Medical Devices; Ophthalmic Devices; Classification of the Scleral Plug, 5327-5330 [2013-01447]
Download as PDF
Federal Register / Vol. 78, No. 17 / Friday, January 25, 2013 / Proposed Rules
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA–2012–N–1238]
Medical Devices; Ophthalmic Devices;
Classification of the Scleral Plug
AGENCY:
Food and Drug Administration,
HHS.
ACTION:
Proposed rule.
The Food and Drug
Administration (FDA or Agency) is
proposing to classify the scleral plug
into class II (special controls), and
proposing to exempt the scleral plugs
composed of surgical grade stainless
steel (with or without coating in gold,
silver, or titanium) from premarket
notification (510(k)) and to continue to
require premarket notification (510(k))
for all other scleral plugs in order to
provide a reasonable assurance of safety
and effectiveness of the device. The
scleral plug is a prescription device
used to provide temporary closure of a
scleral incision during an ophthalmic
surgical procedure.
DATES: Submit either electronic or
written comments by April 25, 2013.
See section IV of this document for the
proposed effective date of a final rule
that may issue based on this proposal.
ADDRESSES: You may submit comments,
identified by Docket No. FDA–2012–N–
1238, by any of the following methods:
SUMMARY:
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
sroberts on DSK5SPTVN1PROD with
Written Submissions
Submit written submissions in the
following way:
• Mail/Hand delivery/Courier (for
paper or CD–ROM submissions):
Division of Dockets Management (HFA–
305), Food and Drug Administration,
5630 Fishers Lane, Rm. 1061, Rockville,
MD 20852.
Instructions: All submissions received
must include the Agency name and
Docket No. FDA–2012–N–1238 for this
rulemaking. All comments received may
be posted without change to https://
www.regulations.gov, including any
personal information provided. For
additional information on submitting
comments, see the ‘‘Comments’’ heading
of the SUPPLEMENTARY INFORMATION
section of this document.
VerDate Mar<15>2010
17:46 Jan 24, 2013
Jkt 229001
Docket: For access to the docket to
read background documents or
comments received, go to https://
www.regulations.gov and insert the
docket number, found in brackets in the
heading of this document, into the
‘‘Search’’ box and follow the prompts
and/or go to the Division of Dockets
Management, 5630 Fishers Lane, Rm.
1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Tina
Kiang, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2414, Silver Spring,
MD 20993–0002, 301–796–6860,
Tina.Kiang@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
A. Statutory and Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act) (21 U.S.C. 301 et
seq.), as amended by the Medical Device
Amendments of 1976 (the 1976
amendments) (Pub. L. 94–295), the Safe
Medical Devices Act of 1990 (Pub. L.
101–629), the Food and Drug
Administration Modernization Act of
1997 (FDAMA) (Pub. L. 105–115),
among other amendments, established a
comprehensive system for the regulation
of medical devices intended for human
use. Section 513 of the FD&C Act (21
U.S.C. 360c) established three categories
(classes) of devices, depending on the
regulatory controls needed to provide
reasonable assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Under section 513 of the FD&C Act,
FDA refers to devices that were in
commercial distribution before May 28,
1976 (the date of enactment of the 1976
amendments), as ‘‘preamendments
devices.’’ FDA classifies these devices
after the Agency takes the following
steps: (1) Receives a recommendation
from a device classification panel (an
FDA advisory committee); (2) publishes
the panel’s recommendation for
comment, along with a proposed
regulation classifying the device; and (3)
publishes a final regulation classifying
the device. FDA has classified most
preamendments devices under these
procedures.
FDA refers to devices that were not in
commercial distribution before May 28,
1976, as ‘‘postamendments devices.’’
These devices are classified
automatically by statute (section 513(f)
of the FD&C Act (21 U.S.C. 360c(f)) into
class III without any FDA rulemaking
process. These devices remain in class
III and require premarket approval,
PO 00000
Frm 00008
Fmt 4702
Sfmt 4702
5327
unless and until: (1) FDA reclassifies the
device into class I or II; (2) FDA issues
an order classifying the device into class
I or II in accordance with section
513(f)(2) of the FD&C Act, as amended
by FDAMA; or (3) FDA issues an order
finding the device to be substantially
equivalent, under section 513(i) of the
FD&C Act (21 U.S.C. 360c(i)), to a
predicate device that does not require
premarket approval. The Agency
determines whether new devices are
substantially equivalent to predicate
devices by means of premarket
notification procedures in section 510(k)
of the FD&C Act and part 807 of the
regulations (21 CFR Part 807).
A person may market a
preamendments device that has been
classified into class III through
premarket notification procedures,
without submission of a premarket
approval application (PMA) until FDA
issues a final regulation under section
515(b) of the FD&C Act (21 U.S.C.
360e(b)) requiring premarket approval.
Section 510(m) of the FD&C Act
provides that a class II device may be
exempted from the premarket
notification requirements under section
510(k) of the FD&C Act, if the Agency
determines that premarket notification
is not necessary to assure the safety and
effectiveness of the device. FDA has
determined that premarket notification
is not necessary to assure the safety and
effectiveness of scleral plugs if the
material is a surgical grade stainless
steel with or without a gold, silver, or
titanium coating.
B. Regulatory History of the Device
After the enactment of the Medical
Device Amendments of 1976, FDA
commenced to identify and classify all
preamendments devices, in accordance
with section 513(b) (21 U.S.C. 360c(b))
of the FD&C Act. In the Federal Register
of September 2, 1987 (52 FR 33346),
FDA classified a total of 109 generic
types of ophthalmic devices. The scleral
plug was not identified in this initial
effort. FDA has regulated scleral plugs
as devices requiring premarket
notification (510(k)). Scleral plugs
currently on the market have been
determined to be substantially
equivalent to devices that were in
commercial distribution prior to May
28, 1976. Currently, FDA regulates
scleral plugs as devices requiring
premarket notification (510(k)). There
have been ten 510(k) submissions
received and cleared under product
code LXP (scleral plugs).
