Dental Devices; Reclassification of Blade-Form Endosseous Dental Implant, 2647-2650 [2013-00388]
Download as PDF
Federal Register / Vol. 78, No. 9 / Monday, January 14, 2013 / Proposed Rules
‘‘Proposed Modification of Class B
Airspace; Las Vegas, NV’’ (77 FR 65332).
The FAA requested that comments on
that proposal be received on or before
December 26, 2012. By letter dated
December 7, 2012, the Aircraft Owners
and Pilots Association (AOPA)
requested that the FAA extend the
comment period for at least 30 days.
AOPA stated that the original comment
period encompassed two Federal
holidays and that no comments had
been posted to the docket as of the date
of their letter. AOPA added that an
extension would provide additional
time for the public to review the NPRM
and submit substantive comments on
the proposal.
Reopening of Comment Period
The FAA has reviewed AOPA’s
request for additional time to comment
on the NPRM and has determined that
reopening of the comment period is
consistent with the public interest and
that good cause exists for taking this
action.
Accordingly, the comment period for
Docket No. FAA–2012–0966; Airspace
Docket No. 12–AWA–5, is reopened as
indicated in the DATES section, above.
Issued in Washington, DC, on January 8,
2013.
Gary A. Norek,
Manager, Airspace Policy and ATC
Procedures Group.
[FR Doc. 2013–00646 Filed 1–10–13; 4:15 pm]
BILLING CODE P
DEPARTMENT OF HEALTH AND
HUMAN SERVICES
Food and Drug Administration
21 CFR Part 872
[Docket No. FDA–2012–N–0677]
Dental Devices; Reclassification of
Blade-Form Endosseous Dental
Implant
AGENCY:
Food and Drug Administration,
Proposed order.
The Food and Drug
Administration (FDA) is proposing to
reclassify the blade- form endosseous
dental implant, a preamendments class
III device, into class II (special controls).
On its own initiative, based on new
information, FDA is proposing to revise
the classification of blade-form
endosseous dental implants.
DATES: Submit either electronic or
written comments on this proposed
order by April 15, 2013. See section XI
srobinson on DSK4SPTVN1PROD with
SUMMARY:
VerDate Mar<15>2010
16:19 Jan 11, 2013
Jkt 229001
Electronic Submissions
Submit electronic comments in the
following way:
• Federal eRulemaking Portal: https://
www.regulations.gov. Follow the
instructions for submitting comments.
Written Submissions
Submit written submissions in the
following ways:
• Mail/Hand delivery/Courier (for
paper or CD–ROM submissions):
Division of Dockets Management (HFA–
305), Food and Drug Administration,
5630 Fishers Lane, Rm. 1061, Rockville,
MD 20852.
Instructions: All submissions received
must include the Agency name and
Docket No. FDA–2012–N–0677 for this
rulemaking. All comments received may
be posted without change to https://
www.regulations.gov, including any
personal information provided. For
additional information on submitting
comments, see the ‘‘Comments’’ heading
of the SUPPLEMENTARY INFORMATION
section of this document.
Docket: For access to the docket to
read background documents or
comments received, go to https://
www.regulations.gov and insert Docket
No. FDA–2012–N–0677 into the
‘‘Search’’ box and follow the prompts
and/or go to the Division of Dockets
Management, 5630 Fishers Lane, Rm.
1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT:
Melissa Burns, Center for Devices and
Radiological Health, Food and Drug
Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1646, Silver Spring,
MD 20993, 301–796–5616,
melissa.burns@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
HHS.
ACTION:
of this document for the proposed
effective date of a final order based on
this proposed order.
ADDRESSES: You may submit comments,
identified by Docket No. FDA–2012–N–
0677, by any of the following methods:
I. Background—Regulatory Authorities
The Federal Food, Drug, and Cosmetic
Act (the FD&C Act), as amended by the
Medical Device Amendments of 1976
(the 1976 amendments) (Pub. L. 94–
295), the Safe Medical Devices Act of
1990 (Pub. L. 101–629), the Food and
Drug Administration Modernization Act
of 1997 (FDAMA) (Pub. L. 105–115), the
Medical Device User Fee and
Modernization Act of 2002 (Pub. L. 107–
250), the Medical Devices Technical
Corrections Act (Public Law 108–214),
the Food and Drug Administration
Amendments Act of 2007 (Pub. L. 110–
85), and the Food and Drug
PO 00000
Frm 00021
Fmt 4702
Sfmt 4702
2647
Administration Safety and Innovation
Act (FDASIA) (Pub. L. 112–144),
establish a comprehensive system for
the regulation of medical devices
intended for human use. Section 513 of
the FD&C Act (21 U.S.C. 360c)
established three categories (classes) of
devices, reflecting the regulatory
controls needed to provide reasonable
assurance of their safety and
effectiveness. The three categories of
devices are class I (general controls),
class II (special controls), and class III
(premarket approval).
Under section 513 of the FD&C Act,
devices that were in commercial
distribution before the enactment of the
1976 amendments, May 28, 1976
(generally referred to as preamendments
devices), are classified after FDA has: (1)
Received a recommendation from a
device classification panel (an FDA
advisory committee); (2) published the
panel’s recommendation for comment,
along with a proposed regulation
classifying the device; and (3) published
a final regulation classifying the device.
FDA has classified most
preamendments devices under these
procedures.
Devices that were not in commercial
distribution prior to May 28, 1976
(generally referred to as
postamendments devices), are
automatically classified by section
513(f) of the FD&C Act into class III
without any FDA rulemaking process.
Those devices remain in class III and
require premarket approval unless, and
until, the device is reclassified into class
I or II or FDA issues an order finding the
device to be substantially equivalent, in
accordance with section 513(i) of the
FD&C Act, to a predicate device that
does not require premarket approval.
