Medical Devices; General Hospital and Personal Use Devices; Classification of the Intravascular Administration Set, Automated Air Removal System, 28404-28407 [2014-11174]
Download as PDF
TKELLEY on DSK3SPTVN1PROD with RULES
28404
Federal Register / Vol. 79, No. 95 / Friday, May 16, 2014 / Rules and Regulations
must be tested and detailed protocols
must be provided for each test:
(i) Bite test to ensure that the capsule
can withstand extreme cases of biting.
(ii) pH resistance test to evaluate
integrity of the capsule when exposed to
a range of pH values.
(iii) Battery life test to demonstrate
that the capsule’s operating time is not
constrained by the battery capacity.
(iv) Shelf-life testing to demonstrate
that the device performs as intended at
the proposed shelf-life date.
(v) Optical testing to evaluate
fundamental image quality
characteristics such as resolution, field
of view, depth of field, distortion,
signal-to-noise ratio, uniformity, and
image artifacts. A test must be
performed to evaluate the potential of
scratches, caused by travelling through
the gastrointestinal tract, on the
transparent window of the capsule and
their impact on the optical and color
performance.
(vi) An optical safety analysis must be
performed based on maximum (worstcase) light exposure to internal
gastrointestinal mucosa, and covering
ultraviolet, visible, and near-infrared
ranges, as appropriate. A mitigation
analysis must be provided.
(vii) A color performance test must be
provided to compare the color
differences between the input scene and
output image.
(viii) The video viewer must clearly
present the temporal or spatial
relationship between any two frames as
a real-time lapse or a travel distance.
The video viewer must alert the user
when the specific video interval is
captured at a frame rate lower than the
nominal one due to communication
errors.
(ix) A performance test evaluating the
latency caused by any adaptive
algorithm such as adjustable frame rate
must be provided.
(x) If the capsule includes a
localization module, a localization
performance test must be performed to
verify the accuracy and precision of
locating the capsule position within the
colon.
(xi) A data transmission test must be
performed to verify the robustness of the
data transmission between the capsule
and the recorder. Controlled signal
attenuation should be included for
simulating a non-ideal environment.
(xii) Software validation, verification,
and hazards analysis must be provided.
(xiii) Electrical equipment safety,
including thermal and mechanical
safety and electromagnetic compatibility
(EMC) testing must be performed. If the
environments of intended use include
locations outside of hospitals and
VerDate Mar<15>2010
20:35 May 15, 2014
Jkt 232001
clinics, appropriate higher immunity
test levels must be used. Labeling must
include appropriate EMC information.
(xiv) Information demonstrating
immunity from wireless hazards.
(3) The clinical performance
characteristics of the device for the
detection of colon polyps must be
established. Demonstration of the
performance characteristics must
include assessment of positive percent
agreement and negative percent
agreement compared to a clinically
acceptable alternative structural imaging
method.
(4) Clinician labeling must include:
(i) Specific instructions and the
clinical and technical expertise needed
for the safe use of the device.
(ii) A detailed summary of the clinical
testing pertinent to use of the device,
including the percentage of patients in
which a polyp was correctly identified
by capsule endoscopy, but also the
percent of patients in which the capsule
either missed or falsely identified a
polyp with respect to the clinically
acceptable alternative structural imaging
method.
(iii) The colon cleansing procedure.
(iv) A detailed summary of the device
technical parameters.
(v) A detailed summary of the deviceand procedure-related complications
pertinent to use of the device.
(vi) An expiration date/shelf life.
(5) Patient labeling must include:
(i) An explanation of the device and
the mechanism of operation.
(ii) Patient preparation procedure.
(iii) A brief summary of the clinical
study. The summary should not only
include the percentage of patients in
which a polyp was correctly identified
by capsule endoscopy, but also the
percent of patients in which the capsule
either missed or falsely identified a
polyp with respect to the clinically
acceptable alternative structural imaging
method.
(iv) A summary of the device- and
procedure-related complications
pertinent to use of the device.
Dated: May 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014–11173 Filed 5–15–14; 8:45 am]
BILLING CODE 4160–01–P
PO 00000
Frm 00012
Fmt 4700
Sfmt 4700
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA–2014–N–0438]
Medical Devices; General Hospital and
Personal Use Devices; Classification
of the Intravascular Administration Set,
Automated Air Removal System
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Final order.