Consistent with the FD&C Act and the
regulations, FDA consulted with the
Ophthalmic Devices Panel (the Panel),
an FDA advisory committee, regarding
E:\FR\FM\25JAP1.SGM
25JAP1
5328
Federal Register / Vol. 78, No. 17 / Friday, January 25, 2013 / Proposed Rules
the classification of this device type on
January 22, 1996 (Ref. 1). At the panel
meeting, the Panel recommended scleral
plugs as classification as class I, 510(k)
exempt. Two 510(k) submissions have
been cleared since the panel meeting.
II. Recommendation of the Panel
During a public meeting which was
held on January 22, 1996, the Panel
made recommendations regarding the
classification and regulatory controls for
the scleral plug. FDA is proposing the
following identification based on the
Panel’s recommendations and the
Agency’s review:
sroberts on DSK5SPTVN1PROD with
A. Identification
A scleral plug is a prescription device
intended to provide temporary closure
of a scleral incision during an
ophthalmic surgical procedure. These
plugs prevent intraocular fluid and
pressure loss when instruments are
withdrawn from the eye. Scleral plugs
include a head portion remaining above
the sclera, which can be gripped for
insertion and removal, and a shaft that
fits inside the scleral incision. Scleral
plugs are removed before completing
the surgery. Therefore, they are
generally only used in operating rooms.
These devices are often made of surgical
grade stainless steel and can be coated
in gold, silver, or titanium.
Scleral plugs have a long and
established history of clinical use. They
are routinely used in many ophthalmic
surgeries (specifically, vitreoretinal
surgeries). One common type of
vitreoretinal surgery is vitrectomy.
Vitrectomy is estimated to be the third
most frequently performed ophthalmic
surgical operation, after cataract and
excimer laser refractive surgery (Ref. 2).
Approximately 225,000 vitrectomies are
done in the United States each year (Ref.
2).
B. Recommended Classification of the
Panel
Although the Panel was informed that
scleral plugs have historically been
treated as class II devices, the Panel
recommended that a scleral plug made
of a material previously used in legally
marketed devices be classified into class
I (general controls) and be exempt from
premarket notification because the
biocompatibility and ability to be
sterilized have already been established.
The Panel’s rationale for suggesting that
the scleral plug be classified into class
I was because general controls would
provide reasonable assurance of the
safety and effectiveness of the device
type if it is made from a material
established to be readily sterilized and
biocompatible. During the panel
VerDate Mar<15>2010
17:46 Jan 24, 2013
Jkt 229001
discussion, a distinction was made that
scleral plugs consisting of other
materials (i.e., materials that are not
already included in legally marketed
medical devices to the date of the
classification regulation) should be
classified into class II and require
biocompatibility testing as a special
control.
As a result of the distinction between
materials used for this device, the Panel
recommended that, unless new
materials are proposed, the device
should be exempt from premarket
notification.
C. Summary of Reasons for
Recommendation
The Panel considered FDA’s extensive
regulatory experience with the device
type and the Panel members’ personal
knowledge of and clinical experience
with the device type. The Panel also
considered the long history of safety and
effectiveness of the device over many
years of clinical use. The Panel
recommended that the scleral plug be
classified into class I because it
concluded that general controls would
provide reasonable assurance of the
safety and effectiveness of the device
type if it was made from a material
established to be readily sterilized and
biocompatible. The Panel also
recommended that scleral plugs be
exempt from premarket notification
requirements if the proposed device
does not introduce new materials (i.e.,
materials that are not established to be
safe for this type of application).
However, FDA believes that a class II
classification is appropriate and
consistent with the intent of the Panel
to establish requirements (such as
biocompatibility and sterility) for these
devices. Although the Panel identified
potential risks and the measures that
could be taken to mitigate these risks,
the Panel’s recommendation of class I
would not permit FDA to establish as
special controls the mitigation measures
discussed (biocompatibility, sterility).
Therefore, while FDA is not adopting
the Panel’s recommendation of
classification into class I, the Agency
agrees with the concerns and mitigation
measures discussed by the Panel that
would support a classification under
class II.
mitigate these risks are identified in
table 1 of this document.
TABLE 1—HEALTH RISKS AND MITIGATION MEASURES FOR THE SCLERAL
PLUG MADE FROM SURGICAL
GRADE STAINLESS STEEL
[With or without a gold, silver, or titanium
coating]
Identified
risk
Mitigation
measures
Infection .....................
Adverse Tissue Reaction.
Loss, breakage, or
migration of the
plug.
Sterility Testing.
Biocompatibility Testing.
Labeling.
For scleral plugs that are made of
surgical grade stainless steel (with or
without a gold, silver, or titanium
coating) the following special controls,
in addition to general controls, can
address the risks to health in table 1 of
this document and provide reasonable
assurance of safety and effectiveness of
the device: (1) Performance data must
demonstrate the sterility and shelf life of
the device; (2) the device must be
demonstrated to be biocompatible; and
(3) labeling must include all information
required for the safe and effective use of
the device, including specific
instructions regarding the proper
incision size, placement, and removal of
the device.