The Agency determines whether new
devices are substantially equivalent to
predicate devices by means of
premarket notification procedures in
section 510(k) of the FD&C Act (21
U.S.C. 360(k)) and 21 CFR part 807.
A preamendments device that has
been classified into class III may be
marketed by means of premarket
notification procedures (510(k) process)
without submission of a premarket
approval application (PMA) until FDA
issues a final order under section 515(b)
of the FD&C Act (21 U.S.C. 360e(b))
requiring premarket approval or until
the device is subsequently reclassified
into class I or class II.
On July 9, 2012, FDASIA was enacted.
Section 608(a) of FDASIA (126 Stat.
1056) amended section 513(e) of the
FD&C Act changing the process for
reclassifying a preamendments device
from rulemaking to an administrative
order.
E:\FR\FM\14JAP1.SGM
14JAP1
srobinson on DSK4SPTVN1PROD with
2648
Federal Register / Vol. 78, No. 9 / Monday, January 14, 2013 / Proposed Rules
Section 513(e) of the FD&C Act
governs reclassification of classified
preamendments devices. This section
provides that FDA may, by
administrative order, reclassify a device
based upon ‘‘new information.’’ FDA
can initiate a reclassification under
section 513(e) of the FD&C Act or an
interested person may petition FDA to
reclassify a preamendments device. The
term ‘‘new information,’’ as used in
section 513(e) of the FD&C Act, includes
information developed as a result of a
reevaluation of the data before the
Agency when the device was originally
classified, as well as information not
presented, not available, or not
developed at that time. (See, e.g.,
Holland Rantos v. United States
Department of Health, Education, and
Welfare, 587 F.2d 1173, 1174 n.1 (D.C.
Cir. 1978); Upjohn v. Finch, 422 F.2d
944 (6th Cir. 1970); Bell v. Goddard, 366
F.2d 177 (7th Cir. 1966).)
Reevaluation of the data previously
before the Agency is an appropriate
basis for subsequent regulatory action
where the reevaluation is made in light
of newly available regulatory authority
(see Bell v. Goddard, supra, 366 F.2d at
181; Ethicon, Inc. v. FDA, 762 F.Supp.
382, 389–391 (D.D.C. 1991)) or in light
of changes in ‘‘medical science.’’ (See
Upjohn v. Finch, supra, 422 F.2d at
951). Whether data before the Agency
are past or new data, the ‘‘new
information’’ to support reclassification
under section 513(e) of the FD&C Act
must be ‘‘valid scientific evidence,’’ as
defined in section 513(a)(3) and 21 CFR
860.7(c)(2). (See, e.g., General Medical
Co. v. FDA, 770 F.2d 214 (D.C. Cir.
1985); Contact Lens Assoc. v. FDA, 766
F.2d 592 (D.C. Cir.), cert. denied, 474
U.S. 1062 (1985).)
FDA relies upon ‘‘valid scientific
evidence’’ in the classification process
to determine the level of regulation for
devices. To be considered in the
reclassification process, the valid
scientific evidence upon which the
Agency relies must be publicly
available. Publicly available information
excludes trade secret and/or
confidential commercial information,
e.g., the contents of a pending PMA.
(See section 520(c) of the FD&C Act (21
U.S.C. 360j(c)).) Section 520(h)(4) of the
FD&C Act, added by FDAMA, provides
that FDA may use, for reclassification of
a device, certain information in a PMA
6 years after the application has been
approved. This includes information
from clinical and preclinical tests or
studies that demonstrate the safety or
effectiveness of the device but does not
include descriptions of methods of
manufacture or product composition
and other trade secrets.
VerDate Mar<15>2010
16:19 Jan 11, 2013
Jkt 229001
Section 513(e)(1) of the FD&C Act sets
forth the process for issuing a final
order. Specifically, prior to the issuance
of a final order reclassifying a device,
the following must occur: Publication of
a proposed order in the Federal
Register, a meeting of a device
classification panel described in section
513(b) of the FD&C Act; and
consideration of comments to a public
docket.
FDAMA added a new section 510(m)
to the FD&C Act. Section 510(m) of the
FD&C Act provides that a class II device
may be exempted from the premarket
notification requirements under section
510(k) of the FD&C Act if the Agency
determines that premarket notification
is not necessary to assure the safety and
effectiveness of the device.
II. Regulatory History of the Device
On December 30, 1980 (45 FR 86025),
FDA published a proposed rule for
classification of endosseous dental
implants (without distinguishing
implants based on geometry) as class III
requiring premarket approval. The panel
recommended class III because the
device is implanted in the body and
presents a potential unreasonable risk of
illness or injury including risks of
abnormal spontaneous pain due to
nerve impingement and a risk of
perforation of the lingual and labial
bony plates of the upper and lower jaws.
On August 12, 1987 (52 FR 30082), a
final rule was published for endosseous
dental implants (again without
distinguishing implants based on
geometry) classifying these devices as
class III. On December 7, 1989 (54 FR
50592), FDA published a proposed rule
to require PMA submissions for all
dental implants. A reclassification
petition was subsequently submitted
requesting reclassification of dental
implants.
FDA held a reclassification panel
meeting on October 24, 1991, and the
panel voted to deny the reclassification
petition. At the request of FDA,
additional panel meetings were held on
November 4, 1997, and January 13,
1998, during which FDA presented new
information regarding root-form
endosseous dental implants. During the
January 1998 panel meeting, the panel
stated that sufficient clinical
information was presented to the panel
to justify reclassification of root-form
implants, implants with special
retention features, and temporary
implants, as class II (special controls)
requiring a 510(k) premarket
notification. However, the panel also
stated that sufficient evidence had not
yet been presented to reclassify blade-
PO 00000
Frm 00022
Fmt 4702
Sfmt 4702
form endosseous dental implants to
class II.