The Food and Drug
Administration (FDA) is classifying the
intravascular administration set,
automated air removal system into class
II (special controls). The special controls
that will apply to the device are
identified in this order and will be part
of the codified language for the
intravascular administration set,
automated air removal system’s
classification. The Agency is classifying
the device into class II (special controls)
in order to provide a reasonable
assurance of safety and effectiveness of
the device.
DATES: This order is effective June 16,
2014. The classification was effective on
March 4, 2014.
FOR FURTHER INFORMATION CONTACT:
Alan Stevens, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave. Bldg. 66, Rm. 2561, Silver Spring,
MD 20993–0002, 301–796–6294.
SUPPLEMENTARY INFORMATION:
SUMMARY:
I. Background
In accordance with section 513(f)(1) of
the Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C.
360c(f)(1)), devices that were not in
commercial distribution before May 28,
1976 (the date of enactment of the
Medical Device Amendments of 1976),
generally referred to as postamendments
devices, are classified automatically by
statute into class III without any FDA
rulemaking process. These devices
remain in class III and require
premarket approval, unless and until
the device is classified or reclassified
into class I or II, or FDA issues an order
finding the device to be substantially
equivalent, in accordance with section
513(i), to a predicate device that does
not require premarket approval. The
Agency determines whether new
devices are substantially equivalent to
predicate devices by means of
premarket notification procedures in
section 510(k) of the FD&C Act (21
E:\FR\FM\16MYR1.SGM
16MYR1
Federal Register / Vol. 79, No. 95 / Friday, May 16, 2014 / Rules and Regulations
U.S.C. 360(k)) and part 807 (21 CFR part
807) of the regulations.
Section 513(f)(2) of the FD&C Act, as
amended by section 607 of the Food and
Drug Administration Safety and
Innovation Act (Pub. L. 112–144, July 9,
2012), provides two procedures by
which a person may request FDA to
classify a device under the criteria set
forth in section 513(a)(1). Under the first
procedure, the person submits a
premarket notification under section
510(k) for a device that has not
previously been classified and, within
30 days of receiving an order classifying
the device into class III under section
513(f)(1), the person requests a
classification under section 513(f)(2).
Under the second procedure, rather than
first submitting a premarket notification
under section 510(k) and then a request
for classification under the first
procedure, the person determines that
there is no legally marketed device upon
which to base a determination of
substantial equivalence and requests a
classification under section 513(f)(2). If
the person submits a request to classify
the device under this second procedure,
FDA may decline to undertake the
classification request if FDA identifies a
legally marketed device that could
provide a reasonable basis for review of
substantial equivalence with the device
or if FDA determines that the device
submitted is not of ‘‘low-moderate risk’’
or that general controls would be
inadequate to control the risks and
special controls to mitigate the risks
cannot be developed.
In response to a request to classify a
device under either procedure provided
by section 513(f)(2) of the FD&C Act,
FDA will classify the device by written
order within 120 days. This
classification will be the initial
classification of the device.
In accordance with section 513(f)(1) of
the FD&C Act, FDA issued an order on
October 23, 2008, classifying the
AirPurge System into class III, because
it was not substantially equivalent to a
device that was introduced or delivered
for introduction into interstate
commerce for commercial distribution
before May 28, 1976, or a device which
was subsequently reclassified into class
I or class II. On October 29, 2008,
Anesthesia Safety Products, LLC
submitted a request requesting
classification of the AirPurge System
under section 513(f)(2) of the FD&C Act.
The manufacturer recommended that
the device be classified into class II (Ref.
1).
In accordance with section 513(f)(2) of
the FD&C Act, FDA reviewed the
request in order to classify the device
under the criteria for classification set
forth in section 513(a)(1) of the FD&C
Act. FDA classifies devices into class II
if general controls by themselves are
insufficient to provide reasonable
assurance of safety and effectiveness,
but there is sufficient information to
establish special controls to provide
reasonable assurance of the safety and
effectiveness of the device for its
intended use. After review of the
information submitted in the request,
28405
FDA determined that the device can be
classified into class II with the
establishment of special controls. FDA
believes these special controls, in
addition to general controls, will
provide reasonable assurance of the
safety and effectiveness of the device.