Because of the varying properties of
other materials and the potential impact
on safety and effectiveness, FDA has
identified additional special controls for
devices made of materials other than
surgical grade stainless steel. Based on
the Panel’s discussion and
recommendations and FDA’s experience
with the device, the risks to health
associated with the scleral plugs made
from materials other than surgical grade
stainless steel and the proposed
measures to mitigate these risks are
identified in table 2 of this document.
TABLE 2—HEALTH RISKS AND MITIGATION
MEASURES FOR SCLERAL
PLUGS MADE FROM MATERIALS
OTHER THAN SURGICAL GRADE
STAINLESS STEEL
Identified
risk
Mitigation
measures
Infection .................
Sterility Testing.
Shelf-life Testing.
Biocompatibility testing.
Material characterization.
Performance testing to
determine the level of
extractables.
D. Risks to Health
Based on the Panel’s discussion and
recommendations and FDA’s experience
with the device, the risks to health
associated with the scleral plugs made
from surgical grade stainless steel with
or without gold, silver, or titanium
coating and the proposed measures to
PO 00000
Frm 00009
Fmt 4702
Sfmt 4702
Adverse Tissue Reaction.
E:\FR\FM\25JAP1.SGM
25JAP1
Federal Register / Vol. 78, No. 17 / Friday, January 25, 2013 / Proposed Rules
TABLE 2—HEALTH RISKS AND MITIGATION
MEASURES FOR SCLERAL
PLUGS MADE FROM MATERIALS
OTHER THAN SURGICAL GRADE
STAINLESS STEEL—Continued
Identified
risk
Mitigation
measures
Loss, breakage, or
migration of the
plug.
Material characterization.
Performance testing for
Mechanical Properties.
Labeling.
The Agency believes that the
following special controls, in addition
to general controls, will provide
reasonable assurance of safety and
effectiveness for scleral plugs that are
composed of a material other than
surgical grade stainless steel, as outlined
in table 2: (1) Performance data must
demonstrate the sterility and shelf life of
the device; (2) the device must be
demonstrated to be biocompatible; (3)
characterization of the device materials
must be performed; (4) performance
data must demonstrate acceptable
mechanical properties under simulated
clinical use conditions including
insertion and removal of the device; (5)
performance data must demonstrate
adequately low levels of the extractables
or residues from manufacturing (or
processing) of the device; and (6)
labeling must include all information
required for the safe and effective use of
the device, including specific
instructions regarding the proper
incision size, placement, and removal of
the device. In addition, the scleral plug
is a prescription device and must be
used in accordance with 21 CFR
801.109.
sroberts on DSK5SPTVN1PROD with
III. Proposed Classification and FDA’s
Finding
Adverse events involving scleral
plugs are rare, as evidenced by the fact
that FDA identified only a single
adverse event in our reporting systems
and two adverse events in the published
literature (Refs. 3 and 4). The one
adverse event reported to FDA resulted
in no persistent adverse effects to the
patient and, according to the report, this
specific type of non-metallic scleral
plug was discontinued and replaced
with surgical grade stainless steel scleral
plugs.
FDA believes that a class II
classification is consistent with the
intent of the Panel to establish
requirements (such as biocompatibility
and sterility) for these devices. The
identified special controls mitigate the
known risks of the device that were
VerDate Mar<15>2010
17:46 Jan 24, 2013
Jkt 229001
identified by the Panel. However, the
FDA does not agree with the Panel that
all materials included in legally
marketed scleral devices can be
exempted from 510(k) due to the
potential for safety concerns in some
materials that will require specific
material information and performance
data to provide a reasonable assurance
of safety and effectiveness. FDA believes
this type of information should be
reviewed by FDA prior to a device being
marketed in the United States.
FDA proposes the scleral plug be
classified into class II. The special
controls, in addition to general controls,
will provide reasonable assurance of the
safety and effectiveness of the device.
FDA also agrees, in part, with the
Panel’s recommendation that premarket
notification is not necessary to assure
the safety and effectiveness of scleral
plugs if new materials are not
introduced and, therefore, the Agency is
giving notice of intent to exempt the
scleral plug device from premarket
notification requirements if the device is
made from surgical grade stainless steel
(with or without a gold, silver, or
titanium coating).
IV. Proposed Effective Date
FDA proposes that any final
regulation based on this proposal
become effective 30 days after its date
of publication in the Federal Register.
V. Environmental Impact
The Agency has determined under 21
CFR 25.34(b) that this action is of a type
that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
VI. Analysis of Impacts
FDA has examined the impacts of the
proposed rule under Executive Order
12866 and Executive Order 13563, the
Regulatory Flexibility Act (5 U.S.C.
601–612), and the Unfunded Mandates
Reform Act of 1995 (Pub. L. 104–4).
Executive Orders 12866 and 13563
direct Agencies to assess all costs and
benefits of available regulatory
alternatives and, when regulation is
necessary, to select regulatory
approaches that maximize net benefits
(including potential economic,
environmental, public health and safety,
and other advantages; distributive
impacts; and equity). The Agency
believes that this proposed rule is not a
significant regulatory action as defined
by Executive Order 12866.
The Regulatory Flexibility Act
requires Agencies to analyze regulatory
PO 00000
Frm 00010
Fmt 4702
Sfmt 4702
5329
options that would minimize any
significant impact of a rule on small
entities. Because the proposed
regulation will classify a previously
unclassified pre-Amendment device
type, there are only five registered
establishments listed in the
Establishment Registration and Device
Listing database, and the proposed
regulation designating the classification
of scleral plugs as class II is consistent
with the historical regulatory oversight
given to this device type, the Agency
proposes to certify that the final rule
will not have a significant economic
impact on a substantial number of small
entities.