On May 14, 2002 (67 FR 34416), and
May 12, 2004 (69 FR 26302), proposed
and final rules respectively were issued
reclassifying only root-form implants
into class II. Blade-form endosseous
dental implant remained class III.
In 2009, FDA published an order
under sections 515(i) and 519 of the
FD&C Act (21 U.S.C. 360i) for the
submission of information on bladeform endosseous dental implants (74 FR
16214, April 9, 2009). In response to
that order, FDA received information
from one device manufacturer; however,
the information was related to other
types of dental implants and was not
relevant for this proposed rule.
III. Device Description
The blade-form endosseous dental
implant is a device placed into the
maxilla or mandible and composed of
biocompatible material, such as
titanium alloy or commercially pure
(c.p.) titanium, with sufficient strength
to support a dental restoration, such as
a crown, bridge, or denture, intended for
the purpose of replacing tooth (or teeth)
roots and extending a support post
through the gingival tissue into the oral
cavity to restore chewing function. The
blade-form implants are either one-piece
or two-piece implants designed with
one to three cylindrical abutment posts
extending from the coronal aspect of the
blade through the soft tissue and into
the oral cavity. For the two-piece
design, the separate abutment post is
retained to the blade implant with a
screw.
The blade-form implant is generally a
rectangular shape or rounded corner
rectangle shape (in the mesio-distal
plane) with a narrow tapered (narrow at
the apical edge) edge (in the buccolingual plane) similar in shape to a razor
blade. Other blade designs, such as
square, V-shaped, and triangles have
also been used. The blade generally
contains open vents of various shapes
and various sizes.
IV. Proposed Reclassification
FDA is proposing that the device
subject to this proposal be reclassified
from class III to class II. In this proposed
order, the Agency has identified special
controls under section 513(a)(1)(B) of
the FD&C Act that, together with general
controls applicable to the devices,
would provide reasonable assurance of
their safety and effectiveness. FDA
believes that the identified special
controls in this proposed order, if
finalized, together with general controls
applicable to the device, would provide
reasonable assurance of safety and
E:\FR\FM\14JAP1.SGM
14JAP1
Federal Register / Vol. 78, No. 9 / Monday, January 14, 2013 / Proposed Rules
srobinson on DSK4SPTVN1PROD with
effectiveness. Absent the special
controls identified in this proposed
order, general controls applicable to the
device are insufficient to provide
reasonable assurance of the safety and
effectiveness of the device.
Therefore, in accordance with
sections 513(e) and 515(i) of the FD&C
Act and 21 CFR 860.130, based on new
information with respect to the devices
and taking into account the public
health benefit of the use of the device
and the nature and known incidence of
the risk of the device, FDA, on its own
initiative, is proposing to reclassify this
preamendments class III device into
class II. FDA believes that this new
information is sufficient to demonstrate
that the proposed special controls can
effectively mitigate the risks to health
identified in the next section, and that
these special controls, together with
general controls, will provide a
reasonable assurance of safety and
effectiveness for blade-form endosseous
dental implant devices.
FDA has also considered blade-form
endosseous dental implant devices in
accordance with the reserved criteria set
forth in section 513(a) of the FD&C Act
and decided that the device does require
premarket notification. The Agency
does not intend to exempt this proposed
class II device from premarket
notification (510(k)) submission as
allowed under section 510(m) of the
FD&C Act.
V. Risks to Health
After considering available
information, including the
recommendations of the advisory
committees (panels) for the
classification of these devices, FDA has
evaluated the risks to health associated
with the use of blade-form endosseous
dental implant devices and determined
that the following risks to health are
associated with its use:
• Local tissue or existing dentition
degeneration: Localized tissue and
existing dentition degeneration may be
caused by endosseous implants due to
excessive mobility, loss of integration,
incompatibility of device components,
or structural failure of the device.
• Pain: Nerve impingement by the
device may cause pain.
• Bone or nerve damage: Improper
design or use of the device may cause
injury during surgery related to sinus
perforation, alveolar plate perforation,
or nerve damage resulting in transient or
chronic pain/facial nerve paresis.
• Infection: Implantable devices may
introduce microorganisms that may
cause local or systemic infections.
• Adverse tissue reaction: Inadequate
tissue compatibility of the materials
VerDate Mar<15>2010
16:19 Jan 11, 2013
Jkt 229001
used in this device could cause an
immune reaction.
• Migration or thermal injury:
Incompatibility with magnetic
resonance imaging may cause the device
to migrate or heat.
VI. Summary of Reasons for
Reclassification
If properly manufactured and used,
blade-form endosseous dental implants
can help restore the patient’s chewing
function by replacing tooth roots and
extending a support post through the
gingival tissue into the oral cavity in
order to support a dental restoration,
such as a crown, bridge, or denture.
FDA believes that blade-form
endosseous dental implant devices
should be reclassified into class II
because special controls, together with
general controls, can be established to
provide reasonable assurance of the
safety and effectiveness of the device,
and because general controls themselves
are insufficient to provide reasonable
assurance of its safety and effectiveness.
In addition, there is now adequate
effectiveness information sufficient to
establish special controls to provide
such assurance.
VII. Summary of Data Upon Which the
Reclassification Is Based
Since the time of the panel
recommendation, sufficient evidence
has been developed to support a
reclassification of blade-form
endosseous dental implants to class II
with special controls. FDA has been
reviewing these devices for many years
and their risks are well known. A
review of the applicable clinical
literature indicates that the device has a
high success rate (remaining implanted/
not removed) and that few relevant
adverse events have been reported in
the case of these devices or related
devices suggesting that the device has a
high long-term safety profile. FDA
believes that the special controls
identified in this proposed order, if
finalized, together with general controls,
can provide a reasonable assurance of
the safety and effectiveness of bladeform endosseous dental implants.