Therefore, on March 4, 2014, FDA
issued an order to the requestor
classifying the device into class II. FDA
is codifying the classification of the
device by adding 21 CFR 880.5445.
Following the effective date of this
final classification order, any firm
submitting a premarket notification
(510(k)) for an intravascular
administration set, automated air
removal system will need to comply
with the special controls named in this
final order. The device is assigned the
generic name intravascular
administration set, automated air
removal system, and it is identified as
a prescription device used to detect and
automatically remove air from an
intravascular administration set with
minimal to no interruption in the flow
of the intravascular fluid. The device
may include an air identification
mechanism, software, an air removal
mechanism, tubing, apparatus to collect
removed air, and safety control
mechanisms to address hazardous
situations.
FDA has identified the following risks
to health associated specifically with
this type of device, as well as the
mitigation measures required to mitigate
these risks.
TABLE 1—IDENTIFIED RISKS AND MITIGATION MEASURES
Identified risk
Mitigation measures
Embolus ..................................................................................................................
Infusion Delivery Error ............................................................................................
Electric Shock .........................................................................................................
Adverse Tissue Reaction ........................................................................................
TKELLEY on DSK3SPTVN1PROD with RULES
Infection ...................................................................................................................
FDA believes that the following
special controls, in combination with
the general controls, address these risks
to health and provide reasonable
VerDate Mar<15>2010
20:35 May 15, 2014
Jkt 232001
Hazard Argument.
Software.
Electromagnetic Compatibility.
Human Factors.
Labeling.
Nonclinical Performance Testing.
Hazard Argument.
Software.
Electromagnetic Compatibility.
Human Factors.
Labeling.
Nonclinical Performance Testing.
Hazard Argument.
Electrical Safety.
Electromagnetic Compatibility.
Hazard Argument.
Biocompatibility.
Sterilization.
Shelf Life.
assurance of the safety and
effectiveness:
1. Provide an argument demonstrating
that all reasonably foreseeable hazards
PO 00000
Frm 00013
Fmt 4700
Sfmt 4700
have been adequately addressed with
respect to the persons for whose use the
device is represented or intended and
E:\FR\FM\16MYR1.SGM
16MYR1
TKELLEY on DSK3SPTVN1PROD with RULES
28406
Federal Register / Vol. 79, No. 95 / Friday, May 16, 2014 / Rules and Regulations
the conditions of use for the device,
which includes the following:
• Description of the device
indications for use, design, and
technology, use environments, and
users in sufficient detail to determine
that the device complies with all special
controls.
• Demonstrate that controls are
implemented to address device system
hazards and their causes.
• Include a justification supporting
the acceptability criteria for each hazard
control.
• A traceability analysis
demonstrating that all credible hazards
have at least one corresponding control
and that all controls have been verified
and validated in the final device design.
2. Appropriate software verification,
validation, and hazard analysis must be
performed.
3. The device parts that directly or
indirectly contact the patient must be
demonstrated to be biocompatible.
4. Performance data must demonstrate
the sterility of fluid path contacting
components and the shelf life of these
components.
5. The device must be designed and
tested for electrical safety and
electromagnetic compatibility (EMC).
6. Nonclinical performance testing
data must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
• Device system and component
reliability testing must be conducted.
• Fluid ingress protection testing
must be conducted.
• Testing of safety controls must be
performed to demonstrate adequate
mitigation of hazardous situations,
including sensor failure, flow control
failure, improper device position,
device malfunction, infusion delivery
error, and release of air to the patient.
7. A human factors validation study
must demonstrate that use hazards are
adequately addressed.
8. The labeling must include the
following:
• The device’s air identification and
removal response time.
• The device’s minimum air volume
identification sensitivity.
• The minimum and maximum flow
rates at which the device is capable of
reliably detecting and removing air.
• Quantification of any fluid loss
during device air removal operations as
a function of flow rate.
Intravascular administration set,
automated air removal systems are
prescription devices restricted to patient
use only upon the authorization of a
practitioner licensed by law to
VerDate Mar<15>2010
20:35 May 15, 2014
Jkt 232001
administer or use the device (21 CFR
880.5445(a); see section 520(e) of the
FD&C Act (21 U.S.C. 360j(e)) and 21
CFR 801.109 (Prescription devices.)).