Section 202(a) of the Unfunded
Mandates Reform Act of 1995 requires
that Agencies prepare a written
statement, which includes an
assessment of anticipated costs and
benefits, before proposing ‘‘any rule that
includes any Federal mandate that may
result in the expenditure by State, local,
and tribal governments, in the aggregate,
or by the private sector, of $100,000,000
or more (adjusted annually for inflation)
in any one year.’’ The current threshold
after adjustment for inflation is $139
million, using the most current (2011)
Implicit Price Deflator for the Gross
Domestic Product. FDA does not expect
this proposed rule to result in any 1year expenditure that would meet or
exceed this amount.
The proposed rule would impact
current manufacturers if they were to
make changes to their existing products
and any manufacture wanting to market
a new scleral plug. If the new or
changed product is made of surgical
grade stainless steel with or without
gold, silver, or titanium coating,
manufacturers could begin marketing
after they complied with the proposed
special controls. They would not need
to submit an application to the Agency
for preapproval. There would be no
change from current requirements for
new products made of alternative
materials; they would need premarket
notification before marketing.
VII. Paperwork Reduction Act of 1995
This proposed rule establishes special
controls that refer to currently approved
collections of information found in
other FDA regulations. These
collections of information are subject to
review by the Office of Management and
Budget (OMB) under the Paperwork
Reduction Act of 1995 (44 U.S.C. 3501–
3520). The collections of information in
21 CFR part 807, subpart E, have been
approved under OMB control number
0910–0120; the collections of
information in 21 CFR part 801 have
been approved under OMB control
E:\FR\FM\25JAP1.SGM
25JAP1
5330
Federal Register / Vol. 78, No. 17 / Friday, January 25, 2013 / Proposed Rules
§ 886.4155
number 0910–0485; the collections of
information in 21 CFR part 807 have
been approved under OMB control
number 0910–0387.
VIII. Comments
Interested persons may submit to the
Division of Dockets Management (see
ADDRESSES) either electronic or written
comments regarding this document. It is
only necessary to send one set of
comments. Identify comments with the
docket number found in brackets in the
heading of this document. Received
comments may be seen in the Division
of Dockets Management between 9 a.m.
and 4 p.m., Monday through Friday.
IX. References
The following references have been
placed on display in the Division of
Dockets Management (see ADDRESSES)
and may be seen by interested persons
between 9 a.m. and 4 p.m., Monday
through Friday and are available
electronically at https://
www.regulations.gov. (FDA has verified
the Web site addresses of the following
references, but FDA is not responsible
for any subsequent changes to the Web
site after this document publishes in the
Federal Register.)
1. Transcript from the Food and Drug
Administration Ophthalmic Devices Panel
Meeting, January 22, 1996.
2. Chang, Stanley, ‘‘LXII Edward Jackson
Lecture: Open Angle Glaucoma After
Vitrectomy,’’ American Journal of
Ophthalmology, vol. 141(6): pp. 1033–1043,
June 2006, available at https://
www.sciencedirect.com/science/article/pii/
S0002939406002546.
3. Stewart, M. W., ‘‘Intraoperative
Radiographic Detection of a ‘Lost’ Scleral
Plug,’’ Retina, vol. 25(4): pp. 526–527, June
2005.
4. Bovino, J. A. and D. F. Marcus,
‘‘Intraocular Foreign-Body Hazard During
Vitrectomy,’’ American Journal of
Ophthalmology, vol. 93 (3): p. 366, March
1982.
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods
and services.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, FDA proposes to
amend part 886 as follows:
PART 886—OPHTHALMIC DEVICES
1. The authority citation for 21 CFR
part 886 continues to read as follows:
sroberts on DSK5SPTVN1PROD with
■
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. In subpart E, add § 886.4155 to read
as follows:
■
VerDate Mar<15>2010
17:46 Jan 24, 2013
Jkt 229001
Scleral plug.
(a) Identification. A scleral plug is a
prescription device intended to provide
temporary closure of a scleral incision
during an ophthalmic surgical
procedure. These plugs prevent
intraocular fluid and pressure loss when
instruments are withdrawn from the
eye. Scleral plugs include a head
portion remaining above the sclera,
which can be gripped for insertion and
removal, and a shaft that fits inside the
scleral incision. Scleral plugs are
removed before completing the surgery.
(b) Classification. Class II (special
controls). The special controls for the
scleral plug are:
(1) The device is exempt from the
premarket notification procedures in
subpart E of part 807 of this chapter
subject to the limitations in § 886.9 if
the material is a surgical grade stainless
steel with or without a gold, silver, or
titanium coating. The special controls
for the surgical grade stainless steel
scleral plug (with or without a gold,
silver, or titanium coating) are:
(i) The device must be demonstrated
to be sterile during the labeled shelf life;
(ii) The device must be demonstrated
to be biocompatible; and
(iii) Labeling must include all
information required for the safe and
effective use of the device, including
specific instructions regarding the
proper sizing, placement, and removal
of the device.
(2) The device is not exempt from
premarket notification procedures if it is
composed of a material other than
surgical grade stainless steel (with or
without a gold, silver, or titanium
coating). The special controls for scleral
plugs made of other materials are:
(i) The device must be demonstrated
to be sterile during the labeled shelf life;
(ii) The device must be demonstrated
to be biocompatible;
(iii) Characterization of the device
materials must be performed;
(iv) Performance data must
demonstrate acceptable mechanical
properties under simulated clinical use
conditions including insertion and
removal of the device;
(v) Performance data must
demonstrate adequately low levels of
the extractables or residues from
manufacturing (or processing) of the
device; and
(vi) Labeling must include all
information required for the safe and
effective use of the device, including
specific instructions regarding the
proper sizing, placement, and removal
of the device.