VIII. Proposed Special Controls
FDA believes that the following
special controls, together with general
controls, are sufficient to mitigate the
risks to health described in section V of
this document:
• The design characteristics of the
device must ensure that the geometry
and material composition are consistent
with the intended use;
• Mechanical performance (fatigue)
testing under simulated physiological
PO 00000
Frm 00023
Fmt 4702
Sfmt 4702
2649
conditions to demonstrate maximum
load (endurance limit) when the device
is subjected to compressive and shear
loads;
• Corrosion testing under simulated
physiological conditions to demonstrate
corrosion potential of each metal or
alloy, couple potential for an assembled
dissimilar metal implant system, and
corrosion rate for an assembled
dissimilar metal implant system;
• The device must be demonstrated to
be biocompatible;
• Sterility testing must demonstrate
the sterility of the device;
• Performance testing to evaluate the
compatibility of the device in a
magnetic resonance (MR) environment;
• Labeling must include a clear
description of the technological
features, how the device should be used
in patients, detailed surgical protocol
and restoration procedures, and relevant
precautions and warnings based on the
clinical use of the device;
• Patient labeling must contain a
description of how the devices works,
how the device is placed, how the
patient needs to care for the implant,
possible adverse events and how to
report any complications; and
• Documented clinical experience
must demonstrate safe and effective use
and capture any adverse events
observed during clinical use.
Blade-form endosseous dental
implants are prescription devices
restricted to patient use only upon the
authorization of a practitioner licensed
by law to administer or use the device.
(Proposed 21 CFR 872.3640(a); see
section 520(e) of the FD&C Act and 21
CFR 801.109 (Prescription devices)).
Prescription-use restrictions are a type
of general controls authorized under
section 520(e) of the FD&C Act and
defined as a general control in section
513(a)(1)(A)(i) of the FD&C Act; and
under 21 CFR 807.81, the device would
continue to be subject to 510(k)
notification requirements.
IX. Environmental Impact
The Agency has determined under 21
CFR 25.34(b)) that this action is of a
type that does not individually or
cumulatively have a significant effect on
the human environment. Therefore,
neither an environmental assessment
nor an environmental impact statement
is required.
X. Paperwork Reduction Act of 1995
This proposed rule refers to
previously approved collections of
information found in FDA regulations.
These collections of information are
subject to review by the Office of
Management and Budget (OMB) under
E:\FR\FM\14JAP1.SGM
14JAP1
2650
Federal Register / Vol. 78, No. 9 / Monday, January 14, 2013 / Proposed Rules
the Paperwork Reduction Act of 1995
(44 U.S.C. 3501–3520). The collections
of information in 21 CFR part 812 have
been approved under OMB control
number 0910–0078; the collections of
information in part 807, subpart E, have
been approved under OMB control
number 0910–0120; the collections of
information in 21 CFR part 814, subpart
B, have been approved under OMB
control number 0910–0231; and the
collections of information under 21 CFR
part 801 have been approved under
OMB control number 0910–0485.
XI. Proposed Effective Date
FDA is proposing that any final order
based on this proposal become effective
on the date of its publication in the
Federal Register or at a later date if
stated in the final order.
XII. Comments
Interested persons may submit either
electronic comments regarding this
document to https://www.regulations.gov
or written comments to the Division of
Dockets Management (see ADDRESSES). It
is only necessary to send one set of
comments. Identify comments with the
docket number found in brackets in the
heading of this document. Received
comments may be seen in the Division
of Dockets Management between 9 a.m.
and 4 p.m., Monday through Friday, and
will be posted to the docket at https://
www.regulations.gov.
List of Subjects in 21 CFR Part 872
Medical devices.
Therefore, under the Federal Food,
Drug, and Cosmetic Act and under
authority delegated to the Commissioner
of Food and Drugs, it is proposed that
21 CFR part 872 be amended as follows:
PART 872—DENTAL DEVICES
BILLING CODE 4160–01–P
Authority: 21 U.S.C. 351, 360, 360c, 360e,
360j, 371.
2. Section 872.3640 is amended by
revising paragraphs (a) and (b)(2) to read
as follows:
■
srobinson on DSK4SPTVN1PROD with
Endosseous dental implant.
(a) Identification. An endosseous
dental implant is a prescription device
made of a material such as titanium or
titanium alloy that is intended to be
surgically placed in the bone of the
upper or lower jaw arches to provide
support for prosthetic devices, such as
artificial teeth, in order to restore a
patient’s chewing function.
(b) * * *
(2) Class II (special controls). The
device is classified as class II if it is a
VerDate Mar<15>2010
16:19 Jan 11, 2013
Jkt 229001
Dated: January 4, 2013.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2013–00388 Filed 1–11–13; 8:45 am]
1. The authority citation for 21 CFR
part 872 continues to read as follows:
■
§ 872.3640
blade-form endosseous dental implant.
The special controls for this device are:
(i) The design characteristics of the
device must ensure that the geometry
and material composition are consistent
with the intended use;
(ii) Mechanical performance (fatigue)
testing under simulated physiological
conditions to demonstrate maximum
load (endurance limit) when the device
is subjected to compressive and shear
loads;
(iii) Corrosion testing under simulated
physiological conditions to demonstrate
corrosion potential of each metal or
alloy, couple potential for an assembled
dissimilar metal implant system, and
corrosion rate for an assembled
dissimilar metal implant system;
(iv) The device must be demonstrated
to be biocompatible;
(v) Sterility testing must demonstrate
the sterility of the device;
(vi) Performance testing to evaluate
the compatibility of the device in a
magnetic resonance (MR) environment;
(vii) Labeling must include a clear
description of the technological
features, how the device should be used
in patients, detailed surgical protocol
and restoration procedures, and relevant
precautions and warnings based on the
clinical use of the device;
(viii) Patient labeling must contain a
description of how the devices works,
how the device is placed, how the
patient needs to care for the implant,
possible adverse events and how to
report any complications; and
(ix) Documented clinical experience
must demonstrate safe and effective use
and capture any adverse events
observed during clinical use.