Prescription-use restrictions are a type
of general controls as defined in section
513(a)(1)(A)(i) of the FD&C Act.
Section 510(m) of the FD&C Act
provides that FDA may exempt a class
II device from the premarket notification
requirements under section 510(k) of the
FD&C Act, if FDA determines that
premarket notification is not necessary
to provide reasonable assurance of the
safety and effectiveness of the device.
For this type of device, FDA has
determined that premarket notification
is necessary to provide reasonable
assurance of the safety and effectiveness
of the device. Therefore, this device
type is not exempt from premarket
notification requirements. Persons who
intend to market this type of device
must submit to FDA a premarket
notification, prior to marketing the
device, which contains information
about the intravascular administration
set, automated air removal system they
intend to market.
II. Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
III. Paperwork Reduction Act of 1995
This final order establishes special
controls that refer to previously
approved collections of information
found in other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
part 807, subpart E, regarding premarket
notification submissions have been
approved under OMB control number
0910–0120, and the collections of
information in 21 CFR part 801,
regarding labeling have been approved
under OMB control number 0910–0485.
IV. Reference
The following reference has been
placed on display in the Division of
Dockets Management (HFA–305), Food
and Drug Administration, 5630 Fishers
Lane, Rm. 1061, Rockville, MD 20852,
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday, and is available
electronically at https://
www.regulations.gov.
PO 00000
Frm 00014
Fmt 4700
Sfmt 4700
1. K080644: De Novo Request per 513(f)(2)
pursuant to the Agency’s not
substantially equivalent (NSE)
determination, dated October 23, 2008,
from Anesthesia Safety Products, LLC,
dated October 29, 2008.
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, 21 CFR part 880 is
amended as follows:
PART 880—GENERAL HOSPITAL AND
PERSONAL USE DEVICES
1. The authority citation for 21 CFR
part 880 continues to read as follows:
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Add § 880.5445 to subpart F to read
as follows:
■
§ 880.5445 Intravascular Administration
Set, Automated Air Removal System.
(a) Identification. An intravascular
administration set, automated air
removal system, is a prescription device
used to detect and automatically remove
air from an intravascular administration
set with minimal to no interruption in
the flow of the intravascular fluid. The
device may include an air identification
mechanism, software, an air removal
mechanism, tubing, apparatus to collect
removed air, and safety control
mechanisms to address hazardous
situations.
(b) Classification. Class II (special
controls). The special controls for this
device are:
(1) Provide an argument
demonstrating that all reasonably
foreseeable hazards have been
adequately addressed with respect to
the persons for whose use the device is
represented or intended and the
conditions of use for the device, which
includes the following:
(i) Description of the device
indications for use, design, and
technology, use environments, and
users in sufficient detail to determine
that the device complies with all special
controls.
(ii) Demonstrate that controls are
implemented to address device system
hazards and their causes.
(iii) Include a justification supporting
the acceptability criteria for each hazard
control.
(iv) A traceability analysis
demonstrating that all credible hazards
have at least one corresponding control
and that all controls have been verified
and validated in the final device design.
E:\FR\FM\16MYR1.SGM
16MYR1
Federal Register / Vol. 79, No. 95 / Friday, May 16, 2014 / Rules and Regulations
(2) Appropriate software verification,
validation, and hazard analysis must be
performed.
(3) The device parts that directly or
indirectly contact the patient must be
demonstrated to be biocompatible.
(4) Performance data must
demonstrate the sterility of fluid path
contacting components and the shelf life
of these components.
(5) The device must be designed and
tested for electrical safety and
electromagnetic compatibility (EMC).
(6) Nonclinical performance testing
data must demonstrate that the device
performs as intended under anticipated
conditions of use. The following
performance characteristics must be
tested:
(i) Device system and component
reliability testing must be conducted.
(ii) Fluid ingress protection testing
must be conducted.
(iii) Testing of safety controls must be
performed to demonstrate adequate
mitigation of hazardous situations,
including sensor failure, flow control
failure, improper device position,
device malfunction, infusion delivery
error, and release of air to the patient.
(7) A human factors validation study
must demonstrate that use hazards are
adequately addressed.