PO 00000
Frm 00011
Fmt 4702
Sfmt 4702
Dated: January 17, 2013.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2013–01447 Filed 1–24–13; 8:45 am]
BILLING CODE 4160–01–P
DEPARTMENT OF EDUCATION
34 CFR Chapter III
Proposed Priorities and Definitions—
NIDRR DRRP—Community Living and
Participation, Health and Function, and
Employment of Individuals With
Disabilities
Office of Special Education and
Rehabilitative Services, Department of
Education.
ACTION: Proposed priorities and
definitions.
AGENCY:
CFDA Numbers: 84.133A–3, 84.133A–4, and
84.133A–5.
The Assistant Secretary for
Special Education and Rehabilitative
Services proposes funding priorities and
definitions for the Disability and
Rehabilitation Research Projects and
Centers Program administered by the
National Institute on Disability and
Rehabilitation Research (NIDRR).
Specifically, this document proposes
priorities for a Disability and
Rehabilitation Research Project (DRRP)
on Community Living and Participation
of Individuals with Disabilities
(Proposed Priority 1), a DRRP on Health
and Function of Individuals with
Disabilities (Proposed Priority 2), and a
DRRP on Employment of Individuals
with Disabilities (Proposed Priority 3). If
an applicant proposes to conduct
research under these priorities, the
research must be focused on one of the
four stages of research. This document
proposes definitions for the four stages
of research: exploration and discovery,
intervention development, intervention
efficacy, and scale-up evaluation. The
Assistant Secretary may use one or more
of these priorities and definitions for
competitions in fiscal year (FY) 2013
and later years. We take this action to
focus research attention on areas of
national need. We intend these
priorities and definitions to contribute
to improved employment and
independent living outcomes for
individuals with disabilities.
DATES: We must receive your comments
on or before February 25, 2013.
ADDRESSES: Address all comments about
this document to Marlene Spencer, U.S.
Department of Education, 400 Maryland
Avenue SW., Room 5133, Potomac
SUMMARY:
E:\FR\FM\25JAP1.SGM
25JAP1
Agencies
[Federal Register Volume 78, Number 17 (Friday, January 25, 2013)]
[Proposed Rules]
[Pages 5327-5330]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2013-01447]
[[Page 5327]]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 886
[Docket No. FDA-2012-N-1238]
Medical Devices; Ophthalmic Devices; Classification of the
Scleral Plug
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed rule.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA or Agency) is proposing
to classify the scleral plug into class II (special controls), and
proposing to exempt the scleral plugs composed of surgical grade
stainless steel (with or without coating in gold, silver, or titanium)
from premarket notification (510(k)) and to continue to require
premarket notification (510(k)) for all other scleral plugs in order to
provide a reasonable assurance of safety and effectiveness of the
device. The scleral plug is a prescription device used to provide
temporary closure of a scleral incision during an ophthalmic surgical
procedure.
DATES: Submit either electronic or written comments by April 25, 2013.
See section IV of this document for the proposed effective date of a
final rule that may issue based on this proposal.
ADDRESSES: You may submit comments, identified by Docket No. FDA-2012-
N-1238, by any of the following methods:
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments.
Written Submissions
Submit written submissions in the following way:
Mail/Hand delivery/Courier (for paper or CD-ROM
submissions): Division of Dockets Management (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
Instructions: All submissions received must include the Agency name
and Docket No. FDA-2012-N-1238 for this rulemaking. All comments
received may be posted without change to https://www.regulations.gov,
including any personal information provided. For additional information
on submitting comments, see the ``Comments'' heading of the
SUPPLEMENTARY INFORMATION section of this document.
Docket: For access to the docket to read background documents or
comments received, go to https://www.regulations.gov and insert the
docket number, found in brackets in the heading of this document, into
the ``Search'' box and follow the prompts and/or go to the Division of
Dockets Management, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Tina Kiang, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 2414, Silver Spring, MD 20993-0002, 301-796-6860,
Tina.Kiang@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background
A. Statutory and Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act) (21 U.S.C.
301 et seq.), as amended by the Medical Device Amendments of 1976 (the
1976 amendments) (Pub. L. 94-295), the Safe Medical Devices Act of 1990
(Pub. L. 101-629), the Food and Drug Administration Modernization Act
of 1997 (FDAMA) (Pub. L. 105-115), among other amendments, established
a comprehensive system for the regulation of medical devices intended
for human use. Section 513 of the FD&C Act (21 U.S.C. 360c) established
three categories (classes) of devices, depending on the regulatory
controls needed to provide reasonable assurance of their safety and
effectiveness. The three categories of devices are class I (general
controls), class II (special controls), and class III (premarket
approval).
Under section 513 of the FD&C Act, FDA refers to devices that were
in commercial distribution before May 28, 1976 (the date of enactment
of the 1976 amendments), as ``preamendments devices.'' FDA classifies
these devices after the Agency takes the following steps: (1) Receives
a recommendation from a device classification panel (an FDA advisory
committee); (2) publishes the panel's recommendation for comment, along
with a proposed regulation classifying the device; and (3) publishes a
final regulation classifying the device. FDA has classified most
preamendments devices under these procedures.
FDA refers to devices that were not in commercial distribution
before May 28, 1976, as ``postamendments devices.'' These devices are
classified automatically by statute (section 513(f) of the FD&C Act (21
U.S.C. 360c(f)) into class III without any FDA rulemaking process.