DEPARTMENT OF HOMELAND
SECURITY
Coast Guard
[Docket Number USCG–2012–1082]
RIN 1625–AA00
Safety Zone, Atlantic Intracoastal
Waterway; Wrightsville Beach, NC
Coast Guard, DHS.
Notice of Proposed Rulemaking.
AGENCY:
The Coast Guard proposes to
extend the temporary safety zone
SUMMARY:
PO 00000
Frm 00024
Fmt 4702
Table of Acronyms
DHS Department of Homeland Security
FR Federal Register
NPRM Notice of Proposed Rulemaking
33 CFR Part 165
ACTION:
established on the waters of the Atlantic
Intracoastal Waterway at Wrightsville
Beach, North Carolina. The safety zone
is necessary to provide for the safety of
mariners on navigable waters during
maintenance on the US 74/76 Bascule
Bridge crossing the Atlantic Intracoastal
Waterway, mile 283.1, at Wrightsville
Beach, North Carolina. The safety zone
extension will temporarily restrict
vessel movement within the designated
area starting on May 1, 2013 through
July 27, 2013.
DATES: Comments and related material
must be received by the Coast Guard on
or before February 13, 2013.
ADDRESSES: You may submit comments
identified by docket number using any
one of the following methods:
(1) Federal eRulemaking Portal:
https://www.regulations.gov.
(2) Fax: 202–493–2251.
(3) Mail or Delivery: Docket
Management Facility (M–30), U.S.
Department of Transportation, West
Building Ground Floor, Room W12–140,
1200 New Jersey Avenue SE.,
Washington, DC 20590–0001. Deliveries
accepted between 9 a.m. and 5 p.m.,
Monday through Friday, except federal
holidays. The telephone number is 202–
366–9329.
See the ‘‘Public Participation and
Request for Comments’’ portion of the
SUPPLEMENTARY INFORMATION section
below for further instructions on
submitting comments. To avoid
duplication, please use only one of
these three methods.
FOR FURTHER INFORMATION CONTACT: If
you have questions on this proposed
rule, call or email CWO4 Joseph M.
Edge, U.S. Coast Guard Sector North
Carolina; telephone 252–247–4525,
email Joseph.M.Edge@uscg.mil. If you
have questions on viewing or submitting
material to the docket, call Renee V.
Wright, Program Manager, Docket
Operations, telephone (202) 366–9826.
SUPPLEMENTARY INFORMATION:
Sfmt 4702
A. Public Participation and Request for
Comments
We encourage you to participate in
this rulemaking by submitting
comments and related materials. All
comments received will be posted
without change to https://
www.regulations.gov and will include
any personal information you have
provided.
E:\FR\FM\14JAP1.SGM
14JAP1
Agencies
[Federal Register Volume 78, Number 9 (Monday, January 14, 2013)]
[Proposed Rules]
[Pages 2647-2650]
From the Federal Register Online via the Government Printing Office [www.gpo.gov]
[FR Doc No: 2013-00388]
=======================================================================
-----------------------------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
21 CFR Part 872
[Docket No. FDA-2012-N-0677]
Dental Devices; Reclassification of Blade-Form Endosseous Dental
Implant
AGENCY: Food and Drug Administration, HHS.
ACTION: Proposed order.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is proposing to
reclassify the blade- form endosseous dental implant, a preamendments
class III device, into class II (special controls). On its own
initiative, based on new information, FDA is proposing to revise the
classification of blade-form endosseous dental implants.
DATES: Submit either electronic or written comments on this proposed
order by April 15, 2013. See section XI of this document for the
proposed effective date of a final order based on this proposed order.
ADDRESSES: You may submit comments, identified by Docket No. FDA-2012-
N-0677, by any of the following methods:
Electronic Submissions
Submit electronic comments in the following way:
Federal eRulemaking Portal: https://www.regulations.gov.
Follow the instructions for submitting comments.
Written Submissions
Submit written submissions in the following ways:
Mail/Hand delivery/Courier (for paper or CD-ROM
submissions): Division of Dockets Management (HFA-305), Food and Drug
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
Instructions: All submissions received must include the Agency name
and Docket No. FDA-2012-N-0677 for this rulemaking. All comments
received may be posted without change to https://www.regulations.gov,
including any personal information provided. For additional information
on submitting comments, see the ``Comments'' heading of the
SUPPLEMENTARY INFORMATION section of this document.
Docket: For access to the docket to read background documents or
comments received, go to https://www.regulations.gov and insert Docket
No. FDA-2012-N-0677 into the ``Search'' box and follow the prompts and/
or go to the Division of Dockets Management, 5630 Fishers Lane, Rm.
1061, Rockville, MD 20852.
FOR FURTHER INFORMATION CONTACT: Melissa Burns, Center for Devices and
Radiological Health, Food and Drug Administration, 10903 New Hampshire
Ave., Bldg. 66, Rm. 1646, Silver Spring, MD 20993, 301-796-5616,
melissa.burns@fda.hhs.gov.
SUPPLEMENTARY INFORMATION:
I. Background--Regulatory Authorities
The Federal Food, Drug, and Cosmetic Act (the FD&C Act), as amended
by the Medical Device Amendments of 1976 (the 1976 amendments) (Pub. L.