(8) The labeling must include the
following:
(i) The device’s air identification and
removal response time.
(ii) The device’s minimum air volume
identification sensitivity.
(iii) The minimum and maximum
flow rates at which the device is capable
of reliably detecting and removing air.
(iv) Quantification of any fluid loss
during device air removal operations as
a function of flow rate.
Dated: May 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014–11174 Filed 5–15–14; 8:45 am]
BILLING CODE 4160–01–P
DEPARTMENT OF DEFENSE
Office of the Secretary
32 CFR Part 79
[Docket ID: DOD–2011–OS–0124]
TKELLEY on DSK3SPTVN1PROD with RULES
RIN 0790–AI81
Child Development Programs (CDPs)
Office of the Secretary,
Department of Defense (DoD).
ACTION: Interim final rule.
AGENCY:
This interim final rule
updates policy, responsibilities, and
SUMMARY:
VerDate Mar<15>2010
20:35 May 15, 2014
Jkt 232001
procedures for providing care to minor
children birth through age 12 years of
individuals who are eligible for care in
DoD CDPs to include center-based care,
family child care (FCC), school-age care
(SAC), supplemental child care, and
community based care; authorizes the
publication of supporting guidance for
the implementation of CDP policies and
responsibilities, including child
development training modules, program
aids, and other management tools; and
establishes the DoD Effectiveness Rating
and Improvement System (ERIS).
DATES: Effective date: This rule is
effective May 16, 2014.
Comment date: Comments must be
received by July 15, 2014.
ADDRESSES: You may submit comments,
identified by docket number and/or RIN
number and title, by any of the
following methods:
• Federal Rulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
• Mail: Federal Docket Management
System Office, 4800 Mark Center Drive,
2nd Floor, East Tower, Suite 02G09,
Alexandria, VA 22350–3100.
Instructions: All submissions received
must include the agency name and
docket number or Regulatory
Information Number (RIN) for this
Federal Register document. The general
policy for comments and other
submissions from members of the public
is to make these submissions available
for public viewing on the Internet at
https://www.regulations.gov as they are
received without change, including any
personal identifiers or contact
information.
FOR FURTHER INFORMATION CONTACT:
Eddy Mentzer, 571–372–0857.
SUPPLEMENTARY INFORMATION:
Justification for Interim Final Rule
This interim final rule provides
overarching policy to the Military
Departments in the execution of their
roles in providing quality child
development programs that ensure the
safety and well-being of children in the
DoD’s care. A 2012 Secretary of Defense
directed audit of criminal background
check processes for all DoD Child and
Youth Services personnel revealed the
need areas for all applicable directives
to be updated to ensure current and
accurate policy is incorporated. The
White House and Secretary of Defense
directed a priority review of the
management and oversight of child and
youth programs in 2013. The review
noted variation in Service-level
approaches to oversight inspections
including headquarters-level
comprehensive inspections and
PO 00000
Frm 00015
Fmt 4700
Sfmt 4700
28407
installation-level fire, health, and safety
inspections. The report recommended
the OSD promulgate guidance to ensure
standardization and clarity. Defense
child development program staff and
leadership have committed to the
SECDEF and White House that they are
committed to improving the consistency
by which these services are delivered
and to ensure the safety and well-being
of children in our care. This interim
final rule addresses these
recommendations and creates a stronger
environment of standardization across
the services.
This interim final rule identifies the
applicability of 32 CFR part 56,
‘‘Nondiscrimination on the Basis of
Handicap in Programs and Activities
Assisted or conducted by the
Department of Defense’’ that implement
section 504 of the Rehabilitation Act for
federally conducted and federally
assisted programs as they apply to
children and youth with special needs.
This interim final rule expands previous
policy by (1) Requiring procedures for
reviewing and making reasonable
accommodation of children with special
needs that do not fundamentally alter
the nature of the program; (2)
considering the needs of the child, the
disability, and the environment of group
care in child development facilities or
home-based care, staffing needs and
training requirements, and resources of
the program; and (3) including Child
Development Programs as part of the
multi-disciplinary Inclusion Action
Team that supports families of children
with special needs.