These devices remain in class III and require premarket approval,
unless and until: (1) FDA reclassifies the device into class I or II;
(2) FDA issues an order classifying the device into class I or II in
accordance with section 513(f)(2) of the FD&C Act, as amended by FDAMA;
or (3) FDA issues an order finding the device to be substantially
equivalent, under section 513(i) of the FD&C Act (21 U.S.C. 360c(i)),
to a predicate device that does not require premarket approval. The
Agency determines whether new devices are substantially equivalent to
predicate devices by means of premarket notification procedures in
section 510(k) of the FD&C Act and part 807 of the regulations (21 CFR
Part 807).
A person may market a preamendments device that has been classified
into class III through premarket notification procedures, without
submission of a premarket approval application (PMA) until FDA issues a
final regulation under section 515(b) of the FD&C Act (21 U.S.C.
360e(b)) requiring premarket approval.
Section 510(m) of the FD&C Act provides that a class II device may
be exempted from the premarket notification requirements under section
510(k) of the FD&C Act, if the Agency determines that premarket
notification is not necessary to assure the safety and effectiveness of
the device. FDA has determined that premarket notification is not
necessary to assure the safety and effectiveness of scleral plugs if
the material is a surgical grade stainless steel with or without a
gold, silver, or titanium coating.
B. Regulatory History of the Device
After the enactment of the Medical Device Amendments of 1976, FDA
commenced to identify and classify all preamendments devices, in
accordance with section 513(b) (21 U.S.C. 360c(b)) of the FD&C Act. In
the Federal Register of September 2, 1987 (52 FR 33346), FDA classified
a total of 109 generic types of ophthalmic devices. The scleral plug
was not identified in this initial effort. FDA has regulated scleral
plugs as devices requiring premarket notification (510(k)). Scleral
plugs currently on the market have been determined to be substantially
equivalent to devices that were in commercial distribution prior to May
28, 1976. Currently, FDA regulates scleral plugs as devices requiring
premarket notification (510(k)). There have been ten 510(k) submissions
received and cleared under product code LXP (scleral plugs).
Consistent with the FD&C Act and the regulations, FDA consulted
with the Ophthalmic Devices Panel (the Panel), an FDA advisory
committee, regarding
[[Page 5328]]
the classification of this device type on January 22, 1996 (Ref. 1). At
the panel meeting, the Panel recommended scleral plugs as
classification as class I, 510(k) exempt. Two 510(k) submissions have
been cleared since the panel meeting.
II. Recommendation of the Panel
During a public meeting which was held on January 22, 1996, the
Panel made recommendations regarding the classification and regulatory
controls for the scleral plug. FDA is proposing the following
identification based on the Panel's recommendations and the Agency's
review:
A. Identification
A scleral plug is a prescription device intended to provide
temporary closure of a scleral incision during an ophthalmic surgical
procedure. These plugs prevent intraocular fluid and pressure loss when
instruments are withdrawn from the eye. Scleral plugs include a head
portion remaining above the sclera, which can be gripped for insertion
and removal, and a shaft that fits inside the scleral incision. Scleral
plugs are removed before completing the surgery. Therefore, they are
generally only used in operating rooms. These devices are often made of
surgical grade stainless steel and can be coated in gold, silver, or
titanium.
Scleral plugs have a long and established history of clinical use.
They are routinely used in many ophthalmic surgeries (specifically,
vitreoretinal surgeries). One common type of vitreoretinal surgery is
vitrectomy. Vitrectomy is estimated to be the third most frequently
performed ophthalmic surgical operation, after cataract and excimer
laser refractive surgery (Ref. 2). Approximately 225,000 vitrectomies
are done in the United States each year (Ref. 2).
B. Recommended Classification of the Panel
Although the Panel was informed that scleral plugs have
historically been treated as class II devices, the Panel recommended
that a scleral plug made of a material previously used in legally
marketed devices be classified into class I (general controls) and be
exempt from premarket notification because the biocompatibility and
ability to be sterilized have already been established. The Panel's
rationale for suggesting that the scleral plug be classified into class
I was because general controls would provide reasonable assurance of
the safety and effectiveness of the device type if it is made from a
material established to be readily sterilized and biocompatible. During
the panel discussion, a distinction was made that scleral plugs
consisting of other materials (i.e., materials that are not already
included in legally marketed medical devices to the date of the
classification regulation) should be classified into class II and
require biocompatibility testing as a special control.
As a result of the distinction between materials used for this
device, the Panel recommended that, unless new materials are proposed,
the device should be exempt from premarket notification.
C. Summary of Reasons for Recommendation
The Panel considered FDA's extensive regulatory experience with the
device type and the Panel members' personal knowledge of and clinical
experience with the device type. The Panel also considered the long
history of safety and effectiveness of the device over many years of
clinical use. The Panel recommended that the scleral plug be classified
into class I because it concluded that general controls would provide
reasonable assurance of the safety and effectiveness of the device type
if it was made from a material established to be readily sterilized and
biocompatible. The Panel also recommended that scleral plugs be exempt
from premarket notification requirements if the proposed device does
not introduce new materials (i.e., materials that are not established
to be safe for this type of application).
However, FDA believes that a class II classification is appropriate
and consistent with the intent of the Panel to establish requirements
(such as biocompatibility and sterility) for these devices. Although
the Panel identified potential risks and the measures that could be
taken to mitigate these risks, the Panel's recommendation of class I
would not permit FDA to establish as special controls the mitigation
measures discussed (biocompatibility, sterility). Therefore, while FDA
is not adopting the Panel's recommendation of classification into class
I, the Agency agrees with the concerns and mitigation measures
discussed by the Panel that would support a classification under class
II.
D. Risks to Health
Based on the Panel's discussion and recommendations and FDA's
experience with the device, the risks to health associated with the
scleral plugs made from surgical grade stainless steel with or without
gold, silver, or titanium coating and the proposed measures to mitigate
these risks are identified in table 1 of this document.