94-295), the Safe Medical Devices Act of 1990 (Pub. L. 101-629), the
Food and Drug Administration Modernization Act of 1997 (FDAMA) (Pub. L.
105-115), the Medical Device User Fee and Modernization Act of 2002
(Pub. L. 107-250), the Medical Devices Technical Corrections Act
(Public Law 108-214), the Food and Drug Administration Amendments Act
of 2007 (Pub. L. 110-85), and the Food and Drug Administration Safety
and Innovation Act (FDASIA) (Pub. L. 112-144), establish a
comprehensive system for the regulation of medical devices intended for
human use. Section 513 of the FD&C Act (21 U.S.C. 360c) established
three categories (classes) of devices, reflecting the regulatory
controls needed to provide reasonable assurance of their safety and
effectiveness. The three categories of devices are class I (general
controls), class II (special controls), and class III (premarket
approval).
Under section 513 of the FD&C Act, devices that were in commercial
distribution before the enactment of the 1976 amendments, May 28, 1976
(generally referred to as preamendments devices), are classified after
FDA has: (1) Received a recommendation from a device classification
panel (an FDA advisory committee); (2) published the panel's
recommendation for comment, along with a proposed regulation
classifying the device; and (3) published a final regulation
classifying the device. FDA has classified most preamendments devices
under these procedures.
Devices that were not in commercial distribution prior to May 28,
1976 (generally referred to as postamendments devices), are
automatically classified by section 513(f) of the FD&C Act into class
III without any FDA rulemaking process. Those devices remain in class
III and require premarket approval unless, and until, the device is
reclassified into class I or II or FDA issues an order finding the
device to be substantially equivalent, in accordance with section
513(i) of the FD&C Act, to a predicate device that does not require
premarket approval. The Agency determines whether new devices are
substantially equivalent to predicate devices by means of premarket
notification procedures in section 510(k) of the FD&C Act (21 U.S.C.
360(k)) and 21 CFR part 807.
A preamendments device that has been classified into class III may
be marketed by means of premarket notification procedures (510(k)
process) without submission of a premarket approval application (PMA)
until FDA issues a final order under section 515(b) of the FD&C Act (21
U.S.C. 360e(b)) requiring premarket approval or until the device is
subsequently reclassified into class I or class II.
On July 9, 2012, FDASIA was enacted. Section 608(a) of FDASIA (126
Stat. 1056) amended section 513(e) of the FD&C Act changing the process
for reclassifying a preamendments device from rulemaking to an
administrative order.
[[Page 2648]]
Section 513(e) of the FD&C Act governs reclassification of
classified preamendments devices. This section provides that FDA may,
by administrative order, reclassify a device based upon ``new
information.'' FDA can initiate a reclassification under section 513(e)
of the FD&C Act or an interested person may petition FDA to reclassify
a preamendments device. The term ``new information,'' as used in
section 513(e) of the FD&C Act, includes information developed as a
result of a reevaluation of the data before the Agency when the device
was originally classified, as well as information not presented, not
available, or not developed at that time. (See, e.g., Holland Rantos v.
United States Department of Health, Education, and Welfare, 587 F.2d
1173, 1174 n.1 (D.C. Cir. 1978); Upjohn v. Finch, 422 F.2d 944 (6th
Cir. 1970); Bell v. Goddard, 366 F.2d 177 (7th Cir. 1966).)
Reevaluation of the data previously before the Agency is an
appropriate basis for subsequent regulatory action where the
reevaluation is made in light of newly available regulatory authority
(see Bell v. Goddard, supra, 366 F.2d at 181; Ethicon, Inc. v. FDA, 762
F.Supp. 382, 389-391 (D.D.C. 1991)) or in light of changes in ``medical
science.'' (See Upjohn v. Finch, supra, 422 F.2d at 951). Whether data
before the Agency are past or new data, the ``new information'' to
support reclassification under section 513(e) of the FD&C Act must be
``valid scientific evidence,'' as defined in section 513(a)(3) and 21
CFR 860.7(c)(2). (See, e.g., General Medical Co. v. FDA, 770 F.2d 214
(D.C. Cir. 1985); Contact Lens Assoc. v. FDA, 766 F.2d 592 (D.C. Cir.),
cert. denied, 474 U.S. 1062 (1985).)
FDA relies upon ``valid scientific evidence'' in the classification
process to determine the level of regulation for devices. To be
considered in the reclassification process, the valid scientific
evidence upon which the Agency relies must be publicly available.
Publicly available information excludes trade secret and/or
confidential commercial information, e.g., the contents of a pending
PMA. (See section 520(c) of the FD&C Act (21 U.S.C. 360j(c)).) Section
520(h)(4) of the FD&C Act, added by FDAMA, provides that FDA may use,
for reclassification of a device, certain information in a PMA 6 years
after the application has been approved. This includes information from
clinical and preclinical tests or studies that demonstrate the safety
or effectiveness of the device but does not include descriptions of
methods of manufacture or product composition and other trade secrets.
Section 513(e)(1) of the FD&C Act sets forth the process for
issuing a final order. Specifically, prior to the issuance of a final
order reclassifying a device, the following must occur: Publication of
a proposed order in the Federal Register, a meeting of a device
classification panel described in section 513(b) of the FD&C Act; and
consideration of comments to a public docket.
FDAMA added a new section 510(m) to the FD&C Act. Section 510(m) of
the FD&C Act provides that a class II device may be exempted from the
premarket notification requirements under section 510(k) of the FD&C
Act if the Agency determines that premarket notification is not
necessary to assure the safety and effectiveness of the device.