This interim final rule extends child
care benefits to same-sex spouse of
Military Service members. At the
direction of the President, the
Department has conducted a careful and
deliberative review of benefits currently
provided. The Department has now
identified family member and
dependent benefits that we can lawfully
provide to same-sex spouse and their
children through changes in DoD
policies and regulations. These benefits
shall be extended to same-sex spouse
and, where applicable, children of
same-sex spouses.
Executive Summary
I. Purpose of the Regulatory Action
a. This interim final rule proposes to:
(a) update policy, responsibilities, and
procedures for providing care to minor
children birth through age 12 years of
individuals who are eligible for care in
Department of Defense Child
Development Programs (CDP) to include
center-based care, family child care
(FCC), school-age care (SAC),
E:\FR\FM\16MYR1.SGM
16MYR1
Agencies
[Federal Register Volume 79, Number 95 (Friday, May 16, 2014)]
[Rules and Regulations]
[Pages 28404-28407]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2014-11174]
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 880
[Docket No. FDA-2014-N-0438]
Medical Devices; General Hospital and Personal Use Devices;
Classification of the Intravascular Administration Set, Automated Air
Removal System
AGENCY: Food and Drug Administration, HHS.
ACTION: Final order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is classifying the
intravascular administration set, automated air removal system into
class II (special controls). The special controls that will apply to
the device are identified in this order and will be part of the
codified language for the intravascular administration set, automated
air removal system's classification. The Agency is classifying the
device into class II (special controls) in order to provide a
reasonable assurance of safety and effectiveness of the device.
DATES: This order is effective June 16, 2014. The classification was
effective on March 4, 2014.
FOR FURTHER INFORMATION CONTACT: Alan Stevens, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave. Bldg. 66, Rm. 2561, Silver Spring, MD 20993-0002, 301-796-6294.
SUPPLEMENTARY INFORMATION:
I. Background
In accordance with section 513(f)(1) of the Federal Food, Drug, and
Cosmetic Act (the FD&C Act) (21 U.S.C. 360c(f)(1)), devices that were
not in commercial distribution before May 28, 1976 (the date of
enactment of the Medical Device Amendments of 1976), generally referred
to as postamendments devices, are classified automatically by statute
into class III without any FDA rulemaking process. These devices remain
in class III and require premarket approval, unless and until the
device is classified or reclassified into class I or II, or FDA issues
an order finding the device to be substantially equivalent, in
accordance with section 513(i), to a predicate device that does not
require premarket approval. The Agency determines whether new devices
are substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21
[[Page 28405]]
U.S.C. 360(k)) and part 807 (21 CFR part 807) of the regulations.
Section 513(f)(2) of the FD&C Act, as amended by section 607 of the
Food and Drug Administration Safety and Innovation Act (Pub. L. 112-
144, July 9, 2012), provides two procedures by which a person may
request FDA to classify a device under the criteria set forth in
section 513(a)(1). Under the first procedure, the person submits a
premarket notification under section 510(k) for a device that has not
previously been classified and, within 30 days of receiving an order
classifying the device into class III under section 513(f)(1), the
person requests a classification under section 513(f)(2). Under the
second procedure, rather than first submitting a premarket notification
under section 510(k) and then a request for classification under the
first procedure, the person determines that there is no legally
marketed device upon which to base a determination of substantial
equivalence and requests a classification under section 513(f)(2). If
the person submits a request to classify the device under this second
procedure, FDA may decline to undertake the classification request if
FDA identifies a legally marketed device that could provide a
reasonable basis for review of substantial equivalence with the device
or if FDA determines that the device submitted is not of ``low-moderate
risk'' or that general controls would be inadequate to control the
risks and special controls to mitigate the risks cannot be developed.
In response to a request to classify a device under either
procedure provided by section 513(f)(2) of the FD&C Act, FDA will
classify the device by written order within 120 days. This
classification will be the initial classification of the device.
In accordance with section 513(f)(1) of the FD&C Act, FDA issued an
order on October 23, 2008, classifying the AirPurge System into class
III, because it was not substantially equivalent to a device that was
introduced or delivered for introduction into interstate commerce for
commercial distribution before May 28, 1976, or a device which was
subsequently reclassified into class I or class II. On October 29,
2008, Anesthesia Safety Products, LLC submitted a request requesting
classification of the AirPurge System under section 513(f)(2) of the
FD&C Act. The manufacturer recommended that the device be classified
into class II (Ref. 1).