Table 1--Health Risks and Mitigation Measures for the Scleral Plug Made
From Surgical Grade Stainless Steel
[With or without a gold, silver, or titanium coating]
------------------------------------------------------------------------
Identified risk Mitigation measures
------------------------------------------------------------------------
Infection................................. Sterility Testing.
Adverse Tissue Reaction................... Biocompatibility Testing.
Loss, breakage, or migration of the plug.. Labeling.
------------------------------------------------------------------------
For scleral plugs that are made of surgical grade stainless steel
(with or without a gold, silver, or titanium coating) the following
special controls, in addition to general controls, can address the
risks to health in table 1 of this document and provide reasonable
assurance of safety and effectiveness of the device: (1) Performance
data must demonstrate the sterility and shelf life of the device; (2)
the device must be demonstrated to be biocompatible; and (3) labeling
must include all information required for the safe and effective use of
the device, including specific instructions regarding the proper
incision size, placement, and removal of the device.
Because of the varying properties of other materials and the
potential impact on safety and effectiveness, FDA has identified
additional special controls for devices made of materials other than
surgical grade stainless steel. Based on the Panel's discussion and
recommendations and FDA's experience with the device, the risks to
health associated with the scleral plugs made from materials other than
surgical grade stainless steel and the proposed measures to mitigate
these risks are identified in table 2 of this document.
Table 2--Health Risks and Mitigation Measures for Scleral Plugs Made
From Materials Other than Surgical Grade Stainless Steel
------------------------------------------------------------------------
Identified risk Mitigation measures
------------------------------------------------------------------------
Infection................................. Sterility Testing.
Shelf-life Testing.
Adverse Tissue Reaction................... Biocompatibility testing.
Material characterization.
Performance testing to
determine the level of
extractables.
[[Page 5329]]
Loss, breakage, or migration of the plug.. Material characterization.
Performance testing for
Mechanical Properties.
Labeling.
------------------------------------------------------------------------
The Agency believes that the following special controls, in
addition to general controls, will provide reasonable assurance of
safety and effectiveness for scleral plugs that are composed of a
material other than surgical grade stainless steel, as outlined in
table 2: (1) Performance data must demonstrate the sterility and shelf
life of the device; (2) the device must be demonstrated to be
biocompatible; (3) characterization of the device materials must be
performed; (4) performance data must demonstrate acceptable mechanical
properties under simulated clinical use conditions including insertion
and removal of the device; (5) performance data must demonstrate
adequately low levels of the extractables or residues from
manufacturing (or processing) of the device; and (6) labeling must
include all information required for the safe and effective use of the
device, including specific instructions regarding the proper incision
size, placement, and removal of the device. In addition, the scleral
plug is a prescription device and must be used in accordance with 21
CFR 801.109.
III. Proposed Classification and FDA's Finding
Adverse events involving scleral plugs are rare, as evidenced by
the fact that FDA identified only a single adverse event in our
reporting systems and two adverse events in the published literature
(Refs. 3 and 4). The one adverse event reported to FDA resulted in no
persistent adverse effects to the patient and, according to the report,
this specific type of non-metallic scleral plug was discontinued and
replaced with surgical grade stainless steel scleral plugs.
FDA believes that a class II classification is consistent with the
intent of the Panel to establish requirements (such as biocompatibility
and sterility) for these devices. The identified special controls
mitigate the known risks of the device that were identified by the
Panel. However, the FDA does not agree with the Panel that all
materials included in legally marketed scleral devices can be exempted
from 510(k) due to the potential for safety concerns in some materials
that will require specific material information and performance data to
provide a reasonable assurance of safety and effectiveness. FDA
believes this type of information should be reviewed by FDA prior to a
device being marketed in the United States.
FDA proposes the scleral plug be classified into class II. The
special controls, in addition to general controls, will provide
reasonable assurance of the safety and effectiveness of the device. FDA
also agrees, in part, with the Panel's recommendation that premarket
notification is not necessary to assure the safety and effectiveness of
scleral plugs if new materials are not introduced and, therefore, the
Agency is giving notice of intent to exempt the scleral plug device
from premarket notification requirements if the device is made from
surgical grade stainless steel (with or without a gold, silver, or
titanium coating).
IV. Proposed Effective Date
FDA proposes that any final regulation based on this proposal
become effective 30 days after its date of publication in the Federal
Register.
V. Environmental Impact
The Agency has determined under 21 CFR 25.34(b) that this action is
of a type that does not individually or cumulatively have a significant
effect on the human environment. Therefore, neither an environmental
assessment nor an environmental impact statement is required.
VI. Analysis of Impacts
FDA has examined the impacts of the proposed rule under Executive
Order 12866 and Executive Order 13563, the Regulatory Flexibility Act
(5 U.S.C. 601-612), and the Unfunded Mandates Reform Act of 1995 (Pub.
L. 104-4). Executive Orders 12866 and 13563 direct Agencies to assess
all costs and benefits of available regulatory alternatives and, when
regulation is necessary, to select regulatory approaches that maximize
net benefits (including potential economic, environmental, public
health and safety, and other advantages; distributive impacts; and
equity). The Agency believes that this proposed rule is not a
significant regulatory action as defined by Executive Order 12866.
The Regulatory Flexibility Act requires Agencies to analyze
regulatory options that would minimize any significant impact of a rule
on small entities. Because the proposed regulation will classify a
previously unclassified pre-Amendment device type, there are only five
registered establishments listed in the Establishment Registration and
Device Listing database, and the proposed regulation designating the
classification of scleral plugs as class II is consistent with the
historical regulatory oversight given to this device type, the Agency
proposes to certify that the final rule will not have a significant
economic impact on a substantial number of small entities.
