classification will be the initial classification of the device.
In accordance with section 513(f)(1) of the FD&C Act, FDA issued an
order on March 14, 2013, classifying the COBAS INTEGRA 800 Tina-quant
HbA1cDx Gen.2 assay into class III, because it was not substantially
equivalent to a device that was introduced or delivered for
introduction into interstate commerce for commercial distribution
before May 28, 1976, or a device which was subsequently reclassified
into class I or class II. On March 29, 2013, Roche Diagnostics
Corporation submitted a request for classification of the COBAS INTEGRA
800 Tina-quant HbA1cDx Gen.2 assay under section 513(f)(2) of the FD&C
Act. The manufacturer recommended that the device be classified into
class II.
In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed
the request in order to classify the device under the criteria for
classification set forth in section 513(a)(1). FDA classifies devices
into class II if general controls by themselves are insufficient to
provide reasonable assurance of safety and effectiveness, but there is
sufficient information to establish special controls to provide
reasonable assurance of the safety and effectiveness of the device for
its intended use. After review of the information submitted in the de
novo request, FDA determined that the device can be classified into
class II with the establishment of special controls. FDA believes these
special controls, in addition to general controls, will provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, on May 23, 2013, FDA issued an order to the requestor
classifying the device into class II. FDA is codifying the
classification of the device by adding Sec. 862.1373.
Following the effective date of this final classification
administrative order, any firm submitting a premarket notification
(510(k)) for a hemoglobin A1c test system will need to comply with the
special controls named in the final administrative order.
The device is assigned the generic name hemoglobin A1c test system,
and it is identified as a device used to measure the percentage
concentration of hemoglobin A1c in blood. Measurement of hemoglobin A1c
is used as an aid in the diagnosis of diabetes mellitus and as an aid
in the identification of patients at risk for developing diabetes
mellitus.
FDA has identified the following risks to health associated with
this type of device and the measures required to mitigate these risks
in table 1:
Table 1--Identified Risks and Required Mitigations
----------------------------------------------------------------------------------------------------------------
Identified risks Required mitigations
----------------------------------------------------------------------------------------------------------------
False negative result due to inadequate device Special controls (1) and (2).
performance.
False positive result due to inadequate device Special controls (1) and (2).
performance.
Use of the test for patients with hemoglobin variants Special control (3).
that may interfere with the test system, and lead to
incorrect results.
----------------------------------------------------------------------------------------------------------------
FDA believes that the following special controls, in addition to
the general controls, address these risks to health and provide
reasonable assurance of safety and effectiveness:
1. The device must have initial and annual standardization
verification by a certifying glycohemoglobin standardization
organization deemed acceptable by FDA.
2. The premarket notification submission must include performance
testing to evaluate precision, accuracy, linearity and interference,
including the following:
Performance testing of device precision must, at a
minimum, use blood samples with concentrations near 5.0 percent, 6.5
percent, 8.0 percent and 12 percent hemoglobin A1c. This testing must
evaluate precision over a minimum of 20 days using at least three lots
of the device and three instruments, as applicable.
Performance testing of device accuracy must include a
minimum of 120 blood samples that span the measuring interval of the
device and compare results of the new device to results of a
standardized test method. Results must demonstrate little or no bias
versus the standardized method.
Total error of the new device must be evaluated using
single measurements by the new device compared to results of the
standardized test method, and this evaluation must demonstrate a total
error less than or equal to 6 percent.
Performance testing must demonstrate that there is little
to no interference from common hemoglobin variants, including
Hemoglobin C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin
S.
3. When assay interference from Hemoglobin F or interference with
other hemoglobin variants with low frequency in the population is
observed, a warning statement must be placed in a black box and must
appear in all labeling material for these devices describing the
interference and any affected populations.
Hemoglobin A1c test system devices are prescription devices
restricted to patient use only upon the authorization of a practitioner
licensed by law to administer or use the device. (See section 520(e) of
the FD&C Act (21 U.S.C. 360j(e)) and 21 CFR 801.109 (Prescription
devices).). Prescription-use restrictions are a type of general control
as defined in section 513(a)(1)(A)(i) of the FD&C Act.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) if FDA determines that premarket notification is not necessary
to provide reasonable assurance of the safety and effectiveness of the
device. For this type of device, FDA has determined that premarket
notification is necessary to provide reasonable assurance of the safety
and effectiveness of the device. Therefore, this device type is not
exempt from premarket notification requirements. Persons who intend to
market this type of device must submit to FDA a premarket notification,
prior to marketing the device, which contains information about the
hemoglobin A1c test system they intend to market.
II. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment
[[Page 50551]]
nor an environmental impact statement is required.
III. Paperwork Reduction Act of 1995
This final administrative order establishes special controls that
refer to previously approved collections of information found in other
FDA regulations. These collections of information are subject to review
by the Office of Management and Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501-3520). The collections of
information in part 807, subpart E, regarding premarket notification
submissions have been approved under OMB control number 0910-0120; the
collections of information in 21 CFR part 820 have been approved under
OMB control number 0910-0073; and the collections of information in 21
CFR parts 801 and 809 regarding labeling have been approved under OMB
control number 0910-0485.
List of Subjects in 21 CFR Part 862
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
862 is amended as follows:
PART 862--CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES
0
1. The authority citation for 21 CFR part 862 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Add Sec. 862.1373 to subpart B to read as follows:
Sec. 862.1373 Hemoglobin A1c test system.
(a) Identification. A hemoglobin A1c test system is a device used
to measure the percentage concentration of hemoglobin A1c in blood.
Measurement of hemoglobin A1c is used as an aid in the diagnosis of
diabetes mellitus and as an aid in the identification of patients at
risk for developing diabetes mellitus.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) The device must have initial and annual standardization
verification by a certifying glycohemoglobin standardization
organization deemed acceptable by FDA.
(2) The premarket notification submission must include performance
testing to evaluate precision, accuracy, linearity, and interference,
including the following:
(i) Performance testing of device precision must, at a minimum, use
blood samples with concentrations near 5.0 percent, 6.5 percent, 8.0
percent, and 12 percent hemoglobin A1c. This testing must evaluate
precision over a minimum of 20 days using at least three lots of the
device and three instruments, as applicable.
(ii) Performance testing of device accuracy must include a minimum
of 120 blood samples that span the measuring interval of the device and
compare results of the new device to results of a standardized test
method. Results must demonstrate little or no bias versus the
standardized method.
(iii) Total error of the new device must be evaluated using single
measurements by the new device compared to results of the standardized
test method, and this evaluation must demonstrate a total error less
than or equal to 6 percent.
(iv) Performance testing must demonstrate that there is little to
no interference from common hemoglobin variants, including Hemoglobin
C, Hemoglobin D, Hemoglobin E, Hemoglobin A2, and Hemoglobin S.
(3) When assay interference from Hemoglobin F or interference with
other hemoglobin variants with low frequency in the population is
observed, a warning statement must be placed in a black box and must
appear in all labeling material for these devices describing the
interference and any affected populations.
Dated: August 19, 2014.
Peter Lurie,
Associate Commissioner for Policy and Planning.
[FR Doc. 2014-20022 Filed 8-22-14; 8:45 am]
BILLING CODE 4164-01-P