In accordance with section 513(f)(2) of the FD&C Act, FDA reviewed
the request in order to classify the device under the criteria for
classification set forth in section 513(a)(1) of the FD&C Act. FDA
classifies devices into class II if general controls by themselves are
insufficient to provide reasonable assurance of safety and
effectiveness, but there is sufficient information to establish special
controls to provide reasonable assurance of the safety and
effectiveness of the device for its intended use. After review of the
information submitted in the request, FDA determined that the device
can be classified into class II with the establishment of special
controls. FDA believes these special controls, in addition to general
controls, will provide reasonable assurance of the safety and
effectiveness of the device.
Therefore, on March 4, 2014, FDA issued an order to the requestor
classifying the device into class II. FDA is codifying the
classification of the device by adding 21 CFR 880.5445.
Following the effective date of this final classification order,
any firm submitting a premarket notification (510(k)) for an
intravascular administration set, automated air removal system will
need to comply with the special controls named in this final order. The
device is assigned the generic name intravascular administration set,
automated air removal system, and it is identified as a prescription
device used to detect and automatically remove air from an
intravascular administration set with minimal to no interruption in the
flow of the intravascular fluid. The device may include an air
identification mechanism, software, an air removal mechanism, tubing,
apparatus to collect removed air, and safety control mechanisms to
address hazardous situations.
FDA has identified the following risks to health associated
specifically with this type of device, as well as the mitigation
measures required to mitigate these risks.
Table 1--Identified Risks and Mitigation Measures
----------------------------------------------------------------------------------------------------------------
Identified risk Mitigation measures
----------------------------------------------------------------------------------------------------------------
Embolus......................... Hazard Argument.
Software.
Electromagnetic Compatibility.
Human Factors.
Labeling.
Nonclinical Performance Testing.
Infusion Delivery Error......... Hazard Argument.
Software.
Electromagnetic Compatibility.
Human Factors.
Labeling.
Nonclinical Performance Testing.
Electric Shock.................. Hazard Argument.
Electrical Safety.
Electromagnetic Compatibility.
Adverse Tissue Reaction......... Hazard Argument.
Biocompatibility.
Infection....................... Sterilization.
Shelf Life.
----------------------------------------------------------------------------------------------------------------
FDA believes that the following special controls, in combination
with the general controls, address these risks to health and provide
reasonable assurance of the safety and effectiveness:
1. Provide an argument demonstrating that all reasonably
foreseeable hazards have been adequately addressed with respect to the
persons for whose use the device is represented or intended and
[[Page 28406]]
the conditions of use for the device, which includes the following:
Description of the device indications for use, design, and
technology, use environments, and users in sufficient detail to
determine that the device complies with all special controls.
Demonstrate that controls are implemented to address
device system hazards and their causes.
Include a justification supporting the acceptability
criteria for each hazard control.
A traceability analysis demonstrating that all credible
hazards have at least one corresponding control and that all controls
have been verified and validated in the final device design.
2. Appropriate software verification, validation, and hazard
analysis must be performed.
3. The device parts that directly or indirectly contact the patient
must be demonstrated to be biocompatible.
4. Performance data must demonstrate the sterility of fluid path
contacting components and the shelf life of these components.
5. The device must be designed and tested for electrical safety and
electromagnetic compatibility (EMC).
6. Nonclinical performance testing data must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance characteristics must be tested:
Device system and component reliability testing must be
conducted.
Fluid ingress protection testing must be conducted.
Testing of safety controls must be performed to
demonstrate adequate mitigation of hazardous situations, including
sensor failure, flow control failure, improper device position, device
malfunction, infusion delivery error, and release of air to the
patient.
7. A human factors validation study must demonstrate that use
hazards are adequately addressed.
8. The labeling must include the following:
The device's air identification and removal response time.
The device's minimum air volume identification
sensitivity.
The minimum and maximum flow rates at which the device is
capable of reliably detecting and removing air.
Quantification of any fluid loss during device air removal
operations as a function of flow rate.
Intravascular administration set, automated air removal systems are
prescription devices restricted to patient use only upon the
authorization of a practitioner licensed by law to administer or use
the device (21 CFR 880.5445(a); see section 520(e) of the FD&C Act (21
U.S.C. 360j(e)) and 21 CFR 801.109 (Prescription devices.)).