Section 202(a) of the Unfunded Mandates Reform Act of 1995 requires
that Agencies prepare a written statement, which includes an assessment
of anticipated costs and benefits, before proposing ``any rule that
includes any Federal mandate that may result in the expenditure by
State, local, and tribal governments, in the aggregate, or by the
private sector, of $100,000,000 or more (adjusted annually for
inflation) in any one year.'' The current threshold after adjustment
for inflation is $139 million, using the most current (2011) Implicit
Price Deflator for the Gross Domestic Product. FDA does not expect this
proposed rule to result in any 1-year expenditure that would meet or
exceed this amount.
The proposed rule would impact current manufacturers if they were
to make changes to their existing products and any manufacture wanting
to market a new scleral plug. If the new or changed product is made of
surgical grade stainless steel with or without gold, silver, or
titanium coating, manufacturers could begin marketing after they
complied with the proposed special controls. They would not need to
submit an application to the Agency for preapproval. There would be no
change from current requirements for new products made of alternative
materials; they would need premarket notification before marketing.
VII. Paperwork Reduction Act of 1995
This proposed rule establishes special controls that refer to
currently approved collections of information found in other FDA
regulations. These collections of information are subject to review by
the Office of Management and Budget (OMB) under the Paperwork Reduction
Act of 1995 (44 U.S.C. 3501-3520). The collections of information in 21
CFR part 807, subpart E, have been approved under OMB control number
0910-0120; the collections of information in 21 CFR part 801 have been
approved under OMB control
[[Page 5330]]
number 0910-0485; the collections of information in 21 CFR part 807
have been approved under OMB control number 0910-0387.
VIII. Comments
Interested persons may submit to the Division of Dockets Management
(see ADDRESSES) either electronic or written comments regarding this
document. It is only necessary to send one set of comments. Identify
comments with the docket number found in brackets in the heading of
this document. Received comments may be seen in the Division of Dockets
Management between 9 a.m. and 4 p.m., Monday through Friday.
IX. References
The following references have been placed on display in the
Division of Dockets Management (see ADDRESSES) and may be seen by
interested persons between 9 a.m. and 4 p.m., Monday through Friday and
are available electronically at https://www.regulations.gov. (FDA has
verified the Web site addresses of the following references, but FDA is
not responsible for any subsequent changes to the Web site after this
document publishes in the Federal Register.)
1. Transcript from the Food and Drug Administration Ophthalmic
Devices Panel Meeting, January 22, 1996.
2. Chang, Stanley, ``LXII Edward Jackson Lecture: Open Angle
Glaucoma After Vitrectomy,'' American Journal of Ophthalmology, vol.
141(6): pp. 1033-1043, June 2006, available at https://www.sciencedirect.com/science/article/pii/S0002939406002546.
3. Stewart, M. W., ``Intraoperative Radiographic Detection of a
`Lost' Scleral Plug,'' Retina, vol. 25(4): pp. 526-527, June 2005.
4. Bovino, J. A. and D. F. Marcus, ``Intraocular Foreign-Body
Hazard During Vitrectomy,'' American Journal of Ophthalmology, vol.
93 (3): p. 366, March 1982.
List of Subjects in 21 CFR Part 886
Medical devices, Ophthalmic goods and services.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, FDA proposes
to amend part 886 as follows:
PART 886--OPHTHALMIC DEVICES
0
1. The authority citation for 21 CFR part 886 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. In subpart E, add Sec. 886.4155 to read as follows:
Sec. 886.4155 Scleral plug.
(a) Identification. A scleral plug is a prescription device
intended to provide temporary closure of a scleral incision during an
ophthalmic surgical procedure. These plugs prevent intraocular fluid
and pressure loss when instruments are withdrawn from the eye. Scleral
plugs include a head portion remaining above the sclera, which can be
gripped for insertion and removal, and a shaft that fits inside the
scleral incision. Scleral plugs are removed before completing the
surgery.
(b) Classification. Class II (special controls). The special
controls for the scleral plug are:
(1) The device is exempt from the premarket notification procedures
in subpart E of part 807 of this chapter subject to the limitations in
Sec. 886.9 if the material is a surgical grade stainless steel with or
without a gold, silver, or titanium coating. The special controls for
the surgical grade stainless steel scleral plug (with or without a
gold, silver, or titanium coating) are:
(i) The device must be demonstrated to be sterile during the
labeled shelf life;
(ii) The device must be demonstrated to be biocompatible; and
(iii) Labeling must include all information required for the safe
and effective use of the device, including specific instructions
regarding the proper sizing, placement, and removal of the device.
(2) The device is not exempt from premarket notification procedures
if it is composed of a material other than surgical grade stainless
steel (with or without a gold, silver, or titanium coating). The
special controls for scleral plugs made of other materials are:
(i) The device must be demonstrated to be sterile during the
labeled shelf life;
(ii) The device must be demonstrated to be biocompatible;
(iii) Characterization of the device materials must be performed;
(iv) Performance data must demonstrate acceptable mechanical
properties under simulated clinical use conditions including insertion
and removal of the device;
(v) Performance data must demonstrate adequately low levels of the
extractables or residues from manufacturing (or processing) of the
device; and
(vi) Labeling must include all information required for the safe
and effective use of the device, including specific instructions
regarding the proper sizing, placement, and removal of the device.
Dated: January 17, 2013.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2013-01447 Filed 1-24-13; 8:45 am]
BILLING CODE 4160-01-P