II. Regulatory History of the Device
On December 30, 1980 (45 FR 86025), FDA published a proposed rule
for classification of endosseous dental implants (without
distinguishing implants based on geometry) as class III requiring
premarket approval. The panel recommended class III because the device
is implanted in the body and presents a potential unreasonable risk of
illness or injury including risks of abnormal spontaneous pain due to
nerve impingement and a risk of perforation of the lingual and labial
bony plates of the upper and lower jaws. On August 12, 1987 (52 FR
30082), a final rule was published for endosseous dental implants
(again without distinguishing implants based on geometry) classifying
these devices as class III. On December 7, 1989 (54 FR 50592), FDA
published a proposed rule to require PMA submissions for all dental
implants. A reclassification petition was subsequently submitted
requesting reclassification of dental implants.
FDA held a reclassification panel meeting on October 24, 1991, and
the panel voted to deny the reclassification petition. At the request
of FDA, additional panel meetings were held on November 4, 1997, and
January 13, 1998, during which FDA presented new information regarding
root-form endosseous dental implants. During the January 1998 panel
meeting, the panel stated that sufficient clinical information was
presented to the panel to justify reclassification of root-form
implants, implants with special retention features, and temporary
implants, as class II (special controls) requiring a 510(k) premarket
notification. However, the panel also stated that sufficient evidence
had not yet been presented to reclassify blade-form endosseous dental
implants to class II.
On May 14, 2002 (67 FR 34416), and May 12, 2004 (69 FR 26302),
proposed and final rules respectively were issued reclassifying only
root-form implants into class II. Blade-form endosseous dental implant
remained class III.
In 2009, FDA published an order under sections 515(i) and 519 of
the FD&C Act (21 U.S.C. 360i) for the submission of information on
blade-form endosseous dental implants (74 FR 16214, April 9, 2009). In
response to that order, FDA received information from one device
manufacturer; however, the information was related to other types of
dental implants and was not relevant for this proposed rule.
III. Device Description
The blade-form endosseous dental implant is a device placed into
the maxilla or mandible and composed of biocompatible material, such as
titanium alloy or commercially pure (c.p.) titanium, with sufficient
strength to support a dental restoration, such as a crown, bridge, or
denture, intended for the purpose of replacing tooth (or teeth) roots
and extending a support post through the gingival tissue into the oral
cavity to restore chewing function. The blade-form implants are either
one-piece or two-piece implants designed with one to three cylindrical
abutment posts extending from the coronal aspect of the blade through
the soft tissue and into the oral cavity. For the two-piece design, the
separate abutment post is retained to the blade implant with a screw.
The blade-form implant is generally a rectangular shape or rounded
corner rectangle shape (in the mesio-distal plane) with a narrow
tapered (narrow at the apical edge) edge (in the bucco-lingual plane)
similar in shape to a razor blade. Other blade designs, such as square,
V-shaped, and triangles have also been used. The blade generally
contains open vents of various shapes and various sizes.
IV. Proposed Reclassification
FDA is proposing that the device subject to this proposal be
reclassified from class III to class II. In this proposed order, the
Agency has identified special controls under section 513(a)(1)(B) of
the FD&C Act that, together with general controls applicable to the
devices, would provide reasonable assurance of their safety and
effectiveness. FDA believes that the identified special controls in
this proposed order, if finalized, together with general controls
applicable to the device, would provide reasonable assurance of safety
and
[[Page 2649]]
effectiveness. Absent the special controls identified in this proposed
order, general controls applicable to the device are insufficient to
provide reasonable assurance of the safety and effectiveness of the
device.
Therefore, in accordance with sections 513(e) and 515(i) of the
FD&C Act and 21 CFR 860.130, based on new information with respect to
the devices and taking into account the public health benefit of the
use of the device and the nature and known incidence of the risk of the
device, FDA, on its own initiative, is proposing to reclassify this
preamendments class III device into class II. FDA believes that this
new information is sufficient to demonstrate that the proposed special
controls can effectively mitigate the risks to health identified in the
next section, and that these special controls, together with general
controls, will provide a reasonable assurance of safety and
effectiveness for blade-form endosseous dental implant devices.
FDA has also considered blade-form endosseous dental implant
devices in accordance with the reserved criteria set forth in section
513(a) of the FD&C Act and decided that the device does require
premarket notification. The Agency does not intend to exempt this
proposed class II device from premarket notification (510(k))
submission as allowed under section 510(m) of the FD&C Act.
V. Risks to Health
After considering available information, including the
recommendations of the advisory committees (panels) for the
classification of these devices, FDA has evaluated the risks to health
associated with the use of blade-form endosseous dental implant devices
and determined that the following risks to health are associated with
its use:
Local tissue or existing dentition degeneration: Localized
tissue and existing dentition degeneration may be caused by endosseous
implants due to excessive mobility, loss of integration,
incompatibility of device components, or structural failure of the
device.
Pain: Nerve impingement by the device may cause pain.
Bone or nerve damage: Improper design or use of the device
may cause injury during surgery related to sinus perforation, alveolar
plate perforation, or nerve damage resulting in transient or chronic
pain/facial nerve paresis.
Infection: Implantable devices may introduce
microorganisms that may cause local or systemic infections.
Adverse tissue reaction: Inadequate tissue compatibility
of the materials used in this device could cause an immune reaction.
Migration or thermal injury: Incompatibility with magnetic
resonance imaging may cause the device to migrate or heat.
VI. Summary of Reasons for Reclassification
If properly manufactured and used, blade-form endosseous dental
implants can help restore the patient's chewing function by replacing
tooth roots and extending a support post through the gingival tissue
into the oral cavity in order to support a dental restoration, such as
a crown, bridge, or denture. FDA believes that blade-form endosseous
dental implant devices should be reclassified into class II because
special controls, together with general controls, can be established to
provide reasonable assurance of the safety and effectiveness of the
device, and because general controls themselves are insufficient to
provide reasonable assurance of its safety and effectiveness. In
addition, there is now adequate effectiveness information sufficient to
establish special controls to provide such assurance.