Prescription-use restrictions are a type of general controls as defined
in section 513(a)(1)(A)(i) of the FD&C Act.
Section 510(m) of the FD&C Act provides that FDA may exempt a class
II device from the premarket notification requirements under section
510(k) of the FD&C Act, if FDA determines that premarket notification
is not necessary to provide reasonable assurance of the safety and
effectiveness of the device. For this type of device, FDA has
determined that premarket notification is necessary to provide
reasonable assurance of the safety and effectiveness of the device.
Therefore, this device type is not exempt from premarket notification
requirements. Persons who intend to market this type of device must
submit to FDA a premarket notification, prior to marketing the device,
which contains information about the intravascular administration set,
automated air removal system they intend to market.
II. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
III. Paperwork Reduction Act of 1995
This final order establishes special controls that refer to
previously approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in
part 807, subpart E, regarding premarket notification submissions have
been approved under OMB control number 0910-0120, and the collections
of information in 21 CFR part 801, regarding labeling have been
approved under OMB control number 0910-0485.
IV. Reference
The following reference has been placed on display in the Division
of Dockets Management (HFA-305), Food and Drug Administration, 5630
Fishers Lane, Rm. 1061, Rockville, MD 20852, and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday,
and is available electronically at https://www.regulations.gov.
1. K080644: De Novo Request per 513(f)(2) pursuant to the Agency's
not substantially equivalent (NSE) determination, dated October 23,
2008, from Anesthesia Safety Products, LLC, dated October 29, 2008.
List of Subjects in 21 CFR Part 880
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, 21 CFR part
880 is amended as follows:
PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES
0
1. The authority citation for 21 CFR part 880 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Add Sec. 880.5445 to subpart F to read as follows:
Sec. 880.5445 Intravascular Administration Set, Automated Air Removal
System.
(a) Identification. An intravascular administration set, automated
air removal system, is a prescription device used to detect and
automatically remove air from an intravascular administration set with
minimal to no interruption in the flow of the intravascular fluid. The
device may include an air identification mechanism, software, an air
removal mechanism, tubing, apparatus to collect removed air, and safety
control mechanisms to address hazardous situations.
(b) Classification. Class II (special controls). The special
controls for this device are:
(1) Provide an argument demonstrating that all reasonably
foreseeable hazards have been adequately addressed with respect to the
persons for whose use the device is represented or intended and the
conditions of use for the device, which includes the following:
(i) Description of the device indications for use, design, and
technology, use environments, and users in sufficient detail to
determine that the device complies with all special controls.
(ii) Demonstrate that controls are implemented to address device
system hazards and their causes.
(iii) Include a justification supporting the acceptability criteria
for each hazard control.
(iv) A traceability analysis demonstrating that all credible
hazards have at least one corresponding control and that all controls
have been verified and validated in the final device design.
[[Page 28407]]
(2) Appropriate software verification, validation, and hazard
analysis must be performed.
(3) The device parts that directly or indirectly contact the
patient must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility of fluid path
contacting components and the shelf life of these components.
(5) The device must be designed and tested for electrical safety
and electromagnetic compatibility (EMC).
(6) Nonclinical performance testing data must demonstrate that the
device performs as intended under anticipated conditions of use. The
following performance characteristics must be tested:
(i) Device system and component reliability testing must be
conducted.
(ii) Fluid ingress protection testing must be conducted.
(iii) Testing of safety controls must be performed to demonstrate
adequate mitigation of hazardous situations, including sensor failure,
flow control failure, improper device position, device malfunction,
infusion delivery error, and release of air to the patient.
(7) A human factors validation study must demonstrate that use
hazards are adequately addressed.
(8) The labeling must include the following:
(i) The device's air identification and removal response time.
(ii) The device's minimum air volume identification sensitivity.
(iii) The minimum and maximum flow rates at which the device is
capable of reliably detecting and removing air.
(iv) Quantification of any fluid loss during device air removal
operations as a function of flow rate.
Dated: May 9, 2014.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2014-11174 Filed 5-15-14; 8:45 am]
BILLING CODE 4160-01-P