VII. Summary of Data Upon Which the Reclassification Is Based
Since the time of the panel recommendation, sufficient evidence has
been developed to support a reclassification of blade-form endosseous
dental implants to class II with special controls. FDA has been
reviewing these devices for many years and their risks are well known.
A review of the applicable clinical literature indicates that the
device has a high success rate (remaining implanted/not removed) and
that few relevant adverse events have been reported in the case of
these devices or related devices suggesting that the device has a high
long-term safety profile. FDA believes that the special controls
identified in this proposed order, if finalized, together with general
controls, can provide a reasonable assurance of the safety and
effectiveness of blade-form endosseous dental implants.
VIII. Proposed Special Controls
FDA believes that the following special controls, together with
general controls, are sufficient to mitigate the risks to health
described in section V of this document:
The design characteristics of the device must ensure that
the geometry and material composition are consistent with the intended
use;
Mechanical performance (fatigue) testing under simulated
physiological conditions to demonstrate maximum load (endurance limit)
when the device is subjected to compressive and shear loads;
Corrosion testing under simulated physiological conditions
to demonstrate corrosion potential of each metal or alloy, couple
potential for an assembled dissimilar metal implant system, and
corrosion rate for an assembled dissimilar metal implant system;
The device must be demonstrated to be biocompatible;
Sterility testing must demonstrate the sterility of the
device;
Performance testing to evaluate the compatibility of the
device in a magnetic resonance (MR) environment;
Labeling must include a clear description of the
technological features, how the device should be used in patients,
detailed surgical protocol and restoration procedures, and relevant
precautions and warnings based on the clinical use of the device;
Patient labeling must contain a description of how the
devices works, how the device is placed, how the patient needs to care
for the implant, possible adverse events and how to report any
complications; and
Documented clinical experience must demonstrate safe and
effective use and capture any adverse events observed during clinical
use.
Blade-form endosseous dental implants are prescription devices
restricted to patient use only upon the authorization of a practitioner
licensed by law to administer or use the device. (Proposed 21 CFR
872.3640(a); see section 520(e) of the FD&C Act and 21 CFR 801.109
(Prescription devices)). Prescription-use restrictions are a type of
general controls authorized under section 520(e) of the FD&C Act and
defined as a general control in section 513(a)(1)(A)(i) of the FD&C
Act; and under 21 CFR 807.81, the device would continue to be subject
to 510(k) notification requirements.
IX. Environmental Impact
The Agency has determined under 21 CFR 25.34(b)) that this action
is of a type that does not individually or cumulatively have a
significant effect on the human environment. Therefore, neither an
environmental assessment nor an environmental impact statement is
required.
X. Paperwork Reduction Act of 1995
This proposed rule refers to previously approved collections of
information found in FDA regulations. These collections of information
are subject to review by the Office of Management and Budget (OMB)
under
[[Page 2650]]
the Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3520). The
collections of information in 21 CFR part 812 have been approved under
OMB control number 0910-0078; the collections of information in part
807, subpart E, have been approved under OMB control number 0910-0120;
the collections of information in 21 CFR part 814, subpart B, have been
approved under OMB control number 0910-0231; and the collections of
information under 21 CFR part 801 have been approved under OMB control
number 0910-0485.
XI. Proposed Effective Date
FDA is proposing that any final order based on this proposal become
effective on the date of its publication in the Federal Register or at
a later date if stated in the final order.
XII. Comments
Interested persons may submit either electronic comments regarding
this document to https://www.regulations.gov or written comments to the
Division of Dockets Management (see ADDRESSES). It is only necessary to
send one set of comments. Identify comments with the docket number
found in brackets in the heading of this document. Received comments
may be seen in the Division of Dockets Management between 9 a.m. and 4
p.m., Monday through Friday, and will be posted to the docket at https://www.regulations.gov.
List of Subjects in 21 CFR Part 872
Medical devices.
Therefore, under the Federal Food, Drug, and Cosmetic Act and under
authority delegated to the Commissioner of Food and Drugs, it is
proposed that 21 CFR part 872 be amended as follows:
PART 872--DENTAL DEVICES
0
1. The authority citation for 21 CFR part 872 continues to read as
follows:
Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 371.
0
2. Section 872.3640 is amended by revising paragraphs (a) and (b)(2) to
read as follows:
Sec. 872.3640 Endosseous dental implant.
(a) Identification. An endosseous dental implant is a prescription
device made of a material such as titanium or titanium alloy that is
intended to be surgically placed in the bone of the upper or lower jaw
arches to provide support for prosthetic devices, such as artificial
teeth, in order to restore a patient's chewing function.
(b) * * *
(2) Class II (special controls). The device is classified as class
II if it is a blade-form endosseous dental implant. The special
controls for this device are:
(i) The design characteristics of the device must ensure that the
geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated
physiological conditions to demonstrate maximum load (endurance limit)
when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to
demonstrate corrosion potential of each metal or alloy, couple
potential for an assembled dissimilar metal implant system, and
corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the
device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the
technological features, how the device should be used in patients,
detailed surgical protocol and restoration procedures, and relevant
precautions and warnings based on the clinical use of the device;
(viii) Patient labeling must contain a description of how the
devices works, how the device is placed, how the patient needs to care
for the implant, possible adverse events and how to report any
complications; and
(ix) Documented clinical experience must demonstrate safe and
effective use and capture any adverse events observed during clinical
use.
Dated: January 4, 2013.
Leslie Kux,
Assistant Commissioner for Policy.
[FR Doc. 2013-00388 Filed 1-11-13; 8:45 am]
BILLING CODE 4160-01-P